Johnson & Johnson DePuy Synthes Battery Power Line II, BPL II Instructions For Use Manual

Instructions for Use
Battery Power Line II (BPL II).
Battery driven power tool system for
orthopedics and traumatology.
This publication is not intended for distribution in the USA.
Content
Introduction
Operating Instructions
Care and Maintenance
General Information Drive Units
Battery Pack Battery Reamer/Drill II Attachments for Battery Reamer/Drill II
Battery Oscillator II Battery Reciprocator II
General Information Preparation prior to Cleaning Manual Cleaning Instructions
Automated Cleaning Instructions with Manual Pre-cleaning
Lubrication Inspection and Function Test Packaging, Sterilization and Storage
3
8
14
20
21
35
37
39
40
41
44
48
54
55
Troubleshooting
System Specifications
Electromagnetic Compatibility
Ordering Information
Warning
This description alone does not provide sufficient background for the direct use of the product. Instruction by a surgeon experienced in handling this product is highly recommended.
Repairs and Technical Service Disposal of Waste
Accompanying Documents
58
59
60
63
67
71
Battery Power Line II Instructions for Use DePuy Synthes 1
2 DePuy Synthes Battery Power Line II Instructions for Use
Introduction
General Information

Intended use

The Battery Power Line II is a heavy-duty battery­driven system intended for orthopedic and trauma applications including:
Drilling Reaming Inserting / removing Kirschner wires and pins – Cutting bone

Battery Reamer/Drill II

Drilling
Reaming
Kirschner wire insertion

Battery Oscillator II

Oscillating sawing
Fixing of cutting block with a pin

Battery Reciprocator II

Reciprocating sawing
Battery Power Line II Instructions for Use DePuy Synthes 3
Introduction General Information

Safety instructions

The Battery Power Line II (BPL II) is only to be used for patient treatment after careful consultation of the instructions for use. It is recommended that an alter­native system is available to use during application, as technical problems can never be completely ruled out.
The Battery Power Line II is designed for use only by physicians and trained medical personnel.
DO NOT use any component if damage is apparent.
DO NOT use any component if the packaging is damaged.
DO NOT use this equipment in the presence of oxy­gen, nitrous oxide or a mixture consisting of flamma­ble anesthetic and air.
To ensure proper operation of the tool, only use Syn­thes original accessories.
Before first and every use and prior returning for ser­vice power tools and their accessories / attachments, excluding the battery, have to be run through the complete reprocessing procedure. Protective covers and films must be fully removed before sterilization.
For the tool to function properly, DePuy Synthes rec­ommends that it is cleaned and serviced after each use in accordance with the process recommended in the “Care and Maintenance” section. Compliance with these specifications can considerably extend the service life of the tool. Only use Synthes Special Oil (519.970) to lubricate the tool.
We recommend using new Synthes cutting tools for every surgical procedure. Efficiently working cutting tools are the basis for successful surgery. Therefore, check used cutting tools after every use for wear and/or damage and replace them if necessary. Cutting tools must be cooled with irrigation fluid to prevent heat necrosis.
The user of the product is responsible for proper use of the equipment during surgery.
Check proper operation of the tool before using it on the patient.
To prevent overheating, always respect the specified duty cycles on page 64.
If the Battery Power Line II is used in conjunction with an implant system make sure to consult the cor­responding Technique Guide.
For important information regarding electromagnetic compatibility (EMC) please refer to the “Electromag­netic Compatibility” section in this manual.
The tool is classified as type BF against electrical shock and leakage current. The tool is suitable for use on patients in accordance with IEC 60601-1.
To ensure proper operation of the tool, annual maintenance by a DePuy Synthes service center is necessary. The manufacturer shall assume no responsibility for damage resulting from im­proper operation, neglected or unauthorized maintenance of the tool.
4 DePuy Synthes Battery Power Line II Instructions for Use
Notes:
Always wear personal protective equipment
(PPE) including safety goggles when handling with the BPL II system.
To avoid injuries, the locking mechanism of the
tool has to be activated before every manipula­tion and when laying the tool down, i.e. the mode switch has to be in the off position.
Only place the tool in an upright position when
changing attachments or cutting tools intraopera­tively. The handpiece must be laid on its side when not in use in order to avoid the risk of be­ing dropped or contaminating other instruments.
If the machine has been dropped it should be
checked carefully for damage. In the event that any damage is visible, do not use it anymore and send it to the DePuy Synthes service center.
The tool may only be operated with a fully
charged battery. Therefore, ensure that the bat­tery is charged in good time. We recommend that the battery is placed into the charger imme­diately after surgery.
The batteries must never be sterilized, washed,
rinsed or dropped. This would destroy the battery with possible secondary damage (explosion hazard!). Only use original Synthes batteries. Further information can be found on page 18ff.
Never place the BPL II in a magnetic environ-
ment since the machine might start unintention­ally.
Battery Power Line II Instructions for Use DePuy Synthes 5
Introduction General Information

Accessories/scope of delivery

The Battery Power Line II consists of three different handpieces, a battery casing, a battery and a range of attachments designed for the system.
To charge the batteries only use the corresponding Synthes Universal Battery Charger II (05.001.204).
To reach the specified performance only Synthes cutting tools should be used. These are optimized to meet the specific requirements of the tool. Non Synthes saw blades can considerably reduce the lifetime of the system
Special auxiliaries such as cleaning brushes (516.101) and Synthes Special Oil (519.970) are available for cleaning and servicing the system.
No oils from other manufacturers may be used. Only Synthes Special Oil (519.970) must be used to lubri­cate the power tools and attachments.
Lubricants with other compositions may cause jam­ming, have a toxic effect or have a negative impact on the sterilization results. Only lubricate the power tool, the battery casing and the attachments when clean.
DePuy Synthes recommends the use of the specifi­cally designed Synthes Washing Baskets (68.001.620, 68.001.625) or Synthes Vario Case (689.202) to sterilize and store the system. Further­more, the Washing Baskets (68.001.620,
68.001.625) can be used for the automated cleaning procedure. Further information can be found on page 46ff.
Please refer to the “Ordering Information” section on page 71ff for an overview of the components of the system.

Storage and transport

Only use the transport as otherwise damage may occur. If this is not possible, please contact your local DePuy Syn­thes office.
For storage and transport environmental conditions please refer to the “Packaging, Sterilization and Storage” section on page 55ff.
original packaging for dispatch and

Warranty

The wa cover damage of any kind resulting from wear, im­proper use, damaged seal, use of non Synthes cut­ting tools and lubricants or improper storage and transport.
The manufacturer shall assume no responsibility for damage resulting from improper operation, neglected or unauthorized maintenance of the tool.
For further information on the warranty please con­tact your local DePuy Synthes office.
rranty for the tools and accessories does not
6 DePuy Synthes Battery Power Line II Instructions for Use

Explanation of symbols used

The following symbols are applied to the device or individual components.
Caution
Consult the provided instructions for use before operating the device.
The device is classified as type BF against electrical shock and leakage current. The device is suitable for use on patients according to the standards defined by IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) ANSI/AAMI ES60601-1:2005, Dated December 2005 and CAN/CSA-C22.2 No. 60601-1:08, Dated February 2008 Do not immerse device in liquids.
Accredited test house for IEC 60601
10PB
0123
Li-Ion
S9 IP X4
The device meets the requirements of directive 93/42/EEC for medical de­vices. It is authorized by an independ­ent notified body for which it bears the CE symbol.
The European Battery Directive 2006/66/EC applies to this device. See section “Disposal” on page 60. This device contains Lithium-Ion batteries that should be disposed of in accordance with environment protection requirements. This symbol indicates that this battery contains rechargeable Lithium-Ion cells. The battery is rechargeable.
Duty cycle type according to IEC 60034-1. Ingress protection rating according to IEC 60529. Locked symbol. Drive Unit is off for safety
Date of manufacture
Non sterile
Temperature
Relative humidity
Atmospheric pressure
Do not use if package is damaged
Battery Power Line II Instructions for Use DePuy Synthes 7
Introduction
Drive Units

Battery Reamer/Drill II (530.705)

Speed (without attachment)
Torque (without attachment)
Weight of handpiece (including battery pack) Cannulation Ø 4.0 mm Protection against electric shock Protection against water ingress Cleaning brush (516.101) and Synthes Special Oil (519.970) included
0 – 340 rpm (maximum speed varies with attachment) 0 –15 Nm (maximum torque varies with attachment) 1565 g / 3.4 lbs
BF
IP X4
Technical data is subject to tolerances.
8 DePuy Synthes Battery Power Line II Instructions for Use

Battery Oscillator II (530.710)

Speed 0 – 12’000 oscillations per
minute
Deflection 4.5° (0° +/ - 2.25°) Weight of handpiece (including battery pack) Protection against electric shock Protection against wa­ter ingress Synthes Special Oil (519.970) included
1685 g / 3.7 lbs
BF
IP X4
Technical data is subject to tolerances.
Battery Power Line II Instructions for Use DePuy Synthes 9
Introduction Drive Units

Battery Reciprocator II (530.715)

Speed 0 –14’000 oscillations
per minute Stroke 4 mm Weight of handpiece (including battery pack) Protection against electric shock Protection against water ingress Synthes Special Oil (519.970) included
1675 g / 3.6 lbs
BF
IP X4
Technical data is subject to tolerances.
10 DePuy Synthes Battery Power Line II Instructions for Use

Battery for Battery Power Line II

Art. no. Type Li-Ion (Lithium Ion) Voltage 14.8 V Capacity 1.5 Ah / 22.2 Wh Charging Time typically < 60 minutes
Technical data is subject to tolerances.
Note:
For further information on the correct method of charging, storing and using the battery, please refer to page 18ff.
530.630
Battery Power Line II Instructions for Use DePuy Synthes 11
Introduction Drive Units

Compatibility between BPL and BPL II batteries

Existing BP BPL II batteries
The existing
530.615) can be used with the new BPL II battery (530.630), battery casing (530.690) and sterile cover (530.660) as seen in Fig. 1.
Existing BPL batteries are compatible with BPL II handpieces
The existing BPL battery (530.620), battery casing (530.680) and sterile cover (530.650) can be used with the new BPL II handpieces (530.705, 530.710,
530.715) as seen in Fig. 2.
Note:
Technical data is subject to tolerances and may vary when combining the two systems.
L handpieces are compatible with
BPL handpieces (530.605, 530.610,
Figure 1
Figure 2
530.605
530.690
530.705
530.680
12 DePuy Synthes Battery Power Line II Instructions for Use
Introduction
Universal Battery Charger II
The Universal Battery Charger II (05.001.204) in­clud
es four independent charging bays. Each charg­ing bay has three slots; the Battery Power Line II battery (530.630) fits into the top slot.
Note:
In order that the BPL II battery can be recognized and charged by the UBC II, firmware version 14.0* is required. If necessary, send the charger to a DePuy Synthes representative for a firmware update.
*Sticker as shown on the underside of the charger with firmware version 14.0:
SW-Rev. 14.0
For further information on the Universal Battery Charger II please consult the relevant Instructions for Use (036.000.500) or contact your local DePuy Syn­thes office.
The BPL II battery cannot be charged with the Uni­versal Battery Charger (item number 530.600 or
530.601).
1 Charging bays (4) 2 Symbols for battery type 3 ON/OFF display 4 Control display for each charging bay 5 Ventilation holes 6 Ventilation holes 7 Power switch 8 Fuses: 2 x 5 AT / 250 V 9 Power cord connection 10 Symbol for BPL I and II batteries (530.620 or
2013/01/09
530.630)
10
Battery Power Line II Instructions for Use DePuy Synthes 13
Operating Instructions
Battery Pack (battery casing with inserted battery)
Synthes non-sterile batteries and advanced charging technology optimize intraoperative battery capacity, maximize battery lifespan and shorten turnaround time. One Universal Battery Charger II (05.001.204) for multiple Synthes battery-driven systems simplifies the charging process. Simple aseptic technique pre­serves the sterile field when assembling the battery pack.
Instruments
530.630 Battery for Battery Power Line II
530.660 Sterile Cover for Battery Power Line II
530.690 Battery Casing for Battery Power Line II

Assembling and inserting the battery pack

Scrubbed p
Open the lid of the battery casing as shown in Fig. 1. Ensure the lid of the battery casing is fully open (Fig. 2). Ensure that the lid of the battery casing is facing towards the scrubbed person (Fig. 3). Position the sterile cover securely on top of the bat­tery casing (Fig. 4).
Notes:
The sterile cover helps to guide the battery into
the battery casing and prevents contamination of the sterile casing by the non-sterile battery.
Sterilize the sterile cover after each use to en-
sure aseptic conditions when inserting the un­sterile battery into the sterile battery casing.
If the non-sterile battery contacts the outside of
the battery casing, the battery casing must be cleaned and resterilized before being used in the operating room.
Do not insert the non-sterile battery into the bat-
tery casing whilst a handpiece is attached.
erson
Figure 1
Figure 2
Figure 3
Scrubbed person
530.690
Scrubbed person
Scrubbed person
Scrubbed person
530.660
14 DePuy Synthes Battery Power Line II Instructions for Use
Figure 4
530.690
Circulating person
Insert the non-sterile battery through the sterile cover into the battery casing (Fig. 5a). Press down on the battery to ensure it is fully seated (Fig. 5b).
Note:
The shape of the battery ensures that it is inserted with the correct pole alignment. The circulating per­son must not touch the outside of the battery casing.
Remove the sterile cover from the battery casing (Fig. 6).
Precaution:
Avoid all contact with the outside of the battery casing in order not to contaminate it. Should the non­sterile battery or the circulating person’s hand come into contact with the outside of the battery casing, it must be cleaned and resterilized before being used in the operating room.
Circulating person
Figure 5a
Circulating person
530.630
530.660
530.690
Scrubbed person
Scrubbed person
Figure 5b
Circulating person
Figure 6
Battery Power Line II Instructions for Use DePuy Synthes 15
Scrubbed person
Operating Instructions Battery Pack (battery casing with inserted battery)
Scrubbed person
Close the battery casing (Figs. 7a and 7b).
Note:
Ensure that both battery casing locks engage and that the lid of the battery casing is closed properly. Always ensure that the lid of the battery casing is totally closed before using the system.
Precaution:
Do not contact the non-sterile battery or the inside of the battery casing in order to avoid con­tamination. Should the scrubbed person come into contact with either the non-sterile battery or the in­side of the battery casing, they must be scrubbed again. Should the battery casing be contaminated it must be cleaned and resterilized before being used in the operating room.
Notes:
Normally, one fully charged battery has sufficient
capacity for an entire operation. As a precaution, a second battery pack (battery casing with in­serted battery) should be kept ready, so that the battery pack can be quickly exchanged under sterile conditions during surgery if necessary.
Never open a battery casing intraoperatively to
insert a new battery. Always replace the whole battery pack by another battery pack prepared before the start of the surgery.
Insert the battery pack into the drive unit, ensuring the contacts on the battery pack align with the con­tacts in the recess of the drive unit (Fig. 8). Press firmly to ensure the battery pack is engaged cor­rectly, and check by pulling lightly downward on the battery pack.
Notes:
For safety reasons, the battery pack can only be
inserted fully when it is in the correct orientation.
To prevent injuries, the mode switch of the drive
unit should always be in the off position when in­serting or removing the battery pack.
Installing the battery pack just before use pre-
vents unwanted discharge of battery capacity.
Figure 7a
Figure 7b
Figure 8
Scrubbed person
Scrubbed person
Scrubbed person
530.705
530.690
16 DePuy Synthes Battery Power Line II Instructions for Use

Removing and disassembling the battery pack

Press both release buttons simultaneously on the drive unit to remove the battery pack (Fig. 9). Open the casing by pressing both battery casing locks and remove the battery or hold open the bat­tery casing to allow another person to remove the battery (Fig. 10). Ensure that the battery does not touch the exterior of the battery casing in order to avoid contaminating the battery. If this occurs follow the information in the “Care and Maintenance” section starting on page 39. Store battery in Universal Battery Charger II (05.001.204) when not in use (Fig. 11).
Precaution:
Do not wash, rinse, sterilize, drop or apply force to the battery (530.630). This will destroy it with possi­ble secondary damage.
Figure 9
Circulating person
Scrubbed person
Scrubbed person
Figure 10
Figure 11
Battery Power Line II Instructions for Use DePuy Synthes 17
Operating Instructions Battery Pack (battery casing with inserted battery)

Charging, storing and using batteries

Charging
Only use the Synthes Universal Battery Charger II (05.001.204) to charge the battery. Using a charger that does not originate from Synthes can damage the battery. In order that the BPL II battery can be recognized and charged by the UBC II, firmware version 14.0 is required. If necessary, send the charger to a DePuy Synthes representative for a firmware update. The BPL II battery cannot be charged with the Uni­versal Battery Charger (item number 530.600 or
530.601). The batteries should always be charged before use. Place the battery into the charger immediately after surgery. Charge the batteries within an ambient temperature range of 10°C / 50°F to max. 40°C / 104°F. Keep the charger and the batteries clean and in a cool and dry place. Detailed information on the Universal Battery Charger II can be found in the Instructions for Use
(036.000.500).
Storage
Always recharge the Li-Ion Battery for Battery Power Line II (530.630) after each use. Do not store an empty battery as this will shorten the life span and will not be covered by warranty. When the battery is not in use, store it in the Synthes Universal Battery Charger II (05.001.204). This will guarantee that the battery is always fully charged and ready to use. The Universal Battery Charger II should always be turned on when a battery is in the charging bay. This ensures availability of charged batteries. Do not expose batteries to heat or fire. Avoid storage in direct sunlight. Keep the charger and the batteries clean and in a cool and dry place.
Use
Do not remove battery from its original packaging until required for use. Do not drop or apply force to the battery. This will destroy it with possible secondary damage. Do not use any battery which is not designed for use with the equipment. Do not use a faulty or damaged battery, as this may damage the power tool. Do not short-circuit a battery. Do not store or transport batteries haphazardly in a box or drawer where they may short-circuit each other or be short-circuited by other metal objects. This can damage the batteries and generate heat, which can cause burns. Batteries give their best performance when they are operated at normal room temperature (20°C / 68°F +/- 5°C / 9°F). The power tool may only be operated with a fully charged battery. Therefore the batteries should al­ways be charged before use. Place the battery into the charger immediately after surgery. Only insert the battery pack directly before using the power tool. This saves battery energy and prevents having to change it during surgery. Follow the information in the “Care and Mainte­nance” section starting on page 39 as well as the Instructions for Use of the Synthes Universal Battery Charger II (036.000.500).
18 DePuy Synthes Battery Power Line II Instructions for Use
Notes:
Generally, medical power tools can heat up if in
constant use. The cool down times should be
observed, see “Duty Cycle” section on page 64,
in order to prevent the power tool from exceed-
ing its acceptable surface temperature. – In case of cell leakage, do not allow the leaking
fluid to come into contact with skin or eyes. In
case of contact, wash the affected area with co-
pious amounts of water and seek medical ad-
vice. – Faulty batteries may not be reused and should
be disposed of in an environmentally friendly
manner and in accordance with national regula-
tions.
Warnings:
Never expose batteries to temperatures over
60°C / 140°F. The maximum exposure time at
60°C / 140°F is 72 hours. – Do not dismantle, open or shred batteries.
Battery Power Line II Instructions for Use DePuy Synthes 19
Operating Instructions
Battery Reamer/Drill II (530.705)
For clockwise rotation, turn the mode switch to the “FWD” position.
For counterclockwise rotation, turn the mode switch to the “REV” position.
The single variable-speed trigger allows control of the speed from 0 to the maximum rpm. Maximum torque and speed vary, depending on the attachment (see pages 21 ff). Ensure that the correct attachment is used for each operation in terms of speed and torque.
For further information on the System Specifications and Duty Cycle please refer to page 63f.
For safety, turn the mode switch to the OFF position.
For clockwise rotation, turn the mode switch to the FWD position.
For counterclockwise rotation turn the mode switch to the REV position.
20 DePuy Synthes Battery Power Line II Instructions for Use
Operating Instructions
Attachments for Battery Reamer/Drill II

Instrument

530.705 Battery Reamer/Drill II
Precaution:
To prevent injuries, the mode switch of the drive unit should always be in the “OFF” position when inserting or removing attachments and cutting tools.
Please observe the safety instructions and warnings stated in the instructions when working with attach­ments. Only use original Synthes attachments. Damage resulting from the use of attachments from other manufacturers is not covered by the warranty.
Insert attachment
Insert the attachment into the coupling of the Battery Reamer/Drill II, aligning the positioning pins of the at­tachment with the grooves on the attachment release ring (Fig. 1). Turn the attachment release ring in the direction of the arrow and push the attachment until it engages in place (Fig. 2). If the attachment does not engage properly then rotate the attachment gently until the drive shaft engages. Check that the attachment coupling is closed properly by lightly pulling the attachment.
Remove attachment
Turn the attachment release ring in the direction of the arrow and remove the attachment.
Note:
Properly functioning tools are essential to the success of an operation. For this reason, used tools must be checked for wear and/or damage after each use and replaced if necessary.
Figure 1
Figure 2
Battery Power Line II Instructions for Use DePuy Synthes 21
Loading...