Kaise SK-4033, SK-4035 Instruction Manual

Output Terminal
Connecting to the external inspection
equipment is possible by using the
equipped photoMOS relay output terminal.
BENCH-TOP DIGITAL MULTIMETER
INSTRUCTION MANUAL
CONTENTS
1 1 2 2 3 4 4 6 6 7 9
9 11 12 13 14 14 15 15 16 17 17 18 18
FOR SAFETY MEASUREMENTS
1. UNPACKING AND INSPECTIONS
2. SPECIFICATIONS
2-1. General Specifications
2-2. Measurement Specifications
3. SAFETY PRECAUTIONS
3-1. Warnings
3-2. AC Adapter
3-3. General Warnings and Cautions
4. NAME ILLUSTRATION
5. MEASUREMENT PROCEDURES
5-1. Preparation for Use
5-2. Voltage Measurement
5-3. Resistance Measurement
5-4. Current Measurement
6. COMPARATOR FUNCTION
6-1. Comparator Settings
6-2. Buzzer Settings
6-3. Comparator Measurement
6-4. Comparator Relay Output
7. MAINTENANCE
7-1. Fuse Replacement
7-2. Periodical Check and Calibration
7-3. Repair
Thank you for purchasing KAISE MODEL
SK-4033/4035 BENCH-TOP DIGITAL
MULTIMETERS. To obtain the maximum
performance of this instrument, read this
Instruction Manual carefully, and take
safe measurement.
Features
Comparator Function
Useful for the quality check in the
production line. Comparator relay output is
also possible by using the output terminal.
True RMS
SK-4035 assures the accurate measurement
by True RMS.
422 Hayashinogo, Ueda City, Nagano Pref., 386-0156 Japan TEL : +81-268-35-1601 / FAX : +81-268-35-1603 E-mail : sales@kaise.com http://www.kaise.com
- 1 -
To prevent an electrical shock hazard to the operator and/or damage to the instruments, read this
instruction manual carefully before using the Insulation Tester. WARNINGS with the symbol
on the Insulation Tester and this instruction manual are highly important.
FOR SAFETY MEASUREMENTS
WARNING
Do not measure High Power Line (High Energy Circuits). High Power Line is
very dangerous and sometimes includes High Surge Voltage that could
cause explosive short in the instrument and could result in serious injury to
the operator. This instrument is for Low Power Line measurement. Even in
the Low Power Line, pay careful attention when measuring high voltage line.
The symbol listed in IEC 61010-1 and ISO 3864 means "Caution (refer to instruction manual)".
The symbol in this manual advises the user of an electrical shock hazard that could cause injury or material damages.
The symbol in this manual advises the user of an electrical shock hazard that could result in serious injury or even death.
WARNING
CAUTION
1. UNPACKING AND INSPECTIONS
Confirm that the following items are contained in the package in good condition. If there is any
damage or missing items, ask your local dealer for replacement.
1. Bench-top Digital Multimeter
2. Test Lead (100-57)
3. AC Adapter (897)
4. Spare Fuse (0.5A/250V)
5. Instruction Manual
1 pce.
1 set
1 pce.
1 pce.
1 pce
- 2 -
2. SPECIFICATIONS
1. DISPLAY (LCD)
(1) Main LCD
a. Numerical display 4000 count, maximum reading 4000, 21mm high b. Units and symbols
(2) Comparator setting LCD
a. Numerical display Maximum reading 3999, 7.5mm high
b. Units and symbols : Bz on, H, G, L
2. OPERATING PRINCIPLE Σ⊿ Conversion
3. AC MEASUREMENT SK-4033 : Average rectification, SK-4035 : True RMS (AC coupling)
4. RANGE SELECTION Auto range / manual range
5. FUNCTION SELECTION Manual (key operation)
6. POLARITY Auto ("" symbol is shown in minus)
7. OVERLOAD WARNING "OL" indication when exceeding the maximum reading see "4. Overload Warning" (p. 10) for details.
8. SAMPLING RATE 10 times/second
9. COMPARATOR
10. DIELECTRIC STRENGTH ±500V DC (between COM terminal and ground)
11. OPERATING TEMPERATURE & HUMIDITY
12. STORAGE TEMPERATURE & HUMIDITY
-20 to 60, 70RH or less in non-condensin
g
13. TEMPERATURE COEFFICIENT Accuracy at 23℃±5℃×0.2/
14. POWER SUPPLY AC adapter
15. CURRENT CONSUMPTION
2-1. General Specifications
AUTO, , , , Ω, kΩ, MΩ, mV, V, μA, mA, A,
and decimal point
a. Setting display LCD (sub-display) b. Setting method Key operation c. Setting range 3999 to 3999 d. Test method L lower limit G higher limit H e. Comparator display LED (L : red,G: green, H : red) f. Comparator relay output photoMOS relay
g. Buzzer OFFGOLOHIHI and LO are selectable
Loading voltage : 250V DC, 250V AC MAX. / ON resistance : 35Ω MAX.
Continuous loading current : 120mA MAX. Output terminal see "2. Dimensions of Recommended Solderless Terminal" (p. 16) for details.
Input : 100V to 240V AC, 50/60Hz Output : 9V DC Switching type
Approx. 18mA or less (normal), approx. 32mA or less (comparator),
approx. 57mA or less (comparator & buzzer)
0 to 35, 80RH or less in non-condensing 35 to 50, 70RH or less in non-condensing
- 3 -
16. FUSE
17. DIMENSIONS & WEIGHT 95(H)×200(W)×260(D)mm, approx. 1,140g
18. ACCESSORIES
19. OPTIONAL ACCESSORIES
2-2. Measurement Specifications
100-57 Test Lead, 897 AC Adapter, F31 Spare Fuse (0.5A/250V for μA/mA
terminal), Instruction Manual
660 AC/DC Clamp Adapter, 821 AC Clamp Adapter, 100-41 Test Lead Kit, 100-62 Test Lead Set, 940 Alligator Clip, 731 BNC Conversion Adapter
(23℃±5, <80RH in non-condensing)
1. DC Voltage (DC. V)
Range
Resolution
Accuracy
Input resistance
Maximum input Overload protection
Range selection : auto range / manual range
400.0mV
4.000V
40.00V
400.0V 600V
±0.3%rdg±2dgt
0.1mV 1mV 10mV 100mV 1V
600V DC
(for 10 sec.)
600V DC
600V DC, 450V AC rms for 1 min.
1000V DC
1000V AC rms
for 1 minute
2. AC Voltage (AC. V)
±1.0%rdg±5dgt
(50Hz to 500Hz)
SK-4033 : Average rectification / SK-4035 : True RMS
Range selection : auto range / manual range SK-4035 : Crest factor : 3:1 of the full scale (in 600V range, 3:1 at 300V or less)
Accuracy assured for the input of 5% or more of the full scale value.
100MΩ
11MΩ
10MΩ
Range
Resolution
Accuracy
Input resistance
Maximum input Overload protection
400.0mV
4.000V
40.00V
400.0V 600V
0.1mV 1mV 10mV 100mV 1V
600V AC
(for 10 sec.)
600V AC rms
600V DC, 450V AC rms for 1 min.
1000V DC
1000V AC rms
for 1 minute
100MΩ
11MΩ
10MΩ
3. DC Current (DC. A)
400.0μA
4.000mA
40.00mA
400.0mA
10.00A
0.05V0.25V0.1V0.6V0.5V
±0.75%rdg±2dgt
±1.5%rdg±2dgt
0.1μA 1μA 10μA 100μA 10mA
400μA DC
400mA DC
10A DC
0.5A/250V fuse
(input terminal)
3A/600V fuse (circuit)
10A/600V fuse (circuit)
Manual
Range
Resolution
Accuracy
Voltage drop
Maximum input
Overload protection
Range selection
10A range (p/n F30) : 10A/600V (φ6.3×32mm), μA/mA range (p/n F29) : 3A/600V (φ6.3× 32mm), power supply (p/n F31) : 0.5A/250V (φ5.2×20mm), μA/mA terminal (p/n F31) :
0.5A/250V (φ5.2×20mm)
- 4 -
SK-4033 : Average rectification / SK-4035 : True RMS
5. Resistance ( Ω )
400.0Ω
4.000kΩ
40.00kΩ
400.0kΩ
4.000MΩ
40.00MΩ
2.4V
1.2V
±0.3%rdg±3dgt
±0.3%rdg±2dgt
±1%rdg±2dgt ±3%rdg±4dgt
0.1Ω 1Ω 10Ω 100Ω 1kΩ 10kΩ
1.1mA110μA12.0μA1.2μA0.12μA0.12μA
400V DC
280V AC rms
for 1 minute
SK-4035 : Crest factor : 3:1 of the full scale
Accuracy assured for the input of 5% or more of the full scale value.
4. AC Current (AC. A)
400.0μA
4.000mA
40.00mA
400.0mA
10.00A
0.05V0.25V0.1V0.6V0.5V
±1.5%rdg±6dgt
(50Hz to 500Hz)
0.1μA 1μA 10μA 100μA 10mA
400μA AC rms
400mA AC rms
10A AC rms
Auto / Manual
0.5A/250V fuse
(input terminal)
3A/600V fuse (circuit)
10A/600V fuse (circuit)
Manual
Range
Resolution
Accuracy
Voltage drop
Maximum input Overload protection
Range selection
Range
Resolution
Accuracy
Max. open circuit voltage
Test current
Overload protection Range selection
3. SAFETY PRECAUTIONS
Correct knowledge of electric measurements is essential to avoid unexpected danger such as operator's injury or damage to the instrument. Read the following precautions carefully for safety measurements.
Before measurement, check if there are no damage to the instrument and the test leads. Dust, grease and moisture must be removed.
3-1. Warnings
WARNING 1. Checks of the Instrument and Test Leads
WARNING 2. Prohibition of High Power Line Measurement
Do not measure High Power Line (High Energy Circuits) such as Distribution Transformers, Bus Bars and Large Motors. High Power Line sometimes includes High Surge Voltage that could cause explosive short in the instrument and could result in shock hazard. Generally, shock hazard could occur when the current between the circuit, that involves more than 30V AC or 42.4V DC, and ground goes up to 0.5mA or more.
- 5 -
Even for Low Power Circuits of electric/electronic appliances, such as heating elements, small
motors, line cords and plugs, High Voltage Measurements are very dangerous. To avoid electric
shock hazard, pay careful attention not to touch any part of the circuit.
WARNING 3. Warning for High Voltage Measurement
WARNING 4. Warning for Dangerous Voltage Measurement
Do not hold the instrument in your hands.
Keep safety distance from the circuit to be measured and the test leads not to touch the
dangerous voltage.
Attach black and red alligator clips to the test lead pins.
Turn off the circuit to be measured when connecting the test leads.
After finishing the measurement, turn off the circuit to be measured again and discharge the
all capacitors. Then, detach alligator clips (test leads) from the circuit.
For dangerous high voltage measurement, strictly observe the warnings below (see Fig. 1).
In case of live-line measurement, strictly observe the warnings below (see Fig. 2) :
Do not hold the instrument in your hands.
Keep safety distance from the circuit to be measured not to touch the dangerous voltage.
Black test lead : Attach black alligator clip and connect to − (earth) side of the circuit.
Red test lead : Connect to (positive) side of the circuit.
Alligator Clips
Fig. 1
Red
Black
Fig. 2
Alligator Clip
Red
Black