Automatic Levels
Model No. 40-6926, 40-6960, 40-6962
Instruction Manual
Congratulations on your choice of this Automatic Level. We suggest you read this instruction manual thoroughly before using the instrument. Save this instruction manual for future use.
©2011 Johnson Level & Tool - Rev. 1 1
Page 2
4579H_Manuals 4/4/11 8:37 AM Page 2
Table of Contents
1. Kit Contents
2. Features and Functions
3. Location of Parts/Components
4. Operating Instructions
5. Calibration
6. Technical Specifications
7. Care and Handling
8. Product Warranty
9. Product Registration
10. Accessories
1. Kit Contents
Description for Model 40-6926 Qty.
26X Automatic Level 1 Instruction Manual with Warranty Card 1 Hard Shell Carrying Case 1
Description for Model 40-6960
30X Automatic Level 1 Instruction Manual with Warranty Card 1 Hard Shell Carrying Case 1
Description for Model 40-6962
32X Automatic Level 1 Instruction Manual with Warranty Card 1 Hard Shell Carrying Case 1
Qty.
Qty.
Page 3
4579H_Manuals 4/4/11 8:37 AM Page 3
2. Features and Functions
• Water resistant, sealed construction for use in most weather conditions
• Wire-hung, magnetically dampened compensator
• Large, easy-to-use focusing knob
• 360º horizontal circle
• Penta prism for bubble viewing
• Top-mounted optical sight for quick reference
• Fine adjustment knobs on left and right sides with endless horizontal drive
• 1:100 stadia for distance estimation
• 5/8” x 11 threads to fit standard tripods
©2011 Johnson Level & Tool - Rev. 1 3
Page 4
4579H_Manuals 4/4/11 8:37 AM Page 4
3. Location of Part/Components
Circular bubble vial observation prism
Circular bubble vial
Adjusting screws for circular bubble vial
Compensator check button
Optical sight
Base plate
Eyepiece cover
Eyepiece focusing knob
Horizontal circle
Leveling screw
Horizontal circle reference mark
Focusing knob
Objective lens
Horizontal tangent knob
Page 5
4579H_Manuals 4/4/11 8:37 AM Page 5
4. Operating Instructions
IMPORTANT: It is the responsibility of the user to verify the
calibration of the instrument before each use.
Set-Up
Set the tripod up level and in a comfortable position to view through the telescope. Connect the instrument to the tripod with the 5/8”-11 central screw. Using the leveling screws, center the circular bubble.
Aiming and Focusing
Focus the cross hairs by turning the eyepiece focusing knob until the cross hairs are sharp and black.
Focus the telescope by locating a target, such as a leveling rod, using the optical sight. Looking through the eyepiece, use the focusing knob to bring the target into sharp focus. Center the cross hair to the target using the horizontal tangent knobs on either side of the instrument.
©2011 Johnson Level & Tool - Rev. 1 5
Page 6
4579H_Manuals 4/4/11 8:37 AM Page 6
Reading measurements using a Leveling Rod
Height reading Read the rod where it is intersected by the horizontal hair. For example, the height reading in Figure 1 is 2.0 ft.
Fig. 1
Distance Measurement
Read the rod where it is intersected by the upper and lower stadia hairs. In Figure 1, these readings are at 1.9 ft. and 2.1 ft. The stadia ration is 1:100, therefore the distance from the instrument to the rod is: (2.1-1.9) x 100 = 20ft.
Angle Measurement
As shown in figure 2, sight point A and note the reading on the horizontal circle. Rotate the level and sight point B. The angle turned is the difference between A and B. The angle A0B = xº = A - B.
Fig. 2
Page 7
4579H_Manuals 4/4/11 8:37 AM Page 7
5. Calibration
5.1 Compensator Check Button
Check the compensator for proper operation before use or anytime the operation of the instrument is in question. Push and release the compensator check button to shake the compensator. The compen­sator should return to the exact horizontal position sighted before the check button was pressed.
5.2 Circular Bubble Vial
Center the vial bubble using the leveling screws, then rotate the instrument 180º. The bubble should remain centered (Fig. 3). If the bubble moved from the center, the vial needs adjustment (Fig. 4).
Fig. 3 Fig. 4
Turn the leveling screws to bring the bubble halfway to center (Fig. 5). Using the allen wrench, turn the two vial adjustment screws to center the bubble (Fig. 6).
Repeat the above procedure until the bubble remain centered when the level is rotated 180º.
Fig. 5
©2011 Johnson Level & Tool - Rev. 1 7
Fig. 6
Page 8
4579H_Manuals 4/4/11 8:37 AM Page 8
5.3 Line-of-Sight
The line-of-sight needs to be horizontal within 3mm of level to be accurate.
Set up and level the instrument on a tripod midway between two leveling rods set approximately 100 ft. to 160 ft. (30m to 50m) apart. Sight rods A and B; the height readings are a1 and b1 (Fig. 7). The value “H” is equal to (a1 - b1). Move the instrument to within 6 ft. (2m) of rod A and re-level. Again sight rods A and B; these height readings are a2 and b2 (Fig. 8).
Fig. 7
Fig. 8
If a1 - b1 = a2 - b2 = H, the line-of sight is horizontal. If not, the level should be adjusted as follows.
Because the instrument is set halfway between A and B, any error in the line-of-sight causes both readings to be erroneous by the same amount. Error e cancels out, so the value a1 - b1 = H is correct. Therefore, a2 - H = b3, the adjusting value.
Fig. 9
To adjust, unscrew the eyepiece cover. Turn the adjusting screw with the adjusting pin (Fig. 9) until the horizontal cross hair gives the reading b3 on rod B. Repeat the above procedure until {(a1 - b1) - (a2 - b2)}
3mm.
Page 9
4579H_Manuals 4/4/11 8:37 AM Page 9
6. Technical Specifications
Telescope Erect Magnification 26X (40-6926), 30X (40-6960),
Leveling accuracy ± 1/16"/200 ft. (±1.5mm/160m) Working range Up to 350’ (100m) (40-6926)
Clear objective aperture 40 mm Field of view 1º 30’ Shortest focusing distance 0.0315” (0.8m) Stadia ratio 100 Stadia addition 0 Water resistant Yes Sensitivity of bubble 8’ (2mm) Circle graduation 1º Standard deviation for 1km double-run leveling 0.0689” (1.75mm)
Compensator: Working range ± 15’ Setting accuracy ± 0.8” Dimensions 7.520” x 5.197” x 5.394”
Weight 3.395 lbs (1.54Kg) Center screw thread 5/8” - 11
32X (40-6962)
Up to 400’ (120m) (40-6960) Up to 450’ (130m) (40-6962)
(191 x 132 x 137mm)
©2011 Johnson Level & Tool - Rev. 1 9
Page 10
4579H_Manuals 4/4/11 8:37 AM Page 10
7. Care and Handling
Care must be taken to maintain the accuracy of the instrument.
After each use, the instrument should be wiped clean and kept in
its carrying case.
• Remove dust from the lenses with a soft brush or a nonabrasive wipe. Never touch the lenses with your fingers.
• Store the instrument in a dust-free area with low humidity.
• A bag of silica gel dryer is included with each instrument.
8. Product Warranty
Johnson Level & Tool offers a three year limited warranty on each of its products. You can obtain a copy of the limited warranty for a Johnson Level & Tool product by contacting Johnson Level & Tool's Customer Service Department, as provided below, or by visiting our web site at www.johnsonlevel.com. The limited warranty for each product contains various limitations and exclusions.
Do not return this product to the store/retailer or place of purchase. Non-warranty repairs and course calibration must be done by an authorized Johnson warranty, if applicable, will be void and there will be NO WARRANTY. Contact one of our service centers for all non-warranty repairs. A list of service centers can be found on our web site at www.johnsonlevel.com or by calling our Customer Service Department. Contact our Customer Service Department for Return Material Authorization (RMA) for warranty repairs (manufacturing defects only). Proof of purchase is required.
®
service center or Johnson Level & Tool's limited
10 ©2011 Johnson Level & Tool - Rev. 1
Page 11
4579H_Manuals 4/4/11 8:37 AM Page 11
NOTE: The user is responsible for the proper use and care of the product. It is the responsibility of the user to verify the calibration of the instrument before each use.
For further assistance, or if you experience problems with this product that are not addressed in this instruction manual, please contact our Customer Service Dept.
In the U.S., contact Johnson Level & Tool’s Customer Service Department at 888-9-LEVELS.
In Canada, contact Johnson Level & Tool’s Customer Service Department at 800-346-6682.
9. Product Registration
Enclosed with this instruction manual you will find a warranty registration card to be completed for your product. You will need to locate the serial number for your product that is located on the bottom of the unit. PLEASE NOTE THAT IN ADDITION TO ANY OTHER
LIMITATIONS OR CONDITIONS OF JOHNSON LEVEL & TOOL'S LIMITED WARRANTY, JOHNSON LEVEL & TOOL MUST HAVE RECEIVED YOUR PROPERLY COMPLETED WARRANTY CARD AND PROOF OF PURCHASE WITHIN 30 DAYS OF YOUR PURCHASE OF THE PRODUCT OR ANY LIMITED WARRANTY THAT MAY APPLY SHALL NOT APPLY AND THERE SHALL BE NO WARRANTY.
©2011 Johnson Level & Tool - Rev. 1 11
Page 12
4579H_Manuals 4/4/11 8:37 AM Page 12
10. Accessories
Johnson®accessories are available for purchase through authorized
®
Johnson applicable limited warranty and there will be NO WARRANTY. If you need any assistance in locating any accessories, please contact our Customer Service Department.
In the U.S., contact Johnson Level & Tool’s Customer Service Department at 888-9-LEVELS.
In Canada, contact Johnson Level & Tool’s Customer Service Department at 800-346-6682.
dealers. Use of non-Johnson®accessories will void any