Joy-It JT-DMSO2D72 User guide

Page 1
JT-DMSO2D72
Portable 3-in-one oscilloscope, signal generator and multimeter
1. GENERAL INFORMATION
Dear customer, Thank you for choosing our product. Please read the following im-
portant information before putting the device into operation. This oscilloscope is compact, portable and flexible. It has a TFT-LCD
backlight brightness. The maximum real time sampling rate can be up to 250 MSa/s. USB
memory devices are also supported, the user can upgrade the firm­ware via USB, the edge trigger function is automatically detected and it has a user-defined fast oset calibration with A/D converter that can sample each channel simultaneously.
3
Page 2
2. SAFETY INSTRUCTIONS
Safety Terms and Symbols:
following terms may appear on product: Danger It represents that harms may be caused to you at once if you per-
form the operation.
Warning It represents that latent harms may be caused to you if you per-
form the operation.
Notice It represents that damage may possibly caused to the product or
other properties if you perform the operation.
Symbols on the product: following symbols may appear on product:
General Safety Summary
Read the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To evade potential haz­ards, use this product only as specified.
Only qualified personnel should perform maintenance.
Use suitable power cord. Use only the power cord specified for this prod-
uct and certified for the country of use. Connect and disconnect properly. Connect a probe with the oscilloscope
before it is connected to measured circuits; disconnect the probe from the oscilloscope aer it is disconnected from measured circuits.
Ground the product. This product is grounded through the grounding
conductor of the power cord. To avoid electric shock, the grounding con-
ductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Notice
Please read
the manual
Protective
ground
terminal
Measuring
ground
terminal
Chassis
ground
terminal
Page 3
Connect the probe the right way. The probe ground lead is at ground po-
tential. Do not connect the ground lead to an elevated voltage. Check all terminal ratings. To avoid fire or shock hazard, check all ratings
and markings on the product. Refer to the product manual for detailed in­formation about ratings before making connections to the product.
Do not operate without covers. Do not operate this product with covers or panels removed.
Avoid exposed circuitry. Do not touch exposed connections and compo-
nents when power is present. Do not operate with suspected failures. If you suspect there is damage to
this product, have it inspected by qualified service personnel.
Assure good ventilation. Do not operate in wet/damp environments. Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
3. TECHNICAL SPECIFICATIONS
Channels 2 Bandwidth 70 MHz Sampling Rate 250 MSa/s Single channel
125 MSa/s Dual channel
Oscillospe Specifications: -Horizontal
Rising time ≤ 5 ns
Waveform Interpolation (sin x)/x
Record length 6000 samples Single channel
3000 samples Dual channel
SEC/DIV Range of 1, 2, 5 Sequence 5 ns/div - 500s/div
Oscilloscope Specifications: -Vertical
A/D converter 8-bit resolution
VOLTS/DIV range 10mV/div –10V/div at BNC input
Measurement Range ± 5 div Rising time at BNC input, typical ≤ 5 ns
Page 4
Selectable analog bandwidth limit,
typical
Typical 20 MHz (reduced to 6 MHz when
using a 1x probe)
Low frequency response (-3db) ≤ 10 Hz
DC gain accuracy 10V/div - 10mV/div ±3% for normal or
Trigger
Type Edge
Mode Auto, Normal, Single
Level ± 4 divisions from centre of screen
Trigger level accuracy 0,2 div*volts/div within 4 div.
Slope Rising, falling, rising & falling
Source CH1, CH2
Input
Coupling DC, AC or GND
Input impedance (DC coupled)
25pF ±3pF, 1MΩ ±2%
Probe attenuation 1X, 10X
Supported probe attenuation factors 1X, 10X, 100X, 1000X
Input protection voltage 150 V
RMS
Measurement
Values between cursors Voltage Dierence: △V
Time Dierence: △T
Automatic measurement Frequency, Amplitude
General specifications
Display type 2.6 inch 64K color TFT-LCD
Display resolution 320 horizontal x 240 vertical pixel
Display contrast Adjustable
Power supply voltage 100-240V AC, 50-60Hz OR 5V DC, 2A
Power consumption < 2.5 W
Fuse T, 3 A
Battery 2 x 2600 mAh
Operating temperature 0– 50°C Storage temperature -20-60°C
Page 5
Allowed relative humidity ≤ 40°C: ≤90% RH, 40-50°C: ≤60% RH
Cooling Method Convection
Altitude:
-Operating and Nonoperating 3 000 m (10 000 feet)
-Random vibration 0.31 g
RMS
from 50 - 500Hz, 10 minutes on each
axis
-Nonoperating 2,46 g
RMS
from 5 - 500 Hz, 10 minutes on each
axis
Mechanical shock in operation 50 g, 11 ms, half sine
Dimensions 199 x 98 x 40 mm Weight 624 g
Arbitrary Waveform Generator:
Waveform Frequency Sine: 1Hz-25MHz; Square: 1Hz-10MHz;
Ramp: 1Hz-1MHz; Expon: 1Hz-5MHz
Sampling rate 250 MSa/s
Amplitude 2.5 Vpp (50Ω)/5Vpp (High impedance)
Frequency resolution 0,1 %
Channel CH1 waveform output
Waveform depth 512Sa
Vertical resolution 12 bit
Output impedance 50 Ω
Square Characteristics:
Pulse width accuracy ± 10 ns
Digital Multimeter:
Maximum resolution 4000 Counts
Testing Modes: Voltage, Current, Resistance, Capacity, Diode,
On-O
Maximum Input Voltage AC: 600 V/ DC: 600 V
Maximum Input Current AC: 10 A / DC: 10 A
Page 6
Range Accuracy Resolution
DC Voltage
400.00mV
±(0.8% + 5)
100uV
4.000V 1mV
40.00V 10mV
400.0V 100mV
600.0V
±(1% + 2)
1V
Overload protection: 400mV: 250V, other: 600Vrms.
AC Voltage
4.000V
±(1.2% + 5)
1mV
40.00V 10mV
400.0V 100mV
600.0V
±(1.5% + 5)
1V
Frequency: 40Hz~400Hz; Frequency of 400V and 600V: 40Hz~100Hz
DC Current
40.00mA
±(1% + 2)
10uA
200.0mA
±(1.5% + 2)
100uA
4.000A
±(1.8% + 2)
1mA
10.00A
±(3% + 2)
10mA
Overload protection: self restoring fuse: 200mA/250V, 4A and 10A range no fuse.
AC Current
40.00mA
±(1.3% + 2)
10uA
400.0mA
±(1.8% + 2)
100uA
4.000A
±(2% + 3)
1mA
10.00A
±(3% + 5)
10mA
Frequency: 40Hz~400Hz; self restoring fuse: 200mA/250V, 4A and 10A range no fuse.
Resistance
400.0Ω
±(1% + 3)
0.1Ω
4.000KΩ
±(1.2% + 5)
1Ω
60.00KΩ 10Ω
400.0KΩ 100Ω
4.000MΩ 1KΩ
40.00MΩ
±(1.5%±3)
10KΩ
Overload protection: 220Vrms
Capacity
40.00nF
±(3% + 5)
10pF
400.0nF 100pF
4.000uF 1nF
40.00uF 10nF
100.0uF 100nF Overload protection: 220Vrms
Diode
0V~1.0V
On-O
<50Ω
Page 7
All specifications herein mentioned apply to the series oscilloscopes. Before checking an oscilloscope to see if it complies with these specifications, make sure it meets the following conditions:
The oscilloscope must have been operating continuously for twenty minutes un­der the specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the oper­ating temperature changes by more than 5°C.
The oscilloscope must be within the factory calibration interval. All specifications are guaranteed unless noted ‘typical’.
This Oscilloscope is a small, lightweight portable instrument which provides the user with a convenient and easy to operate front panel with which you can per­form basic tests.
• General Inspection:
Please check the instrument as following steps aer receiving: Check the shipping container for damage: Keep the damaged shipping contain-
er or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electri­cally.
Check the accessories: Accessories supplied with the instrument are listed in "Accessories" in this manual. If the contents are incomplete or damaged, please notify the franchiser.
Check the instrument: In case there is any mechanical damage or defect, or the instrument does not operate properly or fails performance tests, please notify your reseller.
• Use of safety keyhole:
A safety keyhole is reserved on the back shell of the oscilloscope. Users need to purchase the safety lock by themselves. Wrap one end of the safety lock around a hard-to-move object, insert the other end into the safety lock hole, turn the key clockwise to lock the instrument, and then pull out the key. In this way,
the most basic anti-the requirements can be achieved.
4. GETTING STARTED
Page 8
Adjust the bracket
When using the instrument, you can open the support foot to tilt the instrument upward for easy operation and observation. When the instrument is not in use, the user can close the support foot to facilitate placement or han­dling. Aer adjusting the rack, the instrument can be suspended on the vertical plane.
• Front Panel
The following diagram briefly describes the front panel of this oscilloscope.
Page 9
• The user interface
• Functional Check
Follow the steps below to perform a quick functional check to your oscilloscope.
Power
Press the power key and the device starts. Press the power key again, and the de-
vice will shut down. Before starting it, please confirm that the battery has enough
power. The oscilloscope is supplied with a power adapter with USB-Type-C plug. The input AC power supply is 100~240V, 50~60Hz. The output is 5V@2A. The power adapter can be used to supply the oscilloscope or charge the battery.
When connect the power adapter to the oscilloscope, if the battery is not installed inside the oscilloscope, the power key backlight is red and flash; if the battery is installed inside the oscilloscope and the battery power is not filled, the power key backlight is red; if the battery is installed inside the oscilloscope and the battery power is full, the backlight of the source key is extinguishes.
Observe the waveform
1. Set the switch on the probe to 1X and connect the probe to Channel 1 on the
oscilloscope. First, align the slot in the probe connector with the protuberance on
the CH1 BNC and push to connect; then, turn to right to lock the probe in place.
2. If you use the probe hook-tip, the hook end should be removed, the probe pin should be inserted into the Gen Out output terminal, and the probe grounding clamp should be clamped on the metal outer ring of the Gen Out output terminal. Recommended input ~2V@1KHz peak-peak square wave.
Page 10
3. Press the Auto button and you should see within a few seconds a square wave
of about 2V peak-to-peak at 1KHz in the display. Repeat the steps to observe CH2.
• Probe check
Safety
When using the probe, keep your fingers behind the guard on the probe body to avoid electric shock. Do not touch metallic portions of the probe head while it is connected to a voltage source. Connect the probe to the oscilloscope and connect the ground terminal to
ground before you start any measurements.
Manual Probe Compensation
Upon the first connection of a probe and an input channel, you should manually perform this adjustment to match the probe to the input channel. Uncompensated or wrong compensated probes may lead to errors or faults in measurement. To adjust the probe compensation, follow the steps below.
1. Press Channel button to enter channel setting menu, Set the Probe option at­tenuation in the channel menu to 10X. Set the switch on the probe to 10X and con­nect the probe to Channel 1 on the oscilloscope. If you use the probe hook-tip, the hook end should be removed, the probe pin should be inserted into the Gen Out output terminal, and the probe grounding clamp should be clamped on the metal outer ring of the Gen Out output terminal. The oscilloscope with signal generator function needs to set output signal as 2V @ 1KHz square wave; The Gen Out termi­nal of the oscilloscope without signal generator function automatically outputs 2V @ 1KHz square wave. Press the Auto button.
2. Check the shape of the displayed waveform.
3. If necessary, use a nonmetallic screwdriver to adjust the variable capacity of your probe until the shape of the waveform turns to be the same as the above figure. Repeat this step
as necessary. See the figure below for the way
of adjustment. Probe Attenuation Setting: Probes are of
various attenuation factors which aect the vertical scale of the signal. The Probe Check function is used to verify if the Probe attenua­tion option matches the attenuation of the probe.
Compensated correctly
Overcompensated
Undercompensated
Page 11
You can press Channel button to enter the channel setting menu and select CH1, and select the Probe option that matches the attenuation factor of your probe.
Make sure that the Attenuation switch on the probe matches the Probe option in the oscillo­scope. Switch settings are 1X and 10X.
When the Attenuation switch is set to 1X, the probe limits the bandwidth of the oscilloscope to 6MHz. To use the full bandwidth of the oscil­loscope, be sure to set the switch to 10X.
Menu and Control Keys
All the keys are described as follows:
Scope: Oscilloscope mode DMM:Multimeter mode
AWG:Waveform generator Menu:Function menu
Trig:Trigger setting menu
Enter: In scope, it saves the user-
defined settings of the oscilloscope; In generator, press the button to
confirm aer entering the character. Auto: It adjusts the horizontal and
vertical scales of the oscilloscope automatically and sets the trigger coupling, type, position, slope, level and mode, etc., to acquire a stable waveform display.
Channel: Channel setting menu Time: Horizontal setting menu
Zoom and move keys:
In the trigger menu, the le and down keys push the trigger level down, and the right and upper direction keys push the trigger level up;
In the channel menu, the upper and lower direction keys change the zero level posi­tion of the channel, and the le and right direction keys change the volt/div of the
channel;
In the time base menu, the upper and lower direction keys change the time/div, and the le and right direction keys change the horizontal trigger position;
In DMM, switch measurement function; In generator, aer choosing a parameter, the le and lower direction key will reduce
the parameter value, the right and up direction key will increase the parameter val­ue; it is also used for the digital selection of the virtual keyboard.
5. FUNCTION INTRODUCTION
Page 12
F1/F2/F3/F4: Multi function key, in each menu mode, is responsible for selecting
corresponding menu items in the screen.
• Connectors
Gen Out: The Gen Out terminal of the oscilloscope without signal genera-
tor function always outputs 2V @ 1KHz square wave.
Shortcut keys. Long press this button to enter the menu, and choose shortcut key function; Aer setting, press this button sin­gle time to respond to the corresponding function.
In scope, stop or run the waveform acquisition; In DMM, hold the measuring data or update data;
In generator, turn on or o the waveform output.
Power key.
Page 13
• Automatic set
Auto set is one of the advantages digital oscilloscopes have. When you push the Auto button, the oscilloscope will identify the type of waveform (sine or square wave) and adjust controls according to input signals so that it can accurately display the wave­form of the input signal.
The Auto function examines all channels for signals and displays corresponding waveforms. Auto determines the trigger source according to the following conditions :
If multiply channels get signals, the oscilloscope will use the channel with the lowest frequen­cy signal as the trigger source.
If no signals are found, the oscil­loscope will use the lowest-
numbered channel displayed in Auto Scale as the trigger source.
If no signals are found and no channels are displayed, the oscilloscope will display
and use Channel 1 as the trigger source.
• Default settings
Functions Settings Cursor O Display Format Set to YT Horizontal Position Adjusted
SEC/DIV Adjusted
Trigger Level Set to 50% Trigger Mode Auto Trigger Source Adjusted Trigger Slope Adjusted Trigger Type Edge Vertical Bandwidth Full Vertical Coupling Unchanged VOLTS/DIV Adjusted
Menu or Sys­tem
Option, Button or Knob
Default Setting
Cursor
Type Off Source CH1
Horizontal (amplitude) ±4div
Vertical (time) ±4div
Display Format YT
Horizontal
Position 0.00s SEC/DIV 500μs
Measure On or Off Off
Trigger (Edge)
Source CH1 Slope Rising
Mode Auto Level 0.00v
Vertical Sys­tem,
All Channels
Bandwidth Limit Unlimited Coupling AC Probe Attenuation 1X Position 0.00div (0.00V) VOLTS/DIV 1V
Page 14
Long press button, to enter Short key setting menu, and select Default.
Press button, pops up the prompt to recall the default settings on the screen, now press F1 to confirm. The oscilloscope will display the CH1 waveform and remove all the others. Press F4 to cancel. The table below gives the options, buttons and con­trols that change settings at default setup. The following settings do not change when you recall default settings: Language options, saved settings, saved waveforms, saved reference waveforms, calibration data.
• Horizontal System
Press Time button to enter horizontal system menu, use the direction keys to change the horizontal scale (time base) and the horizontal trigger position. When you change the horizontal scale, the waveform will expand or contract to the screen center.
1. SEC/DIV Knob: Used to change the horizontal time scale to magnify or compress the waveform horizontally. If the waveform acquisition is stopped (by using button),
press Time button and Up or Down buttons to expand or compress the waveform.
2. Horizontal Position Knob: Used to control the trigger position against the screen center. Press Time button and Right or Le buttons to move the waveform right or le. The key resolution vary according to time base. Press "AUTO" key can make the horizontal position return to zero.
3. Mode: Y-T, X-Y, Roll, Scan. Y-T: YT format shows the vertical voltage in relation to time (horizontal scale). Press
Time->Mode to set. X-Y: The XY mode is used to analyze phase dierences, such as those represented by
Lissajous patterns. The format plots the voltage on CH1 against the voltage on CH2, where CH1 is the horizontal axis and CH2 is the vertical axis. You may view the same waveform in XY mode. To perform this operation, stop the acquisition and press Time
->Mode to change the display mode to X-Y. Roll: In Roll mode, the waveform display rolls from right to le. Trigger or horizontal
oset control of waveforms is available during Roll Mode, and it’s only available when set to 100ms/div or slower. Press Time->Mode to change the display mode to Roll, the
time/div will be changed to 100ms/div automatically. Scan: In Scan Mode, the waveform display scan updates from le to right. In Scan
mode, trigger and horizontal oset control of waveforms are available during Scan
Mode. This mode is only available when set to 100ms/div or slower. Generally, used
for measuring low frequency signal. When the time/div is 100ms/div or slower, the oscilloscope automatically enters the scan mode.
Page 15
Vertical System
Vertical system can be used to adjust the vertical scale and location and other set­tings of the channel. Each channel has a separate vertical menu, and each channel can be set separately.
1.Vertical Position
Press Channel->F1 to select the channel, and press Up or Down direction keys to move the vertical position of the selected channel.
2. VOLTS/DIV Settings
The range of Volt/div is 10mV/div-10V/div (1X), or 100mV/div-100V/div (10X), 1V/div­1000V/div (100X), step by 1-2-5. Press Channel->F1 to select the channel, and press Right or Le direction keys to change the Volt/div of the selected channel.
3. Channel Setting Menu
• Trigger System
The trigger determines when the oscilloscope begins to acquire data and display a waveform. Once a trigger is properly set up, the oscilloscope can convert unstable
displays or blank screens to meaningful waveforms.
The triggering mode of this series oscilloscope is edge trigger. Edge trigger distinguishes the trigger points by seeking the specified edge (rising, falling, rising & falling) and trigger level. Press Trig button to enter trigger menu.
Trigger Source: Select the trigger source signal to CH1 or CH2. You can use the trig­ger source options to select the signal that the oscilloscope uses as a trigger.
Options Settings Comments
On/O
On O
Turn on the waveform display. Turn o the waveform display.
Coupling
DC
AC
GND
DC passes both DC and AC components of the input signal.
AC blocks the DC component of the input signal and attenu­ates signals below 10Hz.
GND disconnects the input signal.
Probe
1X
10X
100X
1000X
Selects a value according to the probe attenuation factor so as to ensure correct vertical readouts. Reduce bandwidth to 6MHz when using a 1X probe.
BW
Limit
On O
Limits the bandwidth to reduce display noise; filters the signal to eliminate noise and other unnecessary HF compo-
nents.
Invert
On O
The invert function turns the displayed waveform 180 de­grees, with respect to the ground level. When the oscillo­scope is triggered on the inverted signal, the trigger is also inverted.
Page 16
Slope: Select the trigger slope to rising, falling, rising &
falling. Trigger Mode: You can select the Auto or Normal mode to define how the oscil­loscope acquires data when it does not detect a trigger condition. Auto Mode per­forms the acquisition freely in absence of valid trigger. It allows the generation of untriggered waveforms with the time base set to 100ms/div or slower. Normal Mode updates the displayed waveforms only when the oscilloscope detects a valid trigger
condition. Before this update, the oscilloscope still displays the old waveforms. This
mode shall be used when you only want to view the eectively triggered wave­forms. In this mode, the oscilloscope displays waveforms only aer the first trigger. To perform single trigger acquisition, the trigger mode can be set to "single". When triggered, a single waveform is acquired and then stop.
Force Trigger: Used to complete an acquisition regardless of an adequate trigger signal. This button becomes useless if the acquisition is already stopped.
Trigger Level: It sets the amplitude level the signal must cross to cause an acquisiti­on when using the Edge or Pulse Width trigger. Press Trig button to enter, and press Up or Down direction keys to change the trigger level.
• Save Waveforms
Press Menu button to enter, and select Save to enter store waveform menu. The waveform can be saved to the oscilloscope inside, also can be recalled to view.
The Save Menu Table:
To save the waveform, follow these steps:
1. Press F1 to select the storage position of the waveform.
2. Press F2 to save the waveform data to the specified position. Also, users can long
press button to enter, and select F4->F2 (Data Save) as shortcut keys. Press button one time to save the current waveform data.
3. Press F3, to recall the saved waveform. The waveform can be enlarged or reduced
in the Channel or Time menu by using the direction keys.
Attention:
1. Waveform storage can not only save the waveform of the current channel, but
also save the current state settings at the same time.
2. Users can permanently store 6 waveforms in the oscilloscope memory and rewrite them at any time.
Menu Setting Description
Position 1,2,3,4,5,6 Select the internal storage position. Save Save waveform data. Recall
Recall the waveform.
Page 17
Reference Waveform
The REF channel is used to display the reference waveform, which can compare the actual waveforms with the reference waveforms to find out the dierences.
Press Menu button to enter, and select Ref to enter reference Waveform menu. REF Menu Table:
To save the reference waveform, follow these steps: Press F1 to select the position of the reference waveform. Press F2 to open the REF channel. Press F3 to select the source to CH1 (or CH2), and only select the open channel. Press F4 to save the current waveform to the specified location.
• Measurement
Scale measurement
Graticule: This method allows you to make a quick, visual estimate and take a simp-
le measurement through the graticule divisions and the scale factor. For example, you can take simple measurements by counting the major and minor
graticule divisions involved and multiplying by the scale factor. If you counted 6 ma­jor vertical graticule divisions between the minimum and maximum values of a wave­form and knew you had a scale factor of 50mV/division, you could easily calculate
your peak-to-peak voltage as follows: 6 divisions x 50mV/division = 300mV.
Cursor measurement
The cursor measurement has two parallel lines on the screen, and move two lines to measure the time and voltage parameters of the input signal. The result of cursor measurement will be displayed on the second page of the cursor menu. Before use cursor measurement, make sure that the measured source is the signal you need to measure.
Press Menu button to enter, and select Cursor to enter cursor measurement.
Menu Setting Description
Position
Ref-A
The REF waveform is saved to the oscilloscope Ref-A or Ref-B.
Enable
On
O
Open the REF waveform.
Close the REF waveform.
Source
CH1 CH2
Select CH1 to save as a REF waveform. Select CH2 to save as a REF waveform
Save Save the REF waveform.
Menu Setting Description
Enable
On
O
Open the cursor measurement. Close the cursor measurement.
Type
Voltage Time
The horizontal line is shown to measure the voltage parame­ters. The vertical line is shown to measure the time parameters.
Source
CH1 CH2
Select the measured source.
Cursor1
Select Cursor1 and press up, down, le and right keys to move the Cursor1 position, and display the Cursor1 value.
Cursor2
Select Cursor2 and press up, down, le and right keys to move the Cursor2 position, and display the Cursor2 value.
Incre-
ment
The dierence between the Cursor1 and the Cursor2.
Page 18
To do cursor measurement, follow these steps:
1. Press F1 to open the cursor measurement;
2. Press F2 to select a type of cursor measurement.
3. Press F3 to select the channel that needs to be measured.
4. Press F4 to enter the second page, press F1 or F2 to select Cursor1 or Cursor2, press
up, down, le and right to move Cursor1 or Cursor1;
5. The result of cursor measurement will be displayed on the cursor menu. Automatic Measurement
The oscilloscope provides 2 kinds of automatic measurements, including frequency
and amplitude.
Press Menu button to enter, select Measure to enter automatic measurement. Press F1 to open, the measurement results will be displayed in the top le corner of the screen.
Attention: In the measurement result, the yellow font is the result of CH1, and the green font is
the result of CH2.
• Utility
Press Menu Button to enter.
Menu Description
Frequency
Measure the frequency of the signal.
Maximum Voltage
Measure the maximum voltage of the signal.
Minimum Voltage
Measure the minimum voltage of the signal.
Menu Setting Description
Language
German English
Set the menu language.
Sound
On
O
Open the key sound. Close the key sound.
Backlight luminance
1~10 Set backlight brightness of the screen.
Backlight time
30s 60s 90s 120s Unlimited
Set the screen backlight time.
System infor-
mation
Display system information, such as soware version or PCB version.
Automatic shutdown
5 Minute 10 Minute 20 Minute 30 Minute Unlimited
Set automatic shutdown time.
Calibration
Start Return
Start the self calibration. Exit the self calibration.
Page 19
Note:
1. Backlight time and automatic shutdown time will not be executed when the os­cilloscope is plugged in with an external charging device or connected to a comput­er via a USB cable.
2. Shutdown automatically saves last setup. Self calibration
The self calibration routine helps optimize the oscilloscope signal path for maxi­mum measurement accuracy. You can run the routine at any time but should al­ways run it if the ambient temperature changes by 5°C or more. For a more accu­rate calibration, please power on the oscilloscope and wait for 20 minutes until it has adequately warmed up. To do self calibration, follow these steps:
1. Ensure that no input signal is input, otherwise it may damage the instrument.
2. Press Menu button to select the self calibration function.
Shortcut key
Long press button to enter shortcut key settings menu, and choose shortcut key function; Aer setting, press this button single time to respond to the corre­sponding function.
Default settings: Long press button to enter, and select F1 (Default) as the shortcut key, then press button one time and select F1 TO confirm to recall
the default settings Custom settings: Long press button to enter, and select F2 or F3 (custom 1 or
2) as shortcut keys. Now press Enter button and click F1 to confirm to save the cur­rent settings to custom settings. Press button one time and click F1 to con-
firm to recall the custom settings. Save data: Long press button to enter, and select F4->F1 (Data Save) as
shortcut keys. Press Menu -> Save to enter the save menu, select a location, press
button one time to save the current waveform data.
Interface
This chapter introduces the multimeter function.
The measuring types
include DC voltage (V,
mV), AC voltage, DC cur­rent (A, mA), AC current (A, mA), resistance, ca­pacitance, diode, and on
-o test.
6. DIGITAL MULTIMETER
Page 20
Measurement
1. DC and AC voltage measurement
Press the power button to turn on, then press the "DMM" button to enter to the multi meter function interface; Press the up, down, le and right direction keys or F1, F2, F3, F4 multi-function keys to select "DC V", "DC mV" or "AC V"; Insert the black pen into the input port of the COM banana socket, and insert the red pen into the input port of the V/Ω/C banana port; Connect the red and black forms to the measured point. The voltage value of the measured point will be displayed on the screen.
2. DC and AC current measurement
Press the power button to turn on, then press the "DMM" button to enter to the multi meter function interface; To measure the DC current greater than 200mA, press the up, down, le and right keys or F1, F2, F3, F4 multi function keys to select "DC A" or "AC A", and insert the black pen to the input end of the COM banana socket, and insert the red pen into the input port of the “A” banana socket; To measure the DC current less than 200mA, press the up, down, le and right keys or F1, F2, F3, F4 multi function keys to select "DC mA" or "AC mA", and insert the black pen to the input end of the COM banana socket, and insert the red pen into the input port of the “mA” banana socket.;
Connect the red and black forms to the measured point. The current value of the
measured point will be displayed on the screen.
3. Resistance measurement Press the power button to turn on, then press the "DMM" button to enter to the multi meter function interface; Press the up, down, le and right direction keys or F1, F2, F3, F4 multi-function keys to select "OHM"; Insert the black pen into the input port of the COM banana socket, and insert the red pen into the input port of the V/Ω/C banana port; Connect the red and black forms to the measured point. The resistance value of the
measured point will be displayed on the screen.
4. Capacity measurement Press the power button to turn on, then press the "DMM" button to enter to the multi meter function interface; Press the up, down, le and right direction keys or F1, F2, F3, F4 multi-function keys to select " "; Insert the black pen into the input port of the COM banana socket, and insert the red pen into the input port of the V/Ω/C banana port; Connect the red and black forms to the measured point. The capacitance value of the measured point will be displayed on the screen.
5. Diode measurement Press the power button to turn on, then press the "DMM" button to enter to the multi meter function interface; Press the up, down, le and right direction keys or F1, F2, F3, F4 multi-function keys to select " "; Insert the black pen into the input port of the COM banana socket, and insert the red pen into the input port of the V/Ω/C banana port; Connect the red and black forms to the measured point. The diode value of the mea-
Page 21
sured point will be displayed on the screen.
6. Buzzer measurement Press the power button to turn on, then press the "DMM" button to enter to the multi meter function interface; Press the up, down, le and right direction keys or F1, F2, F3, F4 multi-function keys to select "Buzzer"; Insert the black pen into the input port of the COM banana socket, and insert the red pen into the input port of the V/Ω/C banana port; Connect the red and black forms to the measured point. If the resistance of the mea­sured point is less than 50 ohms, the instrument will emit "drop" sound.
7. Data hold function Press the button on the instrument, and the displayed data will hold on the dis­play, even if the input signal changes or eliminates, the value will not change.
Note: Please select the required measuring gear correctly and then measure again.
This chapter introduces the function of wave­form generator.
Operation description
Press the power button to turn on, and then press the "AWG" button
to enter the waveform
generator function in­terface.
1. Set type Press F1 button to select the desired signal waveform, the optional waveform type includes square, triangle, sine, trapezoidal and four arbitrary.
2. Set frequency
Press F2 button to select Frequency, then use the up, down, le and right direction keys to adjust the frequency, press F2 button to open the digital keyboard again, use the up, down, le, right direction keys and Enter key to set frequency parameter, select "OK" and press Enter button to confirm.
3. Set amplitude Press F3 button to select Amplitude, then use the up, down, le and right direction
keys to adjust the frequency, press F3 button to open the digital keyboard again, use the up, down, le, right direction keys and Enter key to set frequency parameter, select "OK" and press Enter button to confirm.
7. SIGNAL GENERATOR
Page 22
Note that the entered voltage is the half of the peak-to-peak voltage.
4. Set oset Press F4 button to enter the second page. Press F2 to select Oset, then use the up, down, le and right direction keys to adjust the frequency, press F2 button to open the digital keyboard again, use the up, down, le, right direction keys and "Enter" key to set frequency parameter, select "OK" and press Enter button to confirm.
5. Set duty cycle Enter the second page. Press F3 to select Duty, then use the up, down, le and right direction keys to adjust the frequency, press F3 button to open the digital keyboard again, use the up, down, le, right direction keys and Enter key to set frequency pa­rameter, select "OK" and press Enter button to confirm.
6. Generate an arbitrary waveform It is necessary to edit the arbitrary waveform in connection with the soware and download it into the machine. There are 4 arbitrary wave positions, and each positi­on can permanently store an arbitrary wave.
7. Aer setting the waveform parameters, press the button to turn on or o the
signal output. The output waveform of the signal generator can be observed through
an oscilloscope.
• Output the sine waveform
Output a sine waveform with 10KHz/2.5Vpp as follows:
1.Press AWG button to enter the waveform generator function interface.
2.Press F1 to select “Sine”
3. Frequency: First press F2 button to select Frequency, then use the up, down, le
and right direction keys to adjust the frequency. Second, press F2 button to open the digital keyboard again, use the up, down, le, right direction keys and "Enter" key to set frequency parameter, select "OK" and press Enter button to confirm. Set the frequency to 10KHz
4. Amplitude: First press F3 button to select Amplitude, then use the up, down, le and right direction keys to adjust the amplitude. Second, press F3 button to open the digital keyboard again, use the up, down, le, right direction keys and Enter key to set amplitude parameter, select "OK" and press Enter button to confirm. Set the amplitude to 2.5V
5. Oset: Press F4 to enter the second page. First press F1 button to select O­set, then use the up, down, le and right direction keys to adjust the oset.
Second, press F1 button to open the digital keyboard again, use the up, down, le, right direction keys and "Enter" key to set oset parameter, select "OK"
Page 23
and press Enter button to confirm.
6. Press button, the backlight of the button turns green, i.e. output sine waveform.
7. The waveform obser­ved by an oscilloscope is as follows:
• Output the arb waveform
1. Install the soware
Download the latest soware on the oicial website, double click Setup.exe to install. The link is as follows:
https://joy-it.net/files/files/Produkte/JT-DMSO2D72/DMSO2D72_Soware.zip
2. Install the driver Connect the oscilloscope to the computer via the USB cable. Open the device manager of the computer and find the device. If the device icon is shown in the following pic-
ture, you need to install the driver manually.
The drivers are located in the soware package in the Driver folder. Right click the device, and select “Update the driver soware”, and select the driver
files path to install. Aer successful installation of the driver, it is shown as follows:
Page 24
3.Double-click the Hantek2xx2 icon to open the soware and select "DDS" in the
right control bar to enter the signal generator control bar.
4. Put "√" in the box in front of "on/o" to open the signal output.
5. Select "signal type" as "arb", and set the corresponding frequency and amplitude.
6. Select "Arb Channel" as Arb1/Arb2/Arb3/Arb4. Each arb channel can save only one
arbitrary waveform which was downloaded last time. Turn on again aer shutdown, and automatically recall.
7. Draw arbitrary waves in waveform area with mouse.
8. Connect Gen Out connector to oscilloscope for observation.
Page 25
When the battery on the screen is displayed as blank, it indicates that the battery is about to run out. When the battery power is too low, the oscilloscope will prompt “Power o aer 5s”. In order to avoid the automatic shutdown of the oscilloscope due to insuicient power supply, please charge it in time.
If the power button is pressed, the oscilloscope will not react, indicating that the battery power may be exhausted. You can charge the oscilloscope in the following way:
Charge the oscilloscope through charger: Connect the oscilloscope to the power socket through the USB data line and charger distributed by the instrument for charging. Charge the oscilloscope through the USB interface: Connect the oscilloscope to a computer or other equipment through the USB data line. When charging, turn on the oscilloscope and the battery frame on the screen will change. When the battery is full, the oscilloscope will automatically stop charging.
• Storage and Replacement of the battery
Storage
Lithium batteries must be stored in clean, dry and ventilated rooms. Contact with corrosive substances should be avoided and the have to be kept away from fire and heat sources. If the lithium battery is not used for a long time (for example, more than 6 months), it should be charged with 50%-70% electricity, and removed from the instrument, and stored in a dry and cool environment. If the lithium battery rusts, leaks, bulges and other phenomena, it should be remo­ved immediately and scrapped.
Replacement
Batteries can be recharged repeatedly, but they are vulnerable to wear and tear. If standby time is found to be greatly re­duced, batteries need to be replaced. The battery specification is 18650 lithium battery, 3.7V, 2600mA. Please refer to the following to replace it. Unfold the supporting leg, there are two screws. Remove the screws and remove the battery cover, you will see the batteries.
Now you can remove the battery and re-
place. Attention: Pay attention to the positive and negative poles of batteries when repla­cing baterries.
8. CHARGE
Page 26
1. If the oscilloscope does not start up at power on, follow these steps:
Check whether the battery is installed and confirm whether the battery level is enough. If the battery level is not enough, use the power adapter to charge. Restart the instrument aer the battery level is enough.
Contact your local distributor or directly keep touch with Joy-IT Support depart-
ment if the oscilloscope still can not be turned on normally.
2. If there is no display of waveforms on the screen when the oscilloscope is turned on, follow these steps:
Check the probe to assure its proper connection to the input BNC. Check the channel switch (Channel button) to make sure it has been turned on. Check the input signal to verify it has been connected to the probe correctly. Airm that all measured circuits have signals to output. Turn up the magnitude for DC signals with large magnitude.
In addition, you may press the Auto button to perform an automatic detection of signals at first.
Contact Joy-IT Technical Support department in time if there is still no display of waveforms.
3. If the waveform of the input signal is distorted seriously, follow these steps:
Check the probe to assure its proper connection to the channel BNC. Check the probe to assure its good connection to the measured object. Check the probe to verify it has been well calibrated. Otherwise, refer to the content about calibration described in this manual.
4. If the waveform is rolling continuously on the screen but can not be trigge­red, follow these steps:
Check the trigger source to make sure it consistent with the input channel. Check the trigger level to assure its correct adjustment.
You may press TRIGGER button to enter trigger menu, and use Up, Down, Right, Le direction keys to adjust trigger level to return to signal.
9. TROUBLESHOOTING
Page 27
General Care
Do not put or leave the device in a place where the LCD display will be exposed to direct sunlight for long periods of time.
Note: To avoid damage to the oscilloscope or probes, do not expose them to sprays, liquids, or solvents.
Cleaning
Examine the oscilloscope and probes as oen as operating conditions require. To
clean the exterior surface, perform the following steps: Use a lint-free cloth to remove floating dust on the outside of the oscilloscope and
probes. Take care to avoid scratching the display. Use a so cloth dampened with water to clean the oscilloscope. For more eicient
cleaning, you may use an aqueous solution of 75% isopropyl alcohol. Note: To avoid damage to the surface of the oscilloscope or probes, do not use any
corrosive or chemical cleaning agents.
All specifications herein mentioned apply to the series oscilloscopes. Before check­ing an oscilloscope to see if it complies with these specifications, make sure it meets the following conditions: The oscilloscope must have been operating continuously for twenty minutes under
the specified operating temperature. The Do Self Cal operation must be performed through the Utility menu if the operat-
ing temperature changes by more than 5°C. The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted ‘typical’.
All the following accessories are available by contacting your local Distributor.
• 1x Passive probe oscilloscope (1:1) / (1:10)
• 1x Power adapter
• 1x USB-A-to-USB-C cable
• 1x BNC-to-Crocodile clip cable
• 2x Test prod cable multimeter
10. GENERAL CARE AND CLEANING
11. DISCLAIMER OF TECHNICAL SPECIFICATIONS
12. ACCESSORIES
Page 28
www.joy-it.net
Pascalstr. 8 47506 Neukirchen-Vluyn
14. SUPPORT
If there are still any issues pending or problems arising aer your purchase, we are available for you by e-mail, telephone and with our ticket support system.
E-Mail: [email protected] Ticket system: http://support.joy-it.net Telephone: +49 (0)2845 98469-66 (10-17 o‘clock)
For further information please visit our website:
www.joy-it.net
13. PRODUCT RECYCLING
Our informaon and take-back obligaons according to the Electrical and Electronic Equipment Act (ElektroG) Symbol on electrical and electronic equipment:
2
This crossed-out dustbin means that electrical and electronic appliances do not belong in the household waste. You must return the old appliances to a collec­on point. Before handing over waste baeries and accumulators that are not enclosed by waste equipment must be separated from it.
Return opons: As an end user, you can return your old device (which essenally performs the same funcon as the new device purchased from us) for disposal
free of charge when you purchase a new device. Small appliances with no exter-
nal dimensions greater than 25 cm can be disposed of in normal household quanes independently of the purchase of a new appliance.
Possibility of return at our company locaon during opening hours: Simac GmbH, Pascalstr. 8, D-47506 Neukirchen-Vluyn, Germany
Possibility of return in your area: We will send you a parcel stamp with which you can return the device to us free of charge. Please contact us by e-mail at Service@joy-it.net or by telephone.
Informaon on packaging: Please pack your old appliance safely for transport, if you do not have suitable packaging material or do not wish to use your own, please contact us and we will send you suitable packaging.
Loading...