Kane 9206 Quintox User Manual

Page 1
1
KANE9206
Quintox Flue Gas Analyser
© Kane International Ltd
Page 2
2
CONTENTS
Page No:
SAFETY WARNING 7
KANE9206 OVERVIEW 7
STANDARD FEATURES 7
HANDSET 7 MAIN ANALYSER UNIT 8
OPTIONS 9
ELECTROCHEMICAL SENSORS 9 OTHER OPTIONS 9 PROBE OPTIONS 9
SPARE PARTS LIST 9
KANE9206 WITH KANE LIVE PC SOFTWARE 10
ANALYSER LAYOUT AND FEATURES 11
HANDSET FEATURES 11 ANALYSER LAYOUT 12 ANALYSER LAYOUT WITH KMDM110/230 SAMPLE CONDITIONING UNIT FITTED 13 TYPICAL PROBE CONFIGURATION (KMCP6) 14 ANALYSER CONNECTIONS 15
GETTING STARTED 16
BEFORE TAKING READINGS 16
MENU: ALL THE OPTIONS 17
STATUS: ESTABLISHING THE ANALYSER’S SET-UP 18
Page 3
3
SETUP: CONFIGURING THE ANALYSER’S SET-UP 20
LANGUAGE 20 MAIN PURGE 21 PRINTER 21 AUTO SET TIME 22 SET TIME 22 SET DATE 23 HEATER STATUS 23 SELECT ANALYSER UNIT SERIAL NUMBER 23 WIRELESS SET UP 24 WIRELESS PASSKEY 24
ANALYSER UNITS: CONFIGURES ALL THE DATA SOURCES AND SETTINGS 25
FUEL ORIGIN 25 FUEL TYPE 25 EFFICIENCY 25 GAS UNITS 26 TEMPERATURE 26 PRESSURE 26 SET PERCENTAGE REFERENCE 02 26 SET NOx CALCULATION 26 SET COMPENSATION 26 CONVERSION FACTORS 26
CO ALARM 27
CO ALARM SET 27 CO ALARM LEVEL 27
SCREEN 28
CONTRAST 28 B’LIGHT 28 MODE 28 LINES 29
Page 4
4
REPORTS: CONFIGURES REPORTS 30
VIEW REPORTS 30 DELETE ALL REPORTS 31 AUTO LOG TIME 31 START AUTO LOG 31 START AUTO PRINTING 31 HEADER 1 31 HEADER 2 31 FORM FEED 31
SERVICE 32
MANUAL AIR ZERO 32
MANUAL PRESSURE ZERO 32
BEFORE USING THE ANALYSER FOR THE FIRST TIME 33
SAFETY WARNING 33
FIRST TIME USE 33
NORMAL START UP SEQUENCE 33
EVERY TIME YOU USE THE ANALYSER 33 AUTOMATIC CALIBRATION 34
MAIN MEASUREMENT SCREEN 36
IN SMALL FONT MODE 36 IN LARGE FONT MODE 37
SAMPLING THE FLUE GAS 39
LONG TERM MONITORING 39
KMDM110/230 SAMPLE CONDITIONING UNIT 40
SETTING UP 40
Page 5
5
MAINTENANCE 42
EMPTYING AND CLEANING THE IN-LINE WATER TRAP 42 CHANGING THE PARTICLE FILTER 42 CHARGING THE BATTERY 43 CHANGING THE PAPER ROLL 43 TO START PAPER FEED 43 CHANGING THE PRINTER RIBBON 43
PROBLEM SOLVING 45
HOW TO GET EXPERT HELP 45
ANALYSER ANNUAL RECALIBRATION AND SERVICE 46
RETURNING YOUR ANALYSER TO KANE 47
PACKING YOUR ANALYSER 47 SENDING YOUR ANALYSER 47 WHEN WE RECEIVE YOUR ANALYSER 47
SERVICE RETURNS 48
PRODUCT SPECIFICATION 50
UNIT 50 OPTIONAL IR MODULE 51 HANDSET 51 EXTENSION CABLE 52 MAIN BATTERY AND OPTIONAL HEATER BATTERY 52 BATTERY CHARGER 52 PUMP 52 INTEGRAL PRINTER 52 AMBIENT OPERATING RANGE 52 KMHL3000: HEATED SAMPLE LINE 53 KMHP1200: HEATED PROBE 53 KMDM110/230: SAMPLE CONDITIONING UNIT 53 PROBE 53 OPTIONAL PORTABLE PRINTERS 53
Page 6
6
ELECTROMAGNETIC COMPATIBILITY (CE) STATEMENT 54
END OF LIFE DISPOSAL 54 BATTERY DISPOSAL 54
EN50379 REGULATED INSTRUCTIONS 55
APPENDICES 60
A. PARAMETER MEANINGS 60
B. NOx CALCULATIONS 63
ONLY AN NO SENSOR FITTED 63
WORKING IN PPM: NOX REFERENCED TO NO 63 WORKING IN MG/M3: NOX REFERENCED TO NO OR NO2 63 WORKING IN MG/M3: REFERENCED TO NO 63 WORKING IN MG/M3: NOX REFERENCED TO NO2 64 NORMALISING READINGS 64
BOTH NO AND NO2 SENSORS FITTED 64
WORKING IN PPM: NOX = NO + NO2 64 WORKING IN MG/M3 64 NOX = SUM 64 NOX = NO 64 NOX = NO2 64 NORMALISING READINGS 65
ONLY AN NO2 SENSOR FITTED 65
NORMALISING READINGS 65
HIGH CO PURGE OPERATION 65
C. COMBUSTION EFFICIENCY CALCULATION 66
D. CALCULATION OF FUEL DATA 68
PRODUCT REGISTRATION 69
Page 7
7
SAFETY WARNING
This analyser must only be used in well-ventilated locations by trained and competent persons after due consideration of all the potential hazards and with regard to local and National regulations and guidelines.
The analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted from the bottom of the instrument.
Users or portable gas detectors are recommended to conduct a “bump” check before relying on the unit to verify an atmosphere is free from hazard.
A “bump” test is a means of verifying that an instrument is working within acceptable limits by briefly exposing to a known gas mixture formulated to change the output of all the sensors present. (This is different from a calibration where the instrument is also exposed to a known gas mixture but is allowed to settle to a steady figure and the reading adjusted to the stated gas concentration of the test gas).
KANE9206 OVERVIEW
The KANE9206 is broadly based on the KM9106 and whilst retaining many of its core features has been significantly enhanced. The most visible difference is the large graphical display on the handset. Up to 15 lines of text/data can be displayed. The handset links to the main analyser unit using wireless communications or the normal cable. It also has a USB connector to link to a PC via a cable and has an infra-red output to link to the portable Kane KMIRP-2 printer.
The main analyser unit also contains significant enhancements over the KM9106.
STANDARD FEATURES:
19301 Battery charger 18277 UK mains lead 18276 EU mains lead 18275 US mains lead 19332 Instruction manual
Kane 'LIVE' software download from Kane website
HANDSET: KBHS
Wireless and cable connectivity to analyser unit. Wireless and USB connectivity to PC GPS location IR connectivity to Kane IRP portable printers Monster data storage memory (64k records) Graphical display with choice of large or small fonts. Battery rechargeable via main unit or mains charger
Page 8
8
MAIN ANALYSER UNIT: AS STANDARD KANE9206
Measures:
Oxygen Carbon monoxide Ambient temperature Atmospheric pressure Inlet temperature Flue temperature Differential pressure
Features:
Main purge Flow control* NiMh battery packs Plain paper printer
*Flow control
To compensate for different suction levels in flues, hose lengths and filter contamination levels, all of which can affect the flow of sample gases, there is an active flow control system fitted to the KANE9206.
It operates as follows:
Every time the instrument is turned on and finishes its first fresh air purge cycle from inside the analyser, it measures and records the pump pressure just prior to the sensor manifold whilst the pump is at 100% flow rate. During service calibration & normal use, the pump speed is automatically adjusted to 70% of the purge flow rate to maintain consistent flow.
The flow control can cope with typically 100 mbar suction in a flue and still maintain the same nominal flow as that present under ambient conditions.
Page 9
9
OPTIONS:
ELECTROCHEMICAL SENSORS : (UP TO 5 SENSORS)
CHOOSE FROM:
KNO1L/Q Nitric oxide (low range) KNO1H/Q Nitric oxide (high Range) KNO2/Q Nitrogen dioxide KSO2L/Q Sulphur dioxide (low range) KSO2H/Q Sulphur dioxide (high range) KH2S/Q Hydrogen sulphide
OTHER OPTIONS:
KHSA Heater for toxic sensors KHC/Q IR triple bench (CO, HC, CO2) KHPUR/Q High/low CO protection (solenoid and pump) WTS9206P Pumped water trap KMHL3000 Heated line KMHP1200 Heated probe KMDM220 Gas Conditioning Module
PROBE OPTIONS:
KMCHLP6 High temperature 1 metre removable shaft KMCHP6 High temperature 285mm removable shaft KMCHSLP6 Smoke probe with high temperature 1 metre removable shaft KMCHSP6 Smoke probe high temperature 285mm removable shaft KMCSP6 Smoke probe with 285mm removable shaft KMCLP6 1 metre removable shaft KMCP6 285mm removable shaft
SPARE PARTS LIST:
OS11 Oxygen Sensor KCO1/Q Carbon monoxide (H2 compensated) IMP3R Printer ribbon IMP10P Printer paper SF1/5 Chemical filter AF2 Particle filter assembly PF2/10 Particle filter filter pack WN8 Water trap filter pack BP9206 Battery pack WTS9106 Water trap assembly 19403 Peri pump tubing
Page 10
10
KANE9206 WITH KANELIVE PC SOFTWARE
KANELIVE is a free download that runs on Windows based PCs and allows live display and graphing of data. It can be downloaded from the Kane website (www.kane.co.uk) once an analyser has been registered in the MY KANE section of the website.
In the current configuration, the handset needs to be connected to the main unit by a cable and the wireless setting for the handset needs to be changed to TO PC using MENU, SETUP, WIRELESS SETUP.
Once this has been selected, go to the PC and select DEVICES & PRINTERS.
Click ADD A DEVICE.
All devices within range will be displayed in icon form. The KANE9206 handset will be displayed as:
KANE9206HS
999999107
where the 9 digit number is the serial number of the handset.
Double click on this. Now follow the instructions.
Then enter the passkey: 1111
Click on the TICK BOX and then AT SERIAL CONFIGURE and then FINISH.
Now click on KANELIVE to initiate the programme.
Select your analyser type by clicking on the analyser name displayed on the middle of the bottom line of the screen. If more than one analyser has been enabled a drop down will appear. Click on the analyser of your choice.
Then click CONNECT and wait for its colour to change to green.
Then click START which will change to green.
After a few seconds live data will be displayed.
Page 11
11
ANALYSER LAYOUT AND FEATURES
HANDSET FEATURES
INFRA RED LED
REMOTE LEAD SOCKET (8 PIN DIN)
PUMP ON/OFF KEY
PRINT KEY
ENTER KEY
ON/OFF KEY
MENU KEY
STORE KEY
SCROLL UP KEY
SCROLL DOWN KEY
GRAPHICAL DISPLY
PROTECTIVE COVER
BATTERY COMPARTMENT (BACK)
SERIAL USB CONNECTION
Page 12
12
ANALYSER LAYOUT
OXYGEN SENSOR
FILTER BRIDGE OR SULPHUR FILTER
PARTICLE FILTER
BATTERY ENCLOSURE
PRINTER UNIT
HANDSET (STORED IN POCKET)
CONNECTION LEAD
FOR OXYGEN SENSOR
DUAL PRESSURE
PORTS
WATERTRAP
CONNECTION
WATERTRAP LOCATED ON SIDE OF INSTRUMENT
ACCESSORY STORAGE SPACE (LEADS, WATERTRAP ETC)
INSTRUMENT CASE
Page 13
13
ANALYSER LAYOUT WITH KMDM110/230GAS CONDITIONING MODULE FITTED
OXYGEN SENSOR
PARTICLE FILTER
PRINTER UNIT
HEATED LINE CONNECTION
GAS DRYER ASSEMBLY
INSTRUMENT CASE
WATERTRAP CONNECTION
DUAL PRESSURE PORTS
OXYGEN SENSOR
CONNECTOR
FLUE SOCKET
INLET SOCKET
AUX CONNECTOR (25 PIN 'D')
ANALYSER 'ON' SWITCH
REMOTE HANDSET
CONNECTION
(8 PIN DIN)
RED POWER LED
CHARGER INPUT SOCKET
REMOTE HANDSET CONNECTION
(8 PIN DIN)
Page 14
14
TYPICAL PROBE CONFIGURATION (KMCP6)
THERMOCOUPLE CONNECTION (FLUE)
INLET SOCKET FOR OPTIONAL AIR TEMPERATURE PROBE
PROBE GAS CONNECTION TO WATERTRAP
WATERTRAP LOCATED ON SPIGOT IN CASE
DEPTH STOP CONE
SAMPLING PROBE
Page 15
15
ANALYSER CONNECTIONS
CHARGER INPUT SOCKET
DUAL PRESSURE
PORTS
(8 PIN DIN)
OXYGEN SENSOR
CONNECTOR
FLUE SOCKET
INLET SOCKET
AUX CONNECTOR(25 PIN 'D')
ANALYSER 'ON' SWITCH
REMOTE HANDSET CONNECTION (8 PIN DIN)
RED POWER LED
Page 16
16
GETTING STARTED
Check that you have all the items you have ordered.
Before attempting to use the analyser to take readings it is recommended that the batteries are fully charged.
When the charger is connected to the analyser and powered up the red LED by the charger socket will flash until the batteries are fully charged. Once the batteries are fully charged the LED will no longer be illuminated.
To charge the handset; connect the handset to the main analyser unit using its cable. Whilst switched off, but charging, the display will show the Kane logo and a battery charging icon in the bottom right hand corner of the handset screen. Note: the handset battery is charged via the external battery charger and not from the analyser’s internal battery. Once the handset battery is fully charged the icon will disappear from the screen.
The handset can also be charged directly using the mains charger as used for the analyser.
BEFORE TAKING READINGS
You need to establish the current set up of the analyser and then make the changes that you need to suit exactly what you want to do.
So press the MENU key and then select STATUS as described below.
Page 17
17
MENU: ALL THE OPTIONS
Press the MENU key.
----------------------------------MENU---------------------------------
> STATUS SETUP ANALYSER UNITS CO ALARM SCREEN REPORTS SERVICE MANUAL AIR ZERO MANUAL PRESSURE ZERO
11:33:56 23/03/13 S04
The > symbol acts as the cursor. It can be moved up or down by pressing the UP or DOWN keys.
Having made a selection press the ENTER key.
The bottom line of this screen lists:
the time  the date  the number of satellites being received  the wireless connection status  the estimated charge in the handset battery
Page 18
18
STATUS: ESTABLISHING THE ANALYSER'S SET-UP
Press MENU and then select STATUS by pressing ENTER.
----------------------------------MENU---------------------------------
HANDSET SW19170 V1.01 SERIAL NO. 999999107
ANALYSER SW19171 V1.01 SERIAL NO, 999999207
MAIN BATTERY 98% HEATER BATTERY 100%
CAL DATE = 331
11:33:56 23/03/13 S04
This screen lists :
the software version in the handset  the handset serial number  the analyser unit software version number  the analyser unit serial number  the estimated charge in the main battery  the estimated charge in the heater battery (if fitted)  the number of days before annual re-calibration is due
Page 19
19
Press DOWN to move to the next screen.
----------------------------------MENU---------------------------------
REFERENCE O2 = 3.0% NOx CALCULATION = SUM REFERENCE NOx = 5%
EFFICIENCY = NET FUEL = NATURAL GAS FUEL SOURCE = UK
CO ALARM SET = NO CO ALARM LEVEL = 400ppm
11:33:56 23/03/13 S04
Press DOWN to move to the next screen.
----------------------------------MENU--------------------------------­AUTO PURGE = YES AUTO ZERO = NO AUTO PUMP FLOW = NO MAIN PURGE DURATION TIME = 2 mins MAIN PURGE INTERVAL TIME = 30 mins
LOG 0006 AUTO LOG/PRINT TIME = 30 mins START AUTO LOG = NO START AUTO PRINTING = NO
11:33:56 23/03/13 S04
Press DOWN to move back to the first screen.
Press PRINT to print this status on the analyser’s printer.
Press MENU to EXIT.
Page 20
20
SETUP: CONFIGURING THE ANALYSER’S SETUP
----------------------------------MENU---------------------------------
> LANGUAGE MAIN PURGE PRINTER AUTO SET TIME SET TIME SET DATE HEATER STATUS SELECT ANALYSER BOX SERIAL No. WIRELESS SETUP WIRELESS PASS KEY BACK
11:33:56 23/03/13 S04
LANGUAGE:
Align cursor using UP or DOWN keys, then press ENTER.
Use UP or DOWN keys to scroll through the selection.
Press ENTER to select.
Page 21
21
MAIN PURGE:
----------------------------------MENU---------------------------------
> AUTO PURGE MAIN PURGE DURATION TIME MAIN PURGE INTERVAL TIME AUTO ZERO CAL AUTO PUMP FLOW BACK
11:33:56 23/03/13 S04
AUTO PURGE: Set YES or NO
AUTO PURGE DURATION TIME: Set fresh air purge duration to between 2 and 30
minutes
MAIN PURGE INTERVAL TIME : Set the interval between fresh air purges to
between 10 and 120 minutes.
AUTO ZERO CAL: Set YES or NO to automatically re-zero sensors at
the end of a main purge cycle.
AUTO PUMP FLOW: Set YES or NO to automatically control the pump
flow. This may be required for some regulatory test protocols
After switch on, the first purge interal is automatically set to 10 minutes if the optional IR module is fitted. Changes in the purge interval are implemented after completion of the next purge cycle. To implement a change immediately do a “MANUAL AIR ZERO”.
PRINTER:
This sets the destination for outputs from the handset
Page 22
22
The choices of outputs are:
KANEIRP  KANEIRP-2  ANALYSER PRINTER  SERIAL  WIRELESS
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
AUTO SET TIME:
This function is locked off if reports have been stored. To allow the function to operate, delete the reports.
This allows the time to be set automatically from the GPS signals. Select NO to maintain manual setting
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET TIME:
This function is locked off if “AUTO SET TIME” is activated or reports have been logged. If this function is locked, delete the reports.
----------------------------------MENU---------------------------------
> SET TIME
HH:MM:SS
13:59:47
11:33:56 23/03/13 S04
If manual setting is enabled use UP or DOWN keys to change values.
Press ENTER to select.
Page 23
23
SET DATE:
This function can only be set manually, not by GPS.
If this function is locked, delete the reports.
----------------------------------MENU---------------------------------
SET DATE
DD:MM:YY
26:04:13
11:33:56 23/03/13 S04
If manual setting is enabled use UP or DOWN keys to change values.
Press ENTER to select.
HEATER STATUS:
Allows the heaters to be switched off totally or to operate automatically.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SELECT ANALYSER UNIT SERIAL NUMBER:
If more than one analyser unit is within Bluetooth range the handset needs to be set to communicate with the correct unit,
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
Page 24
24
WIRELESS SET UP:
The handset can communicate with an analyser unit or a PC. Wireless can also be switched off and a cable can be used.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
WIRELESS PASSKEY:
This confirms the Passkey setting.
The wireless Passkey is 1111.
Page 25
25
ANALYSER UNITS: CONFIGURES ALL THE DATA SOURCES AND SETTINGS
----------------------------------MENU---------------------------------
FUEL ORIGIN FUEL TYPE EFFICIENCY GAS UNITS COMPENSATION TEMPERATURE PRESSURE SET PERCENTAGE REFERENCE O2 SET NOx CALCULATION SET COMPENSATION CONVERSION FACTORS BACK
11:33:56 23/03/13 S04
FUEL ORIGIN:
Select from a list of country specific fuel tables.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
FUEL TYPE:
Select from the list of fuel types associated with the chosen origin. The K values for the selected fuel are also shown.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
There are 5 User Fuels that can be loaded from a PC. The title of these user fuels can be edited using the keypad on the handset.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
EFFICIENCY:
Select NET or GROSS efficiency calculation
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
Page 26
26
GAS UNITS:
Select ppm or ppm(n) or mg/m3 or mg/m3(n).
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
TEMPERATURE:
Allows the selection of Fahrenheit or Celsius.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
PRESSURE:
Allows the selection of pressure units.
Select from: mbar, In H2O, mm H2O, hPa, psi,.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET PERCENTAGE REFERENCE O2:
Can be set between 0% (equivalent to OFF) and 10%.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET NOX CALCULATION:
Select from: SUM, NO2 or NO
Set REFERENCE NOx to the percentage required or defined by local regulations. Typically 5% NO2 is added to an NO reading. The value can be user set.
SUM adds the readings from an NO sensor and an NO2 sensor when fitted
NO calculates an NOx reading from the NO reading where NOx = NO x 1.1 NO2 calculates an NOx reading from the NO2 reading where NOx = NO2 x 2.05
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
SET COMPENSATION:
Select from YES or NO
CONVERSION FACTORS:
Display the Propane Equivalency Factor PEF and the Methane Equivalency Factor for the IR module (if fitted) and the pitot factor – change from 0.10 to 1.00 as determined by the pitot tube being used.
Page 27
27
CO ALARM:
----------------------------------MENU---------------------------------
CO ALARM SET CO ALARM LEVEL BACK
11:33:56 23/03/13 S04
CO ALARM SET:
Switch the alarm ON or OFF.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
CO ALARM LEVEL:
Allows a specific CO level in ppm to be set as the alarm trigger point.
Use UP or DOWN keys to change the digits. Press ENTER to select and move to the next digit.
The display will show ‘++++’ when the CO alarm is triggered.
Page 28
28
SCREEN:
The screen display is fully configurable.
----------------------------------MENU---------------------------------
CONTRAST B’LIGHT MODE LINES 1 - 6 LINES 7 - 12 LINES 13 - 18 LINES 19 - 24 LINES 25 - 30 BACK
11:33:56 23/03/13 S04
CONTRAST:
Allows the display to be darkened or lightened. Default value is 14.
Use UP or DOWN keys to change the digits. Press ENTER to select and move to the next digit.
B'LIGHT:
The switch off time for the backlight can be set for between 30 and 300 seconds,
Use UP or DOWN keys to change the digits. Press ENTER to select and move to the next digit.
During normal measurements press ENTER to switch the backlight on.
MODE:
The main display can be set for SMALL font or LARGE font.
Page 29
29
LINES:
This feature allows users to customise the screen display to suit their own requirements.
Use UP or DOWN keys to change the selection. Press ENTER to select and move to the next digit.
So for Lines 1-6, the screen shows:
----------------------------------MENU---------------------------------
LINE 1 = CO 0ppm LINE 2 = HC 0ppm LINE 3 = NO 0ppm LINE 4 = NO2 0ppm LINE 5 = SO2 0ppm LINE 6 = NOx 0ppm BACK
11:33:56 23/03/13 S04
The parameter to be displayed on each line can be individually selected.
Use UP or DOWN keys to change the selection. Press ENTER to select and move to the next digit.
Page 30
30
REPORTS: CONFIGURES REPORTS
----------------------------------MENU---------------------------------
VIEW REPORTS DELETE ALL REPORTS AUTO LOG TIME START AUTO LOG START AUTO PRINTING HEADER 1 HEADER 2 FORM FEED
BACK
11:33:56 23/03/13 S04
VIEW REPORTS:
This selection displays a 'main screen' with a LOG No near the top left hand of the display. This number can be changed using UP or DOWN and the display automatically changes.
----------------------------------MENU---------------------------------
LOG 0000 DATE 00/00/00 TIME 12:00:00AM ORIGIN UK FUEL NATURAL GAS
O2 0.00% CO2 0.0% CO 0ppm NO 0ppm NO2 0ppm NOx 0ppm SO2 0ppm H2S ----ppm O2 19.98% FLUE 0.0deg C INLET 0.0deg C AMBIENT 0.0deg C NETT 0.0deg C LOSS 10 DRY 0 WET 10 11:33:56 23/03/13 S04
Page 31
31
DELETE ALL REPORTS:
All reports can be deleted. A confirmation YES is required
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
AUTO LOG TIME:
Automatic logging/printing can be selected for intervals between 10 seconds and 90 minutes
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
START AUTO LOG:
Select YES or NO.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
START AUTO PRINTING:
Select YES or NO.
Use UP or DOWN keys to scroll through the selection. Press ENTER to select.
HEADER 1: 16 CHARACTERS
Allows the printed header line: YOUR COMPANY to be changed.
Use UP or DOWN keys to change the characters. Press ENTER to select and move to the next character.
HEADER 2: 16 CHARACTERS
Allows the printed header line: NAME & PHONE No. to be changed.
Use UP or DOWN keys to change the characters. Press ENTER to select and move to the next character.
FORM FEED:
Allows remote paper feeding on printers.
Page 32
32
SERVICE:
CODE: For use by authorised service agents.
CAL DATE = number of days before annual re-calibration is due.
MANUAL AIR ZERO:
Select this and press ENTER to initiate a fresh air purge and sensor zeroing
MANUAL PRESSURE ZERO:
Select this and press ENTER to re-zero the pressure sensor
Page 33
33
BEFORE USING THE ANALYSER FOR THE FIRST TIME.
SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted from the bottom of the instrument. This instrument must only be used in well ventilated locations. It must only be used by trained and competent persons after due consideration of all the potential hazards.
FIRST TIME USE
Charge the batteries for 12 hours, following this an overnight charge should be sufficient for an average 8 hour day. There may be three battery packs that need charging, the main battery, the optional heater battery and the handset battery. The handset battery can use the same charger as the analyser unit or can be charged via the main unit using a handset lead. All batteries are NiMh.
Whilst charging the red LED will flash.
We offer a wide choice of probes which are not supplied as standard and must be ordered as a separate item.
Take time to read this manual fully.
TIP: Take a look at the Spare Parts list and order some replacement filters and
paper rolls now.
NORMAL START UP SEQUENCE
EVERY TIME YOU USE THE ANALYSER
BEFORE SWITCH-ON CHECK THAT:
the oxygen sensor is connected
the particle filter is not dirty
the sulphur filter is fitted for heavy oil or coal
the water trap and probe line are empty of water
all hose connections, etc, are properly made
Page 34
34
the paper roll is fitted
the analyser unit is in fresh air
the water trap is vertical
the flue temperature is connected
the instrument is placed on a clean, flat, level surface
Switch ON the analyser by pressing ON/OFF on the handset. You also need to press the ON/OFF switch on the analyser main unit.
AUTOMATIC CALIBRATION
During this sequence the analyser pumps fresh air into the sensors to allow toxic sensors to be set to zero and the oxygen sensor to be set to 20.9 %.
During this sequence the handset display will show the following:
KANE QUINTOX
SW 19170 v1.01
SERIAL NO. 999999107
GPS CONNECTED
WIRELESS CONNECTING ……..
Note: The software version number and serial number are examples only.
Page 35
35
If there is no wireless communication between the handset and the main unit the following will appear on the screen.
WIRELESS COMMS ERROR TO ANALYSER UNIT
REFER TO INSTRUCTIONS
PRESS ENTER TO CONTINUE
11:33:56 23/03/13 S04
Press ENTER to continue and follow the instructions.
Once wireless communication is established the main measurement screen will appear:
Page 36
36
MAIN MEASUREMENT SCREEN
IN SMALL FONT MODE
ZERO TIME 60M
CO2 ----% NETT -N\F-deg C HC ----ppm LOSS ----­CO 0ppm DRY ----­NO 0ppm WET -N\F­NO2 0ppm CO LOSS 0% NOx 0ppm P INDEX 0.00% SO2 0ppm CO/CO2 R0.0000 H2S ----ppm EFF (G) ----% O2 19.98% XAIR ----% FLUE -N\F-deg C PRESSURE 0.00mbar INLET 20.3 deg C ATM 986.6mbar AMBIENT
GPS (X) -00011.458
11:33:56 23/03/13 S04
TOP LINE: shows status messages
GASES: O2 and CO2 are shown in %
other gases shown in ppm or other user selected units CO can be shown in % if optional IR bench fitted
TEMPERATURES: displayed in C or F. N\F = not fitted
---- occurs for calculations when N\F applies
---- occurs when a calculation cannot be made due to an out of range value (Eg zero)
Atmospheric pressure (ATM) is always displayed in mbar.
Page 37
37
IN LARGE FONT MODE
There are 5 screens that are accessed using the UP or DOWN keys
-----------------------------------MENU----------------------------------
> LINE 1= CO2 ----% LINE 2= HC ----ppm LINE 3= CO 0ppm LINE 4= NO 0ppm LINE 5= NO2 0ppm LINE 6= NOx 0ppm BACK
11:33:56 23/03/13 S04
-----------------------------------MENU----------------------------------
> LINE 7= SO2 0ppm LINE 8= H2S ----ppm LINE 9= O2 19.98% LINE 10= FLUE -N\F-deg C LINE 11= INLET -N\F-deg C LINE 12= AMBIENT 20.3deg C BACK
11:33:56 23/03/13 S04
Page 38
38
-----------------------------------MENU----------------------------------
> LINE 13= NETT -N\F-deg C LINE 14= LOSS ----­ LINE 15= DRY ----­ LINE 16= WET -N\F­ LINE 17= CO LOSS 0% LINE 18= P INDEX 0.00% BACK
11:33:56 23/03/13 S04
-----------------------------------MENU----------------------------------
> LINE 19= CO/CO2 R0.0000 LINE 20= EFF (G) ----% LINE 21= XAIR ----% LINE 22= PRESSURE 0.00mbar LINE 23= ATM 986.6mbar LINE 24= GPS (X) -00011.458 BACK
11:33:56 23/03/13 S04
-----------------------------------MENU----------------------------------
> LINE 25= GPS (Y) +5148.0957 LINE 26= MAIN BAT 24% LINE 27= HEAT BAT 100% LINE 28= LINE 29= LINE 30= BACK
11:33:56 23/03/13 S04
Page 39
39
SAMPLING THE FLUE GAS
Once the automatic calibration procedure has been completed and the specific fuel has been selected the probe can be inserted into the desired sampling point.
It is recommended that the sampling point be located at least two flue diameters downstream of any bend and that the probe tip is in the centre of the flue (this is normally the point of the hottest temperature). With balanced flues and other industrial units the probe should be positioned far enough into the flue so that no air can ‘back flush’ into the probe.
The probe depth stop cone provided with the instrument allows the probe to be used in holes whose diameters range from 8 mm to 21 mm (5/16 to 4/5 inch).
The standard probe is rated at 650C/1202F. Temperatures of up to 1100C/2012F can be accommodated using an optional high temperature probe.
TIP: To conserve battery power, switch off the pump when you are not taking a
measurement. Use the pump key to turn the pump ON and OFF.
LONG TERM MONITORING
There are a number of things that need to be considered for successful long term unattended monitoring:
The provision of enough power for the duration of the test
The capability to empty the water trap
Regular fresh air purging of the sensors
Protection from rain or water spray from the process being monitored.
If a mains power source is being used it is strongly recommended that the supply
cable is protected by a suitable Residual Current Device (RCD).
Unless the water trap is to be regularly inspected then a pumped water trap should be
Page 40
40
fitted.
Electrochemical sensors need regular refreshing with fresh air, prefereably at around 50% RH. They also need a small percentage of oxygen to be present in the sampled gas. If there is zero oxygen the output from the sensors will decay over time (10 mins or so). In such circumstances, fresh air purge should be programmed for a 50% duty cycle every 10 minutes.
The longest sampling time without purging should be limited to 2 hours and then purge for 30 mintues.
When the KHC infra-red module is fitted, for maximum accuracy it is recommended that purging occurs every 30 minutes.
KMDM110/230 SAMPLE CONDITIONING UNIT
This module is fitted in the front compartment of a standard KANE9206 carry case and comprises a Peltier fan cooled chiller assembly, a peristaltic pump to automatically remove condensate, the control electronics and a power supply module. The module is supported on an aluminium alloy chassis.
The chiller is connected to a flue mounted electrically heated probe (KMHP1200) by a 3 metre long heated line with automatic temperature control (KMHL3000). Because the gas that is extracted from the flue is maintained at 120oC no condensation occurs in the probe or the hose and so no sample gas is lost in the condensate. The chiller flash cools the sample gas to below the ambient dewpoint and any water in the gas immediately condenses. The condensate is then pumped away using a peristaltic pump. Because the gas has no chance to remain in contact with the condensate, volatile sample gas is not lost into the condensate. The chilled gas then naturally warms up as it passes through the sampling pump to the sensors and as it does so its humidity reduces and there is no risk of further condensation.
SETTING UP
The heated sample probe must be connected to the top of the heated line and a gas tight connection be made without over tightening the connections. This joint must then be thermally insulated. Both the heated probe and the heated line must be connected to a mains power source via a suitable Residual Current Device (RCD) and be left powered up for 20 minutes to achieve their operating temperature before attempting to extract sample gas.
When the probe is inserted into the sampling point is must be suitably supported to prevent bending and unnecessary strain. Likewise the heated line should be carefully supported and never be twisted or kinked as this may damage the internal heating elements.
The heated line is connected to the chiller by attaching the end of the line to the through bulkhead connection in the KANE9206 carry case. A short flexible connection then links the hose to the chiller. The chiller needs to be powered up for at least 10 minutes before it
Page 41
41
is used.
The peristaltic pump operates automatically. Always check that the drain of the peristaltic pump is clear and that there are no blockages. The peristaltic pump needs to have its flexible rotor replaced after every 1000 hours of operation.
To operate efficiently the chiller needs to be well ventilated so the case lid must be removed, however the unit must be protected externally to prevent ingress of water from either the plant being tested or from rainfall.
Page 42
42
MAINTENANCE
EMPTYING AND CLEANING THE IN-LINE WATER TRAP
The water trap should be checked and emptied on a regular basis. Water vapour will condense and gather in the probe line this may move suddenly to the trap when the probe is moved. Care should be taken at all time.
Emptying of the water trap is detailed below :-
Carefully remove the end cap from the in-line housing. Dispose of the condensate in a suitable drain, care must be taken as it could be acidic. If condensate spills onto the skin or clothing, clean off immediately using fresh water, seek medical advice if problems occur.
CHANGING THE PARTICLE FILTER
This is a very important part of the analyser and should be changed regularly. It prevents dust and dirty particles entering the pump and sensors and hence causing damage. The filter MUST be changed when it appears discoloured.
Remove the end cap from the filter housing. Carefully remove the paper filter element and dispose of it. Clean the inside of the filter housing with a suitable soft cloth. Insert a new filter element onto the spigot on the filter end cap and carefully insert it into the filter body.
Page 43
43
CHARGING THE BATTERY
It is important that the battery is charged on a regular basis. The instrument constantly powers the internal sensors and may flatten the battery if left unattended for some months. Connect the charger supplied with the instrument to the correct mains supply.
Note: The correct charger type may be required for your local voltage i.e. 110 or 220
volts AC
Insert the plug in the socket marked CHARGER INPUT SOCKET The CHARGER ON RED LED will flash showing the instrument is charging.
CHANGING THE PAPER ROLL
To change the paper roll remove the printer cover by loosening the two screws holding it down. Remove the old paper roll core and insert the new roll so that it sits as follows :-
Feed the free end of paper into the printer through the metal slot beneath the printer ribbon. Start the paper feed sequence until the paper has emerged from the top of the printer, feed the loose end through the cover and refit.
TO START PAPER FEED
Go to MENU, REPORTS, FORM FEED
CHANGING THE PRINTER RIBBON
The printer ribbon cartridge will last for approximately two rolls of paper. Remove the printer cover as detailed above.
Marked on one end of the cartridge is PUSH. Gently press down on this end and the ribbon cartridge will pop up at the other end. Remove the cartridge and dispose of.
Page 44
44
Fit a new ribbon guiding the paper roll between the exposed ribbon and cartridge body.
Refit printer cover.
Page 45
45
PROBLEM SOLVING
The following is a list of problems that may occur on the instrument through its operating life. If the cause of the fault is not easy to identify then we advise you to contact the Kane International Service Department or an International Distributor for expert advice.
Fault symptom Causes Oxygen too high CO2 too low
Air leaking into probe, tubing, water trap, connectors or internal to instrument.
Oxygen cell needs replacing. Analyser not holding charge Analyser not charging
Battery exhausted.
AC charger not giving correct output.
Fuse blown in charger plug. Analyser does not respond to flue gas
Particle filter blocked.
Probe or tubing blocked.
Pump not working or damaged with
contaminants. Flue temperature readings erratic Temperature plug reversed in socket.
Faulty connection or break in cable or plug. Analyser automatically switches off in operation.
Battery below alarm level.
Battery quickly discharging and is faulty. Display is blank. The contrast setting has been lost and
requires resetting. Disconnect handset
lead and reconnect. Set contrast as in
MENU : SCREEN : CONTRAST
HOW TO GET EXPERT HELP
There will be occasions when despite having read the manual there will be problems that you cannot resolve and so you need external help.
Before calling Kane International or one of its International Distributors please first check the following:
Find the serial number of the instrument. It is located on the label close to where the charger and handset leads plug into the analyser. Also make a note of which sensor are fitted by observing the tick unites on the same label.
If the handset and analyser are operating you can also determine the issue of software loaded in the analyser and its handset by viewing STATUS. If you can, take a printout of STATUS and a printout of the measurement screen so that they can be faxed or emailed to your technical support advisor.
Page 46
46
ANALYSER ANNUAL RECALIBRATION AND SERVICE
The analyser should be re-calibrated and serviced annually to stop any long-term sensor or electronics drift or accidental damage.
Local regulations may require more frequent re-calibration.
In the UK Kane International has service facilities at Atherton near Manchester (Tel: 01942-873434), the primary service centre for UK customers and at Welwyn Garden City in Hertfordshire (Tel: 01707-384834), the primary service centre for non-UK customers.
By sending your analyser back to Kane for an annual service (check www.kane.co.uk for details) you have the opportunity to extend the warranty on your analyser to 5 years.
Page 47
47
RETURNING YOUR ANALYSER TO KANE
When returning your KANE9206, please always ensure that you enclose:
Your full contact details A daytime telephone number Details of faults you might have experienced Any relevant accessories (eg. Probe, printer, adaptor and leak detectors). Any
accessories that are returned will be checked. If an accessory has failed then we will quote you for a repair or a replacement.
PACKING YOUR ANALYSER
When returning your analyser, please pack it appropriately to prevent any damage during transit.
Before sealing your package, please ensure that you have enclosed the items listed above and that it is clearly marked for the attention of:
For UK customers:
For non-UK customers:
Northern Service Centre
Kane International Ltd
Gibfield Park Avenue
Atherton
Manchester
M46 0SY
Southern Service Centre
Kane International Ltd
Kane House, Swallowfield
Welwyn Garden City
Hertfordshire
AL7 1JG
SENDING YOUR ANALYSER
Once the analyser has been securely packed then your package is ready for shipment back to Kane. If you do not have an account with a courier company you can take your package to your local Post Office. It is advisable to send the package by Special Delivery so that it is insured and traceable while in transit.
WHEN WE RECEIVE YOUR ANALYSER
On receipt of your package, our Service Engineers will inspect the analyser and any accessories and confirm to you the total service cost. Once you have accepted this the work will be carried out, and upon completion the analyser returned to you.
If you have any questions that we haven’t answered, please feel free to contact our Southern Service Centre:
For UK customers:
For non-UK customers:
Tel: 01942 873434
Fax: 01942 873558
Email: nservice@kane.co.uk
Tel: 01707 384834
Fax: 01707 384833
Email: sservice@kane.co.uk
Page 48
48
Service Returns
(Simply cut out and attach to your package)
Southern Service Centre
Kane International Ltd
Kane House, Swallowfield
Welwyn Garden City
Hertfordshire
AL7 1JG
Northern Service Centre
Kane International Ltd
Gibfield Park Avenue
Atherton
Manchester
M46 0SY
Southern Service Centre
Kane International Ltd
Kane House, Swallowfield
Welwyn Garden City
Hertfordshire
AL7 1JG
Page 49
49
Page 50
50
PRODUCT SPECIFICATION
UNIT
Temp Measurement
Resolution
Range Accuracy
Flue Temperature
0.1o (C/F) 0-1100°C 32-2140°F * Use high temperature probe for gases
>600°C/1112°F
1.0 o C +0.3% of reading
Inlet Temperature 0.1o (C/F) 0-600°C
0-999°F
1.0 o C +0.3% of reading
Gas Measurement*1 Resolution Range Overrange Reading Accuracy
Oxygen (02): 0.01% 0-25% 30% - -0.1% +0.2%
Carbon monoxide (CO): (standard: H
compensated)
1ppm 2000 4000 <100ppm
>100ppm <2000ppm >2000ppm <4000ppm
+/-5ppm +/-5% of reading +/-10% reading
Nitric oxide (NO): (high range0)
1ppm 1000 5000 <100ppm
>100ppm <1000ppm >1000ppm <5000ppm
+/-5ppm +/-5% of reading +/-10% reading
Nitric oxide (NO) (low range)
1ppm 100 300 <100ppm
>100ppm <300ppm
+/-5ppm +/-10% of reading
Nitrogen dioxide (NO2):
1ppm 100 1000 <100ppm
>100ppm <1000ppm
+/-5ppm +/-10% of reading
Sulphur dioxide (SO2) (low range):
1ppm 100 500 <100ppm
>100ppm <500ppm
+/-5ppm +/-10% of reading
Sulphur dioxide (SO2) (high range):
1ppm 2000 5000 <100ppm
>100ppm <2000ppm >2000ppm <5000ppm
+/-5ppm +/-5% of reading +/-10% reading
Hydrogen sulphide (H2S):
1ppm 200 300 <100ppm
>100ppm <200ppm >200ppm <300ppm
+/-5ppm +/-5% reading +/-10% of reading
Gas Measurement*1 Resolution Range Accuracy
Pressure 0.01mbar 0-150 mbar + 0.5% Full Scale
Carbon dioxide (CO
2
)*2
0.1% 0 – Fuel Value + 0.3%
Efficiency *2 0.1% 0-100% + 1%
*1
using dry test gases at STP
*2
calculated
NB: all ppm reading can be displayed in mg/m3 and can be normalised
Page 51
51
OPTIONAL IR MODULE
Hydrocarbons:
Range: Overrange Accuracy Resolution 0-5,000ppm 10,000ppm +5% of reading and + 12ppm vol. 1ppm
CO2:
Range: Overrange Accuracy Resolution 0-20% 40% +5% of reading and + 0.5% vol. 0.1%
CO:
Range: Overrange Accuracy Resolution 0-10% 20% +5% of reading and + 0.2% vol.
<30%
0.1%
Response time T90: 30 seconds Warm up time: 3 minutes Operating temperature range: 5 to 50 deg C. Operating humidity: 10-80% non condensing Power: Supplied by KANE9206
Conversion Factors from a Hexane Calibration
Hexane multiply by 1 Propane multiply by 0.5 Methane multiply by 0.05
HANDSET
Dimensions 240 mm long
135 mm high
70 mm wide Keypad tactile keys Display graphical with backlight and contrast control
ANALYSER
Dimensions 420 mm long
345 mm high
225 mm wide
Page 52
52
EXTENSION CABLE
Specification: 8 pin DIN cable Cable lengths: 10m Standard
5-20m-Optional
MAIN BATTERY AND OPTIONAL HEATER BATTERY
Type: NiMH Rechargeable (12V, 2AH) Life: 8 hours from full charge Charge time: 12 hours trickle
4 hours fast charge
BATTERY CHARGER
Input: 100V-240V AC 60 watts Output: 15V DC @ 4 amps
PUMP
Flow rate: 2 Litres/Minute nominal
500 mbar static suction
INTEGRAL PRINTER
16 character dot matrix. Plain paper
AMBIENT OPERATING RANGE
-10°C to + 55°C
< 85% RH non condensing
Storage: -10°C to 55°C
For regulated testing to EN50379, ambient temperature range: 0°C to + 55°C
Page 53
53
KMHL3000: HEATED SAMPLE LINE
Power supply: 220V ac @ maximum 300 watts
KMHP1200: HEATED PROBE: KMHP1200
Power supply: 220V ac @ maximum 100 watts 1200mm insertion length, 8 mm diameter rated to 1000 o C
KMDM110/230: SAMPLE CONDITIONING UNIT
Power supply: 220 Vac @ 5 amps peak.
PROBE
Choose from a range of probe options. See probe leaflet.
OPTIONAL PORTABLE PRINTERS
Compatible with KMIRP-2
Page 54
54
ELECTROMAGNETIC COMPATIBILITY (CE) STATEMENT
This product has been tested for
compliance with the following generic
standards:
EN 61000-6-3 : 2011 EN 61000-6-1 : 2007
and is certified to be compliant
Specification EC/EMC/KI/KANE9206
details the specific test configuration,
performance and conditions of use.
SAFETY STANDARD
This product complies with the EN61010
Safety Standard
(Safety requirements for electrical
equipment for measurement, control
and laboratory use):
EN61010-1 : 2010
Protection Class 3 (SELV)
END OF LIFE DISPOSAL
The Waste Electrical or Electronic Equipment (WEEE) Directive requires countries in the EU to maximise collection and environmentally responsible processing of these items.
Products are now labelled with a crossed out wheeled bin symbol to remind you that they can be recycled.
BATTERY DISPOSAL
All the user replaceable batteries used in this product are NiMh and are suitable for recycling through any local waste portable battery recycling scheme.
Please note: Batteries used in this instrument should be disposed of in accordance with current legislation and local guidelines.
Page 55
55
EN 50379 REGULATED INSTRUCTIONS
EN 50379 Section 4.3.3 “Instructions” defines a number of specific points that must be included in the relevant instruction manuals. The paragraph numbering below relates to that section of EN 50379.
a) The KANE9206 is compliant the EN 50379 Part 2 as detailed in the third party
approvals issued by TÜV.
b) The KANE9206 is intended to be used with the following fuels:
Natural gas, Natural gas 2, Light oil (28/35 sec), Heavy oil, Coal, Anthractite, Coke, Propane, Butane, Gascor, Kinsale gas, LPG, Bio gas, Wood pellets and 5 user fuels.
c) The KANE9206 handset is designed for use with either non-rechargeable alkaline AA
cells or rechargeable NiMH AA cells. Four cells are needed. Types cannot be mixed. Under no circumstances should any attempt be made to recharge alkaline cells.
The KANE9206 analyser is designed for use with a KANE9206 rechargeable NiMH battery pack. Under no circumstances should any attempt be made to use another type of battery pack other than the manufacture specified one.
The battery charger supplied with the KANE9206 is rated for indoor use only. Its
voltage input must be in the range 100 – 240 V ac at 50 – 60 Hz with a current capability of 1.5A. The chargers output voltage is 15 V dc at a maximum of 4A.
The charger has no user serviceable components.
Only a correctly specified and rated charger must be used with the KANE9206.
d) The KANE9206 is not designed for continuous use and is not suitable for use as a
fixed safety alarm.
e) An explanation of all the symbols used on the analyser’s display is given in Appendix
A of this manual.
f) The recommended minimum time required to perform one complete measurement
cycle and achieve correct indication of the measured values in EN 50379 Part 2 is 110 seconds. This is based on the T90 times defined in the standard, always assuming that parameters being measured have reached stability. This time is the summation of the times for a draught test (10 secs) and a combustion test (90 secs) plus the time to move the hose connection from the pressure input to the water trap (10 secs)
g) This section is non-applicable.
Page 56
56
h) Some commonly occurring materials, vapour or gases may affect the operation of the
KANE9206 in the long or the short term though in normal use Kane International Ltd is not aware of any specific issues that have affected the product. The following list is included to satisfy the stated requirements of EN 50379:
Solvents Cleaning fluids Polishes Paints Petrochemicals Corrosive gases
i) The KANE9206 can be fitted with up to 6 electrochemical sensors although only CO,
NO, SO2 & O2 are covered under EN50379. They have an expected life of more than 2 years. The calibration of these sensors must be confirmed on an annual basis.
The batteries have an expected operational life of more than 500 re-charge cycles.
j) The KANE9206 is designed to operate at ambient temperatures in the range -10oC to
+55oC with relative humidity of 10% to 90% non-condensing although the analyser was only tested under EN 50379 with ambient temperatures in the range -0oC to +55oC. Whilst it is recommended that the analyser is given the protection of a carry case during transportation it is not required for normal operation.
k) The KANE9206 has an initial start-up delay following switch on of approx. 180
seconds. There is no additional delay after battery replacement.
l) Most sensors used in combustion analysers give a zero output when they fail and it is
widely recommended that analysers are regularly checked (also known as a bump test) using either a can of test gas or a known source of combustion products.
The KANE9206 must have its calibration checked on an annual basis and be issued
with a traceable Certificate of Calibration.
The sensor within the KANE9206 can only be replaced by Kane International Ltd or
one of its trained and approved service partners.
The water trap should be checked on a regular basis whilst the analyser is in use
(every few minutes) as the amount of condensate generated varies with the fuel type, atmospheric conditions and the appliances operating characteristics.
The particle filter should be checked at least on a daily basis when using ‘clean’ fuels
and more often when using liquid or solid fuels.
Detailed instructions regarding the changing of the filter and the emptying of the
water trap are given in Section Maintenance of this manual.
Page 57
57
m) WARNING! When using a KANE9206 to test an appliance a full visual inspection of the
appliance, in accordance with its manufacturer’s instructions, must also be carried out.
n) WARNING!
When using a KANE9206, a regular inspection of the water trap and filter must be carried out as blockages can lead to inaccurate measurements.
o) Instructions for testing for leaks and blockages of the gas sample system is covered
in the Section Problem Solving of this manual
p) The KANE9206 changeable parameters and their valid ranges are:
Languages: English, French, German & Dutch Auto purge: Yes & No Main purge duration time: 2 to 60 minutes Main purge interval time: 10 to 120 minutes Auto zero: Yes & No Auto pump flow: Yes & No Printer: KANE IRP, KANE IRP-2, Serial, Wireless, Analyser
printer Auto set time: Yes & No Set time: 00:00:00 to 23:59:59 Set date: 01/01/01 to 31/12/99 Heater status: Auto & Off Wireless setup: Off, To PC & to Analyser Wireless passkey: 0000 to 9999 Fuel origin: UK, France, Germany, Netherlands, Italy, Spain,
Hungary, Poland, Finland, Sweden, China, North
America. Fuel type: Natural gas, Natural gas 2, Light oil (28/35 sec),
Heavy oil, Coal, Anthractite, Coke, Propane, Butane,
Gascor, Kinsale gas, LPG, Bio gas, Wood pellets and
5 user fuels. Efficiency: Net, Gross, Condensing Net & Condensing Gross. Gas units: ppm, ppm(n), mg/m3 & mg/m3(n) Temperature: Celsius and Fahrenheit Pressure: psi, hPA, mm H2O, In H2O, mbar & m/s Set percentage reference O2: 0.0% to 10.0% Set NOx Calculation: NO, NO2 & SUM Reference NOx: 0% to 20% Set compensation: on & off Pitot: 0.10 to 1.00 CO alarm set: on & off CO alarm level: 0000 to 9999 Display contrast: 0 to 15 Display backlight: 30 to 300 seconds Display mode: small font & large font
Page 58
58
Display Lines (1 to 30): Analyser main battery, Analyser heater battery, CO2
reading, HC reading, CO reading, NO reading, NO2
reading, NOx reading, SO2 reading, H2S reading,
O2 reading, flue temperature reading, inlet
temperature reading, ambient temperature reading,
nett temperature reading, total loss reading, dry loss
reading, wet loss reading, CO loss reading,
poison index reading, CO_CO2 ratio reading,
efficiency reading, excessive air reading, pressure
reading, atmospheric pressure reading, GPS
longitude reading, GPS latitude reading, blank &
lambda reading. Auto log /print time: 10seconds to 90 minutes Start auto log: yes & no Start auto printing: yes & no Printer header line 1: 16 alpha numeric characters Printer header line 2: 16 alpha numeric characters
q) The KANE9206 data storage based on EEPROM & flash technology and has a data
retention if the battery is removed of greater than 100 years
Page 59
59
Page 60
60
APPENDICES
A – PARAMETER MEANINGS
The parameters and their meanings are detailed as follows : -
DATE : Analyser date.
TIME : Analyser time.
MAIN BATTERY/ Displays the battery level from 0-100%. The analyser will flash HEATER BATTERY: RECHARGE BATTERY at less than 10 % of charge. The analyser
may show levels greater than 100% when the charger is connected.
----- : Displayed when a calculation cannot be performed because a probe is not fitted or a parameter is out of range.
FUEL : The fuel used in calculation of efficiency and carbon dioxide.
K1g: Gross calorific fuel constant. See Appendix for calculation.
K1n : Gross calorific fuel constant. See Appendix for calculation.
K2 : Percentage Maximum theoretical CO2 (dry basis).
K3: Percentage wet loss.
K4 : Percentage unburnt carbon loss.
O2r : Toxic gas measurements can be referenced to defined oxygen levels.
Oxygen referencing is required by some regulations such as TA­LUFT. If a reference value is selected the toxic gas measurements will be displayed with the symbol n attached to the units. i.e.ppmn What does oxygen reference mean ? If 3 % O2 reference is selected and 5 % O2 is measured in the flue then toxic gas values will be recalculated as if 3 % were measured. The equation for referencing is detailed in the Appendix. oxygen referencing prevents false readings being submitted, e.g. allowing more air into the boiler will increase the oxygen level in the flue and hence dilute any toxic gas reading. Oxygen referencing gives readings as if they were undiluted.
NETT : Nett temperature calculated by deducting the internal AMBIENT
temperature from the measured FLUE temperature. Displays in either Centigrade C or Fahrenheit F and will display NOT FITTED if flue probe not connected.
If an external INLET probe is used then INLET is deducted from
FLUE.
Page 61
61
O2 : Oxygen reading in percentage %.
CO : Carbon monoxide reading indicated in ppm or mg/m3. If the figures
are referenced to oxygen then the display will show ppmn or mg/m3n. Note with a high CO sensor fitted the reading will be displayed in percentage %.
EFF (G) : Combustion Efficiency calculation displayed in percentage. Gross (G)
or Net (N) can be set. The calculation is determined by fuel type see Appendix for calculation. The efficiency is displayed during a combustion test, 00.0 is displayed while in fresh air.
CO2: Carbon dioxide reading in percentage % when measured , not
calculated
CO2c: Carbon dioxide calculation determined by the type of fuel. This only
shows a reading when a combustion test is being carried out. Zero (0.0) is displayed while in fresh air.
FLUE : Temperature measured by flue gas probe in Centigrade or
Fahrenheit. Will show ambient temperature after fresh air calibration and N\F if probe disconnected.
INLET : Temperature measured by the optional inlet air probe or stored using
the Flue probe. The air probe is plugged into the instrument through the INLET socket. This figure is used to calculate the NET temperature instead of AMBIENT when fitted. Will show N\F if not fitted.
AMBIENT : Temperature measured by the internal sensor, used in the NET
temperature
CO/CO2 R : The CO/CO2 ratio, is the ratio of measured CO divided by CO2. It gives an indication of the following :­ How good a gas sample the instrument is reading.
How clean the boiler is running.
For example : A new or clean domestic boiler will display a ratio of
less than 0.004, a unit in need of cleaning 0.0040-0.0080 and a unit in need of major overhaul will show greater than 0.008.
This only shows a reading when a combustion test is being carried
out. 0.0000 is displayed while in fresh air.
P INDEX : The CO/CO2 ratio expressed as a percentage %, called the ‘Poison
Index” i.e. P INDEX % = 100 x CO/CO2. 0.00 is displayed while in fresh air.
XAIR % : Excess air calculated from the measured oxygen and type of fuel
used.
Displays reading during a combustion test +++ is displayed while in
fresh air.
Page 62
62
PRESSURE: Pressure reading. Units can be changed to different scales.
NO: Nitric oxide reading in ppm or mg/m3. Displayed when nitric oxide
sensor fitted. Reading can also be referenced to oxygen ppmn or mg/m3n.
NO2: Nitrogen dioxide reading in ppm or mg/m3. Displayed when nitrogen
dioxide sensor fitted. Reading can also be referenced to oxygen ppmn or mg/m3n.
NOx : Calculated total nitric oxides displayed in ppm or mg/m3. Reading can also be referenced to oxygen ppmn or mg/m3n.
SO2 : Sulphur dioxide reading in ppm or mg/m3. Displayed when sulphur
dioxide sensor fitted. Reading can also be referenced to oxygen ppmn or mg/m3n.
H2S: Hydrogen sulphide reading in ppm or mg/m3. Displayed when
Hydrogen sulphide fitted. Reading can also be referenced to oxygen ppmn or mg/m3n.
HC : Unburnt Hydrocarbon reading ppm of hexane, the sensor is
calibrated with hexane. Displayed when an infra red module is fitted.
Use equivalent factors for propane and methane.
LOSS : Total losses calculated from Combustion Theory. This is the
summation of the next three parameters.
DRY : Calculated heat lost in turning the carbon in the fuel to carbon ioxide
(CO2).
WET : Calculated heat lost in turning the hydrogen in the fuel into water
(H2O).
CO LOSS % : Calculated loss due to partially burnt carbon. Any carbon monoxide
(CO) in the flue has the potential to be turned into carbon dioxide and release more heat, hence this heat is lost up the flue.
GPS (Y): Latitude DDMM.MMM 5148.1060
GPS (X): Longitude DDDMM.MMM -00011.450
ATM: Atmospheric pressure in mbar
Page 63
63
B. NOx CALCULATIONS
ONLY AN NO SENSOR FITTED
WORKING IN PPM: NOX REFERENCED TO NO
The user can select the assumed NO2 percentage and the O2 normalised level
then: NOx in ppm = NO in ppm multiplied by (1 + assumed NO2 percentage)
in this setup NOx can only be displayed as NOx = NO
then normalising:
NO in ppmn = NO in ppm multiplied by (21 minus the O2norm setting) and then divided by (21 minus the actual O2 reading)
For a worked example assume:
NO is 1000ppm NO2 is 5% of NO O2norm is set to 3% actual O2 is zero
NOx in ppm = 1000 x (1 +5/100) =1000 x1.05 = 1050 ppm
NO ppmn = 1000 x (21 - 3)/(21-0) = 1000 x 18 / 21 = 857 ppmn
NOx ppmn = 1050 x 18 / 21 = 900 ppmn
or
NOx ppmn = 857 x 1.05 = 900 ppmn
WORKING IN mg/m3: NOX REFERENCED TO NO OR NO
2
The user can select the assumed NO2 percentage, the O2 reference level and whether the NOx reading is referenced to NO or NO2
WORKING IN mg/m3: REFERENCED TO NO
NO in mg/m3 = NO in ppm multiplied by 1.34
NOx in mg/m3 = NO in mg/m3 multiplied by (1 + assumed NO2 percentage)
Page 64
64
WORKING IN mg/m3: NOX REFERENCED TO NO2
NOx in mg/m3 = NO in ppm multiplied by 2.05 multiplied by (1 + assumed NO2 percentage)
or
NOx in mg/m3 = NO in mg/m3 divided by 1.34, multiplied by 2.05 and multiplied by (1 + assumed NO2 percentage)
NORMALISING READINGS
normalised reading = initial reading multiplied by (21 minus the O
2norm
setting) and then
divided by (21 minus the actual O2 reading)
BOTH NO AND NO2 SENSORS FITTED
WORKING IN PPM: NOX = NO + NO2
normalising readings
ppmn = initial reading in ppm multiplied by (21 minus the O
2norm
setting) and then divided
by (21 minus the actual O2 reading)
WORKING IN MG/M3
The user can select how the readings are referenced.
NOx = SUM NOx = NO NOx = NO
2
NOX = SUM
NOx in mg/m3 = NO in ppm multiplied by 1.34 plus NO2 in ppm multiplied by 2.05
NOX = NO
NOx in mg/m3 = (NO in ppm plus NO2 in ppm) multiplied by 1.34
NOX = NO2
NOx in mg/m3 = (NO in ppm plus NO2 in ppm) multiplied by 2.05
Page 65
65
NORMALISING READINGS
ppmn = initial reading in ppm multiplied by (21 minus the O
2norm
setting) and then divided
by (21 minus the actual O2 reading)
mg/m3n = initial reading in mg/m3 multiplied by (21 minus the O
2norm
setting) and then
divided by (21 minus the actual O2 reading)
ONLY AN NO2 SENSOR FITTED
When there is only an NO2 sensor fitted the NOx function is disabled
NO2 in mg/m3 = NO2 in ppm multiplied by 2.05
NORMALISING READINGS
ppmn = initial reading in ppm multiplied by (21 minus the O2norm setting) and then divided by (21 minus the actual O2 reading)
mg/m3n = initial reading in mg/m3 multiplied by (21 minus the O2norm setting) and then divided by (21 minus the actual O2 reading)
HIGH CO PURGE OPERATION
If there is a requirement to measure CO to concentrations above 10,000ppm then a High Purge module should be fitted (this comprises both a purge pump and a solenoid) in addition to the IR triple gas bench.
The CO measurement reading uses the electrochemical sensor’s reading from 0 to 4000ppm and the IR gas module takes over at 4000ppm to instruments upper limit.
When the electrochemical sensor’s reading passes 5500ppm, the solenoid operates and the high CO purge pump switches on and pumps fresh air across the electrochemical CO sensor. When the IR gas module reading goes below 5500ppm, the solenoid is deactivated and the high purge pump is stopped.
Page 66
66
C. COMBUSTION EFFICIENCY CALCULATION
The efficiency calculation is based upon British Standard BS845.
This identifies three sources of loss associated with fuel burning:
LOSSES DUE TO FLUE GASSES:
Dry Flue gas loss, Moisture and hydrogen Sensible heat of water vapour Unburned gas
LOSSES DUE TO REFUSE:
Combustible in ash Combustible in riddlings Combustible in dust
OTHER LOSSES:
Radiation Convection Conduction Other unmeasured losses
Net efficiency calculations assume that the energy contained in the water vapour (formed as a product of combustion and from wet fuel) is recovered and the wet loss term is zero. Gross efficiency calculations assume that the energy contained in the water vapour is not recovered.
Since the fuel air mixture is never consistent there is the possibility of unburned/partially unburned fuel passing through the flue. This is represented by the unburned carbon loss.
Losses due to combustible matter in ashes, riddlings, dust and grit, radiation, convection and conduction are not included.
EFFICIENCY CALCULATION:
Known Data - Fuel: Qgr = Gross Calorific Value (kJ/kg) Qnet = Net Calorific Value (kJ/kg) K1 = Constant based on Gross or Net Calorific Value: K1g = ( 255 x %carbon in fuel )/Qgr
K1n = ( 255 x %carbon in fuel )/Q
net K2 = % max theoretical CO2 (dry basis) K3 = % Wet Loss
Measured Data: Tf = Flue Temperature Ti = Inlet Temperature O2m = % Oxygen in flue gas
Calculated data: Tnet = Net Temperature % CO2 content in flue gas % Dry Flue Gas losses % Wet losses % Unburned carbon loss % Efficiency
Page 67
67
%CO2 = (20.9 - %O2m) x K2 / 20.9
Tnet = Flue Temperature - Inlet Temperature
Dry flue gas loss = 20.9 x K1n x (Tnet) / K2 x (20.9 - %O2m)
Wet loss = 9 x %H2 + %H2O / Qgr x [2488 + 2.1Tf - 4.2 Ti]
simplified = [(9 x %H2 + %H2O) / Qgr] x 2425 x [1 + 0.001 Tnet]
Wet loss = K3(1+0.001xTnet)
Where K3 = [(9 x %H2 + %H2O) / Qgr] x 2425
Net Efficiency = 100% - dry flue gas losses
= 100% - 20.9 x K1n x (Tnet) / K2 x (20.9 - % O2m)
Gross Efficiency = 100% - {dry flue gas losses + wet losses}
= 100% - [20.9 x K1g x (Tnet) / K2 x (20.9 - %O2m)] + [K3 x (1 + 0.001 x Tnett)]
Excess Air % = [(20.9% / (20.9% - 02m%)) – 1] x 100%
Air Index = 20.9% / (20.9% - 02m%)
NB: Either Excess Air or Air Index can be referred to as LAMBDA in the context of flue gas analysis dependent on local preferences.
For typical condensing gas boiler Excess Air = 31% and Air Index = 1.31
CO2% = [(20.9% - O2m%) x K2% / 20.9%]
Unburned fuel Loss = K4 x CO% / ( CO% + CO2% )
Where K4 = 70 for coke = 65 for anthracite = 63 for Bituminous coal = 62 for coal tar fuel = 48 for liquid petroleum fuel = 32 for natural gas
The formula for K4 is based on the gross calorific value Qgr. To obtain the loss based on net calorific value multiply by Qgr/Qnet. Since this loss is usually small this conversion has been ignored.
OXYGEN REFERENCE
CO(n) = CO x
(20.9 - O2r)
(20.9 - O2m)
Page 68
68
D. CALCULATION OF FUEL DATA
For any fuel not specified by Kane International the net calorific value, gross calorific value and composition should be obtained from the fuel supplier.
The following fuel data has been calculated with reference to the efficiency calculation.
Example 1:
Chemical composition: C 25% H2 3% H2O 50% Q
net
8.35 MJ/kg Qg 9.3 MJ/kg * Max CO2 20.4%
K1n = (255 x % carbon in fuel) / Q
net
(kJ/Kg)
= (255 x 25) / 8350 = 0.763
K1g = (255 x % carbon in fuel) / Qg (kJ/Kg) = (255 x 25) / 9300 = 0.685
K2 = Max % CO2 = 20.40
K3 = Wet Loss = [(9 x %H2 + %H2O) / 9300] x 2425
= [(9 x 3 + 50) / 9300] x 2425 = (77 / 9300) x 2425 = 20.08
K4 = 65 (an approximation for wood) *
The fuel values to program into the Analyser are as follows:
NATURAL GAS K1g K_2 K_4
:
0.763
: 20.4 : 65
K1n K_3 O2r
:
0.685
: 20.08 : 8.0
* Assumed values in the absence of supplied data. See previous appendix for other fuels.
Page 69
69
PRODUCT REGISTRATION
Please complete, detach and return to: Kane International Ltd Kane House, Swallowfield, Welwyn Garden City, Hertfordshire, AL7 1JG
Your Details
Name:
Job Title:
Company Name:
Company Address 1:
Address 2:
Town/City:
County:
Postcode:
Country:
Phone Number:
Fax Number:
Mobile Number:
Email Address:
Product Details
Note: Proof of Purchase may be required for warranty claims.
Date Purchased:
as numbers (28.01.14):
Purchased From:
Model Number:
KANE9206
Product Serial Number:
located on the rear product label beneath the protective rubber sleeve
Page 70
70
Why did you buy a Kane Product?
Made in the UK
Previous Owner
Value for Money
Our Fixed Price Servicing Programme
Kane Brand
Dealer Recommendation
Not your Decision
Other:
What brand was your previous analyser?
How did you hear about Kane?
Magazine Advert
Trade Counter
Training School
Previous Owner
Personal Recommendation
Internet Search
Exhibition
Other:
Your feedback is important to us, please add any additional comments
you would like to make with regard to your recent Kane purchase:
Thank you for completing this survey.
All the information we have collected is confidential.
We do not sell or share data with any other company or organisation
Page 71
71
Page 72
Thank you for reading this data sheet.
For pricing or for further information, please contact us at our UK Office, using the details