JUMO 902830 User Manual

JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 e-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 e-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-JUMO
1-800-554-JUMO Fax: 315-697-5867 e-mail: info@jumo.us Internet: www.jumo.us
Push-in resistance thermometers in steam-tight version
for temperatures from -70 to +200°C
as single, twin or triple resistance thermometer
in 2-wire, 3-wire or 4-wire circuit
connecting cable in
Push-in resistance thermometers are preferably used for temperature measurement in sterilizers. Thanks to their special construction, they are highly suitable for application in pressurized atmospheres containing steam. Other applications include plant engineering and laboratories.
The PTFE connecting cable is suitable for use in humid areas within the temperature range
-190 to +260°C, the FEP cable from -70 to +200°C. The cable connection incorporates strain relief.
The measuring insert is normally fitted with a Pt100 temperature sensor to EN 60 751, Class A in 2-wire circuit. Versions with 2 or 3 measuring circuits are also available. 3-and 4-wire circuit connections can be provided.
PTFE, FEP and silicone
Data Sheet 90.2830 Page 1/3
Technical data
Connection cable ends available as: bare wires, with ferrules, receptacles or multipole connector. Connecting cable PTFE, ambient temperature -190 to +260°C
FEP, ambient temperature -70 to +200°C silicone, ambient temperature -50 to +150°C connecting cable available with shielding (option)
Protection tube stainless steel 1.4571, 4mm, 4.5mm, 6mm dia. Measuring insert Pt100 temperature sensor, EN 60 751, Cl. A, 2-/3-/4-wire circuit Protection IP69
08.05/00386335
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 e-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 e-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-JUMO
1-800-554-JUMO Fax: 315-697-5867 e-mail: info@jumo.us Internet: www.jumo.us
Dimensions
Data Sheet 90.2830 Page 2/3
Type 902830/10 Type 902830/30
08.05/00386335
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 e-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 e-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-JUMO
1-800-554-JUMO Fax: 315-697-5867 e-mail: info@jumo.us Internet: www.jumo.us
Order details: Push-in resistance thermometers in steam-tight version
(1) Basic version
Data Sheet 90.2830 Page 3/3
902830/10
902830/30
Push-in resistance thermometer, steam-tight
Push-in resistance thermometer, steam-tight, pressure-tight up to 5.0bar
(2) Operating temperature in °C
x 302 -70 to +200°C / FEP connecting cable x 370 -50 to +150°C / PTFE connecting cable x 371 -50 to +150°C / silicone connecting cable
(3) Measuring insert
x 1001 1 x Pt100 in 3-wire circuit x 1003 1 x Pt100 in 2-wire circuit x x 1011 1 x Pt100 in 4-wire circuit x 2001 2 x Pt100 in 3-wire circuit x x 2003 2 x Pt100 in 2-wire circuit x x 2011 2 x Pt100 in 4-wire circuit (not in conjunction with 371) x 3028 3 x Pt100 1 x 2-wire circuit, 2 x 3-wire circuit (not in conjunction with 371)
(4) Tolerance class to EN 60 751
xx 2 Class A
(5) Protection tube diameter D in mm
x44mm
x 4.5 4.5mm with 2 measuring circuits in 4-wire circuit x66mm
(6) Fitting length EL in mm (50 EL 500)
x5050mm x x 100 100mm x x 150 150mm x x 200 200mm x x ... please specify in plain text (in 50mm steps)
(7) Connecting cable end
x x 03 bare cable ends x x 11 ferrules to DIN 46 228 Part 4 (standard) x 13 receptacle 6.3 to DIN 46 247 x x 80 multipole connector (specify type in plain text)
(8) Connecting cable length AL in mm (500 AL ≤ 500000)
x x 2500 2500mm (standard) x x ... please specify in plain text (in 500mm steps)
(9) Extra codes
x x 000 no extra code
x 304 without protection tube (sensor enclosed in PTFE sleeve) x x 312 curved protection tube R 13mm (only 4.5mm dia. on Type 902830/30) (please specify dimensions in plain text) x 317 shielded connecting cable
Order code Order example 902830/10 - 370 - 1011 - 2 - 6 - 100 - 11 - 2500 /
-------/ ,...
1
000
1. List extra codes in sequence, separated by commas.
Stock versions
(1) (2) (3) (4) (5) (6) (7) (8) (9)
(1) (2) (3) (4) (5) (6) (7) (8) (9) Sales No.
902830/10 - 370 - 1011 - 2 - 6 - 100 - 11 - 2500 / 317 90/00306448 902830/10 - 370 - 2001 - 2 - 6 - 100 - 11 - 2500 / 000 90/00306449
902830/30 - 302 - 1011 - 2 - 4 - 50 - 11 - 2500 / 000 90/00445084 902830/30 - 302 - 2011 - 2 - 4.5 - 50 - 11 - 2500 / 000 90/00445085
08.05/00386335
-- ----- /
Loading...