Kane 100-1 User Manual

KANE 100-1
Gas Analyzer and Logger
for
Ambient Carbon Monoxide
and
Ambient Carbon Dioxide
Kane International Ltd Kane House, Swallowfield Welwyn Garden City Hertfordshire, AL7 1JG
Tel: +44 (0) 1707 375550 Fax: +44 (0) 1707 393277 E-mail: sales@kane.co.uk www.kane.co.uk
Stock No: 18259 December 2005
© Kane International Ltd
The Kane 100-1 tests for ambient CO and ambient CO2 concentrations.
The large display shows 2 readings at a time and all data can be printed via an optional infrared printer. The printed data can be 'live' data or 'stored data'.
The KANE 100-1 is controlled using 4 buttons and a rotary dial.
The four buttons (from left to right)
switch the instrument on and off, print actual or saved data, switch the backlight on and off, switch the pump on and off, change the top line data display, save data or temporarily hold or ‘freeze’ the current reading.
The buttons with UP, DOWN and ENTER arrows also change settings such as date, time, when in menu mode.
nd
The rotary dial changes the display’s 2
line and selects access to the
menu to make changes to the date, time, fuel, etc.
The KANE 100-1 can store up to 255 sets of readings.
The Kane 100-1 is powered by internal re-chargeable NiMH batteries. The internal battery will power the analyser for about 6Hr, from full charge.
The KANE 100-1 has a protective rubber sleeve with a magnet for “hands–free” operation.
NOTE: Fresh air is referred to in this manual. Sometimes it may be convenient to use thin flexible tubing (min ID. 3mm.) to supply fresh outside air, if a tube is used ensure that the inlet is at least 1 meter away from any possible source of CO
such as an open window.
2
Fuel combustion of any sort may cause high and fluctuating levels of CO over surprisingly large areas. Exhaled breath can easily reach 10,000 to
2
20,000ppm.
Page 2
English 18259
1. BEFORE USING THE ANALYSER FOR THE FIRST TIME:
The analyser may be supplied with either or both of the following chargers. To fully charge the battery, you may use a Kane trickle charger, plugged into the base of the unit for 12-24 hours. Alternatively the Kane fast charger which plugs into the side will take about 1-2 hours, after which it will switch to low trickle charge. Do not fast charge the unit unless the battery charge indicator reads less than 70% otherwise over heating could cause an internal thermal cutout to operate. The unit will then not operate on its internal battery until it has cooled. Avoid leaving the unit on charge and switched off for more than 1 –2 days at a time or the internal battery life could be reduced.
Set the correct time, date etc, after it is switched on and calibrated – See USING
THE ROTARY DIAL below. These are stored when the analyser is switched off.
2. BEFORE USING THE ANALYSER EVERY TIME:
Check the water trap is empty and the particle filter is not dirty:
- To empty the water trap, unplug its rubber stopper and re-plug once it is empty.
- To change the filter, remove protective rubber sleeve, pull out the water trap , remove the water trap’s particle filter from the spigot and replace. Reconnect the water trap and rubber protective sleeve.
After switch on check that date and time are correct and battery power is sufficient.
Avoid exposing the instrument to sudden large temperature changes, and ensure that the unit reads zero for CO and approx. 400 ppm CO
in fresh air before use. If
2
necessary re-zero (see: Menu Selection).
Avoid very close proximity to radio transmitting devices, otherwise readings may be affected.
SAFETY WARNING
This analyser can be used to extract co mbustion gases that may be toxic in relatively low concentrations. These gases are exhausted from the back of the instrument. This
analyser must only be used in well-ventilated locations by trained and competent persons after due consideration of all the potential hazards.
Page 3
English 18259
3. USING THE ANALYSER AND ITS FOUR BUTTONS:
Switching ON the analyser
Switching OFF the analyser
Press the button to switch the unit ON in fresh
air. This allows the analyser’s sensors to be calibrated. This will set CO to zero and CO
may be set to 400 ppm if required.
2
At switch on, the analyser beeps four times and displays the model number. The bottom line counts down from typically 45 .
The Analyser will then indicate AIR ZERO NO?
Select NO? using the push button marked with an enter symbol, the
setting will not be altered, CO is always set to zero at this stage.
CO
2
Select YES only if you are in fresh air and wish to adjust the CO zero offset. Use the UP/DOWN buttons to set the required value (this should be 400ppm in fresh air), allow a few seconds after each button press for readings to settle, and then press ENTER to confirm.
After ENTER is pressed it will take about 4 seconds before the unit is ready to respond to the controls. The sensors are now ready to use. If the analyser will not calibrate, one or more of the sensors may need to be replaced or recalibrated by an authorized repair centre.
When countdown and zero is finished the display’s top line shows the last selected function and the bottom line displays whatever the rotary dial is turned to.
Press the button to switch the analyser OFF. The display counts down from 30 with the pump on to allow time for the sensors to be cleaned or purged with fresh air. If the probe is connected, make sure both analyser and probe are in fresh air.
Press HOLD / ENTER if you want to stop the count-down to off and return to making measurements.
2
Page 4
English 18259
Using UP / DOWN / ENTER Buttons
Enter key to Hold or Store readings
Switching on and off the backlight.
Once the “on” countdown has been completed:
Press the DOWN button for more than 1 second after the short tone or “beep” to change the display’s top line, if you hold this button down there is an “auto repeat” on this function . The Rotary Dial changes the display’s bottom line.
The top line display options are:
Time Date CO
2
CO TL ( -- or Logging interval in minutes if set) BAT %
For example, the analyser will display CO
on the top line and CO on
2
the bottom line if you turn the rotary dial to CO and hold down the DOWN button until CO
is displayed on the top line.
2
Use the UP / DOWN / ENTER keys to change settings (such as time and date ) when the rotary dial is turned to MENU.
In normal operation (not MENU) the ENTER key is used to either HOLD the readings for printing (a short button press of about 1 second release the button on the first beep, the display flashes when held) or to save readings if memory locations are available (a long button press … wait until after the second beep). The display will flash LOG xxx (or FULL) to give the store location of the saved data or will continue to flash if held until reset by a short button press. Note: the HOLD function is inhibited whilst AUTOSTORE is operating.
Press PRINT / BACKLIGHT for about 2 sec. and release to switch backlight on or off. Note: use of the backlight increases the drain on the battery.
Page 5
English 18259
Printing Data
Switching PUMP on / off
“Freezing” the display
Press print key quickly for about 1 sec. to start the analyser printing. The analyser displays "PRINTING" until this is completed.
Make sure the printer is switched on, ready to accept data and its infrared receiver is in line with the emitter on top of the analyser.
The analyser normally operates with the pump on. Press the PUMP key quickly to switch the pump off and on.
When the pump is switched off, the analyser displays "PUMP OFF" approx every 20 seconds.
Press HOLD / ENTER to freeze all readings. The display flashes and can be printed by pressing the PRINT key. Press HOLD / ENTER again for “live” measurements.
Note: the HOLD function is inhibited whilst AUTOSTORE is operating
Page 6
English 18259
Loading...
+ 13 hidden pages