John Bunn JB0112-090, Neb-u-Lite EV2 User manual

Neb-u-Lite® EV2
Nebulizer Compressor
Item JB0112-090
User Manual
Read this manual before operating the nebulizer.
Save this manual for future reference.
Federal Law (USA) restricts this device to sale by or
on the order of a physician. This device should not be used unless the operator has been instructed by a
qualied healthcare professional.
JB0112-090-INS-LAB-RevA18
CONTENTS
INTRODUCTION .............................................................................................3
INTENDED USE ........................................................................................3
IMPORTANT SAFETY PRECAUTIONS ...........................................................4
PRODUCT DESCRIPTION ............................................................................... 7
FEATURES AND COMPONENTS .............................................................7
SETUP .............................................................................................................. 8
OPERATION ...................................................................................................10
NEBULIZER OPERATION ....................................................................... 10
SAFETY OVERLOAD ..............................................................................11
MAINTENANCE ............................................................................................. 12
CLEANING THE COMPRESSOR’S OUTER CASE ...................................12
CLEANING THE NEBULIZER CUP ..........................................................12
REPLACING THE FILTER ........................................................................13
STORAGE ................................................................................................13
TROUBLESHOOTING ....................................................................................14
SPECIFICATIONS ..........................................................................................15
LIMITED WARRANTY ....................................................................................16
COMPRESOR DE NEBULIZADOR MANUAL DE USUARIO EN ESPAÑOL
JB0112-090-INS-LAB-RevA18 2
.....19
INTRODUCTION
is user manual contains important information and safety precautions for the John Bunn Neb-u-Lite EV2 Nebulizer Compressor.
Before using your Nebulizer Compressor, please read and understand this entire user manual. Take special note of all safety precautions marked DANGER and WARNING. Save this user manual for future reference.
Intended Use
e Neb-u-Lite EV2 Nebulizer Compressor System is designed to provide a source of compressed air for aerosol therapy. It is used in conjunction with a jet (pneumatic) nebulizer to produce medicated aerosols for inhalation by patients with respiratory symptoms.
Indications for therapy include asthma, chronic bronchitis, infection of the upper respiratory tract, chronic obstructive pulmonary disease, and other respiratory disorders in accordance with a physician’s prescription. Do not use this product for any other purpose than the preceding. is device can be used with pediatric or adult patients under physician’s prescription.
WARNING: This product contains small parts that may
present a choking hazard to small children, and the tubing may present a strangulation hazard. Always use close supervision when this device is used by or near children or those who require close supervision.
WARNING: Children under the age of 3, or any patient
who is unable to use a mouthpiece properly under supervision, should use a mask.
JB0112-090-INS-LAB-RevA18 3
IMPORTANT SAFETY PRECAUTIONS
When using electrical products, especially when children are present, always follow basic safety precautions, including the following:
DANGER: Indicates a potential hazard situation or unsafe
practice that, if not avoided, will result in death or serious personal injury.
WARNING: Indicates a potential hazard situation or
unsafe practice that, if not avoided, could result in death or serious personal injury.
CAUTION: Indicates a potential hazard or unsafe practice
that, if not avoided, could result in minor or moderate personal injury.
s NOTICE: Indicates a potential hazard or unsafe practice
that, if not avoided, could result in product or property damage.
Info: Provides application recommendations or other useful information to ensure that you get the most from your product.
READ ALL INSTRUCTIONS BEFORE USING
DANGER: To reduce the risk of electrocution:
1. Always unplug this product immediately after use.
2. Do not use this product while bathing, showering, washing dishes, or close to water sources of any kind.
3. Do not place or store device where it can fall or be pulled into a tub or sink.
4. Do not place device in or drop into water or other liquid.
5. Do not reach for a device that has fallen into water. Unplug it immediately.
JB0112-090-INS-LAB-RevA18 4
WARNING: To reduce the risk of burns, electrocution, re
or personal injury:
1. Never leave this device unattended when plugged in.
2. is product contains small parts that may present a choking hazard to small children, and the tubing may present a strangulation hazard. Always use close supervision when this
device is used by or near children or those who require close supervision.
3. Children under the age of 3, or any patient who is unable to use a mouthpiece properly under supervision, should use a mask.
4. Use this device only as intended and described in this manual. Do not use attachments or accessories not recommended by GF Health Products, Inc.
5. Never operate this product if: a. it has a damaged cord or plug, b. it is not working properly, c. it has been dropped or damaged, or d. it has been dropped into water.
Return the product to your Graham-Field authorized
distributor for examination and repair.
6. Keep the cord away from heated surfaces.
7. Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the air openings may be blocked. Keep the air openings free of lint, hair and the like.
8. Never drop or insert any object into any openings or hose.
9. Never use this device while drowsy or sleeping.
10. Do not use this device outdoors or operate where aerosol (spray) products are being used.
JB0112-090-INS-LAB-RevA18 5
11. Do not use this device in the presence of ammable anesthesia or where oxygen or nitrous oxide is being administered. This device has no AP or APG protection.
12. Connect this device to a properly grounded outlet only.
13. This device has a polarized plug (one blade is wider than the other). As a safety feature, this plug will t into a polarized outlet only one way. If the plug does not t fully into the outlet, reverse the plug. If it still does not t, contact a qualied electrician. Do not modify the plug in any way. Do not attempt to defeat this safety feature.
14. Unplug the device before lling or cleaning the nebulizer.
15. This device is oil-free. Do not lubricate.
16. Do not remove the jet nebulizer while the device is operating.
17. Do not leave medication inside the jet nebulizer when it is not in use.
18. Use the nebulizer intermittently; the maximum operating period is 30 minutes. The nebulizer must stand unused for 30 minutes before use.
19. The mouthpiece and masks are for use by one person only; in order to avoid infection, do not share them.
20. WARNING: Cancer and Reproductive Harm -
www.p65warnings.ca.gov
JB0112-090-INS-LAB-RevA18 6
PRODUCT DESCRIPTION
adult mask
Nebulizer Compressor
ON / OFF switch
Features and Components
Please note nebulizer features and components, labeled in the illustration below.
nebulizer / compressor
pediatric mask
nebulizer cup
filter cap
mouthpiece
air outlet connector for tubing
JB0112-090 Neb-u-Lite EV2
tubing
Info: Neb-u-Lite EV2 replacement parts are available at www.grahamfield.com or from your Graham-Field authorized distributor.
JB0112-090-INS-LAB-RevA18 7
SETUP
install mask
on cap
mask
install cap
on cup
cap
cup
mouthpiece
oncap
mouthpiece
Info: Before initial operation, clean the nebulizer cup assembly following the instructions given in MAINTENANCE/Cleaning the Nebulizer Cup.
WARNING: Ensure that the nebulizer compressor is
unplugged during setup.
1. Before each use, inspect the compressor and nebulizer cup assembly for damage or wear; replace as needed.
2. Place the compressor on a table or other at stable surface (not on the oor), in an area that is clean and dust-free. Be sure you can easily reach the controls when seated.
3. Open the nebulizer medication cup: Twist the cap counter­clockwise until it stops, then lift cap o.
4. Pour the prescribed medication into the nebulizer medication cup. Do not ll with more than 8 ml per treatment.
5. Install cap on cup. Turn clockwise to tighten.
6. If using the mouthpiece for treatment,
7. Install mouthpiece onto top of nebulizer
JB0112-090-INS-LAB-RevA18 8
proceed to step 7. If using a mask, install the bottom part of the mask directly onto the top of the nebulizer medication cup cap as shown at right. Proceed to step 8.
medication cup cap.
cap
cup
install
8. Connect the tubing to the adapter on the bottom of the nebulizer medication cup.
9. Connect the tubing’s other end to the compressor air outlet connector.
WARNING: Before use, check that all parts are securely
connected. Use of an improperly assembled nebulizer could diminish or prevent adequate delivery of medication and the eectiveness of the treatment.
JB0112-090-INS-LAB-RevA18 9
OPERATION Nebulizer Operation s NOTICE: The Neb-u-Lite EV2 Nebulizer Compressor is
designed for intermittent use only. Do not operate it continuously for more than 30 minutes for a single use without turning it o and allowing a cooling period of at least 30 minutes.
1. Ensure that the ON/OFF switch is in the OFF position.
2. Plug the power cord into a properly grounded electrical wall outlet.
3. Move the ON/OFF switch to the ON position to start the compressor.
4. Place the nebulizer mouthpiece between teeth and close mouth (or install the mask on face).
5. As aerosol starts owing, inhale deeply and slowly through mouthpiece (or mask).
WARNING: Never use a nebulizer with a clogged nozzle.
If the nozzle is clogged, the aerosol mist will be reduced, altered or stopped, which will diminish or eliminate the eectiveness of the treatment. If clogging occurs, replace the nebulizer.
Info: If you need to interrupt treatment, move the ON/OFF switch to the OFF position.
6. When the treatment is complete, move the ON/OFF switch to the OFF position to turn o the compressor.
7. Unplug the compressor from power outlet.
8. Disconnect nebulizer from tubing.
9. Clean nebulizer as instructed in MAINTENANCE/Cleaning the Nebulizer Cup.
JB0112-090-INS-LAB-RevA18 10
Safety Overload s NOTICE: Do not exceed 30 minutes of continuous
operation.
1. e motor of this device has a built-in thermal overload protector. Should the motor overheat, the protector will automatically shut o the motor. Should this occur, move the ON/OFF switch to the OFF position and allow the motor to cool down for approximately 30 minutes before turning it on again.
2. If the overload protector shuts o the motor frequently, you may have an unstable voltage situation.
3. If the device shuts down and cannot restart, it may need to be replaced. Call your Graham-Field authorized distributor immediately.
JB0112-090-INS-LAB-RevA18 11