The user manual sets out the operating instructions for the
instrument. It is designed to provide a quick reference guide for
use in the field.
The reference manual is aimed at users who are familiar with
AMETEK calibrators, as well as those who are not. The manual
is divided into 10 chapters which describe how to set up,
operate, service and maintain the calibrator. The technical
specifications are described and accessories may be ordered
from the list of accessories.
•Safety symbols
is manual contains a number of safety symbols designed to
Th
draw your attention to instructions which must be followed when
using the instrument, as well as any risks involved.
Warning
Conditions and actions that may compromise the
safe use of the instrument and result in
Caution…
Conditions and actions that may compromise the
safe use of the instrument and result in slight
personal or material damage.
Read this manual carefully before using the
instrument!
In order to avoid any personal injuries and/or damage to
the instrument all safety instructions and warnings must be
observed.
Disposal – WEEE Directive
These calibrators contain Electrical and Electronic circuits
and must be recycled or disposed of properly (in
accordance with the WEEE Directive 2002/96/EC).
Warning
About the use:
• The calibrator mu
than those described in this manual, as it might cause a
hazard.
• The calibrator has been designed for indoor use only
and is not to be used in wet locations.
• The calibrator is not to be used in hazardous areas,
where vapour or gas leaks, etc. may constitute a
danger of explosion.
• The calibrator is not designed for operation in altitudes
above 2000 meters.
• The calibrator is a CLASS I product and must be
connected to a mains outlet with a protective earth
connection. Ensure the ground connection of the
calibrator is properly connected to the protective earth
before switching on the calibrator. Always use a mains
power cable with a mains plug that connects to the
protective earth.
• To ensure the connection to protective earth any
extension cord used must also have a protective earth
conductor.
• Only use a mains power cord with a current rating as
specified by the calibrator and which is approved for the
voltage and plug configuration in your area.
• Before switching on the calibrator make sure that it is
set to the voltage of the mains electricity supply.
•Always position the calibrator to enable easy and quick
disconnection of the power source (mains inlet socket).
• The calibrator must be kept clear within an area of 20
cm on all sides and 1 metre above the calibrator due to
fire hazard.
•Never use heat transfer fluids such as silicone, oil,
paste, etc. in the calibrators. These fluids may
penetrate the calibrator and cause electrical hazard,
damage or create poisonous fumes.
• The calibrator must be switched off before any attempt
to service the instrument is made. There are no user
serviceable parts inside the calibrator.
• When cleaning the well or insertion tube, REMEMBER
to wear goggles when using compressed air!
About the insertion tubes and well:
•Never leave hot insertion tubes which have been
emoved from the calibrator unsupervised – they may
r
constitute a fire hazard or personal injury.
• If you intend to store the calibrator in the optional
aluminium carrying case after use, you must ensure that
the instrument has cooled down to a temperature below 100°C/212°F before placing it in the carrying case.
• Never place a hot insertion tube in the optional carrying
case.
•Do not touch the well, the insertion tube or the grid
plate, as the calibrator is heating up / has been heated
up – they may be very hot and cause burns.
•Do not touch the tip of the sensor when it is removed
from the insertion tube/well – it may be very hot and
cause burns.
•Over 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
•Do not remove the insert from the calibrator before the
insert has cooled down to less than 50°C/122°F.
Caution – Cold surface
Below 0°C/32°F (applies only to the ETC-125 A models)
•Do not touch the well or insertion tube when these are
below 0°C/32°F - they might create frostbite.
• If the calibrator has reached a temperature below
0°C/32°F, ice crystals may form on the insertion tube
and the well. This, in turn, may cause the material
surfaces to oxidize
To prevent this from happening, simply heat up the
calibrator to 100°C/212°F until all water left has
evaporated.
It is very important that humidity in the well and insertion
tube is removed to prevent corrosion and frost
expansion damages.
• Do not use the instrument if the fan is out of order.
• Before cleaning the calibrator, you must switch it off,
allow it to cool down and remove all cables.
About the well, insertion tube and grid plate:
• The well and the insertion tube must be clean before
use.
•Do not pour any form of liquids into the well. It might
damage the well or cause a hazard.
• Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes
carefully when not in use.
• The insertion tube must never be forced into the well.
The well could be damaged as a result, and the
insertion tube may get stuck.
• The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation
on the insertion tube inside the instrument. There is a
risk that the insertion tube may get stuck if this is
allowed to happen.
Note…
The product liability only applies if the instrument is
subject to a manufacturing defect. This liability becomes
void if the user fails to follow the instructions set out in this
manual or uses unauthorised spare parts.
• The calibrator must not be used for any purposes other
than those described in this manual, as it might cause a
hazard.
• The calibrator has been designed for indoor use only
and is not to be used in wet locations.
• The calibrator is not to be used in hazardous areas,
where vapour or gas leaks, etc. may constitute a
danger of explosion.
• The calibrator is not designed for operation in altitudes
above 2000 meters.
• The calibrator is a CLASS I product and must be
connected to a mains outlet with a protective earth
connection. Ensure the ground connection of the
calibrator is properly connected to the protective earth
before switching on the calibrator. Always use a mains
power cable with a mains plug that connects to the
protective earth.
• To ensure the connection to protective earth any
extension cord used must also have a protective earth
conductor.
• Only use a mains power cord with a current rating as
specified by the calibrator and which is approved for the
voltage and plug configuration in your area.
• Before switching on the calibrator make sure that it is
set to the voltage of the mains electricity supply.
•Always position the calibrator to enable easy and quick
disconnection of the power source (mains inlet socket).
• The calibrator must be kept clear within an area of 20
cm on all sides and 1 metre above the calibrator due to
fire hazard.
Follow the instructions below before using the calibrator (cf. Fig. 1):
1. Place the calibrator on an even horizontal surface away from
all draughts.
Caution…
2. Check that the voltage shown on the label above the power
3. Plug in the cable below (pos. 3) the power control switch
4. The selected insertion tube is inserted into the calibrator
Do not use the instrument if the fan is out of order. Ensure
a free supply of air to the fan (pos. 5).
control switch is identical to the mains voltage (ETC-400
only). If not, the instrument should be returned to AMETEK
Denmark A/S or the distributor for replacement.
(pos. 4) and check that the earth connection is present.
(ETC-125 A only).
Caution…
• The well and the insertion tube must be clean before
use.
• The insertion tube must never be forced into the well.
The well could be damaged as a result, and the
insertion tube may get stuck.
•Do not pour any form of liquids into the well. It might
•Do not touch the well, the insertion tube or the grid
plate - they may be very hot and cause burns.
•Do not touch the tip of the sensor when it is removed
from the insertion tube/well – it may be very hot and
cause burns.
•Over 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
Warning……
•Never l
removed from the calibrator unsupervised – they may
constitute a fire hazard or personal injury.
eave hot insertion tubes which have been
•Do not remove the insertion tube from the calibrator
before the insertion tube has cooled down to less than
14 123943 08
Caution…
• The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation
on the insertion tube inside the instrument. There is a
risk that the insertion tube may get stuck if this is
allowed to happen.
• If the calibrator is to be transported, the insertion tube
must be removed from the well to avoid damage to the
instrument.
• Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes
carefully when not in use.
4.1 Switching off the calibrator
The following routine must be observed before the insertion tube is
removed and the instrument turned off (cf. Fig. 1) :
1. If the calibrator has been heated up to temperatures above
50°C/122°F, you must wait until the instrument reaches a
temperature below 50°C/122°F before you switch it off.
2. If the calibrator has reached a temperature below 0°C/32°F,
it should be heated momentarily to a temperature of
100°C/212°F.
3. Turn off the calibrator using the power control switch (pos.
4).
4. Remove the insertion tube from the calibrator using the tool
supplied with the instrument.
Congratulations on your new AMETEK Jofra ETC
Calibrator!
With the AMETEK Jofra calibrator, you have chosen an extremely
effective instrument which we hope will live up to all your
expectations. Over the past many years, we have acquired extensive
knowledge of industrial temperature calibration. This expertise is
reflected in our products which are all designed for daily use in an
industrial environment. Please note that we would be very interested
in hearing from you if you have any ideas or suggestions for changes
to our products.
This reference manual applies to the following instruments:
• Jofra ETC-125 A
• Jofra ETC-400 A
• Jofra ETC-400 R
ISO-9001 certified
-9001 certified in September 1994
by Bureau Veritas Certification Denmark.
CE-label
*
ETC-125 A from serial no. xxxxxx-01328, ETC-400 A / ETC-400 R from serial no. xxxxxx-01798
4 2019-04-12 123943 08
Your new temperature calibrator bears the CE label
and conforms to the Electromagnetic Compatibility
(EMC) Directive 2014/30/EU and the Low Voltage
Directive 2014/35/EU and RoHS Recast (RoHS II)
Directive 2011/65/EU*.
Please contact the dealer from whom you acquired the instrument if
you require technical assistance.
1.1 Warranty
This instrument is warranted against defects in workmanship,
material and design for two (2) years from date of delivery to the
extent that AMETEK will, at its sole option, repair or replace the
instrument or any part thereof which is defective, provided, however,
that this warranty shall not apply to instruments subjected to
tampering or, abuse, or exposed to highly corrosive conditions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES
WHETHER EXPRESS OR IMPLIED AND AMETEK HEREBY
DISCLAIMS ALL OTHER WARRANTIES, INCLUDING, WITHOUT
LIMITATION, ANY WARRANTY OF FITNESS FOR A PARTICULAR
PURPOSE OR MERCHANTABILITY. AMETEK SHALL NOT BE
LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES,
INCLUDING, BUT NOT LIMITED TO, ANY ANTICIPATED OR LOST
PROFITS.
This warranty is voidable if the purchaser fails to follow any and all
instructions, warnings or cautions in the instrument’s User Manual.
If a manufacturing defect is found, AMETEK will replace or repair the
instrument or replace any defective part thereof without charge;
however, AMETEK’s obligation hereunder does not include the cost
of transportation, which must be borne by the customer. AMETEK
assumes no responsibility for damage in transit, and any claims for
such damage should be presented to the carrier by the purchaser.
Read this manual carefully before using
the instrument!
Please follow the instructions and procedures described in
this manual. They are designed to allow you to get the
most out of your calibrator and avoid any personal injuries
and/or damage to the instrument.
Disposal – WEEE Directive
Th
ese calibrators contain Electrical and Electronic circuits
and must be recycled or disposed of properly (in
accordance with the WEEE Directive 2002/96/EC).
Warning
About the use:
• The calibrator mu
other than those described in this manual, as it might
cause a hazard.
• The calibrator has been designed for indoor use only
and is not to be used in wet locations.
• The calibrator is not to be used in hazardous areas,
where vapour or gas leaks, etc. may constitute a
danger of explosion.
• The calibrator is a CLASS I product and must be
connected to a mains outlet with a protective earth
connection. Ensure the ground connection of the
calibrator is properly connected to the protective earth
before switching on the calibrator. Always use a mains
power cable with a mains plug that connects to the
protective earth.
• To ensure the connection to protective earth any
extension cord used must also have a protective earth
conductor.
• Only use a mains power cord with a current rating as
specified by the calibrator and which is approved for
the voltage and plug configuration in your area.
• Before switching on the calibrator make sure that it is
set to the voltage of the mains electricity supply.
•Always position the calibrator to enable easy and quick
disconnection of the power source (mains inlet socket).
• The calibrator must be kept free within an area of 20
cm on all sides and 1 metre above the calibrator due to
fire hazard.
•Never use heat transfer fluids such as silicone, oil,
paste, etc. in the calibrators. These fluids may
penetrate the calibrator and cause electrical hazard,
damage or create poisonous fumes.
• The calibrator must be switched off before any attempt
to service the instrument is made. There are no user
serviceable parts inside the calibrator.
• When cleaning the well or the insertion tube,
REMEMBER to wear goggles when using compressed
air in the dry-block calibrator.
•Avoid knocking, bumping or dropping the instrument.
This can cause permanent damage to the instrument
and loss of accuracy.
•Do not use this instrument for any application other
than calibration work.
• The calibration instruments should only be used by
TRAINED PERSONNEL.
About the insertion tubes and well:
•Never leave hot insertion tubes which have been
emoved from the calibrator unsupervised – they may
r
constitute a fire hazard or personal injury.
• If you intend to store the calibrator in the optional
aluminium carrying case after use, you must ensure
that the instrument has cooled down to a temperature
below 100°C/212°F before placing it in the carrying
case.
• Never place a hot insertion tube in the optional carrying
case.
Caution – Hot surface
This symbol is engraved in the well of the ETC-400 A and
in the grid plate of the ETC-125 A / 400 R.
•Do not touch the well, the insertion tube or the grid
plate, as the calibrator is heating up / has been heated
up – they may be very hot and cause burns.
•Do not touch the tip of the sensor when it is removed
from the insertion tube/well – it may be very hot and
cause burns.
•Over 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
•Do not remove the insert from the calibrator before the
insert has cooled down to less than 50°C/122°F.
Caution – Cold surface
Below 0°C/32°F (applies only to the ETC-125 A models)
Do not touch the well or insertion tube when these are
•
below 0°C/32°F - they might create frostbite.
• If the calibrator has reached a temperature below
0°C/32°F, ice crystals may form on the insertion tube
and the well. This, in turn, may cause the material
surfaces to oxidize
To prevent this from happening, simply heat up the
calibrator to 100°C/212°F until all water left has
evaporated.
It is very important that humidity in the well and insertion
tube is removed to prevent corrosion and frost
expansion damages.
• Do not use the instrument if the fan is out of order.
• Before cleaning the calibrator, you must switch it off,
allow it to cool down and remove all cables.
About the well, insertion tube and grid plate:
• The well and the insertion tube must be clean before
use.
•Do not pour any form of liquids into the well. It might
damage the well or cause a hazard.
• Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes
carefully when not in use.
• The insertion tube must never be forced into the well.
The well could be damaged as a result, and the
insertion tube may get stuck.
• The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation
on the insertion tube inside the instrument. There is a
risk that the insertion tube may get stuck if this is
allowed to happen.
123943 08 2019-04-12 9
Note…
The product liability onl
to a manufacturing defect. This liability becomes void if the
user fails to follow the maintenance instructions set out in
this manual or uses unauthorised spare parts.
• The calibrator has been designed for indoor use only
and is not to be used in wet locations.
• The calibrator is not to be used in hazardous areas,
where vapour or gas leaks, etc. may constitute a
danger of explosion.
• The calibrator is not designed for operation in altitudes
above 2000 meters.
• The calibrator is a CLASS I product and must be
connected to a mains outlet with a protective earth
connection. Ensure the ground connection of the
calibrator is properly connected to the protective earth
before switching on the calibrator. Always use a mains
power cable with a mains plug that connects to the
protective earth.
• To ensure the connection to protective earth any
extension cord used must also have a protective earth
conductor.
• Only use a mains power cord with a current rating as
specified by the calibrator and which is approved for the
voltage and plug configuration in your area.
• Before switching on the calibrator make sure that it is
set to the voltage of the mains electricity supply.
•Always position the calibrator to enable easy and quick
disconnection of the power source (mains inlet socket).
• The calibrator must be kept clear within an area of 20
cm on all sides and 1 metre above the calibrator due to
fire hazard.
•Never use heat transfer fluids such as silicone, oil,
paste, etc. in the calibrators. These fluids may
penetrate the calibrator and cause electrical hazard,
damage or create poisonous fumes.
When setting up the calibrator, you must…(see fig. 1)
diameter and
125 A only)
or use one of the bores in the well that best suits your needs.
See section 3.3 for information on how to select the insertion
Place the calibrator on an even horizontal surface in the spot
you intend to use it.
Caution…
The calibrator is now ready for use.
Section 3.3 and 3.4 are for ETC-125 / 400 A only
Do not use the instrument if the fan is out oforder.
Ensure a free supply of air to the fan.
Check that the voltage shown on the label above the power
control switch is identical to the mains voltage (ETC-400 A/R
only). If not, the instrument should be returned to AMETEK
Denmark A/S or the distributor for replacement.
Check that the earth connection for the instrument is present
d attach the cable.
an
Select an insertion tube with the correct bore
place it in the largest bore of the instrument (ETC-
tube/ correct bore in the well.
3.3 Choice of insertion tube / correct bore in
well.
Caution…
To get the best results out of your calibrator, the insertion
tube dimensions, tolerance and material are critical. We
highly advise using the JOFRA insertion tubes, as they
guarantee trouble free operation. Use of other insertion
tubes may reduce performance of the calibrator and cause
the insertion tube to get stuck.
www.GlobalTestSupply.com
Page 42
Caution…
Before using new insertion tubes for calibration in the ETC-
Insertion tubes are only available for the ETC-125 A.
An insertion tube is selected on the basis of the diameter of the
sensor to be calibrated. You may choose between an 8 mm / 5/16
inch insertion tube, an 3/8 inch insertion tube or alternatively, you
may order an undrilled insertion tube and drill the required hole
yourself. The finished dimension should be as follows:
• Sensor diameter +0.2 +0.05/-0 mm.
bore is selected on the basis of the diameter of the sensor to be
A
calibrated. The sensor can now be inserted directly into the bore of
the well. See Fig. 2 for various design combinations.
125 instrument, the insertion tubes must be heated up to
maximum temperature 125°C (257°F) for a period of
minimum 30 minutes.
•The well and the insertion tube must be clean before
use.
•Scratches and other damage to the insertion tube
should be avoided by storing the insertion tube carefully
when not in use.
•The insertion tube must never be forced into the well.
The well could be damaged as a result, and the
insertion tube may get stuck.
Caution – Hot surface
•Do not touch
tube while the calibrator is heating up – they may be
very hot and cause burns.
•Do not touch the tip of the sensor when it is removed
from the insertion tube – it may be very hot and cause
burns.
•Do not remove the insert from the calibrator before the
insert has cooled down to less than 50°C/122°F.
Caution – Cold surface
Below 0°C/32°F (applies only to the ETC-125 A models)
•Do not touch the well or insertion tube when these are
below 0°C/32°F - they might create frostbite.
• If the calibrator has reached a temperature below
0°C/32°F, ice crystals may form on the insertion tube
and the well. This, in turn, may cause the material
surfaces to oxidize
To prevent this from happening, simply heat up the
calibrator to 100°C/212°F until all water left has
evaporated.
CHECKMARK displayed when the calibrator is stable.
AUTO STEP symbol used to indicate that the function is
active (symbol flashes repeatedly).
Used to display set-temperatures, time-until-stable and
parameter values in the menu system.
Minute time unit for bottom display.
Celsius temperature unit for bottom display.
Fahrenheit temperature unit for bottom display.
Fahrenheit temperature unit for top display.
Celsius temperature unit for top display.
Used to display Read-temperature and parameters in the
The top display continuously shows the read-temperature.
°C
°C
The bottom display shows either the set-temperature or the
estimated time in whole minutes until the calibrator will be
stable:
min
When the calibrator is stable the display will show the
checkmark symbol. The instrument will emit an audible
alarm and the estimated time until stable will be replaced by
the set-temperature:
AUTO STEP is used to step automatically between a range of
different calibration temperatures. This is useful when calibrating
sensors in places which are hard to reach, and when calibrating
sensors for which the output is displayed in a different location.
The function can be illustrated using the following example:
Fig. 6
The symbol for AUTO STEP flashes to indicate that
Press.
the function is active.
The function can be cancelled at any time by pressing
The following will be displayed for one second to indicate that the
calibrator is ready to work towards the set-temperature:
°C
The calibrator will now work towards the given set-temperature. An
audible alarm will be emitted once the calibrator is stable.
The calibrator will wait the specified amount of extra time. The
instrument indicates this by counting down the amount of time
remaining:
min
The calibrator will then go to the next step. The procedure is the
same as for the first step. This process will be repeated until the last
step has been executed and the function has been completed.
If the current set-temperature is higher than the new max-
Press or to set the max-temperature in steps
of 0.1°C or 0.1°F:
°C
temperature, you will need to adjust the set-temperature.
The instrument will immediately begin to cool (if required) as
soon as the new max-temperature is accepted.
ess
Pr
If you wish to exit SETUP, simply press
to accept your selection.
ESC
MENU
.
4.5.3 Adjusting the extra stability time
The extra stability time is the amount of extra time you wish to elapse
before the checkmark symbol is displayed after the calibrator has
stabilised.
When the calibrator is stable, the display will show the
checkmark symbol. The instrument will emit an audible
alarm and the calibrator is now ready for use.
°C
Select the emission factor of the target unit on the IR
thermometer to be calibrated. If your IR-thermometer has a
fixed emissivity setting, use the IRLAB software to calculate
corrections.
Point the IR thermometer perpendicular towards the target
unit surface.
It is very important that the thermometer is held close to the
target unit surface.
Ensure the thermometer’s “field of vision” is centred and contained
within the surface area of the target unit. See fig. 7.
If the calibration is not carried out as described, the result of the
calibration will not be accurate.
Caution…
Avoid contact to the target unit surface with the
thermometer during operation, as the thermometer
accuracy will be affected and the thermometer may
overheat or be damaged. Recommended operating
distance is between 2,5 – 5 cm / 1 – 2”.
Switch off the calibrator using the power control switch.
Note that the calibration procedure may be interrupted at any time
using the power control switch. Switching off the calibrator during the
calibration process will not damage either the instrument or the
sensor.
Caution…
The following guidelines should always be observed when
storing and transporting the calibrator. This will ensure that
the instrument and the sensor remain in good working
order.
ETC-125 A only.
The following routine must be observed before the insertion tube is
removed and the instrument switched off:
Caution – Hot surface
•Over 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
•Do not remove the insertion tube from the calibrator
before the insertion tube has cooled down to less than
50°C/122°F
•Never leave hot insertion tubes which have been removed
from the calibrator unsupervised – unauthorized
personnel might pick up insertion tubes causing severe
burns.
•Never place a hot insertion tube in the optional carrying
case.
Below 0°C/32°F
•Do not touch the well or insertion tube when these are
below 0°C/32°F - they might create frostbite.
• If the calibrator has reached a temperature below
0°C/32°F, ice crystals may form on the insertion tube and
on the well. This, in turn, may cause the material
surfaces to oxidize.
To prevent this from happening, the insertion tube and
the well must be dried. This is done by heating up the
calibrator to min. 100°C/212°F until all water left has
evaporated.
It is very important that humidity in the well and insertion
tube is removed to prevent corrosion and frost expansion
damages.
Remove the insertion tube from the calibrator using the tool for
insertion tube supplied with the instrument as shown in Fig. 8.
ETC-400 R only
When the ETC-400 R is not in use, it is advisable to store the
calibrator in the carrying case.
Caution…
• The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidationon
the insertion tube inside the instrument. There is a risk
that the insertion tube may get stuck if this is allowed to
happen.
• The insertion tube must be removed to avoid damage to
the instrument if the calibrator is to be transported long
distances.
Caution – Hot surface
• Ove
• You must ensure that the instrument has cooled down to
36 2019-04-12 123943 08
r 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
a temperature below 100°C/212°F before placing it in
the carrying case.
When returning the calibrator to the manufacturer for service, please
enclose a fully completed service information form. Simply copy the
form on the following page and fill in the required information.
The calibrator should be returned in the original packing.
Customer data:Date:
Customer name and address:___________________________________________
Attention and Dept.:___________________________________________________
Fax no./Phone no.:____________________________________________________
Your order no.:_______________________________________________________
Delivery address:_____________________________________________________
Distributor name:_____________________________________________________
Instrument data:
Model and Serial no.:__________________________________________________
Warranty claimed Yes:____ No:_____ Original invoice no.:_________________
___ Calibration as left ___ Check
___ Calibration as found and as left ___ Service
___ Accredited calibration as left ___ Repair
___ Accredited calibration as found and as left.
Before cleaning the calibrator, you must switch it off, allow
it to cool down and remove all cables.
Caution – Hot surface
•Over 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
Caution – Hot surface – ETC-125 A only
• Do
• Never leave hot insertion tubes which have been removed
• Never place a hot insertion tube in the optional carrying
Below 0°C/32°F
not remove the insertion tube from the calibrator
before the insertion tube has cooled down to less than
50°C/122°F
from the calibrator unsupervised – unauthorized
personnel might pick up insertion tubes causing severe
burns.
case.
•Do not touch the well or insertion tube when these are
below 0°C/32°F - they might create frostbite.
• If the calibrator has reached a temperature below
0°C/32°F, ice crystals may form on the insertion tube and
on the well. This, in turn, may cause the material
surfaces to oxidize.
To prevent this from happening, the insertion tube and
the well must be dried. This is done by heating up the
calibrator to min. 100°C/212°F until all water left has
evaporated.
It is very important that humidity in the well and insertion
tube is removed to prevent corrosion and frost expansion
damages.
Caution – ETC-125 A only
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation on the
insertion tube inside the instrument. There is a risk that the
insertion tube may get stuck if this is allowed to happen.
Users should/must carry out the following cleaning procedures as
and when required:
•The exterior of the instrument - Clean using water and a soft
cloth.
The cloth should be wrung out hard to avoid any water
penetrating the calibrator and causing damage.
The keyboard may be cleaned using isopropyl alcohol when
heavily soiled.
•The insertion tube (ETC-125 A only) - Must always be clean
and should be regularly wiped using a soft, lint-free, dry cloth.
You must ensure there are no textile fibres on the insertion tube
when it is inserted in the well. The fibres may adhere to the well
and damage it.
•The well - Must always be clean. Dust and textile fibres should
be removed from the well using e.g. compressed air.
Warning
REMEMBER to wear goggles when using compressed air!
•The target unit surface (ETC-400 R only) - Must always be
clean. Dust and textile fibres should carefully be blown from the
surface with clean air. Do not use fluid to clean the target unit
surface.
If it is impossible to clean the surface with clean air only the
target unit could alternatively be heated up to max. temperature.
The dust and fibres will then be burned off the surface. There is,
however, the risk that the dust and fibres could permanently burn
into the surface, altering the emission factor.
You are advised to return the calibrator to AMETEK Denmark A/S or
an accredited laboratory at least once a year for calibration and
adjustment.
Alternatively, you can calibrate/adjust the calibrator yourself. You will
need a reference thermometer and a reference sensor with a
traceable calibration certificate. Please follow the instructions given
below.
Connect the calibrator to an external precision instrument (e.g. a DTI)
as shown in Fig. 9:
DTI
Sensor
Fig. 9
ESC
Hold down the button while pressing the on/off power
control switch.
If the calibrator is found to be within the permitted tolerances, the
instrument will display the text =OK at the top of the display. The text
Cont. will flash in the bottom of the display to indicate that you may
continue without adjustments:
If the instrument detects excessive deviations for one or more steps,
it will show a screen reading =ERR. in the top of the display. The text
AdJ. will flash in the bottom of the display to indicate that an
adjustment is required. Accept by pressing .
Press
Press
to repeat an adjustment step when it is shown on the
display.
Press
display.
Press
coefficients. Next, repeat the entire calibration/adjustment
procedure.
In order to calibrate the inner thermometer of the calibrator, a 3 mm
insertion hole for the reference sensor has been drilled 1 mm
beneath the target unit surface. See fig. 10.
Insertion hole for
th
e reference sensor
Fig. 10
For the calibration it is recommendable to use the RTD reference
sensor 65-Pt100-150 SPEC. which AMETEK DENMARK A/S has
developed specifically for this instrument.
In principle any sensor fitting the insertion hole is useable, however,
there is the risk of incorrect temperature reading due to the limited
insertion depth of the reference sensor hole.
All parts listed in the list of accessories can be obtained from the
factory through our dealers.
Please contact your dealer for assistance if you require parts which
do not appear on the list.
List of accessories
Accessories Parts no.
User and reference manual 123943
Tool for insertion tube (ETC-125 A only) 60F172
Aluminium carrying case 124094
Mains cable, 115V, US, type B 60F135
Mains cable, 240V, UK, type C 60F136
Mains cable, 220V, South Africa, type D 60F137
Mains cable, 220V, Italy, type E 60F138
Mains cable, 240V, Australia, type F 60F139
Mains cable, 230V, Europe, type A 60F140
Mains cable, 230V, Denmark, type G 60F141
Mains cable, 220V, Switzerland, type H 60F142
Mains cable, 230V, Israel, type I 60F143
RS232 serial cable 2 m 123958
5 x undrilled insertion tubes for ETC-125 A 123939
8 mm / 5/16 inch insertion tube for ETC-125 A 123938
3/8 inch insertion tube for ETC-125 A 124045
Shoulder strap with snap hooks 124004
JOFRACAL PC software 124915
JOFRA IR-LAB calibration software for ETC-400 R 124591
Correction table for emissivity for ETC-400 R 124592
Pt100 sensor ø3mm x 150mm for ETC-400 R 65-PT100-150 SPEC