1 INTRODUCTION ...............................................................................................................................3
INTENDED USE OF THIS DEVICE ..............................................................................................
2 SAFETY INFORMATION....................................................................................................................4
DANGER: ....................................................................................................................................
WARNING: ..................................................................................................................................5
CAUTION: ...................................................................................................................................5
3 HANDLING ........................................................................................................................................6
GF HEALTH PRODUCTS, INC. FREIGHT POLICY .......................................................................6
APPARENT LOSS OR DAMAGE .............................................................................................6
CONCEALED LOSS OR DAMAGE ..........................................................................................
UNPACKING ...............................................................................................................................
INSPECTION .........................................................................................................................6
STORAGE ...................................................................................................................................6
4 INSTALLATION .................................................................................................................................7
INSTALLATION LOCATION SAFETY INFORMATION ..................................................................7
MAINTENANCE...........................................................................................................................8
RADIO FREQUENCY INTERFERENCE ........................................................................................8
PRODUCT DESCRIPTION ...........................................................................................................9
CONCENTRATOR ILLUSTRATION ........................................................................................9
5 OPERATION ....................................................................................................................................10
SETUP ......................................................................................................................................10
TURNING THE CONCENTRATOR ON .......................................................................................10
CIRCUIT BREAKER (RESET BUTTON) .....................................................................................11
TURNING THE CONCENTRATOR OFF ......................................................................................11
OPERATION WITH A HUMIDIFIER BOTTLE (IF SO PRESCRIBED) ...........................................12
LED INDICATOR TABLE ............................................................................................................13
3
4
6 6
6 TRANSPORTING THE OXYGEN CONCENTRATOR .........................................................................14
USING THE LUGGAGE HANDLE ...............................................................................................14
SAFETY DURING TRANSPORT ................................................................................................14
7 MAINTENANCE ...............................................................................................................................15
REPLACEMENT PARTS ............................................................................................................15
USER MAINTENANCE ..............................................................................................................15
CLEANING THE CABINET FILTER .....................................................................................15
CLEANING THE CABINET ..................................................................................................16
CLEANING THE (OPTIONAL) HUMIDIFIER BOTTLE ..........................................................16
PREVENTIVE MAINTENANCE RECORD ...................................................................................17
8 TROUBLESHOOTING ......................................................................................................................18
TROUBLESHOOTING TABLE ....................................................................................................19
9 SPECIFICATIONS ...........................................................................................................................20
10 LIMITED WARRANTY .....................................................................................................................21
11 INDEX .............................................................................................................................................23
Graham-Field and John Bunn are registered trademarks, and Optium is a trademark, of GF Health Products, Inc. Packaging, warranties, products, and specications are subject to change without notice.
GF Health Products, Inc. is not responsible for typographical errors.
JB0180-INS-LAB-RevA08 2
Federal Law (USA) restricts this device to sale by or on the order of a physician. This product should not be used unless the operator has been instructed by a
qualied healthcare professional.
WARNING: Users who require continuous oxygenation must plan for alternate reserve sources of power and oxygen in the event of a failure or loss of power and oxygen. This device is to be used ONLY as an oxygen supplement device and is NOT intended to be a life-supporting or life-sustaining device.
1 INTRODUCTION
This user manual contains important information and safety precautions for the John Bunn Optium Oxygen Concentrator.
Before using your Oxygen Concentrator, please read and understand this entire user manual. Take special note of all safety precautions marked DANGER and WARNING. Save this user manual for future reference.
Note: The most current version of this manual can be found online at www.grahamfield.com.
INTENDED USE OF THIS DEVICE
The John Bunn Optium Oxygen Concentrator is intended for use as an oxygen supple­ment device in a home or care facility. Use the concentrator only as prescribed by a physician. This user manual serves as a reference for your oxygen concentrator.
WARNING: THIS DEVICE IS NOT INTENDED FOR LIFE SUPPORT. Power interruption could prevent the device from operating.
There are different accessories that can be used with an oxygen concentrator. Such ac­cessories include oxygen tubing, humidiers, cannulas and masks. Your physician or provider will recommend the different types of accessories to use with your oxygen con­centrator. John Bunn accessories are available and recommended for use with the John Bunn Optium Oxygen Concentrator.
WARNING: Failure to observe the safety statements in this manual can result in
severe re, property damage, physical injury, and/or DEATH.
WARNING: Under certain diagnoses, oxygen therapy can be hazardous; use this
concentrator only as prescribed by a physician.
WARNING: To ensure user safety, use this device only with accessories recommended by Graham-Field, your Graham-Field authorized distributor, your physician, or your respiratory therapist.
JB0180-INS-LAB-RevA08 3
2 SAFETY INFORMATION
The safety statements presented in this chapter refer to the basic safety information that the operator of the oxygen concentrator shall pay attention to and abide by. There are additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specic to the operations. Please note the following special statements, used throughout this manual, and their signicance:
DANGER: Indicates an imminent hazard situation that, if not avoided, will result in death or serious personal injury.
WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in death or serious personal injury.
Caution: Indicates a potential hazard or unsafe practice that, if not avoided, could
s
result in minor personal injury or product/property damage.
Note: Provides application recommendations or other useful information to ensure that you get the most from your product.
DANGER:
To reduce the risk of Electrocution, Fire and/or Explosion
DANGER: Do not use this device while bathing. If continuous usage is required by
the direction of a physician, the concentrator must be located in another room at
least 9 feet (2.8m) from the bath.
DANGER: Do not place or store this device where it can fall or be pulled into a tub or sink.
DANGER: Do not place this device in or drop it into water or other liquid.
DANGER: Do not reach for a device that has fallen into water—unplug it immediately!
DANGER: DO NOT come in contact with the concentrator when you are wet.
DANGER: The oxygen concentrator contains no serviceable parts—do not open
the cabinet! Opening of the concentrator cabinet by anyone other than authorized service personnel could void the warranty and create a hazardous condition, which could result in death or serious personal injury.
DANGER: Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.
JB0180-INS-LAB-RevA08 4
WARNING:
To reduce the risk of burns and/or serious personal injury
WARNING: Federal Law (USA) restricts this device to sale by or on the order of a physician.
WARNING: Use this device only for its intended use as described in this manual.
WARNING: This device is to be used only in accordance with the prescription of a
physician and this user manual. NEVER make adjustments to the ow rate unless
prescribed by a physician.
WARNING: If at any time you feel that you are not receiving the correct amount of
oxygen, contact your provider and/or physician immediately.
WARNING: DO NOT allow any powder or powder-like substance near the oxygen
concentrator and/or its tubing; the powder could cause contamination of the device, and the possible resulting device malfunction could cause oxygen to drop
to ineffective levels.
WARNING: Supervise use of this device closely when it is used by, on, or near
children or those who require extra care.
WARNING: Always unplug this device immediately after use.
WARNING: Do not leave this device unattended when it is plugged in.
CAUTION:
To reduce the risk of minor personal injury and/or product/property damage
Caution: During transport, always keep the concentrator upright to prevent dam-
s
age. If the concentrator is transported in the trunk of a car, ensure that the concen­trator is secured so that it will remain upright.
Caution: Ensure that the temperature at the concentrator's location during trans-
s
port does not fall below -40°F (-40°C) or exceed 158°F (70°C).
JB0180-INS-LAB-RevA08 5
3 HANDLING
GF HEALTH PRODUCTS, INC. FREIGHT POLICY
For your protection, read carefully
The concentrator should ALWAYS be kept in the upright position to prevent damage while being transported. The Optium packaging has been designed to assure maximum protection of the concentrator.
The carrier accepted this merchandise "in good condition" and is responsible for safe delivery. Before signing the freight bill, inspect the shipment carefully for damage or missing pieces.
Apparent loss or damage
Should visual inspection show loss or damage, this MUST be noted on the freight bill and signed by the carrier's agent. Failure to do so may result in the carrier failing to honor the claim. Please contact the carrier to obtain the paperwork necessary to le a claim or contact Graham-Field Customer Service at the number on the back cover.
Concealed loss or damage
If damage is discovered after delivery is made, a concealed damage claim must be en­tered with the freight carrier. When this occurs, make a written request to the carrier for inspection. This request for inspection must be made within 15 days of delivery. The car­rier will provide all paperwork necessary to le a concealed damage or loss claim, since such damage or loss is the carrier's responsibility.
UNPACKING
Note: Unless the oxygen concentrator is to be used immediately, retain containers and packing materials for storage until concentrator use is required.
1. Check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field authorized distributor.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
Inspection
Examine exterior of the oxygen concentrator for nicks, dents, scratches or other dam­ages. Inspect all components.
STORAGE
Store the repackaged oxygen concentrator in a dry area. DO NOT place anything on top of the repackaged concentrator.
Always store the concentrator upright to prevent damage.
s
Caution: Ensure that the temperature at the concentrator's location during storage
s
does not fall below -40°F (-40°C) or exceed 158°F (70°C).
JB0180-INS-LAB-RevA08 6
4 INSTALLATION
INSTALLATION LOCATION SAFETY INFORMATION
Select a well-ventilated, fume-free room in your house where using the oxygen concen­trator would be most convenient. The concentrator can be easily moved from room to room on its casters.
To Reduce the Risk of Electrocution, Fire, Explosion Burns, Serious Personal Injury, and/or Property Damage
WARNING: Keep the oxygen concentrator away from ammable and explosive
areas.
WARNING: To reduce the risk of re, special care must be taken when using an oxygen concentrator. DO NOT SMOKE while using this device. Keep all matches, lighted cigarettes, and other sources of ignition out of the room in which this
concentrator is located. In public settings or when used in hospital or long-term
care environments, prominently display NO SMOKING signs.
WARNING: A spontaneous and violent ignition could occur if oil, grease, or greasy
substances come in contact with oxygen under pressure. ALWAYS keep these substances away from the oxygen concentrator and all other oxygen equipment.
WARNING: Textiles and materials that normally would not easily burn can
ignite and burn with great intensity in oxygen-enriched air; ALWAYS keep these
materials away from the oxygen concentrator and all other oxygen equipment.
WARNING: The oxygen concentrator is intended to be rolled from room to
room. Do not place the oxygen concentrator in surroundings where its airow is
obstructed.
WARNING: Be certain to place the concentrator so that all sides are at least 12
inches (30cm) away from walls, draperies, furniture, or other obstructions. Do not place the concentrator in a conned area.
WARNING: The oxygen concentrator MUST be kept away from heat, re, ammable and explosive areas, and excessive water sources and conditions. Keep all oxygen tubing, cords and oxygen concentrator away from heated or hot surfaces, including space heaters, blankets, stoves and similar electrical
appliances.
WARNING: Do not use a household extension cord if the AC cord does not reach
the power outlet. Use of an extension cord could result in re and electric shock.
WARNING: DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.
WARNING: DO NOT plug the concentrator into the same outlet as any other major
appliance.
JB0180-INS-LAB-RevA08 7
WARNING: Always place the concentrator on a hard surface. Never place the
concentrator on a surface such as a bed or couch, where the concentrator could
tip or fall.
WARNING: Do not place items on top of the concentrator.
WARNING: NEVER block the concentrator's air openings.
WARNING: Keep the concentrator openings free from lint, hair, etc.
Caution: The concentrator should be appropriately placed to keep it from falling,
s
strong vibration or other mechanical damage. Caution: Only use stable and safe electrical power sources.
s
Caution: Do not pull the oxygen concentrator by the power cord.
s
Caution: If the oxygen concentrator has been subjected to sub-freezing tempera-
s
tures for an extended period of time, allow it to warm up to the stated operating temperatures before power up. Failure to do so could result in improper perfor­mance and or alarm conditions until the device reaches normal operating tempera­tures.
MAINTENANCE
Proper care and maintenance are essential to keep your oxygen concentrator in safe working condition. When you believe that a component or part of your concentrator is not functioning properly, contact your Graham-Field authorized distributor immediately, as a potentially hazardous condition could result. Only excellent condition is acceptable where safety is concerned.
RADIO FREQUENCY INTERFERENCE
Most electronic equipment is inuenced by Radio Frequency Interference (RFI). ALWAYS exercise caution with regard to the use of portable communication equipment in the area around the oxygen concentrator.
JB0180-INS-LAB-RevA08 8
PRODUCT DESCRIPTION
flowmeter
circuit breaker (reset button)handle
oxygen tubing
Optium
Oxygen Concentrator
front cabinet
(JB0180 shown)
caster
base
cabinet
humidifier compartment
for humidifier bottle
(optional accessory)
oxygen outlet (connector to
oxygen tubing or
humidifier bottle)
flowmeter knob
power switch
power LED (green)
alarm LED (red)
LCD panel
Optium Oxygen Concentrator
rear cabinet
* Service to be performed by authorized service personnel ONLY
** Velcro is a registered trademark of Velcro Industries B.V. Corporation
cabinet filter
luggage handle
compressor intake filter *
(behind the access door)
power cord with
AC connector
bacteria HEPA filter *
(inside the rear cabinet)
Velcro®strap **
Concentrator Illustration
Concentrator features are shown below.
JB0180-INS-LAB-RevA08 9
5 OPERATION
Optium
Oxygen Concentrator
connections
oxygen outlet
tubing
mask
nasal cannula
WARNING: Read the safety section and operating instructions in this manual before use.
SETUP
1. Before proceeding, ensure that all instructions in Section 4, Installation, have been followed, and that you have read the SAFETY section of this manual.
2. Ensure that the concentrator power switch is in the OFF position. Plug the concentra­tor's AC connector into a properly grounded power outlet.
WARNING: As a safety feature, this device has a polarized plug (one blade is wider than the other). This plug will t in a polarized outlet only one way. If the plug does not t completely in the outlet, reverse the plug. DO NOT attempt to defeat this safety feature. Contact a qualied electrician or your Graham-Field authorized distributor if the plug does not t in the polarized outlet.
TURNING THE CONCENTRATOR ON
Note: For use with a humidifier bottle, please see OPERATION WITH A HUMIDIFIER BOT- TLE (IF SO PRESCRIBED) on the following page.
1. Connect the oxygen tubing securely to the oxygen outlet; connect the other end to the patient's cannula or mask outlet (these may already be connected).
2. Move the concentrator's ON/OFF switch to the ON position to turn on concentra­tor power.
Model JB0180: The green (power)
LED will illuminate.
Model JB0180S: the yellow low oxygen LED will illuminate for approximately
three minutes while the system performs an Oxygen Purity test, then the green (power) LED and appropriate O2 LED will illuminate.
Ensure that oxygen is flowing through the tubing before proceeding.
WARNING: If the ow rate on the owmeter ever falls below 0.5 LPM, check tubing or accessories for blocked or kinked tubing or for a defective humidier bottle.
3. Turn the owmeter knob to center the owmeter ball on the LPM prescribed. The owmeter can be adjusted only after the unit is powered on.
WARNING: Do not change LPM setting unless prescribed by a physician.
Note: Oxygenation times and flow rate ranges are prescribed by your physician.
JB0180-INS-LAB-RevA08 10
CIRCUIT BREAKER (RESET BUTTON)
Your oxygen concentrator is equipped with a circuit breaker (reset button), to protect the electrical circuits from overload, next to the power switch. In the event of an over­load, the circuit breaker button will pop out. To reset the circuit breaker, push the but­ton in.
TURNING THE CONCENTRATOR OFF
After using the concentrator for the time prescribed by your physician, move the con­centrator's ON/OFF switch to the OFF position to turn off concentrator power. Unplug the concentrator's AC connector from the power outlet.
WARNING: Never operate this device if it has a damaged cord or plug; is not working properly; or has been dropped, damaged, or dropped into water. Return the device to your Graham-Field authorized distributor for examination and repair.
JB0180-INS-LAB-RevA08 11
OPERATION WITH A HUMIDIFIER BOTTLE (IF SO PRESCRIBED)
Optium
Oxygen Concentrator
connections when using
a humidifier bottle
tubing connector
tubing
humidifier compartment
humidifier bottle
mask
nasal cannula
oxygen outlet nut
WARNING: The use of a humidier bottle with a concentrator must be prescribed
by a physician.
WARNING: DO NOT overll the humidier bottle.
WARNING: DO NOT reverse the oxygen inlet and output connections, or water from the humidier bottle will travel through the oxygen hose back to the patient.
1. Unscrew the humidier bottle cover.
2. Fill the humidier bottle with puried or distilled water to a point between the maximum and minimum water level lines. If needed, add medication to the water as prescribed by your physician.
3. Replace the cover on the humidier bottle and tighten it securely.
4. Place the humidier bottle in the oxy­gen concentrator's humidier compart­ment, attaching the humidier bottle cover nut to the oxygen outlet.
5. Turn the nut counter-clockwise to se­cure it to the oxygen outlet. Ensure
that the humidifier bottle cover nut is securely attached to the oxygen outlet be­fore continuing.
6. Attach oxygen tubing from the humidier bottle to the patient's cannula or mask out
-
let (these may already be connected).
7. Ensure that the power switch is in the OFF position. Plug the concentrator's AC con
-
nector into a properly grounded power outlet.
8. Turn the concentrator on: See the previous section, TURNING THE CONCENTRATOR ON, and follow steps 2-3.
JB0180-INS-LAB-RevA08 12
LED INDICATOR TABLE
JB0180S LEDs
POWER NORMAL
OXYGEN
LOW
OXYGEN
SERVICE
REQUIRED
(GREEN) (GREEN) (YELLOW) (RED)
JB0180 LEDs
POWER SERVICE
REQUIRED
(GREEN) (RED)
JB0180S LEDs
POWER NORMAL
OXYGEN
LOW
OXYGEN
SERVICE
REQUIRED
(GREEN) (GREEN) (YELLOW) (RED)
The following LED Indicator Table lists LED indicator patterns and causes so that you can become familiar with the concentrator's normal indicators. If you are having a prob­lem with the concentrator, please see Section 8, TROUBLESHOOTING, where there is a list of symptoms and actions.
Model JB0180 only
Indicator LED(s) and/or Audible Alarm Probable Cause
Power POWER LED illuminates nonashing Power is on
Model JB0180S only
Alarm SERVICE REQUIRED LED illuminates;
audible alarm sounds
1)Power failure
2)Concentrator failure
3)High pressure
4)Low pressure
5)O2 < 72% - do not use concentrator, O2 too low
Contact your GF authorized
distributor
Alert SERVICE REQUIRED LED illuminates;
No oxygen ow
intermittent audible alarm sounds
Indicator LED(s) and/or Audible Alarm Probable Cause
Power POWER LED illuminates nonashing Power is on
Alarm SERVICE REQUIRED LED illuminates;
audible alarm sounds
1)Power failure
2)Concentrator failure
3)High pressure
4)Low pressure
5)O2 < 72% - do not use concentrator, O2 too low
Contact your GF authorized
distributor
Alert SERVICE REQUIRED LED illuminates;
O2 Indicator NORMAL OXYGEN LED illuminates O2 > 85%
JB0180-INS-LAB-RevA08 13
intermittent audible alarm sounds
LOW OXYGEN LED illuminates;
SERVICE REQUIRED LED ashes every
two seconds; audible alarm sounds
LOW OXYGEN LED illuminates for
approximately 3 min. at startup; then POWER LED and appropriate O
2
LED
illuminate
No oxygen ow
O
72%, < 85%
2 >
O
Purity Test performs
2
automatically at startup
6 TRANSPORTING THE OXYGEN CONCENTRATOR
luggage handle
in
DOWN position
luggage handle
in
UP position
luggage
handle
The Optium Oxygen Concentrator is equipped with a luggage handle and casters for transport. To transport the concentrator:
1. Remove and store any accessories, such as tubing, cannulas, masks, and humidier bottle, if so equipped.
2. Fold and wrap the power cord and attach it to the concentrator cabinet using the Velcro strap, as shown below.
USING THE LUGGAGE HANDLE
Extend the luggage handle: Pull the handle up completely. Ensure that the handle is completely extended and securely in place.
Retract the luggage handle: Lower the handle completely. Ensure that the handle is completely retracted and securely in place.
Caution: Do not lift oxygen concentrator by
s
the luggage handle.
SAFETY DURING TRANSPORT
Stow the oxygen concentrator upright in a safe place, observing the following:
WARNING: Keep the oxygen concentrator away from ammable and explosive
areas.
WARNING: To reduce the risk of re, DO NOT SMOKE while transporting this device. Keep all matches, lighted cigarettes, and other sources of ignition out of
the vehicle or room in which this concentrator is located.
WARNING: A spontaneous and violent ignition could occur if oil, grease, or greasy
substances come in contact with oxygen under pressure. ALWAYS keep these substances away from the oxygen concentrator and all other oxygen equipment.
WARNING: Textiles and other materials that normally would not easily burn can
ignite and burn with great intensity in oxygen-enriched air; ALWAYS keep these
materials away from the oxygen concentrator and all other oxygen equipment.
Caution: During transport, always keep the concentrator upright to prevent dam-
s
age. If the concentrator is transported in the trunk of a car, ensure that the concen­trator is secured so that it will remain upright.
Caution: Ensure that the temperature at the concentrator's location during trans-
s
port does not fall below -40°F (-40°C) or exceed 158°F (70°C).
JB0180-INS-LAB-RevA08 14
7 MAINTENANCE
cabinet
filter
Note: There is a Preventive Maintenance Record on the next page. When service is per­formed on your concentrator, please make this record available to the service personnel performing the service, to ensure that accurate maintenance records are kept.
Note: All maintenance except Cleaning the cabinet filter, Cleaning the cabinet, and Cleaning the (optional) humidifier bottle, described in those sections directly following, is to be performed only by authorized service personnel.
REPLACEMENT PARTS
Replacement parts and accessories are available through your Graham-Field authorized distributor. For a list of parts, please visit www.grahameld.com.
USER MAINTENANCE
DANGER: The concentrator was specically designed to minimize routine preventive maintenance. Preventive maintenance or adjustments to the concentrator are to be performed only by authorized service personnel; the performance of such maintenance or adjustment by any other personnel could void the warranty and create a hazardous condition, which could result in death or serious personal injury.
WARNING: Ensure that the concentrator is unplugged from the AC power source and the power switch is in the OFF position before performing maintenance on the concentrator.
Cleaning the cabinet lter
Clean the cabinet lter weekly, or more often if it gets dirty or your environment has high dust or pollutant levels.
1. Ensure that the power cord is unplugged and the power
switch is in the OFF position.
2. Remove the lter carefully from the rear of the cabinet
(see picture at right).
3. Vacuum the lter carefully or wash it with a mild detergent
and tap water. If washed, allow the lter to air dry before reinstalling it in the concentrator.
4. Reinstall lter in concentrator.
Caution: Do not operate the concentrator without the
s
cabinet filter installed, or while the filter is wet. These actions could permanently damage the concentrator.
JB0180-INS-LAB-RevA08 15