Important Notice: The equipment must be installed
and operated in the manner for which it was intended.
Facility staff/user is responsible for reading and
understanding the product user manual and contacting
Joerns Healthcare if anything in this manual is unclear.
Joerns will not be held responsible for any injuries
resulting from failure to comply with the instructions
and precautions in this manual.
Warning: Joerns specialty support surfaces are
designed as mattress replacement systems. The
risk of entrapment may occur when the equipment is
placed on bed frames that leave gaps of even a few
inches between the mattress and the head panel, foot
panel, and bed or side rails. The equipment is NOT to
be used when such gaps are present.
Facility staff/user is responsible for ensuring that all
mattresses properly t the bed frames. Joerns is not
responsible for the placement of its equipment on bed
frames that leave gaps between the mattress and
the head panel, foot panel or bed or side rails which
present a risk of harm to patients.
Warning: An optimal bed system assessment should
be conducted on each patient by a qualied clinician
or medical provider to ensure maximum safety of
the patient. The assessment should be conducted
within the context of, and in compliance with, the
state and federal guidelines related to the use of
restraints and bed system entrapment guidance,
including the Clinical Guidance for the Assessment and Implementation of Side Rails published by the
Hospital Bed Safety Workgroup of the U.S. Food
and Drug Administration. Further information can be
obtained at the following web address: http://www.fda.
gov/MedicalDevices/ProductsandMedicalProcedures/
GeneralHospitalDevicesandSupplies/HospitalBeds/
default.htm.
When using the mattress system, always ensure that
the patient is positioned properly within the connes
of the bed. Do not let any extremities protrude over
the side or between the bed rails when the mattress is
being used.
Danger Explosion Hazard: Do not use in the
presence of ammable anesthetics. Do not use in
the presence of smoking materials or open ame.
Air owing through the air mattress will support
combustion.
Danger: To reduce the risk of shock, adhere to the
following instructions. Failure to do so could result in
personal injury or equipment damage.
• Immediately after using the ProCair Plus, unplug it
from its power source.
• Do not place or store the product where it can fall or
be pulled into a tub or sink.
• Do not place or drop the product into water or other
liquid.
• Do not remove the back of the control unit. Refer
servicing to Joerns Healthcare.
Warning: To reduce the risk of burns, shock, re, or
personal injury, adhere to the following instructions.
Failure to do so could result in personal injury or
equipment damage.
1. Use this product only for its intended purpose as
described in this manual. Only use attachments
and/or accessories that are recommended by the
manufacturer.
2. If this product has a damaged power cord or plug,
is not working properly, has been dropped or
damaged, or has been dropped into water, do not
operate it. For examination and repair, return the
product to Joerns.
3. Keep the control unit and power cord away from
heated surfaces, e.g. space heaters.
4. Never block the air openings of the product. Do
not place the control unit on a surface, such as a
bed or couch, where the air opening and/or lter
compartment, located on the back of the control
unit, may be blocked. Keep the air openings free
of lint and hair.
5. Never drop or insert any object into any opening or
hose.
6. Do not spill food or liquids onto the control unit. If
a spill does occur, turn off the unit, disconnect it
from its power supply and allow at least 24 hours
for drying.
7. Do not use the product outdoors, or where
aerosol-spray products are used.
8. Plug this product only into a properly grounded
outlet. Refer to “Grounding Instructions”.
9. Ensure nothing is placed on the power cord and
ensure it is not located where it can be stepped on
or tripped over.
10. Do not attempt to service the control unit. Please
call Joerns Healthcare for any service requests.
11. The therapy pad (top cover) of this product is not
air permeable and may present a suffocation risk.
It is the responsibility of the caregiver to ensure
that the patient can use this product safely.
In April 1999, the U.S. Food and Drug Administration
(FDA) in partnership with representatives from the
hospital and post-acute bed industry, including
Joerns Healthcare, national healthcare organizations,
patient advocacy groups, and other federal agencies
formed the Hospital Bed Safety Workgroup (HBSW).
The workgroup’s goal is to improve the safety of
bed frames for patients and residents in all health
care settings who are most vulnerable to the risk of
entrapment. The efforts of the FDA and the HBSW
culminated in the FDA’s release of recommended
guidelines intended to reduce the risk of entrapment,
including dimensional limits for critical gaps and
spaces between bed system components and clinical
guidance for assessment and implementation of bed
side rails in various health care settings.
Entrapment zones involve the relationship of
bed components often directly assembled by the
healthcare facility rather than the manufacturer.
Therefore, compliance is the responsibility of the
facility.
As the leading manufacturer of long-term care beds
and a frontrunner in addressing this critical issue,
Joerns Healthcare can offer you the expertise,
assistance and products to bring your facility into
compliance.
Joerns® Compliance Solutions
Matching the right bed components in order to meet
regulatory guidelines can be complex.
That is why Joerns offers a wide array of compliance
options. We assist customers in selecting compliant
accessories recommended for their specic bed
model.
Creating a Safer Care Environment
While the guidelines apply to all healthcare settings
(hospitals, nursing homes and home care), long-term
care facilities have particular exposure since serious
entrapment events typically involve frail, elderly or
dementia patients.
For More Information
To learn more about compliance options with
Joerns products, visit our website at www.joerns.
com, or contact our Customer Care representatives
at 800.826.0270 and ask for free informational
publications.
To learn more about entrapment zones, assessment
methods, and guidelines concerning entrapment,
contact Joerns Healthcare at 800.826.0270 or
consult the FDA website: http://www.fda.gov/
MedicalDevices/ProductsandMedicalProcedures/
GeneralHospitalDevicesandSupplies/HospitalBeds/
default.htm.
Important Precautions .................................................................................................................................................. 2
Pressure Redistribution
Shear and Friction Reduction
Indications for Use ........................................................................................................................................................ 5
Pressure Redistribution
Pain Management
Features ....................................................................................................................................................................... 6
Mattress System Features
ES Therapy Control Unit Features
CBC Therapy Control Unit Features
Storage and Care ....................................................................................................................................................... 12
Support Surface
Control Unit
System Specications ................................................................................................................................................ 12
Weight
Dimensions
Electrical Specications
Environmental Conditions
Agency Approvals
Call for Assistance ...................................................................................................................................................... 12
The ProCair Plus, provided by Joerns Healthcare, is a
convertible dynamic pressure redistribution mattress
replacement system. Pressure redistribution and
alternating pressure therapy have been demonstrated
to reduce the risk of pressure ulcers and as being a
valuable aid in the treatment of pressure ulcers.
Warning: The risk of entrapment can arise when
equipment is placed on bed frames that leave gaps of
even a few inches between the mattress and the head
panel, foot panel, and bed or side rails. The equipment
is NOT to be used when such gaps are present. See
“Important Precautions” section of this manual.
The ProCair Plus replacement system is suitable
for at-risk patients and treatment for Stage I and
II pressure ulcers. The ProCair Plus may also be
indicated for additional therapeutic intervention based
on patient’s specic assessment.
With the addition of an optional control unit, the
therapy can take three forms:
• Alternating– where the mattress therapy cells
alternate in pressure.
• Powered– where the mattress internal pressures are
increased
• Autorm– provides maximum air ination that
may be used to assist during some patient care
procedures or to aid in patient transfer
Pressure Redistribution Optimization (P.R.O.)
Technology is a unique, patent pending, air control
technology. The P.R.O. Technology system requires no
adjustments or manual ination devices and features
a four-zone inner air core that automatically adjusts
to meet the needs of each patient based on body
prole and weight. The head zone remains static and
is comprised of high density foam for maximum patient
comfort while the shoulder, torso and foot zones
are optimized independently to maximize pressure
redistribution.
We have ensured that the ProCair Plus addresses key
areas in the treatment of compromised skin: pressure
redistribution and reduction in both friction and
shearing forces.
Pressure Redistribution
ProCair Plus is a convertible dynamic pressure
redistribution mattress system that has three modes
of operation: non-powered, alternating pressure and
static. The alternating and static modes require the
use of an optional control unit (ProTech ES2 or CBC).
The non-powered mode requires only that a control
unit not be attached and that any external hoses are
disconnected from the mattress. In the non-powered
mode ProCair Plus functions as dynamic pressure
redistribution mattress.
The alternating mode increases circulation to
the patient while providing maximum pressure
redistribution by alternating pressure between adjacent
therapy cells throughout the length of the mattress.
Maximum pressure redistribution is achieved through
delivering a specic amount of air to each therapy cell.
This distributes the patient’s weight evenly over a wide
surface area resulting in average pressure readings
well below capillary closure levels.
Shear and Friction Reduction
Shearing occurs when the skin is stationary in relation
to the support surface, while the underlying tissues
and vessels are stretched and damaged. When
a patient’s skin rubs against another surface, the
result is friction. The ProCair Plus therapy cover is
constructed of a breathable, non-plasticizing/moisture
proof nylon with a scratch resistant rubber backing
with low friction and shear properties to protect the
patient’s skin from these damaging forces.
Indications for Use
Pressure Redistribution
• Pressure Ulcers
• Rehabilitation
• Neurology
• Dermatology
• Burns
• Amputations
The selection of a pressure redistribution surface
needs to be based on each individual patient’s clinical
condition or diagnosis and co-morbidities. The ProCair
Plus provides average pressure readings well below
capillary closure levels (32mm of Mercury) and allows
for adequate perfusion to promote healing.
The ProCair Plus provides uniform distribution of
weight over a wide surface area, which relieves
pressure against bony prominences and provides
a soft, gentle therapy surface on which to lie. For
patients experiencing severe pain and discomfort due
to pressure and/or positioning limitations, consider
the ProCair Plus as an adjunct to pain management
interventions.
Note: The above are conditions and diagnoses
for which the ProCair Plus may be indicated.
Occasionally, there are orthopedic and neurological
patients that require body positioning to be maintained
in specic alignment. The ProCair Plus has safety
features and an inner therapy core to prevent deation
of the therapy cells and to keep your patients in
otation at all times. However, in the event of a
puncture or malfunction, the therapy cells may deate
and not provide the necessary alignment. The use
of the ProCair Plus for these patients should be
considered on an individual basis and discussed with
the attending physician.
Features
Mattress Features
• Four-zone (head, shoulder, torso and foot) dynamic
pressure redistribution prole with 15 cell internal
design automatically adjusts to each patient’s body
prole and weight
• The head zone remains static and is comprised of
high density foam for maximum patient comfort while
the shoulder, torso and foot zones are optimized
independently to maximize pressure redistribution
• Dynamic, non-powered mattress easily converts
to powered alternating and static modes with
connected control unit (ProTech ES2 or CBC)
• Automatic re-ination of air zones through P.R.O.
Technology after patient removal - no adjustment or
manual ination required
• Viscoelastic one-inch foam topper provides
maximum pressure redistribution, patient support
and increased comfort
• Integrated foot zone helps to protect delicate heel
area
• The surrounding rm perimeter provide stability
during patient care and transfer and helps support
patient safety
• Five inch deep internal cells are constructed of
highly durable therapy cells, specially designed for
automatic re-ination without the use of a powered
system
• 360° zippered, two-way stretch, therapy cover
is constructed of a breathable, non-plasticizing/
moisture proof nylon, aiding in the prevention of
friction and shearing
• Scratch resistant rubber backing
• Internal non-kinking hose sets
• Top cover is uid, stain and odor resistant and is
treated with a highly effective bacteriostatic agent to
inhibit the growth of bacteria and fungus
• Internal fully enclosed re barrier meets
BFD IX-11, 16CFR Part1633 and the expired
CA TB603 ammability standards
• Supports patients weighing up to 500 lbs.*
ProTechnology ES2 Control Unit Features
• Two therapy modes of operation – Alternating and
Powered (static)
• Alternating mode alternates the air pressure in
adjacent cells throughout the mattress (excluding the
head section)
• Three alternating pressure cycle times: 5, 10 and 15
minutes
• Autorm setting of unit provides maximum air
ination designed to assist both patients and
caregivers during patient care and transport
• Eight comfort control settings to maximize patient
compliance and promote healing.
• Compact lightweight control unit is quiet and energy
efcient
• Crisp, easy to read graphics for intuitive set up and
therapy control
• Closed loop pressure sensor control system
eliminates concerns of changes in mattress interface
pressure due to ambient temperature and pressure
changes
• Integrated swing out hanging brackets for xing to
most bed types
• Quick disconnect hoses allow for rapid attachment at
the mattress (xed at the control unit)
*Mattress weight capacity only; total weight must not exceed bed frame
manufacturers’speciedloadcapacity.
Page 7
Joerns® Therapeutic Support Surfaces
ProCair Plus
CBC Therapy Control Unit Features
• Alternating feature alternates the air pressure in
adjacent cells throughout the mattress (excluding the
head section)
• Therapy modes of operation: Alternate and
Static powered
• Ten comfort control settings to maximize patient
compliance and promote healing
• Compact lightweight control unit is quiet and energy
efcient
• Crisp, easy to read graphics for intuitive set up and
therapy control
• Integrated swing out brackets for xing to most bed
types
• Quick disconnect hoses allow for rapid attachment at
the mattress (xed at the control unit)
Grounding Instructions
Warning: Use a properly grounded, three-prong, 120V
AC outlet for this product. Failure to use a grounded
outlet could result in personal injury or damage to
equipment or house wiring, including risk of re. A
qualied electrician should be contacted to correct the
wiring and ensure a properly grounded outlet.
How to Determine if Outlet has the
Proper Grounding
Most hardware stores sell circuit testers (Figure 1) that
can be used to test an outlet for proper grounding.
The tester plugs into an outlet and by observing
the indicator lights you can determine if the outlet is
properly grounded. For a higher level of assurance, an
electrician should be requested to thoroughly test the
electrical system with more reliable equipment.
If repair or replacement of the cord or plug is
necessary, please contact Joerns Healthcare for
assistance.
Operation
Warning: For important safeguards, see page two.
Caution: The patient’s head should be positioned in
the center of the top section of the mattress. When
using the mattress system always ensure that the
patient is positioned properly within the connes of the
bed. Do not let any extremities protrude over the side
or between the bed rails when the mattress is being
used.
Before installing this product, have the electrical
system checked to make sure the electrical circuits
and the electrical service are properly grounded.
Having a three-prong outlet does not necessarily
mean it is grounded. Sometimes two-prong outlets are
replaced with a three-prong type even though there is
no ground wire.
There is always a chance of a loose connection or
poor installation of a ground wire that causes the loss
of proper ground at the outlet. Inadequate grounding
at electrical outlets can occur even if there is a
ground wire. Wires can become loose over time at the
connection to the outlet.
Note: To install new wires on a circuit requires a
qualied electrician.
Non-Powered Mode
The ProCair Plus in its non-powered mode is a
dynamic pressure distribution mattress replacement
system consisting of four zones: the head, shoulder,
torso and foot. The head zone remains static while
the cells in the shoulder, torso and foot areas are
connected to the P.R.O. Technology valve that allows
air to automatically be drawn into the cells in order to
provide the optimal amount of support for each area.
Based on the average patient’s body weight
distribution, the volume of air in each of these zones
was developed to provide the precise amount of air/
foam mix to ensure optimal clinical outcomes, with
average pressure readings well below capillary closure
levels (32mm of Mercury).
Over a period of time the air in the therapy cells will
naturally diffuse and the cells will deate. If no control
unit is used, it is recommended that the patient be
removed from the mattress periodically to allow the
compressed cells to automatically re-expand to their
natural state without the need of an external power
system.
The Mode key is used to place the control unit in one
of the desired modes of operation.
Comfort Adjust
The Comfort Adjust is located on the right of the
control panel. The ES control unit can be customized
to meet individual patient needs within a therapeutic
window. Use the up and down keys to increase or
decrease pressure in the mattress.
Lockout
The unit is designed to lock out all the adjustment
controls after the patient has been positioned correctly.
Approximately three minutes after the last button is
pushed, the Power light begins to ash indicating Lockout is enabled. This feature is to prevent any
unauthorized changes to the patient settings. To
unlock and make adjustments to the settings, press
both up and down Comfort Adjust arrows at the same
time. Lockout mode will return after approximately
three minutes of inactivity.
Cycle Key
When the unit is placed in Alternating mode, the ve
minute time light illuminates and the mattress therapy
cells will alternate in pressure on a ve minute cycle.
The cycle times can be adjusted to ten minute and
fteen minute intervals by pressing the cycle key.
ES2 Control Unit (Figure 2)
Figure 2
Standby Light
The Standby light illuminates when the unit is initially
plugged in, indicating power is available. If the unit is
turned off by pressing the Power button, the standby
light re-illuminates. If the Standby light illuminates
without a caregiver pressing the Power button, it
indicates that there has been a power interruption and
the therapy control unit is ready to be turned back on.
Press the Power button and reset the preferred mode
of therapy and comfort level.
Mode Selection
The Mode key is used to place the control unit in one
of the desired modes of operation.
The ES2 Control unit has three therapy modes:
Autorm, Therapy, and Alternate.
• Autorm: The Autorm light illuminates when in
Autorm mode. The mattress is manually inated
and can be used for patient transfer, or for a
specic patient procedure where a rm mattress is
preferable. The unit is designed to return to Therapy
mode after approximately 10 minutes in the Autorm
mode. This feature ensures the patient is not left in
the Autorm mode for an extended period of time.
• Therapy: The Therapy light is illuminated when in
Therapy mode. This mode should be used along with the Comfort Adjust keys to increase or decrease
the general rmness of the P.R.O. Matt® Plus
mattress.
• Alternate: The Alternate light is illuminated when
in the Alternate mode. The pressure automatically
alternates between adjacent cells. The alternate
frequency is carried out at preset time intervals set
by the Cycle Time key.
Power button is used to turn the power on and off and
comfort level.
Comfort Adjust
The Comfort Adjust buttons are located in the center
of the control panel. Use the up and down keys to
increase or decrease general rmness in the mattress.
Lockout
The unit is designed to lockout all the adjustment
controls after the patient has been positioned correctly.
Approximately three minutes after the last button is
pushed, the Power light begins to ash indicating Lockout is enabled. This feature is to prevent any
unauthorized changes to the patient settings. To
unlock and make adjustments to the settings, press
both of the up and down Comfort Adjust arrows
at the same time. Lockout mode will return after
approximately three minutes of inactivity.
Cycle Time
When the unit is placed in Alternate mode, the ve
minute time light illuminates and the mattress therapy
cells will alternate in pressure on a ve minute cycle.
The cycle times can be adjusted to 10 minute and 15
minute intervals by pressing the Cycle Time key.
The THERAPY PRESSURE light indicates the system has
the appropriate amount of pressure in the system.
Transport
Figure 3
CBC Control Unit (Figure 3)
The CBC control unit has two therapy modes: Alternating
and Static.
• Powered Static: Used along with the rm/soft dial to
increase the general rmness of the P.R.O. Matt
mattress. After a period of 0-5 minutes, the amber mode
switch will illuminate when in the powered Static mode.
Note: The very softest setting in the Powered Static mode
is still rmer than the non-powered mode of the P.R.O. Matt
Plus mattress.
• Alternating: The pressure automatically alternates
between adjacent therapy cells throughout the length of
the P.R.O. Matt Plus mattress on a ten minute cycle.
Power
The green Power switch is used to turn the CBC control
unit on and off. When in the on position, the Power light is
illuminated.
Mode Selection
The Mode switch is used to place the control unit in either
the Powered Static or Alternating modes.
Comfort Adjust
The Comfort Adjust dial is located on the left side of the
control panel. The CBC control unit can be customized to
meet individual patient needs within a therapeutic window.
Use the comfort dial to increase or decrease pressure in the
mattress.
Low Pressure Alarm
The LOW PRESSURE light will illuminate if there is not
enough pressure in the inner air cells. If this occurs, check
the hose connection to the mattress to ensure the hoses are
tightly connected without air leakage.
®
Plus
Patients being transported on the P.R.O. Matt Plus mattress
should be transported in the non-powered mode of
operation.
• Turn the control unit off
• Disconnect the air hoses from the mattress inlet valve
• Unplug the power cord from line power
After transport, if powered therapy is desired, reconnect
hoses and system power according to setup instructions.
Power Failure
In case of an extended power failure, the P.R.O. Matt Plus
mattress system should be disconnected from the control
unit. The patient can be left on the mattress which will force
out air from the internal alternating cells. In the non-powered
state, the mattress will continue to function as a dynamic
pressure redistribution mattress replacement system.
Note: ES2 Control Unit Only: In the event of a power failure,
if power is restored within approximately one hour, the
control unit will return to previous settings. If the duration of
a power failure extends beyond an hour, the control unit will
default to Standby mode when power is restored.
Mattress Connection
All of the control units have a hose set permanently
mounted to the side of the control unit. The connector
that attaches to the mattress has three ports. The
left and right are of the same connector type and are
interchangeable. The center connection is different
and must be connected to the center port.
To disconnect the control unit from the mattress, locate
the ports at the foot end of the mattress (Figure 4).
Disconnect the hose set by pressing down and pulling
the left and right connectors (Figure 5). Repeat with
the middle connector to fully disconnect the mattress
from the control unit (Figure 6).
Note: If the control unit is not powered for a long
period of time the connector should be disconnected
from the ProCair Plus mattress and the control unit
should be stored (see “Storage and Care” section).
Troubleshooting
ProCair Plus surface is not alternating or
increasing in pressure:
1. Ensure the hose connection from the therapy
mattress system (mattress) to the control unit is
securely connected.
2. Ensure that the control unit is plugged into an
AC outlet.
3. Ensure that the Power light is illuminated.
4. If steps 1-3 aren’t successful, contact Joerns
Healthcare for tech support at 800.826.0270.
Nursing Procedures
Recommended Linen:
Special linens are not necessary for the ProCair Plus,
and there is no need for a bottom sheet. The mattress
cover should be covering the therapy cells at all times.
Based upon the patient’s specic needs, the following
may be utilized:
• Draw or slide sheet in positioning and to further
minimize friction and shearing.
• Incontinence barrier pad for patients incontinent of
urine and stool, and patients with highly draining
wounds.
• Add use top sheets, blankets and/or bedspreads as
needed for patient comfort.
• Keep the amount of padding between the patient
and bed to a minimum for optimum performance.
Patient Positioning and Comfort
patients should be repositioned after elevation. The
knee gatch or knee section of the bed may be elevated
rst, to help prevent the patient from sliding when the
back section is elevated.
Contractures
Contractures and foot drop are a concern for all
bedridden and immobilized patients. Physical therapy
and any prescribed exercises may be performed on
the ProCair Plus as is done on any traditional hospital
bed.
Incontinence
Moisture against the skin surface leads to maceration,
or softening of the tissues. To prevent maceration, we
recommend you use an incontinence barrier pad to
absorb the excess moisture.
In the event of incontinence or excess drainage, you
should wipe off the excess uid from the bed surface.
Safety Information
Patient Migration
Specialty bed products are designed to reduce
pressure and the shearing/friction forces on the
patient’s skin. The risk of gradual movement and/or
sinking into hazardous positions of entrapment and/
or inadvertent bed exit may be increased due to the
nature of these products.
Traction
With any traction or unstable fractures, maintain
physician-directed angle of articulation and guard
against risks of patient migration or inadvertent
deation of patient surface.
General Repositioning
Patients should be turned and repositioned per an
individual turning schedule or per facility policy.
If it is not counter indicated, it is desirable to keep the
back section of the bed in the at position to provide
optimal pressure redistribution and minimize the risk of
shearing injuries.
Elevating Patient into Sitting Position
The special properties of the ProCair Plus reduce the
opportunity for shear and friction that may occur when
raising the back section of other bed systems. As
with any surface, sliding can be expected; therefore,
Monitor skin conditions regularly, particularly in areas
where incontinence and drainage occur or collect.
Early intervention may be essential to preventing
serious skin breakdown.
Bed Height
To minimize the risks of falls or injury, the mattress
support platform should always be in the lowest
practical position when the patient is unattended.
Make sure areas under and around the frame are clear
of objects, persons and parts of body before adjusting
height.
Page 11
Joerns® Therapeutic Support Surfaces
ProCair Plus
Cleaning
Warning: Unplug the control unit from its power
source. Failure to do so could result in personal injury
or equipment damage.
Warning: Do not expose the mattress or control unit to
excessive moisture that would allow for liquid pooling.
Personal injury or equipment damage could occur.
Caution: Do not use harsh cleansers/detergents, such
as scouring pads and heavy-duty grease removers, or
solvents, such as acetone. Equipment damage could
occur.
Control Unit Cleaning
Wipe off dust. If necessary, clean the housing exterior
with a disinfectant solution or a mild detergent and a
damp cloth, then wipe dry. Wipe power cord.
General Cleaning
If there is no visible soilage with possible body uids,
we recommend that you clean the mattress system
with a mild detergent and warm water. If disinfection
is desired, you may use a combination cleanser/
disinfectant as explained in “Disinfecting”.
Disinfecting
When there is visible soilage, and between patients,
we recommend that you disinfect the unit and mattress
with a tuberculocidal disinfectant. Disinfectant should
be registered with the Environmental Protection
Agency (EPA).
1. Use rubber gloves and eye protection
2. Prepare detergent/disinfectant (registered by EPA
as hospital disinfectant) solution according to
instructions on label for correct use-dilution
3. Thoroughly wipe down entire mattress
4. Remove gloves and dispose; wash hands
Filter Cleaning
When using an optional control unit, check the air lter
on the rear of the unit regularly for buildup of dust/dirt.
If buildup is visible, turn off the control
unit and disconnect the power cord from the wall
outlet. Remove the lter by grasping the lter and
pulling outward. Replace with the second supplied
lter. Ensure the replaced lter covers the entire
lter region.
Hand-wash the removed lter in warm soapy water
and allow to air dry. When dry, store the lter in a safe
place for the next lter maintenance.
1. Patient care equipment that does not come in
direct contact with the patient requires only lowlevel disinfection. Wiping surfaces with a properly
prepared detergent/disinfectant carries out lowlevel disinfecting.
2. Processing of dirty patient care equipment should
take place in a designated area away from clean
or sterile supplies and food preparation areas.
3. Detergent/disinfectants should not be mixed with
other germicides or detergents. Using the proper
dilution ensures the most effective killing power of
the disinfectant.
4. Patient care equipment that is used in isolation
areas should be disinfected in accordance with all
internal policies and procedures regarding such
equipment.
Steam Cleaning
Do not use any steam cleaning device on the unit.
Excessive moisture can damage mechanisms in this
unit.
Do not use any steam cleaning device on the mattress.
Excessive moisture can damage mechanisms in this
mattress
Maintenance
Warning: Only authorized personnel trained by Joerns
Healthcare should perform preventative maintenance.
Preventative maintenance performed by unauthorized
personnel could result in personal injury and/or
equipment damage.
Any maintenance done without Joerns Healthcare’s
authorization will void any warranties on this product.
When the product is not in use, store the power
cord properly. Failure to do so could result in personal
injury.
Note: Clean the ProCair Plus as described in the
previous section prior to storage.
Support Surface
• Thoroughly wipe down outside of the support
surface and allow to air dry.
• Cover with plastic and return to storage area. It is
recommended not to fold the mattress and to avoid
storage of the mattress other than in a
at format.
Control Unit
The permanently mounted hoses may be stored using
the included hose management strap.
The power cord may be wrapped around the unit for
convenience. Wrap the unit in a plastic bag for dust
resistance then store the unit in an area appropriate
for an electronic medical device.
Joerns Healthcare, warrants the ProCair Plus mattress
to be sold free from defects in workmanship and
materials, under normal and proper use, for a period of
ve (5) years, three (3) years non-prorated. The cover
warranty period is one (1) year. Damages arising from
improper use will not be covered by this warranty.
The Pro Technology ES2 control unit will be covered
for a period of two (2) years and the CBC control unit
will be covered for a period of one (1) year.
Improper use is dened as those caused by:
• Burns
• Chemical
• Needle punctures, cuts or abrasions
• Excessive loads
• Staining
• Negligent or excessive usage
• Improper maintenance, handling and/or cleaning
• Failure to use in the manner indicated in the
ProCair Plus user manual
Any modication, repair or alteration done to the
ProCair Plus that were not authorized in writing by
Joerns Healthcare will void this warranty.
Joerns Healthcare will pay shipping and handling costs
incurred in relation to this warranty for the period of
one (1) year. Thereafter, those fees will be the sole
responsibility of the purchaser. Any claims must be
submitted to Joerns Healthcare in writing within the
dened warranty period.
Joerns Healthcare reserves the right to repair or
replace the mattress or mattress components free
of charge during the warranty period. Substituted
components will be of equal or greater quality. No
returns, allowances, credits, discounts, charge-backs
or other deductions will be made without Joerns
Healthcare’s prior written authorization.
This warranty is the only warranty applicable to the
ProCair Plus and there are no other warranties,
expressed or implied, and no one other than Joerns
Healthcare has the authority to modify this warranty.
Joerns Healthcare’s liability will not exceed the
purchase price (plus any shipping/handling fees)
of the mattress. Joerns disclaims any liability for
consequential damages arising from a breach of this
warranty.
All warranty claims must have an assigned Joerns
Healthcare Return Authorization (RA) number.
Returned products are subject to inspection. It is
the sole discretion of Joerns Healthcare personnel
to determine if the claim is billable, or a non-charge
warranty replacement.
Note to Purchaser: please be advised that some
fabric will stretch and all foam (regardless of chemical
composition) and padding will compress during the