JUMO AQUIS 500 CR User Manual

JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 Email: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road East Syracuse, NY 13057, USA Phone: 315-437-5866
1-800-554-5866 Fax: 315-437-5860 Email: info.us@jumo.net Internet: www.jumousa.com
JUMO AQUIS 500 CR
Transmitter/controller
2 analog inputs
Input 1: Conductivity / resistivity
1 binary input
For floating contact Functions:
- key inhibit
- alarm stop
- HOLD
Supply voltage
110—240 V AC 12—24 V DC 20—30 V AC/DC
Analog outputs (option)
Output 1 + 2: 0(4)—20 mA or 0—10 V
option
Setup interface
Switching outputs (option)
Outputs 3 + 4:
- relay, changeover (SPDT)
User-friendly configuration Reloading of languages Plant documentation
Configurable as analog process value output and/or continuous controller output (PID action)
Configurable as
- limit controller
- pulse width output (PID action)
- pulse frequency output (PID action)
- modulating controller (PID action)
- USP contact
- "Purified water" contact as per Ph.Eur.
- calibration timer run down
standard
Input 2: Temperature manual input or automatic Pt100 / Pt1000 / 4 kΩ
Transmitter/Controller for conductivity, TDS, resistivity and temperature
Brief description
The instrument is used for the conductive measurement/control of electrolytic conductivity, resistivity or the TDS value. In addition, the JUMO AQUIS 500 CR also offers the possibility of showing the measured conductivity according to a customer-specific table. Conductive two-electrode cells as well as four-electrode cells can be connected to the instrument. Temperature serves as the second input variable, measured by a Pt100/1000 probe. Depending on the measured variable, it is therefore possible to implement specific, automatic temperature compensation. The instrument is operated using keys and a large LC graphics display on which the measurements are clearly legible. The plain-text presentation of the parameters makes it easier for the user to configure the instrument, and also helps in programming it correctly. Thanks to its modular design, the instrument can be perfectly matched to the particular application requirement. Up to four outputs are available (see the block diagram for the functions).
Typical areas of application
Universally applicable in water and wastewater engineering, service/process water and wastewater, drinking water and well/surface water, pure and high-purity water as well as for pharmaceutical water (e.g. as per USP, Ph.Eur., WFI), water quality measurements, TDS measurements (ppm or mg/l).
Block diagram
Approvals/approval marks (see Technical data)
V1.00/EN/00477047
Data Sheet 202565
h
t
i
w
t
n
a
>
i
l
5
4
p
6
m
<
o
P
C
S
U
Key features
• Direct changeover to
- conductivity (µS/cm or mS/cm)
- resistivity (kΩ x cm or MΩ x cm)
- TDS measurement (ppm or mg/l)
- customer-specific table
• Automatic temperature compensation: off (e.g. USP), linear, ASTM, natural water (EN 27888/ISO 7888)
• Large LC graphics display with background lighting
• Choice of display: large numbers, bar graph or trend display
• Calibration options according to measured variable: cell constant and temperature coefficient
• Calibration logbook
• Two-electrode cells (as standard) or four-electrode cells can be connected
• Pollution detection can be activated
• Auto-range operation
• IP67 enclosure protection (in surface mountable housing) IP65 enclosure protection (for panel mounting)
• Language changeover: German, English, French; further languages can be loaded through the setup program
• Using the setup program: user-friendly programming, plant documentation, additional languages can be loaded
Page 1/12
Type 202565
Page 2/12Data Sheet 202565
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 Email: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road East Syracuse, NY 13057, USA Phone: 315-437-5866
1-800-554-5866 Fax: 315-437-5860 Email: info.us@jumo.net Internet: www.jumousa.com
(1)
(2)
(4)
(5)
(6)
(7)
(8)
(9)
(3)
Functional description
The instrument is designed for use on site. A rugged housing protects the electronics and the electrical connections from corrosive environmental conditions (IP67). As an alternative, the instrument can also be installed in a control panel, and is then protected to IP65 on the front. The electrical connection is made by easy-to-fit pluggable screw terminals.
Transmitter
Two-electrode cells (standard) as well as four­electrode cells can be used for measurement.
Two-electrode cells can be connected, in the usual increments for cell constants (K=0.01;
0.1; 1.0; 3.0 and 10.0). Thanks to the widely adjustable relative cell constant, it is also possible to connect sensors with different cell constants (e.g. K=0.2).
In the case of the 4-electrode cells, the values K=0.5 and 1.0 have been predefined for the cell constant. Here too, the instrument can be matched to sensors with different cell constants (e.g. K=0.4).
The instrument can perform automatic temperature compensation, by acquiring the temperature of the sample solution.
Displays and controls
(1) Switching output 1 or 2 is active (2) Binary input 1 is actuated (3) Keypad is inhibited (4) Alarm has been activated (5) Instrument is in manual mode (6) Instrument status (7) Temperature of medium (8) Principal measurement (9) Unit of principal measurement
The user can define what is to be shown in positions (7) and (8) of the display:
•no display
• compensated or uncompensated measurement
•temperature
• output level 1 or 2
•setpoint 1 or 2
V1.00/EN/00477047
Operation
For easy programming and operation, all parameters are arranged in clearly structured levels and shown in plain text. Operation is protected by a code word. This facilitates individual adaptation of the operation, since parameters can be generally enabled or specifically assigned to the protected area.
As an alternative to configuration from the keys, the instrument can also be configured through the convenient setup program for PC (option).
Display modes
Three display modes are available:
Large numbers
In this display mode, the measurements are shown in digits, as usual.
Tre n d display
The numerical value is supplemented by a symbol which indicates the change direction and change speed of the measurement. This can, for instance, be very useful during controller optimization.
from left to right: fast, medium and slow rise, stable,
slow, medium and fast fall.
Bar graph
This display mode allows the user to see at a glance in which region the measurement is at present.
The bar graph can be freely scaled.
Function modes
Electrolytic conductivity
Display/control, unit µS/cm or mS/cm.
Resistivity (high-purity water)
Display/control, unit kΩ x cm or MΩ x cm.
TDS
Display/ control with ppm for the unit. In this mode, the specific TDS factor can be entered in addition.
Customer-specific table
In this mode, the input value (conductivity or resistivity) can be displayed in accordance with a table (up to 20 value pairs). Thanks to this function, it is possible to implement simple concentration measurements, for example. The values in the table can only be entered through the optional setup program.
Calibration
Cell constant
Because of manufacturing tolerances, the cell constant of a conductivity cell may deviate slightly from its nominal value. In addition, the cell constant may change during operation (due to deposits or wear, for example). This results in a change of the output signal from the cell. The instrument provides the user with the possibility of compensating any deviation from the nominal value of the cell constant by manual entry or automatic calibration of the relative cell constant. A manual entry is used, for instance, for calibration during high-purity water measurement.
Temperature coefficient
The conductivity of almost all solutions depends on the temperature. To ensure correct measurement, it is therefore necessary to know both the temperature and the temperature coefficient [%/°C] of the sample solution. The temperature can either be measured automatically, with a Pt100 or Pt1000 temperature probe, or it has to be set manually by the user. The temperature coefficient can be automatically determined by the instrument, or it can be entered manually.
Calibration logbook
The five most recent successful calibrations can be called up in the calibration logbook. This makes it possible to evaluate the ageing of the sensor that is connected.
Calibration timer
The calibration timer indicates (if required) when the next routine calibration is due. The calibration timer is activated by entering a number of days, after which recalibration has to be carried out (plant or operator requirement).
MIN / MAX value memory
This memory acquires the minimum or maximum input variables that have occurred. This information serves, for example, to decide whether the sensor that is connected is suited to the values that are actually present.
Page 3/12Data Sheet 202565
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 Email: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road East Syracuse, NY 13057, USA Phone: 315-437-5866
1-800-554-5866 Fax: 315-437-5860 Email: info.us@jumo.net Internet: www.jumousa.com
Hysteresis
Setpoint
pH
On
Off
Limit
Hysteresis
Limit
pH
On
Off
Setpoint
Hysteresis
Setpoint
pH
On
Off
Limit
Distance
Hysteresis
Setpoint
pH
On
Off
Limit
Distance
Time
On
Off
Pulse duration
Pulse contact
Time
On
Off
Trigger condition
Detection of deposits
Deposit detection can be activated for 4­electrode cells. During normal operation, it can happen that deposits form on electrodes. This has the result that a lower concentration is displayed than actually present. With activated “deposit detection” function, the instrument tells you when the cell needs to be serviced.
Auto-range
For some processes, the availability of two measuring ranges is advantageous, for instance for rinsing or regeneration processes. What is usually required here, is the precise acquisition of a low conductivity. Rinsing or regeneration, however, involves a much higher conductivity, which could lead to an out-of-range condition (error). This situation is not just unsatisfactory, but may even be dangerous. Thanks to the auto-range function, two measuring ranges can be determined. The instrument then switches between them in a defined manner.
Binary input
The following functions can be activated through the binary input:
• Activate key inhibit When this function has been activated, operation from the keys is no longer possible.
• Activate HOLD mode After activating this function, the outputs (analog and relay) adopt the states that have previously been defined.
• Alarm suppression (controller alarm only) This function temporarily deactivates the alarm generation via the relay (has to be configured accordingly).
Linking the corresponding terminals by means of a floating contact (e. g. relay) will activate the pre-defined function.
Control functions
The relays can have functions assigned that are configured via parameters. The control function is freely programmable as P, PI, PD or PID action.
Relay outputs
Two relay changeover contacts are available for the principle measurement variable and/or temperature. The following functions can be programmed:
• Switching direction (min/max)
• Limit controller (pull-in/drop-out delay, hysteresis)
• Pulse width output (see control functions)
• Pulse frequency output (see control functions)
• Modulating controller function (see control functions)
• Limit comparators (pull-in/drop-out delay, hysteresis)
• Pulse function The output switches on briefly when reaching the switching point and then off again.
•Alarm
• Sensor or range error
• Response to alarm, over/underrange, calibration and HOLD
Contact functions
MAX limit comparator
MIN limit comparator
Alarm window 1
Alarm window 2
Pulse contact Trigger condition longer than pulse duration
Analog Outputs
There are up to 2 analog outputs available. The following functions can be selected:
Output Analog process value output Continuous controller
Main variable Temperature 1X-X 2-XX
With the analog process value output, the range start and end values are freely selectable. The response of the outputs to over/underrange, alarm and calibration is freely programmable. Simulation function: The analog process value outputs can be freely set in the manual ("Hand")
Application: "Dry run" start-up of the plant, troubleshooting, servicing.
V1.00/EN/00477047
mode.
main value
Page 4/12Data Sheet 202565
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 Email: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road East Syracuse, NY 13057, USA Phone: 315-437-5866
1-800-554-5866 Fax: 315-437-5860 Email: info.us@jumo.net Internet: www.jumousa.com
Time
On
Off
Pulse duration
Pulse contact
Time
On
Off
Trigger condition
Process value X
Setpoint W
Proportional band X
P
100%
50%
0%
Output level y
Switching period
10%
90%
90%
10%
t
On
50%
50%
t
Off
10%
90%
X
P
X - W
0
1
100%
50%
0%
No pulses
50% of pulse frequency
Maximum pulse frequency
Setpoint W
Proportional band X
P
X
P
X - W
0
1
Output level y
Process value X
Pulse contact Trigger condition shorter than pulse
Pulse width controller (output is active with X > W and P action)
duration
If the process value X exceeds the setpoint W, the P controller will control proportionally to the control deviation. On going outside the proportional band, the controller operates with an output level of 100 % (100 % duty cycle).
Measuring ranges / cell constants
This state-of-the art instrument offers a far wider dynamic range on the input side than can be managed physically or chemically by the conductivity cells. For this reason, the range must be matched to the operating range of the cell.
Examples of ranges for combination with 2-electrode cells
Cell constant (K) Recommended/practical measuring span
0.01 1/cm 0.05 µS/cm to 20 µs/cm
0.1 1/cm 1 µS/cm to 1000 µs/cm
1.0 1/cm 0.01 mS/cm to 100 ms/cm
3.01/cm 0.1mS/cmto30ms/cm
10.0 1/cm 0.1 mS/cm to 200 ms/cm
(depending on the conductivity cell)
Pulse frequency controller (output is active with X > W and P action)
If the process value X exceeds the setpoint W, the P controller will control proportionally to the control deviation. On going outside the proportional band, the controller operates with an output level of 100 % (maximum switching frequency).
It is assumed that the measuring span of the instrument is always larger than the recommended or practically usable range of the conductivity cell. The smaller range (instrument or conductivity cell) determines the maximum range that can be used.
Example
Which span can the instrument cover with a predefined cell constant?
The predefined cell constant is K=0.4 The span of the instrument is
0.1 µS/cm x 0.4 1/cm to 1250 mS/cm x 0.4 1/cm 0.04 µS/cm — 500 mS/cm
Example
A measurement is to be carried out in the 10 µS/cm to 500 µS/cm range. A conductivity cell with the cell constant K = 0.1 1/cm is chosen. The unit µS/cm without a decimal place is configured on the instrument.
Combination with 4-electrode cells and 2-electrode cells having cell constants that deviate from the above graduation
This requires taking a closer look at the instrument technology and considering both the uncompensated and the temperature-compensated measuring span.
The uncompensated measuring span of the instrument is calculated according to the formula: Measuring span = 0.1 µs/cm x cell constant (K) to 2500 mS x cell constant (K).
After taking account of the temperature compensation range, the following compensated measuring span (approx.) will remain: Measuring span = 0.1 µs/cm x cell constant (K) to 1250 mS x cell constant (K).
Cell constant (K) Measuring span covered by instrument
0.01 0.001 µS/cm to 1.25 ms/cm
0.1 0.01 µS/cm to 12.5 ms/cm
1.0 0.1 µS/cm to 125 ms/cm
3.0 0.3 µS/cm to 375 ms/cm
10.0 0.1 mS/cm to 1250 ms/cm
V1.00/EN/00477047
(temperature-compensated)
Loading...
+ 8 hidden pages