Kaman SMU9200 User Manual

User Manual
SMU-9000
EC DECLARATION OF CONFORMITY
This apparatus, when installed and operated per the manufacturer’s recommendations, conforms with the protection requirements of EC Council Directive 89/336/EEC on the approximation of the laws of the member states relating to Electromagnetic Compatibility, as required by the Technical Construction File Route to Conformity.
The Certificate has been issued in accordance with the conditions of regulation No. 53 of the U.K. Statutory Instruments 1992 No: 2372 - The Electromagnetic Compatibility Regulations.
DO NOT MAKE ANY MODIFICATIONS TO CABLE LENGTH, SENSOR
OR CALIBRATED TARGET MATERIALS WITHOUT PRIOR
CONSULTATION WITH A KAMAN APPLICATION ENGINEER
Copyright © 2000 PART NO: 860127 Rev. E
********
********
PROPRIETARY INFORMATION
Any reproduction, disclosure or other use of this information is expressly prohibited except
as Kaman Precision Products may otherwise agree to by writing.
Part IX: Warranty
Kaman Precision Products Standard Limited Warranty
Products of Kaman Precision Products are warranted to be free from defects in materials and workmanship when installed and operated in accord with instructions outlined in the instruction manual.
Kaman Precision Products' obligation under this warranty shall be limited to repair or replacement (at the discretion of Kaman Precision Products of the defective goods returned to Kaman's plant within one (1) year from date of shipment. Extreme environment sensors are limited to the maximum operating temperature as specified within the most current Kaman Precision Products Extreme Environment Systems data sheets.
This warranty is valid except when the products have been subject to misuse, accident, negligent damage in transit or handling, or operation outside the conditions prescribed in the data sheet or instruction manual. This will be determined by Kaman Precision Products personnel.
In no event shall Kaman be liable for incidental or consequential damages, including commercial loss, resulting from any article sold under this Agreement.
In the event Buyer fails to limit to Kaman's warranty set forth above, any express or implied warranty Buyer may make with respect to any product of which any article sold thereunder is a component, Buyer shall indemnify and hold Kaman harmless from any and all liability, costs and expenses to which Kaman may be subjected as a result of Buyer's failure to so limit its express or implied warranties.
THIS WARRANTY IS EXCLUSIVE AND IS MADE IN LIEU OF ALL OTHER WARRANTIES; AND THOSE IMPLIED WARRANTIES, INCLUDING SPECIFICALLY THE WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY E X PRESSLY LIMITED TO ONE (1) YEAR DURATION.
NO MODIFICATION OR ALTERATION OF THE FOREGOING WARRANTY AND LIMITATION OR REMEDIES PROVISIONS SHALL BE VALID OR ENFORCEABLE UNLESS SET FORTH IN A WRITTEN AGREEMENT SIGNED BY KAMAN AND THE BUYER.
Kaman Precision Products Warranty No. 7A
SMU-9000 User Manual 23 ••• Part IX: Warranty

Part I: Welcome

Introduction

Ka SMU-9000 measuring systems use inductive (eddy current) technology to measure position without contacting the target. SMU-9000 measuring systems are ideally suited for applications requiring high resolution and economical performance-to-price ratio. Based on Kaman’s proprietary Pulse Width Modulated electronics, the SMU-9000 systems incorporate the latest surface-mount components with printed circuit board designs. The use of pulse width modulation techniques virtuall y eliminates the noise of traditional linear amplifiers, thereby “unmasking” position/motion sensing down into sub-atomic dimensions. This technology requires only simple, single ended supply voltage and low power circuitry.
SMU-9000 User Manual 1 ••• Part I: Welcome

Features Summary / Advantages

Extremely high resolution, to 1Å
Signal-to-noise ratio: 1 PPM
Precision of 0.1% of full scale, using computer correction.
Small size for efficient system integration: 2 channels in less
than 5.5 cubic inches (91cc)
Flexible packaging and performance.
Multiple configurations available: 9000 single channel, 9200
dual or balanced differential operation, and three and four channel models.
Wide range of sensor options.
Low power consumption: less than 40 mW per channel.

The SMU-9000

Figure 1. Single Channel SMU-9000.(Poly-carbonate Enclosure shown)
SMU-9000 User Manual 2 ••• Part I: Welcome

SMU-9000 Configurations

The SMU-9000 measuring system is available in several configurations. These configurations include: Singl e Channel, Dual Channel, Three Channel, and Four Channel systems shown in figures 2 through 4. The system electronics is housed in an enclosure made of die cast aluminum or poly-carbonate material. Sensors for the SMU-9000 come in a wide variety. The selection of a specific sensor is dependent upon the application in which the system is used.
Figure 2. Outline Drawing, Single and Dual Channel SMU-9000
SMU-9000 User Manual 3 ••• Part I: Welcome
Figure 3. Outline Drawing, 3 Channel SMU-9000
Figure 4. Outline Drawing, 4 Channel SMU-9000
SMU-9000 User Manual 4 ••• Part I: Welcome

Part II: Setting Up the System

Power Supply Requirements

The critical circuit voltages needed for operation of the SMU­9000 are internally regulated. Therefore, any single ended DC voltage power supply which can furnish from 6.5 to 30 volts at 20ma per channel may be used.
Note: Avoid using switching power supplies if high resolution (low noise) output is required.
The output amplifier runs directly from the supply voltage (positive) and an internally generated negative voltage. The system can be calibrated to have an output voltage swing anywhere from -5 volts to within 1.5 volts below the supply voltage. T ypi cal output voltage ranges and the r equired input supply voltages are as follows:
Output Voltage Range Required Input Supply Voltage
-5 Vdc to +5 Vdc Requires a minimum 6.5 Volt supply
0 Vdc to +10 Vdc Requires a minimum 11.5 Volt
supply
0.2 Vdc to 3.5 Vdc Requires a 5 Volt supply and special 3.5 Volt internal requlator to achieve minimum power drain of 40 mW per channel
-10 Vdc to +10 Vdc
balanced differential
Table 1. Output Voltage Range and Required Input Voltages for the SMU-9000
Requires differential SMU-9200 with ± 5 Volt output from each channel and a minimum 6.5 Volt supply