JUMO CTI-500 Operating Instructions Manual

JUMO CTI-500
Inductive Conductivity/Concentration and
Temperature Transmitter with switch contacts
Type 202755
Operating Instructions
20275500T90Z001K000
V5.01/EN/00444870
WARNING:
All the nececssary settings are described in this manual. However, if any difficul­ties should arise during start-up, please do not carry out any unauthorized mani­pulations. You could endanger your righs under the instrument warranty!
Please contact the nearest subsidiary or the head office in such a case.
Resetting the LC display
If the brightness/contrast setting is such that the text in the display is not reada­ble, the basic setting can be restored as follows:
Switch off the supply voltage.Switch on the supply voltage and immediately keep the keys and held
down.
Resetting the operating language to "English"
If the operating language has been set and you cannot understand the text of the display, the language can be set to "English" with the Administrator password
7485. Thereafter, the desired language can be set in ADMINISTRATOR LEVEL /
DEVICE DATA / ....
Contents
1 Typographical conventions ...................................................... 5
1.1 Warning signs ..............................................................................................5
1.2 Note signs ....................................................................................................5
2 General ....................................................................................... 6
2.1 Preface .........................................................................................................6
2.2 Design of the measuring transmitter ............................................................6
3 Inductive conductivity measurement ...................................... 7
3.1 Area of application .......................................................................................7
3.2 Function .......................................................................................................8
4 Identifying the device version .................................................. 9
4.1 Nameplate ....................................................................................................9
4.2 Order details ...............................................................................................10
5 Device description .................................................................. 13
5.1 Transmitter technical data ......................................................................13
6 Mounting .................................................................................. 18
6.1 General .......................................................................................................18
6.2 Head-mounted transmitter .........................................................................19
6.3 Split version (separate sensor) ...................................................................20
7 Installation ............................................................................... 25
7.1 General ......................................................................................................26
8 Setup program ........................................................................ 29
8.1 Function .....................................................................................................29
9 Commissioning ....................................................................... 30
9.1 Head-mounted transmitter or split version ................................................30
9.2 Replacement sensor ..................................................................................30
10 Operation ................................................................................. 31
10.1 Controls ......................................................................................................31
Contents
10.2 Principle of operation .................................................................................33
10.3 Principle of operation .................................................................................35
10.4 Measurement mode ...................................................................................36
10.5 Operator level .............................................................................................36
10.6 Administrator level .....................................................................................44
10.7 Calibration level ..........................................................................................46
10.8 The dilution function ...................................................................................47
11 Calibration ............................................................................... 51
11.1 General .......................................................................................................51
11.2 Calibrating the relative cell constant ..........................................................51
11.3 Calibrating the temp. coefficient of the sample solution ...........................52
12 Maintenance ............................................................................ 60
12.1 Cleaning the conductivity sensor ...............................................................60
13 Eliminating faults and malfunctions ...................................... 61
13.1 Checking the device ..................................................................................61
14 Appendix .................................................................................. 66
14.1 Before configuration ...................................................................................66
15 China RoHS ............................................................................. 70

1.1 Warning signs

Danger
This symbol is used when there may be danger to personnel if the instructions are ignored or not followed correctly!
Caution
This symbol is used when there may be damage to equipment or data if the instructions are ignored or not followed correctly!

1.2 Note signs

Note This symbol is used when your special attention is drawn to a remark.

1 Typographical conventions

1
abc
Action instruction
Footnote
Footnotes are remarks that refer to specific points in the text. Footnotes consist of two parts:
A marker in the text, and the footnote text. The markers in the text are arranged as continuous superscript numbers.
This symbol indicates that an action to be performed is described. The individual steps are marked by this asterisk. Example: Remove crosspoint screws.
5

2 General

(2)
(1)
(3)
Head-mounted version:
Transmitter combined with
conductivity sensor,
Type 202755/xx...
Split version:
Transmitter with
separate sensor,
Type 202755/xx...
(1)
(2)
(4)
(5)
(4)
(3)

2.1 Preface

Please read these operating instructions before commissioning the instrument. Keep the manual in a place that is accessible to all users at all times.

2.2 Design of the measuring transmitter

Examples
(1) Transmitter (4) Inductive conductivity sensor
(2) Process connection (5) With or without graphics LC display
(3) Exposed temperature sensor
6

3 Inductive conductivity measurement

3.1 Area of application

General The inductive measurement method permits largely maintenance-free
acquisition of the specific conductivity, even in difficult media conditions. Unlike the conductive measurement method, problems such as electrode decomposition and polarization do not occur.
Brief description
The instrument is used for the measurement/control of conductivity or concentration in liquid media. It is particularly recommended for use in media where severe deposits of dirt, oil, grease or gypsum/lime precipitates are to be expected. The integrated temperature measurement enables fast and accurate temperature compensation, which is of particular importance when measuring conductivity.
Two built-in switching outputs can be freely programmed to monitor limits for conductivity / concentration and / or temperature. It is also possible to assign alarm and control functions (dilution).
The instrument is operated either from the membrane keypad and plain-text graphics display (operator language can be changed over) or through the user­friendly PC setup program. Simply rotating the housing cover makes it possible to read the display, regardless of whether the installation is in horizontally or vertically arranged pipes. By using the setup program, the instrument configuration data for plant documentation can be saved and printed out. To prevent any tampering, the instrument can also be supplied without keypad or display. In this case, the setup program is needed for programming.
The measuring transmitter is available either as a combined unit (transmitter and measuring cell together in one unit) or as a split version (transmitter and cell connected by cable). The split version is particularly suitable for plant subjected to strong vibration and/or significant heat radiation at the point of measurement, or for installation on sites that are difficult to access. Immersion models up to 2000 mm are available for application in open containers or sluices.
Typical areas of application
- freshwater and wastewater engineering
- HVAC systems and cooling tower monitoring (dilution control)
- flushing baths (e.g. monitoring electroplating baths)
- inlet and final control in on-site sewage treatment plants
- concentration monitoring
- vehicle washing plant
7
3 Inductive conductivity measurement
(2)
(1)
(3)
(4)
(1a)

3.2 Function

of the transmitter
of the measuring cell
The transmitter has been designed for use on site. A rugged housing protects the electronics and the electrical connections from corrosive environmental conditions (IP67 enclosure). As standard, the device has one analog signal output each for conductivity/concentration and temperature respectively. Further processing of the standard signals can take place in a suitable display/ control device, or, for example, directly in a PLC.
The output signals are electrically isolated from one another and from the medium.
The conductivity is measured using an inductive probe. A sinusoidal a.c. voltage feeds the transmitting coil. Depending on the conductivity of the liquid to be measured, a current is induced in the receiver coil. This current is proportional to the conductivity of the medium. The cell constant of the inductive probe depends on its geometry. The cell constant can also be affected by components in the immediate vicinity.
(1) Temperature sensor, exposed (1a) optionally:
internal temperature sensor (2) Measuring cell body in PP (3) Measurement coils (4) Liquid loop
8

4.1 Nameplate

On the transmitter
The nameplate is affixed to the case.
On the separate sensor
The nameplate (flag tag) is affixed to the connecting cable.
Contents
The nameplate contains important information. This includes:

4 Identifying the device version

Description Description on the
Device type Typ 202755/15-168-0-82/000 Fabrication number F-Nr 0220465201018240001 Voltage supply - DC 19 ... 31 V
Device type (Typ)
Compare the specifications on the nameplate with the order. Identify the supplied device version using the order details (order code).
Fabrication number (F-Nr)
Among other things, the fabrication number contains the date of production (year/week). Example: F-Nr = 0220465201018240001 The characters in question are in positions 12 to 15 (from the left). The device was therefore produced in the year 2018 in its 24th calendar week.
For devices with a separate sensor (type code extensions /60 or /65), the transmitter and detached sensor are matched to one another at the factory! When connecting the components, please note that the serial number of the external sensor (marked on the flag tag attached to the connecting cable) must match the serial number marked on the nameplate of the transmitter!
Example
nameplate
9
4 Identifying the device version

4.2 Order details

4.2.1 Head-mounted transmitter

(1) Basic types
202755/10 JUMO CTI-500 Head transmitter without display/keyboard, consisting of
transmitter with permanently mounted sensor
202755/15 JUMO CTI-500 Head transmitter with display/keyboard
(2) Process connection
168 PVC union nut G 1 ½ A 169 Stainless steel union nut G 1 ½ A
2,6
2
607 Taper socket with union nut DN 50 DIN 11851 (dairy compression fitting) 617 Clamping socket (clamp) 2 ½" similar to DIN 32676 690 SMS DN 2"
(3) Immersion length
0 see chapter 6.2 "Head-mounted transmitter", page 19
(4) Electrical connection
82 cable glands 83 M12 plug / socket connector (instead of cable glands) 84 two M16 cable glands and one blanking plug
(5) Extra codes
4
000 no extra code 062 with DNV GL approval
8
268 internal temperature sensor 580 1 set M12 plug/socket connectors 768 Measurement cell material PVDF
5
844 Voltage supply AC 24 V
1
7
3
(1) (2) (3) (4) (5) (5) Order code / - - / , ... Order example 202755/10 / 168 - 0 - 82 / 000 ,
1
The PC setup program is required for programming the device.
2
Special tee is not included in delivery.
3
If required, order extra code /580
4
Indicate the type code extension in succession separated by a comma.
5
Only in case of process connection 168 and 169 in combination with type code extension 268.
6
Maximum medium temperature 60°C.
7
Assembly material (holding clamp) not part of the supplied kit. If required, please order additionally.
8
Not possible in conjunction with extra code 768 and/or 844.
10
4 Identifying the device version

4.2.2 Transmitter with separate sensor

(2) Basic types
202755/20 JUMO CTI-500 transmitter without display/keyboard (without sensor) 202755/25 JUMO CTI-500 transmitter with display/keyboard (without sensor) 202755/60
202755/65
202755/80 JUMO CTI-500 replacement sensor with 10 m cable without transmitter,
000 not available 168 PVC union nut G 1 ½ A 169 Stainless steel union nut G 1 ½ A 607 Taper socket with union nut DN 50 DIN 11851 (dairy compression fitting) 617 Clamping socket (clamp) 2 ½" similar to DIN 32676 690 SMS DN 2" 706 immersion version
0 not available 0500 0500 mm 1000 1000 mm 1500 1500 mm 2000 2000 mm (maximum)
xxxx special length (in 250 mm steps; e.g. 0250; 0750; 1250; 1750)
21 fixed cable with M12 socket connector on separate sensor 82 cable glands on the operating unit 83 M12 plug/socket connectors on operating unit 84 two M16 cable glands and one blanking plug
000 no extra code 268 internal temperature sensor 580 1 set M12 plug/socket connectors 768 Measurement cell material PVDF 844 Voltage supply AC 24 V
JUMO CTI-500 transmitter without display/keyboard with sensor (cable length: 10 m)
1
JUMO CTI-500 transmitter with display/keyboard with sensor (cable length: 10 m)
including calibration set
3,7
(3) Process connection
2, 8
2
(4) Immersion length
(5) Electrical connection
4
(6) Extra codes
5
6
3
1
7
(1) (2) (3) (4) (5) (5) Order code / - - / , ... Order example 202755/65 / 706 - 1000 - 82 / 000 ,
11
4 Identifying the device version
1
The PC setup program is required for programming the instrument.
2
A tee is not included in delivery.
3
Mounting items (union/ring nuts, holding clamp) are not included in delivery. If required, please order additionally.
4
If required, order extra code /580
5
Show type extension codes in succession separated by a comma.
6
Only in case of process extension 168 and 169 in combination with type extension code 268.
7
A balancing set is urgently required for commissioning. If required, please order additionally.
8
Maximum medium temperature 60 °C.
12

5.1 Transmitter technical data

5.1.1 General

A/D converter resolution: 15-bit
sampling time: 500 msec = 2 measurements/sec
Supply For operation in SELV- and PELV-circuits!
As standard: 19 — 31 V DC (24 V DC nominal), with reverse-polarity protection
extra code 844: AC 24 V ±10%, 50...60 Hz
ripple: < 5% power consumption with display: 3 W power consumption without display: 2.6 W

5 Device description

Rating of the solid-state relays
Electrical connection
Display (option) graphics LCD with background lighting; adjustable contrast
Permissible ambient temperature (transmitter)
Permissible storage temperature (transmitter)
Enclosure protection (transmitter)
U DC 45 V U AC 30 V I 200 mA
plug-in screw terminals 2.5 mm
dimensions: 62 x 23 mm
-5 to +50°C max. 93% relative humidity, no condensation
-20 to +75°C max. 93% relative humidity, no condensation
IP67
2
or M12 plug/socket connector
Housing Polyamide
Weight depending on version and process connection
approx. 0.3 — 2 kg
13
5 Device description

5.1.2 Conductivity/ concentration transmitter

Concentration measurement
(implemented in the device software)
Calibration timer
Output signal Conductivity/ concentration
Burden 500Ω for current output
Analog output for “Alarm”
- NaOH (caustic soda) 0 — 15 % by weight or 25 — 50 % by weight
- HNO
- customer-specific concentration curve
adjustable: 0 — 999 days (0 = off)
0 — 10 V / 10 — 0 V 2 — 10 V / 10 — 2 V 0 — 20 mA / 20 — 0 mA 4 — 20 mA / 20 — 4 mA
The output signal is freely scalable.
2kΩ for voltage output
Low (0 mA / 0 V / 3.4 mA / 1.4 V) or High (22.0 mA / 10.7 V) or a value with a fixed setting (safe value)
(nitric acid); check chemical resistance of the sensor !
3
0 — 25 % by weight or 36 — 82 % by weight
freely programmable through the setup program (see “special functions”)
Measuring ranges
Four ranges can be selected. One of these ranges can be activated via an external switch or a PLC.
Measurement ranges Tran sm it te r
0 — 500 µS/cm 0 — 1000 µS/cm 0 — 2000 µS/cm 0 — 5000 µS/cm 0 — 10 mS/cm 0 — 20 mS/cm 0 — 50 mS/cm 0 — 100 mS/cm 0 — 200 mS/cm 0 — 500 mS/cm 0 — 1000 mS/cm 0 — 2000 mS/cm
1
not compensated for temperature
Note: The overall tolerance is made up of the tolerance of the transmitter + the tolerance of the sensor.
1
To ler an ce
(in % of range
span)
0.5%
14

5.1.3 Temperature transmitter

5 Device description
Temperature acquisition
Temperature measurement range
Characteristic linear
To ler an ce 0.5% of range
Output signal for temperature
Burden 500Ω for current output
manually -20.0 — 25.0 — 150°C/°F or
automatically
-20 — 150°C/°F
0 — 10 V / 10 — 0 V 2 — 10 V / 10 — 2 V 0 — 20 mA / 20 — 0 mA 4 — 20 mA / 20 — 4 mA
The output signal is im Bereich -20...+200°C freely scalable. The sensor can be used in the range -10...+100°C.
2 kΩ for voltage output
Analog output for “Alarm”
Low (0 mA / 0 V / 3.4 mA / 1.4 V) or High (22.0 mA / 10.7 V) or a value with a fixed setting (safe value)

5.1.4 Temperature compensation

Reference temperature
Temperature coefficient
Compensation range
Function - Linear compensation (constant temperature coefficient).
15 to 30°C, adjustable
0.0 to 5.5 %/°C, adjustable
-20 to 150°C
This type of compensation can be used with normal water with an acceptable level of accuracy. The temperature coefficient used is then about 2,2 %/K.
- Natural water (DIN EN27888 or ISO 7888 as the case may be). In this case, a so-called non-linear temperature compensation is used. According to the above standard, the corresponding type of compensation can be applied in the case of natural ground water, mountain spring water and surface warter.
15
5 Device description
The conductivity of the water is compensated in the range from 0°C to 36°C.
- non-linear (learning function, see special functions) Here, the actual graph of the temperature coefficient during a heating-up or cooling-down process is determined by the transmitter.

5.1.5 Sensor

Material PP (polypropylene)
Note:
Temperature, pressure and sample medium affect the cell operating life.
Temperature of the sample medium
Process connection maximum medium temperature
168 706 169 607 617 690
60 °C
80 °C
short-term100 °C
Note the limiting values (ambient temperature) of the device.
Pressure 10 bar max. at 20°C
06 bar max. at 60°C
Measurement range Sensor
0 — 500 µS/cm 0 — 1000 µS/cm 0 — 2000 µS/cm 0 — 5000 µS/cm 0 — 10 mS/cm 0 — 20 mS/cm 0 — 50 mS/cm 0 — 100 mS/cm 0 — 200 mS/cm 0 — 500 mS/cm 0 — 1000 mS/cm 0 — 2000 mS/cm
To le ra n ce
(in % of range span)
1%
0.5%
1
1%
16
1
not compensated for temperature

5.1.6 Approvals/approval marks

Approval mark DNV GL
Testing agency DNV GL
5 Device description
Certificate/ certification number
Inspection basis
Valid for Type 202755/10
TAA00001W9
DNV GL Class Guideline CG-0339
Type 202755/15
17

6 Mounting

min. 20 mm
min. 20 mm
(1)

6.1 General

Mounting site Make sure that the site is readily accessible, for calibration at a later time.
The fixing must be secure and free from vibration. Avoid direct sunlight! Take care that there is adequate flow through and around the sensor (1) ! If the device is to be mounted in a pipeline, there must be at least 20 mm
clearance between the sensor and the wall of the pipe. If it is not possible to achieve this minimum clearance, then a limited compensation can be made through the “Mounting factor” parameter.
For submerged operation in basins, a location must be chosen that is representative of the typical conductivity or concentration.
Mounting position
Screwing in and unscrewing the detached sensor
Sensor details
The measuring transmitter can be mounted in any position.
The cable must not be twisted!
Avoid putting tension on the cable. In particular, avoid tugging it.
18

6.2 Head-mounted transmitter

Installation variations
6 Mounting
19
6 Mounting

6.3 Split version (separate sensor)

Tran sm it ter head
Drilling jig for wall-mounting
20
Sensor component
6 Mounting
21
6 Mounting
Feststellschraube
Flansch verschiebbar
60
18
3
100
+5
0
Ø18
Ø125 Ø165

6.3.1 Separate sensor as immersion model

Optional accessory: Flange DN 32, part no. 00083375
Optional accessory: Flange DN 50, part no. 00083376
Fixing screw
Sliding flange
22

6.3.2 Examples of installation

Version with
process connection 168
Union nut (PVC) Insert nut G1 / " (PVC)
1
2
Tee 90° DN32 (PVC)
Union nut (PVC)
Insert nut G1 / "
1
2
Tee 90° DN40 (PVC)
Optional accessories part no.: 00439247
Optional accessories Sales No.: 20/00439249
Version with
process connection 168
3URFHVVFRQQHFWLRQ
0.'1 
7KUHDGHGSLSH DGDSWRU',16'1 
5HGXFHU7HHWREHSURYLGHGE\WKHSODQWRSHUDWRU OPUTVQQMJFECZUIFEFWJDFNBOVGBDUVSFS ',1VKRUW666'1'1'1
6ORWWHGXQLRQQXW '1 
9HUVLRQZLWK
SURFHVVFRQQHFWLRQ
6FUHZHGSLSHILWWLQJ'1
',1PLONFRQH
'1
'1
'1
6 Mounting
Weld-on threaded connection
DN50, DIN 11 851
(Counterpart for process connection -607)
part no.: 00085020
23
6 Mounting
30...50 mm
(2)
(1)
(1)

6.3.3 Pipe assembly set

The screws (1) M5 x 30 are used for pipe diameters from 30 to 40 mm. The screws (2) M5 x 40 are used for pipe diameters from 40 to 50 mm. The pipe assembly set is also suitable for horizontal pipes..
24

7 Installation

The electrical connection must only be carried out by properly qualified personnel!
The choice of cable, the installation and the electrical connection must
conform to the requirements of VDE 0100 “Regulations on the Installation of Power Circuits with Nominal Voltages below 1000 V” or the appropriate local regulations.
The electrical connection must only be carried out by qualified personnel.If contact with live parts is possible while working on the device, it must be
completely disconnected from the electrical supply.
The electromagnetic compatibility conforms to EN 61326.Run input, output and supply cables separately and not parallel to one
another.
The device is not suitable for use in areas with an explosion hazard (Ex
areas).
Apart from faulty installation, incorrect settings on the instrument may also
affect the proper functioning of the subsequent process or lead to damage.
25
7 Installation
(1)
(1)
(2)

7.1 General

Opening the operating unit
It is only necessary to open the housing for devices with cable glands.
Devices with M12 plug/socket connectors should not be opened!
Connecting up the cables
Remove four screws (1) and take off the cover
26
To connect the single conductors, pull off the pluggable screw terminals (1) in the operating unit.
Pass the connecting cables through the cable glands (2).
Wiring
Caution:
On devices with a separate sensor and M12 plug/ socket connectors, the screw terminals are sealed inside the device.
Removal of this sealing will invalidate the warranty!
7 Installation
For devices with a separate sensor (type code extensions (2) /60 or /65), the transmitter and detached sensor are matched to one another at the factory! When connecting the components, please note that the serial number of the external sensor (marked on the label attached to the connecting cable) must match the serial number marked on the nameplate of the transmitter!
27
7 Installation
8
7
10
9
11
12
13
14
Connections for the transmitter
Connections Screw
Supply voltage
Supply voltage
19 — 31 V DC
(with reverse-polarity protection)
Outputs
Analog signal output
Conductivity/concentration
0 — 20 mA resp. 20 — 0 mA or
4 — 20 mA resp. 20 — 4 mA
or
0 — 10 V resp. 10 — 0 V or
2 — 10 V resp. 10 — 2 V
(electrically isolated)
Analog signal output
Te mp er at ur e
0 — 20 mA resp. 20 — 0 mA or
4 — 20 mA resp. 20 — 4 mA
or
0 — 10 V resp. 10 — 0 V or
2 — 10 V resp. 10 — 2 V
terminals
1 L+
2 L-
3 +
4 -
5+
6-
Conn./pin
I / 1 I / 2
I / 3 I / 4
II / 1 II / 2
(electrically isolated)
Switching output K1
(floating)
7 8
Status indication
LED K1
Switching output K2
Status indication
9
10
LED K2
Binary inputs
Binary input E1 11
12
Binary input E2 13
14
II / 3 II / 4
II / 5 II / 6
II / 7
I / 5
II / 8
I / 5
28

8.1 Function

Interface cable (95/00350260)
Plug-on adapter
(included in setup set)
Supply voltage

8 Setup program

Configurable parameters
Connections
The setup program, which is available as an option, can be used for easy adaptation of the transmitter to specific requirements.
- Setting the measurement range and the range limits.
- Setting the response of the output to an out-of-range signal.
- Setting the functions of the switched outputs K1 and K2.
- Setting the functions of the binary inputs E1 and E2.
- Setting up special functions (e.g. the dilution function).
- Setting up a customer-specific characteristic,
-etc.
Data transmission from or to the transmitter can only take place when it is connected to the electrical supply see Chapter 7 "Installation", Page 25ff.
The setup interface is not electrically isolated. When connecting the PC interface cable, it is therefore absolutely essential to ensure that either the supply of the transmitter or of the PC is not electrically earthed (for instance, use a battery-powered notebook).
29

9 Commissioning

The transmitter has been tested in the factory for fault-free functioning, and is delivered ready for operation.

9.1 Head-mounted transmitter or split version

Mounting the device, see "Mounting", Page 18.Connecting the device, see "Installation", Page 25.
For devices with a separate sensor (type code extensions (2) /60 or /65), the transmitter and detached sensor are matched to one another at the factory! When connecting the components, please note that the serial number of the external sensor (marked on the label attached to the connecting cable) must match the serial number marked on the nameplate of the transmitter!
The separate sensor must be connected before switching on the transmitter!

9.2 Replacement sensor

Connect up the sensor as described in the operating instructions for the
replacement sensor.
Calibrate the sensor as described in the operating instructions for the
replacement sensor.
After changing the separate sensor, the voltage supply of the transmitter (with connected sensor) must be switched off and on again!
30

10.1 Controls

(6)
(2)
(1)
(3)
(4)
(6) (5)
PGM
EXIT
Device without LC display

10 Operation

Device with LC display
(1) Grafik LC display, back-lit (2) key, confirm entries/select menu (3) key, cancel entry without saving/cancel calibration go back one
menu level (4) key, increase value/step on in selection (5) key, reduce value/step on in selection (6) LEDs K1 and K2 show the states of the switched outputs.
In normal operation, the LED lights up if the corresponding output is
active.
If the pulse function is active, the LED only indicates the status.
The K1 LED blinks during calibration.
In fault condition, the LED K1 and LED K2 blink.
31
10 Operation
(1) (2) (3) (4) (6)(5) (7)
(8)
(9)(10)
(11)
LC display
(1) (2) (3)
(4)
(5) (6)
Output K1 is active Output K2 is active Binary input 1 is
activated Binary input 2 is
activated Keypad is inhibited Device status (indications)
- Alarm (e.g overrange)
- Calib blinking (calibration timer has run down
- Calib (customer calibration is active)
Output mode
(7)
- Hand (manual operation)
- Hold (hold operation) Conductivity/concentration
(8)
measurement Unit for conductivity/
(9)
concentration measurement Temperature of the medium
(10)
Device status e.g.
(11)
- Measurement (normal)
- Dilution (dilution function)
- Dosing (dilution function)
- Inhibited (dilution function)
- Calibration status
32

10.3 Principle of operation

10.3.1 Operation in levels

Measurement mode, see Chapter 10.4 "Principle of operation" page 35
OPERATOR LEVEL, see chapter 10.6 "Operator level", page 36
INPUT CONDUCTIVITY MEASUREMENT RANGE 1...4
TEMPERATURE COMPENSATION TEMP. COEFFICIENT 1...4 REFERENCE TEMPERATURE REL. CELL CONSTANT MOUNTING FACTOR CONCENTR. MEASUREMENT CONCENTR. RANGE OFFSET FILTER TIME CALIBR. INTERVAL
OUTPUT CONDUCTIVITY SIGNAL TYPE
SCALING START 1...4 SCALING END1...4 IN CASE OF ALARM DURING CALIBRATION SAFETY VALUE MANUAL OPERATION MANUAL VALUE
INPUT TEMPERATURE UNIT
MEAS. VALUE ACQUISITION MANUEL SPECIFICATION OFFSET FILTER TIME
OUTPUT TEMPERATURE SIGNAL TYPE
SCALING START SCALING END IN CASE OF ALARM DURING CALIBRATION SAFETY VALUE MANUAL OPERATION MANUAL VALUE
OUTPUT BINARY 1 FUNCTION
LIMIT VALUE HYSTERESIS DISTANCE MANUAL OPERATION DURING HOLD IN CASE OF ALARM/CALIBR. SWITCH-ON DELAY SWITCH-OFF DELAY PULSE DURATION
OUTPUT BINARY 2 FUNCTION
LIMIT VALUE HYSTERESIS DISTANCE MANUAL OPERATION DURING HOLD IN CASE OF ALARM/CALIBR. SWITCH-ON DELAY SWITCH-OFF DELAY
PULSE DURATION INPUT BINARY 1 FUNCTION INPUT BINARY 2 FUNCTION DILUTION REDUCTION
DOSING TIME
LOCK TIME
10 Operation
33
10 Operation
DEVICE DATA LANGUAGE
CONTRAST
LIGHTING
INVERTING LCD
ADMINISTR. LEVEL, see chapter 10.6, "Operator level", page 36
Password
PAR AMET ER LEVEL
tivity input) page 37
INPUT CONDUCTIVITY
OUTPUT CONDUCTIVITY
INPUT TEMPERATURE
OUTPUT TEMPERATURE
OUTPUT BINARY 1
OUTPUT BINARY 2
INPUT BINARY 1
INPUT BINARY 2
DILUTION FUNCTION
DEVICE DATA
, see chapter 10.6.1 "CONDUCTIVITY IN" (conduc-
CALIBRATION LEVEL
REL. CELL CONSTANT TEMP. COEFF. LINEAR TEMP. CO. NON-LINEAR
DILUTION FUNCTION
REDUCTION DOSING TIME LOCK TIME
, see chapter 10.7 "Administrator level" page 44
ENABLE LEVEL
vity output), page 38
INPUT CONDUCTIVITY
OUTPUT CONDUCTIVITY
INPUT TEMPERATURE
OUTPUT TEMPERATURE
OUTPUT BINARY 1
OUTPUT BINARY 2
INPUT BINARY 1
INPUT BINARY 2
DILUTION FUNCTION
DEVICE DATA
CALIBRATION ENABLE
39
REL. CELL CONSTANT
TEMP. COEFF. LINEAR
TEMP. CO. NON-LINEAR
, see chapter 10.8 "Calibration level" page 46
, see chapter 10.6.2 "CONDUCTIVITY OUT" (conducti-
, see chapter 10.6.3 "TEMPERATURE IN", page
34

10.4 Principle of operation

Operation in levels
10 Operation
35
10 Operation
PGM

10.5 Measurement mode

Representation In measurement mode, the conductivity is shown (compensated for the
reference temperature) or the concentration and temperature of the medium being measured.
(1)
(2)
(1) MEASUREMENT -> Measurement mode (2) 20.5°C -> Temperature of the sample medium (3) 203 mS/cm -> conductivity of the medium (compensated for the
reference/ comparison temperature – usually 25°C)
(3)

10.6 Operator level

All the parameters that have been enabled by the administrator (administrator level) can be edited in this level. All other parameters (marked by a key ) can only be read.
Press the key for at least 3 seconds.Select OPERATOR LEVEL.
36

10.6.1 CONDUCTIVITY IN (Conductivity input)

10 Operation
RANGE 1 — 4
1
0 — 500 µS/cm
0 — 1000 µS/cm
0 — 2000 µS/cm 0 — 5000 µS/cm 0 — 10 mS/cm 0 — 20 mS/cm 0 — 50 mS/cm 0 — 100 mS/cm 0 — 200 mS/cm 0 — 500 mS/cm 0 — 1000 mS/cm 0 — 2000 mS/cm UNC
1
Measurement ranges 2, 3 and 4 are only used if BINARY INPUT is configured to RANGE/TEMPCO
2
This measurement range is not temperature-compensated.
TEMP. COMPENSATION
LINEAR
NON-LINEAR (see "Non-linear temperature coefficient (ALPHA)", Page 55) NATURAL WATER (permissible temperature range 0 to 36°C to EN 27 888)
2
TEMPCO 1 — 4
1
0 — 2.20 — 5.5%
1
Ranges 2, 3 and 4 are only used if BINARY INPUT is configured to RANGE/TEMPCO.
REFERENCE TEMP.
15.0 to 25.0 to 30°C
REL. CELL CONSTANT
80.0 — 100.0 — 120%
MOUNTING FACTOR
80.0 — 100.0 — 120% If it is not possible to achieve the minimum clearance of 20 mm between the
sensor and the outer wall, then a limited compensation can be made through this parameter.
CONC. MEAS. TYPE
NO FUNCTION
NaOH HNO3
37
10 Operation
CUSTOMIZED (values can only be entered by using the optional
setup program)
CONC. MEAS. RANGE
For HNO
3
0 — 25 % BY WEIGHT
36 — 82 % BY WEIGHT For NaOH
0 — 15 % BY WEIGHT
25 — 50 % BY WEIGHT
OFFSET
-100 to 0 to +100 mS/cm (+/- 10% of range)
FILTER TIME
00:00:00 — 00:00:01 — 00:00:25 H:M:S
CALIB. INTERVAL
0 — 999 DAYS (0 = switched off)

10.6.2 CONDUCTIVITY OUT (conductivity output)

SIGNAL TYPE
0 — 20 mA
4 — 20 mA
20 — 0 mA 20 — 4 mA 0 — 10 V 2 — 10 V 10 — 0 V 10 — 2 V
SCALING START 1 — 4
0 µS/cm = 4 mA
Can be set in the range being used, depending on the signal type.
1
Ranges 2, 3 and 4 are only used if
BINARY INPUT is configured to RANGE/TEMPCO.
SCALING END 1 — 4
1
1000 µS/cm = 20 mA
Can be set in the range being used, depending on the signal type.
1
Ranges 2, 3 and 4 are only used if
BINARY INPUT is configured to RANGE/TEMPCO.
1
38
DURING ALARM
LOW (0 mA / 0 V / 3.4 mA / 1.4 V)
HIGH (22 mA / 10.7 V) SAFE VALUE (depending on the signal type)
DURING CALIBRATION
MOVING
FROZEN SAFE VALUE
SAFE VALUE
0.0 — 4.0 — 22.0 mA (depending on the signal type) 0 — 10.7 V
MANUAL MODE
OFF
ON
10 Operation
MAN. VALUE
0.0 — 4.0 — 22.0 mA (depending on the signal type) 0 — 10.7 V

10.6.3 TEMPERATURE IN

DIMENS. UNIT
°C
°F
MEAS. MODE
SENSOR
MANUAL
MANUAL VALUE
-20.0 to 25.0 to 150°C
OFFSET
FILTER TIME
-15.0 to 0.0 to 15.0°C
00:00:00 — 00:00:01 — 00:00:25 H:M:S
39
10 Operation

10.6.4 TEMPERATURE OUT

SIGNAL TYPE
0 — 20 mA
4 — 20 mA
20 — 0 mA 20 — 4 mA 0 — 10 V 2 — 10 V 10 — 0 V 10 — 2 V
SCALING START
-20.0 to 0.0°C = 4 mA (depending on the signal type)
SCALING END
+200 to 150.0°C = 20 mA (depending on the signal type)
DURING ALARM
LOW (0 mA / 0 V / 3.4 mA / 1.4 V)
HIGH (22 mA / 10.7 V) SAFE VALUE (depending on the signal type)
DURING CALIBRATION
MOVING
FROZEN SAFE VALUE
SAFE VALUE
0.0 — 4.0 — 22.0 mA (depending on the signal type) 0 — 10.7 V
MANUAL MODE
OFF
ON
MAN. VALUE
0.0 — 4.0 — 22.0 mA (depending on the signal type) 0 — 10.7 V
40

10.6.5 BINARY OUTPUT 1 and BINARY OUTPUT 2

FUNCTION
NO FUNCTION
MIN. CONDUCT. MAX. CONDUCT. LK1 CONDUCT. LK2 CONDUCT. MIN. TEMP. MAX. TEMP. LK1 TEMP. LK2 TEMP. CALIB. TIMER ALARM
10 Operation
MAX limit comparator MIN limit comparator
LK1 alarm window LK2 alarm window
Pulse contact Trigger condition longer than pulse duration
Pulse contact Trigger condition shorter than pulse duration
41
10 Operation
LIMIT
-20.0 — 0.0 — 999.0 (depending on the function, see above)
HYSTERESIS
0.0 — 0.5 — 999.0 (depending on the function, see above)
SPACING
0.0 — 999.0 (depending on the function, see above)
MANUAL MODE
OFF
ON
FOR HOLD
INACTIVE
ACTIVE FROZEN
FOR ALARM / CALIB.
INACTIVE
ACTIVE FROZEN
ON-DELAY
00:00:00 — 01:00:00 H:M:S
OFF-DELAY
00:00:00 — 01:00:00 H:M:S
PULSE DURATION
00:00:00 — 01:00:00 H:M:S (see above: “Function, Pulse contact”)
42
10 Operation

10.6.6 BINARY INPUT 1 and BINARY INPUT 2

FUNCTION
NO FUNCTION
HOLD/LOCK KEY RANGE/TEMPCO. DILUTION
Setting parameters Binary input 1 Binary input 2
Range / temperature coefficient changeover
Lock keys closed X Hold function X closed Start dilution function close (0 - 1 edge) open Stop dilution function open close (0 - 1 edge)
Range 1 / TC 1
Range 2 / TC 2
Range 3 / TC 3
Range 4 / TC 4
open open
closed open
open closed
closed closed

10.6.7 DILUTION (description: see "The dilution function", Page 47)

REDUCE
0 — 10 — 50%
DOSING TIME
LOCK TIME
0:00:00— 00:01:00— 18:00:00 H:M:S
0:00:00— 00:01:00— 18:00:00 H:M:S
43
10 Operation
PGM
PGM

10.6.8 DEVICE DATA

LANGUAGE
GERMAN
ENGLISH FRENCH ITALIAN DUTCH POLISH PORTUGUESE RUSSIAN SWEDISH SPANISH
CONTRAST 0 — 6 — 11
Entering the password 7485 in the administrator level will reset the operating language to English.
LIGHTING
OFF ON
IF OPERATED (approx. 50 s after the last key operation:
LCD INVERSE
OFF
ON

10.7 Administrator level

- All parameters can be edited (altered) in this level.
- In this level, you can also define which parameters can be edited (altered) by a “normal” user, and/or which calibration actions are permitted. Editable parameters can be edited in the operator level. Non-editable parameters are marked in the operator level by a key symbol
the lighting will be switched off)
You can access the administrator level as follows:
Press the key for at least 3 seconds.Use the or key to select ADMINISTRATOR LEVEL.Use or to enter the password 300. Press the key.
44
Levels within the administrator level
10 Operation

10.7.1 Parameter level

The administrator can edit all parameters for the operator level in this level. The structure “Parameter level” within the administrator level is identical to the operator level, see "Operator level", Page 36 and the following.

10.7.2 Enable level

In this level, the administrator can define which parameters can be altered/ edited by the operator in the operator level. The available options are READ ONLY and EDIT. The structure “Parameter level” within the administrator level is identical to the operator level, see "Operator level", Page 36 and the following.

10.7.3 Calibration enable (CALIB. ENABLE)

In this level, the administrator can define whether the operator can access
- the relative cell constant
- the linear temperature coefficients
- the non-linear temperature coefficients
for calibration i.e. alteration.
45
10 Operation
PGM

10.8 Calibration level

All the calibrations that have been enabled by the administrator (administrator level) can be carried out in this level.
Press the key for at least 3 seconds.Use the or key to select CALIBRATION LEVEL.

10.8.1 REL. CELL CONSTANT (relative cell constant)

If this function has been enabled by the administrator, then the operator can calibrate the relative cell constant of the device here; see "Calibrating the relative cell constant", Page 51.

10.8.2 TEMPCO LINEAR (linear temperature coefficient)

If this function has been enabled by the administrator, then the operator can calibrate the device for liquids with a linear temperature coefficient; see "Linear temperature coefficient (ALPHA)", Page 52.

10.8.3 TEMPCO NON-LIN. (non-linear temperature coefficient)

If this function has been enabled by the administrator, then the operator can calibrate the device for liquids with a non-linear temperature coefficient; see "Non-linear temperature coefficient (ALPHA)", Page 55.
46

10.9 The dilution function

EXIT
10 Operation
Brief description
Addition of biocide
Dilution before biocide addition
For cooling water, the conductivity is used to deduce the total salt content. If a conductivity limit is reached (at the maximum permissible salt content/ concentration), then the cooling water must be diluted. A dilution valve is opened, the concentrated water flows out, and is replaced by fresh water. When the conductivity of the cooling water has fallen below the limit, the dilution valve is closed again.
A biocide is added to the cooling water, to prevent biological growth in the cooling system. There is no ideal setting for the amount used and the timing of the biocide dosing. In most cases, the dosing time is used as the controlled variable. The dosing quantity is therefore defined by the pumping rate and duration (system-specific). The success of the biocide treatment must be checked at regular intervals.
If a biocide that increases conductivity is added to the cooling water, this could increase the conductivity to beyond the limit. This would cause the dilution valve to be opened, and a portion of the added biocide would be discharged into the waste water (possibly contravening regulations!).
To prevent this, the conductivity in the cooling system is reduced by dilution, to, for example, 10% below the limit, before the biocide is added. The dilution valve is then temporarily blocked.
Dilution inhibit After adding the biocide, the dilution should be inhibited for a while, until the
biocide that is present in the cooling system is mostly decomposed (observe the statutory regulations!).
Implementation - The dilution function is only available in the “Conductivity measurement”
mode – not for concentration measurement.
- When the dilution function is activated, all the parameters that are irrelevant for this function are switched off.
- The dilution function can be started through binary input 1 and stopped through binary input 2, see "BINARY INPUT 1 and BINARY INPUT 2", Page
43. The dilution function can also be stopped by using the key.
- The present status of the dilution function will be shown in the display.
- The dilution valve is controlled by output K1.
- The addition of biocide valve is controlled by output K2.
- After dilution, K1 goes to the configured hold state (dilution inhibit).
47
10 Operation
PGM
PGM
PGM
- The dilution factor can be adjusted through binary input 1, over a range 1 — 50% below the limit value. The preset value is 10% below the limit.

10.9.1 Stop dilution

All the parameters are system-dependent, and must be adjusted to suit system requirements.
Press the key for at least 3 seconds.Use the or key to select OPERATOR LEVEL; use the key to
confirm the selection.
Use the or key to select BINARY INPUT; use the key to confirm
the selection.
48
10 Operation
PGM
EXIT
PGM
PGM
PGM
PGM
Use the or key to select DILUTION; use the key to confirm the
selection.
Change to the operator level, using the key.Use the key to select DILUTION.
Confirm the selection with the key.
Use the and/or keys to set the dilution factor in the range from
1 — 10 — 50% below the limit value.
Confirm the selection with the key.Use the or key to select DOSING TIME; use the key to confirm
the selection.
Set the dosing time with the and keys in the range from
0:00:00 — 00:01:00— 18:00:00 H:M:S.
Confirm the setting with the key.
49
10 Operation
PGM
PGM
Use the or key to select LOCK TIME; use the key to confirm the
Set the lock time with the and keys in the range from
Confirm the setting with the key.
selection.
0:00:00 — 00:01:00— 18:00:00 H:M:S.
If there is an interruption in the supply voltage during dilution, the function will be canceled.
The dilution function will have to be restarted if it is to be continued.
50

11.1 General

PGM
PGM
The device offers various calibration options to increase the precision.
The conductivity sensor should be cleaned and calibrated at regular intervals (depending on the medium being measured).
The K1 LED blinks during calibration.

11.2 Calibrating the relative cell constant

In order to meet enhanced demands for precision, the cell constant must first be calibrated.
Requirements - The supply voltage for the device must be present.
see Chapter 7 "Installation", Page 25ff.
- The sensor must be connected to the transmitter (applies to the split version).
- The transmitter is in the measurement mode.

11 Calibration

Immerse the conductivity sensor in a reference solution with a known
conductivity.
The temperature of the sample solution must remain constant during calibration!
Press the key for at least 3 seconds.Use the and keys to select CALIBRATION LEVEL; use the key to
confirm the selection.
51
11 Calibration
PGM
PGM
PGM
PGM
EXIT
EXIT
Use the and keys to select REL. CELL CONSTANT; use the key
to confirm the selection.
When the measurement is stable, press the key.Use the and keys to correct the indicated uncompensated
conductivity to match the known value for the reference solution.
Press .
The relative cell constant calculated by the device is displayed.
To accept the relative cell constant that has been determined -> press the
key for at least 3 seconds or
to reject the value -> press the key.
The transmitter is in the calibration menu. Press the key;
The transmitter is now in the measurement mode, and shows the compensated conductivity of the reference solution.

11.3 Calibrating the temp. coefficient of the sample solution

11.3.1 Linear temperature coefficient (ALPHA)

The conductivity of any sample solution will change according to its individual temperature coefficient. We therefore recommend carrying out a calibration of the temperature coefficient.
Requirements - The supply voltage for the device must be present.
see Chapter 7 "Installation", Page 25ff.
52
11 Calibration
PGM
PGM
PGM
PGM
- The sensor must be connected to the transmitter (applies to the split version).
- The transmitter is in the measurement mode.
Immerse the conductivity sensor in a sample of the solution to be
measured.
Press the key for at least 3 seconds.Use the and keys to select CALIBRATION LEVEL; use the key to
confirm the selection.
Use the and keys to select TEMPCO LINEAR; use the key to
confirm the selection.
Use the and keys to enter the working temperature; confirm with the
key.
The working temperature must be at least 5°C above or below the reference temperature (25.0°C).
53
11 Calibration
PGM
EXIT
EXIT
(1)
(2)
(3)
The LC display now shows
- at top (1): the selected working temperature (blinking)
- in the middle (2): the reference temperature (blinking)
- below (3): the present sensor temperature (steady)
Warm up the sample medium until both the reference and the working
temperatures have been reached (the corresponding values no longer blink).
During calibration, the rate of change of temperature for the sample solution must not exceed 10°C/min for the device with exposed temperature sensor, or 1°C/min for the device with an internal temperature sensor.
As soon as one of the target temperatures has been reached, its display becomes static (no longer blinking).
Calibration can also be carried out through a cooling procedure (falling temperature). In this case, it starts above the working temperature and finishes below the reference temperature.
The LC display now shows the derived temperature coefficient in %/°C. To accept the temperature coefficient that has been determined -> press the
key for at least 3 seconds or
to reject the value -> press the key. The transmitter is in the calibration menu. Press the key.
The transmitter is now in the measurement mode, and shows the compensated conductivity of the reference solution.
54
11 Calibration

11.3.2 Non-linear temperature coefficient (ALPHA)

General Since the temperature coefficient of some media is not constant over a
sizeable temperature range, the device provides the option of subdividing a
Determining the TC curve
temperature range (T used for compensation in each of these range sections. This “TC curve” can be
- edited with the setup program and transmitted to the device.
- or calibration can be performed automatically on the device.
Start
to T
) into 5 sections. A different TC value can be
End
Calculation of a temperature coefficient
α = temperature coefficient (TC) γ = uncompensated conductivity
55
11 Calibration
TC curve
Temperature compensation with the TC curve
Sequence for automatic calibration
The present temperature of the medium is applied to the TC curve to determine the corresponding temperature coefficient, see "TC curve", Page 56.
Intermediate values, e.g. (α
at T4) are derived through a linear interpolation.
(α
4
The derived TC is used to calculate the compensated conductivity, in the same way as with the linear compensation.
If the measured temperature is lower than the start temperature, the first TC is used for compensation. If the measured temperature is higher than the end temperature, the last TC is used for compensation.
The TC curve is automatically recorded over a temperature range that has been defined by the user. The temperature range between the start and end temperatures is subdivided into 5 sections of equal size.
The temperature range must be larger than 20°C, and cover the reference temperature. Example: Reference temperature 25°C, start temperature 18°C and end temperature 50°C.
at Tx) between two known values (α3 at T3) and
x
56
11 Calibration
PGM
PGM
PGM
The rate of change of the temperature must not exceed
- 10°C / min. for an exposed temperature sensor, and
- 01°C / min for an internal temperature sensor.
Requirements - The supply voltage for the device must be present.
see Chapter 7 "Installation", Page 25ff.
- The sensor must be connected to the transmitter (applies to the split version).
- The transmitter is in the measurement mode.
Immerse the conductivity sensor in a sample of the solution to be
measured.
Press the key for at least 3 seconds.Use the and keys to select CALIBRATION LEVEL; use the key to
confirm the selection.
Use the or key to select TEMPCO NON-LIN.; use the key to
confirm the selection.
57
11 Calibration
PGM
PGM
Use the and keys to enter the start temperature; confirm with the
Use the and keys to enter the end temperature; confirm with the
key.
key.
The start temperature must be lower than the reference temperature (25.0°C).
(1)
(2)
The end temperature must be at least 20°C above the start temperature.
The transmitter will define the fixed temperature points itself. The LC display now shows
- at top (1): the next target temperature (blinking)
- below (2): the present sensor temperature (steady)
Warm up the sample medium until is it above/below the temperature that is
blinking.
58
11 Calibration
PGM
EXIT
EXIT
The next target temperature is displayed as blinking.
During calibration, the rate of change of temperature for the sample solution must not exceed 10°C/min for the device with exposed temperature sensor, or 1°C/min for the device with an internal temperature sensor.
As soon as one of the target temperatures has been reached, its display becomes static (no longer blinking).
Warm up the sample medium until is it above the temperature that is
blinking.
Repeat the procedure as often as required, until the device has determined
all 6 temperature coefficients.
The LC display now shows the derived temperature coefficients in %/°C. To accept the temperature coefficients that have been determined -> press
the key for at least 3 seconds or to reject the values -> press the key.
The transmitter is in the calibration menu. Press the key.
The transmitter is now in the measurement mode, and shows the compensated conductivity of the reference solution.
59

12 Maintenance

12.1 Cleaning the conductivity sensor

Do not use solvents. Hard-to-remove crusts and deposits can be softened and removed
with dilute hydrochloric acid.
Observe the safety regulations!
Deposits Deposits on the sensor section can be removed with a soft brush (e.g. a bottle
brush).
60

13 Eliminating faults and malfunctions

Error possibilities
t
Problem Possible cause Measures
no measurement display or current output
measurement display 000 or current output 4 mA
wrong or unstable measurement display
supply voltage missing
sensor not immersed in medium, reservoir level too low
flow-through fitting is blocked
sensor is faulty see "Checking the
sensor not immersed deeply enough
inadequate mixing ensure good
supply voltage should be checked, also check terminals
top up the reservoir
flow-through fitting should be cleaned
device", Page 61 top up the reservoir
mixing for sensor: all-round free space of approx. 5mm ensure all-round flow
Measurement value dis­play 8888, temperature display "ok", blinking
Measurement value dis­play 8888, temperature display 8888 blinking
air bubbles check mounting site,
see "General", Page 18.
Conductivity measure­ment range overshoo­ting or conductivity mea­surement probe faulty.
Temperature measure­ment range overshoo­ting or undershooting or short circuit or interrupti­on of the temperature sensor.
Select suitable measure­ment range.
Replace conductivity measuring transmitter.
The temperature of the measuring medium must be in the range from
0...150°C. Replace the conductivity
measuring transmitter. Send the device for re-
pairs.

13.1 Checking the device

General The device is calibrated at the factory, and is maintenance-free. If,
nevertheless, measurement deviations appear with no apparent cause, the transmitter can be tested as follows.
61
13 Eliminating faults and malfunctions
R

13.1.1 Resistance loop test

Position of the resistance loop
During calibration, do not touch the sensitive part of the cell or put it down on any surface, otherwise the measurement will be falsified.
Lead a wire through the cell (see diagram)Connect a resistor R to the wire
Calculating the resistance
Example 1 The measuring transmitter is to show 20 mS:
Formula for calculating the resistance of the resistance loop:
2
· K
N
R =
Lf N = No. of turns in the loop
Note: 1 mS/cm = 1·10
1 µS/cm = 1·10
For display values up to 49 mS, the loop must have 1 turn. For display values above 50 mS, the loop must have 3 turns.
The cell constant of the device is 6.25 1/cm.
2
1
R =
·6.25 1/cm
20·10-3 S/cm
= 312.5
Ω
R = Resistance of the resistance loop
K = Cell constant Lf = Required display in S/cm
-3
S/cm
-6
S/cm
62
To achieve a display of 20 mS/cm, the resistance loop (with 1 turn) must have a resistance of 312.5 Ohm.
13 Eliminating faults and malfunctions
Example 2 The measuring transmitter is to show 500 mS:
2
3
R =
To achieve a display of 500 mS/cm, the resistance loop (with 3 turns) must have a resistance of 112.5 Ohm.
Precalculated values
·6.25 1/cm = 112.5
500·10-3 S/cm
Required display
0 µS/cm 0 no resistance
625 µS/cm 1 10000
1000 µS/cm 1 6250 Ω 2000 µS/cm 1 3125 Ω 5000 µS/cm 1 1250 Ω
10 mS/cm 1 625 Ω 20 mS/cm 1 312.5 Ω
50 mS/cm 3 1125 Ω 100 mS/cm 3 562.5 Ω 200 mS/cm 3 281.3 Ω 500 mS/cm 3 112.5 Ω
1000 mS/cm 3 56.3 Ω 2000 mS/cm 3 28.1 Ω
Ω
Number of
turns
Required resistance
Ω
Te st sequence
Calculate the test resistance.Wire up the device, see Chapter 7 "Installation", Page 25. Select the corresponding measurement range, see Chapter 10.6.1
"CONDUCTIVITY IN (Conductivity input)", Page 37 -> RANGE 1 — 4
Set TC to 0%/°C, see Chapter 10.6.1 "CONDUCTIVITY IN (Conductivity
input)", Page 37 -> TEMPCO
Fit the resistance loop as shown in the diagram.
63
13 Eliminating faults and malfunctions

13.1.2 Reference liquid test

Immerse in test solution
Te st sequence
Prepare the conductivity test solution in a container of adequate size.Wire up the device, see Chapter 7 "Installation", Page 25.Select the range appropriate to the conductivity test solution, see Chapter
10.6.1 "CONDUCTIVITY IN (Conductivity input)", Page 37 -> RANGE 1 — 4
Set TC to 0%/°C, see Chapter 10.6.1 "CONDUCTIVITY IN (Conductivity
input)", Page 37 -> TEMPCO
Immerse the cell in the container, and do not move it any more during the
measurement.
64
13 Eliminating faults and malfunctions

13.1.3 Reference measuring instrument test

Immerse in test solution
Te st sequence
Prepare the conductivity test solution in a container of adequate size.Wire up the device, see Chapter 7 "Installation", Page 25.Select the range appropriate to the conductivity test solution, see Chapter
10.6.1 "CONDUCTIVITY IN (Conductivity input)", Page 37 -> RANGE 1 — 4
Set TC to 0%/°C, see Chapter 10.6.1 "CONDUCTIVITY IN (Conductivity
input)", Page 37 -> TEMPCO
Set the TC for the reference instrument to 0%/°C as well (see operating
instructions for the reference instrument). If this is not possible, then the sample liquid must be tempered to the reference temperature for the reference instrument.
Immerse the cell under test and the cell for the reference instrument in the
container, and do not move them any more during the measurement.
The output and display of the device under test or the attached indicator
must match the indication of the reference instrument, taking into account acceptable device deviations.
65

14 Appendix

14.1 Before configuration

If a number of instrument parameters have to be modified in the instrument, then it is advisable to note them in the table below, and then modify these parameters in the sequence given.
The following list shows the maximum number of parameters that can be altered.
Depending on the configuration, some of the parameters will not be alterable (editable) for your device.
Parameter Selection / value range
Factory setting
Conductivity input
Range 1 0 — 500 µS/cm
0 — 1000 µS/cm 0 — 2000 µS/cm 0 — 5000 µS/cm 0 — 10 mS/cm 0 — 20 mS/cm 0 — 50 mS/cm 0 — 100 mS/cm 0 — 200 mS/cm 0 — 500 mS/cm 0 — 1000 mS/cm
0 — 2000 mS/cm (uncompensated) Temperature compensation
Temperature coefficient 1 0.0 to 5.5%/°C 37 Reference temperature 15.0 to 25.0 to 30°C 37 Relative cell constant 80.0 — 100.0 — 120.0% 37 Mounting factor 80.0 — 100.0 —120.0% 37 Concentration
measurement
Offset -200 to 0 to +200 mS/cm 38 Filter time 00:00:00 — 00:00:25 H:M:S 38 Calibration interval 0 — 999 days 38
Conductivity output
Signal type 0 — 20 mA
linear
non-linear
natural water
No function
NaOH
HNO
3
customer-specific
4 — 20 mA
20 — 0 mA
20 — 4 mA
0 — 10 V
2 — 10 V
10 — 0 V
10 — 2 V
New
adjustment
see page
37
37
37
38
66
14 Appendix
Parameter Selection / value range
Factory setting
Scaling start 0 — 90% = 4 mA (e.g.)
of range span Scaling end 100 — 10% = 20 mA (e.g.)
of range span During alarm low
high
safe value During calibration moving
frozen
safe value Safe value 0.0 — 4.0 — 22.0 mA 39 Manual mode off
on Manual value 0.0 — 4.0 — 22.0 mA 39
Temperature input
Unit °C
°F Measurement mode Sensor
manual Manual value -20.0 to 25 to 150°C 39 Offset -15.0 to 0.0 to +15°C 39 Filter time 00:00:00 — 00:00:01 — 00:00:25
H:M:S
Temperature output
Signal type 0 — 20 mA
4 — 20 mA
20 — 0 mA
20 — 4 mA
0 — 10 V
2 — 10 V
10 — 0 V
10 — 2 V Scaling start -20.0 to 183°C = 4 mA
(0 — 90% of range span) Scaling end -3 to 150 to 200°C = 20 mA
(100 — 10% of range span) During alarm low
high
safe value
New
adjustment
see page
38
38
39
39
39
39
39
39
38
38
38
38
67
14 Appendix
Parameter Selection / value range
Factory setting
During calibration moving
frozen
safe value Safe value 0.0 — 4.0 — 22.0 mA 38 Manual mode off
on Manual value 0.0 — 4.0 — 22.0 mA 38
Binary output 1 or binary output 2
Function No function
Conductivity MIN contact
Conductivity MAX contact
Conductivity LK1
Conductivity LK2
Temperature MIN contact
Temperature MAX contact
Temperature LK1
Temperature LK2
Calibration timer
Alarm Limit value -20.0 — 9999.0 42 Hysteresis 0.0 — 0.5 — 999.0 42 Spacing 0.0 — 999.0 42 Manual mode off
on For hold inactive
active
frozen For alarm/calibration inactive
active
frozen Switch-on delay 00:00:00 — 01:00:00
H:M:S Switch-off delay 00:00:00 — 01:00:00
H:M:S Pulse duration 00:00:00 — 01:00:00
H:M:S
Binary input
Function No function
Key lock/hold
Meas. range/temperature
coefficient
Dilution function
Dilution function
Reduction 0 — 10 — 50% 43 Dosing time 00:00:00 — 00:01:00 — 18:00:00
H:M:S Lock time 00:00:00 — 00:01:00 — 18:00:00
H:M:S
New
adjustment
see page
38
38
41
42
42
42
42
42
42
43
43
43
68
14 Appendix
Parameter Selection / value range
Factory setting Device data
Language German
English
French
Italian
Dutch
Polish
Portuguese
Russian
Swedish
Spanish Contrast 0 — 6 — 11 44 Lighting off
on
during operation
LCD inverse off
on
New
adjustment
see page
44
44
44
69

15 China RoHS

żżżżżż
żżżżżż
;żżżżż
;
ℶ❐兓Ⓔ
3URGXFWJURXS
ℶ❐₼㦘⹂䓸德䤓⚜䱿♙⚺摞
&KLQD((3+D]DUGRXV6XEVWDQFHV,QIRUPDWLRQ
捷ↅ⚜䱿
&RPSRQHQW1DPH
3E
+J
&G
⏼ↆ杻
&U9,
⮩䅃勣啾 ᧤3%%
⮩䅃ℛ啾搩
3%'(
ཆ༣
+RXVLQJ
*HKlXVH
䗷〻䘎᧕
3URFHVVFRQQHFWLRQ 3UR]HVVDQVFKOXVV
㷪⇽
1XWV
0XWWHU
㷪ṃ
6FUHZ
6FKUDXEH
żżżżż
㦻嫷㫋∬㗽^:dϭϭϯϲϰ䤓屓⸩冥Ⓟᇭ
7KLVWDEOHLVSUHSDUHGLQDFFRUGDQFHZLWKWKHSURYLVLRQV6-7 ż᧶嫷䯉年㦘⹂䓸德⦷年捷ↅ㓏㦘⧖德㧟㠨₼䤓⚺摞⧖⦷*%7屓⸩䤓棟摞尐㻑ⅴₚᇭ ,QGLFDWHWKHKD]DUGRXVVXEVWDQFHVLQDOOKRPRJHQHRXVPDWHULDOV¶IRUWKHSDUWLVEHORZWKHOLPLWRIWKH *%7
î᧶嫷䯉年㦘⹂䓸德咂⺠⦷年捷ↅ䤓㩟₏⧖德㧟㠨₼䤓⚺摞怔⒉*%7屓⸩䤓棟摞尐㻑ᇭ ,QGLFDWHWKHKD]DUGRXVVXEVWDQFHVLQDWOHDVWRQHKRPRJHQHRXVPDWHULDOV¶RIWKHSDUWLVH[FHHGHGWKH OLPLWRIWKH*%7
70
JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc.
Street address: Moritz-Juchheim-Straße 1 36039 Fulda, Germany Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net
JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk
6733 Myers Road East Syracuse, NY 13057, USA
Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com
Loading...