Kane AUTOplus 4-2, AUTOplus 5-2 User Manual

Page 1
AUTOplus
Automotive Exhaust
Gas Analyzer
December 2013
© Kane International Ltd
P/N: AP-UM/E
User Manual
Page 2
AUTOplus manual
Page 2
CONTENTS
Page No:
Overview
1. ANALYSER LAYOUT AND FEATURES 4-6
1.1 Features and Keypad 4
1.2 Layout (Rear) 5
1.3 Standard Probe Configuration 6
2. SAFETY WARNING 7
3. FIRST TIME USE 8
4. NORMAL START UP SEQUENCE 8
4.1 Every Time You Use The Analyzer 8
4.2 Automatic Zero Calibration 9
4.3 The Keypad 12
4.4 Menu 14
4.5 Sampling the Exhaust Gas 29
4.6 Turning the Pump ON/OFF 29
4.7 Regular Checks During Sampling 30
4.8 Normal Shutdown Sequence 30
5. PRINTING INFORMATION–OPTIONAL EXTRA ONLY 32-33
5.1 Printing a ‘Live’ Test 32
5.2 Standard Printout 32
6. MAINTENANCE 34-35
6.1 Emptying and Cleaning the In-line Water Trap 34
6.2 Changing the Particle Filter 34
6.3 Cleaning 35
6.4 User Serviceable Parts 35
Page 3
AUTOplus manual
Page 3
CONTENTS
Page No:
7. PROBLEM SOLVING 36
8. ZERO CHECKS AND RE-CALIBRATION 37
8.1 Zero Setting 37
9. PRODUCT SPECIFICATION FOR HANDHELD AUTOplus 4-2 and AUTOplus 5-2 39
APPENDICES 40-45
A. MAIN DISPLAY PARAMETERS 40 B. LAMBDA CALCULATION 42
B.1 Oxygen Balance Formula 42
C. ELECTROMAGNETIC COMPATABILITY (CE)
STATEMENT 43
D. END OF LIFE DISPOSAL 43 E. PRODUCT REGISTRATION 45
Page 4
AUTOplus manual
Page 4
1. ANALYSER LAYOUT AND FEATURES
1.1 Features and Keypad
ON/OFF
UP
Scrolls up through options
MENU
Allows access to all menu functions
DOWN
Scrolls down through options
PUMP Turns pump on and off
STORE
Enters data storage menu
ENTER
Accepts a command, i.e. enters a menu option
PRINT
Prints current data
Page 5
AUTOplus manual
Page 5
1.2 Layout (Rear)
Page 6
AUTOplus manual
Page 6
1.3 Standard Probe Configuration
Note: A variety of probes are available, including high temperature probes
necessary for on-road / rolling road vehicle testing when the temperature of the emission gases increase. Please request further details from your distributor or contact Kane International Ltd directly.
3
5
0
m
m
%
I
n
s
e
r
t
i
o
n
%
L
e
n
g
t
h
Sta inle ss% Ste e l%Sh a ft
Pro b e % In le t
Fle xib le %Inse rtion % Pro b e % (Sta nd a rd % 6m m%Petrol%Hose)
Fle xib le % Ho se %Line
Exha ust%Se c uring % Clamp
Wa te r% Tra p Connection
Caution:%Probe%may%be% hot%when%rem o ving %from exhaust.
Probe Seal
Page 7
AUTOplus manual
Page 7
2. SAFETY WARNING
This analyzer extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted from the top of the analyzer. This analyzer must only be used in well-ventilated locations. It must only be used by trained and competent persons after due consideration of all the potential hazards.
Protection Against Electric Shock (in accordance with EN 61010-1 : 2010)
This analyzer is designated as Class III equipment and should only be connected to SELV circuits.
The battery charger is designated as:
Class II equipment Installation category II Pollution degree 2 Indoor use only Altitude to 2000m Ambient temperature 0°C-40°C Maximum relative humidity 80% Mains supply fluctuations not to exceed 10% of the nominal voltage.
Page 8
AUTOplus manual
Page 8
3. FIRST TIME USE
Charge the battery for 12 hours; following this an overnight charge should be sufficient for an average 8-hour day (turning pump off to save power between tests).
The analyzer has a rechargeable NiMh battery, use only the mains charger or 12 volt adapter supplied or damage may occur to the analyzer and battery.
Check that you have all the items you have ordered.
Take time to read this manual fully.
When using the analyzer for the first time you will need to choose from: -
Language selection Time and date Printed header name and telephone number
4. NORMAL START UP SEQUENCE
4.1 Every Time You Use The Analyzer
BEFORE SWITCH-ON CHECK THAT:
The particle filter is dry and not dirty or damaged
The water trap and probe line are empty of water
All hose connections, etc, are properly made
The probe is sampling CLEAN AMBIENT air
The water trap is correctly fitted and the analyzer upright
* DO NOT RUN ANALYSER WITHOUT WATER TRAP FITTED *
Switch ON the analyzer by pressing
Page 9
AUTOplus manual
Page 9
4.2 Automatic Zero Calibration
During this sequence the analyzer pumps fresh air into the sensors to allow them to zero and the oxygen sensor to be set to 20.9 %.
After switch-on the analyzer will briefly display information:
AUTOplus
SW V1.01
Bluetooth
Auto 4 or Auto 5 indicates 4 or 5 gas version SW is the software version Bluetooth indicates Bluetooth fitted and enabled
(If the charger is connected and is pressed the display will show
Page 10
AUTOplus manual
Page 10
Press again and the display header screen will be shown.
NOTE: When the “CHARGING BATTERY” screen is displayed fast charging mode is operating.)
And then the initializing screen is shown:
FRESH
AIR
PURGE
TIME 90
15:13:27
26/04/12
The time will count down in seconds to zero and is pre-set by the analyzer. During countdown the analyzer will initialize the measuring system, do not block the end of the probe or insert into or near the vehicle exhaust. It also displays some setup values including atmospheric pressure and internal ambient temperature.
TIP: Take care to purge the analyzer in really fresh air. Often the air in a
garage or workshop is contaminated with HC up to about 1 meter from the floor with the emissions from past spills of oil, lubricants etc that may have soaked into the floor.
Once the initialization is completed the analyzer will zero the CO, HC, CO2 and NO sensors if fitted and set oxygen to 20.9%. Keep the analyzer and exhaust probe sampling fresh air.
Page 11
AUTOplus manual
Page 11
The next screen is the MAIN DISPLAY of the analyzer: -
ZERO CAL TIME 15
0000
%
CO2
0000
%
CO
20.9
%
O2
0
ppm
HC
0
ppm
NO
----
Λ 15:13:27
26/04/12
The first time to zero is 15 minutes. After that the time to zero is 30 minutes. The analyzer will beep 3 times one minute before a fresh air zeroing is required and then after a further minute beep continuously.
The bottom line of the display shows time, date and battery level.
When a zero is required the top left hand corner of the display will show “ZERO
REQUIRED”. Press the key to initiate zeroing. When zeroing always ensure that the analyzer is sampling fresh air
The analyzer is now ready for use.
NOTE: The backlight comes on automatically whenever a button is pressed
and then switches off after about 30 seconds.
TIP: It has been found that in most workshops there is a background level
of HC due to oil permeation of the floor. For true fresh air you need to be sampling outside the workshop or at least 2 meters above the floor.
Notes:
1) Lambda can be replaced by AFR as described below.
2) If AUTOLOG is active then the log time will be displayed at the top right hand side of the screen
3) If the charger is connected then the display will show the Battery icon filling up.
Page 12
AUTOplus manual
Page 12
4.3 The Keypad
The keypad has 8 keys:
ON/OFF UP DOWN ENTER PUMP toggle
PRINT .... and SEND
STORE MENU
ON/OFF:
Turns the analyzer ON or OFF. If you press OFF by mistake simply press ENTER to cancel.
/ /
UP / DOWN / ENTER:
Mainly used after selecting MENU to change options and make selections.
PUMP:
Turns the pump OFF or ON. Turn the pump off when not sampling to preserve battery life.
PRINT:
Also acts as SEND key for Bluetooth and serial communications. When active PRINTING appears in the top left hand corner of the display
Page 13
AUTOplus manual
Page 13
PRINTING
0000
%
CO2
0000
%
CO
20.9
%
O2
0
ppm
HC
0
ppm
NO
----
Λ
15:13:27
26/04/12
TIP: When an IR printer is selected you can get a duplicate print automatically
by pressing and holding until a double beep is heard.
STORE:
Press to log readings from the screen into memory. The log number of the readings is displayed for a few seconds in the top left hand corner of the screen.
LOG 005
0000
%
CO2
0000
%
CO
20.9
%
O2
0
ppm
HC
0
ppm
NO
----
Λ
15:13:27
26/04/12
You can LOG up to 500 sets of readings. Stored reading can be accessed from the MENU function
Page 14
AUTOplus manual
Page 14
Press and hold for 2 seconds to toggle AUTOLOG on and off.
Note: AUTOLOG is disabled when the pump is switched off.
4.4 Menu
There are 5 main menu categories each with their own sub-menu.
SETUP
UNITS
SCREEN
REPORTS
SERVICE
MANUAL ZERO
LANGUAGE
FUEL
CONTRAST
VIEW
CODE
PUMP
VEHICLE
B’LIGHT
DELETE
CAL DUE
PRINTER
EFF
LINES
AUTOLOG TIME
SET TIME
PEF
BACK
AUTOLOG START
SET DATE
MEF
HEADER 1
PASSKEY
BACK
HEADER 2
BACK
BACK
Note: In SERVICE: CODE is reserved for official service agents only.
Pressing the MENU key causes the following screen to be displayed
MENU
SETUP
UNITS
SCREEN
REPORTS
SERVICE
MANUAL ZERO
12.52.30 am
26/04/12
Page 15
AUTOplus manual
Page 15
Use / to scroll the cursor through the selection and the press to choose. To exit this screen press MENU again.
SETUP
MENU
LANGUAGE
PUMP
PRINTER
SET TIME
SET DATE
PASS KEY
BACK
12.52.30 am
26/04/12
Language:
Use / to change. Use to accept.
MENU
LANGUAGE
= English
= French
= German
= Dutch
12.52.30 am
26/04/12
Page 16
AUTOplus manual
Page 16
PUMP: Pump speed can be set from 1 (slowest ) to 4 (fastest).
Use / to change. Use to accept
MENU
PUMP
= 4
12.52.30 am
26/04/12
PRINTER: Select from the following communication methods
Use / to change. Use to accept
MENU
PRINTER
= KANE IRP-2
= KANE IRP
= USB-SERIAL
= BLUETOOTH
12.52.30 am
26/04/12
Page 17
AUTOplus manual
Page 17
SET TIME:
Use / to change number. Use to accept and move the cursor.
MENU
SET TIME
HH:
MM:
SS
12:
52:
30 am
12.52.30 am
26/04/12
NOTE: If ‘FUNCTION LOCKED’ appears, delete all reports (see p23-25).
SET DATE:
Use / to change number. Use to accept and move the cursor.
Date format can also be changed.
MENU
SET DATE
DD:
MM:
YY
26:
04:
12
12.52.30 am
26/04/12
NOTE: If ‘FUNCTION LOCKED’ appears, delete all reports (see p23-25).
Page 18
AUTOplus manual
Page 18
PASSKEY: This needs to be set to suit particular Bluetooth devices.
Use / to change number. Use to accept and move the cursor.
MENU
PASS KEY
= 2111
12.52.30 am
26/04/12
UNITS:
Use / to change. Use to accept.
MENU
FUEL
VEHICLE
EFF
PEF
MEF
BACK
12.52.30 am
26/04/12
Page 19
AUTOplus manual
Page 19
FUEL: note when diesel is selected HC and Lambda are not displayed.
MENU
FUEL
= gasoline/petrol
= LPG
= CNG
= DIESEL
12.52.30 am
26/04/12
VEHICLE: allows a vehicle registration number to be entered.
Use / to change. Use to accept.
MENU
VEHICLE:
VE05DPZ
12.52.30 am
26/04/12
Page 20
AUTOplus manual
Page 20
EFF: Set Lambda or AFR
Use / to change. Use to accept.
MENU
EFF:
= LAMBDA
= AFR
12.52.30 am
26/04/12
PEF: (Propane equivalency Factor) displays value only.
MENU
PEF:
0.545
12.52.30 am
26/04/12
Page 21
AUTOplus manual
Page 21
MEF: (METHANE equivalency Factor) displays value only.
MENU
MEF:
10.000
12.52.30 am
26/04/12
SCREEN
MENU
CONTRAST
BACKLIGHT
LINES
BACK
12.52.30 am
26/04/12
Use / to change. Use to accept.
Page 22
AUTOplus manual
Page 22
CONTRAST: set contrast from 1 to 15
MENU
CONTRAST
12
12.52.30 am
26/04/12
BACKLIGHT: adjust switch off time from 30 to 300 secs
MENU
B’LIGHT
30S
12.52.30 am
26/04/12
Page 23
AUTOplus manual
Page 23
LINES: Allows the main display to be configured by the user to his preferences
MENU
LINE 1
= CO2%
----%
LINE 2
LINE 3
LINE 4
LINE 5
LINE 6
EXIT
12.52.30 am
26/04/12
NOTE: If ‘FUNCTION LOCKED’ appears, delete all reports (see p23-25).
Use / to change. Use to accept.
Choose from: CO2
NO O2 LAMBDA/AFR COK NOX HC CO
REPORTS
MENU
VIEW REPORTS
DELETE ALL REPORTS
AUTO LOG TIME
START AUTO LOG
HEADER 1
HEADER 2
BACK
12.52.30 am
26/04/12
Page 24
AUTOplus manual
Page 24
VIEW REPORTS
MENU
LOG 13
Date: 07/10/11
Time: 03:35
VEHICLE: VE05DPZ
FUEL: DIESEL
CO2
14.4%
CO
0.05%
O2
1.1%
HC
----
NO
150 ppm
Λ
----
12.52.30 am
26/04/12
Reports can be printed by pressing .
LOG number to be viewed can be changed by pressing / .
Page 25
AUTOplus manual
Page 25
DELETE ALL REPORTS:
MENU
DELETE ALL?
NO
YES
12.52.30 am
26/04/12
MENU
CONFIRM DELETE?
NO
YES
12.52.30 am
26/04/12
Page 26
AUTOplus manual
Page 26
AUTOLOG TIME: Can be set from 5 secs to 300 secs,
MENU
AUTOLOG TIME
= 10
12.52.30 am
26/04/12
Note: If the PUMP is switched off then AUTOLOG is inhibited
START AUTOLOG
TIP: Before starting AUTOLOG you might want to manually re-zero the
analyzer so that you have 30 minutes of logging time before a re-zero is requested.
Switches ON or OFF automatic timed logging
MENU
START AUTOLOG?
YES
12.52.30 am
26/04/12
Page 27
AUTOplus manual
Page 27
To initiate AUTOLOG go back to the main display where the Autolog time will be
displayed in the top right hand corner of the display, and then press . The analyzer will beep and the autolog time will disappear for a second or so.
AUTOLOG can be toggled on and off by pressing and holding until the autostore time in the top right hand side of the screen appears or disappears.
HEADER 1 HEADER 2 allows inputting of company details to appear on two lines of the
printout
MENU
HEADER 1:
Company Name
12.52.30 am
26/04/12
MENU
HEADER 2:
Phone number?
12.52.30 am
26/04/12
Page 28
AUTOplus manual
Page 28
Use / to change. Use to accept.
SERVICE
MENU
CODE
= 0000
CAL DATE
= 369
12.52.30 am
26/04/12
CODE entry is reserved for approved service agents.
CAL DATE is the countdown for days remaining until annual calibration is due.
MANUAL ZERO
This allows the user to manually re-zero the analyzer
REMOVE PROBE FROM EXHAUST
PRESS ENTER TO ZERO
12.52.30 am
26/04/12
Page 29
AUTOplus manual
Page 29
4.5 Sampling the Exhaust Gas
Once the zero calibration and test procedures have been completed and the fuel has been selected the probe can be inserted into the desired vehicle exhaust.
Ensure the probe is inserted into the exhaust pipe so as to not allow air into the probe. The exhaust of a car can pulse, especially at low RPM, drawing air in causing bad readings, ensure the flexible probe is fully inserted and the clip attached to the exhaust pipe.
4.6 Turning the pump ON/OFF
The analyzer is fitted with a pump to draw gas from the vehicle exhaust. To conserve battery power, switch off the pump when you are not taking a measurement. Gas values will not be displayed when the pump is turned off .
Use the key to turn ON and OFF the pump.
Ve hic le Exha ust
Ensu re /C la m p /is/Fixe d to / Ve h ic le / Ex ha u st
Page 30
AUTOplus manual
Page 30
The analyzer will block readings while the pump is off and display ‘----‘ on all gas channels.
It is recommended that the analyzer sample fresh air for 60 seconds before the pump is turned off.
4.7 Regular Checks During Sampling
Care must be taken at all times not to exceed the analyzers operating
specifications, in particular ensure the following:-
DO NOT PLACE THE ANALYSER IN THE ENGINE BAY.
Do not expose the analyzer to temperatures outside its normal operating
range.
DO NOT PLACE THE ANALYSER ON A HOT SURFACE.
Ensure that liquid in the water trap does not go over the level indicator.
Note! The indicator only works while the trap is vertical. Water condenses in the probe line and can quickly fill the water trap when the probe is moved. Take care, watch the water trap closely and empty any water when it is noticed.
Make sure that the in-line particle filter is clean and does not become blocked. If this filter is allowed to become dirty then damage may occur inside the analyzer.
Check the following for water:
Probe line, water trap and particle filter
When the blockage is cleared the analyzer should resume normal operation. If it not possible to clear the problem then internal damage may have occurred and the unit should be returned to a service center.
TIP: After testing hang the sample hose so that both ends are free to drain
any condensate.
4.8 Normal Shutdown Sequence
DO THIS EVERY TIME YOU USE THE ANALYSER
Page 31
AUTOplus manual
Page 31
Remove the probe from the vehicle exhaust - TAKE CARE ! THE PROBE WILL BE HOT - and allow it to cool naturally. Do not immerse the probe in
water as this will be drawn into the analyzer and damage the pump and sensors.
Once the probe is removed from the exhaust allow the readings to return to zero
and press the analyzer will count down from 30 to switch off.
OFF 30
MENU TO ESCAPE
12.52.30 am
26/04/12
If you have not finished but press by mistake, you can press to return to normal operation and not switch OFF.
TIP: It’s a good idea to empty the water trap after testing so that the risk of
water getting on to the particle filter or into the analyzer is minimized.
Page 32
AUTOplus manual
Page 32
5. PRINTING INFORMATION – OPTIONAL EXTRA ONLY
Supplied as an accessory for the analyzer is an infrared thermal printer. Read the manual supplied with the printer prior to operation. Connections to the analyzer are detailed below:
Infrared thermal printer - this does not require a cable to transmit the data
but uses an infrared (IR) link similar to a TV remote control. The IR emitter is positioned on the top of the Analyzer and the bottom of the printer. Ensure they are pointing at each other and within 300 mm, with no obstructions in the way. Data may be lost if transmission is interrupted, a black square is evidence of this. Keep the Analyzer pointing at the printer until the printout has finished.
Data can either be printed from a ‘live’ test or from stored data.
5.1 Printing a ‘Live’ Test
During a vehicle test the Analyzer will print data on request. With the analyzer
showing the MAIN DISPLAY press and hold until the second bleep. Current data will be sent to the printer. If the print button is held until the third bleep, the test results will be printed in duplicate.
5.2 Standard Printout
The standard printout is shown below:
AUTOplus V1.00
SERIAL No. 030912287
Kane Auto Plus
(44) - 1707 - 375550
DATE:
09/07/12
TIME:
08:25:33AM
VEHICLE
AB11 DEF
FUEL
GAS/PETROL
CO2
%
0.0
CO
%
0.1
O2 % 20.95
HC
ppm
0
COK % ----
Lambda ----
Page 33
AUTOplus manual
Page 33
NOTE: The analyzer will stop logging once it has reached 500 readings and
will return to the main display if is pressed. Data can still be viewed and printed.
TIP: Make a note of the location number for your particular test as it may
be useful when printing.
TIP: Stored and displayed with the data are actual time and date of the
test.
Page 34
AUTOplus manual
Page 34
6. MAINTENANCE
6.1 Emptying and Cleaning the In-line Water Trap
The in-line water trap should be checked and emptied on a regular basis. Water vapor will condense and gather in the probe line. This may move suddenly to the trap when the probe is moved. Care should be taken at all times.
Emptying of the water trap is detailed below :-
Water level indicator (do not exceed level while trap is vertical)
Carefully remove the end cap from the in-line housing. Dispose of the condensate in a suitable drain. Clean the inside of the water trap using a soft cloth
6.2 Changing the Particle Filter
This is a very important part of the analyzer and should be changed regularly. It prevents dust and dirty particles entering the pump and sensors and hence causing damage. The filter MUST be changed when it is discolored.
Remove the end cap from the in-line filter housing. Carefully remove the paper filter element and dispose of it. Clean the inside of the filter housing with a suitable soft cloth. Insert a new filter element onto the spigot in the filter housing and carefully replace the end cap.
IF THE FILTER IS NOT CHANGED REGULARLY DAMAGE WILL OCCUR TO THE SENSORS RESULTING IN A CHARGEABLE SERVICE. IT IS EVIDENT
Page 35
AUTOplus manual
Page 35
FROM AN INTERNAL INSPECTION OF THE ANALYSER IF THE FILTER HAS BEEN CHANGED REGULARLY.
6.3 Cleaning
The body of the analyzer, its screen and rubber boot can be cleaned using a damp clothe and standard household detergents. Wipe clean with a damped cloth rinsed in water. Leave the analyzer in a warm place until all surface moisture has evaporated and the analyzer is completely dry.
Under no circumstances should you use solvent-based cleaners as they can cause blooming on the display and degradation of the plastic parts.
6.4 User Serviceable Parts
This product contains NO user serviceable parts. If problems occur seek advice from your nearest approved service center.
Page 36
AUTOplus manual
Page 36
7. PROBLEM SOLVING
The following is a list of problems that may occur on the analyzer through its operating life. If the cause of the fault is not easy to identify then we advise you contact Kane International Technical Support or an International Distributor for expert advice.
Fault symptom
Causes
Oxygen too high
Air leaking into probe, tubing, water
trap, connectors or internal to analyzer.
Oxygen cell needs replacing.
Oxygen Error (FAULT)
Infrared gas Error (FAULT)
Zero calibration fault
Analyzer has been stored in a cold
environment and is not at normal working temperature.
Oxygen cell or infrared bench needs replacing.
Analyzer not holding charge
Analyzer not charging
Battery exhausted.
Charger not giving correct output.
Charger could be faulty.
Analyzer does not respond
to exhaust gas
Particle filter blocked.
Probe or tubing blocked.
Pump not working or damaged with
contaminants.
Analyzer automatically switches off in operation
Battery below alarm level.
Ambient temperature above 50°C.
Battery quickly discharging and is
faulty.
Page 37
AUTOplus manual
Page 37
INTERNAL FILTER
To protect the analyzer from water ingress a filter is installed inside the casing to protect the infrared measuring system. This filter will block if care is not taken during sampling:
Ensure any build-up of water in the probe line and water trap are removed as soon as possible.
The external particle filter is changed regularly.
The analyzer is allowed to sample fresh air on a regular basis.
The analyzer samples fresh air for 3 minutes before switch off.
Do not blow smoke from a cigarette into the analyzer.
If you suspect the internal filter is blocked perform the following:
Remove the probe connection from the water trap.
Empty and clean the water trap with a dry cloth.
Fit a new external particle filter.
Run the analyzer in fresh air (pump ON) for at least one hour.
If the problem does not clear contact a service agent
8. ZERO CHECKS AND RE-CALIBRATION
During normal operation of the analyzer the following checks may be requested as required:
Zero setting of all sensors (can also be selected manually)
8.1 Zero setting
The zero setting function sets the working sensors to zero using fresh air. This function is activated as follows:
Following analyzer switch ON.
On a timed basis. Following the analyzer first being turned ON a zero will
requested automatically at 15 and then 30 minute intervals. Subsequent requests will be every 30 minutes.
Page 38
AUTOplus manual
Page 38
On request by the user from the MENU
The zero sequence is as follows, ensure the on screen commands are followed or the analyzer may not zero correctly.
MENU
ZERO CHECK
REMOVE PROBE
FROM EXHAUST
THEN PRESS ENTER
12.52.30 am
26/04/12
Ensure the probe is removed from the vehicle exhaust and is sampling fresh air, in the garage environment this should be ½ meter or 18 inches above the
ground. Once this is done press to activate the pump.
WARNING: The sensors will only be reset if the probe is sampling fresh air
for at least 60 seconds.
Once the zero is complete the screen will return to the MAIN DISPLAY.
Page 39
AUTOplus manual
Page 39
9. PRODUCT SPECIFICATION FOR HANDHELD
AUTOplus 4-2 & AUTOplus 5-2
Parameter
Resolution
Accuracy
Range
Carbon Monoxide (Infrared)
0.01 %
+/- 5 % of reading *1 +/- 0.06 % volume *1
0-10 % Over-range 20 %
Oxygen (fuel cell)
0.01 %
+/- 5 % of reading
*1
+/- 0.1 % volume *1
0-21 % Over-range 25 %
Hydrocarbon (Infrared)
1 ppm
+/- 5 % of reading
*1
+/- 12 ppm volume *1
0-5000 ppm Over-range: 10,000 ppm
Carbon Dioxide (Infrared))
0.1 %
+/- 5 % of reading
*1
+/- 0.5 % volume *1
0-16 % Over-range: 25%
Nitric Oxide*² (fuel cell)
1 ppm
0-1500ppm +/-5% or 25ppm;
0-1500ppm Over-range: 5000 ppm
Carbon Monoxide Corrected COK
0.01%
Calculated
0-15%
Lambda AFR (Petrol) (LPG)
0.001
00.01
0.8 – 1.2
11.76 – 17.64
12.48 – 18.72
Sensor response T95
Nominal 20 seconds AUTOplus 4-2, 5-2.
Warm up
Less than 2 minutes
Pre-programmed Fuels
Petrol/Gasoline, LPG Diesel and CNG.
PC connection
Via RS 232 port
Data-Logging
500 Tests
Dimensions
Weight Handset Probe
1kg 220mm x 55mm x 120mm Insertion depth 350mm x Diameter 15mm Clip handle to secure to exhaust, 4m long hose Various probes available including high temperature
Ambient Operating Range
+5°C to +45°C/10% to 90% RH non condensing
Storage temperature
Minimum: 0°C Maximum: +50°C
Battery Charger
Input: 100-240 V ~ 47-63 Hz Output: 12 V DC
Analyzer battery run time
>4 hours from full charge with the pump running
*1
Using dry gases at STP
*² Standard on models AUTOplus 5-2
To obtain the quoted specification an analyzer should be calibrated with clean ambient air (normally outside the workshop) at standard temperature and pressure (STP).
Note: The analyzer is not for use with a Dynamometer, unless the high temperature probe is fitted.
Page 40
AUTOplus manual
Page 40
APPENDICES
A. MAIN DISPLAY PARAMETERS
The parameters and their meanings are detailed as follows: -
FUEL: The selected fuel will be displayed, i.e. PETROL.
PETROL – Leaded or Unleaded petrol/gasoline.
LPG – Liquid Petroleum Gas
CNG – Compressed Natural Gas
DIESEL
DATE : Analyzer date. See Set-Up menu to change.
TIME : Analyzer time. Use Set-Up menu to change.
BATTERY : Displays the battery level from 0-100%. During normal data
display operation the analyzer will show LOW BATTERY briefly at intervals of about 45 seconds for less than 30% of charge. The analyzer will turn off if less than 10% charge is detected. With the charger connected the display will read close to 100%. Note: Allowing the battery to discharge fully may destroy it.
O2 : Oxygen measured in the exhaust gas indicated in percentage %.
With the pump off the analyzer will display - - - -. If there is a fault with the oxygen sensor then FLT will be displayed.
CO : Carbon monoxide measured in the exhaust gas indicated in
percentage %. With the pump off the analyzer will display - - - -. If there is a fault with the CO reading then FLT will be displayed.
CO2 : Carbon dioxide measured in the exhaust gas in percentage %.
With the pump off the analyzer will display - - - -. If there is a fault with the CO2 reading then FLT will be displayed.
HC : Hydrocarbons measured in the exhaust gas indicated in ppm
(parts per million) n-hexane (petrol). With the pump off the analyzer will display - - - -. If there is a fault with the HC reading then FLT will be displayed.
Page 41
AUTOplus manual
Page 41
COK : Generally known as corrected CO. This value is calculated and
used for comparison with the actual infrared measured CO value.
COK = (COx15)/(CO + CO2), for normal car exhaust CO + CO2 =
about 15%. In this case CO is approximately equal to COK. If COK is clearly higher than CO this indicates defects such as exhaust leaks. At near zero COK is not valid.
LAMBDA (λ): The value of Lambda gives an indication of the burning
fficiency of the engine. This can be replaced with the Air Fuel Ratio (AFR) below. See section 5.2.3.to change between displays. Appendix B gives the formula used.
When sampling fresh air and lambda is outside operation range
this indication will show ‘-----‘.
AFR : Air Fuel Ratio is another method for displaying the efficiency of an
engine. The calculation for the AFR is Lambda multiplied by 14.7 for Petrol and 15.6 for LPG (typically). When sampling fresh air this indication will show ‘-----‘.
NO : Nitric oxide reading in ppm (parts per millions) of the exhaust gas.
Displayed when Nitric oxide sensor fitted, indicated on the rear label. Displays NOT FITTED or N/F when sensor not fitted and FAULT or FLT if failed.
NOx : A calculated value based on the measured level of Nitric Oxide to
display total oxides of Nitrogen. NOx = NO X 1.05
TIME TO: The analyzer requires to regularly zero the sensors. Once a ZERO has been performed the time to the next zero is displayed in minutes. Check there is sufficient time remaining before starting a test and perform a manual zero if not.
Page 42
AUTOplus manual
Page 42
B. LAMBDA CALCULATION
The value for Lambda is a determinant for the burning efficiency of an engine. The value depends on the composition of the fuel, the air that is used for the combustion and on the combustion products as found in the exhaust gases.
A basic formula, taking into account:
Components of the fuel: carbon, hydrogen, oxygen and water content;
Water content of the air;
Components of the exhaust gases: carbon dioxide, carbon monoxide,
hydrocarbons and nitrogen oxide;
has been developed by J. Brettschneider and published in Bosh Technishe Berichte, Volume 6 (1979), No. 4, page 177-186.
A simplified formula, derived from the basic formula, and based upon the assumption that the water content of the fuel and air and the NOx content in the exhaust gases are negligible, allows the computation of Lambda when certain components of the exhaust are measured.
B.1 Oxygen balance formula
For Lambda calculation, based upon measurements of CO, CO2, HC and O2, the following formula is standardized: Displayed on the analyzer as LAMBDA (O)
λ = CO2 + (CO/2) + O2 + [HCV/4 × {3.5 / (3.5 + CO/CO2)} – OCV/2] × (CO2 + CO)
(1 + HCV/4 - OCV/2) × {(CO2 + CO) + (K1 × HC)}
Where:
CO = Carbon monoxide % volume measured. CO2 = Carbon dioxide % volume measured. HC = Hydrocarbon ppm volume measured. O2 = Oxygen % volume measured. K1 = Conversion factor for HC is expressed in ppm vol n-hexane (C6H14)
equivalent. Its value in this formula is 6.10-4 Hcv = Atomic ratio hydrogen to carbon in the fuel. Nominal value is 1.7261 Ocv = Atomic ratio oxygen to carbon in the fuel. Nominal value is 0.0176
Page 43
AUTOplus manual
Page 43
C. ELECTROMAGNETIC COMPATABILITY (CE)
STATEMENT
This product has been tested for compliance with the following generic standards:
EN 61000-6-3:2011 EN 61000-6-1:2007
and is certified to be compliant
Specification EC/EMC/KI/Autoplus details the specific test configuration, performance and conditions of use.
Please Note: Batteries used in this analyzer should be disposed of in
accordance with current legislation and local guidelines.
E: END OF LIFE DISPOSAL
At the end of its life this analyzer should be sent to the appropriate recycling centre in accordance with current legislation and local guidelines.
Page 44
AUTOplus manual
Page 44
Page 45
AUTOplus manual
Page 45
Product Registration
Please complete, detach and return to:
ANSED Diagnostic Solutions LLC. 420 Lexington Avenue Suite 2631 New York, NY 10170 USA
Your Details
Name:
Job Title:
Company Name:
Company Address 1:
Address 2:
Town/City:
State:
ZipCode:
Country:
Phone Number:
Fax Number:
Mobile Number:
Email Address:
Product Details
Note: Proof of Purchase may be required for warranty claims.
Date Purchased (DD/MM/YYYY):
Purchased From:
Model Number:
Product Serial Number:
Page 46
AUTOplus manual
Page 46
Thank you for buying this analyzer.
Before use, please register on our website:
www.ANSEDdiagnostics.com
Or complete, detach and return the Product Registration page.
Loading...