If the device or a sensor connected to it fails abruptly, it is likely that a
dangerous overdosage has occurred!
For this case, appropriate precautionary measures must be taken.
NOTE!
Read these installation instructions before starting up the device. Store these
installation instructions in a place that is accessible to all users at all times.
This chapter explains in detail all submenus and setting options for inputs/
outputs and display functions in the "Configuration" menu.
Complete configuration of the device is described in the operating manual.
⇨ Operating manual B 202581.0
10.2Important information
WARNING!
After every configuration change, the device restarts functions affected by the
changes. Analog and digital outputs can assume undesired states during startup.
Therefore, configuration changes must never be made while equipment is operating!
CAUTION!
Besides incorrect installation, incorrectly set values on the device can impair
performance of the connected process or cause damage. Therefore, always
provide safety devices independent of the device and allow only qualified personnel to make settings.
10 Configuration
CAUTION!
When changing configuration data that are relevant for the data monitoring and
recording function, data recording is terminated and a new recording session
begun.
NOTE!
Changes to the configuration settings described in this chapter can be made
directly on the device or via the JUMO PC setup program.
NOTE!
Settings in the "Configuration" menu can be changed only if a user with corresponding user rights is logged in.
⇨ Chapter 8.1.1 "Passwords and user rights", page 77
NOTE!
Changes to the configuration go into effect only after you leave the configuration menu ("Exit" menu item or "Close window" button).
9
10 Configuration
10.3Basic settings
Open: Device Menu > Configuration > Basic Settings
Configuration pointSelection/
setting option
Device nameup to 20 text charactersDevice ID, e.g. for identification of exported mea-
National languageGerman
English
Explanation
surement data in the JUMO PCA 3000 evaluation software
Setting the operating language
Additional languages can be installed on the device using the setup program.
⇨ Operating manual B 202581.0
Language select after
power on
Mains frequency50 Hz
Device temp.Degrees Celsius
Interface temperature Degrees Celsius
Memory alarm limit0 to 100 %If the available memory display reaches this val-
Yes
No
60 Hz
Degrees Fahrenheit
Degrees Fahrenheit
Specifies whether a prompt for selecting the operating language should appear when the device
is switched on
Supply frequency of the electric power grid serving the mounting site
The supply frequency must be specified to suppress EMC interference caused by the mains
voltage. Setting the correct supply frequency is
thus also necessary when supplying the device
with direct voltage.
Setting the default on the temperature unit for all
temperature values in the device
Setting the default on the temperature unit for all
temperature values communicated via interfaces
ue, the memory alarm is triggered.
10
10.4Display
10.4.1General
Open: Device Menu > Configuration > Display > General
10 Configuration
Configuration pointSelection/
setting option
Lock touchscreenSelection from
binary selector
Simulation of
inputs
Main screenSelection of an
Display 1 and 2 general screen
Show detailed screen
1to6
Show diagram
1 and 2
Show process screen
Show controller
overview
Show controller
1to4
Show alarmsYes
Yes
No
operation screen from
the operation loop
Yes
No
No
Explanation
Digital signal that blocks operation of the touchscreen (e.g. key switch for locking operation)
When this function is activated, alternating on/off
signals are simulated automatically at the binary
inputs and continuous value changes at the analog inputs
This function is used for troubleshooting. Deactivate it during normal operation.
Selection of the operation screen as main screen
The main screen appears after the device has
been switched on or the "Home" button has been
pressed.
Individual operation screens can be shown or
hidden in the operation loop here.
Activation or deactivation of alarm visualization in
the title bar of the operation screens
11
10 Configuration
10.4.2Screen
Open: Device Menu > Configuration > Display > Screen
Configuration itemSelection/
setting option
Activation of screensaver
Wait time for screensaver
Signal screensaverSelection from
Brightness1 to 10Display brightness (10 levels)
10.4.3Colors
Configuration pointSelection/
Alarm 1 to 2Selection from
inactive
time to switch off
control signal
10 to 32767 sonly with activation of the screensaver after
binary selector
Open: Device Menu > Configuration > Display > Colors
setting option
color palette
Explanation
Type of screensaver activation
wait time:
Time to wait before displaying the screensaver
when there is no activity on the device
only with control signal screensaver activation:
Signal for activating the screensaver
Explanation
Color setting for signaling alarms 1 to 2 of the
measuring inputs
Recording:
Analog channel
1to4
Binary channel
1to3
Background, ana-
logue
Background, binary
Color t-stamp diagr
Color diag. traces
Controller
Background
Setpoint value
Actual value
Output level
Heating contact
Cooling contact
Selection from
color palette
Selection from
color palette
Upon reaching the set alarm values, the measured value displays and bar graphs appear in
the set colors.
Color setting for the visualization elements of the
individual recorder screens
Color setting for the visualization elements of the
individual controller screens
12
10.5Operating loop
10.5.1General screens
Open: Device Menu > Configuration > Operation Loop > General Screen >
General Screen 1 to 2
10 Configuration
Configuration pointSelection/
setting option
Gener. screen type2-part screen
4-part screen
Screen titleup to 31 text charactersTitle of the overview screen
Value title 1 to 2(4)up to 15 text charactersTitles of the individual main value display fields
Signal main value
1to2(4)
Color main value
1to2(4)
Signal sec. value
1to2(4)
Color sec. value
1to2(4)
Title additional valueup to 15 text charactersTitle of the additional value display field
Signal add valueSelection from
Color addit. valueSelection from
Title binary valueup to 15 text charactersTitle of the binary value display field
Signal binary val.
1to3
Selection from
analog selection
Selection from
color palette
Selection from
analog selection
Selection from
color palette
analog selection
color palette
Selection from
binary selector
Explanation
Selection of the type of overview screen;
2-part screen: Display of 2 main values, 2 second values, 1 additional value and 3 binary values;
4-part screen: Display of 4 main values, 4 second values, 1 additional value and 3 binary values
Signal source of the analog value for each main
value displayed
Color of the measured value display for each
main value
Signal source of the analog value for each second value displayed
Color of the measured value display for each
second value
Signal source of the analog value displayed as
additional value
Color of the measured value display for the additional value
Signal sources of the binary values visualized in
the binary value display field
Screen titleup to 31 text charactersTitle of the individual screen
Input signal main val-ueSelection from
analog selection
Color main valueSelection from
color palette
Second value input
signal
Color sec. valueSelection from
Title additional valueup to 15 text charactersTitle of the additional value display field
Additional valueSelection from
Color addit. valueSelection from
Title binary valueup to 15 text charactersTitle of the binary value display field
Signal binary val.
1to3
Selection from
analog selection
color palette
analog selection
color palette
Selection from
binary selector
Explanation
Signal source of the analog value displayed as
the main value and visualized as a bar graph
Color of the measured value display and the bar
graph for the main value
Signal source of the analog value displayed as
the second value
Color of the measured value display for the second value
Signal source of the analog value displayed as
additional value
Color of the measured value display for the additional value
Signal sources of the binary values visualized in
the binary value display field
14
10.6Analog inputs
10.6.1Base unit temperature inputs
Base unit temperature inputs: IN 4/5
Open: Device Menu > Configuration > Analog Inputs >
Temperature Inputs 1 to 2
10 Configuration
Configuration pointSelection/
setting option
Descriptionup to 15 text charactersDesignation for the input
Signal type IN 4/5:
Pt100
Pt1000
400 Ω
4000 Ω
IN 5 :
100 kΩ
NTC 8k55
NTC 22k
Resistance potentiome-
a
ter
Connection type2-wire,
3-wire
Customer-specific
linearization
Selection of a
linearization table
Explanation
Type of sensor connected
For Pt100, Pt1000 and NTC,
corresponding linearizations are provided.
For 400 Ω, 4000 Ω and 100 kΩ, a customer-spe-
cific linearization must be configured.
For resistance transmitters
cific linearization can be configures if necessary.
⇨ "Customer-specific linearization", in this table
only for signal types Pt100, Pt1000, 400 Ω,
4000 Ω, 100 kΩ and NTC:
Connection variants for the connected resistance
thermometer
only for signal types 400 Ω , 4000 Ω , 100 kΩ
or resistance transmitters
Linearization tables contain up to 40 value pairs
in any measurement characteristic line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To create a linearization table, you need the JUMO PC
setup program.
⇨ Operating manual B 202581.0
Start of
display range
Display range end-99999 to +99999
Decimal placeAuto,
Offset-999 to +999
-99999 to +99999
Fixed comma format
b
b
b
Upper/lower limit for labeling the scale when displaying measured values, e.g. in recorder diagrams and bar graphs
Decimal points in the display
Correction value added to measuring value
a
a customer-spe-
a
:
This value can be used, for instance, to compensate for measurement errors resulting from the
resistance of wires.
Filter Time Constant0.0 to 25.0 sOptimization of measuring value updating
The higher the value of the filter time const., the
slower the measured value is updated.
15
10 Configuration
Configuration pointSelection/
setting option
Ra0 to 99999 Ωon ly for IN 5: Resistance value between the slid-
Rs6 to 99999 Ωonly for IN 5: Span of the variable resistance
Re0 to 99999 Ωon ly for IN 5: Resistance value between the slid-
Alarms 1/2Analog input alarms are used to monitor measured values in relation to ad-
justable limit values.
The alarm settings for all analog device functions are explained together.
Explanation
er (S) and start point (A) in a resistance transmitter/potentiometer
point.
value between the slider (S) and start point (A)
er (S) and end point (E) in a resistance transmitter/potentiometer
point.
a
when the slider is at the start
a
when the slider is at the end
⇨ Chapter 10.11.1 "Alarms for analog signals and digital sensors", page 42
a
Resistance transmitter
b
The temperature unit specified in the basic settings appears in the entry field.
⇨ Chapter 10.3 "Basic settings", page 10
16
10.6.2Universal inputs of base unit and optional boards
Base unit universal input: IN 6
Optional board universal inputs: IN 11/12
Open: Device Menu > Configuration > Analog Inputs >
Universal Input 1 to 3 > Configuration
10 Configuration
Configuration pointSelection/
setting option
Descriptionup to 15 text charactersDesignation for the input
Operating modelinear scaling,
linear scaling: Standard signals (for
IN 11/12 and resistance transmitter
measurement characteristic line or customerspecific linearization
For standard signals the scale start point, scale
end point and unit must be specified.
Temp. measurement: Measurement made with
a
resistance thermometer
The type of sensor is selected in the "Signal type"
configuration item. The unit for the temperature is
specified in the "Basic Settings" menu.
⇨ Chapter 10.3 "Basic settings", page 10
a
) with linear
Signal typeIN 6/11/12:
0to20mA
4to20mA
20 to 0 mA
20 to 4 mA
only IN 11/12:
0to10V
10 to 0 V
Pt100
Pt1000
400 Ω
4000 Ω
Resistance potentiometer
pH value, conductivity, and free chlorine: The
values measured by the individual analysis sensors are received in the form of a standard signal.
Outside factors affecting the individual analysis
process variables are compensated. It is thus
necessary to make appropriate compensation
settings in the configuration of the universal input.
Type of sensor connected
Correct scale settings are required for the stan-dard signals.
⇨ "Scale start/end" in this table.
Appropriate linearizations are available for
Pt100, Pt1000 andresistance transmitters
For 400 Ω and 4000 Ω, a customer-specific linearization must be configured.
a
.
⇨ "Customer-specific linearization" in this table
17
10 Configuration
Configuration pointSelection/
Explanation
setting option
Connection type2-wire,
3-wire
only for signal types Pt100, Pt1000, 400 Ω and
4000 Ω:
Connection variants for the connected resistance
thermometer
Customer-specific
linearization
Selection of a
linearization table
Linearization tables contain up to 40
value pairs in any measurement characteristic
line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To create a linearization table, you need the JUMO PC
setup program.
⇨ Operating manual B 202581.0
Unitup to 5 text charactersProcess variable unit
not adjustable for pH measurement
The temperature unit is specified in the basic settings.
⇨ Chapter 10.3 "Basic settings", page 10
Scale start-99999 to +99999
Scale end-99999 to +99999
Start of
-99999 to +99999
display range
Display range end-99999 to +99999
Decimal placeAuto
Fixed comma format
Offset-999 to +999
Filter Time Constant0.0 to 25.0 sOptimization of measured value updating
Ra0 to 4000 Ωonly for IN 11/12: Resistance value between the
b
only for standard signals:
Measured value from sensor (uncomp.) that corresponds to the lower limit of the standard signal
range [0 V or 0(4) mA];
Refer to the technical data for
the sensor.
b
only for standard signals:
only for standard signals:
Measured value from sensor (uncomp.) that corresponds to the upper limit of the standard signal
range [10 V or 20 mA]
Refer to the technical data for
the sensor.
b
Upper/lower limit for labeling the scale when displaying measured values, e.g. in recorder dia-
b
grams and bar graphs
Decimal points in the display
b
only for temperature and
conductivity measurement:
Correction value added to measured value
The higher the value of the filter time const., the
slower the measured value is updated.
slider (S) and start point (A) in a resistance potentiometer when the slider is at the start point
18
10 Configuration
Configuration pointSelection/
Explanation
setting option
Rs6 to 4000 Ωonly for IN 11/12: Span of the variable
resistance value between the slider (S) and
start point (A)
Re0 to 4000 Ωonly for IN 11/12: Resistance value between the
slider (S) and end point (E) in a resistance potentiometer when the slider is at the end point.
Compensation temperature
Selection from
analog selection
Analog input of the compensation thermometer
for temperature-compensated measurement of
pH value, free chlorine or conductivity
CompensationTC linear,
TC-curve,
Type of temperature compensation for conduc-
tivity measurement
natural water,
natural water with
expanded temperature
range,
ASTM neutral,
ASTM acid,
ASTM alkaline,
NaOH 0 to 12 %,
NaOH 25to50%,
HNO
HNO
H
H
H
3
3
2SO4
2SO4
2SO4
0to25%,
36 to 82 %,
0to28%,
36 to 85 %,
92 to 99 %,
HCL0to18%,
HCL 22 to 44 %
Reference temp.15to30°Crequired only for conductivity measurement
with "TC linear" or "TC curve" temperature
compensation:
Temperature at which the (temperature-compensated) conductivity value displayed was set
Compensation
pH value
Selection from
analog selection
Analog input of the pH-value sensor for pH-compensated measurement of free chlorine
Alarms 1/2Analog input alarms are used to monitor measured values
in relation to adjustable limit values. The alarm settings for all analog device
functions are explained together.
⇨ Chapter 10.11.1 "Alarms for analog signals and digital sensors", page 42
a
Resistance transmitter
b
The unit for the particular sensor value appears in the entry field.
10.6.3Calibration timer
Open Calibration Timer Settings Universal Inputs:
Device Menu > Configuration > Analog Inputs >
Universal Input 1 to 3 > Calibration Timer
Calibration timers prompt the user to calibrate sensors on a regular basis. The
settings for all analysis inputs and the universal inputs are explained together.
19
10 Configuration
⇨ Chapter 10.12 "Calibration timer", page 46
20
10.6.4pH/Redox/NH3analysis inputs
Open: Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Configuration
10 Configuration
Configuration pointSelection/
setting option
Descriptionup to 15 text charactersDesignation for the input
Electrode typepH standard
pH antimony
pH ISFET
redox
ammonia
Redox unitmV
percent
Explanation
Type of electrode connected
mV: unit for the redox potential
percent: concentration percentage that can be
derived from the redox measurement
This requires a two-point calibration.
⇨ Chapter 13.2.1 "Calibration methods for Re-
dox sensors", page 63
Filter Time Constant0.0 to 25.0 sOptimization of measured value updating
The higher the value of the filter time const., the
slower the measured value is updated.
Start of
display range
Display range end-99999 to +99999
Compensation tem-
perature
Glass electrode
monitoring
-99999 to +99999
Selection from
analog selection
off
minimum impedance
maximum impedance
min./max. impedance
a
a
Upper/lower limit for labeling the scale when displaying measured values, e.g. in recorder diagrams and bar graphs
Analog input of the temperature sensor for compensation of the effect of temperature on the pHvalue measurement
configurable monitoring of pH glass electrodes
without impedance converter
minimum impedance: monitoring for soft short/
sensor failure
Reference
electrode monitoringOnOff
max.
reference impedance
Alarm/Event listoff
Sensor alarm
delay
0to100kΩupper impedance limit for monitoring a reference
event
alarm
0 to 999 sThe sensor alarm is suppressed for the set dura-
maximum impedance: monitoring for aging/pollution/wire break
Activation of reference electrode impedance
monitoring
A high-impedance symmetrical connection is required.
electrode
Assignment of sensor failure notification to alarm
list or event list
tion of the alarm delay.
21
10 Configuration
Configuration pointSelection/
setting option
Text sensor alarmup to 21 text charactersText for the alarm/event list in case of sensor er-
Alarms 1/2Analog input alarms are used to monitor measured values
in relation to adjustable limit values. The alarm settings for all analog device
functions are explained together.
Explanation
ror
⇨ Chapter 10.11.1 "Alarms for analog signals and digital sensors", page 42
a
The entry field units depend on the configuration items "Electrode type" and
"Redox unit".
10.6.5Calibration timer
Open Calibration Timer Settings Analysis Inputs pH/Redox/NH3:
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Calibration Timer
Calibration timers prompt the user to calibrate sensors on a regular basis. The
settings for all analysis inputs and the universal inputs are explained together.
⇨ Chapter 10.12 "Calibration timer", page 46
NOTE!
The following points must be observed for correct operation of glass electrode
monitoring via impedance measurement (see preceding table):
•Impedance measurements are possible only with glass-based sensors.
•Sensors must be connected directly to an analysis input for pH/Redox/NH
on the device.
•Impedance converters must not be installed in the measuring circuit.
•The maximum admissible cable length between sensor and device is 10 m.
•Fluid resistances have a direct impact on the measurement result. It is
therefore advisable to activate the impedance measurement in liquids at a
minimum conductivity of approx. 100 µS/cm.
Open: Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Configuration
Configuration pointSelection/
Explanation
setting option
Descriptionup to 15 text charactersDesignation for the input
Compensation tem-
perature
Selection from
analog selection
Analog input of the compensation thermometer
for temperature-compensated conductivity measurement
Reference temp.15to30°Crequired only for conductivity measurement
with "TDS", "TC linear" or "TC curve" temperature compensation:
The temperature at which the conductivity value
displayed was set
Filter Time Constant0.0 to 25.0 sOptimization of measured value updating
The higher the value of the filter time const., the
slower the measured value is updated.
nominal
cell constant
for CR: 0.01 to 10 cm
for Ci: 4.00 to 8.00 cm
-1
nominal cell constant of the conductivity sensor
-1
(can be read from the sensor nameplate)
If an ASTM test certificate showing preciselymeasured cell constants is available, in addition
to entering the nominal cell constants the relative
cell constants of all measuring ranges in the calibration values must be manually entered (cf.
Chapter "Manual entry of calibration values",
page 50). The relative cell constants must be calculated from nominal cell constants (nameplate
for the sensor) and measured cell constants
(ASTM test certificate):
(measured cell constant × 100 %) ÷ nominal cell
constant = relative cell constant.
For conductivity sensors with 4 electrodes, pollution detection is made available.
possible only for conductive conductivity
measurement in a 4-wire circuit:
When this function is activated, a sensor alarm is
triggered in case of soiling.
possible only for conductive conductivity
measurement:
When this function is activated, a sensor alarm is
triggered in case of a sensor wire break.
23
10 Configuration
Configuration pointSelection/
setting option
Alarm/Event listoff
event
alarm
Sensor alarm
delay
Text sensor alarmup to 21 text characterspossible only for conductive conductivity
Measuring range selection 1
Measuring range selection 2
Measuring ranges
1to4
0to999spossible only for conductive conductivity
Selection from
binary selector
Selection from
binary selector
-Four ranges each can be configured for conduc-
Explanation
possible only for conductive conductivity
measurement:
Assignment of the sensor failure notification to
alarm list or event list
measurement:
The sensor alarm is suppressed for the set duration of the alarm delay.
measurement:
Text for the alarm/event list in case of sensor error
Measuring range selection permits selection of
the measuring ranges 1 to 4 by controlling the binary signals.
⇨ Chapter "CR/Ci measuring range switching",
page 24
tive/inductive (CR/Ci) conductivity measurements. The settings for all CR-/Ci analysis inputs
are explained together.
⇨ "CR/Ci measuring range configuration", page
26
Alarms 1/2
per measuring range
1to4
Analog input alarms are used to monitor measured values
in relation to adjustable limit values. The alarm settings for all analog device
functions are explained together.
⇨ Chapter 10.11.1 "Alarms for analog signals and digital sensors", page 42
10.6.7Calibration timer
Open Calibration Timer Settings Analysis Inputs CR/Ci:
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Calibration Timer
Calibration timers prompt the user to calibrate sensors on a regular basis. The
settings for all analysis inputs and the universal inputs are explained together.
⇨ Chapter 10.12 "Calibration timer", page 46
10.6.8CR/Ci measuring ranges
Four configurable measuring ranges are available for each CR/Ci analysis input
for measuring electrolytic conductivity. The measuring range is switched by
means of 2 selectable digital signals. These signals are specified in the configuration of the individual conductivity measurement inputs.
The following table shows which binary value combinations activate the individual measuring ranges:
Active measuring
range
Measuring range 100
Measuring range 210
Measuring range 301
Measuring range 411
Binary signal
Measuring range selection 1
Digital signal
Measuring range switching 2
25
10 Configuration
CR/Ci measuring range configuration
Open: Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Measuring Range 1 to 4
Configuration pointSelection/
setting option
TDS factor0.01 to 2.00only for conductive conductivity with
Compensationfor CR/Ci:
Off,
TC linear,
natural water,
natural water with
expanded temperature
range
only for CR:
TDS,
ASTM neutral,
ASTM acid,
ASTM alkaline
only for Ci:
TC curve,
NaOH 0 to 12 %,
NaOH 25to50%,
HNO30to25%,
HNO336to82%,
H2SO40to28%,
H2SO436to85%,
H2SO492to99%,
HCL0to18%,
HCL 22 to 44 %
Unit for
calculation
Unitup to 5 text charactersonly for conductive conductivity measure-
for CR/Ci:
µS/cm
mS/cm
only for CR:
kΩ/cm
MΩ/cm
Explanation
TDS compensation:
Conversion factor from measured conductivity to
display unit (see configuration item "Unit" in this
table)
Type of temperature compensation for conductivity measurement
Unit in which the conductivity is displayed
ment with
TDS compensation:
Unit for the process variable to be displayed for
TDS measurements or when using customerspecific linearization (e.g. ppm or mg/l)
26
10 Configuration
Configuration pointSelection/
setting option
Customer-specific
linearization
Selection of a
linearization table
Explanation
Linearization tables contain up to 40
value pairs in any measurement characteristic
line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To create a linearization table, you need the JUMO PC
setup program.
⇨ Operating manual B 202581.0
Start of
display range
Display range end-99999 to +99999
Decimal placeAuto
Offset-99999 to +99999
Alarms 1/2
per measuring range
1to4
-99999 to +99999
Fixed comma format
Analog input alarms are used to monitor measured values
in relation to adjustable limit values. The alarm settings for all analog device
functions are explained together.
a
a
a
Upper/lower limit for labeling the scale when displaying measured values, e.g. in recorder diagrams and bar graphs
Decimal points in the display
Correction value added to measured value
⇨ Chapter 10.11.1 "Alarms for analog signals and digital sensors", page 42
a
The unit set for the conductivity measurement input appears in the entry field.
27
10 Configuration
10.7Analog outputs of base unit and optional boards
Open: Device Menu > Configuration > Analog Outputs >
Analog Output 1 to 9
Configuration pointSelection/
setting option
Descriptionup to 15 text charactersDesignation for the output
Signal
output value
Analog signal0 to 10 V
Enable manu. modeYes
Selection from
analog selection
0to20mA
4to20mA
10 to 0 V
20 to 0 mA
20 to 4 mA
No
Explanation
Analog signal source of the output
Type of standard signal to be generated
The manual mode for the specific output is enabled/locked here.
The manual mode permits fixed analog values
to be set for the output for testing purposes.
⇨ Chapter 8.2.3 "Functional level", page 93
Safety value 1 to 40 to 10.7 V
or
0to22mA
Scale start 1 to 4-99999 to +99999
Scale end 1 to 4-99999 to +99999
Digital signal for holdSelection from
binary selector
a
a
Specifies an analog value that the output assumes at hold, at calibration or in case of a fault
If an analysis input for conductivity measurement
is set as "Signal output value", the safety values
1 to 4 are assigned to conductivity measuring
ranges 1 to 4. Pairs with the same number belong together.
Otherwise, the safety value 1 applies.
Analog value of the analog signal source (see
configuration item "Signal output value") that corresponds to the lower limit of the standard signal
range generated
Analog value of the analog signal source (see
configuration item "Signal output value") that corresponds to the upper limit of the standard
signal range generated (10 V or 20 mA)
Digital signal for activating the hold function
When the hold function is activated, the analog
output assumes the state defined in the "Response at hold" setting.
[0V or 0(4)mA]
28
10 Configuration
Configuration pointSelection/
setting option
Response at holdlow
high
NAMUR low
NAMUR high
hold
safety value
Response during calibration
Response in case of a
fault
a
The unit for the value set for the "Signal output value" appears in the entry
field.
Moving
Frozen
Safe value
low
high
NAMUR low
NAMUR high
hold
safety value
Explanation
Specification of the analog output value when the
hold function is activated, during calibration of
one of the sensors for the particular output or at
error (overrange/underrange)
low: lower limit of the standard signal value
range
high: upper limit of the standard signal value
range (10 V or 20 mA)
NAMUR low: lower NAMUR limit of the standard
signal
NAMUR high: upper NAMUR limit of the standard signal (10.7 V or 22 mA)
hold: unchanging analog value
Safety value: see configuration item
"Safety value" in this table
[0 V or 0(4) mA]
[0 V or 0(3.4) mA]
10.8Digital inputs of base unit and optional boards
Open: Device Menu > Configuration > Digital Inputs >
Digital Input 1 to 9
Configuration pointSelection/
setting option
Descriptionup to 21 text charactersDesignation for the input
InversionYes
No
ContactBase unit:
potential-free contact,
external voltage source
Optional boards:
potential-free contact
AlarmBinary inputs alarms are used to monitor input-side switching signals. The
alarm settings for all digital device functions are explained together.
⇨ Chapter 10.11.2 "Digital signal alarms", page 45
Explanation
Invert/do not invert state
Type of digital signal connected
29
10 Configuration
10.9Digital outputs of base unit and optional boards
Open: Device Menu > Configuration > Digital Outputs >
Digital Output 1 to 17
Configuration pointSelection/
setting option
Descriptionup to 21 text charactersDesignation for the output
Signal
output value
InversionYes
Enable manu. modeYes
Selection from
binary selector
No
No
Explanation
Digital signal source for the output
Invert/do not invert state
The manual mode for the specific output is enabled/locked here.
Manual mode permits fixed binary values
(switching states) to be set for the output for testing purposes.
⇨ Chapter 8.2.3 "Functional level", page 93
30
10.10Digital sensors
NOTE!
For operation of digital sensors, you need the extra code "JUMO digiLine protocol activated“ (see Chapter 4.2 "Order Details", page 17)
NOTE!
Only one serial interface of the device can be configured for operation of digital
sensors. If your device has 2 serial interfaces (base unit and possibly an optional board), select 1 interface for connection of digital sensors and configure
it appropriately.
NOTE!
Correct functioning of digital sensors depends on correctly setting the interface
to which the digital sensors are connected and are to be operated. Make sure
the selected serial interface is configured correctly.
10.10.1General
10 Configuration
Open General Settings Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6> General
Digital sensors and JUMO digi-Line
electronics can only be linked if of
these settings match the type information of the sensor to be linked. If
this setting is changed during operation of a linked sensor, the sensor involved loses its link and must be put
into operation anew.
31
10 Configuration
10.10.2Configuration
Open Configuration Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > Configuration
Gen. settings for all types of digital sensors
Configuration pointSelection/
setting option
Descriptionup to 15 text charactersPlain text designation for the digital
Alarms 1/2Analog input alarms are used to monitor measured values in relation to ad-
justable limit values.
The alarm settings for all analog device functions are explained together.
Explanation
sensor input
This designation is displayed in
menus, e. g. analog or binary selection.
⇨ Chapter 10.11.1 "Alarms for analog signals and digital sensors", page 42
Only for sensors with JUMO digiLine pH/ORP/T
Configuration pointSelection/
setting option
TAG checkinginactive
active
Sensor TAGup to 20 text characters
Explanation
If desired, this function can be activated to assign JUMO digiLine electronics to measuring points. When it
is activated, the "TAG number" of the
sensor is compared with the entry in
the "Sensor Tag" field of the digital
sensor inputs on connection of a sensor to a master device. If these differ,
the JUMO digiLine electronics are
not linked.
a
Filter Time Constant
Start of display range -99999 to +99999
Display range end-99999 to +99999
0 to 25 sOptimization of measured value up-
32
The "TAG number" of the JUMO digiLine electronics can be entered and
edited in the JUMO digiLine electronics only with the JUMO DSM software.
dating in the JUMO digiLine electronics
The higher the value of the filter time
const., the slower the measured value is updated.
b
b
Upper/lower limit for labeling the
scale when displaying measured values, e.g. in recorder diagrams and
bar graphs
10 Configuration
Configuration pointSelection/
Explanation
setting option
Compensation
a
Fixed compensation temperature
Sensor temperature
Interface
For pH sensors only:
Selection of the signal source for
temperature compensation in the
JUMO digiLine electronics
Fixed compensation temperature:
Compensation with a fixed temperature value that is entered in the configuration item "Compensation
temperature".
Sensor temperature: The integrated temperature probe of the pH sensor supplies the compensation
temperature.
Interface: The AQUIS touch S transmits the compensation temperature
to the JUMO digiLine electronics via
the serial interface. The source of the
compensation temperature is set in
the configuration point "Compensation temperature".
Fixed compensation
temperature
a
-25to+150°Conly if "Compensation" is set to
"Fixed compensation temperature":
constant temperature value, for temperature compensation of the pH value measurement in the JUMO
digiLine electronics
Compensation temperature
Selection from Analog selectiononly if "Compensation" is set to
"Interface":
Selection of the signal source from
the Analog selection of the JUMO
AQUIS touch S for temperature compensation of the pH value measurement in the JUMO digiLine
electronics
Temperature input filter time
a
0 to 25 sOptimization of measured tempera-
ture value updating in the JUMO digiLine electronics
The higher the value of the filter time
const., the slower the measured value is updated.
Temperature offset
a
-10 to+10 °CCorrection value that is added to the
measured temperature value
a
This setting is saved in the configuration of the JUMO digiLine electronics.
b
The unit for the item set in "Parameter" appears in the entry field.
33
10 Configuration
For JUMO ecoLine O-DO sensors only
Configuration pointSelection/
setting option
Oxygen unit% Sat
mg/l
ppm
Salt content0 to 10000 g/kgIndication of the salt content (salinity)
Air pressure500 to 1500 hPaIndication of the ambient air pressure
Compensation
a
Fixed compensation temperature
Sensor temperature
Interface
Explanation
Selection of the unit in which the oxygen concentration is displayed.
of the measurement solution for compensation of the effect of the salinity
on the measured value of the oxygen
concentration in the measurement
solution
for compensation of the effect of the
air pressure on the measured value
of the oxygen concentration in the
measurement solution
Fixed compensation temperature:
Compensation with a fixed temperature value that is entered in the configuration item "Compensation
temperature".
Fixed compensation
temperature
Compensation temperature
a
Sensor temperature: The integrated temperature probe of the O-DO
sensor supplies the compensation
temperature.
Interface: The AQUIS touch S transmits the compensation temperature
to the sensor electronics via the interface. The source of the compensation temperature is set in the
configuration point "Compensation
temperature".
-25to+150°Conly if "Compensation" is set to
"Fixed compensation temperature":
constant temperature value, for temperature compensation of the oxygen measurement in the O-DO
sensor
Selection from Analog selectiononly if "Compensation" is set to
"Interface":
Selection of the signal source from
the Analog selection of the JUMO
AQUIS touch P for temperature compensation of the oxygen measurement in the O-DO sensor
34
10 Configuration
Configuration pointSelection/
Explanation
setting option
Start of display range -99999 to +99999
Display range end-99999 to +99999
a
a
Upper/lower limit for labeling the
scale when displaying measured values, e.g. in recorder diagrams and
bar graphs
Sampling rate1 to 999 sIndication of the length of the interval
between 2 measurements
A slower "sampling rate" contributes
to a longer operating life of the sensor.
With faster "sampling rates, the measured value is updated frequently.
Filter Time Constant0 to 25 sOptimization of measured value up-
dating in the JUMO digiLine electronics
The higher the value of the filter time
const., the slower the measured value is updated.
a
The unit set as the unit in "Oxygen unit" appears in the entry field.
For JUMO ecoLine NTU sensors only
Configuration pointSelection/
Explanation
setting option
Turbidity measuring
range
automatic
0to50NTU
Selection of the measuring range for
the turbidity measurement
0to200NTU
0to1000NTU
0to4000NTU
The user can choose a fixed measuring range or automatic measuring
range selection.
Turbidity unitNTU
FNU
Selection of the unit in which the
measured turbidity value is displayed.
Decimal placeFixed comma formatDecimal points in the display
Filter Time Constant0 to 25 sOptimization of measured value up-
dating in the JUMO digiLine electronics
The higher the value of the filter time
const., the slower the measured value is updated.
Start of display range -99999 to +99999
Display range end-99999 to +99999
a
a
Upper/lower limit for labeling the
scale when displaying measured values, e.g. in recorder diagrams and
bar graphs
a
The unit set as the unit in "Turbidity unit" appears in the entry field.
35
10 Configuration
For digital JUMO tecLine sensors only
Configuration pointSelection/
setting option
Measuring rangeSetting of the measuring range for
Free chlorine pH-dependent
Open chlorine measurement
Total chlorine
Chlorine dioxide
Chlorine dioxide tenside-neutral
Bromide
Free chlorine pH-neutral2 ppm
Ozone tensidedependent
Ozone tensideneutral
Peracetic acid200 ppm
2 ppm
20 ppm
20 ppm
200 ppm
10 ppm
20 ppm
2 ppm
10 ppm
2000 ppm
20000 ppm
Explanation
the various measurands of digital
JUMO tecLine sensors
This is where you set the measuring
range of your digital JUMO tecLine
sensor according to the order details
for the sensor.
Hydrogen peroxide20000 ppm
20 %
Decimal placeAuto
fixed comma format
Filter Time Constant0 to 25 sOptimization of measured value up-
Start of display range0 to 20000
Display range end0 to 20000
pH-compensated chlorine measurement
On
Off
a
a
Decimal points in the display
dating in the JUMO digiLine electronics
The higher the value of the filter time
const., the slower the measured value is updated.
Upper/lower limit for labeling the
scale when displaying measured values, e.g. in recorder diagrams and
bar graphs
available only for pH-dependent
measurements of free chlorine:
Activation/deactivation of the pHcompensation for the free chlorine
measurement
36
10 Configuration
Configuration pointSelection/
setting option
pH compensation sourceSelection from Analog
selection
a
The unit for the particular sensor value appears in the entry field.
Explanation
available only with pH-compensation activated:
Analog input of the pH-value sensor
for pH-compensated measurement
of free chlorine
37
10 Configuration
10.10.3Sensor alarms
Digital sensors cyclically transmit a number of alarm and status bits. In the settings for the sensor alarms, it is possible to set which of these signals should be
indicated on the JUMO AQUIS touch S. The settings for the alarm conditions
themselves are made in the configuration of the individual sensor electronics or
stipulated by the specifications for the particular sensor. More detailed information on configuration can be found in the operating manual for the particular sensor type.
Open Sensor Alarms Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6>
Sensor Alarms
Every alarm transmitted by the digital sensors has the following configuration
parameters:
Configuration pointSelection/
setting option
Alarm/Event listoff
event
alarm
Sensor alarm
delay
Alarm textup to 21 text charactersText for the alarm/event list
0 to 999 stime delay from receipt of the alarm
The range of alarm types that can be transmitted by a digital sensor depends on
the sensor type. The individual alarm lists for the various sensor types are provided in the following.
Explanation
Assignment of alarm notification to
alarm list or event list
signal from the sensor and triggering
of the alarm on the device
Displayed only and cannot be
changed.
38
10 Configuration
JUMO digiLine pH
AlarmExplanation
pH sensor alarmAlarm for pH value out of range
Temperature sensor alarmAlarm for temperature out of range
pH min. warningLower pH limit value pre-alarm from sensor (see "Sensor
monitoring" in the JUMO digiLine pH operating manual)
pH min. alarmLower pH limit value alarm from sensor (see "Sensor
monitoring" in the JUMO digiLine pH operating manual)
pH max. warningUpper pH limit value pre-alarm from sensor (see "Sensor
monitoring" in the JUMO digiLine pH operating manual)
pH max. alarmUpper pH limit value alarm from sensor (see "Sensor
monitoring" in the JUMO digiLine pH operating manual)
Min. temperature warningLower temperature limit value pre-alarm from sensor (see
"Sensor monitoring" in the JUMO digiLine pH operating
manual)
Min. temperature alarmLower temperature limit value alarm from sensor (see
"Sensor monitoring" in the JUMO digiLine pH operating
manual)
Max. temperature warningUpper temperature limit value pre-alarm from sensor (see
"Sensor monitoring" in the JUMO digiLine pH operating
manual)
Max. temperature alarmUpper temperature limit value alarm from sensor (see
"Sensor monitoring" in the JUMO digiLine pH operating
manual)
Calibration timer alarmAlarm for sensor calibration due (see "Calibration data" in
the JUMO digiLine pH operating manual)
CIP/SIP/Autoclaving warningPre-alarm for maximum number of CIP/SIP/autoclaving
cycles (see "Sensor monitoring" in the JUMO digiLine pH
operating manual)
CIP/SIP/Autoclaving alarmAlarm for maximum number of CIP/SIP/autoclaving cy-
cles (see "Sensor monitoring" in the JUMO digiLine pH
operating manual )
Sensor stress warningSensor stress pre-alarm (see "Sensor monitoring" in the
JUMO digiLine pH operating manual)
Sensor stress alarmSensor stress alarm (see "Sensor monitoring" in the
JUMO digiLine pH operating manual)
Digital input stateSignal state of the digital input of the sensor electronics
39
10 Configuration
JUMO digiLine ORP
AlarmExplanation
ORP sensor alarmAlarm for redox value out of range
Calibration timer alarmAlarm for sensor calibration due (see "Calibration data" in
the JUMO digiLine ORP operating manual)
Digital input stateSignal state of the digital input of the sensor electronics
JUMO digiLine T
AlarmExplanation
Temperature sensor alarmAlarm for temperature value out of range
Digital input stateSignal state of the digital input of the sensor electronics
40
10.10.4CIP/SIP definition (JUMO digiLine pH only)
Open CIP/SIP Definition Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > CIP/
SIP Definition
10 Configuration
Configuration pointSelection/
Explanation
setting option
CIP min. tempera-
a
ture
SIP min. temperature
-20 to+150 °CTemperature thresholds for identifying CIP/SIP
cycles
a
If the CIP/SIP cycle takes place above one of
these values during the set duration of a CIP/SIP
cycle, the values identify a successfully completed CIP/SIP cycle and the CIP or SIP counter in
the JUMO digiLine electronics is incremented.
The respective counter is reset only after the val-
ue has dropped below the CIP/SIP temperature.
CIP cycle duration
SIP cycle duration
CIP/SIP alarminginactive
a
0 to 9999 sDuration of a CIP/SIP cycle
a
Setting for alarming on the JUMO AQUIS touch
active
S, if the CIP, SIP or autoclaving counters in the
JUMO digiLine electronics have reached the
maximum number of permissible cycles.
a
This setting is saved in the configuration of the JUMO digiLine electronics.
10.10.5Calibration timer
Open Calibration Timer Settings Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > Calibration Timer
Calibration timers prompt the user to calibrate sensors on a regular basis. The
settings for all analysis inputs and the universal inputs are explained together.
⇨ Chapter 10.12 "Calibration timer", page 46
41
10 Configuration
10.11Input alarm functions
10.11.1Alarms for analog signals and digital sensors
Open Temperature Inputs Alarm Configuration:
Device Menu > Configuration > Analog Inputs >
Temperature Inputs 1 to 2
Open Universal Inputs Alarm Configuration:
Device Menu > Configuration > Analog Inputs >
Universal Input 1 to 3 > Configuration
Open pH/Redox/NH
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Configuration
Open CR / Ci Analysis Inputs Alarm Configuration:
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Configuration > Measuring Range 1 to 4
Open Alarm Configuration Inputs for Digital Sensors:
Device Menu > Configuration > Digital Sensors >
Digital Sensors 1 to 6 > Configuration
Open Alarm Configuration For External Analog Inputs:
Device Menu > Configuration > External Analog Inputs >
External Analog inputs 1 to 8
Open Alarm Configuration For Flow Rate:
Device Menu > Configuration > Flow Rate > Flow Rate 1 to 2
Configuration pointSelection/
setting option
Digital signal for holdSelection from the
binary selector
Response at holdinactive
active
hold
Response at
calibration
Response at errorinactive
inactive
active
hold
normal
active
hold
Analysis Inputs Alarm Configuration:
3
Explanation
Digital signal for activating the hold function
When the hold function is activated, the alarm as-
sumes the state defined in the "Response at
hold" setting.
Specification of the alarm state when the hold
function is activated, at calibration of the particu-
lar input or at error (out of range)
inactive: alarm suppressed
active: alarm forced
hold: alarm state is maintained regardless of
alarm condition changes
42
normal: alarm in acc. with alarm condition
10 Configuration
Configuration pointSelection/
setting option
Alarm typeinactive
minimum alarm
maximum alarm
alarm window
invert alarm window
only for
CR analysis inputs:
USP
Pre-alarm USP
purified water
purified water pre-alarm
Alarm/Event listoff
event
alarm
Alarm textup to 21 text charactersText for the alarm/event list
Limit value-99999 to +99999Limit value for the particular alarm type
Hysteresis0 to 99999Spacing between switch-on and switch-off points
Explanation
Four alarm types (comparator functions) can be
selected to monitor measured values for violation
of limit values.
⇨ Characteristic lines after the table
Limit value alarms to USP <645> or
European Pharmacopoeia (Ph. Eur.) for
cleaned water
Assignment of alarm notification to alarm list or
event list
for the alarm types
⇨ Characteristic lines after the table
Window range0 to 99999Range of the alarm window
⇨ Characteristic lines after the table
Pulse functionYes
No
Pulse time0 to 999 sDuration of alarm with pulse function activated
Alarm delay
On
Alarm delay Off0 to 999 sTime delay between ending of the alarm condi-
0 to 999 sTime delay between occurrence of the alarm
Time limit for alarm with the wiper time as maximum alarm duration
condition and triggering of the alarm
tion and clearing of the alarm
43
10 Configuration
1
0
Binary
value
Input signal
Limit value
Hysteresis
1
0
Binary
value
Input signal
Limit value
Hysteresis
1
0
Binary
value
Input signal
Limit value
Window range
Hysteresis
Hysteresis
1
0
Binary
value
Input signal
HysteresisHysteresis
Limit value
Window range
Minimum alarm (On-signal when value drops below lower limit)
Maximum alarm(On-signal when value exceeds upper limit)
Alarm window (On-signal within a configurable value range
)
Invert alarm window (On-signal outside a configurable value range)
44
10.11.2Digital signal alarms
Open Alarm Configuration For Binary Inputs:
Device Menu > Configuration > Binary Inputs >
Binary Inputs 1 to 9
Open Alarm Configuration For External Binary Inputs:
Device Menu > Configuration > External Binary Inputs >
External Binary Inputs 1 to 8
Open Alarm Configuration For Logic Formula:
Device Menu > Configuration > Logic Formula > Formula 1 to 30
10 Configuration
Configuration pointSelection/
setting option
Digital signal for holdSelection from the
binary selector
Response at holdinactive
active
hold
normal
Alarm typeactive
inactive
Alarm/Event listoff
event
alarm
Alarm textup to 21 text charactersText for the alarm/event list
Alarm active athigh
low
Alarm delay 0 to 999 sTime delay between occurrence of the alarm
Explanation
Digital signal for activating the hold function
When the hold function is activated, the alarm assumes the state defined in the "Response at
hold" setting.
Specification of the alarm state when the hold
function is activated
inactive: alarm suppressed
active: alarm forced
hold: alarm state is maintained regardless of
alarm condition changes
normal: alarm in acc. with alarm condition
Arming or disarming an alarm
Assignment of alarm notification to alarm list or
event list
Digital input alarm condition
condition and triggering of the alarm
45
10 Configuration
10.12Calibration timer
Every input for analysis measurands has its own calibration timer. Inputs for
temperature sensors do not have a calibration timer, since they do not need to
be calibrated. Calibration timers signal when sensor calibration is due via a calibration alarm. Once a particular input has been calibrated successfully, its calibration timer is reset. The signal for a due calibration can be generated, for
instance, by binary outputs with external external indicator lights or the alarm/
event list. The structure of the calibration timer configuration of sensors with
JUMO digiLine electronics differs from that of all other analysis sensors. In this
case, there is only the "Calibration interval" parameter. Observe the following
configuration table.
10.12.1Configuration of the calibration timers
Open Calibration Timer Configuration Universal Inputs:
Device Menu > Configuration > Analog Inputs >
Universal Input 1 to 3 > Calibration Timer
Open Calibration Timer Configuration Analysis Inputs:
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Calibration Timer
Open Calibration Timer Settings Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > Calibration Timer
Configuration pointSelection/
Explanation
setting option
Function
Alarm/Event list
a
a
inactive
active
off
event
Activation/deactivation of the calibration timer
for an analog input
Assignment of calibration timer timeout notification to alarm list or event list
alarm
Alarm text
a
21 text charactersText for the alarm/event list upon timeout of cali-
bration timers
Calibration interval
a
0 to 9999 daysTime from one calibration to the next. When a
calibration is due is indicated by the calibration
alarm on the JUMO AQUIS touch S.
a
For sensors with JUMO digiLine electronics, only the calibration interval can be set. With pH and redox
sensors with JUMO digiLine electronics, the calibration alarm is active automatically. The texts for alarm
and event list are preset.
.
46
10.13Serial interfaces
The serial interface settings of all participating devices on a bus
must match.
Open: Device Menu > Configuration > Serial Interfaces >
Serial Interfaces 1 to 2
10 Configuration
Configuration pointSelection/
setting option
ProtocolModbus slave
Modbus digital sensors
Baud rate9600
19200
38400
Explanation
Communication protocol of the bus system
Modbus Slave: For operation of the device as
a slave in a Modbus system
Modbus digital sensors: For operation of
digiLine sensors on the serial interface (see
order details for extra code "JUMO digiLine
protocol activated")
In the JUMO AQUIS touch P, either the interface on the base unit or the optional serial interface (if present) can be configured for
digital sensors (JUMO digiLine operation). Simultaneous use of both interfaces for JUMO
digiLine operation is not possible.
Transmission speed (symbol rate) of the serial
interface
For all bus users (device and digital sensors)
to communicate with one another, they must
have the same baud rate.
a
The baud rates of JUMO tecLine and JUMO
digiLine sensors are automatically set by the
JUMO digiLine master-device during the
scanning process.
When connecting digital JUMO ecoLine sen-sors, the baud rate of the JUMO AQUIS
touch S must be set to "9600" prior to startup.
Otherwise, the sensors will not begin operation.
supported baud rates of JUMO digital sensors
•JUMO ecoLine: 9600
•JUMO tecLine and JUMO digiLine:
9600,19200, 38400
47
10 Configuration
Configuration pointSelection/
setting option
Data format8 - 1 - no parity
8 - 1 - odd parity
8 - 1 - even parity
Minimum response
time
0 to 500 msMinimum time from receiving a query to
Explanation
Format of the data word
For all bus users (device and digital sensors)
to communicate with one another, they must
have the same data format.
The data formats of JUMO tecLine and JUMO
digiLine sensors are automatically set by the
JUMO digiLine master-device during the
scanning process.
When connecting digital JUMO ecoLine sensors, the data format of the JUMO
AQUIS touch S must be set to "8-1-no parity"
prior to startup. Otherwise, the sensors will not
begin operation.
Format: Useful bit – stop bit – parity bit
sending a response
a
This parameter is used to adjust the response
speed of the device to slower bus users.
Device address1 to 254For "Modbus Slave" protocol only:
Unambiguous specification of a bus participant
a
For all users to communicate with one another, this setting be the same on all users.
48
11.1Important information
WARNING!
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
⇨ Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
11.2General information
The actual electrical characteristics of analysis sensors always deviate somewhat from the nominal specifications. The reasons for this include:
•Like every measuring instrument, analysis sensors always have a certain
uncertainty of measurement that results from manufacturing tolerances.
•During use, analysis sensors are exposed to chemical processes. Deposits
and wear phenomena caused by these processes result in changes of the
electrical characteristics of sensors.
To optimize the accuracy of measurements, analysis sensors must be calibrated. Calibrations are required:
•during installation or when changing a sensor
•regularly at time intervals that must be specified by the user
•if implausible measured values are displayed
•if process conditions change, e. g. as the result of equipment modification
11 Calibration in general
Calibration timers can be configured to provide a regular reminder of when calibrations are due.
⇨ Chapter 10.12 "Calibration timer", page 46
Each successfully completed calibration is recorded in the calibration
logbook.
⇨ Chapter 11.3 "Calibration logbook", page 51
11.2.1General procedure for calibration
True calibration (calibrating with routines)
After opening one of the calibration routines in the device, the user is guided
through a process with measurements and entries. In the course of this process,
the calibration values are determined and saved automatically.
Suitable calibration routines are available for every type of analysis sensor. The
individual calibration routines for the various sensor types are described in separate chapters.
⇨ Chapter 12 "Calibrating a pH measuring chain", page 55 to Chapter 16 "Cali-
brating Ci conductivity sensors", page 79
Execution of calibration routines requires that the following prerequisites be met:
49
11 Calibration in general
•You must be logged in as a user with the right to perform calibrations. Factory-preset users have all of these rights.
⇨ "Passwords and user rights", page 77
•You must ensure that the calibration default settings for the individual analysis inputs and, possibly, the universal inputs are set correctly. The explanations of the calibration default settings can be found in the chapters on
calibration of the various analysis sensors.
•In the case of Ci analysis inputs, it should be noted that the optional boards
must undergo a basic calibration during the initial commissioning. If this has
not yet been performed, it is necessary to do so prior to any other calibration.
⇨ Chapter 9.3 "Ci base calibration", page 121
Manual entry of calibration values
NOTE!
Incorrectly entered calibration values result in incorrect measured values.
Correct measurements are essential for control systems and limit
monitoring.
If calibration values are known, they can also be entered manually. This may be
the case with temperature-compensated conductivity measurements, for instance, when the temperature coefficient of a liquid is known. Another important
application case is the manual entry of relative cell constants of conductivity
sensors. If an ASTM test certificate showing precisely-measured cell constants
is available, in addition to entering the nominal cell constants in the configuration
of the conductivity sensor concerned (cf. Chapter 10.6.6 "CR/Ci analysis inputs
(conductive/inductive conductivity)", page 23), the relative cell constants of all
measuring ranges in the calibration values must be manually entered. The relative cell constants must be calculated from nominal cell constants (nameplate
for the sensor) and measured cell constants (ASTM test certificate):
(measured cell constant × 100 %) ÷ nominal cell constant = relative cell constant.
Calculation example:
nominal cell constant = 0.1 cm-1
measured cell constant = 0.1014 cm
rel. cell constant =
(0.1014 cm
Known calibration values are entered manually under:
Device menu > Calibration > Select Analog Input > Calibration Values
-1
× 100 %) ÷ 0.1 cm-1= 101.4 %
-1
50
11.3Calibration logbook
A separate logbook is maintained for each analysis and universal input.
The last 10 successful calibrations of the input concerned are saved in the cal-
ibration logbook. Canceled or failed calibrations (calibrations outside the permissible limits) are not saved in the logbook, but rather noted in the event list.
Manual changes of calibration values on the device are also documented. The
following data are retained in the logbook:
•Heading with description of the measurement input and calibration method
•Date and time
•Measurand
•Calibration assessment (assessment of the calibration values determined
during the true calibration)
•Calibration values determined or entered
•Reference values used
•Calibration mode (true calibration/manual entry of calibration values)
Since this information does not yet fit in a screen line, the logbook entries are
listed in abbreviated form with the date and calibration results for the time being.
More exact information can be accessed for every entry via the detail view.
11 Calibration in general
Example of a calibration logbook
Calibration assessment symbols
Calibration values are valid:
Sensor is OK
The calibration values determined are critical.
It is recommended that the sensor be cleaned.
Manual value input
For Ci analysis inputs (inductive conductivity) and universal inputs that have
been configured as a conductivity measurement input, a "TC curve" button is
also displayed. Tapping this button opens a list with the temperature coefficients
determined from the last "TC curve calibration".
51
11 Calibration in general
Example of a detail view of a logbook entry
The calibration logbook provides an overview of the calibrations performed.
Tapping the "Details" button opens the selected logbook entry in the detail view.
The detail view displays a table with all calibration values from a calibration procedure. The "Service" button is used for diagnostic purposes by trained personnel or JUMO Service.
52
11 Calibration in general
Assessment criteria
pH calibrations
(glass electrodes and ISFET connected to analysis measuring inputs as well as standard signals connected to universal inputs)
Calibration value [unit]
Zero point [pH]...<5≤...<6to8<...≤9<...
Slope [%]...<75≤...<89.6to103.1<...≤110<...
pH calibrations (antimony electrodes connected to analysis measuring inputs)
Calibration value [unit]
Zero point [pH]...<-2 to +2<...
Slope [%]...<10 to 110<...
Redox zero-point calibration
Calibration value [unit]
Zero point [mV]...< -200 ≤...<-120 to +120<...≤ +200 <...
NOTE!
There is no assessment of the calibration values in the case of a redox 2-point
calibration.
——
——
——
Ammonia calibration
Calibration value [unit]
Zero point [mV]...< -612 ≤...<-312 to +588<...≤ +888 <...
——
53
11 Calibration in general
vCalibration of conductivity sensors
(analysis measuring inputs and standard signals connected to universal inputs)
Calibration value [unit]
——
Rel. cell constant (CR) [%]...<50≤...<75 to 125<...≤150<...
Rel. cell constant (Ci) [%]...<80≤...<90 to 110<...≤120<...
Calibration value [unit]
——
Temperature coefficient
(CR) [%/K]
Temperature coefficient
(Ci) [%/K]
...<0to8<...
...<0to5.5<...
NOTE!
No assessment of the calibration values is performed for universal inputs in the
"linear scaling" operation mode.
54
12 Calibrating a pH measuring chain
12.1Important information
WARNING!
During the calibration, the relays and the analog output signals assume the
states configured under the respective "Response at calibration' configuration
points for the analog and digital outputs!
12.2General information
The calibration of pH electrodes is based on measurements in buffer solutions
with a defined pH-value. The pH values of the buffer solutions used are specified
either via entry of fixed values into the calibration default settings, entered during
the calibration or recognized automatically by "automatic buffer recognition"
during the calibration process. For "automatic buffer recognition", a buffer set table must be selected in the calibration default settings. In this case, the buffer
solutions used must be listed in the provided buffer set table. Since pH value
measurement of liquids is temperature-dependent, the temperature of the buffer
solution must be sensed in order to compensate for its effect on the result of the
measurement. This requires either manual entry or measurement with the aid of
a temperature sensor.
12.2.1Calibration methods for pH sensors
Zero-point calibration
This calibration method is used to determine the pH zero point on the measurement characteristic curve. The slope is retained.
A buffer solution with a defined pH value is needed as a reference.
Two-point Calibration
By measuring 2 different buffer solutions with defined pH values, the pH zero
point and pH slope of the measuring chain are established.
The pH values of the buffer solutions must be at least 2 pH apart. This calibration
method is recommended for most applications.
Three-point calibration
With a three-point calibration, the pH zero point as well as the pH slope in the
acidic region and the pH slope in the alkaline region are established. The threepoint calibration can be performed only for pH sensors on analysis inputs. It is
not available for JUMO digiLine pH sensors.
This method requires 3 buffer solutions with defined pH-values as references.
One of these must be acidic, one neutral and one alkaline. The pH values of the
buffer solution must be at least 2 pH apart from one another. This calibration
method is recommended for applications with more demanding accuracy requirements when performing measurements in both the alkaline and acidic regions.
55
12 Calibrating a pH measuring chain
Sample screen:
pH calibration default
settings
12.2.2Calibration default settings for pH sensors
Before you can perform a calibration, you must first enter the necessary calibration default settings. The possible settings for the pH Calibration are described
in the following.
Open the calibration default settings:
Device Menu > Calibration > Select Analysis Input or Digital Sensor Input for
pH/redox/NH3 > Calibration Presets
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in. The "Calibration default settings" menu for a digital sensor is visible for the digital sensor concerned only
when it is linked.
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
ParameterPossible settingsExplanation
Buffer set selectionBuffer set 1 to 3
Buffer 1 pH-value-2 to +16 pHmanual entry of the pH-values of the
Buffer 2 pH-value-2 to +16 pH
Buffer 3 pH-value-2 to +16 pH
j
factory-preset:
•Buffer set 1:
Reference buffer solutions for
calibrating pH measuring instruments acc. to DIN 19266
•Buffer set 2: Technical buffer
solutions, preferably for calibrating and adjusting technical
pH measuring instruments
acc. to DIN 19267
Buffer set tables contain pH values for
selected buffer solutions as a function of
temperature. These tables can be prepared/edited using commercially available standard solutions (DIN 19266,
NIST; otherwise, technical buffer solutions etc.) or on the basis of customerspecific information. With their aid,
buffer solutions can be recognized automatically when calibrating. The pH value data for the buffer solutions used
must be contained in the buffer set table
selected.
When a buffer set is selected, this activates automatic buffer recognition and
the entry fields for the settings "pH buffer 1 to 3" are hidden.
You need the JUMO PC setup program
to edit buffer set tables.
buffer solutions being used for the calibration
CompensationFixed compensation temperature
Temperature input
Interface
Depending on the selected Calibration
routine, the corresponding entry fields
for "pH buffer 1 to 3" are displayed.
The pH-values of the buffer solutions
used must be at least 2 pH apart.
Fixed compensation temperature:
Compensation with a fixed temperature
value that is entered in the configuration
item "Compensation temperature".
Temperature input: The integrated
temperature probe of the pH sensor
supplies the compensation temperature.
Interface: The AQUIS touch S transmits the compensation temperature to
the sensor electronics via the interface.
The source of the compensation temperature is set in the configuration point
"Compensation temperature".
57
12 Calibrating a pH measuring chain
ParameterPossible settingsExplanation
Temperature compensation
Selection from Analog selectionTemperature input for automatic acqui-
sition of the test/sample solution temperature during the calibration
58
12 Calibrating a pH measuring chain
tap button to
enter
temperature
12.3pH Calibration routines
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
⇨ Chapter 8.2.1 "Log-on/Log-out", page 91
NOTE!
Digital sensors need to be linked before they can be calibrated?
⇨ Chapter 8.2.7 "Digital sensors", page 97
12.3.1Zero-point calibration
StepAction
1Start the Zero-point calibration.
for pH sensors on analysis inputs:
Device Menu > Calibration > Select Analysis Input for pH/redox/
NH
> Open Zero-Point Calibration
3
for pH sensors with JUMO digiLine electronics:
Device Menu > Calibration > Digital Sensor 1 to 6 > Open ZeroPoint Calibration
2If temperature compensation was not specified in the calibration
default settings, enter the temperature of the buffer solution here
manually.
If temperature compensation was specified, the
temperature of the buffer solution is determined automatically.
3Clean the pH electrode and immerse it in the buffer solution.
59
12 Calibrating a pH measuring chain
tap button to enter pH-value of
the buffer
field
StepAction
4Entry of the pH-value of the buffer solution
•without buffer recognition:
Check whether the "pH buffer 1" matches the pH-value of the
buffer solution used. If a buffer set table was not specified, the
"buffer 1 pH" value is taken from the calibration default settings.
It is still possible to change this manually here.
•with buffer recognition:
A prerequisite here is that a buffer set table has been selected
in the buffer set table and the pH-value of the buffer solution
used is contained in this buffer set table. If these prerequisites
are satisfied, the pH value of the buffer solution is determined
automatically during the calibration.
5Wait until the measured value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK"
6A Protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
7Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
60
12 Calibrating a pH measuring chain
To enter the temperature, tap the
button
12.3.2Two-point and three-point calibration
StepAction
1Start the desired Calibration routine.
for pH sensors on analysis inputs:
Device Menu > Calibration > Analysis Input Select Analysis Input
for pH/redox/NH
for pH sensors with JUMO digiLine electronics:
Device Menu > Calibration > Digital Sensor 1 to 6 > Open TwoPoint Calibration
2If temperature compensation was not specified in the calibration
default settings, enter the temperatures of the buffer solution here
manually.
If temperature compensation was specified, the
temperature of the buffer solution is determined automatically.
3
> Open Two-Point or Three-Point Calibration
3Clean the pH electrode and immerse it in one of the buffer solu-
tions.
For a Two-point calibration, you need 2 buffer solutions.
For a three-point calibration, you need 3 buffer solutions (acidic,
neutral, and alkaline).
61
12 Calibrating a pH measuring chain
tap button to enter pH-value of
the buffer
field
StepAction
4Entry of the pH-value of the buffer solution
•without buffer recognition:
Check whether the "pH buffer 1" matches the pH-value of the
buffer solution used. If a buffer set table was not specified, the
"buffer 1 pH" value is taken from the calibration default settings.
It is still possible to change this manually here.
•with buffer recognition:
A prerequisite here is that a buffer set table has been selected
in the buffer set table and the pH-value of the buffer solution
used is contained in this buffer set table. If these prerequisites
are satisfied, the pH value of the buffer solution is determined
automatically during the calibration.
5Wait until the measured value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".
6For each additional calibration point, repeat steps 3 to 5 with the
required buffer solutions.
7A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
8Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
62
13 Calibrating redox sensors
13.1Important information
WARNING!
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
⇨ Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
13.2General information
The calibration of Redox sensors is based on measurements in test solutions
with a defined Redox potential.
13.2.1Calibration methods for Redox sensors
Zero-point calibration
This calibration method is used to determine the redox zero point.
A test solution with a defined redox potential is needed as a reference.
"mV" must be set as the redox unit in the configuration of the redox measure-
ment input (analysis input or input for digital sensors).
This calibration is used to establish an application-specific measurement characteristic curve where redox potentials are represented as a percentage of concentration values. The redox potentials of 2 solutions are measured.
Concentration values in percent are then assigned to the measured values by
the user.
Two process-typical sample solutions are needed as calibration solutions as a
reference.
"Percent" must be set as the redox unit in the configuration of the redox measurement input (analysis input or input for digital sensors).
Example: In a detoxification plant, the toxicity of a liquid is to be measured on
the basis of the Redox potential. The calibration requires 2 solutions:
•the concentration of the highly toxic solution is given as 80 % by the user, for
•the concentration of the detoxified solution is given as 10% by the user, for
Using the redox potential, the toxicity can now be measured and displayed in
percent.
13.2.2Calibration default settings for Redox sensors
Before you can perform a calibration, you must first enter the necessary calibration default settings. The possible settings for the Redox calibration are described in the following.
Open the calibration default settings:
Device Menu > Calibration > Select Analysis Input for pH/redox/NH
tion Presets
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in. The "Calibration default settings" menu for a digital sensor is visible for the digital sensor concerned only
when it is linked.
Redox test solution -1500 to +1500 mVManual entry of the Redox potential of
64
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
the test solution being used for calibration
NOTE!
Remember that the configuration of the redox measurement input must be set
to the redox unit "mV" for the zero-point calibration and to "Percent" for the twopoint calibration.
Device Menu > Calibration > Select Analysis Input for pH/redox/
NH
> Zero-Point Calibration
3
for redox sensors with JUMO digiLine electronics:
Device Menu > Calibration > Digital Sensor 1 to 6 > Zero-Point
Calibration
3Check that the "Redox test solution" value displayed matches the
Redox value of the test solution.
The "Redox test solution" value is taken from the calibration default
settings. It is still possible to change this manually here.
65
13 Calibrating redox sensors
StepAction
4Clean the Redox electrode and immerse it in the test solution. Wait
until the measured value displayed stabilizes and then confirm the
result of the measurement by pressing "OK".
5A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
66
13.3.2Two-point Calibration
Tap button to
enter field
StepAction
1Ensure that
•the calibration default settings are correct
•"Percent" is set as the Redox unit in the configuration of the Re-
⇨ Chapter 13.2.2 "Calibration default settings for Redox sensors",
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
⇨ Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
14.2General information
The calibration of ammonia sensors is based on measurements in ammoniafree test solutions.
14.2.1Calibration methods for ammonia sensors
Zero-point calibration
This calibration method is used to determine the ammonia zero point.
An ammonia-free test solution (e.g. water) is needed as a reference.
14.2.2Calibration default settings for ammonia sensors
In the ammonia sensor calibration default settings, the zero-point calibration is
enabled and preconfigured as the only available calibration routine.
Open the calibration default settings:
Device Menu > Calibration > Select Analysis Input for pH/redox/NH
tion Presets
> Calibra-
3
69
14 Calibrating ammonia sensors
14.3Ammonia calibration routines
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
⇨ Chapter 8.2.1 "Log-on/Log-out", page 91
14.3.1Zero-point calibration
StepAction
1Start the Zero-point calibration.
Device Menu > Calibration > Select Analysis Input for pH/redox/
NH
> Zero-Point Calibration
3
2Clean the ammonia electrode and immerse it in the ammonia-free
test solution. Wait until the measured value displayed stabilizes
and then confirm the result of the measurement by pressing "OK".
3A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
4Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
70
15 Calibrating CR conductivity sensors
15.1Important information
WARNING!
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
⇨ Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
15.2General information
The calibration of CR sensors is based on measurements in test solutions with
a defined electrolytic conductivity. Since the electrolytic conductivity of liquids is
temperature dependent, the temperature of the test solution must be sensed.
This requires either manual entry or measurement with the aid of a temperature
sensor.
15.2.1Calibration methods for CR conductivity sensors (conductive)
Rel. cell constant
The deviation from the nominal cell constant of a CR sensor is described by the
relative cell constant. The rel. cell constant is determined by making a measurement in a test solution with a defined conductivity.
Temperature coefficient
The temperature coefficient is a measure of the temperature dependence of the
electrolytic conductivity of a liquid. It is used to compensate for the effect of temperature when measuring the electrolytic conductivity. When performing a temperature-compensated conductivity measurement, the conductivity value
measured is always indicated with reference to the fixed reference temperature.
With the aid of the temperature coefficient, the value of the electrolytic conductivity displayed at the reference temperature is calculated from the current measured values of conductivity and temperature of the liquid.
Reference temp. is set in the configuration of the individual CR analysis input.
page 23
The temperature coefficient is determined from 2 measurements in a test solution at different temperatures (reference and operation temp.).
NOTE!
If the temperature coefficient of a sample solution is known, it can be entered
directly.
⇨ Chapter 11.2.1 "General procedure for calibration", page 49
71
15 Calibrating CR conductivity sensors
Sample screen:
CR calibration presets
15.2.2Calibration presets for CR conductivity sensors
Before you can perform a calibration, you must first enter the necessary calibration default settings. The possible settings for the CR calibration are described
in the following.
Open the calibration default settings:
Device menu > Calibration > Select CR Analysis Input >
Calibration Presets
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
⇨ Chapter 8.2.1 "Log-on/Log-out", page 91
72
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
15 Calibrating CR conductivity sensors
Calibration presets for calibrating the relative cell constant
ParameterPossible settingsExplanation
Reference conductivi-ty0 to 9999 mS/cmConductivity of the reference solution
Calibration presets for calibrating the temperature coefficient
ParameterPossible settingsExplanation
Temperature compensation
Reference temperature
Operation temp.-50 to +150 °C
Selection from
analog selection
-50 to +150 °CThe conductivities of a sample solution at reference
Temperature input for automatic acquisition of the
test/sample solution temperature during the calibration
temperature and working temperature are captured
during the calibration process. This yields 2 value
pairs (temperature/conductivity). These value pairs
provide the basis for calculating the temperature coefficient.
The operation temp. must differ from the reference
temperature by at least 5 °C.
73
15 Calibrating CR conductivity sensors
Tap button to
enter measuring
range
15.3CR calibration routines
NOTE!
Conductivity measuring inputs can be configured with measuring range changeover. Accordingly, calibrations must be performed for all "accessible measuring
ranges".
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
⇨ Chapter 8.2.1 "Log-on/Log-out", page 91
15.3.1Calibrating the relative cell constant
StepAction
1Start calibration of the relative cell constant.
2Clean the sensor and immerse it in the sample solution. Ensure
that the rel. cell constant is calibrated correctly (if necessary, make
a trial measurement with a test solution).
For CR optional boards, continue with step 3;
for universal inputs set to the "conductivity measurement" operation mode, continue with step 4
3Enter one of the measuring ranges 1 to 4.. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected measuring range.
76
15 Calibrating CR conductivity sensors
requested
Temperature values
current temperature
value
Display after
acquisition of first
value
remaining
requested
Temperature value
StepAction
4•with temperature acquisition
A prerequisite is that temperature compensation was specified in
the calibration the presets.
Bring the temperature of the sample solution to the requested values of the reference and operation temperatures in succession.
The order does not matter. Acquisition of the individual values
takes place automatically.
•without temperature acquisition
If temperature compensation was not specified in the calibration
presets, you must control acquisition of the value manually. Bring
the temperature of the sample solution first to the value of the reference temperature and confirm by pressing "OK".
Then proceed in the same manner for the operation temperature.
5A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
77
15 Calibrating CR conductivity sensors
78
16 Calibrating Ci conductivity sensors
16.1Important information
WARNING!
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
⇨ Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
16.2General information
The calibration of Ci sensors is based on measurements in test solutions with a
defined electrolytic conductivity. Since the electrolytic conductivity of liquids is
temperature dependent, the temperature of the test solution must be sensed.
This requires either manual entry or measurement with the aid of a temperature
sensor.
16.2.1Calibration methods for Ci conductivity sensors (inductive)
Rel. cell constant
The deviation from the nominal cell constant of a Ci sensor is described by the
relative cell constant. The rel. cell constant is determined by making a measurement in a test solution with a defined conductivity.
Temperature coefficient
The temperature coefficient is a measure of the temperature dependence of the
electrolytic conductivity of a liquid. It is used to compensate for the effect of temperature when measuring the electrolytic conductivity. When performing a temperature-compensated conductivity measurement, the conductivity value
measured is always indicated with reference to the fixed reference temperature.
With the aid of the temperature coefficient, the value of the electrolytic conductivity displayed at the reference temperature is calculated from the current measured values of conductivity and temperature of the liquid.
Reference temperature is set in the configuration of the individual Ci analysis input.
page 23
The temperature coefficient is determined from 2 measurements in a test solution at different temperatures (reference and operation temp.).
NOTE!
If the temperature coefficient of a sample solution is known, it can be entered
directly.
⇨ Chapter 11.2.1 "General procedure for calibration", page 49
79
16 Calibrating Ci conductivity sensors
Conductivity
6 calibration points
5 intervals
Temperature
TC curve (for nonlinear temperature coefficients)
If the conductivity of a liquid whose temperature coefficient changes with temperature has to be measured, this method can determine 5 temperature coefficients for 5 temperature intervals. In this way, it is possible to determine a good
approximation of the temperature coefficient curve. While the operator brings
the sample solution to the temperature values requested by the device, the device determines the temperature coefficient for each interval. This requires installation of a temperature sensor that the device can use to sense the
temperature of the sample solution.
16.2.2Calibration presets for Ci conductivity sensors
Before you can perform a calibration, you must first enter the necessary calibration default settings. The possible settings for the Ci calibration are described in
the following.
Open the calibration default settings:
Device menu > Calibration > Select Ci Analysis Input or Universal Input >
Calibration Presets
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
⇨ Chapter 8.2.1 "Log-on/Log-out", page 91
80
16 Calibrating Ci conductivity sensors
Sample screen:
Ci calibration presets
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
Calibration presets for calibrating the relative cell constant
ParameterPossible settingsExplanation
Reference conductivi-ty0 to 9999 mS/cmConductivity of the reference solution
Calibration presets for calibrating the temperature coefficient
ParameterPossible settingsExplanation
Temperature compensation
Reference temperature
Operation temp.-50 to +150 °C
Selection from
analog selection
-50 to +150 °CThe conductivities of a sample solution at reference
Temperature input for automatic acquisition of the
test/sample solution temperature during the calibration
temperature and working temperature are captured
during the calibration process. This yields 2 value
pairs (temperature/conductivity). These value pairs
provide the basis for calculating the temperature coefficient.
The operation temp. must differ from the reference
temperature by at least 5 °C.
81
16 Calibrating Ci conductivity sensors
Calibration presets for calibrating the TC curve
ParameterPossible settingsExplanation
Temperature compensation
Starting temperature-50 to +250 °CThe starting and final temperatures of the range for
Final temperature-50 to +250 °C
Selection from
analog selection
NOTE!
Calibration of the TC curve is possible only with automatic temperature acquisition.
Temperature input for automatic acquisition of the
test/sample solution temperature during the calibration
which a temperature coefficient curve is to be determined.
The starting temperature must be at least 20 °C lower than the final temperature. The reference temperature for the measurement input must lie
between the starting and final temperatures, and differ from the starting and final temperatures by at
least 2 °C.
82
16 Calibrating Ci conductivity sensors
tap button to
enter measuring
field
16.3Ci calibration routines
NOTE!
Conductivity measuring inputs can be configured with measuring range changeover. Accordingly, calibrations must be performed for all "accessible measuring
ranges".
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
⇨ Chapter 8.2.1 "Log-on/Log-out", page 91
NOTE!
Analysis inputs for inductive conductivity measurements (Ci) must undergo a
basic Ci calibration in the course of commissioning. Subsequent calibration is
not possible without this initial basic Ci calibration.
⇨ Chapter 9.3 "Ci base calibration", page 121
16.3.1Calibrating the relative cell constant
StepAction
1Start calibration of the relative cell constant.
Device menu > Calibration > Select Ci analysis input or universal
input > Relative time constant calibration
with Ci optional board, proceed with Step 2;
for universal inputs set to the "conductivity measurement" operating mode, continue with step 3;
2Enter one of the measuring ranges 1 to 4.. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected measuring range.
83
16 Calibrating Ci conductivity sensors
tap button to
change
reference
conductivity
field
StepAction
3Ensure that
•the sensor has been cleaned and is immersed in the test solution,
•the set reference conductivity matches the conductivity value of
the test solution.
Wait until the measured value displayed stabilizes and then confirm the result of the measurement by pressing "OK".
The preset reference conductivity can be changed here manually
if necessary.
4A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
5Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
84
16 Calibrating Ci conductivity sensors
tap button to
enter measuring
field
16.3.2Calibrating the temperature coefficient (TC)
StepAction
1Start calibration of the temperature coefficient.
Device menu > Calibration > Select Ci analysis input or universal
input > TC calibration
2Clean the sensor and immerse it in the sample solution. Ensure
that the rel. cell constant is calibrated correctly (if necessary, make
a trial measurement with a test solution).
for Ci optional boards, continue with step 3;
for universal inputs set to the "conductivity measurement" operating mode, continue with step 4
3Enter one of the measuring ranges 1 to 4.. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected measuring range.
85
16 Calibrating Ci conductivity sensors
requested
Temperature values
current temperature
value
Display after
acquisition of first
value
remaining
requested
Temperature value
StepAction
4•with temperature acquisition
A prerequisite is that temperature compensation was specified in
the calibration the presets.
Bring the temperature of the sample solution to the requested values of the reference and operation temperatures in succession.
The order does not matter. Acquisition of the individual values
takes place automatically.
86
•without temperature acquisition
If temperature compensation was not specified in the calibration
presets, you must control acquisition of the value manually. Bring
the temperature of the sample solution first to the value of the reference temperature and confirm by pressing "OK".
Then proceed in the same manner for the operation temperature.
5A Protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
16 Calibrating Ci conductivity sensors
tap button to
enter measuring
field
requested
Temperature values
current temperature
value
16.3.3Calibrating the TC curve
StepAction
1Start the desired calibration of the TC curve.
Device menu > Calibration > Analysis input 1 to 4 (Ci) or universal
input 1 to 3
2Clean the sensor and immerse it in the sample solution. Ensure
that the rel. cell constant is calibrated correctly (if necessary, make
a trial measurement with a test solution).
for Ci optional boards, continue with step 3;
for universal inputs set to the "conductivity measurement" operating mode, continue with step 4
3Enter one of the measuring ranges 1 to 4.. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected measuring range.
rTC curve
4Ring the temperature of the sample solution to the requested set-
points in succession. Six temperatures are requested.
5A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
87
16 Calibrating Ci conductivity sensors
88
17 Calibrating universal inputs
Universal
input Operation modes
Calibration
routine
17.1Important information
WARNING!
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
⇨ Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
17.2General information
17.2.1Calibration methods for universal inputs
Universal inputs can be configured with various operation modes for a number
of different process variables (see table below).
Detailed information on the possible configurations:
⇨ Chapter 10.6.2 "Universal inputs of base unit and optional boards", page 17
Appropriate calibration routines for each operation mode of a universal input can
be enabled in the calibration default settings.
The following table lists the availability of calibration routines for the individual
operation modes.
linear
scaling
Zero-point calibrationXX
Slope calibrationXX
Two-point CalibrationXX
Rel. cell constant X
Temperature coefficientX
TC curveX
Zero-point calibration
This calibration method is used to determine the zero point on the measurement
characteristic curve. The slope is retained.
A test solution with a defined value of the respective measurand is needed as a
reference.
Slope calibration
This calibration method is used to determine the slope of the measurement
characteristic curve. The zero point is retained.
A test solution with a defined value of the respective measurand is needed as a
reference.
Temp. measurement
pH value
measurement
Conductivity
measurement
free chlorine,
pH/T-compensated
89
17 Calibrating universal inputs
Two-point Calibration
The zero point and slope of the measuring characteristic curve are determined
with the aid of 2 measurings of 2 different reference solutions.
Two test solutions with defined values of the respective measurand are needed
as a references.
Calibrating rel. cell constants, temperature coefficients, and TC curves
The calibration routines for conductivity measuring with the universal input correspond to those for the Ci calibration.
⇨ Chapter 16 "Calibrating Ci conductivity sensors", page 79
90
17 Calibrating universal inputs
Sample screen:
Universal input calibration default settings in
the "linear scaling" operation mode
Which calibration default settings are available depends on the configuration
settings of the universal input.
Open the calibration default settings:
Device menu > Calibration > Select Universal Input >
Calibration default settings
Calibration default settings for the individual operation modes
•linear scaling
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
The "linear scaling" operation mode permits adjustment of the zero point, slope,
and two-point calibration.
•pH value measurement
The calibration default settings correspond to those for the zero-point and twopoint calibration for pH/redox/NH
analysis inputs.
3
⇨ "pH calibration default settings", page 57
•Conductivity measurement
The calibration default settings correspond to those for the Ci calibration.
⇨ Chapter 16.2.2 "Calibration presets for Ci conductivity sensors", page 80
91
17 Calibrating universal inputs
Sample screen:
Universal input calibration default settings in
the
"free chlorine, pH/Tcompensated" operation mode
•free chlorine, pH/Temp.-compensated
In the universal inputs calibration default settings for the "free chlorine, pH/Tcompensated" operation mode, the slope calibration is enabled and preconfigured as the only available calibration routine.
Additional calibration default settings for calibration of chlorine sensors are explained in the following table.
ParameterPossible settingsExplanation
Temperature compensation
pH compensation
source
Selection from Analog selectionTemperature input for automatic acqui-
sition of the test/sample solution temperature during the calibration
Selection from Analog selectionpH measurement input for automatic
sensing of the test/sample solution pH
value during the calibration
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
⇨ Chapter 8.2.1 "Log-on/Log-out", page 91
92
17 Calibrating universal inputs
17.3Universal input calibration routines
This chapter explains the universal inputs calibration routines for the "linear scaling" and "free chlorine, pH/T-compensated" operation modes.
The explanations in the corresponding calibration chapters for the "pH value
measurement" and "Conductivity measurement" operation modes apply, except
that three-point calibration for pH sensors is not available for universal inputs
(see Chapter 17.2.1 "Calibration methods for universal inputs", page 89).
bration
2Clean the sensor and immerse it in the test solution.
3Check the displayed values for the pH value and
temperature. Automatic sensing can be configured in the calibra-
tion default settings for both measurands independently of one an-
other. With automatic acquisition, the particular influencing
variable is only displayed and can no longer be changed here.
Without automatic acquisition, the particular influencing variable
must be entered manually here.
Wait until the measured value displayed stabilizes and then check
the pH value displayed. Then confirm by pressing "OK".
Example with temperature acquisition and without pH value acqui-
sition
4Enter the concentration value of the test solution. Confirm by
pressing "OK"
97
17 Calibrating universal inputs
StepAction
5A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
98
18 Calibrating O-DO sensors
18.1Important information
WARNING!
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
⇨ Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
18.2General information
Depending on the calibration method, O-DO sensors are calibrated on the basis
of measurements in water and water sulfite solutions (sulfite
concentration < 2 %).
NOTE!
As the manufacturing of air-saturated water requires a high level of effort and is
difficult to reproduce, the easier process of calibration in water vapor-saturated
air is recommended for the operational calibration.
Positioning the sensor in the water vapor-saturated air
To achieve successful calibration, the following points must be taken into account:
•The sensor must be kept dry during the calibration process. Drops of water
adhering to the sensor membrane could distort the measurement result.
•The air pressure and temperature must remain constant during the calibration.
99
18 Calibrating O-DO sensors
Sample screen:
O-DO
calibration default settings
18.2.1Calibration methods for O-DO sensors
End value calibration
With end value calibration, the slope of the sensor is calibrated beyond the defined state of 100 % oxygen saturation. This state can in principle be achieved
in two ways:
•By positioning the sensor in water vapor-saturated air (for example, directly
over a water surface).
•By positioning the sensor in air-saturated water (air is directed through water
until the water is saturated with it).
Two-point Calibration
With two-point calibration, the zero point and slope of the sensor are calibrated.
This calibration method offers the greatest possible level of accuracy and is particularly recommended for measurements of small oxygen concentrations.
18.2.2Calibration default settings for O-DO sensors
Before you can perform a calibration, you must first enter the necessary calibration default settings. The possible settings for the O-DO calibration are described in the following.
Open the calibration default settings:
Device Menu > Calibration > Digital Sensor 1 to 6 >
Calibration Default Settings
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in. The "Calibration default settings" menu for a digital sensor is visible for the digital sensor concerned only
when it is linked.