Kahn Easidew I.S. User Manual

Easidew I.S.
Dew-Point Transmitter
User’s Manual
Kahn Instruments, Inc. 2015 This document is the property of Kahn Instruments and may not, either in part or whole, be copied or otherwise reproduced, communicated in any way to third parties nor stored in any data processing system, without the written authority of Kahn Instruments, Inc.
Easidew I.S.
For Kahn Instruments' contact information please go to
www.kahn.com
© 2015 Kahn Instruments
This document is the property of Kahn Instruments, Inc. and may not be copied or
otherwise reproduced, communicated in any way to third parties, nor stored in any Data
Processing System without the express written authorization of Kahn Instruments, Inc.
Easidew I.S. User’s Manual
Kahn Instruments 3
Contents
Safety ................................................................................................................................vi
Electrical Safety ...........................................................................................................vi
Pressure Safety ............................................................................................................vi
Toxic Materials .............................................................................................................vi
Repair and Maintenance ...............................................................................................vi
Calibration ...................................................................................................................vi
Safety Conformity ........................................................................................................vi
Abbreviations ..................................................................................................................... vii
Warnings ...........................................................................................................................vii
1 INTRODUCTION ................................................................................................1
1.1 Features ............................................................................................................ 1
2 INSTALLATION ..................................................................................................2
2.1 Unpacking the Instrument ................................................................................... 2
2.2 Preparation of the Sensor Cable .......................................................................... 3
2.3 Cable Connection ............................................................................................... 5
2.4 Electrical Schematic ............................................................................................ 5
2.4.1 Electrical Boundaries ..................................................................................... 6
2.5 Transmitter Mounting .......................................................................................... 6
2.5.1 Transmitter Mounting - Sample Block (Optional) .............................................. 7
2.5.2 Transmitter Mounting - Direct Pipeline Connection ........................................... 8
2.5.3 Transmitter Mounting - With Additional Process Connection Adapter ................ 9
3 OPERATION ....................................................................................................10
4 MAINTENANCE ................................................................................................11
4.1 O-Ring Replacement ......................................................................................... 12
Figures
Figure 1 Transmitter Unpacking Method ....................................................................2
Figure 2 Connector Terminal Block Removal ..............................................................3
Figure 3 Bare Wires .................................................................................................3
Figure 4 Crimped Wires ...........................................................................................3
Figure 5 Cut to 5mm ...............................................................................................4
Figure 6 Connection to Connector Terminal Block .......................................................4
Figure 7 Wiring Connections .....................................................................................4
Figure 8 Connector Installation .................................................................................5
Figure 9 2-Wire Connection Diagram .........................................................................5
Figure 10 Electrical Connections .................................................................................5
Figure 11 Maximum Load of Easidew - Including Cable Resistance ................................6
Figure 12 Transmitter Mounting ..................................................................................7
Figure 13 Transmitter Mounting - Pipe or Duct.............................................................8
Figure 14 Transmitter Mounting with Adapter .............................................................9
Figure 15 Installation Location .................................................................................10
Figure 16 Indication of Dead Space ..........................................................................10
Figure 17 Replacement of HDPE Guard .....................................................................11
Figure 18 Dimensions .............................................................................................15
Easidew I.S. User’s Manual
4 97099 Issue 13.4, August 2015
Appendices
Appendix A Technical Specications ............................................................................. 14
A.1 Dimensions .................................................................................15
Appendix B System Drawings ....................................................................................... 17
B.1 Baseefa Approved System Drawing ...............................................17
B.2 FM Approved System Drawing ......................................................18
B.3 CSA Approved System Drawing .....................................................19
Appendix C Hazardous Area Certication ......................................................................21
C.1 Product Standards .......................................................................21
C.2 Product Certication ....................................................................21
C.3 Global Certicates/Approvals ........................................................21
C.4 Terminal Parameters ....................................................................21
C.5 Special Conditions of Use .............................................................22
C.6 Maintenance and Installation ........................................................ 22
Appendix D EC Declaration of Conformity ...................................................................... 24
Easidew I.S. User’s Manual
Kahn Instruments 5
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. The user must not use this equipment for any other purpose than that stated. Do not apply values greater than the maximum value stated.
This manual contains operating and safety instructions, which must be followed to insure the safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage. Use competent personnel using good engineering practice for all procedures in this manual.
Electrical Safety
The instrument is designed to be completely safe when used with options and accessories supplied by the manufacturer for use with the instrument.
Pressure Safety
DO NOT permit pressures greater than the safe working pressure to be applied to the instrument.
The specied safe working pressure is 6500 psig (450 barg / 45 MPa). Refer to the Technical Specications in Appendix A.
Toxic Materials
The use of hazardous materials in the construction of this instrument has been minimized. During normal operation it is not possible for the user to come into contact with any hazardous substance which might be employed in the construction of the instrument. Care should, however, be exercised during maintenance and the disposal of certain parts.
Repair and Maintenance
The instrument must be maintained by Kahn Instruments, Inc.. Refer to www.kahn.com for contact details.
Calibration
The recommended calibration interval for this instrument is 12 months unless it is to be used in a mission-critical application or in a dirty or contaminated environment in which case the calibration interval should be reduced accordingly. The instrument should be returned to the manufacturer, Kahn Instruments Inc., or one of their accredited service agents for re-calibration.
Safety Conformity
This product meets the essential protection requirements of the relevant US and European standards
and directives. Further details of applied standards may be found in the Technical Specications in
Appendix A.
Easidew I.S. User’s Manual
6 97099 Issue 13.4, August 2015
Abbreviations
The following abbreviations are used in this manual:
barg pressure unit (=100 kP or 0.987 atm) (bar gauge) °C degrees Celsius °F degrees Fahrenheit DC direct current dp dew point fps feet per second ft-lbs feet per pounds g grams
lbs/in pounds per inch
µm micrometer
m/sec meters per second
mA milliampere max maximum mm millimetres MPa megapascal
Nl/min normal liters per minute
Nm Newton meter oz ounces ppm
V
parts per million by volume psig pounds per square inch RH relative humidity scfh standard cubic feet per hour V Volts
Ω Ohms
ø diameter “ inch(es)
Warnings
The following general warning listed below is applicable to this instrument. It is repeated in the text in the appropriate locations.
Where this hazard warning symbol appears in the following
sections it is used to indicate areas where potentially hazardous
operations need to be carried out.
Easidew I.S. User’s Manual
Kahn Instruments 1
INTRODUCTION
1 INTRODUCTION
The Easidew I.S. dew-point transmitter has been manufactured, tested and calibrated to the highest available standards and should be in perfect working order, ready for installation into a gas measurement application. If, after reading this manual, there are any questions about the instrument or how to install and operate it, please contact Kahn Instruments. Inc. Refer to www.kahn.com for contact details.
This manual covers the following Easidew I.S. (Intrinsically Safe) dew-point products:
Easidew I.S. with 5/8” UNF thread
Easidew I.S. with 3/4” UNF thread
Easidew I.S. with G 1/2” BSP thread
1.1 Features
The Easidew I.S. dew-point transmitter is a continuous, on-line, 4-20 mA transmitter for the measurement of dew-point temperature or moisture content in air and other
non-corrosive gases. It is designed specically for use within Zone 0, 1 and 2 hazardous
areas.
The key features are:
IECEx, CSA, FM, ATEX certied transmitter for use in hazardous areas
5/8” or 3/4” UNF, G1/2” BSP, process connection
Dew-point or ppmV moisture content
2-wire loop powered connection
Rugged 316 stainless steel IP66 construction
Measurement ranges -148 to +68°Fdp (-100 to +20°Cdp)
-166 to +68°Fdp (-110 to +20°Cdp)
Accuracy ±2°Cdp
Calibration Certicate (NPL, NIST)
Easidew I.S. User’s Manual
2 97099 Issue 13.4, August 2015
INSTALLATION
2 INSTALLATION
2.1 Unpacking the Instrument
On delivery, check that all the following standard components are in the packing tube:
Easidew I.S. Transmitter
5/8” UNF and G 1/2” BSP version - Bonded seal tted over threaded part of
transmitter body
OR
3/4” UNF version - Recessed sensor O-ring seal
Quantity 3 cable crimps
Sensor connector
Packing tube, and foam protectors (not needed for operation - save all the packing materials for the purpose of returning the instrument for re­calibration or any warranty claims)
Certicate of calibration
NOTE: If any component is not present in the packing box, contact Kahn Instruments immediately.
Unpack the dew-point transmitter packing tube as follows (see
Figure 1).
G 1/2BSP (black)
3/4” UNF (green)
5/8” UNF (blue)
98764 5321
Figure 1
Transmitter Unpacking Method
1. Unscrew the cap (1) from the packing tube (9).
2. Remove the bag containing 3 cable crimps (2).
3. Remove the foam block (4) containing the connector (3).
4. Pull out the transmitter (6) from the tube, complete with the two foam covers (7) and (8) and the red protective cap (5).
5. Remove the foam covers from the transmitter but leave the green, blue or black plastic protective cover (6) and the red cap (5) in place until ready for installation.
NOTE: The transmitter sensing element is protected while in transit by a black, green or blue cover containing a small desiccant capsule. The connection pins are protected by a red plastic cap. None of these plastic items are required for the operation of the transmitter.
NOTE: Keep the connector (3) in a safe place until the transmitter is ready for wiring.
Easidew I.S. User’s Manual
Kahn Instruments 3
INSTALLATION
2.2 Preparation of the Sensor Cable
The sensor cable is NOT supplied as standard. A cable can be obtained by contacting Kahn Instruments (see www.kahn.com for details).
The crimps supplied must be tted onto any cable installed
into the connector in order to comply with Hazardous Area
Certication of the product.
If making a cable assembly it is important that the cable is correctly terminated. See
Figures 3 to 6
.
Cable connection to the Easidew I.S. transmitter is made via the removable connector. Removing the central screw enables the connector terminal block to be removed from the outer housing by using a small screwdriver to pry it off.
O-ring
and washer
Figure 2
Connector Terminal Block Removal
Caution: When removing the central screw insure that the small
sealing O-ring and the washer are retained on the screw and
are present during re-installation.
NOTE:
Figure 3
to
Figure 6
shown below, should be followed in detail. The crimps should be applied such that there is no possibility of a conductor strand of a core becoming free (see
Figure 4)
.
Figure 3
Bare Wires
Figure 4
Crimped Wires
Easidew I.S. User’s Manual
4 97099 Issue 13.4, August 2015
INSTALLATION
When the crimp is made it should have a minimum of 2 positions of crimping. After the crimp is made it should be trimmed to a length of 5mm (see
Figure 5)
. When the crimps are installed into the connector terminal block insure they are fully inserted, as shown in
Figure 6
, before tightening the terminal clamping screw.
10
mm
1
2
3 4
Figure 5
Cut to 5mm
Figure 6
Connection to Connector Terminal Block
When all wire connections are made, insure that there is a minimum clearance distance
and a minimum creepage distance in air of 0.8” (2mm) between each terminal.
For the transmitter to work properly, and to achieve maximum performance, the sensor cable must be connected to the sensor connector as shown in the drawing below.
Note: The drawing below shows the identity of the connector terminals and wiring connections of the cable supplied by Kahn Instruments.
SCALE 2:1
SHORT
AS POSSIBLE
1
3
2
4
BLUE
BRAID
BRAID
GREEN - 4-20 mA
(SOURCE)
SCREEN
RED
+POWER
GND
VIEW ON REAR
OF CONNECTOR
GREEN
SIGNAL (SOURCE)
GREEN
YELLOW
BLUE
RED
BLUE - SCREEN RED + POWER
GREEN - 4-20 mA
BLUE - SCREEN
RED + POWER
GN
RD
BL
Figure 7
Wiring Connections
Always connect the 4-20 mA return signal to a suitable load
(see
Figure 7)
before the power is applied. Without this
connection, the transmitter may be damaged if allowed to
operate for prolonged periods.
Easidew I.S. User’s Manual
Kahn Instruments 5
INSTALLATION
2.3 Cable Connection
When installing the connector, and to insure that full ingress protection is achieved, the securing screw (with the O-ring and washer) must be tightened to a minimum torque setting of 3.4 Nm (2.5 ft-lbs). The sensor cable used must be a minimum diameter of
4.6mm (0.2”).
O-ring
and washer
Figure 8
Connector Installation
2.4 Electrical Schematic
NOTE: The screen/shield should be connected for maximum performance and to avoid interference.
3
1
Max Load 250R @ 12V 500R @ 24V
Supply 12V Min 28V Max
+
Screen/ Shield
Figure 9
2-Wire Connection Diagram
DEW-POINT TRANSMITTER CERTIFICATION No’s: Baseefa06ATEX0330X IECEx BAS 06.0090X
TRANSMITTER VERSION
TERMINAL NUMBER
EASIDEW I.S.
3
1
(+)
(+)
(-)
(RETURN)
HAZARDOUS AREA
SAFE AREA
GALVANIC ISOLATION INTERFACE
KFD2-STC4-Ex1 H
KFD0-CS-Ex2.50p
KFD2-CR-Ex1.20200
KFD2-CR-Ex1.30200
KFD0-CS-Ex1.50P
MTL5041
MTL5040
MTL5541
+
4-20 mA
LOAD
+VS (20 - 35 V DC)
VS -
-
Figure 10
Electrical Connections
Easidew I.S. User’s Manual
6 97099 Issue 13.4, August 2015
INSTALLATION
2.4.1 Electrical Boundaries
100
200
300
400
500
600
12 14 16 18 20 22 24 26 28
Resistance (ohms)
Supply Voltage
Figure 11
Maximum Load of Easidew - Including Cable Resistance
2.5 Transmitter Mounting
Prior to installation of the transmitter, unscrew and remove the black, green or blue plastic cover and retain for future use. Take care to prevent any contamination of the sensor before installation (handle the transmitter by the main body only,
avoiding contact with the sensor guard).
The Easidew I.S. can be mounted either into a ow-through sensor sampling block
(optional) or directly into a pipe or duct. It can be operated at pressures of up to 6500
psig (450 barg / 45 MPa) when tted with the bonded seal or O-ring provided.
The recommended gas ow rate, when mounted in the optional sampling block, is 2.1 to 10.6 scfh (1 to 5 Nl/min). However, for direct insertion applications, gas ow can be from static to 32.8 fps (10 m/sec).
NOTE: Pass the seal over the mounting thread and assemble into the sampling
location, by hand, using the wrench ats only. DO NOT grip and twist the
sensor cover when installing the sensor.
When installed, fully tighten using a wrench until the seal is fully compressed and to the following torque settings:
5/8” - 18 UNF 22.5 ft-lbs (30.5 Nm)
3/4” - 16 UNF 40 Nm (29.5 ft-lbs)
G 1/2” BSP 56 Nm (41.3 ft-lbs)
Easidew I.S. User’s Manual
Kahn Instruments 7
INSTALLATION
2.5.1 Transmitter Mounting - Sample Block (Optional)
The following procedure must be carried out by a qualied
installation engineer.
To mount the transmitter into the sensor block (preferred method), proceed as follows, refer to
Figure 12.
1. Insure that the green, blue or black protective cover (2), and its
desiccant capsule (2a), have been removed from the tip of the transmitter.
2. G 1/2” and 5/8” Versions - Insure that the bonded seal (2) is over the
threaded part of the transmitter body.
3/4” Version - Insure the O-ring is fully seated in the recess.
Under no circumstances should the sensor guard be handled
with the ngers.
3. Screw the transmitter (1) into the sample block (3) and tighten to the
appropriate torque setting (see Section 2.5). NOTE: Use the ats of
the hexagonal nut and not the sensor body.
4. Fit the transmitter cable/connector assembly to the plug located on the
base of the transmitter and tighten the securing screw (see Section 2.3).
1
2
3
2a
4
4
Figure 12
Transmitter Mounting
Easidew I.S. User’s Manual
8 97099 Issue 13.4, August 2015
INSTALLATION
2.5.2 Transmitter Mounting - Direct Pipeline Connection
The transmitter may be directly mounted into a pipe or duct as shown in
Figure 13.
Caution: Do not mount the transmitter too close to the bottom
of a bend where any condensate in the pipeline might collect
and saturate the probe.
The pipe or duct will require a thread to match the transmitter body thread. Mounting dimensions are shown in
Figure 13.
For circular pipework, to insure the integrity of
a gas tight seal, a mounting ange will be required on the pipe or tubing in order to provide a at surface to seal against.
The following procedure must be carried out by competent
personnel.
1. Insure that the protective cover (and its desiccant capsule) has
been removed from the tip of the transmitter.
WARNING: Under no circumstances should the sensor guard
be handled with the ngers.
2. G 1/2” and 5/8” Versions - Insure that the bonded seal (2) is over the
threaded part of the transmitter body.
3/4” Version - Insure the O-ring is fully seated in the recess.
3. Screw the transmitter (3) into the pipe (1). Tighten enough to obtain a
gas tight seal. NOTE: Do not overtighten or the thread on the pipe or tubing may be stripped.
14
15
16
17
18
18
20
21
22
23
2
3
4
5
6
7
8
9
10
11
(1.9”)
1
1
2
2
3
3
Optional
display
(available
on request)
Optional
cable
(available
on request)
48mm
Figure 13
Transmitter Mounting - Pipe or Duct
Easidew I.S. User’s Manual
Kahn Instruments 9
INSTALLATION
2.5.3 Transmitter Mounting - With Additional Process Connection Adapter
Applicable to the Easidew I.S. 5/8” Version ONLY
The following procedure must be carried out by a qualied
installation engineer.
To mount the adapter into the transmitter, proceed as follows (see
Figure 14)
:
1. Insure that the protective cover (2), and its desiccant capsule (2a), have been removed from the tip of the transmitter.
2. Place the bonded seal (3) over the threaded part of the transmitter body.
3. Screw the adapter (4) onto the threaded part of the transmitter and tighten to 22.5 ft-lbs (30.5 Nm). NOTE: Use the ats of the hexagonal nut and
not the sensor body.
WARNING: Under no circumstances should the sensor guard be
handled with the ngers.
4. Screw the transmitter (1) with its seal (3) and adapter (4) into the sample block (see Section 2.5.1 or see Section 2.5.2) and fully tighten using a wrench until the seal is fully compressed and to the following torque settings:
G 1/2” BSP 41.3 ft-lbs (56 Nm)
3/4” - 16 UNF ` 29.5 ft-lbs (40 Nm)
1/2” NPT Use a suitable sealant e.g. PTFE tape using
correct taping procedures
NOTE: Use the ats of the hexagonal nut and not the sensor body.
1
2
3
4
2a
Figure 14
Transmitter Mounting with Adapter
Easidew I.S. User’s Manual
10 97099 Issue 13.4, August 2015
OPERATION
3 OPERATION
Operation is very simple, assuming the following installation techniques are adhered to:
Sampling Hints
Be Sure the Sample is Representative of the Gas Under Test:
The sample point should be as close to the critical measurement point as possible. Also, never sample from the bottom of a pipe as entrained liquids may be drawn into the sensing element.
Figure 15
Installation Location
Minimize Dead Space in Sample Lines:
Dead space causes moisture entrapment points, increased system response times and measurement errors, as a result of the trapped moisture being released into the passing sample gas and causing an increase in partial vapor pressure.
Deadspace
Figure 16
Indication of Dead Space
Remove Any Particulate Matter or Oil from the Gas Sample:
Particulate matter at high velocity can damage the sensing element and similarly, at low velocity, they may ‘blind’ the sensing element and reduce its response speed. If particulate, such as degraded desiccant, pipe scale or rust is present in the sample gas,
use an in-line lter, as a minimum level of protection. For more demanding applications
Kahn Instruments offers a range of sampling systems (for more information contact www.kahn.com).
Use High Quality Sample Tube and Fittings:
Kahn Instruments recommends that, wherever possible, stainless steel tubing and
ttings should be used. This is particularly important at low dew points since other
materials have hygroscopic characteristics and adsorb moisture on the tube walls, slowing down response and, in extreme circumstances, giving false readings. For temporary applications, or where stainless steel tubing is not practical, use high quality thick walled PTFE tubing.
Position Transmitter away from Heat Source:
It is recommended, as good instrumentation practice, that the transmitter is placed
away from any heat source to avoid adsorption/desorption.
Easidew I.S. User’s Manual
Kahn Instruments 11
MAINTENANCE
4 MAINTENANCE
Calibration
Routine maintenance of the Easidew I.S. is conned to regular re-calibration by exposure
of the transmitter to sample gases of known moisture content to insure that the stated accuracy is maintained. Calibration services traceable to the US
National Institute of
Standards and Technology
(NIST) are provided by Kahn Instruments.
Kahn Instruments offers a variety of re-calibration and service exchange schemes to
suit specic needs. A Kahn representative can provide detailed, custom advice (refer to
www.kahn.com for details of Kahn Instruments’ contact information).
The following procedure must be carried out by a qualied
installation engineer.
Sensor Guard Replacement
The sensor is supplied with either a white HDPE or a stainless steel guard. The method of replacement is the same for both types.
HDPE Guard
The HDPE guard provides <10μm protection to the dew-point sensor. It is designed
to show any contamination and the guard should be changed if the surface becomes discolored.
When replacing the guard, care should be taken to handle the guard by the bottom part only. Replacement guards (EA2-HDPE) - pack of 10 - can be obtained by contacting Kahn Instruments (www.kahn.com) or your local distributor.
r
0
9
0
6
HANDLE,
USING
GLOVES, BY
BLACK PART
ON LY
Figure 17
Replacement of HDPE Guard
Stainless Steel Guard
The stainless steel guard provides <80μm protection to the dew-point sensor. It is
designed to show any contamination and the guard should be changed if the surface becomes discolored.
When replacing the guard, care should be taken to handle the guard by the bottom part only. A replacement guard (SSG) can be obtained by contacting Kahn Instruments (www.kahn.com).
Easidew I.S. User’s Manual
12 97099 Issue 13.4, August 2015
MAINTENANCE
Bonded Seal
If the installed bonded seal gets damaged or lost, a pack of 5 replacement bonded seals
(1/2-BS (for G 1/2 -BSP) or 5/8-BS (for 5/8” -18 UNF)) can be obtained by contacting
Kahn Instruments, or your local distributor.
4.1 O-Ring Replacement
If the installed O-ring gets damaged or lost, a pack of 5 replacement O-rings (3/4­OR (for 3/4” - 16 UNF)) can be obtained by contacting Kahn Instruments.
Do not touch the lter with bare hands
1. Identify the O-ring to be removed, as shown below.
BS116 (3/4” x 3/32”) viton,
75 shore
2. Carefully slide tweezers, thin bladed screwdriver or a blunt needle under the outer edge of the O-ring. NOTE: Take care not to scratch any of
the surfaces of the surrounding metal component.
3. Move the tool around the circumference to assist the extraction process.
Slide the O-ring clear of the thread and lter.
4. Make sure the groove has no scratches and is free from grease, dirt
or debris. Slide the new O-ring over the lter and thread and into the
groove. NOTE: Do not touch the lter with bare hands.
Easidew I.S. User’s Manual
Kahn Instruments 13
APPENDIX A
Appendix A
Technical Specications
Easidew I.S. User’s Manual
14 97099 Issue 13.4, August 2015
APPENDIX A
Appendix A Technical Specications
Performance
Measurement Range (dew point)
-148 to +68°Fdp (-100 to +20°Cdp)
-166 to +68°Fdp (-110 to +20°Cdp) Accuracy (dew point) ±3.6°Fdp (±2°Cdp) Repeatability 0.9°Fdp (0.5°Cdp) Response Time 5 mins to T95 (dry to wet) Calibration 13-point calibration with traceable 7-point calibration certicate
Electrical Specications
Output Signal
4-20 mA (2-wire connection current source)
User-congurable over range
Output Dew point or moisture content for ppm
V
Analog Output Scaled Range
Dew point: -148 to +68ºF (-100 to +20ºC) OR Moisture content in gas: 0 - 3000 ppm
V
Non-standard available upon request
Supply Voltage 12 to 28 V DC
Load Resistance
Max 250 Ω @ 12 V (500 Ω @ 24 V)
Current Consumption 20 mA max CE Marked Certied
Operating Specications
Operating Temperature -40 to +158ºF (-40 to +70ºC)
Operating Pressure
6500 psig (450 barg / 45 MPa) max Qualied over-pressure rating: (2 x operating pressure) 13053 psig (900 barg / 90 MPa)
Flow Rate
2.1 to 10.6 scfh (1 to 5 Nl/min) mounted in standard sampling block
0 to 32.8 fps (0 to 10 m/sec) direct insertion
Temperature Coefcient Temperature compensated across operating temperature range
Mechanical Specications
Ingress Protection
NEMA 4 in protection accordance with standard NEMA 250-2003 IP66 in accordance with standard BS EN60529:1992
Housing Material 316 stainless steel Dimensions Transmitter plus connector: L=5.19” x ø 1.77” (132mm x ø 45mm)
Sensor Guard
Standard: HDPE Guard < 10µm Optional: 316 stainless steel sintered guard < 80µm
Process Connection & Material
G 1/2” BSP; 3/4” - 16 UNF; 5/8” - 18 UNF
Material - 316 stainless steel Weight 5.29oz (150g) Interchangeability Fully interchangeable transmitter Electrical Connection Hirschmann GDS series (DIN 4350-C)
Diagnostic Conditions (factory programmed)
Condition
Sensor fault
Under-range dew point
Over-range dew point
Output
23 mA 4 mA 20 mA
Easidew I.S. User’s Manual
Kahn Instruments 15
APPENDIX A
Approved Galvanic Isolators
KFD2-CR-EX1.20200
KFD2-CR-EX1.30200
KFD0-CS-EX1.50P
KFD0-CS-EX2.50P KFD2-STC4-EX1.H MTL5041, MTL5040
Hazardous Area Certication
Certication Codes
ATEX II 1G Ex ia IIC T4 Ga (-20°C to +70°C)
IECEx Ex ia IIC T4 Ga (-20°C to +70°C)
TC TR Ex 0Ex ia IIC T4 Ga (-20°C to +70°C)
FM Class I, Division 1, Groups A B C D, T4
CSA Class I, Division 1, Groups A B C D, T4
A.1 Dimensions
132mm
(5.19”)
46mm (1.81”)
10mm (0.39”)
G1/2” BSP
10mm
(0.39”)
ø27mm
(1.06”)
SENSOR
132mm
(5.19”)
46mm (1.81”)
10mm (0.39”)
5/8” UNF
10mm
(0.39”)
ø27mm
(1.06”)
5/8” - 18 UNF Bonded Seal
ø25.4 x 2mm
(ø1 x 0.07”)
SENSOR
132mm
(5.19”)
46mm (1.81”)
10mm (0.39”)
10mm
(0.39”)
27mm (1.06”)
A/F
45mm (1.77”)
27mm (1.06”)
A/F
45mm (1.77”)
27mm (1.06”)
A/F
45mm (1.77”)
ø27mm
(1.06”)
3/4” UNF
3/4” - 16 UNF O-Ring
ø18.72 x 2.62mm
(ø0.75 x 0.09”)
SENSOR
G1/2” BSP Bonded Seal
ø28.65 x 2.61mm
(ø1.12 x 0.10”)
G1/2” Process Connection
3/4” Process Connection
5/8” Process Connection
Figure 18
Dimensions
Easidew I.S. User’s Manual
16 97099 Issue 13.4, August 2015
APPENDIX B
Appendix B
System Drawings
Easidew I.S. User’s Manual
Kahn Instruments 17
APPENDIX B
Appendix B System Drawings
B.1 Baseefa Approved System Drawing
Easidew I.S. User’s Manual
18 97099 Issue 13.4, August 2015
APPENDIX B
B.2 FM Approved System Drawing
10/03/06
MSB
Ex90385
SHEET 1 OF 1
NTS
SCALE
SIGN
A3
DATE
MOD. No.ISSUE
DRAWING NUMBER
TOLERANCES:
UNLESS OTHERWISE STATED
FINISH
DIMENSIONS:
MATERIAL
TITLE
USED ON
APPROVEDCHECKED
DATEDATE
DRAWN
DATE
MICHELL INSTRUMENT S LTD. 01/11/05 DOF03
mm
DRAWING
UNITS
THIS DOCUMENT IS THE PR OPERTY OF MICHELL INSTRUMENTS LTD.
AND MUST NOT BE COPIED NOR D ISCLOSED TO A THIRD PARTY
WITHOUT T HE CONSENT OF MICHELL INS TRUMENTS.
EASIDEW I.S. DEWPOINT TRANSMITTER
FM SYSTEM DRAWING
3rd ANGLE
PROJECTION
MICHELL INS TRUMENTS LTD. CAMBRIDGE ©
ANGLES: ±0.5°
HOLE Ø:
+0.1
-0.0
0 DEC. PLACE: ± 0.
5
1 DEC. PLACE: ± 0.2
2 DEC. PLACE: ± 0.1
01 CERT ISS 16/07/07 IMA
4 Inches
100mm
02 CERT ISS 23/12/08 IMA
03 CERT ISS 21/01/09 IMA
04 CERT ISS 24/03/09 IMA
05 11081 06/04/11 IMA
KFD2-CR-Ex1.20200
BAS 00 ATEX 7164
000
KFD2-CR-Ex1.30200
BAS 00 ATEX 7164
000
KFD0-CS-Ex1.50P
BAS 98 ATEX 7343
000
MTL5041
BAS 01 ATEX 7155
000
MTL5040
BAS 98 ATEX 2227
000
KFDO-CS-Ex2.50P
BAS98ATEX7343

HAZARDOUS LOCATION
CLASS 1, DIVISION 1, GROUPS A,B,C, & D
NON-HAZARDOUS LOCATION
+VS (20 TO 35V DC)
VS -
LOAD
4/20mA
+
-
(+)
(-)
(RETURN)
(+)
DEWPOINT
TRANSMITTER
CERTIFICATION No's:
Baseefa06ATEX0330X
IECEX BAS 06.0090X
3
1
TRANSMITTER VERSION
TERMINAL NUMBER
EASIDEW
I.S.
EASIDEW
PRO I.S.
2
4
3
THE CAPACITANCE AND EITHER THE INDUCTANCE OR THE INDUCTANCE TO RESISTANCE RATIO
(L/R) OF THE CABLE MUST NOT EXCEED THE FOLLOWING VALUES:
DCAB
33mH
12.6 mH
4.2mH
GROUP CAPACITANCE INDUCTANCE OR L/R RATIO
( F) ( mH) ( H/ohm)
THE ISOLATION OF THE SIGNAL WIRES WI TH THE EASIDEW DISCONNECTED, MUST BE ABLE TO
WITHSTAND A 500V AC INSULATION TEST.
THE INSTALLATION MUST COMPLY WITH THE INSTALLATION PRACTICES OF THE COUNTRY OF
USE, i.e. ANSI/ISARP12.6(INSTALLATION OF INTRINSICALLY SAFE SYSTEMS FOR HAZARDOUS
[CLASSIFIED] LOCATIONS) AND THE NATIONAL ELECTRICAL CODE ANSI/NFPA 70.
THE CAPACITANCE AND THE INDUCTANCE OF TEH HAZARDOUS AREA CABLES MUST NOT
EXCEED THE VALUES GIVEN IN TABLE 1.
2.11 F
613 nF
46 nF
435 H/
217 H/
54 H/
Easidew I.S. User’s Manual
Kahn Instruments 19
APPENDIX B
B.3 CSA Approved System Drawing
10/03/06
MSB
Ex90385CSA
SHEET 1 OF 1
NTS
SCALE
SIGN
A3
DATE
MOD. No.ISSUE
DRAWING NUMBER
TOLERANCES:
UNLESS OTHERWISE STATED
FINISH
DIMENSIONS:
MATERIAL
TITLE
USED ON
APPROVEDCHECKED
DATEDATE
DRAWN
DATE
MICHELL INSTRUMENTS LTD. 01/11/05 DOF0 3
mm
DRAWING
UNITS
THIS DOCUMENT IS THE PROPERTY OF MICHELL INSTRUMENTS LTD.
AND MUST NOT B E COPIED NOR DISCLO SED TO A THIRD PART Y
WITHOUT THE CONSENT OF MIC HELL INSTRUMENTS.
EASIDEW I.S. & EASIDEW PRO I.S.
DEWPOINT TRANSMITTER
SYSTEM DRAWIN. CSA
3rd ANGLE
PROJECTION
MICHELL INSTRUMENTS LTD. CAMBRIDGE ©
ANGLES: ±0.5°
HOLE Ø:
+0.1
-0.0
0 DEC. PLACE : ± 0.
5
1 DEC. PLACE: ± 0.2
2 DEC. PLACE: ± 0.1
02 CERT ISS 13/06/08 IMA
03 CERT ISS 16/06/08 IMA
4 Inches
100mm
04 CERT ISS 25/03/09 IMA
05 CERT ISS 15/06/09 IMA
06 11081 06/04/11 IMA
CSA Certified
Barrier
HAZARDOUS LOCATION
CLASS 1, DIVISION 1,
Gps A,B,C, & D
NON-HAZARDOUS LOCATION
S (20 TO 35V DC)
VS -
LOAD
4/20mA
+
-
(+)
(-)
EASIDEW I.S.
DEWPOINT
TRANSMITTE R
Vmax = 28V
Imax = 93mA
Pmax = 651mW
Ci = 37nf
Li = 0
(RETURN)
(+)
3
1
TRANSMITTER VERSION
TERMINAL NUMBER
EASIDEW
I.S.
EASIDEW
PRO I.S.
2
4
3
THE CAPACIT ANCE AND EITHER THE I NDUCTANCE OR THE IN
DUCTANCE TO RESISTANCE RATIO
(L/R) OF THE CABLE MUST NOT EXCEED THE FOLLOWING VA
LUES:
ABCD
4.2mH
12.6 mH
33mH
GROUP CAPACITANCE
INDUCTANCE OR L/R RATIO
(
F) (mH) ( H/ohm)
THE ISOLATION OF THE SIGNAL WIRES WITH THE EASIDEW
DISCONNECTED, MUST BE ABLE TO
WITHSTAND A 500V AC INSULATION TEST.
46 nF
613 nF
2.11 F
54 H/
217 H/
435 H/
THE INSTALLATION MUST COMPLY WITH THE INSTALLATION
PRACTICES OF THE COUNTRY OF
USE. i.e. ANSI/ISA RP12.6 (INSTALLATION OF INTRINSI
CALLY SAFE SYS TEMS FOR HAZARDOUS
[CLASSIFIED] LOCATIONS) AND THE NATIONAL ELECTRICAL
CODE ANSI/NFPA 70.
THE CAPACIT ANCE AND THE INDUCTANCE OF THE HAZARDOUS
AREA CABLES MUST NOT
EXCEED THE VALUES G IVEN IN TABLE 1
Intrinsically safe(entity), Class 1 , Div1, Group A,B,C,D
Hazardous Locat ion Installations
1) Control room equipment may not use or generate over 250Vrm
2) Wi re all circuits for pow er supply per CEC Part 1.
3) Use only e ntity approved safe ty barrier or other associ ated
equipment that satisfy the following conditions:
V
CG
<
V
max
, I
SC
<
I
MAX
, C
a
>
C
i
+ C
CABLE
, L
a
>
L
i
+ L
CABLE
Transmitter entity param eters are as follows:
V max < 2.8Vdc
I max < 93mA
Ci = 37nF
Li = 0uH
4) WARNING: SUBSTITUTION OF COMPONENTS MAY IMPARE
INTRINSIC SAFETY.
5) Ex ia is defined as Intrinsically Safe.
Type
Isolated Repeater
Dual Isolated
Repeater
Transmitter Supply
Isolator
Transmitter Supply
Isolator
Smart Transmitter
Power Supply
Certificate Number
BAS98ATEX7 343
UL Canada E106378CUL
BAS98ATEX7343
UL Canada
E106378CUL
BAS00ATEX7164
UL Canada E106378CUL
BAS00ATEX7164
UL Canada E106378CUL
BAS99ATEX7060
UL Canada E106378CUL
Interface
KFD0-CS-Ex1.50P
KFD0-CS-Ex2.50P
KFD2-CR-Ex1.20200
KFD2-CR-Ex1.30200
KFD2-STC4-Ex1.H
Connection t o Easidew I.S.
Pin 1 (+)
Pin 2 (-)
Channel 1 - Pin 1 (+)
Channel 1 - Pin 2 (-)
Channel 2 - Pin 4 (+)
Channel 2 - Pin 5 (-)
Pin 1 (+)
Pin 3 (-)
Pin 1 (+)
Pin 3 (-)
Pin 1 (+)
Pin 3 (-)
Easidew I.S. User’s Manual
20 97099 Issue 13.4, August 2015
APPENDIX C
Appendix C
Hazardous Area Certication
Easidew I.S. User’s Manual
97099 Issue 13.2, April 2015 21
APPENDIX C
Appendix C Hazardous Area Certication
The Easidew I.S is certied compliant to the ATEX Directive (94/9/EC), and IECEx for use within Zone 0, 1 and 2 hazardous areas and has been assessed so by Baseefa Ltd (Notied Body 1180).
The Easidew I.S is certied compliant to the North American Standards (USA and
Canada) for use within Class I, Division 1, Groups A, B, C & D hazardous locations and has been assessed so by CSA and FM.
C.1 Product Standards
This product conforms to the Standards:
EN60079-0:2012 IEC60079-0:2011 EN60079-11:2012 IEC60079-11:2011
FM Class 3600:1998 CAN/CSA-C22.2 No. 0-10 FM Class 3610:2007 CAN/CSA-C22.2 No. 157-92
FM Class 3810:2005 C22.2 No. 142-M1987
C.2 Product Certication
This product is attributed with the product certication codes:
ATEX & IECEx II 1G Ex ia IIC T4 Ga (-20°C to +70°C)
North American IS, Class I, Division 1, Groups A, B, C & D, T4
C.3 Global Certicates/Approvals
ATEX Baseefa 06ATEX0330X ATEX System Baseefa 07Y0027 IECEx IECEx BAS 06.0090X CSA 2013218 FM 3030238
TC TR Ex RU C-GB. ГБ05.B.00229
These certicates can be viewed or downloaded from our website at: http://www.kahn.com
C.4 Terminal Parameters
Ui
= 28 V
li
= 93 mA
Pi
= 651 mW
Ci
= 37 nF
Li
= 0
Easidew I.S. User’s Manual
22 97099 Issue 13.4, August 2015
APPENDIX C
C.5 Special Conditions of Use
1. The wiring connections to the free socket must be made via crimped connectors in such a way that all the strands of the wire used are held securely by the crimp.
2. The plastic plug and socket create a potential for electrostatic discharge so must not be rubbed with a dry cloth or cleaned with solvents.
3. The Easidew I.S. Dew-Point Transmitter does not withstand the 500 V AC insulation test to frame. This must be taken into account when installing the equipment.
C.6 Maintenance and Installation
The Easidew I.S. must only be installed by suitably qualied personnel and in accordance with the instructions provided and the terms of the applicable product certicates.
Maintenance and servicing of the product must only be carried out by suitably trained personnel or returned to Kahn Instruments.
Easidew I.S. User’s Manual
Kahn Instruments 23
APPENDIX D
Appendix D
EC Declaration of Conformity
Easidew I.S. User’s Manual
24 97099 Issue 13.4, August 2015
APPENDIX D
Appendix D EC Declaration of Conformity
http://www.kahn.com
Loading...