This manual contains information that must be observed in the interest of your own safety and
to avoid damage to assets. This information is supported by symbols which are used in this
manual as follows.
Please read this manual before commissioning the device. Keep the manual in a place accessible to all users at all times.
If difficulties occur during commissioning, please refrain from carrying out any manipulations
that could jeopardize your warranty rights.
Warning signs
DANGER!
This symbol indicates that injury to persons caused by electrical shock may occur, if the
respective protective measures are not carried out.
WARNING!
This symbol in connection with the signal word indicates that injury of persons may occur,
if the respective protective measures are not carried out.
1 Introduction
Note signs
CAUTION!
This symbol in combination with the signal word indicates that damage to assets or data
loss will occur if suitable precautions are not taken.
CAUTION!
This sign indicates that components could be destroyed by electrostatic discharge
(ESD = Electro Static Discharge), if the respective cautionary measures are not taken.
Only use the ESD packages intended for this purpose to return device inserts, assembly
groups or assembly components.
TIP!
This symbol refers to important information about the product or its handling or additional
use.
REFERENCE!
This symbol refers to further information in other sections, chapters or manuals.
DISPOSAL!
This device and, if installed, the batteries must not be disposed of in the garbage can after
use! Please ensure that they are disposed of properly and in an environmentally friendly manner.
7
1 Introduction
1.2General information
(1)Locking screw for operation
•The device measures the temperature in liquids and gases.
•The temperature is displayed digital.
•Depending on the version, the following outputs are available:
- 1x PNP switching output
- 2x PNP switching outputs
- 1x PNP switching output and 1x analog output 0 to 20 mA, configurable
- 1x PNP switching output and 1x analog output 4 to 20 mA, configurable
- 1x PNP switching output and 1x analog output 0 to 10 V, configurable
•The device is also available in a version for use in increased medium temperatures.
•The device is additionally available in a version equipped with M 12x1 connection for RTD
temperature probes.
•The device can be set directly on site or configured via the setup program with a notebook/
PC.
CAUTION!
The protection class specified for the device is only achieved when the locking screw is inserted and tightened.
8
1.3Description
The highly precise, electronic temperature switch comprises a sheath with built-in temperature
probe, a process connection as well as an attached case with LCD display for the electronics.
Depending on the application, the following output variants are available: 1x PNP or 2x PNP
switching output (binary output) or 1x PNP switching output and 1x analog output.
The configuration of the output signal and the measuring range can be customized.
Depending on the version, the electronic temperature switch can be used in an operating tem-
perature range from -50 to +150 °C, -50 to +260 °C and -50 to +500 °C. The analog output signal 4 to 20 mA, 0 to 20 mA, 0 to 10 V or reversed 20 to 4 mA, 20 to 0 mA and 10 to 0 V is
available linearized (temperature linear).
The electronic temperature probe is designed for industrial applications and complies with the
European Standards to guarantee electromagnetic compatibility (EMC).
1 Introduction
9
1 Introduction
1.4Block diagram
10
2.1Type specifications
Position
A laser has been used to label the type specifications on the device display case.
Contents
The specifications contain important information. Amongst others:
DescriptionExample
Basic type902940/10
Voltage supplyDC 14 to 30 V
Output signal4to20mA
Measuring range (MB)-50 to +150 °C
Switching output1x PNP 250 mA
Fabrication No. (F No.)
The serial number is used by the manufacturer to identify the device.
The F-No. is not located on the device but on the package.
The F-No. contains the production date (year and week) on position 12 to 15 (counted from the
left).
Example: F-No. = 0149678701010440
The device was produced in the 44th calendar week in 2010.
2 Device identification
11
2 Device identification
2.2Type description for basic type 902940/10
(1) Basic type
902940/10DELOS T
Electronic temperature switch
with display and analog output
Ambient temperature on the display case: -20 to +75 °C
Parts coming into contact with the medium are electropolished,
roughness depth Ra ≤ 0.8 μm
(2) Version
x8Standard, with factory settings
x9Customer-specific configuration (specifications in plain text)
(3) Operating temperature in °C
x370-50 to +150 °C
(4) Measuring insert
x10131x Pt1000 in 4-wire circuit
(5) Tolerance class as per DIN EN 60751
x2Class A (standard)
x3Class AA (1/3 DIN B)
(6) Output
x4701x PNP switching output
x4712x PNP switching output
x4751x PNP switching output and 1x analog output 4 to 20 mA, configurable
x4761x PNP switching output and 1x analog output 0 to 20 mA, configurable
x4771x PNP switching output and 1x analog output 0 to 10 V, configurable
(7) Sheath diameter D in mm
x6∅ 6mm
(8) Fitting length EL in mm (EL 50 to 1000 mm)
x5050mm
x100100 mm
x150150 mm
x...Please specify in plain text (50 mm steps)
(9) Process connection PA
x000None (plain sheath made of stainless steel 316L)
x102Screw connection G 1/4 (stainless steel 316Ti)
x103Screw connection G 3/8 (stainless steel 316Ti)
x104Screw connection G 1/2 (stainless steel 316Ti)
x118Screw connection M 12x1.5 (stainless steel 316Ti)
x126Screw connection M 18x1.5 (stainless steel 316Ti)
x128Screw connection M 20x1.5 (stainless steel 316Ti)
x997JUMO PEKA with EHEDG certificate (stainless steel 316L)
(compatible process connection adapter, see data sheet 40.9711)
(10) Extra codes
x000None
x310Sheath ∅ 6 mm stepped down to ∅ 3.8 mm (fitting length EL max. 800 mm)
x810Welding socket (only for process connection 380)
(1)(2)(3)(4)(5)(6)(7)(8)(9)(10)
Order code--------/ , ...
Order example902940/10-8-370- 1013 -2-475-6-50-000/000
1
State extra codes one after another, separated by commas.
1
13
2 Device identification
2.2.1Scope of delivery for basic type 902940/10
Our scope of delivery includes:
•1 device in the version ordered
•1 combination tool
(required for operation on the device and to turn the display case through ± 160°)
•1 operating instructions B 90.2940.0
In the event of questions, please contact your supplier.
2.2.2Accessories, process connection for basic type 902940/10
The following articles must be ordered separately and with costs:
ArticlePart No.
Pipe screw connection G 1/4 (stainless steel 316Ti) for sheath screw connections90/00080811
Pipe screw connection G 1/2 (stainless steel 316Ti) for sheath screw connections90/00305445
2.2.3Accessories for basic type 902940/10
The following articles must be ordered separately and with costs:
ArticlePart No.
Setup program on CD-ROM, multilingual90/00550018
PC interface with USB/TTL converter and adapter (USB connection line)
(required for the configuration with the setup program)
Connection line (Y cable)
(required for the configuration with the setup program)
Combination tool
(required for operation on the device and after installation
to turn the display case through ± 160°to the LH or RH side)
Cable box, 4-pin (straight) M 12x1 with PVC connection line length 2000 mm
(can be used for individual assembling)
Cable box, 4-pin (angled) M 12x1 with PVC connection line length 2000 mm
(can be used for individual assembling)
70/00456352
40/00507861
40/00526614
40/00404585
40/00409334
14
2 Device identification
2.3Type description for basic type 902940/30
(1) Basic type
902940/30DELOS T
Electronic temperature switch
with display and analog output
with neck pipe for higher medium temperatures
Ambient temperature on the display case: -20 to +75 °C
Parts coming into contact with the medium are electropolished,
roughness depth Ra ≤ 0.8 μm
(2) Version
x8Standard, with factory settings
x9Customer-specific configuration (specifications in plain text)
(3) Operating temperature in °C
x386-50 to +260 °C
(4) Measuring insert
x10131x Pt1000 in 4-wire circuit
(5) Tolerance class as per DIN EN 60751
x2Class A (standard)
x3Class AA (1/3 DIN B)
(6) Output
x4701x PNP switching output
x4712x PNP switching output
x4751x PNP switching output and 1x analog output 4 to 20 mA, configurable
x4761x PNP switching output and 1x analog output 0 to 20 mA, configurable
x4771x PNP switching output and 1x analog output 0 to 10 V, configurable
(7) Sheath diameter D in mm
x6∅ 6mm
(8) Fitting length EL in mm (EL 50 to 1000 mm)
x5050mm
x100100 mm
x...Please specify in plain text (50 mm steps)
(9) Process connection PA
x000None (plain sheath made of stainless steel 316L)
x102Screw connection G 1/4 (stainless steel 316Ti)
x103Screw connection G 3/8 (stainless steel 316Ti)
x104Screw connection G 1/2 (stainless steel 316Ti)
x118Screw connection M 12x1.5 (stainless steel 316Ti)
x126Screw connection M 18x1.5 (stainless steel 316Ti)
x128Screw connection M 20x1.5 (stainless steel 316Ti)
x997JUMO PEKA with EHEDG certificate (stainless steel 316L)
(compatible process connection adapter, see data sheet 40.9711)
(10) Extra codes
x000None
x310Sheath ∅ 6 mm stepped down to ∅ 3.8 mm (fitting length EL max. 700 mm)
x810Welding socket (only for process connection 380)
(1)(2)(3)(4)(5)(6)(7)(8)(9)(10)
Order code--------/ , ...
Order example902940/30-8-386- 1013 -2-475-6-50-102/000
1
State extra codes one after another, separated by commas.
1
16
2 Device identification
2.3.1Scope of delivery for basic type 902940/30
Our scope of delivery includes:
•1 device in the version ordered
•1 combination tool
(required for operation on the device and to turn the display case through ± 160°)
•1 operating instructions B 90.2940.0
In the event of questions, please contact your supplier.
2.3.2Accessories, process connection for basic type 902940/30
The following articles must be ordered separately and with costs:
ArticlePart No.
Pipe screw connection G 1/4 (stainless steel 316Ti) for sheath screw connections90/00080811
Pipe screw connection G 1/2 (stainless steel 316Ti) for sheath screw connections90/00305445
2.3.3Accessories for basic type 902940/30
The following articles must be ordered separately and with costs:
ArticlePart No.
Setup program on CD-ROM, multilingual90/00550018
PC interface with USB/TTL converter and adapter (USB connection line)
(required for the configuration with the setup program)
Connection line (Y cable)
(required for the configuration with the setup program)
Combination tool
(required for operation on the device and after installation
to turn the display case through ± 160° to the LH or RH side)
Cable box, 4-pin (straight) M 12x1 with PVC connection line length 2000 mm
(can be used for individual assembling)
Cable box, 4-pin (angled) M 12x1 with PVC connection line length 2000 mm
(can be used for individual assembling)
70/00456352
40/00507861
40/00526614
40/00404585
40/00409334
17
2 Device identification
2.4Type description for basic type 902940/50
(1) Basic type
902940/50DELOS T
Electronic temperature switch
with display and analog output
and M 12x1 connection for RTD temperature probes
Ambient temperature on the display case: -20 to +75 °C
(2) Version
x8Standard, with factory settings
x9Customer-specific configuration (specifications in plain text)
(3) Operating temperature in °C
x408-50 to +500 °C
(4) Measuring input
x10131x Pt1000 in 4-wire circuit
(5) Tolerance class as per DIN EN 60751
x0None (not relevant)
(6) Output
x4701x PNP switching output
x4712x PNP switching output
x4751x PNP switching output and 1x analog output 4 to 20 mA, configurable
x4761x PNP switching output and 1x analog output 0 to 20 mA, configurable
x4771x PNP switching output and 1x analog output 0 to 10 V, configurable
(7) Sheath diameter D in mm
x0None (not relevant)
(8) Fitting length EL in mm
x000None (not relevant)
(9) Process connection PA
x000None (not relevant)
(10) Extra codes
x000None
(1)(2)(3)(4)(5)(6)(7)(8)(9)(10)
Order code--------/
Order example902940/50-8-408- 1013-0-475-0-000-000/000
18
2 Device identification
2.4.1Scope of delivery for basic type 902940/50
Our scope of delivery includes:
•1 device in the version ordered
•1 combination tool
(required for operation on the device and to turn the display case through ± 160°)
•1 operating instructions B 90.2940.0
In the event of questions, please contact your supplier.
2.4.2Accessories for basic type 902940/50
The following articles must be ordered separately and with costs:
ArticlePart No.
Setup program on CD-ROM, multilingual90/00550018
PC interface with USB/TTL converter and adapter (USB connection line)
(required for the configuration with the setup program)
Connection line (Y cable)
(required for the configuration with the setup program)
Combination tool
(required for operation on the device and after installation
to turn the display case through ± 160° to the LH or RH side)
Cable box, 4-pin (straight) M 12x1 with PVC connection line length 2000 mm
(can be used for individual assembling)
Cable box, 4-pin (angled) M 12x1 with PVC connection line length 2000 mm
(can be used for individual assembling)
Machine connector M 12x1, 4-pin as per IEC 60947-5-290/00404727
Wall holder for DELOS T with M 12x1 connector90/00555129
Push-in RTD temperature probe with Pt1000 temperature probe
and machine connector M 12x1
902150/99-386-1006-2-6-100-56-2500/315
Sheath diameter 6 mm, fitting length 100 mm,
connection cable length 2500 mm
Push-in RTD temperature probe with Pt1000 temperature probe
and machine connector M 12x1
902150/99-386-1006-2-6-200-56-2500/315
Sheath diameter 6 mm, fitting length 200 mm,
connection cable length 2500 mm
70/00456352
40/00507861
40/00526614
40/00404585
40/00409334
90/00551310
90/00551311
19
2 Device identification
20
3 Electrical connection
3.1Installation notes
CAUTION!
The electrical connection must only be carried out by qualified personnel!
•Ensure that the fuses of the load circuits are suitable for the maximum loads to avoid damage to the device.
•The electromagnetic compatibility meets standard EN 61326.
•Do not connect any additional consumers to the voltage supply of the device.
•The device is not suitable for use in areas with an explosion hazard (Ex areas).
•In addition to a faulty installation, also incorrectly set values on the device could impair the
orderly function of the following process or lead to damage. For this reason, always provide
safety devices/guards independent of the device, the setting of which is restricted to expert
personnel.
3.2Device connection assignment for M 12x1 machine connector
The connection is located on the device rear!
TIP!
The assignment only applies to A-coded standard cables.
1brown (bn)2white (wh)
3blue (bu)4black (bk)
5gray (gy)
21
3 Electrical connection
1
2
3
4
5
L-
K1
L+
3.3Connection assignment for order code 470
The connection is located on the device rear!
1x PNP switching output
Voltage supply
1 L+DC 14 to 30 V
3 L-GND
Output
4 K1Highside open collector
max. 0.25 A
2nc
5Interface
22
3 Electrical connection
1
2
3
4
5
L-
K1K2
L+
3.4Connection assignment for order code 471
The connection is located on the device rear!
2× PNP switching output
Voltage supply
1 L+DC 14 to 30 V
3 L-GND
Output
4 K1Highside open collector
2 K2
5Interface
max. 0.25 A
23
3 Electrical connection
1
2
3
4
5
L-
K1
L+
3.5Connection assignment for order code 475, 476 and 477
The connection is located on the device rear!
1x PNP switching output and 1x analog output
Voltage supply
1 L+DC 14 to 30 V
3 L-GND
Output
4 K1Highside open collector
max. 0.25 A
2 analog0(4)to20mA/0to10V
5Interface
24
3 Electrical connection
12 34
3.6Connection assignment for RTD temperature probe for basic
type 902940/50
TIP!
Please ensure that this connection is not mixed up with the connection on the device rear
(only for analog or switching output, voltage supply and interface)!
Machine connector M 12x1, 4-pole
as per IEC 60947-5-2
Top view to the machine connector on the corresponding RTD temperature probe!
Top view to the installed socket at the device bottom!
Connection assignment for RTD temperature
probe in 4-wire circuit (input)
25
3 Electrical connection
26
4.1General information
CAUTION!
Check that the device is compatible with the medium to be measured.
Installation site
•Ensure that the device is easily accessible for the later operation.
•Ensure that the device is fastened safely and with low vibrations.
•Avoid direct sun radiation!
•Ensure an admissible ambient temperature at the installation site.
Installation position
The device can be installed in any position.
We recommend the "vertical" installation position.
4 Installation
27
4 Installation
4.1.1Turning the LCD display
The device LCD display can be turned through 180°. (Either via the setting on the device or the
optional setup program).
This will facilitate reading when, for example, the device is installed turned through 180°.
Chapter 6.17 "Setting the display alignment (D.Dir)", page 63
4.1.2Turning the display case
The display case of the device can be turned to the LH or RH side through ±160° using the
supplied combination tool (1).
28
(1)Combination tool
4.2Device dimensions
4 Installation
Figure 1:Basic type 902940/10 with process connection (LH)
Basic type 902940/30 with neck pipe and process connection (center)
Basic type 902940/50 with M 12x1 connection for RTD temperature probe (RH)
The 997 JUMO PEKA process connection adapter is EHEDG certified.
For detail information on this process connection adapter, refer to data and price sheet
40.9711.
32
5.1Operating element
5 Operation
(1)Locking screw(2)Operating element
How to proceed:
StepActivity
1Remove the locking screw (1) using a suitable screwdriver.
2Use the supplied combination tool
(or a 0.5 x 3 mm screwdriver or Allen key, width across flats 2)
to "Turn/press" the operating element (2).
33
5 Operation
5.2LCD display
5.2.1Measuring mode (normal display)
Example:
The LCD display is lit yellow.
5.2.2Setting mode
Operation
Example:
The LCD display is lit red.
ContinuePress the combination tool for less than 1 second (< 1s)
Yes (take-over)Press the combination tool for less than 1 second (< 1s)
No (cancel)Press the combination tool for more than 3 seconds (> 3s)
Time outno action for more than 60 seconds (> 60 s)
TIP!
To change over from the setting mode to the measuring mode:
- "No" (Cancel) or
- Wait for "Timeout" = No actions for 60 seconds.
34
5.3Levels
5 Operation
35
5 Operation
5.4Parameters
5.4.1Input
ParametersDisplaySetting range
Temperature unit°C
°F
Decimal place
temperature values
Offset
(zero point correction)
Damping
(filter time constant)
0
1
-100.0 to 0 to +100.0 °C
0.00 to 0.10 to 99.99 s
1
Mains frequency (Hz)50 H
60 H
1
The standard setting is displayed bold.
36
5.4.2Analog output
5 Operation
ParametersDisplaySetting range
Type of signal
(only with analog output)
Scaling start
(only with analog output)
Scaling end
(only with analog output)
Error signal
(only with analog output)
4to20mA
0to20mA
0to10V
-50 °C
150 °C/260 °C/500 °C
Note:
Depending on the device ordered.
3.4 mA or 22 mA
with output signal 4 ... 20 mA
0mA or 22mA
with output signal 0 ... 20 mA
0 V or 10.7 V
with output signal 0 ... 10 V
Chapter 6.10 "Setting the error signal of the
analog output (S.Err)", page 53
Note:
Depending on the configured output signal.
1
The standard setting is displayed bold.
1
37
5 Operation
5.4.3Binary output 1
ParametersDisplaySetting range
Switching function = Hysteresis, N/O contact
1
= Hysteresis, N/C contact
= Window, N/O contact
= Window, N/C contact
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
Switching point100 °C
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
Release point90 °C
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
Hysteresis
(only with configured
switching point and
release point)
Switch-on delay0.00 to 99.99 s
0to1 to 500 °C
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
Note:
Application exclusively with window switching
functions.
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
38
1
The standard setting is displayed bold.
5.4.4Binary output 2
5 Operation
ParametersDisplaySetting range
Switching function
(only with second switching output)
= Hysteresis, N/O contact
= Hysteresis, N/C contact
= Window, N/O contact
= Window, N/C contact
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
Switching point
(only with second switching output)
Release point
(only with second switching output)
Hysteresis
(only with second switching output and configured
switching point and
release point)
Switch-on delay
(only with second switching output)
100 °C
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
90 °C
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
0to1 to 500 °C
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
Note:
Application exclusively with window switching
functions.
0.00 to 99.99 s
Chapter 6.12 "Setting the switching function
(B.Fct)", page 55
1
1
The standard setting is displayed bold.
39
5 Operation
5.4.5Display and operation
ParametersDisplaySetting range
Display alignmentStd = standard (for standard operation)
turn = turned through 180° (for operation turned
through 180°)
1
Chapter 6.17 "Setting the display alignment
(D.Dir)", page 63
Unit of the actual value
display
(only with analog output)
Software version of the
operating unit
(cannot be edited)
Software version of the
signal unit
(cannot be edited)
Uni.T = Temperature unit (Uni.T)
Proz = % from the scaled range (Sc.Lo and Sc.Hi)
Chapter 6.18 "Setting the display unit (D.Uni)",
page 64
Display of the software version of the operating
unit
Chapter 6.19 "Displaying the software version
of the operating unit (SW.Di)", page 65
Display of the software version of the signal unit
Chapter 6.20 "Displaying the software version
of the signal unit (SW.Si)", page 66
Code
(can only be programmed
via the setup program)
1
The standard setting is displayed bold.
0000 to 0072 to 9999
Chapter 6.2 "Unlocking the device (enter the
code).", page 42
40
6.1Quick introduction
TIP!
This is a proposal to be able to configure the device reliably within a short time. When checking the setting possibilities specified in this list prior to starting configuration, timeouts can be
avoided during configuration.
How to proceed:
StepActivity
1Install the device.
Chapter 4 "Installation", page 27
2Connect the device.
Chapter 3 "Electrical connection", page 21
3Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
6 Commissioning/start-up
4Select the unit of the measured value.
Chapter 6.4 "Setting the unit of the measured value (Uni.T)", page 44
5Set the output signal.
Chapter 6.8 "Setting the output signal (S.Type)", page 47
6Set the output signal scaling.
Chapter 6.9 "Setting the output signal scaling", page 48
7Set the switching function.
Chapter 6.12 "Setting the switching function (B.Fct)", page 55
8Set the switching point.
Chapter 6.13 "Setting the switching point (B.Sp)", page 59
9Set the release point.
Chapter 6.14 "Setting the release point (B.RSp)", page 60
41
6 Commissioning/start-up
6.2Unlocking the device (enter the code).
The device is protected against unauthorized operation by a code.
Code is set to 0072 (factory setting). Changes are only possible with the setup program. The
device will become unprotected when the setup program is used to set the code to 0000.
Unlocking
42
(1)Locking screw(2)Combination tool
6 Commissioning/start-up
How to proceed:
StepActivity
1Remove the locking screw (1) using a suitable screwdriver.
2Briefly press the combination tool (2) until the third "0" (counted from the left) flashes.
Display previously changes to "red".
3Turn the combination tool until "7" appears. Briefly press the combination tool.
4Briefly press the combination tool until the fourth "0" (counted from the left) flashes.
5Turn the combination tool until "2" appears. Briefly press the combination tool.
6The device automatically changes over to the parameter level.
If an incorrect code was entered:
The display automatically changes over to the code entry after 3 seconds (or press the
combination tool). Repeat steps 2 to 5.
6.3Operation cancellation
How to proceed:
StepActivity
1Press the combination tool for more than 3 seconds or
2wait for "Timeout" (no action for more than 60 seconds).
43
6 Commissioning/start-up
6.4Setting the unit of the measured value (Uni.T)
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "Uni.T" appears in the bottom line.
3"Press"
The measured temperature is displayed in °C.
4"Turn"
The measured temperature is displayed in °F.
Setting:
°C = Temperature unit °C (factory setting)
°F = Temperature unit °F
5Confirm the setting: "Press" until the display stops flashing.
6.4.1Display and setting possibilities of the device
Measuring rangeTemperature unitDisplay
StartEnd
-50to+150°C°C-50+150
-50to+260°C°C-50+260
-50to+500°C°C-50+500
-58to+302°F°F-58+302
-58to+500°F°F-58+500
-58to+932°F°F-58+932
44
6 Commissioning/start-up
6.5Setting the offset (zero point) (Off.T)
6.5.1Edited offset setting
With this setting, the measured temperature can be increased by an adjustable value.
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "Off.T" appears in the bottom line.
3"Press"
TIP!
"-" means: the offset is negative - the measured temperature is reduced.
The value is entered "digit-by-digit".
45
6 Commissioning/start-up
6.6Setting the filter time constant (damping) (DamP)
The filter time constants (damping) can be used to "damp" the measured value.
Small filter time constant: the measured value is updated faster.
Large filter time constant: the measured value is updated slower.
The value is entered in seconds with two decimal places.
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "DamP" appears in the bottom line.
3"Press"
6.7Setting the mains frequency (Freq)
This setting can be used to change the mains frequency from 50 Hz to 60 Hz.
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "Freq" appears in the bottom line.
3"Press"
Setting:
50H = Mains frequency 50 Hz (factory setting)
60H = Mains frequency 60 Hz
46
6 Commissioning/start-up
6.8Setting the output signal (S.Type)
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "S.Type" appears in the bottom line.
3"Press"
Setting:
4.20A = Output signal 4 to 20 mA
0.20A = Output signal 0 to 20 mA
0.10U = Output signal 0 to 10 V
6.8.1Output performance
The output signal behavior in the event of an abrupt change of the input signal is specified in
the following table and the figure.
OutputDead time t1Time constant t2
Current output< 200 ms at 50 Hz mains frequency
< 320 ms at 60 Hz mains frequency
Voltage output< 200 ms at 50 Hz mains frequency
< 320 ms at 60 Hz mains frequency
70 ms
80 ms
47
6 Commissioning/start-up
6.9Setting the output signal scaling
The output signal scaling describes how the measured temperature is "converted" to an output
signal.
Simple example:
Actual
The device has a measuring range (1) from -50 to +150 °C and the output signal 4 to 20 mA
(3).
Set point
The customer wants that the "Customer measuring range“ (2) -50 to +150 °C (100 % from the
measuring range (1)) complies with the 4 to 20 mA (100 %) output signal (3).
(1)Measuring range(2)Customer specific measuring range
(3)Output signal
48
Customer specific scaling
It is frequently desired that a part of the measuring range is scaled to the output signal.
Example:
Actual
The device has a measuring range (1) from -50 to +150 °C and the output signal 4 to 20 mA
(3).
Set point
The customer wants that the "Customer scaling range“ (2) 0 to 50 °C (25 % from the measuring
range (1)) complies with the 4 to 20 mA (100 %) output signal.
Scaling is 1 : 4 (25 % : 100 %).
6 Commissioning/start-up
(1)Measuring range(2)Customer scaling range
(3)Output signal
49
6 Commissioning/start-up
Inverting the output signal
The device offers the possibility to invert the output signal (3).
During this process the output signal
•0 to 20 mA is inverted to the output signal 20 to 0 mA
•4 to 20 mA is inverted to the output signal 20 to 4 mA
•0 to 10 V is inverted to the output signal 10 to 0 V
Example:
20 to 4 mA
(1)Measuring range(2)Customer scaling range
(3)Output signal
50
6 Commissioning/start-up
6.9.1Setting the scaling start value (Sc.Lo)
TIP!
The output signal can only be scaled on devices with analog output.
Setting range:Measuring range
Factory-setting:Measuring range start
Example:
The device has a measuring range from -50 to +150 °C.
The device output signal is 0 to 20 mA.
Objective:The range from 0 to 100 °C (customer scaling) is to be displayed on the out-
put side by 0to20mA.
Setting:Scaling start value (Sc.Lo) = 0
Scaling end value (Sc.Hi) = 100
Result:At a temperature below 0 °C, the device indicates an error (measuring
range gone below) and provides the respective error signal (0 mA) on the
analog output.
At a temperature of 0 °C, the device provides 0 mA on the analog output.
At a temperature of 100 °C, the device provides 20 mA on the analog output.
At a temperature exceeding 100 °C, the device indicates an error (measuring range exceeded) and provides the respective error signal (22 mA) on
the analog output.
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "Sc.Lo" appears in the bottom line.
3"Press"
TIP!
The value is entered "digit-by-digit“!
51
6 Commissioning/start-up
6.9.2Setting the scaling end value (Sc.Hi)
TIP!
The output signal can only be scaled on devices with analog output.
Setting range:Measuring range
Factory-setting:Measuring range end
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "Sc.Hi" appears in the bottom line.
3"Press"
TIP!
The value is entered "digit-by-digit“!
52
6 Commissioning/start-up
6.10Setting the error signal of the analog output (S.Err)
TIP!
An error signal for overrange or underrange is only transmitted on devices with analog output.
In the event of errors (e.g. probe break or probe short-circuit), the analog output transmits the
signal configured here.
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "S.Err" appears in the bottom line.
3"Press"
Setting:
Type of signalSetting
4 to20 mA
0to20mA
0to10V
= 3.4 mA
= 22 mA
= 0 mA
= 22 mA
= 0 V
= 10.7 V
53
6 Commissioning/start-up
6.11Behavior when the scaling range is exceeded
The standard signal range of the analog output is limited as per recommendation of Namur NE
43.
Type of signallower limitupper limit
4to20mA3.8mA20.5mA
0to20mA0mA20.5mA
0to10V0V10.2V
Example:
4to20mA (factory setting)
54
6 Commissioning/start-up
6.12Setting the switching function (B.Fct)
General information
The switching output behavior of the device can be set.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "B.Fct" appears in the bottom line.
3"Press"
Window function, N/O contact
Window function, N/C contact
58
6 Commissioning/start-up
6.13Setting the switching point (B.Sp)
Chapter 6.12 "Setting the switching function (B.Fct)", page 55
Setting range:Measuring range (> B.RSp)
Factory-setting:100 °C
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "B.Sp" appears in the bottom line.
3"Press"
TIP!
The value is entered "digit-by-digit“!
59
6 Commissioning/start-up
6.14Setting the release point (B.RSp)
Chapter 6.12 "Setting the switching function (B.Fct)", page 55
Setting range:Measuring range (> B.Sp)
Factory-setting:90 °C
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "B.RSp" appears in the bottom line.
3"Press"
TIP!
The value is entered "digit-by-digit“!
60
6.15Setting the hysteresis (B.HyS)
TIP!
Only with the window switching function.
Chapter 6.12 "Setting the switching function (B.Fct)", page 55
Setting range:0 to 500 °C
Factory-setting:1 °C
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "B.HyS" appears in the bottom line.
3"Press"
6 Commissioning/start-up
TIP!
The value is entered "digit-by-digit“!
61
6 Commissioning/start-up
6.16Setting the switch-on delay time (B.Dly)
Setting range:0.00 to 99.99 s
Factory-setting:0.00 s
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "B.Dly" appears in the bottom line.
3"Press"
TIP!
The value is entered "digit-by-digit“!
62
6 Commissioning/start-up
6.17Setting the display alignment (D.Dir)
Setting range:Std = Standard = Device vertical
turn = turned = Device turned through 180°
Factory-setting:Std
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "D.Dir" appears in the bottom line.
3"Press"
or
63
6 Commissioning/start-up
6.18Setting the display unit (D.Uni)
Setting range:Uni.t = Temperature unit set as for "Uni.T"
Chapter 6.4 "Setting the unit of the measured value (Uni.T)", page 44
Pro2 = Percent of the scaled measuring range = "Sc.Hi" minus "Sc.Lo"
Chapter 6.9.1 "Setting the scaling start value (Sc.Lo)", page 51
and
Chapter 6.9.2 "Setting the scaling end value (Sc.Hi)", page 52
Factory-setting:Uni.t
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "D.Uni" appears in the bottom line.
3"Press"
Uni.t = The measured value is displayed in the selected unit.
Chapter 6.4 "Setting the unit of the measured value (Uni.T)", page 44
or
Pro2 = The measured value is displayed in percent of the scaling range.
Chapter 6.9.1 "Setting the scaling start value (Sc.Lo)", page 51
and
Chapter 6.9.2 "Setting the scaling end value (Sc.Hi)", page 52
Example:
The device scaling range was set to 0 to 150 °C.
When the device measures a temperature of 75 °C, 50 % is displayed.
64
6 Commissioning/start-up
6.19Displaying the software version of the operating unit (SW.Di)
Setting range:Read only!
Factory-setting:-
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "SW.Di" appears in the bottom line.
3
65
6 Commissioning/start-up
6.20Displaying the software version of the signal unit (SW.Si)
Setting range:Read only!
Factory-setting:-
How to proceed:
StepActivity
1Unlock the device.
Chapter 6.2 "Unlocking the device (enter the code).", page 42
2"Turn" until "SW.Si" appears in the bottom line.
3
66
7 Setup program
7.1General information about the setup program
The setup program optionally available permits a comfortable and clearly structured setting of
the large number of device parameters. Settings made once can be saved on a data carrier as
a file and transmitted one to one to several devices.
Configurable parameters
Depending on the device, it is possible to set, for example:
•Measuring range and range limits
•Output behavior in the event of an overrange or underrange
•Functions of the K1 and K2 switching outputs
TIP!
For configuration, the device must be connected to the power supply.
Chapter 3 "Electrical connection", page 21
Connection
(1)Rated voltage supply DC 24 V(2)Device
(3)USB/TTL converter
included in VARTN 70/00456352
(a)Cable box 4-pin (straight) M 12x1
with PVC connection cable length 2000 mm
VARTN 40/00404585 or
Cable box 4-pin (angled) M 12x1
with PVC connection cable length 2000 mm
VARTN 40/00409334
(c)PC interface cable (gray)
Part of (3)
(4)Notebook/PC
(b)Connection cable (Y cable)
VARTN 70/00507861
67
7 Setup program
How to proceed:
StepActivity
1Install the setup program on the notebook/PC.
2Screw-fit the connection cable (b) to the connector of the device (2).
3Connect the USB/TTL converter (3) to the connection cable (b) and PC interface cable
(gray) (c).
4Connect the PC interface cable (gray) (c) to the notebook/PC.
5Connect the cable box to the rated voltage supply (1) and the connection cable (b) using
the PVC connection cable (a).
TIP!
During installation, the driver for the USB/TTL converter is also installed on the notebook/PC.
68
8.1Possible errors
DisplayPossible causesMeasure
8 Overcoming errors and malfunctions
•Overrange
or underrange
•Probe break
Device error
•1 = Internal
communication error
•2 = Error, analog output
•3 = Short circuit,
switching output 1
•4 = Short circuit,
switching output 2
•5 = VCC 8 V outside the
working range
•6 ... 8 = Internal
communication error
•9 = Invalid configuration
Display overrun:
Upper display: „- - - -“
Bottom display: Parameter name
Value is smaller than -9999
or exceeds +9999.
Check temperature and/or probe.
1, 6, 7, 8: Contact the service.
2: Check ambient temperature.
Check output for cable break. Output
load too high (for current output) or
too low (for voltage output).
3, 4: Check the switching output.
5: Check voltage supply.
9: Check the configuration.
Check switching output.
Check voltage supply.
69
8 Overcoming errors and malfunctions
70
9 Supplement
9.1Technical data
Electrical connection Machine connector M 12x1, 4-pin as per IEC 60947-5-2
Process connectionsScrew connection G 1/4, G 3/8 and G 1/2
SheathStainless steel 316 L Material No. 1.4404/1.4435
Protection ratingIP65 as per DIN EN 60529 with the machine connector inserted
Response timet
Measuring insertPt1000 temperature probe, DIN EN 60751, class A or AA (1/3 DIN B), 4-wire circuit
Screw connection M 12x1.5; M 18x1.5 and M 20x1.5
Screw connection 1/2-14NPT
Union nut G 3/8
Pipe screw connection G 1/4 and G 1/2
Screw connection G 1/2 with CIP-conforming sealing cone and EHEDG certificate
Conical port with union nut (milk pipe union)
Clamp
Spherical welding socket with clamp screw connection
Welding socket with CIP-conforming sealing cone
Varivent connections with EHEDG certificate
JUMO PEKA with EHEDG certificate
Stainless steel 316 Ti Material No. 1.4571
: 3 s in water 0.4 m/s
0.5
t
: 8 s in water 0.4 m/s
0.9
9.1.1General information
Reference conditionsDIN 16086 and DIN EN 60770
Displaypositively lit LCD display
AlignmentThe display can be mirror-imaged by 180° using the setup program
after installation, the display case can be swivelled to the LH or RH side by ± 160°
(use the combination tool)
SizeDisplay 16 x 26 mm / font size 7 mm / 2x 4-digit
ColorStandard operation: amber
Error: red (text "Err", error code 1 to 9 flashes)
Setup interface occupied: red
Switching status displayK1, K2
Temperature unit°C or °F
Operation
on the devicewith the rev transmitter beneath the locking screw with combination tool or screwdriver
with PCwith the setup program with PC interface
0.5 x 3 mm or Allen key, width across flats 2
9.1.2Input
Measuring input1x Pt1000 temperature probe, 4-wire circuit
Measuring rangeBasic type 902940/10: -50 to +150 °C
Limit value deviation0.15 + 0.002 x t, class A
Basic type 902940/30: -50 to +260 °C
Basic type 902940/50: -50 to +500 °C
0.10 + 0.017 x t, class AA (1/3 DIN B)
71
9 Supplement
9.1.3Measuring circuit monitoring
Probe short-circuit,
probe and cable break,
measuring range gone below,
measuring range exceeded
Analog output 0 to 20 mA, 0 mA or 22 mA configurable
Analog output 4 to 20 mA, 3.4 mA or 22 mA configurable
Analog output 0 to 10 V, 0 V or 10.7 V configurable
Switching outputs, low
Additional error message on the LCD display
9.1.4Outputs
All analog outputs in 3-wire circuit/Open Collector, PNP switching output
Analog output
can be freely configured4 to 20 mA and 1x PNP switching output
Switching output
Number1x PNP switching output
Switching typeN/C / N/O
Switching functionWindow / hysteresis
Switching capacity
- Voltage drop from UBPNP ≤ 2V
- Contact ratingON ≤ 250 mA/OFF ≤ 1mA
- Switching cycles> 10 million
Reaction timeat 50 Hz: ≤ 200 ms
Short-circuit proofYes
Load check, current
- Period duration2 s; T
- Periodic protection circuitry
for overcurrent
Scaling range
Analog outputScaling can be freely selected within the measuring range
Behavior when the scaling range is gone
below
Behavior when the scaling range is
exceeded
Switching output
- Switching pointMeasuring range ( > Release point)
- Release pointMeasuring range ( < Switching point)
- Damping0.00 to 99.99 s
- Switch-on delay0.00 to 99.99 s
Apparent ohmic resistance
4to20mAR
0to20mAR
0to10VR ≥ 10 kΩ
0 to 20 mA and 1x PNP switching output
0 to 10 V and 1x PNP switching output
Analog output 0 to 20 mA, linear drop to 0 mA
Analog output 4 to 20 mA, linear drop to 3.8 mA
Analog output 0 to 10 V, linear drop to 0 V
Analog output 0 to 20 mA, linear increase up to 20.5 mA
Analog output 4 to 20 mA, linear increase up to 20.5 mA
Analog output 0 to 10 V, linear increase up to 10.2 V
≥ (UB - 6.5 V)/0.022 A
I
≥ (UB - 6.5 V)/0.022 A
I
72
9 Supplement
9.1.5Ambient conditions
Admissible temperatures
Ambient temperature, display case-25 to +75 °C
Ambient temperature-50 °C; restricted function
Storage temperature-40 to +85 °C
Admissible air humidity
- during operation100 % including condensation on the device outer case
- during storage90 % without condensation
Admissible mechanical loadrefering to basic type 902940/10 and 902940/30 with fitting length 100 mm
- Vibration resistance10 g, 10 to 2000 Hz as per IEC 60068-2-6
- Shock resistance50 g for 11 ms / 100 g for 1 ms as per IEC 60068-2-27
Electromagnetic compatibility(only with 4-pin connection line and ground case)
- emitted interferenceClass A as per EN 61326
- interference resistancePerformance characteristic A as per EN 61326
Protection ratingIP65 as per DIN EN 60529
Ambient temperature error≤ 0.05 %/Κ
only stationary use, risk of cable break, LCD display without function
9.1.6Auxiliary energy
Voltage supplyDC 14 to 30 V (nominal voltage supply DC 24 V)
- for output 0(4) to 20 mADC 12 to 30 V
- for output 0 to 10 VDC 14 to 30 V
Reverse voltage protectionYes
Power consumption≤ 45 mΑ without load, ≤ 545 mΑ with 2x PNP switching output
Electrical connectionMachine connector M 12x1, 4-pin as per IEC 60947-5-2, A-coded
CircuitSELV
Influence of the voltage supply≤ 0.02 %/V deviation from DC 24 V
Ripple: ensure that the voltage peaks do not exceed or go below the specified values