Julabo CD-1001F Instruction Manual

Immersion / Open Bath / Refrigerated Circulators
Pos: 2 /Technische Dokumentation/Flash CORIO/Einhängethermostat CORIO C/Titelfoto CD @ 0\mod_1395828649188_ 0.docx @ 4299 @ @ 1
=== Ende der Lis te fü r Textm ar ke I nh alt1 = ==
Important: keep original operating manual for future use.
1.953.0801-V3
english
Table of contents
1 Intended use ........................................................................................ 5
2 Description ........................................................................................... 5
3 Explanation of safety information ..................................................... 6
3.1 Explanation of other information ............................................................ 6
4 Safety instructions .............................................................................. 7
5 Operator's responsibility - safety instructions ................................. 9
5.1 Requirements for the operating personnel .............................................. 9
5.2 Operating and ambient conditions for using the unit .............................. 9
5.3 Operating the unit ................................................................................. 10
6 Control and function elements......................................................... 11
6.1 Installation of the circulator .................................................................. 14
6.2 Bath fluids ............................................................................................. 14
7 Commissioning .................................................................................. 16
7.1 Excess temperature and low level safety devices .................................. 17
7.2 Filling the 1001F ................................................................................... 18
7.3 Filling quantities and shut-off criteria ................................................... 18
7.4 Switching on / Start - Stop .................................................................... 19
7.5 Setting the desired temperature value .................................................. 19
Setting the timer ........................................................................................ 20
7.6 ATC - Absolute Temperature Calibration (1-point calibration) .............. 21
7.7 Switch Autostart on and off .................................................................. 22
8 Error messages / Possible causes of faults ...................................... 22
9 Operation of the CORIO CD via the USB interface .......................... 26
10 Emptying the bath tank .................................................................... 28
11 Technical data .................................................................................... 29
11.1 Technical data for circulator .................................................................. 29
11.2 Technical data for refrigeration circulation circulator ............................ 30
11.3 Refrigerant ............................................................................................ 31
12 Materials of parts in contact with the bath fluid ........................... 32
2
Intended use
12.1 Circulator .............................................................................................. 32
13 Accessories ........................................................................................ 33
13.1 For external connection ......................................................................... 33
13.2 For open baths ...................................................................................... 33
14 Maintenance, cleaning, storage ....................................................... 33
14.1 Maintain the refrigeration capacity. ...................................................... 33
14.2 Cleaning ................................................................................................ 34
14.3 Storage .................................................................................................. 34
15 Repair service .................................................................................... 34
15.1 Warranty ............................................................................................... 35
16 Waste disposal .................................................................................. 38
16.1 Packaging.............................................................................................. 38
16.2 Unit ....................................................................................................... 38
16.3 Refrigerant ............................................................................................ 38
Pos: 4 /Technische Dokumentation/Flash CORIO/Einhängethermostat CORIO C/Neue Seite @ 0\mod_1417593782 389_0.d ocx @ 76 30 @ @ 1
3
The operating manual
Pos: 5 /Technische Dokumentation/Flash CORIO/Glückwunsch - Einleitung @ 0\mod_13951368662 01_18.d ocx @ 4161 @ @ 1
Congratulations.
You have made an excellent choice. JULABO thanks you for the trust. This operating manual is designed to familiarize you with the operation
of our units and their possible applications. Please read the operating manual carefully.
Please call us if you have any questions about the operation of the unit or about the operating manual.
Contact:
JULABO USA, Inc.
884 Marcon Boulevard Allentown, PA 18109 Phone: +1(610) 231-0250 Fax: +1(610) 231-0260 info.us@julabo.com www.julabo.com
The JULABO quality management system
The standards for the development, production and distribution of temperature control devices for laboratory and industry use satisfy the requirements of ISO 9001 and ISO 14001. Registration certificate No. 01 100044846.
Unpacking and inspection
If the packaging is damaged or if you discover any concealed transport damage when you have unpacked the devices and the accessories, please notify the supplier in the form of a statement of damage.
• should be kept for future use.
• must be available to operating personnel at all times.
Pos: 6 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/CD-Produktübersicht @ 0\mod_140 930467092 6_18.doc x @ 7330 @ 1 @ 1
4
Intended use
safety instructions at your workstation.

1 Intended use

JULABO circulators are laboratory devices which are designed for the temperature control of certain liquid media in a bath tank or with a refrigeration unit. The bath fluids recommended by JULABO must be used as tempering media. Units with pump connections allow the tempering tasks to be carried out in an external temperature control system. JULABO circulators are not designed for the direct temperature control of foods, semi-luxury foods and tobacco, or pharmaceutical and medical products.
Direct temperature control means unprotected contact between the object and the tempering medium (bath fluid).
The devices are not suitable for use in potentially explosive
Pos: 8 /Technische Dokument

2 Description

Pos: 10 /Tech nis ch e Doku m ent atio n/Fl ash CO RIO /COR IO _CD_ GESA MT/ Ve rhal ten + BA les en @ 0\mod_ 140921 8461856_1 8.docx @ 7203 @ @ 1
environments.
These circulators are operated via the splash-proof keypad. The microprocessor technology allows the setpoint to be set, displayed and saved using the temperature display LED.
The PID temperature control automatically adjusts the heat supply to the requirements in the bath.
ATC - Absolute Temperature Calibration (1-point calibration)
The pump function can be switched from internal to external
circulation by a simple switchover button.
Safety equipment to IEC 61010-2-010:
- Excess temperature protector is a safety device which is
independent of the control circuit whose value is set using a tool (screwdriver).
- A float switch acts as the low level safety device. If these
safety devices trip, the heater and circulating pump are completely shut down.
Interfaces:
- CAN bus to communicate with the refrigeration machine
- USB connector to act as host to update and read data. USB device
for remote control.
Pos: 11 /Tech nis ch e Doku m ent atio n/Fl ash CO RIO /Einh än geth ermo st at C ORIO C/Neue S eite @ 0\mod_ 141759 378238 9_0.doc x @ 763 0 @ @ 1
It is important to follow these safety instructions to prevent personal injury and property damage. These instructions apply in addition to the
It is essential that you read the user information before starting the device.
5
This signal word designates a danger with a low level of risk which, if it
Designates user tips and other useful information.

3 Explanation of safety information

The operating manual contains warnings to increase safety when using the unit. The general warning sign, consisting of an equilateral triangle surrounding an exclamation sign and reproduced in various signal colors, is preceded by the signal words.
"Warning of a dangerous situation". The significance of the danger is classified with a signal word. Read the
instructions carefully and follow them.
This signal word designates a danger with a high level of risk which, if it not prevented, will result in death or serious injury.
This signal word designates a danger with a medium level of risk which, if it not prevented, may result in death or serious injury.
not prevented, may result in minor or moderate injury.
Designates a possibly harmful situation. If it is not prevented, the system or something near it may be damaged.

3.1 Explanation of other information

TIP
Your attention is drawn to something special by this.
Dangers at second glance
Designates states which only occur after the start of an action and could have been prevented if the warning had been heeded.
Informative note
Provides additional information.
Pos: 13 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/ Sich erhe its anw eisu ng en @ 0 \mod_ 1409 218454 274_18.do cx @ 6917 @ 1 @ 4
6
Safety instructions

4 Safety instructions

It is important to follow these safety instructions to prevent personal injury and property damage. These instructions apply in addition to standard safety practices for working places.
It is essential that you read the user information before starting the device
• Use PPE (safety gloves, safety shoes, safety goggles).
Transport the unit carefully. The interior of the unit can also be
damaged by impacts or if it is dropped.
• Do not loiter under the unit during transportation and operation.
• The unit is not intended for use in potentially explosive areas.
Please observe the specifications for the minimum space
requirement when setting up the unit in chapter 12.3 Refrigerant, page 32.
Only operate the unit in rooms that are well-ventilated, dry and free of frost.
• Switch the unit off immediately if there is refrigerant leakage.
Place the unit on a flat surface of non-flammable material.
• Operate the unit under an exhaust hood as much as possible.
• Do not start the unit if it is damaged or leaking.
Compare the mains voltage and frequency with the specifications on
the type plate.
Only connect the unit to a fused mains connection via a FI circuit breaker (Ia=30 mA).
Only connect the unit to a power socket with ground contact (PE – protective earth)!
The power supply plug serves as safe disconnecting device from the power supply network and must be freely accessible at all times.
Check the mains cable regularly for signs of damage.
• Do not start the unit if it has a damaged power cable.
• Keep the mains cable away from hot pump connections.
• Refer to the safety sticker. Parts of the unit can be hot or cold.
• Never use the unit without bath fluid.
Do not reach into the thermal bath fluid.
Check the filling level of the bath fluid at regular intervals. The pump
and heater must always be completely covered with bath fluid.
Set the excess temperature protector below the flash point of the bath fluid.
7
Monitor the heat expansion of the bath oils as the bath temperature rises.
• Prevent water getting into hot bath oils.
• Use suitable tubing.
• Secure the tubing connections to prevent them sliding off.
• Do not bend the bath fluid tubing.
Check the hoses at regular intervals for signs of material fatigue (for
example cracking).
• Do not drain the bath fluid when it is hot.
Check the temperature of the bath fluid before draining it, for
example by switching on the unit briefly.
Switch off the unit and pull the plug before moving the unit or carrying out service or repair work.
Have all service and repair work carried out by authorized specialists only.
Switch off the unit and disconnect it from the power supply before cleaning it.
Pos: 14 /Technische Dokumentation/Flash CORIO/Verantwortung des Betreibers Therm. Kl. 3, brennbar @ 0\mod_1395046754147_18.do cx @ 4 104 @ 122 2 @ 1
• Drain the unit completely before transporting it.
8
Operator's responsibility - safety instructions

5 Operator's responsibility - safety instructions

Products manufactured by JULABO GmbH ensure safe operation when installed, operated and according to common safety regulations. This section explains the potential dangers which may occur when operating the unit and specifies the most important safety measures to prevent these dangers as far as possible.

5.1 Requirements for the operating personnel

The operator is responsible for the qualifications of the personnel operating the unit. Ensure that the personnel who operate the unit are trained in the relevant work application by a trained person.
The operative must receive regular training about the dangers involved with their work and about action to prevent such dangers.
Ensure that everybody involved with the operation, maintenance and installation have read and understood the safety information and the operating manual. The unit may only be configured, installed, maintained and repaired by trained personnel.
If hazardous substances or substances which may become hazardous are used, the unit may only be used by a person who is completely familiar with these substances and the unit. This person must be able to assess the possible dangers in full.

5.2 Operating and ambient conditions for using the unit

Avoid impacts on the housing, vibrations, damage to the operative
keypad (keys and display) and heavy soiling.
Ensure that the product is checked at regular intervals suitable for its
frequency of use to ensure that it is in perfect condition.
Check the proper condition of the mandatory warning, prohibition,
and safety labels at least every 2 years.
Ensure that the mains supply has a low impedance to prevent
influencing of other units powered in the same mains.
The unit is designed for operation in a controlled electromagnetic
environment. This means that in an environment of this nature, transmission equipment such as mobile phones should not be used in the immediate vicinity.
Other units with components which are suceptible to magnetic fields
may be influenced by magnetic radiation. We recommend to maintain a minimum distance of 1 m.
Permissible ambient temperature: max. 40 °C, min. 5 °C.
The relative humidity should not exceed 50 % (40 °C).
• Do not store in an aggressive atmosphere. Protect from dirt.
• Protect from direct sunlight.
9
1
2a
2b

5.3 Operating the unit

The bath may be filled with flammable materials. Fire hazard! Chemical dangers may occur, depending on the bath medium. Refer to all warnings on the substances used (bath fluids) and in the
relevant instructions (safety data sheets). The formation of explosive mixtures is possible if the ventilation is inadequate. Only use the units in well ventilated areas. The unit is not suitable for use in potentially explosive environments. Special substance specifications (bath fluids) must be observed for correct operation. Caustic or corrosive bath fluids must not be used. When using hazardous substances or substances which may be hazardous, the operator must apply the enclosed safety symbols (1 + 2a or 2b) on the control side panel where they are clearly visible:
Warning of a danger zone. Attention! Observe documentation. (Operating manual, safety data sheet)
It is essential that you read the user information prior to operation. Area of validity: EU
It is essential that you read the user information prior to operation. Area of validity: USA, NAFTA
3
As a result of the wide range of operating temperatures, special care and caution is essential.
There are thermal dangers: Burns, scalds, hot steam, hot parts and surfaces which may be touched.
If external units are connected
Refer to the instructions in the manuals for the external units which you connect to the JULABO unit, particularly the safety instructions.
The connection assignment of the plugs and the technical data for the
Pos: 15 /Tech nis ch e Doku m ent atio n/Fl ash CO RIO /Bedi en- und Funktionselemente CD-F300 @ 0\mod_1396265311140_ 18.doc x @ 43 87 @ 1 @ 1
10
products must be observed at all times.
Warning label Proposition 65
Warning about hot surfaces. (The label is applied by JULABO)
Control and function elements

6 Control and function elements

Front Rear
Position Designation
1 Main switch
2 Four-digit temperature display LED, menu display
3 Control indicator - alarm
4 Edit keys
Temperature setpoint increase or decrease Press the key briefly for step-by-step changes. Press and hold the key for fast change of setpoint.
5 OK key 1. Switch the device on/off.
2. Store value / parameter.
6 Pump switchover,
Delivery
external…….internal
7 Adjustable excess temperature safety device
8 Control indicator - cooling
9 Control indicator - heating
10 USB host interface (type A)
11
11
12 Fuses: Miniature circuit-breaker
13 CAN plug for connecting to the refrigeration machine.
14 USB device interface (type B) for data transfer to the PC, for example for control tasks
15 Pump connection: Return, IN M16x1, male
16 Pump connection: Supply, OUT M16x1, male
Refrigerated Circulators
Mains connection: Integrated connector for voltage supply (mains cable included as accessory)
using the EasyTEMP software.
17 Recessed handle (front, rear)
18 CAN connection cable for refrigeration machine circulator
19 CAN plug for connecting to the circulator.
20 Fuses: Miniature circuit-breakers, for refrigeration machine
21 Connection cable: Voltage supply, refrigeration machine circulator
12
Control and function elements
22 Electrical connection: Integrated connector to supply power to the circulator
23
Electrical connection: Integrated connector to supply power to the refrigeration circulator
24 Drain cock and drain port (behind ventilation grille)
25 Caps (connectors for the cooling coil)
26 Equipment castors (front lockable)
27 Ventilation grille
Accessories, included in the supply
1x Main cable for voltage supply for the refrigeration machine (23) and circulator (11) (use one only for refrigeration circulator)
1x Connection cable: Refrigeration machine (22) Circulator (11)
1x CAN connection cable (18, for refrigeration machine circulator)
Pos: 16 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Aufstellen, Achtung @ 0\mod_140921844399 4_18.docx @ 6532 @ 2 @ 4
Mains cable, country-specific, order number, see www.julabo.com.
13
Observe the safety sticker - do not remove!
Evaporation and constant refilling may produce a concentration

6.1 Installation of the circulator

Things to keep in mind during the installation process:
Pos: 25 /Tech nis ch e Doku m ent atio n/Fl ash CO RIO /COR IO _CD_ GESA MT/ Temp er ierf lü ss igkeit , Au sw ahl , br enn bare, Wa rnu ng (CD) @ 0\mod _140921 8456099_18 .docx @ 6983 @ 2 @ 1

6.2 Bath fluids

it - Wasser @ 0\mod_ 140921 845585 0_18.d ocx @ 69 72 @ 3 @ 1
Risk of tipping due to improper transportation Crushing, damage to the unit
Use PPE (safety gloves, safety shoes, safety goggles).
Carry the unit with 2 persons (see the Technical Data for the
weight).
Transport the unit carefully on firm, level ground. The interior of the unit can also be damaged by impacts or if it is dropped.
Do not loiter under the unit during transportation and operation.
The installation site should be a sufficiently large room to ensure
that it does not become too hot due to the heat emission.
The surface for the device should be flat and made of non­flammable material.
Lock the castors, once the device is at the operating location.
A specific room size is prescribed for refrigerated circulators.
At high temperatures, position the unit under an exhaust hood as
much as possible due to potential vapors from the thermal bath fluid.
Water as the bath fluid
If you use water as the bath fluid
Recommended water mixture: 70 % soft/decalcified water and 30 % tap water for a temperature range from 5 °C to 80 °C.
The parts of the bath which come into contact with the bath fluid may be damaged and cause the failure of the device.
The water quality depends on the local conditions.
Hard water is not suitable for temperature control tasks due to its high lime content and will produce lime deposits in the bath.
Ferrous water can cause corrosion, even on stainless steel.
Chloric water can cause pitting corrosion.
Distilled and deionized water is not suitable. Their specific
properties cause corrosion in the bath, even on stainless steel.
Check the quality of the water you use at regular intervals.
14
Control and function elements
of harmful substances in the bath.
You should therefore check the quality of the water in the bath at regular intervals.
Replace the water in the bath in full at regular intervals.
Water bath protection products
The water bath protection product "Aqua-Stabil" is recommended to combat algae, bacteria and fungus formation. Order No. Designation 8 940 006 6x 100 ml bottles
Pos: 27 /Tech nis ch e Doku m ent atio n/Fl ash CO RIO /COR IO _CD_ GESA MT/ Temp er ierf lü ss igkeit en, Ung ee igne te - SHKL 3 @ 0\mod_140921 845744 1_18.d ocx @ 7038 @ @ 1
Pos: 28 /Tech nis ch e Doku m ent atio n/Fl ash CO RIO /COR IO _CD_ GESA MT/I nd ire kt t emp eri ert e Flü ss igkei t -Vo rsich t - @ 0\mo d_140 921845 0811_18 .doc x @ 6774 @ @ 1
8 940 012 12x 100 ml bottles
Unsuitable bath fluids.
JULABO cannot accept any liability for damage caused by the selection of an unsuitable bath fluid.
Unsuitable products include bath fluids which
are highly viscous (much higher than recommended at the relevant working temperature).
tend to crack.
have a toxic, caustic or corrosive effect.
Properties of indirectly temperature-controlled fluids and substances
The intended use of the units includes the indirect temperature control of fluids.
We do not know which substances these are. Many substances are:
inflammable, flammable or explosive
harmful
polluting
In other words: dangerous
The user bears sole responsibility for handling these substances! Use personal protective equipment!
15
Pos: 29 /Technis che Doku m ent atio n/Fl ash COR IO /COR IO_ CD_ GESAM T/ Temp er iers yst em, ext ern, Pum p enset @ 0 \mod_1 409218 458486 _18.doc x @ 70 82 @ 2 @ 1

7 Commissioning

Commissioning the circulator with a refrigeration machine
Connect the circulator and refrigeration machine using the mains lead. Connect them to the voltage supply using the fitted plug on the refrigeration machine and the mains lead. Connect the CAN jacks on
Pos: 32 /Technische Dokumentation/Flash CORIO/Warnung heiße Oberfläche @ 0\mod_1389686088320_18.docx @ 3668 @ @ 1
both devices with the CAN connection cable to transfer data.
The following questions should help to identify possible dangers and minimize risk.
Are hazardous vapors or gases produced when heated? Does operation of the bath has to be conducted in a fume hood?
What should you do if a dangerous substance has been spilled on
or in the device? Obtain information on the substance before starting work and define a decontamination method.
Are all hoses and electrical cables securely connected and routed?
Keywords: Sharp edges, hot surfaces during operation, moving machine parts, etc.
Danger from mains voltage.
Risk of injury from electric power.
Compare the mains voltage and frequency with the details on the model plate.
Only connect the unit to a fused mains connection via a FI circuit breaker (Ia=30 mA).
The device may only be connected to power outlets with a ground contact (PE – protective earth).
The mains plug serves as a safe disconnecting device from the power supply network and must be freely accessible at all times.
• Do not start the device if it has a damaged mains cable.
• Check the mains cable regularly for signs of damage.
We disclaim all liability for damage caused by incorrect line
voltages!
Cold or hot surface temperatures of the unit
Unit parts may develop low or high surface temperatures. A hot surface means it has a temperature of 60 °C / 140 °F or more.
Let the device cool down to an uncritical safe temperature.
Pos: 33.1 /Technische Dokumentation/Flash CORIO/Schutzeinrichtung, Übertemperatur, Unterniveau - Erklärung @ 0\mod_ 1389 688531 828_18.do cx @ 3728 @ 2 @ 1
Use safety gloves.
16
Commissioning

7.1 Excess temperature and low level safety devices

The safety devices are not affected by the control circuit. When they trip all actors are permanently shut down.
The alarm is displayed optically and acoustics with a continuous signal
Pos: 33.2 /Technische Dokumentation/Flash CORIO /Überte mpera tur-, U nt ern ive au-Schutzeinrichtung -Warnung @ 0\mod_1389688599281_18.d oc x @ 3734 @ @ 1
+ +
E 14
Pos: 34 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Füllen - Grundlegende Gef ahren @ 0\mo d_140 921844 9251_18 .doc x @ 6708 @ 2 @ 1
tone and the reason for the alarm is shown on the display as a number.
Danger from damaged safety devices
Possible serious consequences for personnel and working areas.
Check the safety devices at least twice per year.
Excess temperature protector, IEC 61010-2-010
Turn the adjustable excess temperature protector to the cut-out point (actual temperature) using a screwdriver. The actors will be shut down on all poles, the circulator will show error message E 14, the "Alarm" control display will be lit and a continuous signal tone will sound.
Low level safety device, IEC 61010-2-010
The float switch in this device must be moved manually in the bath to test the function, for example using a screwdriver. Push down the float until its reaches the mechanical stop.
The actors will be shut down on all poles, the circulator will show error message E 01, the "Alarm" control display will be lit and a continuous signal tone will sound.
17

7.2 Filling the 1001F

Top up the tempering fluid timely.
Ensure that no tempering fluid gets into the thermostat during filling.
By inserting the material, the tempering fluid is displaced. This should be considered when filling the bath tank.

7.3 Filling quantities and shut-off criteria

Pos: 36 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Pumpeneinstellung @ 0\mod_140921845296 4_18.d ocx @ 6 873 @ 2 @ 4
Pos: 37 /Technische Dokumentation/Flash CORIO/Einschalten/Start-Stopp @ 0\mod_1395238575986_18.doc x @ 42 09 @ 2 @ 1
Filling height without basket
Number of bottles Filling height with
basket an bottles
and bottles
40 mm 5 bottles 20 mm
60 mm 10 bottles 20 mm
78 mm 15 bottles 20 mm
95 mm 20 bottles 20 mm
Fill the bath tank as per table (e.g., with Thermal G). Place the stainless steel basket with bottles into the bath tank. The filling height should now be approx. 20 mm below the edge of the bath tank.
Limit values:
The recommended filling height for water amounts to
• maximum approx. 20 mm below the edge of the bath tank
• minimum approx. 60 mm below the edge of the bath tank
Emptying
• Switch off the unit and pull the plug.
• Place a container under the drain port (24).
• Open the drain cock.
• Close the drain cock again after the tank is completely emptied.
18
Commissioning

7.4 Switching on / Start - Stop

When pressing the keys, it is advisable to hold the circulator head with one hand.
Switching on
• The device is switched on by pressing the power switch (1).
The unit performs a self-test. All the segments of the four digit LED
temperature display will illuminate. The "OFF" signal then indicates that the unit is ready to operate.
OFF
28.30
Pos: 38 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Temp erat ur -Sol lwe rtein st ellu ng @ 0\mod _140 921845 5522_18 .docx @ 6961 @ 2 @ 1

7.5 Setting the desired temperature value

Factory setting: 10 °C
Example: Changing the desired value from
33.00 to 45.50
Start
Press OK for approx. 1 second. The current bath temperature will be shown on LED temperature display.
Stop
Press OK for approx. 1 second.
Switch off the unit with the circulator's mains power switch.
The flashing control indicator "Cooling" after temperature
regulation is complete indicates that a valve is still open. Only turn off the main power switch after this indicator has stopped flashing.
The temperature can be set when the device has been started or stopped.
The set value is saved so that it will be retained even after a
power outage.
1.
Press the edit key or briefly to switch from
displaying the actual value to the desired value. The digits before the decimal point flash.
Change the value:
2. Press the key to set a higher value. Press the key to set a lower value. Press the key briefly for single steps, press and hold the key to adjust the values quickly.
Save the set value as the desired value by pressing the
3.
key. The digits after the decimal point flash.
19
4. Change the value:

Setting the timer

Press the key to set a higher value. Press the key to set a lower value. Press the key briefly for single steps, press and hold the key to adjust the values quickly.
Save the set value as the desired value by pressing the
5. key.
The new desired value will flash three times.
Timer displays
Timer set to 15 minutes.
28 seconds until the timer ends
The timer keeps the set value for a defined period of time. The device then stops (OFF). The timer settings can be set when the device has been started or stopped.
The set value is saved so that it will be retained even after a
power outage.
Setting the timer
1. and simultaneously press “ on the display.
2. Changing the timer value (in minutes): Press the key to set a higher value. Press the key to set a lower value. Press the key briefly for single steps, press and hold the key to adjust the values quickly.
3. Save the set value as the desired value by pressing the
key.
The new timer value will flash three times.
The decimal point flashes until the desired value (±0.1 K) is reached. After the desired value is reached, the timer starts and the display alternates between the actual temperature (for 3 sec.) and the remaining time (for 2 seconds). The remaining time < 1 min is displayed in seconds. Until the desired value is reached, it can still be changed. The timer remains active and will start when the new desired value is reached. If the desired value is changed while the timer is running, the timer will be deactivated.
After the set time elapses, a beep sounds twice and the device goes into the
Stopping the timer
OFF state.
1. Press the key for about 1 sec. -
t” and O. appear
OFF.
Pos: 39 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/ATC - CD @ 1\mod_1432892086111_18. docx @ 8260 @ 2 @ 1
20
Commissioning
If the value is outside a window of Error message:

7.6 ATC - Absolute Temperature Calibration (1-point calibration)

1-point calibration:
The temperature sensor can be calibrated at any value in the working temperature range.
Place a calibrated thermometer (resolution: 0.01°C) in the center of the bath to measure the actual bath temperature.
1. Switch on the circulator at its mains switch
50.00
50
48.90 CAL
.00
2.
Set the required calibration value as described in "Adjusting
temperature setpoint" (example: 50.00°C).
3. Start the unit: Press the key.
4. The bath will be heated to 50.00°C. When the setpoint is reached, allow the bath temperature to settle for approx. 3 minutes.
5. Start the calibration:
Press the “Service” and the key together until the comma in the display starts to flash.
6. Read the bath temperature value on the calibrated thermometer
and and set it. (Example: 48.90 °C)
7. Press the key to confirm.
8. The circulator will confirm the process by displaying "CAL".
Err = Error
±5°C, its entry will be ignored!
- Err
Pos: 40 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Autostart ein / aus @ 0\mod_1409218444899_18.d ocx @ 6 565 @ 2 @ 1
21
1 2
The switchover process is shown briefly on the temperature
Autostart on
Autostart off switched off)

7.7 Switch Autostart on and off

The AUTOSTART function allows the device to be started as soon as the main switch is pressure, which in turn allows you to use a timer.
Uncontrolled device start
If circulators are started using "AUTOSTART", ensure that even if it is started accidentally, for example after a power outage, it does not pose a danger to personnel or equipment.
Ensure that the circulator's safety equipment is set correctly.
. Press and hold the key and
+
. switch on the circulator using the main switch.
display LED.
AOn
AOFF
Pos: 41 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Alarmmeldungen_CD @ 0\mod_14092184426 21_18.d ocx @ 6488 @ 1 @ 1
(With a setpoint change of 50 K, the circulator is also

8 Error messages / Possible causes of faults

The following faults which trigger alarms result in the units's heater
+
E 01
E 05
22
and circulating pump being shut down permanently.
The alarm indicator lights up and a continuous signal tone will sound. The reason for the alarm or warning will be shown on the LED temperature display in coded form.
Warnings are displayed alternately with the actual value on the display.
The signal tone can be muted by pressing the OK key.
The device is being operated with no or too little bath fluid or the level is below the minimum level. Top up the bath fluid.
A hose has burst (bath fluid level too low because it has been pumped out). Replace the hose and top up the bath fluid.
The cable for the working temperature sensor has been interrupted or short-circuited.
Error messages / Possible causes of faults
E 06
E 14
E 33
E 60 E 61
E 63 E 70
Defect of the working or excess temperature sensor. The working and excess temperature sensors report a temperature
difference of more than 20 K.
The cut-out value of the excessive temperature protector is below the defined working temperature . Set the safety temperature to a higher value.
The cable for the overtemperature safety sensor has been broken or short-circuited.
Internal error. Contact JULABO Service Department.
Connection error between Corio CD and refrigeration unit The data communication between the Corio CD and the refrigeration
unit is permanently monitored. If communication cannot be established (e.g. by a defective connection cable), the error message „E 61“ is generated. By pressing the „OK“ key, the alarm is acknowledged. The Corio CD continues operating purely as heating circulator until the next interruption of the power supply (power-off). When the fault has been repaired, the Corio CD controls the refrigeration unit according to the settings in the menu (Off, Auto, On) after the next power-on.
Internal error. Contact JULABO Service Department.
Internal error. Contact JULABO Service Department.
E 72 E 80 E 82
E 83 E108
E116 E118
Internal error. Contact JULABO Service Department.
Warning: Internal error. Contact JULABO Service Department.
Warning: Update error (incorrect hex file). Contact JULABO Service Department.
Warning: Excessive power consumption via USB interface (<300 mA).
The self-locking alarm is still active. Switch off the device at the main switch. Wait for approx. 4 seconds and then switch it on again.
The internal AD converter is defective.
23
E431 E1431
E401 E402 E413 E414 E417 E418 E425
Alarms and warnings for refrigeration machines 200F, 201F, 300F
Maximum compressor current exceeded.
Warning: No compressor current detected.
Alarms and warnings for refrigeration machine 600F, 601F, 900F, 1001F
Temperature sensor evaporator outlet defective (short circuit).
Temperature sensor evaporator outlet defective (break).
Evaporation pressure sensor defective (short circuit).
Evaporation pressure sensor defective (break).
Condensation pressure sensor defective (short circuit).
Condensation pressure sensor defective (break).
Error in refrigeration system.
E426
Error in refrigeration system.
E427
Error in refrigeration system.
E1427
Warning: Error in refrigeration system
E431
Maximum compressor current exceeded.
E432
Error in refrigeration system.
E433
Error in refrigeration system.
E1431
Warning: No compressor current detected.
24
Error messages / Possible causes of faults
Pos: 42 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Betri eb U SB - PC @ 1\mod_14 303928 77254_1 8.doc x @ 8190 @ 1 @ 1
To cancel the alarm state
1. Switch off the device at the main switch.
2. Eliminate the cause of the alarm.
3. Eliminate the cause of the alarm or wait for approx. 4 seconds, depending on error type.
4. Switch on the device again at the mains power switch.
5. If the error occurs again, a remote diagnostic is to be made.
Faults which are not displayed:
Circulating pump motor overload protection. The circulating pump motor is protected from overloads. After a cooling
phase the motor will restart automatically. If necessary the unit should be inspected by a JULABO service technician.
JULABO Technical Service
Phone: (610) 231-0250 option 3 Fax: (610) 231-2060 Email: service.us@julabo.com
25

9 Operation of the CORIO CD via the USB interface

• PC and CORIO CD are switched on.
• Connect the CORIO CD and PC with the USB cable.
The CORIO CD reports to the PC after the installation of the driver with the identifier "STMicroelectronics virtual COM port" COM port in Device Manager
For comfortable operation, you can use the JULABO software EasyTemp. There are also interface commands to the direct query available:
Input Function
STATUS
VERSION
IN_SP_00
IN_PV_00
IN_PV_01
IN_PV_03
IN_PV_04
IN_MODE_05
OUT_SP_00
OUT_MODE_05
current operating status
device name + voltage variant + software version
Setpoint query
current value query
Variable query
Safety sensor query
Safety potentiometer query
Start/Stopp query
Send setpoint
Start/Stop device 1/0
Remote operation
The device can be controlled remotely via the USB interface. To do this, connect the unit to a PC via the interface.
When remote operation is active (press and hold down and simultaneously for approximately 3 seconds) the display alternates between the current bath temperature or “OFF” and the message “-r-”. Switching remote operation on and off is confirmed with a short
r On or “r OFF”.
26
Operation of the CORIO CD via the USB interface
Data logging
Certain characteristic parameters can be read. To do this:
1. Attach the USB flash drive.
2. Start data logging with the and key combination. Hold the keys down until the activation is indicated by LO61. Data logging is also indicated by a flashing dot after the last decimal place (28.33.). Every record is automatically provided with a log file name and stored in a separate txt file.
3. The same key combination can be used to stop data logging. Deactivation is indicated by LO60.
Cold Mode:
You can set the refrigeration unit operating mode. Here you can switch between automatic operation, refrigeration unit always on, and refrigeration always off.
To do this, press the button while switching on the power switch. The mode changes one position each time the power switch is switched on and the button pressed. The display shows:
COn
- „
where the required cooling capacity to maintain the bath temperature is available.
- „
AUtO
where cooling capacity is available if needed
COFF
- „
where no cooling capacity is required.
Reading black box data:
The black box is integrated into the controller and stores all relevant data from the last 20 minutes. In addition, persisting alarms and warnings are logged to an alarm memory. This data can be read.
Installation is easy and is carried out step by step. Please follow the instructions.
To do this, insert the USB flash drive into the thermostat interface and press the “SERVICE button” + . The display then shows
-BB-”.
Refrigeration unit always on,
Automatic operation (factory setting)
refrigeration unit always off,
27
Refer to all regulations for disposing of bath fluids.

10 Emptying the bath tank

Danger of scalds from hot bath fluid or hot drain tap. Please note the following when draining the bath fluid:
Hot bath fluid: Do not drain the bath fluid when it is hot.
Environmental Hazard:
Emptying
Switch off the unit and pull the plug or disconnect the connection to the power supply on all poles. For baths without a drain tap, remove the circulator from the bath tank.
Small bath tanks do not have a drain tap and can be carried for drainging. The temperature of the bath fluid should not exceed 50 °C.
Enclosed baths and refrigeration machines
Connect a suitable hose to the drain port ( 12 mm external).
Route the hose to a vessel or drain.
Open the drain valve with the knurled screw.
To reduce the weight, the bath can be partly emptied using a hose
pump (transfer pump).
Do not empty the bath in temperatures of ≤ 0 °C since the drain
tap may freeze.
Pos: 47 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Technische Daten -Thermostat CD @ 0\mod_14092 184553 04_18.do cx @ 6 950 @ 12 @ 1
28
Technical data
°C
°C
200 V /
50 Hz
200 V /
60 Hz
230 V /
50 Hz
230 V /
60 Hz
kg
°C
A
V / Hz
V / Hz
A
Classification to DIN 12876-1
III (FL)

11 Technical data

11.1 Technical data for circulator

Circulator
Working temperature range
Temperature stability
Temperature setting
Temperature display
Resolution
ATC - Absolute Temperature Calibration
Temperature control
Heating capacity (at 230 V)
Heating capacity (at 115 V)
Heating capacity (at 100 V)
°C
kW
kW
kW
CORIO CD
20 ... 150
±0.03
Digital
LED
0.01
1-point
PID1
2.0
1.0
0.8
Circulating pump:
Delivery rate at 0 bar L/min 12 14 15 17
Pressure at 0 liters bar 0.30 0.33 0.35 0.43
Max. viscosity
Dimensions (WxDxH) without bracket
Useful immersion depth
Weight
Ambient temperature range
Operating temperature range
Mains power connection
Current consumption (230 V)
For CH and GB model (at 230 V)
Mains power connection
Current consumption (at 230 V)
Mains power connection
Current consumption (at 115 V)
Mains power connection
Current consumption (at 100 V)
Pos: 48 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Technische Daten - CD200F-CD600F @ 0\mod_1409218455 054_18.docx @ 6939 @ 2 @ 1
cSt
cm
cm
°C
V / Hz
A
A
V / Hz
A
50
13.2 x 16.0 x 18.4
16.6
2.6
5 ... 40
-40 … 150
230 ±10 % / 50
10
10
230 ±10 % / 60
10
115 ±10 % / 60
10
100 ±10 % / 50-60
9
29
Refrigeration circul. circulator
CD-1001F
Working temperature range
°C
-38 ... 100
Temperature stability
°C
±0.03
Temperature display
LED
Resolution
°C
0.01
ATC – Absolute Temperat. Calibration
1-point
Temperature control
PID1
Refrigeration capacity
°C
20
10 0 -10
-20
-30
(medium ethanol)
kW 1 0.95
0.9
0.63
0.35
0.13
Refrigerant
R452A
Dimensions (WxDxH)
cm
45 x 64 x 95
cm
35 x 41
Bath depth
cm
30
Filling volume, from...to
Liter
42 ... 56
Weight, with circulator
kg
74
Ambient temperature range
°C
5 ... 40
V/ Hz
Current consumption (at 230 V / 50 Hz)
A
Nom. 5 / Tot. 16
at model CH (at 230 V / 50 Hz)
A
Nom. 5 / Tot. 10
at model GB (at 230 V / 50 Hz)
A
Nom. 5 / Tot. 13
Current consumption (at 230 V / 60 Hz)
A
Nom. 5 / Tot. 16
Mains power connection
V/ Hz
200 +10 %; -5 % / 50 / 60
Current consumpt. (at 200 V / 50/60 Hz)
A
Nom. 5 / Tot. 16

11.2 Technical data for refrigeration circulation circulator

Useful bath opening (WxD)
Mains power connection
230 ±5 % / 50 / 60
All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C Technical changes without prior notification reserved.
mgebung @ 0\mod_1409298215054_18 .doc x @ 7311 @ @ 1
Safety precautions to IEC 61010-2-010:
Excess temperature protection, adjustable 0°C ... 170°C Low level protection Float switch Classification to DIN 12876-1 Class III Alarm Optical and audible (permanent)
Ambient conditions to IEC 61010-1:
For indoor use only.
Altitude up to 200 m - normal zero.
Ambient temperature: +5 ... +40 °C
30
Technical data
EMC requirements
The device is an ISM device of group 1 per CISPR 11 (uses HF for internal purposes) and is classified in class A (industrial and commercial sector).
Devices of class A are intended for the use in an industrial electromagnetic environment.
When operating in other electromagnetic environments, their electromagnetic compatibility may be impacted.
Humidity
• Maximum relative humidity 80%, for temperatures up to 31°C,
Linear decrease to 50% relative humidity at a temperature of 40°C
Max. voltage fluctuation of ±10% are permissible.
Protection class to EN 60 529: IP 21 The device complies with Safety class I Overvoltage category II
Pos: 52 /Tech nis ch e Doku m ent atio n/Fl ash CO RIO /COR IO _CD_ GESA MT/ Kä ltem itt el @ 1\mod_ 145752 950509 5_18 .docx @ 9191 @ 2 @ 4
Pollution degree 2

11.3 Refrigerant

In the event of an error in the refrigeration system (leak) a certain room size is specified in standard EN 378 for each kg of refrigerant.
The refrigerant used and the quantity are stated on the type plate.
Refrigerant used in relation to JULABO
Limit value for 1 m3 volume [kg]
Pos: 53 /Technische Dokumentation/Flash CORIO/F-Gas-Verordnung @ 1\mod_1453374700170_18.doc x @ 83 51 @ 3 @ 1
R23 0.68
R134a 0.25
R404A 0.52
R507 0.53
R508B 0.2
R452A 0.423
Propane (R290) 0.008
Ethylene (R1150) 0.007
31
Pos: 54 /Tech nis ch e Doku m ent atio n/Fl ash CO RIO /COR IO _CD_ GESA MT/ Ma teri al, med ium sberü hr ten Te ile, Th er mo sta t @ 0\ mod_14 092184 51950_1 8.doc x @ 6829 @ 12 @ 1
Information about the used refrigerants The Regulation (EU) No. 517/2014 on fluorinated greenhouse
gases applies to all systems which contain fluorinated refrigerants and
replaces (EC) 842/2006. The aim of the Regulation is to protect the environment by reducing
emissions of fluorinated greenhouse gases. Among other things it regulates the emission limits, use and recovery of
these substances. It also contains requirements for operators of systems which require / contain these substances to function.
Under Regulation 517/2014, the operator of a system of this nature has the following duties:
The operator must ensure that the equipment is checked at regular intervals for leaks.
These intervals depend on the CO
equivalent of the system. This is
2
calculated from the refrigerant fill volume and type of refrigerant. The CO
equivalent of your system is shown on the model plate.
2
The operator undertakes to have maintenance, repair, service, recovery and recycling work carried out by certified personnel who have been authorized by JULABO.
All such work must be documented. The operator must keep records and archive them for at least five years. The records must be submitted to the relevant authority on request.
Refer to the text of the Regulation for further information.

12 Materials of parts in contact with the bath fluid

12.1 Circulator

Description Material
Motor 1.4301
Pump PPS
Heater 1.4404 / 316L
Sensor 2xPt 100 metal, fitted 1.4571
Sensor connection 1.4301
Float 1.4401
Float pipe 1.4571
Spring hinge clamp 1.1248, coated
Tubing FPM / FKM
Pos: 55 /Technische Dokumentation/Fl ash COR IO/CO RIO_ CD_GE SAMT/ Zubehö r @ 0\mod_ 142426 746469 1_18.do cx @ 79 00 @ 1 22 @ 1
32
Accessories

13 Accessories

A wide selection of accessories is available for the following products at
www.julabo.com for optimum adaption to your temperature control
task.

13.1 For external connection

• Bath fluids
• Tubing
Shut-off valve
• Barbed fittings
• Adapters

13.2 For open baths

Temperature applications for samples, preparation of samples for serology and clinical chemistry, analysis, etc.
Test tube racks
Pos: 56 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Warten, Reinigen, Lagern @ 0\mod_1409218462355_1 8.doc x @ 7225 @ 1221 1 @ 1
Immersion-height adjustable platforms

14 Maintenance, cleaning, storage

Danger of injury during maintenance, repair and transport Danger from mains voltage.
Have all service and repair work carried out by authorized specialists only.
Switch off the unit and pull the plug,
- before starting any cleaning work,
- before carrying out any service or repair work or
- before moving the unit.
Empty the unit completely before moving it.
Transport the unit carefully.

14.1 Maintain the refrigeration capacity.

The device is designed for continuous operation in normal conditions. No regular maintenance work is required.
The condenser on the front should be cleaned from time to time to maintain the full refrigeration capacity.
33

14.2 Cleaning

14.3 Storage

1. Switch off the device.
2. Pull the plug.
3. Let the unit cool down to room temperature.
4. Carefully extract the contamination at the condenser through the ventilation grille at the front. Make sure that the fins are not damaged.
Use low surface tension water (for example soap suds) to clean the bath and the functional parts of the circulator which are immersed in it. Clean the exterior device with a cloth and low surface tension water.
The circulator is designed for continuous use in normal conditions. No regular maintenance work is required.
The bath tank should only be filled with suitable bath fluid. In the event of contamination, the bath fluid must be replaced from time to time.
Units which are not to be reused must be stored in a dry, place, protected from dust and frost, after cleaning. The system components must be fully emptied and carefully dried, for example using compressed air. Seal the connectors.

15 Repair service

Before asking for a service technician or returning a JULABO unit for repair, please contact our Technical Service Department.
JULABO Technical Service Phone: (610) 231-0250 option 3 Fax: (610) 231-2060 Email: service.us@julabo.com If you return a unit to JULABO:
Clean the unit to avoid any harm to the service personnel.
• It is essential that you enclose a short fault description.
Before returning the device, please complete an online return form
at http://www.julabo.com/us/support/us-rma.
• Ensure careful and adequate packing..
JULABO cannot accept any liability for damage caused by incorrect
packaging.
In the interest of product improvement, JULABO reserves the right to make any necessary technical modifications during the repair to
Pos: 57 /Technische Dokumentation/Flash CORIO/CORIO_CD_GESAMT/Gewährleistung EU @ 0\mod_1409218450062_18.docx @ 6741 @ 2 @ 1
34
ensure the proper functioning of the unit.
Repair service

15.1 Warranty

The following Warranty Provisions shall apply to products sold in North America by Julabo (“Seller”) to the entity shown as buyer (“Buyer”) on Seller’s invoice.
1. Initial Warranty. Upon Seller’s receipt of payment in full for the products and subject to
Buyer’s compliance with the terms of sale and any other agreement with Seller relating to the products, Seller warrants to the Buyer that the products manufactured by the Seller are free from defects in material and workmanship for a period not to exceed two (2) years or ten thousand (10,000) hours of operation, whichever comes first, from the date the product is shipped by Seller to Buyer (the “Initial Warranty”).
2. EXCLUSION OF ALL OTHER EXPRESS WARRANTIES;
EXCLUSION OF ALL IMPLIED WARRANTIES. OTHER THAN THE INITIAL WARRANTY, NO OTHER EXPRESS WARRANTIES ARE MADE. ALL IMPLIED WARRANTIES OF EVERY TYPE AND KIND, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR USE, ARE EXCLUDED IN ALL RESPECTS AND FOR ALL PURPOSES. SELLER DISCLAIMS AND MAKES NO IMPLIED WARRANTIES WHATSOEVER.
3. Exclusions. The Initial Warranty does not include damage to the product resulting
from accident, misuse, improper installation or operation, unauthorized or improper repair, replacement or alteration (including but not limited to repairs, replacements, or alterations made or performed by persons other than Seller’s employees or authorized representatives), failure to provide or use of improper maintenance, unreasonable use or abuse of the product, or failure to follow written installation or operating instructions. Buyer must return the product’s record of purchase to the Seller or one of Seller’s authorized representatives within thirty (30) days of the date the product is shipped by Seller to Buyer in order to make a claim under the Initial Warranty. Notwithstanding anything contained herein to the contrary, all glassware, including but not limited to reference thermometers, are expressly excluded from the Initial Warranty.
4. Buyer’s sole remedies; Limitations on Seller’s Liability. Buyer’s sole and exclusive remedy under the Initial Warranty is strictly limited, in Seller’s sole discretion, to either: (i) repairing
35
defective parts; or (ii) replacing defective parts. In either case, the warranty period for the product receiving a repaired or replaced part pursuant to the terms of the Initial Warranty shall not be extended. All repairs or replacements performed by Seller pursuant to these Warranty Provisions shall be performed at Seller’s facility in Allentown, Pennsylvania, U.S.A. or at the facility of an authorized representative of Seller, which location shall be determined by Seller in its sole discretion; provided, however, that Seller may, in its sole discretion perform such repairs or replacements at Buyer’s facility in which case Buyer shall pay Seller’s travel, living and related expenses incurred by Seller in performing the repairs or replacements at Buyer’s facility. As a condition precedent to Seller’s obligation to repair or replace a product part under the Initial Warranty, Buyer shall (i) promptly notify Seller in writing of any such defect; (ii) shall have returned the product’s record of purchase to Seller or to one of Seller’s authorized representatives within thirty (30) days of the date the product is delivered to Buyer; and (iii) assist Seller in all respects in its attempts to determine the legitimacy and basis of any claims made by or on behalf of Buyer including but not limited to providing Seller with access to the product to check operating conditions. If Buyer does not provide such written notice to Seller within the Initial Warranty period or fails to return the product’s record of purchase as set forth above, Seller shall have no further liability or obligation to Buyer therefore. In no event shall Seller’s liability under the Initial Warranty exceed the original purchase price of the product which is the subject of the alleged defect.
5. THE REMEDIES PROVIDED IN THE INITIAL
WARRANTY ARE THE SOLE AND EXCLUSIVE REMEDIES AVAILABLE TO THE BUYER. NOTWITHSTANDING ANYTHING TO THE CONTRARY CONTAINED HEREIN, AND EVEN IF THE SOLE AND EXCLUSIVE REMEDIES FAIL OF THEIR ESSENTIAL PURPOSE FOR ANY REASON WHATSOEVER, IN NO EVENT SHALL SELLER BE LIABLE FOR BUYER’S MANUFACTURING COSTS, LOST PROFITS, GOODWILL, OR ANY OTHER SPECIAL, INDIRECT, PUNITIVE, INCIDENTAL OR CONSEQUENTIAL DAMAGES TO BUYER OR ANY THIRD PARTY AND ALL SUCH DAMAGES ARE HEREBY DISCLAIMED.
6. Assignment. Buyer shall not assign any of its rights or
obligations hereunder without the prior written approval of Seller; provided, however, that if Buyer is a distributor of Seller, the rights and obligations of Buyer under these Warranty Provisions shall inure to the benefit of and be binding upon Buyer’s customers who provide the product’s proof of purchase to Seller pursuant to the terms set forth herein. Seller may assign any or all of its rights or obligations hereunder without Buyer’s prior consent.
36
Repair service
7. Governing Law. The Warranty Provisions and all questions relating to their validity,
interpretation, performance, and enforcement shall be construed in accordance with, and shall be governed by, the substantive laws of the Commonwealth of Pennsylvania without regard to its principles of conflicts of law.
8. Waiver. Any failure of the part of Seller to insist on strict compliance with the
Warranty Provisions shall no way constitute a waiver of such right. No claim or rights arising out of a breach of the Warranty Provisions by Buyer may be discharged in whole or in part by a waiver of the claim or right, unless the waiver is in writing signed by an authorized representative of Seller. Seller’s waiver or acceptance of any breach by Buyer of any provisions of the Warranty Provisions shall not constitute a waiver of or an excuse for nonperformance as to any other provision of the Warranty Provisions nor as to any prior or subsequent breach of the same provision.
9. Freight. Buyer will arrange and pay for shipping and handling charges for the
unit to be returned to the Seller. Seller will arrange and pay for shipping and handling for the return of the unit to the Buyer.
37

16 Waste disposal

16.1 Packaging

Packaging materials must be disposed of as prescribed by the current local regulations.

16.2 Unit

In the European Economic Area (EEA) the disposal of waste equipment is regulated in the "Directive of the European Parliament and of
the Council on Waste Electrical and Electronic Equipment (WEEE)". The current official journal on this matter is available on the
European Parliament's homepage. The symbol for the separate collection of electrical and electronic
equipment is a crossed-out trash can. Disposal with household waste (unsorted waste) or similar collections of
municipal waste is not permitted! Contact an authorized waste disposal contractor in your country.

16.3 Refrigerant

Refrigerants must be disposed of as prescribed by the current local regulations. They may only be disposed of by trained personnel.
38
JULABO USA, Inc.
884 Marcon Boulevard Allentown, PA 18109 Phone: +1(610) 231-0250 Fax: +1(610) 231-0260 info.us@julabo.com
www.julabo.com
Changes without prior notification reserved. 09.04.2018
Loading...