Joerns Healthcare RC-Ultra-Hi-Lo-850-Manual1 User-Service Manual Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed Spread Base Bariatric Low Bed Model RCUHLB850

Page 1
User-Service Manual
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed
Spread Base Bariatric Low Bed
Model RCUHLB850
To avoid injury, read user’s manual before using.
TM
Page 2
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Important Precautions
Caution: Rotating parts. Keep ngers and arms clear of bed when mechanism is moving.
Caution: Pinch points. Keep objects away from these areas while bed is in motion.
Caution: Electrical hazard may occur if bed is plugged into inadequate power supply. Make sure to use a 115 volt, 3 prong, grounded outlet with a capacity of at least 15 amps.
Caution: Keep feet clear of casters. Make sure casters are locked before transferring patients.
Caution: Make sure that the cords from the control module and pendant are properly positioned before operating the bed. Do not allow either cord to get caught in the moving parts of the bed.
Caution: When using a turning mattress, make sure there is a bolster between the mattress and side rail.
Warning: Keep all objects out from under bed. Certain parts under the bed raise and lower when the bed is in motion.
Warning: Do not place arms or head into frame while operating. Stand clear of the bed frame before
operating.
Warning: Do not operate this device if the power cord or any of the cords coming from the control module are cut, frayed, damaged or loosely connected.
Warning: Do not service the bed without rst disconnecting the power cord.
Warning: The bed must be unplugged from the wall socket prior to using the emergency crank. Failure to do so may result in damage to the bed and injury to the personnel using the crank.
Warning: ESD - electrostatic discharge.
Warning: The bed should be left in the lowest position
when unattended in order to reduce the risk of injury due to falls while getting into or out of bed, or while lying on the bed.
Warning: Possible Fire Hazard. Use nasal mask or ½ bed tent oxygen administering equipment. Oxygen tent should not extend below mattress support platform. Pendant should not be placed in an oxygen enriched environment such as an oxygen tent. Use of electrical circuits in an oxygen-enriched environment could result in a re hazard.
Warning: Possible Injury Or Death. An optimal bed system assessment should be conducted on each resident by a qualied clinician or medical provider to ensure maximum safety of the resident. The assessment should be conducted within the context of, and in compliance with, the state and federal guidelines related to the use of restraints and bed system entrapment guidance, including the Clinical Guidance for the Assessment and Implementation of Side Rails published by the Hospital Bed Safety Workgroup of the U.S. Food and Drug Administration. Further information can be obtained at the following web address: http://www.fda.gov/ MedicalDevices/ProductsandMedicalProcedures/ GeneralHospitalDevicesandSupplies/HospitalBeds/ default.htm.
Warning: Possible Injury Or Death. Use a properly sized mattress in order to minimize the gap between the side of mattress and Side Rails/Assist Devices. This gap must be small enough to prevent resident/ patient from getting his/her head or neck caught in this location. Make sure that raising or lowering bed, or contouring the mattress support platform, does not create any hazardous gaps. Excessive gaps may result in injury or death.
Warning: Possible Injury Or Death. Residents/ patients may become entangled in pendant cord. Patients with reduced mental acuity should not be allowed access to pendant. Unsupervised use of pendant could result in injury or death.
Caution: Before transporting bed, unplug bed from power source, secure Power Cord on cord storage straps on Head Panel and hang Pendant from Head Panel. Failure to do so could result in Power Cord, pendant cord, or bed damage, thereby creating a potential hazard.
Caution: Extend all four sections on each side prior to adjusting bed position or side rails.
Warning: When raising or lowering bed, casters at the foot end should be in the steer position. Head end should be free to move.
2
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Warning: Use Width Extenders only after the resident/ patient is safely positioned in the bed. Failure to do so could result in injury.
Page 3
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
Caution: Drainage Bag Loops are not intended for use as tie downs.
Warning: Possible Injury Or Death. Bed safe working load is 850 pounds. This is total weight counting resident/patient, mattress, bedding and any other equipment or persons likely to be on bed. Do not exceed 850-pound safe working load. Exceeding the safe working load could result in property damage, injury or death.
Warning: Possible Injury Or Death. Do not use any Side Rails/Assist Devices until you verify they are locked in place. Failure to lock assist devices may result in injury or death.
Notice: Make sure the power cord is unplugged from the power outlet prior to disconnecting/connecting any foot panel cables. Failure to ensure bed is not powered when disconnecting/connecting the foot panel cables may result in damage to the bed’s electronic controls.
Warning: Possible Injury Or Death. If using accessories not manufactured, marketed, or provided by Joerns Healthcare for Camtec beds, consult with the manufacturer for compatibility and limitations prior to use. Failure to do so may result in injury or death.
Bed System Entrapment Information
In April 1999, the U.S. Food and Drug Administration (FDA) in partnership with representatives from the hospital and post-acute bed industry, including Joerns Healthcare, national healthcare organizations, resident advocacy groups, and other federal agencies formed the Hospital Bed Safety Workgroup (HBSW). The workgroup’s goal is to improve the safety of bed frames for patients and residents in all health care settings who are most vulnerable to the risk of entrapment. The efforts of the FDA and the HBSW culminated in the FDA’s release of recommended guidelines intended to reduce the risk of entrapment, including dimensional limits for critical gaps and spaces between bed system components and clinical guidance for assessment and implementation of bed side rails in various health care settings.
Entrapment zones involve the relationship of bed components often directly assembled by the healthcare facility rather than the manufacturer. Therefore, compliance is the responsibility of the facility.
As the leading manufacturer of long-term care beds and a frontrunner in addressing this critical issue, Joerns Healthcare can offer you the expertise, assistance and products to bring your facility into compliance.
Joerns® Compliance Solutions
Matching the right bed components in order to meet regulatory guidelines can be complex.
That is why Joerns offers a wide array of compliance options. We assist customers in selecting compliant accessories recommended for their specic bed model.
For More Information
To learn more about compliance options with Joerns products, visit our website at www.joerns.com, or contact our Customer Care reps at 800-826-0270 and ask for free informational publications.
To learn more about entrapment zones, assessment methods and guidelines concerning entrapment, contact Joerns Healthcare at 800-826-0270 or consult the FDA website: http://www.fda.gov/ MedicalDevices/ProductsandMedicalProcedures/ GeneralHospitalDevicesandSupplies/HospitalBeds/ default.htm.
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
3
Page 4
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Table of Contents
Important Precautions .................................................................................................................2
Entrapment Information ...............................................................................................................3
Introduction ..................................................................................................................................4
Specications ..............................................................................................................................5
Operation .....................................................................................................................................6
Cleaning and Maintenance ..........................................................................................................8
Troubleshooting ...........................................................................................................................9
Customer Maintenance Checklist ..............................................................................................10
Appendix A.................................................................................................................................12
Accessories: Scale Instructions and Bed Exit Alarm .................................................................28
Warranty ....................................................................................................................................34
Introduction
The Joerns Camtec Series RC Ultra Hi-Lo Bed (RCUHLB850) has been designed primarily for use in hospitals and other institutions. The frame has a long histrory as the Camtec model 403649-S supporting care of some of the most challenging patients presented within the care continuum. The unique expandable frame design allows the bed to accommodate support surfaces of up to 48" wide. The bed surface can be adjusted to allow it to pass through a 40" door opening.
The frame is constructed of structural steel. The frames lowest position is approximately 9½" from the oor to the mattress support platform. The frame in the highest position is approximately 29".
The frame is designed to support a safe working load of 850 lbs. Side rails at the head end and foot end are standard. The rotation of the side rails assists in patient ingress and egress.
The trapeze is an available option using a full-frame design purpose built to support bariatric patients for this type of bed and allow convenient adjustability for repositioning. The bed is equipped with (3) locking swivel casters, (1) steering swivel caster and 4 small swivel casters to allow for easy bed movement and facilitate patient transport.
The bed contains four electric drive modules, one for the head, one for the knee, one Hi/Lo for the head end and one Hi/Lo for the foot end of the bed. A hand crank is supplied at the head end of the bed that can be used in the event of a power failure.
4
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 5
Specifications
Safe Working Load .................. 850 lbs
Overall Width ........................... 40" expandable to 52"
Overall Length ......................... 94"
Mattress Support Platform ..... Width: Adjustable 36", 40", 42", and 48"
Length: 80"
Mattress Support Platform Height ..... Lowest position (from oor) 9½"
Highest position 29"
Power Requirements ............... 115 VAC 60Hz 10.5 AMP
Operating Conditions
Ambient Temperature .............. +10°C to +30°C
Relative Humidity .................... 30% to 95% Non-Condensing
Atmospheric Pressure ............ 700 hPa to 1060 hPa
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
Storage and Shipping Conditions
Ambient Temperature .............. -30°C to +50°C Relative
Humidity ................................... 10% to 95%
Atmospheric Pressure ............ 500 hPa to 1060 hPa
Bed Weight with Side Rails, End Boards ........ 500 lbs.
Motor Duty Cycle for Intermittent Operation .. 1 minute ON / 10 minutes OFF
Head Fowler ............................. Adjustable up to 70 degrees
Foot/Knee Gatch ...................... Adjustable up to 40 degrees
Trendelenburg .......................... Adjustable up to 12 degrees
Reverse Trendelenburg........... Adjustable up to 12 degrees
Side Rails ................................. Head end and foot end are standard, Rotation type
Casters ..................................... Diameter 4" ...... Width 2⅜"
(3) total lock and (1) steer lock
Diameter 3" ...... Swivel (4)
Options .................................... Trapeze
Mattress Foot Extension
5
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 6
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Operation
1. When raising or lowering the bed, the casters at the head end should be locked and the casters at the foot end should be in the steer position. If the casters are locked at both ends at the same time, when raising or lowering the bed, damage to the bed may result.
2. Plug the power cord into a grounded 115 volt socket having a capacity of at least 15 amps. Make sure light on control module is illuminated.
3. Using the hand pendant, cycle the bed through all of its operations. Make sure that everyone is clear of the moving parts during these operations.
4. Use this User-Service Manual to instruct all personnel who are involved in the use of this model bed.
5. When transferring the patient into or out of the bed all four casters should be in the locked position.
6. Width section adjustments are made by pulling tether and sliding out extension.
Caution: Extend all four sections of each side prior to adjusting bed position or side rails.
7. Operate side rail using green button and rotating rail to ensure locked in place in up position or rotate down as needed.
Notice: Make sure the power cord is unplugged from the power outlet prior to disconnecting/connecting any foot panel cables. Failure to ensure bed is not powered when disconnecting/connecting the foot panel cables may result in damage to the bed’s electronic controls.
Casters
The bed is equipped with (4) locking casters consisting of (3) total lock & (1) steer lock and (4) swivel casters. The steer lock caster is located on the left foot position and will lock the direction of the caster inline (end to end) with the bed when activated. This allows for better control when transporting the patient. For best control the bed should be pushed or pulled from the head end.
Never lock casters at both ends at the same time, when raising or lowering the bed, as permanent damage to the bed may result. All casters should be locked during ingress or egress of patient.
Foot End Caster
6
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Head End Caster
Page 7
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
Five Function Hand Pendant Operation
The HEAD up and down buttons cause the head of the bed to raise and lower.
The FOOT up and down buttons cause the knees and lower portion of the legs (Knee Gatch) to be raised and lowered.
The BED up and down buttons cause the height of the bed to raise and lower.
The SEAT up button causes the bed to go into a comfort chair position. The head section and head end of the bed will raise, the knee section will raise and the foot end of the bed will lower.
The SEAT down button will cause the bed to return to a supine position
The TREND down button will cause the foot end of the bed to raise and the head end to lower with a limiting angle of 12 degrees.
The reverse TREND up button will cause the foot end of the bed to go down and the head end of the bed to raise until reaching a limit of 12 degrees.
The LOCK OUT ACTIVE light indicates that a Staff Control Lockout feature is active. The hand pendant cannot override the Staff Control
When using the SEAT or TREND functions and want to return to a level position, it is often easiest to lower the bed to its lowest point.
The bed contains four electric drive modules. A hand crank is also supplied, at the head end of the bed; which can be used to adjust the bed in the case of a power or actuator failure. Disconnect the power before using the crank. The pendant should not be used during hand cranking.
Sta Control Pendant Operation
Caution: Keep air therapy pumps clear of staff control.
The staff control pendant is located on the outside of the foot end board. These controls let the staff lock out any single, combination or all of the bed functions.
The functions of the Universal Staff Control operate the same as the Five Function Hand Pendant, but there are some additional functions on the Universal Staff Control.
There is a LOCK OUT feature. Press and hold the Lock Out button for 2 seconds to enable or disable the Lock Out feature. The Lock Out activation will be indicated on the Universal Staff Control Pendant.
The CPR feature overrides all Lock Outs and returns the bed to a supine position.
7
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 8
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Cleaning and Maintenance
Cleaning
Warning: High pressure cleaning or power wash are not to be used on any frame part.
Warning: Lock casters when cleaning, inspecting or performing any maintenance on the bed. Lower all mattress
support platform sections unless necessary for service access.
Before cleaning any portion of the bed, disconnect the power cord from the wall outlet
1. Using standard cleaning procedures, wipe down the frame of the bed. Be sure to wear appropriate protection for the type of materials being used.
2. Use cleaning detergents sparingly around the pendant controller, electronic controller and the motors.
Periodic Inspection and Maintenance
(See Customer Maintenance Check List)
The amount of maintenance required by the bed will be dictated by its use. As a minimum the unit should be periodically inspected every 6 months.
1. After a week or two of usage, check all the bolts to be sure that none of them have worked loose and that all pins are in their normal location and securely fastened. Check all welds.
2. At least once every six months the actuator drive screw should be lubricated using a heavy duty grease such as Drydene #5871 or equivalent.
3. Put light machine oil on all pivot points every 6 months.
4. Check all electrical connections for tightness.
5. Check all electrical wiring for fraying, kinking damage and/or deterioration.
IF ANY DISCREPANCIES ARE NOTED DURING INSPECTION, THEY MUST BE CORRECTED BEFORE CONTINUED USE OF THE BED.
Joerns Healthcare Technical Support can provide additional assistance to repair or troubleshoot any problems.
Major service and repair must only be performed by authorized Joerns Healthcare Technical Support, which you may contact at 800.966.6662.
8
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 9
Troubleshooting
Problem Possible Cause Solution
No functions operate No power to unit
Power light does not light
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
Electrical cords not properly plugged
into control module
Check that the power cord is plugged
into receptacle and light on control
module is illuminated.
Check for cord breaks
Check all electrical cords
and connections.
Power light is on but no functions
operate
Three motors operate “OK” One
motor will not operate
If the above does not resolve the problems, call Joerns Healthcare Technical Support at 800.966.6662. Have the
serial number of the bed and make the phone call from as close to the bed as practical.
Bad pendant Replace pendant
Bad control module Replace control module
Bad pendant
Bad control module
Bad motor
Bad wiring
Interchange motor connections on control module. Swap connections
on the module so that problem motor
is plugged into a motor connection
on the module that is working. If the
the motor that was OK now does not
operate, it indicates either the control
Isolate the Problem
problem motor now operates and
module or the pendant is faulty.
Replace as required.
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
9
Page 10
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Customer Maintenance Check List
S/N Bed____________________________________ S/N Controller__________________________________
Bed Model__________________________________ Date_____________________
Initial Date
1 Verify what options are on the bed:
Foot End Side Rails Scale
Head End Side Rails Trapeze
Pads (If Applicable) I.V. Pole
Other______________________________________________________________
2 Verify and record the S/N
3 Check Head Section operation with pendant
4 Check Knee Gatch operation with pendant
5 Check the Bed Elevation operation with pendant
6 Check the Trendelenburg operation with pendant (If Applicable)
7 Check the Seat operation with the pendant (If Applicable)
8 Check casters function: Steer________ Total________
9 Inspect the hardware to ensure that components are t securely
10 Check to ensure that the Electrical components are secure
11 Check Power Cord for the following:
A: Check all strain reliefs and cable clamps are secure
B: Check for worn areas, cuts or frays
C: Check plug for bent or loose prongs
12 Check that Pendant cord has no damage and that it is secured properly to the control box
13 Check to ensure that all decals are afxed at proper location
Head, Foot and Motor decals
Power and Control Cord on Head Board
Power and Control on the Control Box
14 Check the presence and t of the Mattress Stop
15 Check the presence, t and operation of the Emergency Crank
16 Check the operations of all accessory options noted in #1 above
17 Check to ensure that the side rails function properly and lock in place
18 Measure and record the Leakage Current value should be less than 250 microamperes
(UL 60601-1) Use Calibrated Model LT544D-20 Digital Safety Analyzer
Measured Value____________
19 Measure and record Line Grounding Resistance. Value should read equal to or less than 0.1
OHMS as measured between the ground wire in the line cord and the bed frame.
Measured Value____________
20 Visually inspect the bed for paint or assembly defects; touch up as needed
21 Check the Top Expansion (If Applicable) of the Mattress during operation at:
Head Section Right ________ Left ________
Center Section Right ________ Left ________
Upper Knee Gatch Right ________ Left ________
Lower Knee Gatch Right ________ Left ________
10
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 11
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
Customer Maintenance Check List Continued...
22 Lubricate all pivot points with a light machine oil. (This should be done approx. every 6 months)
23 Lubricate the actuator drive screws using a heavy duty grease such as Drydene #5871 or
equivalent. (This should be done approx. every 6 months)
24 If bed is a Scale Bed, check that the scale functions are operating properly
25 Check that all scale/exit alarm connections are secure
26 Test bed exit alarm
Comments
Initial Date
Inspector Signature_______________________________________________ Date_________________________
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
11
Page 12
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Wiring Diagram
12
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 13
Appendix A - Main Frame Assembly
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
13
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 14
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Mattress Support Assembly
14
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 15
Appendix A - Basic Bed Assembly
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
15
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 16
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Side Boards and Hardware
16
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 17
Appendix A - Axle Assembly, Head End
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
17
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 18
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Axle Assembly, Foot End
18
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 19
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Scale System Assembly FG 1250 Base Scale
19
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 20
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Scale System Assembly FG1257 with Bed Exit/Nurse Call
20
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 21
Appendix A - Label Location
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
21
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 22
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Labels Used on Bed
22
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 23
Appendix A - Parts List
Item Part No. Description
1 107-65-0 Plate, Caster Lock Transmission Pivot
2 107068-0 Plate, Caster Lock Link
3 203649-0 Weldment, Caster Cap RH
4 203648-0 Weldment, Caster Cap LH
5
6
7 203526-0 Weldment, Caster Lever Mounting Rod
8 203533-0- Weldment, Caster Lock Transmission Shaft
9 203602-0 Weldment, Axle, Scale
10 24006640 Weldment, Scale Pivot Axle
11 900085-0 Washer, Lock, ¼" Internal Tooth
12 900149-0 Bolt, Hex, ¼-20 x 1-¼
13 900328-S Screw, 10-32 UNF x ¼ SS Pan Head Phillips
14 900358-0 Retaining Ring
15 900411-0 Nut, ½-13 Hex, ESNA
16 900419-0 Washer, Nylon, ½ x ¾ x ¹/
17 903458-0 Caplug, 1¼" SQ
18 903630-0 Screw, 6mm x 10mm Pan Head Phillips
19 903631-0 Bushing, Flanged, ⅜" ID x ½" OD
20 903632-0 Bushing, Flanged, ¾" ID x 1" x ⅝" LG
21 903664-0 Bushing, Sleeve, ½" ID x ¾" OD x ⅝" LG
22 903785-0 Caster, 3" w/Stud, Swivel
23 903931-0 Caster, Total Lock
24 903932-0 Caster, Steer Lock
25 903994-0 Lever, Caster Locking
26 903995-0 Plug, Caster Lever, Red
27 903996-0 Plug, Caster Lever, Green
28 904017-0 Nut, 5/16-18 Hex, JAM, ESNA
29 107312-0 Stud, Load Cell Mounting
30
31 901233-0 Shoulder Bolt, Load Cell, ½" DIA x 1.250" LG
32 903699-0 Screw, ⅜-16 x 1.000" LG Socket Head Cap
33 104592-0 Spacer, ½" x 9/16"
34 903925-0 Casting, Hinge Half
35 106071-0 Spacer, 0.937"
36 106072-0 Spacer, 0.402" x 0.625" OD
37 106141-0 Spacer, 2.032"
38
39 107073-0 Bushing, Knee Gatch Link Mounting
40
41 107137-0 Spacer, Knee Gatch Link Mounting
16
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
23
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 24
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Parts List
Item Part No. Description
42 107382-0 Hanger, Camtec Controller
43 107170-0 Mattress Stop, Flat, Flip-Up
44 107179-0 Link, Knee Gatch Connecting
45 903924-0 Casting, Actuator Mounting
46 203060-0GY Weldment, Lower Knee Gatch
47 203204-1GY Weldment, Emergency Crank, Bed
48 24006635GY Weldment, Leg Link
49 203541-0GY Weldment, Side Board Support, LH Head
50 203542-0GY Weldment, Side Board Support, RH Head
51 203543-0GY Weldment, Mattress Support, Mid Section
52 203544-0GY Weldment, Mattress Support, Upper Knee Gatch
53 203545-0GY Weldment, Side Board Support, LH Lower Knee Gatch
54 230546-0GY Weldment, Side Board Support, RH Lower Knee Gatch
55
56 203544-0GY Weldment, Head Section
57 203555-0GY Weldment, Upper Knee Gatch
58 300587-0 Assembly, Adjustable Side Rail Release Plunger
9003645-0 Pull Tab, Bed Width Adjustment Pin
59 900094-0 Nut, 10-32 Hex, ESNA
60 900129-0 Washer, Lock ⅜" HELICAL
61 900171-0 Washer, #10 Internal Tooth
62 900658-0 Caplug, RER 1 x 2 - 14-20
63 901007-0 Screw, #10-32 x ⅝" LG, Round Head Phillips Stainless
64 902513-0 Bolt, Hex, ⅜-16 x 1" LG
65 902514-0 Bolt, Hex, ⅜-16 x 1½" LG
66 902518-0 Nut, ⅜-16 Hex, ESNA
67 903115-0 Bolt, Hex, ½-13 x 1¼" LG
68 903118-0 Bolt, Hex, ⅜-16 x 2½" LG
69 903152-0 Bushing, SS1620 18, ⅝" x 1⅛"
70 903154-0 Bolt, Hex, ½-13 x 2"
71 903172-0 Spring Clip
72 903173-0 Washer, Nylon, ⅝" x 1⅛" x 1/16"
73 903174-0 Grommet, ¾" x 1⅛"
74 903180-0 Cable Clamp 5/16" Nylon
75
76 903605-0 Washer, Rubber, ½" Screw Size
77 903653-0 Screw, #10-24 x ½" LG Self Tapping Hex Head
78 903950-0 Pendant, 10 Button DC, XL Series
79 904003-0 Actuator Assembly, Knee Gatch
80 904004-0 Actuator Assembly, Fowler
81 904005-0 Actuator Assembly, Hi-Lo Foot
24
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 25
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Parts List
Item Part No. Description
82 904005-0 Actuator Assembly, Hi-Lo Head
83 904009-0 Controller with Molex Connectors, 5 Function
84 904322-0 Cover, Latch, Side Board
85 904323-0 Side Board, Foot End, RH, Spread Base Low Bed
86 904324-0 Side Board, Foot End, LH, Spread Base Low Bed
87 904325-0 Side Board, Head End, LH, Spread Base Low Bed
88 904326-0 Side Board, Head End, RH, Spread Base Low Bed
89 903947-0 Controller, Staff Control
90 904406-0 Strip, Knee Gatch Skid (Old Part Number 107119-0)
91 904329-0 Cover, Side Board Pivot Arm
92 107139-0 Link, Axle Stabilizer
93 203558-0GY Weldment, Stabilizer, LH, Spread Base Low Bed
94 203559-0GY Weldment, Stabilizer, RH, Spread Base Low Bed
95 203654-0GY Weldment, Bolted Side Board MTG Arm, Pivot Left
96 203655-0GY Weldment, Bolted Side Board MTG Arm, Pivot Right
97 904289-0 Casting, Side Board Release
98
99
100 300950-0 Assy, Bolted Side Board, EXP, Pivot Left, LH, Head End
101 300951-0 Assy, Bolted Side Board, EXP, Pivot Right, RH Head End
102 300952-0 Assy, Bolted Side Board, EXP, Pivot Right, LH Foot End
103 300953-0 Assy, Bolted Side Board, EXP, Pivot Left, RH Foot End
104
105 900053-0 Screw, #10-¾" LG Flat Head Phillips
106 900991-0 Washer, Nylon, ⅜" x ¾" x 1/16" Thick
107 902516-0 Washer, Flat, ⅜" Zinc Plated
108 904369-0 Bushing, ½" x ⅝" x ⅞"
109 903155-0 Bolt, ½-13 x 2½", Hex
110 903165-0 Head Board Nut, ¼-20 x 17mm
111 903637-0 Head Board Bolt, ¼-20 x 30mm
112 904000-0 Spring, Compression, 0.480" OD
113
114 904015-0 Ring, Retaining, .0500" Shaft, E-Style
115 904067-0 Rivet, Drive, Polyamide, 0.250" x 0.750" LG
116 104605-0 Foot Board Support
117 104607-0 Head Board Support
118 106459-0 Strap, Cord Hanger, Velcro Loop
119 106460-0 Strap, Cord Hanger, Velcro Hook
120 300591-0 Assembly, Foot Board Support Tube
121 300592-0 Assembly, Head Board, Support Tube
122 900653-0 Caplug, Square 1"
25
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 26
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Parts List
Item Part No. Description
123 903175-0 Washer, Finish, #8 SS
124 903176-0 Screw, #8 x ½" LG, Oval Head, Phillips, SS
125 903451-0 Spring, Compression, 0.600" OD x ¾" LG
126 903452-0 Ball, ⅝" DIA Chrome
127 300242-0 Assembly, Head Board, 39", Meteor Nebula
128 904312-0 Head Board, 39", Meteor Nebula
129 300780-S Assy, FT BRD, 39" W/Staff Cont. and Scale, Meteor Nebula
130 904328-0 Foot Board, 39" W/Staff Control Pend., Meteor Nebula
131 903472-0 Label, Scale Instructions FG1257
131 903768-0 Label, Scale Instructions FG3600
132 107257-0GY Cover, Cord, Leg Link
133 903777-0 Load Cell
134
135
136
137
138 904194-0 Scale Readout, Rotating, FG1257
138 903773-0 Scale Readout, Rotating, FG1000
139 903477-0 Label, Head
140 903478-0 Label, Bed
141 903479-0 Label, Foot
142 903482-0 Label, Camtec Logo Caregivers
143 903489-0 Label, Power Cord Hanger
144 903490-0 Label, End Board Socket
145 903491-0 Label, Trapeze Socket
146 903500-0 Label, Pendant Cord Hanger
147 903502-0 Label, Warning Keep Feet
148 903505-0 Label, In Case of Power Failure
149 903507-0 Label, 2" Gap
150 903508-0 Label, Brake, Caster Cap Total Lock
151 903509-0 Label, Steer, Caster Cap Steer Lock
152 903530-0 Label, Caution Pinch Area
153
154 903569-0 Label, Max. 850 LBS
155 904071-0 Cable Clamp, 3/16" Nylon
156 300828-0 Assembly, Bumper Wheel
157 900021-0 Screw, #10-32 x 1" LG Flat Head Phillips, Stainless Steel
158 904193-0 Junction Box, FG1257 Scale System
159 903235-0 Screw, 5/16-18 x ⅜" LG Socket Set
160
161
26
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 27
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
Appendix A - Parts List
Item Part No. Description
162
163
164
165
166
167 904201-0 Night Light, Under Bed FG1257
168 904392-0 Label, When Raising, Maintain 15" From Wall
169 904438-0 Label, Slide ID, 3649 Left Hand
170 904439-0 Label, Slide ID, 3649 Right Hand
171 904440-0 Label, Slide ID 3649 Left Foot
172 904441-0 Label, Slide ID, 3649 Right Foot
173 904442-0 Label, Slide ID, 3649 Mid Section
174 904443-0 Label, Slide ID, 3649 Upper Knee Gatch
175 903483-0 Label, Do Not Press
180 300966-0 Assembly, Two Side Board Pivot Brackets, Short Rod, Pivot Left Head/Right Foot with
Stabilizer
181 300967-0 Assembly, Two Side Board Pivot Brackets, Short Rod, Pivot Right Head/Left Foot with
Stabilizer
182
183
184 107422-0 Rod, Tapped, Side Board Pivot, Short
185 107427-0 Spacer, Stabilizer Pivot
186 901016-0 Washer, Nylon, 0.312" ID x 0.500" OD x 0.062" Thick
187 904358-0 Screw, ¼-20 UNC x ⅞" LG BUT HD Socket Cap
188 904361-0 Nut, 0.250-20 Hex ESNA, Thin, Zinc Plated
189 904367-0 Screw, 5/16-24 UNF x ⅝" LG BUT HD Socket Cap
190 904368-0 Washer, Flat, Thru Hardened, 0.344" ID x 0.786" OD x 0.056" Thick
191 904407-0 Screw, #4 Sheet Metal x ½" LG Flat Head Phillips, S/S
192 107347-0 Retainer, Staff Control (Old P/N 904414-0)
193 901016-0 Washer, Nylon, 5/16 x ¾ x 1/16
194 900891-0 Bolt, ⅜-16 x ¾" Hex
195 902509-0 Nut, ⅜-16 ESNA, Thin
196 901237-0 Caplug, 2" x 3" x 11 GA
197 903194-0 Screw, #6 x ⅜" LG SS Phillips
198 904179-0 Cordset
199 904072-0 Bolt, Hex, ⅜-16 x ⅝" LG
200
201 107198-0 Bar, Side Board Pendant Support
202 107525-0 Washer, Stabilizer
203 107313-0 Bracket, Wire Harness Mounting
204 904112-0 Cable Harness, FG1257
204 904113-0 Cable Harness, FG3600
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
27
Page 28
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
Accessories FG1250 Base Scale
FG1250 Basic Operational Flow
Initializing System Zero
Note: When power is rst applied to the system, the display will read “AC POWER LOST! - REFER TO MANUAL”.
This indicates that there was an interruption in power and to check the patient weight. To reset the system, press the “FREEZE/RESUME” button.
Before patient is placed on to the bed, you must rst set the system zero. This is done as follows:
Sets the initial system zero reference.
1. Press and hold the “0.0” button. The displayed message will indicate “HOLD TO ZERO” as the digit to the right counts down.
2. When the message changes to “PLEASE WAIT - HANDS OFF”, release the button and do not touch the bed. In a few seconds the display will indicate “WEIGHT = 000.0 LB” The patient may now be placed on the bed.
Important: When the “0.0” button is pressed, everything on the weighing surface will be zeroed out on the display. Make sure that all linens, pillows, etc., are in place before pressing the “0.0” button so that they will not appear as part of the patient’s weight.
Button Functions
After the patient has been placed on the bed, simply press and release the “ON” button. The displayed message will read: “PLEASE WAIT - HANDS OFF”. Do not touch the bed. After a few seconds the patient’s weight will be displayed.
The WEIGHT CHANGE mode is used to monitor the patient’s weight relative to zero.
1. Press the “WT CHG” button. The display will read” “WT CHG = (patient weight)”.
2. Press the “0.0” button to set the relative zero.
Any variation in weight will now be viewed as a “+” or “–” gure relative to zero. Pressing the “WEIGHT” button will switch back to the WEIGHT MODE allowing you to view the patient’s total weight again.
Note: When you rest the system zero, the WEIGHT CHANGE MODE will also be reset to zero.
28
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 29
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
This feature will allow you to make adjustments to the bed without effecting the displayed patient weight.
1. Press the “FREEZE/RESUME” button. The display will read: “FREEZE = (patient weight)”.
2. Make all necessary adjustments to the bed.
While in the FREEZE mode, the “WEIGHT” and “WT/CHG” keys will allow you to manually adjust the displayed weight reading in the event of a know error. Simply press the “WEIGHT” and “WT/CHG” keys to make the proper corrections.
Note: If FREEZE mode is left unattended for more than three minutes, display will read “LEFT IN FREEZE - REFER TO MANUAL”. You must press the “FREEZE/RESUME” button to exit this mode.
Toggles between POUNDS and KILOGRAMS display modes.
Moving Bed and Patient
If it becomes necessary to move the bed and/or patient while the scale system is active, use the following procedure:
1. Press the “FREEZE/RESUME” button. The display will read: “FREEZE = (patient weight)”.
2. Make all necessary adjustments to bed and/or patient.
3. Press the “FREEZE/RESUME” button once again to resume normal weight monitoring.
Loss of Power
If it becomes necessary to disconnect the bed power while the scale is in use:
1. Note the displayed patient weight before disconnecting the power.
2. Unplug the bed and move bed and patient to the new location.
3. When the patient is settled restore the power to the bed.
4. Press the “FREEZE/RESUME” button to WAKE-UP the readout. If the weight reading is correct, the system
is ready.
5. If the displayed reading is not correct:
a. Press the “FREEZE/RESUME” button. The display will read “HOLD (weight) b. Use the “WEIGHT” and “WT/CHG” keys to ramp the displayed value to the correct weight.
c. When done, press the “FREEZE/RESUME” button once again to save the correction and exit to the normal
weighing mode. If you make a mistake, press “ON” to about your changes.
Important: If you do not know the correct patient weight, you must remove the patient from the bed and re-zero the system.
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
29
Page 30
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
System Calibration and Maintenance
Important: The following calibration procedure should only be performed by qualied service personnel. The 1228 load cell, itself, has no user serviceable components and should not be tampered with for any reason.
Recalibration is generally not required, but should be checked periodically to ensure accuracy.
Calibration Procedure
1. Press the “FREEZE/RESUME” button to enter the FREEZE MODE.
2. When the display indicated “FREEZE = (weight)”, press and hold the “LB/KG” button. The display
will read “CAL” as the right hand digit counts down to enter the CAL MODE.
3. When in the CAL MODE, press the “0.0” button to zero the display.
4. Place a known weight on to the weighing surface and compare it to the displayed reading.
5. Use the “WEIGHT” and “WT/CHG” keys to make any necessary adjustments to the displayed value. The displayed value should be within 0.1% of the calibrated weight plus or minus 1 digit of reading.
6. Press the “FREEZE/RESUME” to save changes or “ON” to abort changes.
Calibration Tolerance Table
Low Limit Applied Load High Limit
99.9 100.0 100.1
199.8 200.0 200.2
299.7 300.0 300.0
399.6 400.0 400.4
499.5 500.0 500.5
599.4 600.0 600.6
699.3 700.0 700.7
799.2 800.0 800.8
899.1 900.0 900.9
999.0 1000.0 1001.0
Cleaning and Disinfecting
The readout and transducer for the 1228 Scale System are made of aluminum and steel with a painted surface. You need only exercise caution when cleaning the display window as this is made of clear polyester. We recommend mild soap and water for general cleaning.
30
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 31
Accessories FG1257 Scale with Bed Exit/Nurse Call
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
31
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 32
Joerns® Camtec Series Bed Frames RC Ultra Hi-Lo Bed (RCUHLB850)
32
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 33
Notes
Joerns® Camtec Series Bed Frames
RC Ultra Hi-Lo Bed (RCUHLB850)
33
© 2015 Joerns Healthcare • 6110277 RevB • 15-2770
Page 34
Joerns Healthcare Warranty Program
for Joerns® RC Ultra Hi-Lo Bed, Model RCUHLB850
Joerns RC Ultra Hi-Lo Beds, Model RCUHLB850, are guaranteed for a period of one years from the date of delivery, against defects in materials and workmanship, under normal use and service.
This one-year warranty includes all mechanical and electrical components.
Steel structural components on beds are covered under warranty for a period of ve years from the date of delivery.
Welds are covered under warranty for the lifetime1 of the product.
Damage caused by use in unsuitable environmental conditions, abuse or failure to maintain the product in accordance with user and service instructions is not covered.
Any alteration, modication, or repair unless performed by or authorized in writing by Joerns Healthcare, will void this warranty.
Parts
Joerns beds contain various parts that wear from normal use. These parts are not covered under the stated warranty but are covered for 90 days after date of delivery.
Joerns Healthcare’s obligation under this warranty is limited to supplying replacement parts, servicing or replacing, at its option, any product which is found by Joerns Healthcare to be defective.
Warranty replacement parts are covered by the terms of this warranty until the product’s original stated warranty period expires.
When requested by Joerns Healthcare, parts must be returned for inspection at the customer’s expense. Credit will be issued only after inspection.
Service
Most service requests can be handled by the facility Maintenance Department with assistance from the Joerns Healthcare Product Service Department.
Most parts requested can be shipped next day air at the customer’s expense.
Should a technician be required, one will be provided by Joerns Healthcare, at our discretion. Only the Joerns Healthcare Product Service Department can dispatch authorized technicians.
This warranty is extended to the original purchaser of the equipment.
Manufactured by:
Joerns Healthcare 2100 Design Rd.
1
Weld lifetime dened as 15 years.
Post Acute, Acute, HomeCare
2430 Whitehall Park Dr. Ste 100 Charlotte, NC 28273 (P) 800.826.0270 (F) 800.457.8827
www.joerns.com • email: info@joerns.com © 2015 Joerns Healthcare • 6110277 RevB • 15-2770
VA/Government
19748 Dearborn Street Chatsworth, CA 91311 (P) 800.966.6662 (F) 800.232.9796