Thermowells for Heat Meter Temperature Probes
• For RTD temperature probes
• RTD temperature probes can be replaced without emptying the system
• Thermowells made from various materials
• Compliant with magnetic inductive flow sensors with identification marking
• Available with inspection certificate 3.1
Thermowells are used when it is necessary to replace the RTD temperature probes without emptying the system and/or without pressure resistance. Short thermowells are predominantly
screwed into T-pieces while longer design types are screwed into special welded sockets or
welded directly in the pipe wall.
For nominal values up to DN 50, the protection tube should dip above the middle. For larger diameters it should dip into the cross-section of the pipe so that it is deep enough (≥ 50 mm) to
avoid heat conduction errors.
In addition, one must pay attention to the minimum immersion depth of the used probe.
The specified dimensions for EL refer to the insertion length of the RTD temperature probe; the
dimensions for L provide the insertion lengths of the thermowells.
Page 1/4Data sheet 902440
Technical data
Process connection Screws G 1/4, G 3/8, G 1/2, G 3/4
Protection tube Material:
Stainless steel 1.4571
Stainless steel 1.4435
Brass
Insertion lengths L = 35 to 400 mm
2013-10-25/00386327

JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: 315-437-5866
1-800-554-5866
Fax: 315-437-5860
E-mail: info.us@jumo.net
Internet: www.jumousa.com
Dimensions
Page 2/4Data sheet 902440
Basic type 902440/40 Basic type 902440/41 Basic type 902440/42
Basic type 902440/43 Basic type 902440/44 Basic type 902440/45
2013-10-25/00386327

JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: 315-437-5866
1-800-554-5866
Fax: 315-437-5860
E-mail: info.us@jumo.net
Internet: www.jumousa.com
Order details
(1) Basic type
902440/40 Screw-in thermowell
with sealing screw, straight protection tube,
approved up to 90 °C, PN = max. 25 bar
Page 3/4Data sheet 902440
902440/41 Screw-in thermowell
with internal screw thread M 10 x 1,
straight protection tube,
approved up to 130 °C, PN = max. 25 bar
902440/42 Screw-in thermowell
according to DIN EN 1434 with sealing screw,
External Ø 8 mm stepped down to 7.5 mm,
Internal Ø 6 mm, fit tolerance H10,
approved up to 200 °C, PN = max. 40 bar
902440/43 Screw-in thermowell
with sealing screw,
External Ø 8 mm, internal Ø max. 6.2 mm,
approved up to 200 °C, PN = max. 40 bar
902440/44 Screw-in thermowell
with sealing screw,
External Ø 12 mm stepped down to Ø 9 mm,
Internal Ø 6 mm, fit tolerance H12, PN = max. 40 bar
902440/45 Screw-in thermowell
with internal screw thread 1/4" and sealing sheet,
External Ø 8 mm, internal Ø 6.15 mm,
approved up to 180 °C, PN = max. 40 bar
(2) Internal diameter of protection tube in mm
xx 5 Ø5mm
xxx 5.2 Ø5.2mm
xxxxx 6 Ø6mm
xx 6.2 Ø6.2mm
(3) Insertion length L in mm (85 to 400 mm for basic type 902440/42 and 902440/43)
x 35 35 mm (EL = 46 mm)
x x 40 40 mm (EL = 47 mm)
x x 50 50 mm (EL = 57 mm)
x x x 85 85 mm (EL = 98 mm, for basic type 902440/45 EL = 85.5 mm)
x 92.5 92.5 mm (EL = 93 mm)
xxxx 120 120mm (EL = 133mm, for basic type 902440/45 EL = 120.5 mm)
x 147.5 147.5 mm (EL = 148 mm)
x x x 155 155 mm (EL = 168 mm)
x x x 210 210 mm (EL = 223 mm)
... Specifications in plain text
(4) Process connection GW
x x 102 External thread G 1/4
x 103 External thread G 3/8
x xxxx 104 External thread G 1/2
x 105 External thread G 3/4
(5) Material
xxxx 20 Stainless steel
x x 46 Brass
(1) (2) (3) (4) (5)
Order code ---Order example 902440/40 - 5 - 40 - 102 - 46
2013-10-25/00386327

JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: 315-437-5866
1-800-554-5866
Fax: 315-437-5860
E-mail: info.us@jumo.net
Internet: www.jumousa.com
Stock versions
(1) (2) (3) (4) (5) Part no.
902440/40 - 5 - 35 - 102 - 46 00340460
902440/40 - 5.2 - 35 - 102 - 46 00330518
902440/40 - 5.2 - 50 - 103 - 46 00330497
902440/40 - 5.2 - 50 - 104 - 46 00326403
902440/41 - 5.2 - 40 - 102 - 46 00362827
902440/43 - 6.2 - 85 - 104 - 20 00349100
902440/43 - 6.2 - 120 - 104 - 20 00352465
Page 4/4Data sheet 902440
2013-10-25/00386327