JUMO AQUIS 500 pH, 202560, AQUIS 500 Operating Instructions Manual

JUMO AQUIS 500 pH
Transmitter/Controller for pH, ORP and
NH
3
- (ammonia) concentration Type 202560
B 20.2560.0
2012-04-12/00475451
WARNING:
A sudden malfunction of the instrument, or one of the sensors connected to it, could potentially result in dangerous, imprecise dosing! Suitable preventive measures must be in place to prevent this from happening.
Note:
Please read these Operating Instructions before placing the instrument in operation. Keep the manual in a place which is accessible to all users at all times.
Resetting the brightness of the LC display:
If the brightness/contrast setting has been adjusted so that the display text is no longer legible, the basic setting can be restored as follows:
Switch off the supply voltage. Switch on the supply voltage and immediately press and hold the and
keys simultaneously.
Reset the language to "English":
If the language has been adjusted so that the display text is no longer comprehensible, use the Administrator password, 7485, to reset the language to "English":
Press the key for longer than 3 seconds. Press the key once. Briefly press the key. Enter 7485. Briefly press the key. The required language can then be set in
ADMINISTR. LEVEL / PASSWORD / PARAMETER LEVEL / DISPLAY / LANGUAGE.
PGM
PGM
PGM
Contents
1 Typographical conventions ...................................................... 6
1.1 Warning signs ..............................................................................................6
1.2 Note signs ....................................................................................................6
2 Description ................................................................................ 7
3 Identifying the instrument version ........................................... 8
3.1 Nameplate ....................................................................................................8
3.2 Type designation ..........................................................................................9
3.3 Scope of delivery .........................................................................................9
3.4 Accessories (in delivery package) ..............................................................10
3.5 Accessories (optional) ................................................................................11
4 Electrical connection .............................................................. 12
4.1 Installation notes ........................................................................................12
4.2 Electrical isolation ......................................................................................13
4.3 General .......................................................................................................14
4.4 Connection .................................................................................................14
4.5 Terminal assignments ................................................................................19
4.6 ISFET-pH combination electrode according to data sheet 20.1050 .........21
5 Mounting .................................................................................. 22
5.1 General .......................................................................................................22
5.2 Surface mounting .......................................................................................22
5.3 Pipe installation set / weather protection roof ...........................................23
5.4 DIN rail installation set ...............................................................................23
5.5 Mounting in a panel ....................................................................................24
6 Operation ................................................................................. 26
6.1 Controls ......................................................................................................26
6.2 LC display ..................................................................................................27
6.3 Principle of operation .................................................................................28
6.4 Measurement mode ...................................................................................31
6.5 Input/output information ............................................................................31
6.6 MANUAL mode / simulation mode ............................................................33
6.7 HOLD mode ...............................................................................................37
6.8 Operator level .............................................................................................38
6.9 Administrator level .....................................................................................38
6.10 Device info .................................................................................................44
6.11 Controller function .....................................................................................44
Contents
7 Commissioning ....................................................................... 46
7.1 Fast start ....................................................................................................46
7.2 Setup examples .........................................................................................47
8 Calibration ............................................................................... 68
8.1 pH electrode ..............................................................................................68
8.2 pH antimony electrode ...............................................................................76
8.3 ORP electrode ............................................................................................76
8.4 Ammonia (NH
3
)- cell ..................................................................................80
9 Setup program ........................................................................ 82
9.1 Function .....................................................................................................82
10 Eliminating faults and malfunctions ...................................... 83
10.1 Possible faults ............................................................................................83
11 Appendix .................................................................................. 84
11.1 Operator level parameters .........................................................................84
11.2 Parameter explanations .............................................................................90
11.3 Glossary .....................................................................................................94
12 Instrument description ........................................................... 98
12.1 Technical data ..........................................................................................98
12.2 Panel cut-out ............................................................................................101
Ziffern
1-point calibration 43 1-point calibration, ammonia 80 1-point calibration, pH 68, 76, 78 2-point calibration 43 2-point calibration, pH 70 3-point calibration 43 3-point calibration, pH 72
A
Accessories 1011 Acidity errors 43 Administrator level 38 Administrator rights 40 Alkaline errors 43 Asymmetrical connection 1617, 19, 9495
B
Basic settings 42 Block diagram 7
C
Cable routing 15 Cell zero point 43 Conductor cross-sections 12
D
Date of manufacture 8 Device info 44
E
Electrical connection 12 Electrical isolation 13, 82 Enable level 40
F
Fast start 46 Fixing brackets 22
H
HOLD mode 37
I
Inputs 19 Installation position 22 ISFET sensor 21, 95, 99
L
Logbook 44
M
MANUAL mode 33 MANUAL mode for analog outputs 36 MANUAL mode for switching outputs 33
Measurement mode 27, 31 MIN/MAX values 31 Mounting location 22
N
Nameplate 8 Normal display 27, 31
O
Operator level 38 Output level display 32 Outputs 20 Overview of MANUAL mode 34
P
Panel mounting 24 Parameter level 40 Password 38 Pipe 23 Pipe installation 23 Principle of operation 28
S
Scope of delivery 9 Sensor connection 1618 Setup examples 47 Setup interface 82 Simulation mode 33 Simulation of switching outputs 35 Slope 43 slope 86 Sunlight 22 Supply 19, 21 Surface 22 Surface mounting 22 switching function 90 Symmetrical connection 1819
T
Terminal assignments 19 Type designation 9
W
Wash timer 97 Weather protection roof 23
Z
zero point 86
6
1 Typographical conventions
1.1 Warning signs
1.2 Note signs
Danger
This symbol is used when there may be danger to personnel if the instructions are disregarded or not followed accurately!
Caution
This symbol is used when there may be damage to equipment or data if the instructions are ignored or not followed accurately!
Caution
This symbol is used where special care is required when handling components liable to damage through electrostatic discharge.
Note
This symbol is used when your special attention is drawn to a remark.
abc
1
Footnote
Footnotes are remarks that refer to specific points in the text. Footnotes consist of two parts:
A marker in the text, and the footnote text. The markers in the text are arranged as continuous superscript numbers.
Action instruction
This symbol indicates that an action to be performed is described. The individual steps are marked by this asterisk. Example: Remove crosspoint screws.
7
2 Description
General The instrument is used for measuring/controlling the pH, ORP or NH3-
(ammonia) concentration. The function is switchable on the instrument itself. Depending on the measured variable, combination electrodes (e. g. pH/ORP combination electrodes, gas-sensitive sensors) or split versions (glass/metal electrodes with a separate reference electrode) can be readily connected. Temperature serves as the second input variable, measured by a Pt100/1000 probe, for example. It is therefore possible to implement automatic temperature compensation for the pH and NH
3
variables.
The instruments are operated using unambiguous keys and a large LC graphics display on which the measurements are clearly legible. The plain-text presentation of the parameters makes it easier for the user to configure the instrument, and also helps in programming it correctly.
Thanks to its modular design, the instrument can be perfectly matched to the specific application requirements. Up to four outputs are available (see the block diagram for the functions).
Typical areas of application
Universal application in water and wastewater engineering, service/process water and wastewater, drinking water and well/surface water, leakage monitoring in refrigeration plant.
Block diagram
Transmitter/controller
2 analog inputs
Input 1: pH / ORP / ammonia concentration
1 binary input
For floating contact Functions:
- key inhibit
- alarm stop
- HOLD
Supply
110—240 V 12 — 24 VACDC 20—30 V AC/DC
Analog outputs (option)
Outputs 1 + 2: 0(4) –20 mA or 0 –10 V
option
Setup interface
Switching outputs (option)
Outputs 3 + 4:
- relay, changeover (SPDT)
User-friendly configuration Reloading of languages Plant documentation
Configurable as analog process value output and/or continuous controller output (PID action)
Configurable as
- limit controller
- pulse width output (PID action)
- pulse frequency output (PID action)
- modulating controller (PID action)
- USP contact
- "Purified water" contact as per Ph.Eur.
- calibration timer run down
standard
Input 2: temperature manual input or automatic Pt100 / Pt1000 / 4 kΩ
Supply
for ISFET sensor
8
3 Identifying the instrument version
3.1 Nameplate
on the transmitter
Typ: 202560/10-888-000-000-000-23/000
VARTN: 20/00491200
JUMO
Fulda, Germany www.jumo.net
AQUIS 500 pH
AC 110..240V -15/+10% 48..63Hz
< 14VA
F-Nr.: 0168122901012150001
The date of manufacture is coded in the “F-Nr.” (serial number): 0935 means manufactured in year 2012 / week 15
9
3 Identifying the instrument version
3.2 Type designation
3.3 Scope of delivery
- Transmitter/controller
- 1 bag with accessories
- Operating Instructions
(1) Basic type
202560 JUMO AQUIS 500 pH
Transmitter/controller for pH, ORP, NH
3
- (ammonia) concentration and temperature
(2) Basic type extensions
10 for panel mounting 20 in surface-mountable housing
(3) Output 1 (for principle measurement variable or continuous controller)
000 no output 888 analog output 0(4) — 20 mA or 0 — 10 V
(4)
Output 2 (for temperature measurement variable or continuous controller)
000 no output 888 analog output 0(4) — 20 mA or 0 — 10 V
(5) Output 3
000 no output 310 relay with changeover (SPDT) contact
(6) Output 4
000 no output 310 relay with changeover (SPDT) contact
(7) Supply voltage
23 110 — 240 V AC +10%/-15%, 48 — 63 Hz 25 20 — 30 V AC/DC, 48 — 63 Hz 30 12 — 24 V DC
±15%
(8) Extra codes
000 none
(1) (2) (3) (4) (5) (6) (7) (8)
Order code /-----/ Order example 202560 / 20 - 888 - 000 - 310 - 000 - 23 / 000
3 Identifying the instrument version
10
3.1 Accessories (in delivery package)
Contents Designation
3 x plug-in screw terminals
1 x large plug-in link
1 x small plug-in link
1 x cable clip for cable diameter > 5 mm
2 x cable clips for cable diameter < 5 mm
1 x cable clip for cable diameter < 3 mm
2 x pan head screws 3.5x6.5 4 x round spacers for panel mounting 4 x hexagonal nuts for panel mounting
4 x countersunk screws M6x10
4 x fixings
1 x cable gland M12x1.5
1 x sealing ring for cable gland M12x1.5
1 x reducing sealing ring for cable gland M12x1.5
2 x cable glands M16x1.5
2 x sealing rings for cable gland M16x1.5
1 x multiple seal for cable gland M16x1.5
11
3 Identifying the instrument version
3.4 Accessories (optional)
Type Part n o.
Protection canopy for JUMO AQUIS 500
1
00398161
Pole-mounting kit for JUMO AQUIS 500
2
00483664 Support pillar with pedestal base, arm and chain 00398163 PC setup software 00483602 PC interface, including USB/TTL converter and
adapter (USB connecting cable)
00456352
Fixing for suspended fitting 00453191
1
The pole-mounting kit is needed for mounting the protection canopy.
2
Using the pole-mounting kit, the JUMO AQUIS 500 can be fitted to a pole (e.g. support pillar or railing).
12
4 Electrical connection
4.1 Installation notes
The choice of cable, the installation and the electrical connection must
conform to the requirements of VDE 0100 “Regulations on the Installation of Power Circuits with Nominal Voltages below 1000 V” or the appropriate local regulations.
If contact with live parts is possible while working on the device, it must be
completely disconnected from the electrical supply.
Load circuits must be fused for the maximum relay current in each case, in
order to prevent welding of the relay contacts in the event of a short circuit.
The electromagnetic compatibility conforms to EN 61326.Run input, output and supply cables separately and not parallel to one
another.
Uses shielded sensor cables with twisted conductors. Do not run these
cables close to current-carrying components or cables. Ground shielding at one end.
Sensor leads should be implemented as uninterrupted cables (not routed
via terminal blocks etc.).
Do not connect any additional loads to the supply terminals of the
instrument.
The instrument is not suitable for use in areas with an explosion hazard (Ex
areas).
Apart from faulty installation, incorrect settings on the instrument may also
affect the proper functioning of the subsequent process or lead to damage. Safety devices independent of the instrument should therefore always be provided and should only be capable of adjustment by specialist personnel.
Conductor cross-sections and core-end ferrules
Fitting sizes
The electrical connection must only be carried out by qualified professional persons !
Minimum cross-section
Maximum cross-section
Min. length of core-end ferrule
Without core-end ferrule 0.34mm
2
2.5mm
2
10mm (stripped)
Core-end ferrule, no lip 0.25mm
2
2.5mm
2
10mm
Core-end ferrule, lip up to 1.5mm
2
0.25mm
2
1.5mm
2
10mm
Core-end ferrule, lip above 1.5mm
2
1.5mm
2
2.5mm
2
12mm
Twin ferrule with lip 0.25mm
2
1.5mm
2
12mm
The IP67 enclosure protection for the instrument will only be achieved if each cable is fed into the instrument through a cable gland.
13
4 Electrical connection
4.2 Electrical isolation
3700 V AC
1
Setup interface
Supply
30 V AC 50 V DC
Supply for 2-wire transmitter
3700 V AC
Relay contacts
Binary input
Supply for inductive proximity switch
Supply for ISFET sensor
30 V AC 50 V DC
Main input
Analog output 1
Secondary input (Pt100 / Pt1000)
30 V AC 50 V DC
Analog output 2
1
Not with 12 — 24 VDC supply voltage
4 Electrical connection
14
4.3 General
Opening the instrument
Remove four screws (1) and swing down the top section.
4.4 Connection
4.4.1 pH combination electrode / ORP combination electrode
Fabricating the connecting cable
Strip the cable as shown in the diagram.Insulate the exposed shielding with a shrink-sleeve (1), to prevent short-
circuits.
Apply core ferrules to the ends of the conductors.
Core ferrule dimensions see Chapter 4.1 “Installation notes”, page 12.
(1)
25
55
8
(1)
15
4 Electrical connection
Connecting the cables
The electrical connection for the surface-mountable housing is easily accessible when the instrument is folded out.
The connection cable between sensor and transmitter must be a shielded cable with a diameter of 8 mm max. The instrument contains a guide plate that ensures an optimum cable path. The sensor cables are run to the plug-in screw terminals and must have a strain relief.
The cable clip (3) must only be screwed down (see next page) by a 3.5x6.5 pan head screw! A longer screw may cause a hazardous voltage to be contacted to the shielding!
4 Electrical connection
16
Asymmetrical connection of a combination electrode (standard)
Lead the connecting cables in through the cable glands.Lay the signal cable as shown in the diagram. Use the cable clip (3) to
clamp the signal cable to the shielding.
Connect the cores as assigned below, and see Chapter 4.5 “Terminal
assignments”, page 19.
Push the plug-in terminals for row 1 (1) and row 2 (2) into the sockets in the
instrument.
Sensor connection
(1)
(2)
(3)
(1)
(2)
(3)
(4)
pH combination electrode
SC = shielding (screen)
Row 1
Row 2
In environments with difficult EMC conditions, a coaxial cable with a double shielding must be used. A shielded 2-core cable is required for connecting the temperature probe.
17
4 Electrical connection
Asymmetrical connection of a combination electrode with integrated temperature sensor (SMEK)
Lead the connecting cables in through the cable glands.Lay the signal cable as shown in the diagram. Use the cable clip (3) to
clamp the signal cable to the shielding.
Connect the cores as assigned below, and see Chapter 4.5 “Terminal
assignments”, page 19.
Push the plug-in terminals for row 1 (1) and row 2 (2) into the sockets in the
instrument.
Sensor connection
(1)
(2)
(3)
(1)
(2)
(3)
(4)
?
pH combination electrode
SC = shielding (screen)
Row 1
Row 2
Temperature sensor
4 Electrical connection
18
Symmetrical connection of a combination electrode with separate temperature sensor
Lead the connecting cables in through the cable glands.Lay the signal cables as shown in the diagram. Use the cable clips (3) to
clamp each signal cable to the shielding.
Connect the cores as assigned below, and see Chapter 4.5 “Terminal
assignments”, page 19.
Push the plug-in terminals for row 1 (1) and row 2 (2) into the sockets in the
instrument.
Sensor connection
(1)
(2)
(3)
pH combination electrode
SC = shielding (screen)
Row 1
Row 2
Separate temperature sensor
Ground pin or conductive pipe/
(1)
(2)
(3)
(4)
?
reservoir wall at measurement point
Triaxial cable
! remove plug-in link 3-5 !
In environments with difficult EMC conditions, a coaxial cable with a double shielding must be used. A shielded 2-core cable is required for connecting the temperature probe.
19
4 Electrical connection
4.5 Terminal assignments
Connection Screw
terminals
Row
Supply voltage
Supply voltage (23): 110 — 240 V AC -15/+10%, 48 — 63 Hz
Supply voltage (25): 20—30V AC/DC, 48—63Hz
Supply voltage (30): 12 — 24 V DC +/-15% (permissible only for connection to SELV/PELV circuits)
1 N (L-)
2 L1 (L+)
1
NC 3 NC 7 NC 14 NC 15
Inputs
Glass/metal electrode 1
2
NC 2 Reference electrode 3
NC 4 GND
With asymmetrical connection for pH measurement Link terminals 3 and 5 (Accessory: large plug-in link)
5
FP (liquid potential) With symmetrical connection for pH measurement
6
NC 7 RTD
in 2-wire circuit (Accessory: small plug-in link)
8 9
10
RTD in 3-wire circuit
8 9
10
Binary input 11
12
1
3
ϑ
9
8
10
ϑ
9
8
10
11
12
4 Electrical connection
20
Outputs
Analog output 1 0 — 20 mA resp. 20 — 0 mA or
4 — 20 mA resp. 20 — 4 mA or 0 — 10 V resp. 10 — 0 V
(electrically isolated)
+ 13
- 14
2
Analog output 2 0 — 20 mA resp. 20 — 0 mA or
4 — 20 mA resp. 20 — 4 mA or 0 — 10 V resp. 10 — 0 V
(electrically isolated)
+ 15
- 16
Switching output K1 (floating)
pole 4 break (SPST-NC)5 make (SPST-NO)6
1
Switching output K2 (floating)
pole 8 break (SPST-NC) 9 make (SPST-NO) 10
Connection Screw
terminals
Row
6
4
5
10
8
9
21
4 Electrical connection
4.6 ISFET-pH combination electrode according to data sheet
20.1050
Connection Color Screw
terminals
Row
Cap-
adapter
JUMO AQUIS 500 pH
Supply voltage for ISFET sensor
Supply voltage DC ± 5 V, 5 mA
blue
black
green
11 L+
12
13 L-
1
pH sensor
Sensor Reference
white / black
screen13 + 5 linked
2
RTD in 3-wire circuit
white
red
red / black
10
9 8
Parallel resistance 4.53 kΩ only in conjunction with process connection
615!
red / black
red
8 9
2
ϑ
The orange core of the cap adapter is not connected! The TEMPERATURE INPUT / SENSOR TYPE / CUSTOMIZED parameter must
be configured for process connection 615!
22
5 Mounting
5.1 General
Mounting location
Find a location that ensures easy accessibility for the later calibration. The fastening must be secure and must ensure low vibration for the
instrument. Avoid direct sunlight! Permissible ambient temperature at the installation location: -10 to 55°C with
max. 95% rel. humidity, no condensation.
Installation position
The instrument can be mounted in any position.
5.2 Surface mounting
n
Screw four fixing brackets (1) onto the enclosure.
The fixing brackets can be turned in increments of 90°.
Fasten the enclosure onto the fixing brackets (with screws, plugs, etc.) on a
surface or plate.
Fixing brackets (1) are included with delivery.
30
60
149
134
120
161
140
77
94
94
23
(1)
23
5 Mounting
5.3 Pipe installation set / weather protection roof
The pipe installation set for JUMO AQUIS 500 (sales no.: 20/00483664) can be used to fasten the instrument (and optionally the protective roof for JUMO AQUIS 500, sales no.: 20/00398161) onto pipes or railings with a diameter from 30 to 50 mm.
Screws (1) M5 x 30 for pipe diameters from 30 to 40 mm. Screws (2) M5 x 40 for pipe diameters from 40 to 50 mm. The pipe installation set is also suitable for horizontal pipes.
5.4 DIN rail installation set
The DIN rail installation set for JUMO AQUIS 500 (sales no.:20/00) can be used to attach the instrument to a 35 mm x 7.5 mm DIN rail as defined in DIN EN 60715 A.1.
ø30-50
(2)
(1)
(1)
5 Mounting
24
5.5 Mounting in a panel
Prepare the panel cut-out and holes based on the drill template.Place the control panel (1) in the panel cut-out and fasten it with screws (2)
spacing rollers (3) and nuts (4).
Drilling template See section 12.2 "Panel cut-out", page
101. The panel must be sufficiently thick to achieve the
specified IP65 enclosure protection!
(1)
(2)
(3)
(4)
To ensure electrical safety, the mounting set for panel installation (sales no.: 20/00530470) must be installed, see next page!
25
5 Mounting
The mounting set (sales no.: 20/00530470) consists of parts (1), (2) and (3).
Make the electrical connection.Screw on two stud bolts (1).Fasten the cover (2) with two knurled nuts (3).
Depth behind panel
(1)
(2)
(3)
~55
26
6 Operation
6.1 Controls
(1) Transmitter (3) Control panel
(2) LC display
(2)
(1)
(3)
27
6 Operation
6.2 LC display
6.2.1 Measurement mode (normal display)
(1) (2)
(3)
(4) (5)
Relay K1 is active Relay K2 is active
or AL R1 = alarm, relay K1 AL R2 = alarm, relay K2 ALR12 = alarm, relay K2+K2
Binary input 1 is activated
Keypad is inhibited Device status (indications)
- Alarm (e.g overrange)
- Calib. blinking (calibration timer run down)
- Calib. (customer calibration is active)
(6)
(7) (8)
(9) (10) (11)
Output mode
- Hand (manual operation)
- Hold (Hold operation) Measurement Unit Temperature of medium Operating mode ALR1 = alarm, controller 1
ALR2 = alarm, controller 2 ALR12 = alarm, controllers 1
and 2
(1) (2) (3) (5)(4) (6)
(7)
(8)(9)
(10)
(11)
In order to return to the measurement mode from another display mode: Press the key or wait for the timeout.
EXIT
6 Operation
28
6.3 Principle of operation
Administrator
level
Operator
level
Meas. mode
Calibration
level
EXIT
Calibration
logbook
Device info
PGM
OUTPUT LEVEL
Controller 1
2Controller
MIN/MAX VALUES
> 3 sec
EXIT
or timeout (adjustable)
PGM
< 2 sec
PGM
< 2 sec
EXIT
or timeout ( )adjustable
EXIT
o)r timeout (adjustable
EXIT
o)r timeout (adjustable
EXIT
o)r timeout (adjustable
EXIT
CTRL. SETPOINTS
SETPOINT 1
2SETPOINT
or timeout (adjustable)
EXIT
MANUAL MODE
OVERVIEW
PGM
< 2 s
PGM
< 3 sec
or
29
6 Operation
6.3.1 Operation in levels
Measurement mode (normal display); see Chapter 6.4 “Measurement mode”, page 31
CTRL. SETPOINTS
MIN/MAX values see Chapter 6.5.1 “MIN/ MAX values”, page 31
Output level display see Chapter 6.5.2 “Output level display”, page 32
Manual mode overview see Chapter 6.6 “MANUAL mode / simulation mode”, page 33
OPERATOR LEVEL, see Chapter 6.8 “Operator level”, page 38
INPUT (main variable e. g. pH, ORP, NH
3
) TEMPERATURE INPUT BINARY INPUT CTRL. CHAN. 1 CTRL. CHAN. 2 CTRL.SPEC.FUNCT. SWITCH OUTPUT 1 SWITCH OUTPUT 2 ANALOG OUTPUT 1 ANALOG OUTPUT 2 DISPLAY WASH TIMER
ADMINISTR. LEVEL, see Chapter 6.9 “Administrator level”, page 38
Password
PAR AMETER LEVE L
, see Chapter 6.9.2 “Parameter level”, page 40 INPUT (main variable e. g. pH, ORP, NH3) TEMPERATURE INPUT BINARY INPUT CTRL. CHAN. 1 CTRL. CHAN. 2 CTRL.SPEC.FUNCT. SWITCH OUTPUT 1 SWITCH OUTPUT 2 ANALOG OUTPUT 1 ANALOG OUTPUT 2 DISPLAY
ENABLE LEVEL
, see Chapter 6.9.3 “Enable level”, page 40 CTRL. CHAN. 1 CTRL. CHAN. 2 CTRL.SPEC.FUNCT. CONDUCTIVITY IN TEMPERATURE INPUT BINARY INPUT SWITCH OUTPUT 1 SWITCH OUTPUT 2 ANALOG OUTPUT 1 ANALOG OUTPUT 2 DISPLAY WASH TIMER
6 Operation
30
Measurement mode ADMINISTRATOR LEVEL
BASIC SETTINGS,
see Chapter 6.9.4 “Basic settings”, page 42
SENSOR MONIT. REF. MONIT. GLASS EL. RE-INITIALIZE DEVICE
CALIB. LEVEL
, see Chapter 6.9.5 “Calibration level”, page 43 1-POINT CALIB. 2-POINT CALIB. 3-POINT CALIB.
CALIB. ENABLE
ENABLE 1-POINT CALIB. ENABLE 2-POINT CALIB. ENABLE 3-POINT CALIB.
DELETE LOGBOOK
REALLY DELETE LOGBOOK?
CALIB. LEVEL
1-POINT CALIB. 2-POINT CALIB. 3-POINT CALIB.
CALIB. LOGBOOK
DEVICE INFO
SENSOR MONIT. REF. MONIT. GLASS EL.
Loading...
+ 74 hidden pages