Joerns Healthcare FlapCair User & Service Manual

User-Service Manual
Joerns® Support Surface
FlapCair
To avoid injury, read user’s manual before using.
Joerns® Support Surface FlapCair Model
Important Precautions
Important Notice: The equipment must be installed and operated in the manner for which it was intended. Facility staff/user is responsible for reading and understanding the product user manual and contacting Joerns Healthcare if anything in this manual is unclear. Joerns will not be held responsible for any injuries resulting from failure to comply with the instructions and precautions in this manual.
Warning: Joerns Healthcare’s specialty support surfaces are designed as mattress replacement systems. The risk of entrapment may occur when the equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the headboard, footboard, and bed or side rails. The equipment is NOT to be used when such gaps are present.
Facility staff/user is responsible for ensuring that all
mattresses properly t the bed frames. Joerns is not
responsible for the placement of its equipment on bed frames that leave gaps between the mattress and the headboard, footboard or bed or side rails which present a risk of harm to patients.
Warning: An optimal bed system assessment should
be conducted on each patient by a qualied clinician
or medical provider to ensure maximum safety of the patient. The assessment should be conducted within the context of and in compliance with the state and federal guidelines related to the use of restraints and bed system entrapment guidance, including the Clinical Guidance for the Assessment and Implementation of Side Rails published by the Hospital Bed Safety Workgroup of the U.S. Food and Drug Administration. Further information can be obtained at the following web address: http://www.fda.gov/cdrh/ beds/.
When using the mattress system, always ensure that
the patient is positioned properly within the connes
of the bed. Do not let any extremities protrude over the side or between the bed rails when the mattress is being used.
Danger: Explosion Hazard: Do not use in the
presence of ammable anesthetics. Do not use in the presence of smoking materials or open ame. Air owing through the air mattress will support
combustion.
Danger: To reduce the risk of shock, adhere to the following instructions. Failure to do so could result in personal injury or equipment damage.
Immediately after using the FlapCair, unplug it
from its power source.
Do not place or store the product where it can fall
or be pulled into a tub or sink.
Do not place or drop the product into water or
other liquid.
Do not remove the back of the control unit. Refer
servicing to Joerns Healthcare.
Warning: To reduce the risk of burns, shock, re, or
personal injury, adhere to the following instructions. Failure to do so could result in personal injury or equipment damage.
1. Use this product only for its intended purpose as described in this manual. Only use attachments and/or accessories that are recommended by the manufacturer.
2. If this product has a damaged power cord or plug, is not working properly, has been dropped or damaged, or has been dropped into water, do not operate it. For examination and repair, return the product to Joerns.
3. Keep the control unit and power cord away from heated surfaces, e.g. space heaters.
4. Never block the air openings of the product. Do not place the control unit on a surface, such as a
bed or couch, where the air opening and/or lter
compartment, located on the back of the control unit, may be blocked. Keep the air openings free of lint and hair.
5. Never drop or insert any object into any opening or hose.
6. Do not spill food or liquids onto the control unit. If a spillage does occur, turn off the unit, disconnect it from its power supply and allow at least 24 hours for drying.
7. Do not use the product outdoors, or where aerosol-spray products are used.
8. Plug this product only into a properly grounded outlet. Refer to “Grounding Instructions”.
9. Ensure nothing is placed on the power cord and ensure it is not located where it can be stepped on or tripped over.
10. Do not attempt to service the control unit. Please call Joerns Healthcare for any service requests.
11. The therapy pad (top cover) of this product is not air permeable and may present a suffocation risk. It is the responsibility of the caregiver to ensure that the patient can use this product safely.
2
© 2013 Joerns Healthcare • 6110066 RevD • 13-2631
Joerns® Support Surface
FlapCair Model
Save These Instructions for Future Reference
Bed System Entrapment Information
Although essential in the practice of long-term care, bedside rails, in recent years, have also been a subject of regulatory review and evolution in design and use.
That focus includes not only the challenge of achieving an appropriate balance between resident security and unnecessary restraint, but also the additional safety issue of entrapment.
The U.S. Food and Drug Administration (FDA), working with our company and other industry representatives has addressed the potential danger of entrapment with new safety guidelines for medical beds. These guidelines recommend dimensional limits for critical gaps and spaces between bed system components.
Entrapment zones involve the relationship of components often directly assembled by the healthcare facility rather than the manufacturer. Therefore, compliance is the responsibility of the facility.
Creating a Safer Care Environment
While the guidelines apply to all healthcare settings, (hospitals, nursing homes and at home), long-term care facilities have particular exposure since serious entrapment events typically involve frail, elderly or dementia patients.
For More Information
To learn more about compliance options with Joerns products, visit our website at www.joerns.com, or contact our Customer Care reps at 800-826-0270 and ask for free informational publications.
To learn more about entrapment zones, assessment methods and guidelines concerning entrapment, contact Joerns Healthcare at 800-826-0270 or consult the FDA website: http://www.fda.gov/ MedicalDevices/ProductsandMedicalProcedures/ GeneralHospitalDevicesandSupplies/HospitalBeds/ default.htm.
As the leading manufacturer of long-term care beds and a frontrunner in addressing this critical issue, Joerns Healthcare can offer you the expertise, assistance and products to bring your facility into compliance.
Joerns® Compliance Solutions
Matching the right bed components in order to meet regulatory guidelines can be complex.
That is why Joerns offers a wide array of compliance options. We assist customers in selecting compliant
accessories recommended for their specic bed
model.
3
© 2013 Joerns Healthcare • 6110066 RevD • 13-2631
Joerns® Support Surface FlapCair Model
Table of Contents
Important Precautions ............................................................................................................................... 2
Introduction ................................................................................................................................................. 5
Moisture Control Shear and Friction Reduction
Indications for use ...................................................................................................................................... 5-6
Spinal Cord Injury Pressure Relief Pain Management
Features....................................................................................................................................................... 6
Therapy Control Unit Features Therapy Mattress System Features Therapy Pads
Grounding Instructions .............................................................................................................................. 7
Setup ............................................................................................................................................................ 7-8
To Create Super Flotation Zones
Operation ..................................................................................................................................................... 8
Patient Comfort Controls and Monitoring
Additional Features .................................................................................................................................... 9-10
CPR Transport Power Failure CairRails Optional Accessories
Troubleshooting .........................................................................................................................................10
Therapy Surface Not Inating
Unable to Change Therapy Mode or Adjust Comfort Control
Nursing Procedures ................................................................................................................................... 10-11
Recommended Linen Changing the Therapy Pad Patient Positioning and Comfort Incontinence
Safety Information ...................................................................................................................................... 11
Patient Migration Traction Skin Care Bed Height
Cleaning ...................................................................................................................................................... 11-12
Control Unit General Cleaning Disinfecting Therapy Pad Steam Cleaning Filter Cleaning
Maintenance ................................................................................................................................................ 12
Storage and Care ........................................................................................................................................ 12-13
Control Unit Support Surface
System Specications ............................................................................................................................... 13
Weight Dimensions
Electrical Specications
Environmental Conditions Agency Approvals
Call for Assistance ..................................................................................................................................... 13
Warranty ...................................................................................................................................................... 14
4
© 2013 Joerns Healthcare • 6110066 RevD • 13-2631
Joerns® Support Surface
FlapCair Model
Introduction
The FlapCair, provided by Joerns Healthcare, is a unique therapy system that provides pressure relief by combining low air loss with pulsation. Low air loss therapy has been demonstrated to reduce the risk of pressure ulcers as well as being a valuable aid in the treatment of pressure ulcers. The pulsation feature may provide additional patient comfort and enhance the function of the lymphatic system.
The FlapCair mattress replacement system is suitable for the care and prevention of all stages of pressure ulcers, for patients who have been assessed to be at risk from the complications of immobility, and for
patients with healing grafts and ap sites. Our unique
cellular system allows caregivers to provide super
otation zones to provide extra relief to these sites.
Joerns FlapCair is the rst true low air loss mattress to
take pressure relief to the next level with 160 individual therapy cells instead of the industry standard of 20.
More air cells mean a more exible surface area to
evenly distribute the patient’s weight. The patient’s weight is evenly distributed over a substantial surface area resulting in low average interface pressures.
The caregiver has the exibility to provide additional
pressure reduction on any part of the patient’s body by simply closing off the air supply to any of the 160 cells-
thereby “oating” the wound or surgical site. FlapCair
is truly customizable to meet the needs of your patient.
Warning: The risk of entrapment can arise when equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the headboard, footboard, and bed or side rails. The equipment is NOT to be used when such gaps are present. See “Important Precautions” section of this manual.
The FlapCair is quiet, comfortable and simple enough
for single caregiver installation, featuring rapid ination in just ve minutes or less. The user-friendly controls
allow for easy adjustment of patient comfort.
Additionally, low friction and low shear materials, together with average interface pressures well below capillary closure levels, means that the FlapCair meets the comfort and clinical requirements of your patients up to 350 lbs.
Moisture Control
Patients are at risk for skin maceration if excess moisture is permitted to accumulate beneath the patient. This may be due to perspiration, incontinence or wound drainage.
On the FlapCair, moisture is controlled via the specially
treated breathable, uid-proof, urethane coated nylon
therapy pad. The moisture vapor permeable fabric
of the therapy pad allows a sufcient amount of air
to circulate beneath the pad and wick away excess moisture.
Shear and Friction Reduction
Shearing occurs when the skin is stationary in relation to the support surface, while the underlying tissues and vessels are stretched and damaged. When a patient’s skin rubs against another surface, the result is friction. The top surface of the FlapCair therapy pad is constructed from a very smooth nylon fabric with low friction and low shear properties to protect the patient’s skin from these damaging forces.
Indications for Use
Note: The selection of a pressure-relieving surface needs to be based on each individual patient’s clinical condition, diagnosis and/or co-morbidities. The choice and use of a support surface is one factor in a holistic program of wound care and treatment.
Spinal Cord Injury
The FlapCair is not recommended for use by patients with unstable spinal fractures. Advice should be obtained from the appropriate physician before using the FlapCair system for these patients.
Pressure Relief
Pressure Ulcers Flaps Grafts Rehabilitation Neurology Dermatology Burns Amputations
Pain Management
We have ensured that the FlapCair addresses the four key areas in the treatment of compromised skin: pressure relief, moisture control, and reduction in both friction and shearing forces.
AIDS Arthritis Oncology
© 2013 Joerns Healthcare • 6110066 RevD • 13-2631