JUMO ATH-EXx Operating Instructions Manual

ATH-EXx
Explosion-protected surface-mounted thermostat
for potentially explosive gas atmosphere Zone 1 and
potentially explosive dust atmosphere Zone 21
B 605051.0
Operating Instructions
2011-04-01/00379556
Please read these Operating Instructions before commissioning the instrument. Keep the manual in a place that is accessible to all users at all times. Please assist us to improve these operating instructions, where necessary. Your sug­gestions will be appreciated.
Phone+49 661 6003-0 Fax +49 661 6003-607
All necessary settings and, where appropriate, alterations inside the instrument are described in these operating instructions. If any difficulties should still arise during commissioning, you are asked not to carry out any unauthorized manip­ulations on the unit. You could endanger your rights under the instrument war­ranty! Please contact the nearest subsidiary or the head office in such a case.
Contents
Seite
1 Introduction ............................................................................... 5
1.1 Typographical conventions ...................................................................... 5
1.1.1 Warning signs .............................................................................................. 5
1.1.2 Note signs ................................................................................................... 5
1.2 Application ................................................................................................. 6
1.3 Ex marking ................................................................................................. 6
1.4 Safety notes ............................................................................................... 6
2 Instrument identification .......................................................... 7
2.1 Nameplate .................................................................................................. 7
2.2 Type designation ....................................................................................... 8
3 Mounting .................................................................................... 9
3.1 Dimensions ................................................................................................ 9
3.2 Opening the surface-mounting thermostat .......................................... 10
3.3 Fixing the surface-mounting thermostat .............................................. 10
3.4 Capillary / temperature probe / pocket ................................................. 10
3.4.1 General ...................................................................................................... 10
3.4.2 Approved probes and sheaths .................................................................. 10
3.5 Permissible operating conditions at the pocket .................................. 11
3.5.1 Probe mountings 20 and 22 ...................................................................... 11
3.5.2 Probe mountings 10, 15, 21, 60, 65 .......................................................... 14
3.6 Mounting the probe ................................................................................. 14
4 Installation ............................................................................... 15
4.1 Regulations and notes ............................................................................ 15
4.2 Electrical connection .............................................................................. 16
4.3 Connection diagrams ............................................................................. 16
5 Settings .................................................................................... 17
5.1 Limit setting ............................................................................................. 17
5.1.1 TW, STW, TB (code 7) ............................................................................... 17
5.1.2 STB (code 70) ........................................................................................... 17
5.1.3 TB, STB (code 7-F or 70-F) ....................................................................... 18
5.2 Resetting the TB or STB limiter ............................................................. 19
5.3 Self-monitoring (fail-safe) ....................................................................... 19
5.3.1 Response to a fracture of the measuring system ..................................... 19
5.3.2 Response to low temperature ................................................................... 19
6 Instrument description ........................................................... 20
6.1 Technical data ......................................................................................... 20
7 Appendix .................................................................................. 24
5
1 Introduction
1.1 Typographical conventions
1.1.1 Warning signs
1.1.2 Note signs
abc
1
Footnote
Footnotes are remarks that refer to specific points in the text. Footnotes consist of two parts:
A marker in the text, and the footnote text. The markers in the text are arranged as continuous superscript numbers. The footnote text (in smaller typeface) is placed at the bottom of the page and
starts with a superscript number.
Action
This symbol indicates that an action to be performed is described. The individual steps are marked by this asterisk, e.g. Select setpoint
Danger
This symbol is used when there may be danger to personnel if the instruc­tions are ignored or not followed correctly!
Caution
This symbol is used when there may be damage to equipment if the instruc- tions are ignored or not followed correctly!
Note This symbol is used when your special attention is drawn to a remark.
Reference This symbol refers to further information in other chapters or sections.
1 Introduction
6
1.2 Application
Depending on the version, the explosion-protected surface-mounting thermostat type ATH-EXx can be used as a
TW temperature monitor
TB temperature limiter
STW fail-safe temperature monitor
STB break-proof protection temperature limiter
in areas with an explosion hazard caused by gases, vapors or mists, Zone 1 or in areas with combustible dust or dust/air mixtures, Zone 21.
1.3 Ex marking
II 2G Ex ed IIC T5 bzw. T6
for potentially explosive gas atmospheres
II 2D Ex tD A21 IP65 T95°C bzw.T80°C
for potentially explosive dust atmospheres
Test certificate: PTB 03 ATEX 1187
CE marking and No. of notified body: CE 0044.
1.4 Safety notes
Physical and toxicological properties of the substances that may escape in the event of a system fracture.
– = no; X = yes
If this surface-mounting thermostat is used inside the hazardous area, the relevant regulations for mounting and operating equipment in such an area must be observed.
Cutting through or kinking the capillary of the surface-mounting thermostat, type ATH-EXx, will result in permanent failure of the instrument!
Only use the ATH-EXx with TZ 679 in areas that are classified as areas with low mechanical risk.
Filling liquid may escape in the event of a measuring system fracture. At present there is no restrictive statement from the health authorities concerning any danger to health over short periods and at low concentration, e.g. after a fracture of the measuring system.
End of scale°CDanger-
ous
reaction
Fire and
explosion hazard
Water
conta-
mination
Toxicological data
Ignition
temp.
°C
Explosion
limit
% v/v
irritant danger to
health
toxic
< +200°C +355°C 0,6 - 8 V X X X
+200°C
+350°C
+490°C - - X X X
> 350°C
+500°C
– – – ––––
7
2 Instrument identification
2.1 Nameplate
( 1 ) Type Chapter 2.2 "Type designation", page 8 ( 2 ) Type code ( 3 ) Control range or limit value range / ambient temperature at which this
thermostat was calibrated (option) ( 4 ) Ex-protection marking ( 5 ) Testing laboratory / certification ( 6 ) Switching capacity ( 7 ) Permissible ambient temperature ( 8 ) Serial number ( 9 ) Year of production (10) Week of production (11) Running serial device number (12) Extra code
( 4 )
( 3 )
( 2 )
( 1 )
( 5 )
( 6 )
( 8 )
( 9 )
(10)
(11)
( 7 )
( 12 )
2 Instrument identification
8
2.2 Type designation
ATH- EX x- . . - . ATH Surface-mounting thermostat (with capillary)
ATH- EXx Marking for explosion-protected version ATH- EX x- ..-. Code number for the function:
2=0Temperature monitor (TW)
with changeover contact. Limit adjustable within the housing.
7=0Temperature limiter (TB)
with changeover contact and restart lock-out. Limit adjustable within the housing.
7-F =0Temperature limiter (TB)
with changeover contact and restart lock-out. Limit is factory-set.
20 =0Fail-safe temperature monitor (STW)
with changeover contact. Limit adjustable within the housing.
70 =0Break-proof protection temperature limiter (STB)
with change-over contact and restart inhibit. The limit value can be set by the installing engineer– before commissioning the plant – within the housing.
70-F =0Break-proof protection temperature limiter (STB)
with change-over contact and restart inhibit. Limit value set to a fixed value at the factory.
ATH- EX x- . . - .F=0Designation for factory-set limit setting.
Extra code (TZ) 679: Min. operating temperature: -50°C
9
3 Mounting
3.1 Dimensions
90
d
122
120
82
6.4
10.4
106
L
29
25
( 1 )
( 2 )
( 1.1 )
( 1.2 )
( 7 )
( 5 )
( 6 )
( 4 )
( 3 )
(1) Ex cable gland M 20 x 1.5 x 6-12 (1.1) Screw-in thread Tightening torque: 3,75 Nm (1.2) Compression screw Tightening torque: for cable dia. 06mm=3.5Nm
Tightening torque: for cable dia. 12 mm = 2.5 Nm (2) Cover screws (3) Cover (4) Housing base (5) Drilling diagram (6) Ex cable gland for sealing the capillary exit (7) Temperature probe
3 Mounting
10
3.2 Opening the surface-mounting thermostat
Unscrew the 4 cover screws (2)Remove cover (3) with seal (not visible)
3.3 Fixing the surface-mounting thermostat
Mounting position
To DIN 16257, NL 0 — NL 90 (other NL on request)
Drill fixing holes according to the drilling diagram.Use the 4 screws to secure the housing base.
3.4 Capillary / temperature probe / pocket
3.4.1 General
3.4.2 Approved probes and sheaths
see current data sheet 606710
Make sure that the seal is seated correctly when reassembling!
Cutting through or kinking the capillary of the surface-mounting thermostat will lead to permanent instrument failure!
Minimum permissible bending radius of the capillary is 5 mm. The temperature probe must be mounted in a JUMO pocket, otherwise the
approval of the surface-mounting thermostat becomes invalid.
The temperature probe must completely immersed in the medium to be measured.
To ensure their overall accuracy, the thermostats must only be used together with the pockets supplied by the factory (diameter D = 8, 10 or 15 mm).
Pockets with a diameter D = 10 mm must only be fitted with probes with diameter d = 8 mm.
Fitting several probes into a common pocket is only permitted with 2 or 3 probes (mounting type "10", diameter d = 6 mm) and 15 x 0.75 mm pockets. When fitting 2 probes in a common pocket, the factory-supplied spring clip must be fitted in the pocket.
For operation in air, probe mounting type "10" (without pocket) must be cho­sen.
11
3 Mounting
3.5 Permissible operating conditions at the pocket
3.5.1 Probe mountings 20 and 22
3.5.1.1 Steel pockets
Materials Tube: St 35.8 I
Weld-in nipple: 16 Mo 3 (no turned groove)
Loading
The values given below refer to the maximum loading on the probe mounting concerned. The maximum pressure which can be sealed depends on the mounting conditions and may possibly be lower.
Temperature Tube diameter
8 x 0.75 mm or
conical
10 x 0.75 mm 15 x 0.75 mm
max. permissible
pressure
max. permissi-
ble pressure
max. permissi-
ble pressure 100°C 89 bar 72 bar 48 bar 150°C 83 bar 67 bar 45 bar 200°C 78 bar 63 bar 42 bar 300°C 59 bar 47 bar 32 bar 400°C 46 bar 37 bar 25 bar 450°C 24 bar 19 bar 13 bar
With operating temperatures above 420°C, the permissible operating time is limited to 200,000 hours. TRD 508 must be observed in this respect.
3 Mounting
12
Permissible incident flow velocity
Permissible incident flow velocity “v” at the maximum permissible pressure loading and different immersion tube lengths “S”
Permissible incident flow velocity “v” at the maximum permissible pressure loading and different immersion tube temperatures “t”
Temperature: +200°C
Thermal medium:
air (1) water, oil (2)
Tube diameter D:
08mm
10 mm 15 mm
13
3 Mounting
3.5.1.2 Stainless steel pocket
Materials Tube: X 6 CrNiMoTi 17122
Screw-in/weld-in nipple: X 6 CrNiMoTi 17122
Loading
Permissible incident flow velocity
on request
3.5.1.3 Brass pocket
Material CuZn (nickel-plated)
Loading
Permissible incident flow velocity
on request
Temperature Tube diameter
8 x 0.75 mm or
conical
10 x 0.75 mm 15 x 0.75 mm
max. permissible
pressure
max. permissi-
ble pressure
max. permissi-
ble pressure 100°C 92 bar 74 bar 50 bar 150°C 88 bar 71 bar 48 bar 200°C 83 bar 67 bar 45 bar 300°C 72 bar 58 bar 39 bar 400°C 67 bar 54 bar 36 bar
Temperature Tube diameter
8 x 0.75 mm 10 x 0.75 mm 15 x 0.75 mm
max. permissible
pressure
max. permissi-
ble pressure
max. permissi-
ble pressure 100°C 50 bar 40 bar 27 bar 150°C 48 bar 39 bar 26 bar
3 Mounting
14
3.5.2 Probe mountings 10, 15, 21, 60, 65
Materials Choice of steel, stainless steel or brass
Loading Only for use in unpressurized media.
*
depending on the control range
3.6 Mounting the probe
Probe mounting 15 60, 65 10, 21 Maximum temperature +55°C +200°C
*
+500°C
*
The temperature probe (2) must be immersed in the medium for its entire length, otherwise there will be appreciable deviations from the switching point.
In the case of probe mountings 20, 21 and 23, the temperature probe is secured in the pocket by a clamping clip (1).
( 1 )
( 2 )
15
4 Installation
4.1 Regulations and notes
The electrical connection must only be carried out by qualified personnel.
The choice of cable, the installation and the electrical connection must
conform to the requirements of VDE 0100 “Regulations on the Installation of Power Circuits with Nominal Voltages below 1000 V” or the appropriate local regulations.
If contact with live parts is possible while working on the instrument, it must be completely disconnected from the electrical supply.
Earth the instrument at the PE terminal to the protective earth conductor. This cable must have at least the same cross-section as used for the supply cables. Earthing cables must be wired in a star configuration to a common earth point that is connected to the protective earth conductor of the electrical supply. Do not loop earthing cables, i.e. do not run them from one instrument to another.
Apart from faulty installation, incorrect settings on the thermostat can also affect the proper functioning of the subsequent process or lead to damage. Setting up must therefore be restricted to qualified personnel. Please observe the relevant safety regulations for such matters.
In case of electrical connections in a potentially explosive area, it is necessary to comply with the relevant specifications.
4 Installation
16
4.2 Electrical connection
Open the housing.
"Opening the surface-mounting thermostat", page 9
Pass the connection cable (cable diameter 6 to 12 mm) through the Ex
cable compression gland (1).
"Dimensions", page 8
Make the connection to the terminals (2) in accordance with the connection
diagram.
Fitting type X (no special tools).The connecting cable must have a fixed installation
Position the connecting cable in the housing and tighten the Ex cable com-
pression gland (1), using the tightening torque given below: "Dimensions", page 8.
4.3 Connection diagrams
Connect the protective earth conductor to the PE terminal.
Screw-in thread (1.1) into housing = 3.75 Nm Compression screw (1.2). = 3.50 Nm for 06 mm cable diameter Compression screw (1.2). = 2.50 Nm for 12 mm cable diameter
Ensure that connection line for operating temperatures up to -50°C are low-temperature resistant.
TW / STW
TB / STB
2
4L12
4
1
PE N
12
4
42
L1
N
N
PE
PE
*
1
42
142
(2)
NPE
1
4
2
4L12
17
5 Settings
5.1 Limit setting
5.1.1 TW, STW, TB (code 7)
Open the housing.
"Opening the surface-mounting thermostat", page 9
Set the limit on the setpoint spindle (1).
5.1.2 STB (code 70)
Setting the limit according to the scale
Open the housing.
"Opening the surface-mounting thermostat", page 9
Set the limit on the setpoint spindle (1) against the internal scale.Secure this setting by sealing the setpoint spindle ( 1 ), for instance with
“red” thread sealing wax, temperature resistant to at least +80°C)
05
05
°C
( 1 )
Important note for setup and operation!
When the thermostat is used as a safety device for explosion protection in accordance with EU Directive 94/9/EC Annex II Sec. 1.5, a functional test is required that conforms to the applicable regulations.
The switching point must be fixed by the installation engineer after thermally testing individual items, and secured against alteration. Several aspects must be observed:
- the probe geometry and thermal coupling
- the maximum ambient temperature
- the maximum product temperature
5 Settings
18
Setting the lim­its according to specific system characteristic
Open the housing. "Opening the surface-mounting thermostat", page 9Warm up the temperature probe – in the system – to the intended limit
temperature (allow at least 5 minutes for the temperature to equalize), while using a calibrated reference measuring device to measure and observe the precise temperature at the temperature probe.
Determine the required switching position (circuit 1-4 opens and circuit 1-2
closes) by rotating the setpoint spindle ( 1 ) from the end of the scale in the direction of the start of the scale.
Secure this setting by sealing the setpoint spindle ( 1 ), for instance with
“red” thread sealing wax, temperature resistant to at least +80°C.
A = secure the limit setting by sealing the setpoint spindle
with “red” thread sealing wax.
( 2 ) reset arm
5.1.3 TB, STB (code 7-F or 70-F)
(A)
(1)
(1)
(2)
Important note for setup and operation!
When the thermostat is used as a safety device for explosion protection in accordance with EU Directive 94/9/EC Annex II Sec. 1.5, a functional test is required that conforms to the applicable regulations.
In addition, suitable measures (such as calculation, thermal testing of individual items) must be used to document that the switching point set by the manufacturer prevents the temperature exceeding the given limit. The specific local and operating conditions must be taken into account, such as:
- the probe geometry and thermal coupling
- the maximum ambient temperature
- the maximum product temperature
- The limit setting is fixed at the factory.
- Alteration is not permitted.
19
5 Settings
5.2 Resetting the TB or STB limiter
If the temperature at the probe goes above the limit that has been set, the electrical circuit 1-4 is opened and remains mechanically locked out. When the critical temperature has fallen to about 9-15% below the set limit, the TB or STB can be reset manually.
Open the housing.
"Opening the surface-mounting thermostat", page 9
Press the reset arm ( 2 ) in the direction of the arrow.
"( 2 ) reset arm", page 18
5.3 Self-monitoring (fail-safe)
5.3.1 Response to a fracture of the measuring system
5.3.2 Response to low temperature
On the STW or STB, in the event of a fracture of the measuring system (leakage) the circuit 1-4 will remain permanently open.
On the STB, the microswitch is also mechanically and permanently locked out.
If the probe temperature of approx. -30°C (-60°C for the TZ 679) of the STW and STB is gone below, the power circuit is opened.
When the temperature rises above the minimum probe temperature, the STW resets itself automatically. The STB requires a manual reset.
"Resetting the TB or STB limiter", page 19
20
6 Instrument description
6.1 Technical data
Ex marking
II 2G Ex ed IIC T5 bzw. T6
for potentially explosive gas atmospheres
II 2D Ex tD A21 IP65 T95°C bzw. T80°C
for potentially explosive dust atmospheres
Explosion protection
II 2G Equipment group II, Category 2,
equipment for potentially explosive gas atmospheres
II 2D Equipment group II, Category 2,
equipment for potentially explosive dust atmospheres
Explosion protection type
Ex General requirements
European Standard EN 60079-0
e Increased safety
European Standard EN 60079-7
d Flameproof enclosure
European Standard EN 60079-1
IIC Gas group T5 or T6 Temperature class
Application in combustible dusts
European Standard EN 61241-0
EN 61241-1
IP65 Protection to EN 60529-IP65 T95°C or T80°C max. permissible surface temperature
(application in combustible dusts)
Te st certificate
PTB 03 ATEX 1187
21
6 Instrument description
Permissible ambient tem­perature in operation
Max operating temperature of the cable screw-connection for the TZ 679: 70°C.
Permissible probe tempera­ture
max. limit +15% (+550°C with limit > 450°C)
Permissible stor­age temperature
max. +50°C, min. -20°C max. +50°C, min. -50°C et TZ 679
Connecting cable
diameter 6 to 12 mm, conductor cross-section up to 2.5 mm
2
Switching dif­ferential
in % of control/ limit range
Contact rating
Fusing required see contact rating
Capillary
°C
Thermostat head °C
T6 T5
max. +40 +40
+55 +55
min. -20 (-50 et TZ 679)
Thermostats with end of scale
350°C
Thermostats with end of scale
350°C
Switching
action
Nominal
value
Possible
actual
value
Switching
action
Nominal
value
Possible
actual
value
TW 3 3-5 TW 5 5-9
STW 5 5-7 STW 6 6-11
TB / STB – – – – TB / STB – – – –
TW, STW
Terminals 1-4 (break) Terminals 1-2 (make) AC1: 230 V +10%, 10 A AC1: 230 V +10%, 5 A AC15: 230 V +10%, 2 A AC15: 230 V +10%, 0.8 A
DC1: 230 V +10%, 0.25 A
TB, STB
Terminals 1-4 (break) Terminals 1-2 (make) AC1: 230 V +10%, 16 A AC1: 230 V +10%, 10 A AC15: 230 V +10%, 2 A AC15: 230 V +10%, 1.5 A
DC1: 230 V +10%, 0.25 A
6 Instrument description
22
Enclosure pro­tection
EN 60 529 - IP 65
Operating medium
Water, oil, air, superheated steam
Time constant t
0.632
Mode of opera­tion
according to EN 60 730-1 TW,TB: Type 2BL = automatic action with micro-disconnection in
operation, no auxiliary supply required.
STW: Type 2BKL = automatic action with micro-disconnection in
operation, fail-safe.
STB: Type 2BFKL = automatic action with micro-disconnection in
operation, fail-safe, reset only permissible by means of a tool.
Operating posi­tion (NL)
To DIN 16257, NL 0 — NL 90 (other NL on request)
Weight approx. 1.2 kg
Capillary and probe material
Min. bending radius of the capillary
5mm
in water in oil in air / superheated
steam
45 s 60 s 120 s
End of scale Capillary Probe up to +200°C copper, Mat. Ref. 2.0090
electro-tinned
1.5 mm diameter
copper, Mat. Ref. 2.0090 electro-tinned brazed
up to +300°C copper, Mat. Ref. 2.0090
electro-tinned
1.5 mm diameter
stainless steel, Mat. Ref. 1.4571 brazed
up to +500°C stainless steel,
Mat. Ref. 1.4571
1.5 mm diameter
stainless steel, Mat. Ref. 1.4571 welded
up to +300°C stainless steel,
Mat. Ref. 1.4571
1.5 mm diameter
stainless steel, Mat. Ref. 1.4571 welded
23
6 Instrument description
Switching point accuracy
in % of scale span, referred to the limit value at Ta +22°C.
Mean ambient temperature effect
in % of scale span, referred to the limit value. A deviation of the ambient temperature at the thermostat head and/or the cap-
illary from the +22°C calibration ambient temperature produces a shift in the switching point: higher ambient temperature = lower switching point lower ambient temperature = higher switching point
TW, TB
in upper third of scale
+0
%
-4
at start of scale
+0
%
-6
STW, STB
in upper third of scale
+0
%
-5
at start of scale
+0
%
-7
Surface-mounting thermostats with end of scale
< +200°C +200°C +350°C > 350°C
TW
TB
STW
STB
TW
TB
STW
STB
TW
TB
STW
STB
effect due to the thermostat head, % per °C
0.08%/K 0.17%/K 0.06%/K 0.13%/K 0.14%/K 0.12%/K effect due to capillary, % per °C per meter
0.047%/K 0.054%/K 0.09%/K 0.11%/K 0.04%/K 0.03%/K
24
7 Appendix
25
7 Appendix
7 Appendix
26
27
7 Appendix
7 Appendix
28
29
7 Appendix
7 Appendix
30
JUMO GmbH & Co. KG
Street address: Moritz-Juchheim-Straße 1 36039 Fulda, Germany Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 e-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2TT, UK Phone: +44 1279 635533 Fax: +44 1279 635262 e-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-5866
1-800-554-JUMO Fax: 315-697-5867 e-mail:
info@jumo.us Internet: www.jumo.us
Loading...