JUMO CTI-750 Operating Manual

Inductive
conductivity transmitter
in stainless steel housing
JUMO CTI-750
Type 202756
B 20.2756.0.1
Operating manual
Contents
1 Typographical conventions ...................................................... 8
1.1 Warning signs ..............................................................................................8
1.2 Attention-drawing signs ...............................................................................8
2 General ....................................................................................... 9
2.1 Foreword ......................................................................................................9
2.2 Structure of the JUMO CTI-750..................................................................10
3 Inductive conductivity measurement .................................... 11
3.1 Field of application .....................................................................................11
3.2 Function .....................................................................................................12
4 Identifying the device ............................................................. 13
4.1 Type nameplate .........................................................................................13
4.2 Type declaration ........................................................................................14
4.3 JUMO CTI-750 as "Head transmitter" ........................................................14
4.4 JUMO CTI-750 as "Transmitter with separate sensor" .............................15
5 Device description .................................................................. 17
5.1 Technical data, transmitter ......................................................................17
6 Assembly .................................................................................. 21
6.1 General ...................................................................................................... 21
6.2 Dimensions / Process connections, head transmitter ...............................22
6.3 JUMO CTI-750 with separate sensor ........................................................24
6.4 Assembly examples ...................................................................................28
7 Installation ............................................................................... 30
7.1 General ......................................................................................................31
8 Setup program ........................................................................ 34
8.1 Function .....................................................................................................34
9 Commissioning ....................................................................... 35
9.1 Head transmitter or transmitter with separate sensor ...............................35
9.2 Replacement sensor ..................................................................................35
Contents
10 Operation ................................................................................. 36
10.1 Operating elements ....................................................................................36
10.2 Principle of operation .................................................................................38
10.3 Principle .....................................................................................................40
10.4 Measurement mode ...................................................................................41
10.5 Operator level .............................................................................................41
10.6 Administrator level .....................................................................................49
10.7 Calibration level ..........................................................................................51
10.8 The desalination function ...........................................................................52
11 Calibrating ............................................................................... 56
11.1 General ...................................................................................................... 56
11.2 Calibration of the relative cell constant ......................................................56
11.3 Calibration of the temperature coefficient of the measurement solution ...57
12 Maintenance ............................................................................ 65
12.1 Cleaning the conductivity sensor ...............................................................65
13 Rectifying errors and faults ................................................... 66
13.1 Device inspection .......................................................................................66
14 Annex ....................................................................................... 72
14.1 Before configuring .................................................................................. 72
5
A
A/D-converter 17 Sediments 11, 65 Desalination reduction 53 DESALINATION FUNCTION 48 Desalination function 48, 52, 74 Desalination function: Start 48 Desalination function: Stop 48 Desalination 11 Desalination valve 52 REDUCTION 48 Reduction 74 DISTANCE 47 Distance 74 Alarm window 47 Analog output in case of "Alarm" 1819 Connection 34 Connection layout of the transmitter 33 Weld-on threaded connector28 Display 17 OUTPUT BINARY 46 Output binary 74 OUTPUT CONDUCTIVITY 44 Output conductivity 72 OUTPUT TEMPERATURE 45 Output temperature 73 Outputs 33 Output signal 18 Output signal, temperature 19 SWITCHOFF DELAY 48 Switchoff delay 74
B
Operating in levels 40 Operation principle 38 IN CASE OF ALARM 4445 In case of alarm 73 IN CASE OF ALARM / CALIB. 47 in case of alarm / at the time of calibration 74 IN CASE OF HOLD 47 in case of Hold 74 IN CASE OF CALIBRATION 44, 46
In case of calibration 73, 74
LIGHTING 49 Lighting 75 Calculation of a temperature coefficient 60 REFERENCE TEMPERATURE 42 Reference temperature 72 Binary inputs 33 Biocide 52 Burden 1819
C
CIP 11 CIP-Process 11
D
DOSING TIME 48 Dosing time 74 Pressure 20
E
Levels of the Administrator level 50 MOUNTING FACTOR 43
Mounting factor 21, 72
Installation position 21 Installation variants 22 INPUT BINARY 48 input binary 74 INPUT CONDUCTIVITY 42 Input conductivity 72 INPUT TEMPERATURE 45 Input temperature 73 UNIT 45 Unit 73 SWITCH-OFF DELAY48 Switch-off delay 74 Setting parameters 48 Electrical connection 17, 30 Electromagnetic compatibility 30 Removing the pigmenting 32 Determining the TC-curve 60 explosion-endangered areas 30
F
Error possibilities 66
FILTERING TIME 43, 45 Filtering time 72, 73 FUNCTION 46, 48 Function 20, 74
G
galvanic separation 34 Housing 17 Accuracy 1920 DEVICE DATA 49 Device data 75 Weight 17 LIMITING VALUE 47 Limiting value 74
H
MANUAL OPERATION 44, 4647
6
Manual operation 7374
MANUAL VALUE 44, 46
Manual value 7374 Date of manufacture 13 Attention-drawing signs 8 Hold function 48 HYSTERESIS 47
K
CALIBR.-INTERVAL 43 Calibration interval 72 Calibration timer 18 Characteristic 19 Compensation range 19 Configurable parameters 34 Configuring 72 CONTRAST 49 Contrast 75 CONCENTR. RANGE43 CONCENTR. MEASUREMENT 43 Concentration measurement 18, 72
L
LCD INVERSE 49 LCD inverse 75 Limit comparator 46 Solvent 65
M
MANUAL SPECIFICATION 45 Manual specification 73 Material 20 MEASUREMENT RANGE 42 Measurement range 72 MEASUREMENT RANGE 1 42 Measurement range, sensor 20 Measurement range / Temperature coefficient switching
48
Transmitter head 24 Measurement process 11 MEASUREMENT VALUE ACQUISITION 45 Measurement value acquisition 73 Installation location 21
N
natural 20 natural water 20 non-linear temperature coefficient 60
O
OFFSET 43, 45 Offset 7273
P
Parameters, configurable 34 Password 49 Polarization 11 PULSE DURATION 48 Pulse duration 74
R
Reference fluid 70 Reference measuring instrument 71 Reference temperature 19 REL. CELL CONSTANT 42 Relative cell constant 72
S
Breaking capacity of the semiconductor relay
17
System of protection 17 Sensor part 25 Service-Hotline 9 Setup interface 34
SAFETY VALUE 44, 46 Safety value 73, 74
SIGNAL TYPE 4445
Signal type 7273
SCALING START 4445 Scaling start 73 SCALING END 4445 Scaling end 73 Sunlight radiation 21
Voltage supply 17, 33
LANGUAGE 49 Language 75
T
Keyboard lock 48 TEMP. COEFFICIENT 42 TEMP. COMPENSATION 42 Temperature of the measurement medium 20 Temperature acquisition 19 Temperature coefficient 19, 72 Temperature compensation 72 Temperature compensation with the TC-curve
61
Temperature measurement range 19 TC-curve 61
U
Influence of ambient temperature 1819 reconfigure 72
7
V
LOCKING TIME 48 Locking time 74 Pre-calculated values 69
W
Warning signs 8 Resistant loop 67 Wiping contact 47
Z
CELL CONSTANT 42 Cell constant 72 permissible storage temperature 17 permissible ambient temperature 17
8
1 Typographical conventions
1.1 Warning signs
1.2 Indicative signs
Caution This sign is used if, by carefully following or not following instructions, harm to
persons can occur
Caution This sign is used if, by carefully following or not following instructions, dam-
age to devices or data can occur
Note
This sign is used when your attention is to be drawn to something special.
abc
1
Footnote
Footnotes are remarks that refer. to specific places in the text. Footnotes consist of two parts:
Flag in the text and footnote text. The flagging in the text is done by superscript serial numbers.
Action indication
This sign indicates that an activity to be carried out is being described. The individual work steps are marked by this star. Example: Loosen the crosshead screws..
9
2 General
2.1 Foreword
Please read the operating manual before you commission the device. Store the operating manual at a place that is accessible for all users at all times.
Please help us to improve this operating manual.
We would be grateful for your suggestions. Telephone(06 61) 60 03-7 14
Telefax (06 61) 60 03-6 05
All the required settings are described in this manual. Should there be any difficulties nonetheless at the time of commissioning, we request you not to carry out any impermissible manipulations. You could render your warranty entitlement null and void.
Please get in touch with the nearest branch or with the head office.
In case of technical queries Service-Hotline:
Telephone:(06 61) 60 03-3 00 or (06 61) 60 03-6 53 Telefax: (06 61) 60 03-88 13 00 or (06 61) 60 03-88 16 53 E-mail: Service@jumo.net
2 General
10
2.2 Structure of the JUMO CTI-750
Examples
(1) Transmitter (4) Inductive conductivity measuring
probe
(2) Process connection (5) with or without graphics LC-display
(3) Temperature sensor
Model:
Transmitter and
conductivity measuring
probe combined, type 202756/xx...
Model:
Transmitter w ith
separate sensor,
type 202756/xx...
(2)
(1)
(3)
(1)
(2)
(4)
(5)
(4)
(3)
11
3 Inductive conductivity measurement
3.1 Range of Applications
General The inductive measurement process allows a mostly maintenance-free acqui-
sition of the specific conductivity even in difficult medium conditions. In con­trast to the conductive measurement process, problems like electrode replacement and polarization do not occur.
Brief descrip­tion
The device is used for the measurement / control of the conductivity / concen­trations of liquid media. Using it is particularly recommended in media in which significant deposits from carried dirt, oil grease or of lime and gypsum are expected. The integrated temperature measurement makes exact and fast temperature compensation possible, which is particularly important for the measurement of the conductivity. Additional functions such as the combined switching of the measurement range and temperature coefficient make possi­ble the optimum use in case of CIP-processes.
Two integrated switching outputs can be freely programmed for limit value monitoring or conductivity / concentration and /or temperature. In addition, alarm and control tasks (desalination) can be assigned.
Operation is either via a membrane keyboard and a plain text graphical display (user language can be changed) or via a comfortable PC-Setup program. By simply turning the housing cover, reading the display is possible both in case of installation in vertical or horizontal pipes. By means of the Setup program, the device configuration data can also be saved and printed for plant docu­mentation purposes. To prevent manipulation, the device can also be supplied without a keyboard / display. In this case, the Setup program is required for programming.
The JUMO CTI-750 can be supplied as a combine device (transmitter and measuring cell in one device) or as a shouldered version (transmitter and mea­suring cell connected by cables). The separate version is particularly suitable for plants with intense vibrations and/or intense temperature radiation at the measurement location or for installation at not easily accessible places.
Typical usage fields
- CIP-cleaning (CIP = Clean In Place / Process)
- Concentration monitoring or chemical dosing
- Foodstuffs beverages and pharmaceutical industries
-Product monitoring (phase separation, Product / Product mixture / Water) in the beverage industry, breweries, dairies
- Control (e.g. phase separation of cleaners / rinsing water of cleaning processes e.g.. bottle cleaning plants and in case of container cleaning)
3 Inductive conductivity measurement
12
3.2 Function
of the transmit­ter
The JUMO CTI-750 transmitter is conceived for use at the location. A robust housing protects the electronics and the electrical connections from aggres­sive ambient influences (system of protection IP 67). In the standard version, the device has one analog signal input for the conductivity / concentration and the temperature. The further processing of the standard signals can take place in suitable display / control devices or, e.g. directly in a PLC.
The output signals are galvanically separated from one another and from the measurement medium.
of the measur­ing cell
The measurement of the conductivity takes place with an inductive probe. A sinusoidal AC is supplied to the transmitting coil. Depending on the conductiv­ity of the liquid to be measured, a current is induced in the receiver coil. The current is proportional to the conductivity of the medium. The cell constant of the inductive probe is geometry-dependent. In addition, the cell constant can be influenced by parts in its immediate vicinity.
(1) Plastic body (4) Measurement medium (2) T-shaped flow channel (5) Temperature sensor Pt100 (3) Fluid loop
(2)
(1)
(3)
(4)
(5)
13
4 Identifying the device
4.1 Type nameplate
on the transmitter
on the connect­ing cable (only in case of separate sen­sor)
JUMO GmbH & Co. KG
Fulda, Germany www.jumo.net
JUMO CTI-750 Typ: 202756/16-607-0000-82/000 VARTN: 20/00445843 F-Nr.: 00909467 01 0 0745 0001
DC 19...31 V 3W
F-Nr.: 00909467 01 0 0517 0001
In the case of devices with a separate sensor (base type supplementation (2)
66), for every instrument, the transmitter and the separate sensor are matched
to one another at the factory. When connecting the components, ensure that the production number of the external sensor (on the flag tag on the connecting cable) is identical to the pro­duction number of the transmitter (on the nameplate).
The date of manufacture is encrypted in the "F-No." : 0517 means year of manufacture 2005 / calendar week 17
4 Identifying the device
14
4.2 Type explanation
4.3 JUMO CTI-750 as "Head-mounted transmitter"
1
Only in conjunction with additional code 767 (measuring cell material PEEK).
2
If required, M12-plug / -sockets, additional code 580 should be ordered.
3
Specify additional codes in succession separated by a comma.
4
Assembly material (mounting brackets) not part of the scope of supply.
(1) Basic type
202756 JUMO CTI-750
Inductive transmitter / switching device for conductivity / concentration and temperature
(2) Basic type supplementation
16 Head-mounted transmitter in stainless steel housing
with display / keyboard
(3) Process connection
o 107 Stud thread G1 1/4A o 108 Stud thread G1 1/2A o 110 Stud thread G2A o 606 Screwed pipe connection DN40, DIN 11 851
(MK DN50, milk cone)
o 607 Screwed pipe connection DN50, DIN 11 851
(MK DN40, milk cone)
o 608 Screwed pipe connection DN65, DIN 11 851
(MK DN65, milk cone)
o 609 Screwed pipe connection DN80, DIN 11 851
(MK DN80, milk cone)
o 617 Clamp 2 1/2"
4
o 686 VARIVENT® DN40/50
1
o 690 SMS 2"
(4) Immersion length
o 000 see "Dimensions / Process connections (head-
mounted transmitter"
(5) Electrical connection
o82Cable glands o 83 M12-plug / -sockets
(instead of the cable glands)
2
o 84 Two cable glands M16 and a dummy plug
(6) Extra codes
3
000 None o 268 Temperature sensor inside X 767 Measuring cell material PEEK X 768 Measuring cell material PVDF o 844 Voltage supply AC 24 V
(1) (2) (3) (4) (5) (6)
Ordering key / - - - / , ... Ordering example 202756 / 16 - 607 - 000 - 82 / 000
15
4 Identifying the device
4.4 JUMO CTI-750 as "Transmitter with separate sensor"
(1) Basic type
202756 JUMO CTI-750
Inductive transmitter / switching device for conductivity / concentration and temperature
(2) Basic type supplementation
26 Transmitter in stainless steel housing, with display / keyboard
(without sensor)
5
66 Transmitter in stainless steel housing, with display / keyboard includ-
ing sensor (cable length 10 m)
85 Replacement sensor with 10 m cable
1, 5
for transmitter in stainless steel housing
(3) Process connection
X 000 not available o o o 107 Stud thread G1 1/4A o o o 108 Stud thread G1 1/2A o o o 110 Stud thread G2A o o o 606 Screwed pipe connection DN40, DIN 11 851 (MK DN40, milk cone) o o o 607 Screwed pipe connection DN50, DIN 11 851 (MK DN50, milk cone) o o o 607 Screwed pipe connection DN65, DIN 11 851 (MK DN65, milk cone) o o o 607 Screwed pipe connection DN80, DIN 11 851 (MK DN80, milk cone) o o o 617 Clamp 2 1/2"
1
ooo 686 VARIVENT® DN40/50
2
o o o 690 SMS 2"
5
o o 706 Submerged version
6
(4) Immersion length
X 0000 see "Dimensions / Process connections (separate sensor)" X o o 0500 0500 mm X o o 0000 1000 mm X o o 0000 1500 mm X o o 0000 2000 mm X o o xxxx Special length (in grid of 250 mm; e.g. 0250; 0750; 1250; 1750)
(5) Electrical connection
X 21 Fixed cable with M12-socket at separate sensor o o 82 Cable glands on operating part o o 83 M12-plug / sockets on operating part
3
o o 84 Two cable glands M16 and a dummy plug
(6) Extra codes
4
X X X 000 no extra code
o o 268 Temperature sensor inside o o 580 1 set M12-plug / sockets X X X 767 Measuring cell material PEEK X X X 768 Measuring cell material PVDF o o 844 Voltage supply AC 24 V
4 Identifying the device
16
1
Assembly material (Union / grooved nut, mounting bracket) not part of scope of supply. If required, please order as well, see accessories.
2
Only in conjunction with additional code 767 (measuring cell material PEEK).
3
If required, M12-plug / -sockets, additional code 580 should be ordered.
4
Specify additional codes in succession separated by a comma.
5
An equalizing set is indispensable for commissioning. If not available, please order as well, see Accessories.
6
Only in conjunction with additional code 768 (measuring cell material PVDF).
(1) (2) (3) (4) (5) (6)
Ordering key / - - - / , ... Ordering example 202756 / 66 - 607 - 1000 - 82 / 767
17
5 Device description
5.1 Technical data, transmitter
5.1.1 General
A/D-converter Resolution: 15-bit
sampling time:500 ms = 2 measurements/s
Voltage supply For operation on SELF- or PELV-circuits
Standard production feature: DC 19…31 V (nominal DC 24 V), the device has reverse polarity protection
Ripple: < 5% Power drawn with display: 3 W Power drawn without display: 2.6 W
In case of extra code 844: AC 24 V ±10%
Breaking capacity of the semiconductor relay U< 50 V AC/DC
I 200 mA
electrical con­nection
Threaded plug terminals 2.5 mm
2
or M12-plug/ sockets
Display (option) Graphic-LCD with background lighting; Contrast adjustable
dimensions: 62 x 23 mm
permissible ambient tem­perature (transmitter)
-5…+50°C max. 93% rel. atmospheric humidity, without thawing
permissible storage temper­ature (transmitter)
-10…+65°C max. 93% rel. atmospheric humidity, without thawing
System of pro­tection (transmitter)
IP 67
Housing Stainless steel, material no. 1.4305
Weight depending on the model and process connection
about 2.4 kg
5 Device description
18
5.1.2 Conductivity-/ Concentration transmitter
Concentration measurement
(implemented in the device soft­ware)
- NaOH (sodium hydroxide solution)
0...15 % by weight or 25...50 % by weight
- HNO
3
(nitric acid); pay attention to the chemical resistance of the sensor!
0...25 % by weight or 36...82 .% by weight
- Customer-specific concentration curve freely programable via the setup program (see "Special functions")
Calibration timer
adjustable: 0...999 days (0 = off)
Output ­signal conductivity / concentration
0...10 V / 10...0 V
2...10 V / 10...2 V
0...20 mA / 20...0 mA
4...20 mA / 20...4 mA
The output signal is freely scalable.
Load 500 Ω for current output
2kΩ for voltage output
Influence of ambient tem­perature
0.1%/K
Analog output in case of "Alarm"
Low (0 mA / 0 V / 3.4 mA / 1,4 V) or High (22.0 mA / 10,7 V) or a fixed value that can be set (safety value)
Measurement ranges
Four measurement ranges can be selected. One of these measurement ranges can be activated via an external switch or a PLC .
Measurement ranges, transmitter
Accuracy
(in % of the start -
of measurement range)
0...500 µS/cm
0.5%
0...1000 µS/cm
0...2000 µS/cm
0...5000 µS/cm
0...10 mS/cm
0...20 mS/cm
0...50 mS/cm
0...100 mS/cm
0...200 mS/cm
0...500 mS/cm
0...1000 mS/cm
0...2000 mS/cm
1
19
5 Device description
1
not temperature-compensated
Note: The overall accuracy is formed from the accuracy of the transmitter + the accuracy of the sensor.
5.1.3 Temperature transmitter
Temperature acquisition
manual -20.0...25.0...150°C / °F or automatic
Temperature measurement range
-20...150°C / °F
Characteristic linear
Accuracy 0.5% of the measurement range
Influence of ambient tem­perature
0.1%/K
Output -signal temperature
0...10 V / 10...0 V
2...10 V / 10...2 V
0...20 mA / 20...0 mA
4...20 mA / 20...4 mA
The output signal is freely scalable.
Load 500 Ω for current output
2 kΩ for voltage output
Analog output in case of "Alarm"
Low (0 mA / 0 V / 3.4 mA / 1,4 V) or High (22.0 mA / 10,7 V) or a fixed value that can be set (safety value)
5.1.4 Temperature compensation
Reference tem­perature
15...30°C adjustable
Temperature coefficient
0.0...5.5 %/K adjustable
5 Device description
20
Compensation range
-20...150°C
Function - Linear compensation (constant temperature coefficient).
This type of compensation can be applied with acceptable for many normal types of water. The temperature coefficient used is then about 2.2 %/K.
- Natural water (DIN EN27888 or ISO 7888). In this case, a so-called non-linear temperature compensation is applied. According to the above standard, the corresponding type of compensation can be applied in the case of natural ground waters, spring waters and above-ground waters. The conductivity of the water is compensated in the range from 0°C to 36°C.
- Not linear (for learning function, see special functions) Here, the actual flow of the temperature coefficient during a heating-up or cooling process is determined with the transmitter.
5.1.5 Sensor
Material PEEK (Polyetheretherketone) or
PVDF (Polyvenylidenfluoride)
Note!:
Temperature, pressure and measurement medium influence the life of the measuring cell
Temperature of the measure­ment medium
-10 ... +120°C
short-term 140°C (sterilization)
Pressure max. 10 bar
1
not temperature-compensated
Measurement range Sensor
Accuracy
(in % of the start -
of measurement range)
0...500 µS/cm 1%
0...1000 µS/cm
0...2000 µS/cm
0.5%
0...5000 µS/cm
0...10 mS/cm
0...20 mS/cm
0...50 mS/cm
0...100 mS/cm
0...200 mS/cm
0...500 mS/cm
0...1000 mS/cm 1%
0...2000 mS/cm
1
21
6 Installation
6.1 General
Installation location
The compatibility of the material with the measurement medium must be checked by the customer.
Ensure easy accessibility for subsequent calibration. The fastening must be secure and low-vibration. Avoid direct sunlight. It is necessary to pay attention to a good flow through and around the sensor
(1). When installing in a pipe, a minimum distance of 20 mm must be maintained
from the sensor to the pipe wall. If these minimum distances cannot be maintained, the parameter "Mounting factor" can be used to achieve limited equalization.
In immersion operation in a basin, an installation location that is representative for the typical conductivity or concentration must be provided.
Mounting loca­tion
The JUMO CTI-750 can be mounted in any position.
Screwing in and unscrewing the separate sensor
No cable twisting should occur.
Avoid tensile forces on the cable, especially jerky pulling.
Sensor details
(1)
Stainless steel sleeve: Material
1.4571
Sealing: FPM
6 Installation
22
6.2 Dimensions / Process connections, head transmitter
Installation vari­ants
Process connection (1.4301)
71
130
207
40
84
177
130
71
56
37
Process connection (1.4301)
71
130
224
20
60
104
G
Model with
process connection 607
MK DN50
Model with
process connection 617
Clamp 21/2"
(mounting bracket not part
of scope of supply)
Process connection
Slotted union nut F50 (1.4301)
71
84
40
212
130
Model with process connection
107 = Stud thread G1 1/4A 108 = Stud thread G1 1/2A
110 = Stud thread G2A
Version with
process connection 686
VARIVENT® DN50
(mounting bracket not
in scope of supply)
23
6 Installation
Process connection (1.4301)
Ring nut SMS 2" (1.4301)
130
71
215
84
40
Model with
process connection 690
SMS 2"
6 Installation
24
6.3 JUMO CTI-750 with separate sensor
Tran sm it te r­head
Wall fastening
ø110
84
71
176
100
91
120
ø5.5
120
100
25
6 Installation
Sensor part
Process connection (1.4301)
M12 socket (PBT / PA)
Cable gland M16 (PA)
M12 socket (PBT / PA)
Cable gland M16 (PA)
M12 socket (PBT / PA)
Cable gland M16 (PA)
Process connection 686 VARIVENT DN50 (1.4301)
®
M12 socket (PBT / PA)
Cable gland M16 (PA)
Shouldered version with process connection 607
MK DN50
(union nut not part
of scope of supply)
Shouldered version with process connection 686
VARIVENT
®
DN50
(mounting bracket not
in scope of supply)
Shouldered version with process connection 617
Clamp 21/2"
(mounting bracket not part
of scope of supply)
Shouldered version with
process connection 107 = Stud thread G1 1/4A 108 = Stud thread G1 1/2A
110 = Stud thread G2A
6 Installation
26
M12 socket (PBT / PA)
Cable gland M16 (PA)
Process connection (1.4301)
Shouldered version with process connection 690
SMS 2"
(union nut not part
of scope of supply)
27
6 Installation
6.3.1 Separate sensor as submerged version
M12 socket (PBT/PA)
Cable gland System of protection IP68 up to 0,2 bar) according to EN 60529 (PA)
110
Submerged pipe (PVDF)
ø40
EL = Submerged length
44
Shouldered version with process connection 706
Immersion version
(pipe clamps not part of the scope of supply)
57
18
3
100
+5
0
Fixing screw
Flange mov
a
ø
100
ø
165
ø
18
Fixing screw
Flange movable
60
18
3
100
+5
0
ø
18
ø
125
ø
165
Optional accessories:
Flange DN 32, Sales no.: 20/00083375
Optional accessories:
Flange DN 50, Sales no.: 20/00083376
6 Installation
28
6.4 Installation examples
Process connection 607 Screwed pipe fitting DN50, DIN 11 851 (MK DN50, Milk cone) (1.4301)
Ring nut DN50 (1.4301)
Threded pipe adapter DIN. S, DN50 (Edelstahl)
Reducing Tee DIN, short SSS DN65/50 (e.g. 1.4301) (to be provided by the plant operator supplied bynot JUMO)
Tee DIN 11 852, SSS DN50 Dim. A shortened to 30 mm
Process sonnection 607 Screwed pipe fitting DN50 DIN11 851 (MK DN50, Milk cone) (1.4301)
Flow direction
Process connection 617 Clamp 21/2" (1.4301)
Clamping ring (e.g. 1.4301)
Threaded pipe adapter, S DN2,5" (stainless steel)
Reducing Tee DIN, short SSS DN65/50 (e.g. 1.4301) (to be provided by the plant operator supplied bynot JUMO)
Weld-on threaded connector
DN50, DIN 11 851
(counter-piece to process connection
-607)
Sales no.: 20/00085020
29
6 Installation
Tee DIN, short SSS DN65/50 (e.g. 1.4301) (to be provided by the plant operator supplied bynot JUMO)
30
7 Installation
When selecting the conductor material for the installation and for the elec-
trical connections of the device, compliance is required with the specifica­tions of VDE 0100 "Regulations governing the installation of heavy-current systems with rated voltages below 1000 V" or the relevant country specifi­cations.
The electrical connections may only be set up by technically qualified per-
sonnel.
Completely isolate the device from the main supply, if live parts can be
touched while working.
The electromagnetic compatibility corresponds to EN 61326.The input, output and supply cables must be laid spatially separated from
one another and not parallel to one another.
The device is not suitable for installation in explosion-endangered areas.Apart from faulty installation, wrongly set values on the device can also
have an adverse impact on the following process in its proper functioning, or result in damage.
The device must be grounded:
Head-mounted transmitters: e.g. indirectly above a metallic pipe. Transmitter with separate sensor: e.g. indirectly above a metallic pipe or above the wall fastening.
The electrical connections may only be set up by technically qualified personnel.
31
7 Installation
7.1 General
Open the oper­ating unit
Unscrew the cover (1)
Connect the cables
(1)
For connecting the individual cores, pull out the threaded plug terminals (1) in the operating unit.
Lead the connecting cables through the cable glands (2).
(1)
(2)
SETUP
7 Installation
32
Wiring
Wiring sugges­tion, head­mounted trans­mitter
Wiring sugges­tion Tran sm it te r with separate sensor
Caution:
In the case of devices with a separate sensor and M12-plugs/ -sockets, the threaded terminals in the device are pigmented.
Removing the pigmenting will result in the warranty being rendered null and void.
In the case of devices with a separate sensor (base type supplementation (2) /65 or /60), for every instrument, the transmitter and the separate sensor are matched to one another at the factory. When connecting the components, ensure that the production number of the external sensor (on the flag tag on the connecting cable) is identical to the pro­duction number of the transmitter (on the nameplate).
Cable gland M16 (PA)
Cable gland M12 (PA)
Supply and signal output (conductivity and temperature) Cable gland M12 (PA)
Binary input
Switching outputs
Connector I Supply and signal output (conductivity and concentration) M12 built-in plug conn. 5-pole
Connector II Signal output temperature and binary input Switching outputs M12 built-in socket conn. 8-pole
Connector III Inductive sensor M12 8-pole
built-in socket conn.
33
7 Installation
Connection lay­out of the trans­mitter
Connection Threaded
terminals
Plug / Pin
Voltage supply
Standard production feature: Voltage supply DC 19…31 V (with polarity reversal protection)
In case of extra code 844: AC 24 V
1 L+
2 L-
I / 1 I / 2
Outputs
analog signal output conductivity/ concentration
0...20 mA or 20...0 mA or
4...20 mA or 20...4 mA or
0...10 V or 10...0 V or
2...10 V or 10...2 V (galvanically separated)
3 +
4 -
I / 3 I / 4
analog signal output temperature
0...20 mA or 20...0 mA or
4...20 mA or 20...4 mA or
0...10 V or 10...0 V or
2...10 V or 10...2 V (galvanically separated)
5+
6-
II / 1 II / 2
Switching output K1 (potential-free) Switch position indication LED K1
7 8
II / 3 II / 4
Switching output K2 Switch position indication LED K2
9
10
II / 5 II / 6
Binary inputs
Binary input E1 11
12
II / 7
I / 5
Binary input E2 13
14
II / 8
I / 5
8
7
10
9
11
12
13
14
34
8 Setup-Program
8.1 Function
Configurable parameters
With the optional Setup program that is available, the transmitter can be com­fortably matched to the requirements.
- Setting the measurement range and the measurement range limits.
- Setting the behavior of the outputs in case of overshooting of the measure­ment range.
- Setting the functions of the switching outputs K1 and K2.
- Setting the functions of the binary inputs E1 and E2.
- Setting special functions (e.g. desalination function).
- Setting a customer-specific characteristic
-etc.
Connection
Data transmission from and to the transmitter can only take place if the voltage supply is connected to it see chapter 7 "Installation", page 30ff.
The Setup interface does not have any galvanic isolation. There­fore, when connecting the PC interface cable, it is imperative to ensure that either the voltage supply of the transmitter or of the PC are not electrolytically coupled to the ground (e.g. notebook should be used in battery mode).
SETUP
PC interface cable
TN 70/00350260 or 70/00456352
Inserted adapter
(present in the Setup set)
Voltage supply
35
9 Commissioning
9.1 Head transmitter or transmitter with separate sensor
Install the device, see "Installation", page 21.Connect the device, see "Installation", page 30.
9.2 Replacement sensor
Connect the sensor, see the operating manual of the replacement sensor.Calibrate the sensor, see the operating manual of the replacement sensor.
The measuring transmitters are checked in the factory for flawless operability and shipped ready for operation.
In the case of devices with a separate sensor (base type supplementation (2)
66), for every instrument, the transmitter and the separate sensor are matched
to one another at the factory. When connecting the components, ensure that the production number of the external sensor (on the flag tag on the connecting cable) is identical to the pro­duction number of the transmitter (on the nameplate).
36
10 Operation
10.1 Operating elements
JUMO CTI-750 with and with­out LC-Display
(1) Graphical LC-display,
background lit up.
(2) Key
confirm inputs, select menu.
(3) Key
interrupt inputs without saving / interrupt calibration. One menu-level back.
(4) Key
increase numerical value / forward the selection.
(5) Key
decrease numerical value / forward the selection.
(1)
(2) (3)
(4)
(5)
(5)
PGM
EXIT
(6) LEDs "K1" / "K2" indicate the
status of the switching outputs. In normal operation, the LED flows when the corresponding switching output is active. If the wiper function is acti­vated, the LED only indicates the status.
The LED "K1" flashes during the calibration. In case of an error, LEDs "K1" and LED "K2".flash.
37
10 Operation
LC-Display
(1) (2) (3)
(4)
(5) (6)
Output K1 is active Output K2 is active Binary input 1 is
triggered Binary input 2 is
triggered Keyboard is locked Device status (hints)
- Alarm (e.g. overrange)
- Calib. flashing (calibration timer
run out)
- Calib (customer calibration enabled)
(7)
(8)
(9)
(10) (11)
Output mode
- manual (manual operation)
- hold (hold operation) Conductivity /concentration-
measured value Unit of the conductivity- /con-
centrations measured value Medium temperature Device status e.g.
- Measurement (normal)
- Desalination (desalination function)
- Dosing (desalination function)
- Locked (desalination function)
- Status of the calibration
(1) (2) (3) (4) (6)(5) (7)
(8)
(9)(10)
(11)
10 Operation
38
10.2 Operation principle
10.2.1 Operating in levels
Measurement mode, see chapter 10.4 "Measurement mode", page 41
OPERATOR LEVEL, see chapter 10.5 "Operating level", page 41
INPUT CONDUCTIVITY MEASUREMENT RANGE 1...4
TEMP.COMPENSATION TEMP.COEFFICIENT 1...4 REFERENCE TEMPERATURE REL.CELL CONSTANT MOUNTING FACTOR CONCENTR.MEASUREMENT CONCENTR.RANGE OFFSET FILTER TIME CALIBR.INTERVAL
OUTPUT CONDUCTIVITY SIGNAL TYPE
SCALING START 1...4 SCALING END 1...4 IN CASE OF ALARM IN CASE OF CALIBRATION SAFETY VALUE MANUAL OPERATION MANUAL VALUE
INFLOW TEMPERATURE UNIT
MEASURED VALUE ACQUISITION MANUAL SPEC. OFFSET FILTERING TIME
OUTFLOW TEMPERATURE SIGNAL TYPE
SCALING START SCALING END IN CASE OF ALARM AT TIME O F C A L I B R AT I O N SAFETY VALUE MANUAL OPERATION MANUAL VALUE
OUTPUT BINARY 1 FUNCTION
LIMIT VALUE HYSTERESIS DISTANCE MANUAL OPERATION IN CASE OF HOLD IN CASE OF ALARM / CALIB. SWITCH-ON DELAY SWITCH-OFF DELAY PULSE DURATION
OUTPUT BINARY 2 FUNCTION
LIMIT VALUE HYSTERESIS DISTANCE MANUAL OPERATION IN CASE OF HOLD IN CASE OF ALARM / CALIB. SWITCH-ON DELAY SWITCH-OFF DELAY
PULSE DURATION INPUT BINARY 1 FUNCTION INPUT BINARY 2 FUNCTION DESALINATION FUNCTION REDUCTION
DOSING TIME
LOCKING TIME
10 Operation
39
DEVICE DATA LANGUAGE
CONTRAST
LIGHTING
LCD INVERSE
ADMINISTR.-LEVEL, see chapter 10.6 "Administrator level", page 49
Password
PAR AMET ER-L EVEL
, see chapter 10.6.1 "Parameter level", page 50 INPUT CONDUCTIVITY. OUTPUT CONDUCTIVITY INPUT TEMPERATURE OUTPUT TEMPERATURE OUTPUT BINARY 1 OUTPUT BINARY 2 INPUT BINARY 1 INPUT BINARY 2 DESALINATION FUNCTION DEVICE DATA
ENABLE LEVEL
, see chapter 10.6.2 "Release level", page 50 INPUT CONDUCTIVITY. OUTPUT CONDUCTIVITY INPUT TEMPERATURE OUTPUT TEMPERATURE OUTPUT BINARY 1 OUTPUT BINARY 2 INPUT BINARY 1 INPUT BINARY 2 DESALINATION FUNCTION DEVICE DATA
CALIB ENABLE
, see chapter 10.6.3 "Calibration release (CALIB.­RELEASE)", page 50
REL. CELL CONST. TEMP. COEFF.. LINEAR TEMP. COEFF. N-LINEAR
CALIBR.-LEVEL
, see chapter 10.7 "Calibration level", page 51 REL. CELL CONST. TEMP. COEFF.. LINEAR TEMP. COEFF. N-LINEAR
DESALINATION FUNCTION
, see chapter 10.8 "The desalination function", page 52 REDUCTION DOSING TIME LOCKING TIME
10 Operation
40
10.3 Principle
Level
41
10 Operation
10.4 Measurement mode
Depiction In the measurement mode, the conductivity compensated to the reference
temperature, or the concentration, and the temperature of the measurement medium are displayed.
10.5 Operating level
In this level, all the parameters that have been released by the Administrator (Administrator level) can be edited. All the other parameters (marked by a key
) can only be read.
Press the key for longer than 3 seconds.Select "OPERATOR LEVEL".
(1) MEASUREMENT -> Measurement mode (2) 20.5°C -> Temperature of the measurement medium (3) 203 mS/cm -> compensated conductivity of the measurement medium
(referred to the reference temperature, generally 25°C).
PGM
(1)
(2)
(3)
10 Operation
42
10.5.1 INPUT CONDUCTIVITY (Input conductivity)
MEASUREMENT RANGE 1...4
1
0...500 µS/cm
0...1000 µS/cm
0...2000 µS/cm
0...5000 µS/cm
0...10 mS/cm
0...20 mS/cm
0...50 mS/cm
0...100 mS/cm
0...200 mS/cm
0...500 mS/cm
0...1000 mS/cm
0...2000 mS/cm UNK
2
1
The measurement ranges 2, 3 and 4 are only used if "INPUT BINARY" is configured on "MEASUREMENT RANGE./TEMP C.".
2
This measurement range is not temperature-compensated.
TEMP. COMPENSATION
LINEAR
NON-LINEAR (see "Non-linear temperature coefficient (ALPHA)", page 60) NAT WATERS (permissible temperature range 0...36°C according to EN 27
888)
TEMP. COEFFICIENT 1...4
1
0...2.20...5.5%
1
The ranges 2, 3 and 4 are only used if "INPUT BINARY" is configured on "MEASUREMENT RANGE./TEMP C.".
REFERENCE TEMPERATURE
15.0...25.0...30°C
CELL CONSTANT
2.00...6.80...10.0 1/cm Checking or changing is only necessary if there is a substitute sensor (base
type supplementation 85) connected to the transmitter with a separate sensor. The cell constant is printed on the substitute sensor (K=x,xx).
REL. CELL CONSTANT
80.0...100.0...120%
43
10 Operation
MOUNTING FACTOR
80.0...100.0...120% If the minimum distances (20 mm) of the sensor to the outer wall cannot be
maintained, a limited compensation can be achieved with this parameter.
CONCENTR. MEASUREMENT
NO FUNCT.
NaOH HNO3 CUST. SPEC. (The input of the values is only possible with the optional
Setup program)
CONCENTR. RANGE
In case of HNO
3
0...25 % by WEIGHT
36...82 % by WEIGHT In case of NaOH
0...15 % by WEIGHT
25...50 % by WEIGHT
OFFSET
-100...0...+100 mS/cm (+/- 10% of the measurement range)
FILTERING TIME
00:00:00...00:00:01...00:00:25 H:M:S
CALIBR.-INTERVAL
0...999 DAYS (0 = switched off)
10 Operation
44
10.5.2 OUTPUT CONDUCTIVITY. (Output conductivity)
SIGNAL TYPE
0...20 mA
4...20 mA
20...0 mA
20...4 mA
0...10 V
2...10 V
10...0 V
10...2 V
SCALING START 1...4
1
0 µS/cm = 4 mA
Can be set in the current measurement range, depending on the signal type
1
The ranges 2, 3 and 4 are only used if
"INPUT BINARY" is configured on "MEASUREMENT RANGE./TEMP C.".
SCALING END 1...4
1
1000 µS/cm = 20 mA
Can be set in the current measurement range, depending on the signal type
1
The ranges 2, 3 and 4 are only used if
"INPUT BINARY" is configured on "MEASUREMENT RANGE./TEMP C.".
IN CASE OF ALARM
LOW (0 mA / 0 V / 3,4 mA / 1,4 V)
HIGH (22 mA / 10.7 V) SAFETY VALUE (depending on the signal type)
AT THE TIME OF CALIBRATION
ACCOMPANYING
FROZEN SAFETY VALUE
SAFETY VALUE
0,0...4,0...22,0 mA (depending on the signal type)
0...10,7 V
MANUAL OPERATION
off
on
MANUAL VALUE
0,0...4,0...22,0 mA (depending on the signal type)
0...10,7 V
45
10 Operation
10.5.3 INFLOW TEMPERATURE
UNIT
°C
°F
MEASUREMENT VALUE ACQUISITION
Sensor
manual
MANUAL SPECIFICATION
-20.0...25.0...150.0°C
OFFSET
-15.0...0.0...+15.0°C
FILTERING TIME
00:00:00...00:00:01...00:00:25 H:M:S
10.5.4 OUTFLOW TEMPERATURE
SIGNAL TYPE
0...20 mA
4...20 mA
20...0 mA
20...4 mA
0...10 V
2...10 V
10...0 V
10...2 V
SCALING START
-20 ... 0.0°C = 4 mA (depending on the signal type)
SCALING END
+200 ... 150.0°C = 20 mA (depending on the signal type)
IN CASE OF ALARM
LOW (0 mA / 0 V / 3,4 mA / 1,4 V)
HIGH (22 mA / 10.7 V) SAFETY VALUE (depending on the signal type)
10 Operation
46
AT THE TIME OF CALIBRATION
ACCOMPANYING
FROZEN SAFETY VALUE
SAFETY VALUE
0,0...4,0...22,0 mA (depending on the signal type)
0...10,7 V
MANUAL OPERATION
off
on
MANUAL VALUE
0,0...4,0...22,0 mA (depending on the signal type)
0...10,7 V
10.5.5 OUTPUT BINARY 1 and OUTPUT BINARY 2
FUNCTION
NO FUNCTION
COND. MIN. COND. MAX. COND. LK1 COND. LK2 TEMP. MIN. TEMP. MAX. TEMP. LK1 TEMP. LK2 CALIBR. TIMER ALARM
Max limit comparator Min limit comparator
Hysteresis
Cond.
On
Off
Limit value Setpoint
Hysteresis
Limit value Setpoint
Cond.
On
Off
47
10 Operation
LIMIT VALUE
-20.0... 999.0 (depending on the function, see above)
HYSTERESIS
0.0...1.0...999.0 (depending on the function, see above)
DISTANCE
0.0... 999.0 (depending on the function, see above)
MANUAL OPERATION
OFF
ON
IN CASE OF HOLD
INACTIVE
ACTIVE FROZEN
IN CASE OF ALARM / CALIB.
INACTIVE
ACTIVE FROZEN
Alarm window LK1 Alarm window LK2
Wiper contact Triggering condition longer than pulse duration
Wiper contact Triggering condition shorter than pulse duration
Hysteresis
Cond.
On
Off
Limit value Setpoint
Distance
Hysterese
Sollwert
Lf
Ein
Aus
Grenzwert
Abstand
10 Operation
48
SWITCH-ON DELAY
00:00:00...01:00:00 H:M:S
SWITCH-OFF DELAY
00:00:00...01:00:00 H:M:S
PULSE DURATION
00:00:00...01:00:00 H:M:S (see above: "Function, wiping contact)
10.5.6 INPUT BINARY 1 and INPUT BINARY 2
FUNCTION
NO FUNCTION
KEYBOARD LOCK/HOLD MEASUREMENT RANGE./TEMP C. DESALINATION FUNCTION.
10.5.7 DESALINATION FUNCTION (Description see "The desalination function", page
52)
REDUCTION
0...10...50%
DOSING TIME
0:00:00...00:01:00...18:00:00 H:M:S
LOCKING TIME
0:00:00...00:01:00...18:00:00 H:M:S
Setting parameters binary input 1 binary input 2
Measurement range- / Temperature­coefficient switching
MB1 / Tk1 open open MB2 / Tk2 closed open MB3 / Tk3 open closed
MB4 / Tk4 closed closed Keyboard lock closed X Hold-function X closed Desalination function: Start close (flank 0 - 1) open Desalination function Stop open close (flank 0 - 1)
49
10 Operation
10.5.8 DEVICE DATA
LANGUAGE
GERMAN
ENGLISH FRENCH SPANISH POLISH SWEDISH ITALIAN PORTUGUESE DUTCH RUSSIAN
CONTRAST
0...6...11
LIGHTING
OFF ON FOR OPERATION (about 50 s after the last keystroke,
the lighting gets switched off)
INVERSE LCD
OFF
ON
10.6 Administrator level
- All the parameters can be edited in this level.
- In this level, it is also possible to determine which parameters a "normal" user may edit or which calibrations may be carried out. Editable parameters can be edited in the operator level. Non-editable parameters are marked in the operator level with a key sym­bol .
You can get to the Administrator level as follows:
Press the key for longer than 3 seconds.With the keys or , select "ADMINISTRATOR-LEVEL".Keys or input the password 300.Press the key .
By inputting the password 7485 in the Administrator level, the operating language is reset to English.
PGM
PGM
10 Operation
50
Levels of the Administrator level
10.6.1 Parameter level
In this level, the Administrator can edit any parameter of the operator level. The structure of the parameter level in the Administrator level is identical to the operator level see "Operating level", page 41 and onwards.
10.6.2 Release level
In this level, the Administrator can determine which parameters the operator is permitted to modify in the operator level.. The options "READ ONLY" and "EDIT" are available for the purpose. The structure of the parameter level in the Administrator level is identical to the operator level see "Operating level", page 41 and onwards.
10.6.3 Calibration release (CALIB.-RELEASE)
In this level, the Administrator can specify whether the operator may calibrate
- the relative cell constant
- the linear temperature coefficient
- the non-linear temperature coefficient
51
10 Operation
i.e. change them
10.7 Calibration level
In this level, the calibrations released by the Administrator (Administrator level) can be carried out.
Press the key for longer than 3 seconds.With the keys or , select "CALIBRATION-LEVEL".
10.7.1 REL. CELL CONSTANT. (relative cell constant)
If this function has been released by the Administrator, the operator can cali­brate the relative cell constant of the JUMO CTI-750 here; see "Calibration of the relative cell constant", page 56.
10.7.2 TEMP. COEF. LINEAR (Temperature coefficient linear)
If this function has been released by the Administrator, the operator can cali­brate the JUMO CTI-750 on fluids with linear temperature coefficients; see "Linear temperature coefficient (ALPHA)", page 57.
10.7.3 TEMP. CO. N-LINEAR (Temperature coefficient, non-linear)
If this function has been released by the Administrator, the operator can cali­brate the JUMO CTI-750 on fluids with non-linear temperature coefficients; see "Non-linear temperature coefficient (ALPHA)", page 60.
PGM
10 Operation
52
10.8 The desalination function
Brief descrip­tion
In the case of cooling water, using the conductivity, the total salt content is assessed. Upon reaching a limiting conductivity (at maximum permissible salt concentration / densification), dilution of the cooling water is necessary. For this purpose, a desalination valve is opened, densified water flows out and is supplemented with fresh water. After the conductivity of the cooling water has dropped below the limit value, the desalination valve is closed again.
Addition of bio­cide
In order to prevent biological growth in cooling systems, biocides are added to the cooling water. There is no ideal control variable for the added quantity and time of a biocide dosing. In most cases, the dosing time is used as the control variable. Here, the dosing quantity is obtained from the pump capacity and pump operating time (plant-specific). The success of the biocide treatment must be checked at regular intervals.
Desalination before the addi­tion of biocide
If a conductivity-enhancing biocide is added to the cooling water, the conduc­tivity can cross the limit value as a result. Thereupon, the desalination valve would open and a part of the added biocide would be discharged into the effluent canal (note and comply with the legal conditions!).
To prevent this, before adding the biocide, the conductivity in the cooling sys­tem is reduced by desalination by about 10% below the limit value. Thereafter, the desalination valve is temporarily blocked.
Desalination locking
After the addition of the biocide, the desalination should be blocked till the biocide that has been added has mostly decomposed in the cooling system (not and comply with the legal requirements.).
Realization in the case of the JUMO CTI-750
- The desalination function is only possible in the mode "Conductivity mea­surement". Not in the case of concentration measurement..
- If the desalination function is activated, all the parameters that are not rele­vant for this function are turned off.
- The desalination function can be started via the binary input 1 and stopped using the binary input 2, see "INPUT BINARY 1 and INPUT BINARY 2", page 48. The desalination function can also be stopped with the key.
- The current status of the desalination function is shown in the display.
- The desalination valve is controlled through output K1.
- The addition of biocide is controlled via output K2.
- After the desalination, K1 goes into the configured Hold state (desalination locking).
EXIT
53
10 Operation
- The desalination reduction can be set in the range from 1...50% below the actual limit value of binary input 1. The default setting is 10% below the limit value.
10.8.1 Setting the desalination function
All the parameters are plant-dependent and must be matched to the prevalent conditions.
Press the key for longer than 3 seconds.With the keys or , select the "OPERATOR LEVEL"; with key , con-
firm the selection.
With the keys or , select "INPUT BINARY"; with key , confirm the
selection.
PGM
PGM
PGM
10 Operation
54
With the keys or , select "DESALINATION FUNCT.". with key ,
confirm the selection.
Use the key to switch to the operator level.With the key select "DESALINATION FUNCTION".
With key , confirm the selection.
Set the desalination reduction with the keys or in the range from
1...10...50% below the actual limit value.
with key , confirm the setting.With the keys or , select "DOSING TIME"; with key , confirm the
selection.
Set the dosing time with the keys or in the range from
0:00:00...00:01:00...18:00:00 H:M:S.
with key , confirm the setting.
PGM
EXIT
PGM
PGM
PGM
PGM
55
10 Operation
With the keys or , select "LOCKING TIME"; with key , confirm the
selection.
Set the locking time with the keys or in the range from
0:00:00...00:01:00...18:00:00 H:M:S.
with key , confirm the setting.
If there is a failure of the supply voltage during the running of the desalination function, the function is canceled.
In order that the desalination function can run again, it must be started again.
PGM
PGM
56
11 Calibration
11.1 General
For increasing the accuracy, the device has various calibration options.
11.2 Calibration of the relative cell constant
In case of increased demands on the accuracy, the cell constant must first be calibrated.
Precondition - the JUMO CTI-750 must be supplied with power.
see chapter 7 "Installation", page 30ff.
- The sensor must be connected to the transmitter (in case of "shouldered" construction).
- The transmitter is in "Measurement mode".
Submerge the conductivity sensor in a reference solution of known conduc-
tivity.
Press the key for longer than 3 seconds.With the keys or , select "CALIBRATION-LEVEL". With key , con-
firm the selection.
At regular intervals (depending on the measurement medium), the conductivity sensor must be cleaned and calibrated.
The LED "K1" flashes during the calibration.
During the calibration, the temperature of the measurement solu­tion must remain constant.
PGM
PGM
57
11 Calibration
With the keys or , select "REL. CELL CONSTANT.". with key ,
confirm the selection.
When the measured value is stable, press the key.With the keys or , correct the displayed uncompensated conductiv-
ity value to the conductivity value of the reference solution.
Press the key.
The relative cell constant calculated by the JUMO CTI-750 is displayed.
Accept the relative cell constant determined -> press the key for longer
than 3 seconds, or discard the value -> press the key .
The transmitter is in the "Calibration menu". Press key ;
The transmitter is in "Measurement mode" and displays the compensated conductivity of the reference solution.
11.3 Calibration of the temperature coefficient of the mea­surement solution
11.3.1 Linear temperature coefficient (ALPHA)
The conductivity of any measurement solution changes according it its spe­cific temperature coefficient. We therefore recommend carrying out the calibration of the temperature coef­ficient
PGM
PGM
PGM
PGM
EXIT
EXIT
11 Calibration
58
Precondition - the JUMO CTI-750 must be supplied with power.
see chapter 7 "Installation", page 30ff.
- The sensor must be connected to the transmitter (in case of "shouldered" construction).
- The transmitter is in "Measurement mode".
Submerge the conductivity sensor in a sample of the measurement solu-
tion.
Press the key for longer than 3 seconds.With the keys or , select "CALIBRATION-LEVEL". with key , con-
firm the selection.
With the keys or , select "TEMP. COEFF LINEAR". with key , con-
firm the selection.
With the keys or , input the operating temperature and confirm with
the key .
The operating temperature must be at least 5°C above or below the reference temperature (25.0°C).
PGM
PGM
PGM
PGM
59
11 Calibration
The LC-Display now shows
- at the top (1) the selected operating temperature (flashing)
- below that (2) the reference temperature (flashing)
- below that (3) the current sensor temperature (static)
Heat the measurement medium till both the reference temperature as well
as the operating temperature are reached (the corresponding value does not flash any more).
The LC-display now shows the determined temperature coefficient in %/K. Accept the temperature coefficient determined -> press the key for longer
than 3 seconds, or discard the value -> press the key .
The transmitter is in the "Calibration menu". Press key ;
The transmitter is in "Measurement mode" and displays the compensated conductivity of the reference solution.
During the calibration, the temperature changing speed of the measurement solution of 10 K/min in the case of JUMO CTI-750 with a free-standing tem­perature sensor or 1 K/min in the case of JUMO CTI-750 with an integrated tempera­ture sensor must not be exceeded.
As soon as one of the target temperatures is reached, its display becomes static (not flashing).
Calibration is also possible in the cooling process (under dropping temperature). The start is made above the operating temperature, the end below the reference temperature.
PGM
EXIT
EXIT
(1)
(2)
(3)
11 Calibration
60
11.3.2 Non-linear temperature coefficient (ALPHA)
General Since the temperature coefficient of some media is not constant over a larger
temperature range, the JUMO CTI-750 has the facility to divide a temperature range (T
start
to T
end
) into 5 sub-ranges. In each of these ranges, compensation
can be carried out with different TC values. This so-called TC curve can
- be edited with the Setup program and transmitted to the device
- or calibrated automatically at the device.
Determining the TC-curve
Calculation of a temperature coefficient
α = Temperature coefficient (TC) γ = uncompensated conductivity
γ
Ref
T=T
Start 1
Temperature
Uncompensated
conductivity
γ
2
T
2
T
Ref
T=T
End 6
T
5
T
3
T
4
1
2
3
4
5
61
11 Calibration
TC-curve
Temperature compensation with the TC­curve
With the help of the current medium temperature, the corresponding tempera­ture coefficient is determined from the TC curve,’ see "TC-curve", page 61.
Intermediate values, e.g. (α
x
in the case of Tx) between two determined values
(α
3
in the case of T3) and (α4 in the case of T4) are approximated linearly.
With the TC determined, the compensated electric conductance is calculated as in the case of the linear temperature compensation.
Sequence of the automatic calibration on the JUMO CTI­750
The TC curve is plotted automatically in a temperature range determined by the user. Here, the temperature range from the starting to the final temperature is divided into 5 equal sections.
The temperature range must be greater than 20 Kelvin and overlap the refer­ence temperature. Example: Reference temperature 25°C, starting temperature 18°C and final temperature 50°C.
If the measured temperature is less than the starting temperature, compensation is done with the first TC. If the measured temperature is greater than the final temperature, compensation is done with the last TC.
11 Calibration
62
Precondition - the JUMO CTI-750 must be supplied with power.
see chapter 7 "Installation", page 30ff.
- The sensor must be connected to the transmitter (in case of "shouldered" construction).
- The transmitter is in "Measurement mode".
Submerge the conductivity sensor in a sample of the measurement solu-
tion.
Press the key for longer than 3 seconds.With the keys or , select "CALIBRATION-LEVEL". with key , con-
firm the selection.
With the keys or , select "TEMPCO. N-LINEAR". with key , con-
firm the selection.
The temperature changing speed may not exceed
- 10 K/min in case of a free-standing temperature sensor and
- 01 K/min in case of an integrated temperature sensor !
PGM
PGM
PGM
63
11 Calibration
With the keys or , input the starting temperature and confirm with
the key .
With the keys or , input the final temperature and confirm with the
key .
The transmitter automatically determines the temperature data points. The LC display now shows
- at the top (1) the next temperature to be reached (flashing)
- below that (2), the current sensor temperature (static)
Heat the measurement medium till the flashing temperature is exceeded or
undershot.
The starting temperature must be below the reference temperature (25.0°C).
The final temperature must be at least 20°C above the starting temperature.
PGM
PGM
(1)
(2)
11 Calibration
64
The next temperature to be reached is displayed flashing.
Heat the measurement medium till the flashing temperature is exceeded.Repeat the procedure till all 6 temperature coefficients are determined by
the JUMO CTI-750.
The LC-display now shows the determined temperature coefficients in %/K. Accept the temperature coefficients determined -> press the key for
longer than 3 seconds, or discard the values -> press the key .
The transmitter is in the "Calibration menu". Press key ;
The transmitter is in "Measurement mode" and displays the compensated conductivity of the reference solution.
During the calibration, the temperature changing speed of the measurement solution of 10 K/min in the case of JUMO CTI-750 with a free-standing tem­perature sensor or 1 K/min in the case of JUMO CTI-750 with an integrated tempera­ture sensor must not be exceeded.
As soon as one of the target temperatures is reached, its display becomes static (not flashing).
PGM
EXIT
EXIT
65
12 Maintenance
12.1 Conductivity-clean sensor
Deposits Deposits on the sensor part can be removed with a soft brush (e.g. bottle
brush).
Do not use any solvents. Stubborn sediments or deposits can be dissolved with dilute
hydrochloric acid and removed.
Comply with the safety specifications.
66
13 Rectifying errors and faults
Error possibilities
t
13.1 Device checking
General The device is calibrated at the factory and is maintenance-free. If measure-
ment value variations from unknown causes should occur nonetheless, the transmitter can be checked as follows.
Problem Possible cause Measure
No display of measured value or signal output
Voltage supply absent Check voltage supply,
check terminals
Measured value display 000 or signal output 0% (e.g. 4 mA)
Sensor not submerged in medium; Tank level too low
Fill up tank
Flow meter blocked
Clean the flow meter
Sensor faulty see "Device checking",
page 66
Measured value display 8888 flashing + device status ALARM flashing.
The temperature display is OK or LED 1 + LED 2
flashing
Out of range => the mea­surement/ display range was exceeded or under­shot
select suitable measure­ment range or check the concentration table
Measured value display 8888 flashing + device status ALARM flashing.
The temperature display is 8888 flashing or LED 1 + LED 2
flashing
The temperature sensor is faulty
The transmitter or the conductivity sensor must be replaced.
or set measured value ac-
quisition "Inflow temper­ature" briefly to manual, see "INFLOW TEMPER­ATURE" , page 45.
wrong or varying measured value display
Sensor not submerged deep enough
Fill the tank
no blending taking place ensure good blending
near the sensor, ensure clearance of about 5 mm for circulation
Air bubbles Check installation loca-
tion, see "General", page 21.
67
13 Rectifying errors and faults
13.1.1 Checking with resistance loop
Position of the resistance loop
Lead a wire through the measurement cell (see Figure)Connect the resistance R to the wire
Calculation of the resistance
Formula for calculation of the resistance of the resistance loop:
Remark: 1 mS/cm = 1·10
-3
S/cm
1 µS/cm = 1·10
-6
S/cm
In case of display values up to 20 mS, the resistance loop must have 1 turn. In case of display values from 50 mS onwards, the resistance loop must have 3 turns.
When calibrating the sensitive part of the measurement cell, do not place on or touch any surface, else the measured value will get fal­sified.
R =
N
2
· K
R = Resistance of the resistance loop
Lf N = Number of turns in the loop
K = Cell constant Lf = desired display in S/cm
The cell constant of the JUMO CTI-750 depends on the structural shape.!
The Tee measuring cell has a cell constant of 6.80 1/cm. The VARIVENT
®
-measuring celll has a cell constant of 6.50 1/cm.
13 Rectifying errors and faults
68
Example 1 The JUMO CTI-750 with Tee-shaped measuring cell should display 20 mS:
To get a display of 20 mS/cm, the resistance loop (with 1 turn) must have a re­sistance of 340 Ohm.
Example 2 The JUMO CTI-750 with Tee-shaped measuring cell should display 500 mS:
To get a display of 500 mS/cm, the resistance loop (with 3 turns) must have a resistance of 122.4 Ohm.
R =
1
2
·6.80 1/cm = 340
Ω
20·10-3 S/cm
R =
3
2
·6.80 1/cm = 122.4
Ω
500·10-3 S/cm
69
13 Rectifying errors and faults
Pre-calculated values
The display value 0 is obtained when the sensor is dry and without conducting coats, as well as when there is no resistance loop.
Structural shape/ Cell constant
Material: PEEK K = 6.80 1/cm
Material: PEEK
K = 6.50 1/cm
Display in case of
measurement
range end
Number of
turns
Cell constant
[1/cm]
Required resis-
tance
[Ω]
500 µS/cm 1 6,80 13600,00
1 6,50 13000,00
1000 µS/cm 1 6,80 6800,00
1 6,50 6500,00
2000 µS/cm 1 6,80 3400,00
1 6,50 3250,00
5000 µS/cm 1 6,80 1360,00
1 6,50 1300,00
10 mS/cm 1 6,80 680,00
1 6,50 650,00
20 mS/cm 1 6,80 340,00
1 6,50 325,00
50 mS/cm 3 6,80 1224,00
3 6,50 1170,00
100 mS/cm 3 6,80 612,00
3 6,50 585,00
200 mS/cm 3 6,80 306,00
3 6,50 292,50
500 mS/cm 3 6,80 122,40
3 6,50 117,00
1000 mS/cm 3 6,80 61,20
3 6,50 58,50
2000 mS/cm 3 6,80 30,60
3 6,50 29,25
Material: PVDF K = 5.45 1/cm
13 Rectifying errors and faults
70
Carry out test Determine test resistance.
Connect the JUMO CTI-750 electrically, see chapter 7 "Installation", page
30.
Select the corresponding measurement range, see chapter 10.5.1 "INPUT
CONDUCTIVITY (Input conductivity)", page 42 -> MEASUREMENT RANGE
1...4
Set TC to 0%/K, see chapter 10.5.1 "INPUT CONDUCTIVITY (Input con-
ductivity)", page 42 -> TEMP.COEFFICIENT.
Put the resistance loop in place in accordance with the figure.
13.1.2 Testing with reference fluid
Put in test solu­tion
Carry out test Prepare the conductivity test solution in a sufficiently large container
Connect the device electrically, see chapter 7 "Installation", page 30.Select the measurement range according to the conductivity test solution,
see chapter 10.5.1 "INPUT CONDUCTIVITY (Input conductivity)", page 42 ­> MEASUREMENT RANGE 1...4
Set TC to 0%/K, see chapter 10.5.1 "INPUT CONDUCTIVITY (Input con-
ductivity)", page 42 -> TEMP.COEFFICIENT.
Submerge the measurement cell in the container and do not move any
more during the measurement.
71
13 Rectifying errors and faults
13.1.3 Testing with reference measuring instrument
Put in test solu­tion
Carry out test Prepare the conductivity test solution in a sufficiently large container
Connect the JUMO CTI-750 electrically, see chapter 7 "Installation", page
30.
Select the measurement range according to the conductivity test solution,
see chapter 10.5.1 "INPUT CONDUCTIVITY (Input conductivity)", page 42 ­> MEASUREMENT RANGE 1...4
Set TC to 0%/K, see chapter 10.5.1 "INPUT CONDUCTIVITY (Input con-
ductivity)", page 42 -> TEMP.COEFFICIENT.
Set the TC of the reference instrument also to 0%/K (see operating manual
of the reference instrument). If that is not possible, the liquid sample must be tempered to the reference temperature of the reference instrument.
Submerge the measuring cell to be tested and the measuring cell of the ref-
erence instrument in the tank and do not move them any more at all during measurement.
The output and the display of the instrument to be tested or the display of
the display device connected to it and that of the reference device con­nected to it must match after taking the permissible instrument errors into account.
72
14 Annex
14.1 Before configuring
If many parameters of the device are to be re-configured, it is advisable to make a note of all the parameters to be changed in the following table, and to process the parameters in the given sequence.
The following list shows the maximum number of modifiable parameters.
Depending on the configuration, some of the parameters are not modifiable (editable) in this device.
Parameter Selection / Range of values
Factory setting
New setting see page
Input, conductivity
Measurement range 1...4 0...500 µS/cm
0...1000 µS/cm
0...2000 µS/cm
0...5000 µS/cm
0...10 mS/cm
0...20 mS/cm
0...50 mS/cm
0...100 mS/cm
0...200 mS/cm
0...500 mS/cm
0...1000 mS/cm
0...2000 mS/cm (uncompensated)
42
Temperature compensa­tion
linear
non-linear natural water
42
Temperature coefficient
1...4
0...2.20...5.5%/K 42
Reference temperature 15.0...25.0...30°C 42 Cell constant 2.00...6.80...10.00 1/cm 42 Relative cell constant 80.0...100.0...120.0% 42 Mounting factor 80,0...100.0...120,0% 43 Concentration measure-
ment
no function
NaOH HNO
3
customer-specific
43
Offset -200...0...+200 mS/cm 43 Filtering time 00:00:01...00:00:25 H:M:S 43 Calibration interval 0...999 days 43
Output, conductivity
73
14 Annex
Signal type 0...20 mA
4...20 mA
20...0 mA
20...4 mA
0...10 V
2...10 V
10...0 V
10...2 V
44
Scaling, beginning 0...90% = 4 mA (e.g.)
from the measurement range scope
44
Scaling: End 100...10% = 20 mA (e.g.)
from measurement range scope
44
In case of an alarm low
high safety value
44
At the time of calibration accompanying
frozen safety value
44
Safety value 0.0...4.0...22.0 mA 44 Manual operation off
on
44
Manual value 0.0...4.0...22.0 mA 44
Inflow temperature
Unit °C
°F
45
Measurement value acqui­sition
Sensor
manual
45
Manual specification -20.0...25...150°C 45 Offset -15.0...0.0...+15°C 45 Filtering time 00:00:00...00:00:01...00:00:25 H:M:S 45
Outflow temperature
Signal type 0...20 mA
4...20 mA
20...0 mA
20...4 mA
0...10 V
2...10 V
10...0 V
10...2 V
45
Scaling, beginning -20...0.0...183°C = 4 mA
(0...90% of measurement range scope)
45
Parameter Selection / Range of values
Factory setting
New setting see page
14 Annex
74
Scaling: End -3...150...200°C = 20 mA
(100...10% of measurement range scope)
45
In case of an alarm low
high safety value
44
At the time of calibration accompanying
frozen safety value
44
Safety value 0.0...4.0...22.0 mA 46 Manual operation off
on
44
Manual value 0.0...4.0...22.0 mA 44
Output binary 1 or binary 2
Function no function
Conductivity min.-contact Conductivity Max.-contact Conductivity LK1 Conductivity LK2 Conductivity Min.-contact Conductivity Max.-contact Conductivity LK1 Conductivity LK2 Calibration timer Alarm
46
Limit value 0.0...9999.0 47 Hysteresis 0.0...1.0...999.0 47 Distance 0.0...999.0 47 Manual operation off
on
47
in case of Hold inactive
active frozen
47
in case of alarm / calibra­tion
inactive
active frozen
47
Switch-on delay 00:00:00...01:00:00 H:M:S 48 Switch-off delay 00:00:00...01:00:00 H:M:S 48 Pulse duration 00:00:00...01:00:00 H:M:S 48
Input binary
Parameter Selection / Range of values
Factory setting
New setting see page
75
14 Annex
Function no function
Keyboard locking/ Hold Measurement range / Temperature coefficient desalination function
48
Desalination function
Reduction 0...10...50% 48 Dosing time 00:00:00...00:01:00...18:00:00 H:M:S 48 Locking time 00:00:00...00:01:00...18:00:00 H:M:S 48
Device data
Language German
English French Spanish Polish Swedish Italian Portuguese Dutch Russian
49
Contrast 0...6...11 49 Lighting off
on
at the time of operation
49
Inverse LCD off
on
49
Parameter Selection / Range of values
Factory setting
New setting see page
JUMO GmbH & Co. KG
Street address: Moltkestraße 13 - 31 36039 Fulda, Germany Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 e-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2TT, UK Phone: +44 1279 635533 Fax: +44 1279 635262 e-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-JUMO
1-800-554-JUMO Fax: 315-697-5867 e-mail:
info@jumo.us Internet: www.jumo.us
Loading...