Joerns Healthcare EasyCare User & Service Manual

Page 1
User-Service Manual
Joerns® Bed Frames
EasyCare® Bed
To avoid injury, read user’s manual before using.
Page 2
Joerns® Bed Frames EasyCare®
Important Precautions
Warning: Possible Injury Or Death. This product is intended for use as an adjustable mattress platform that contours for resident/patient comfort with adjustable height for resident/patient and caregiver convenience. Use of this product in a manner for which it was not designed could result in unproven or unsafe conguration, potentially resulting in injury or death.
Warning: Possible Injury. Before adjusting bed,
ensure that area under and near bed perimeter is free of people and obstructions. Failure to do so could
result in injury or damage to the bed.
Warning: Possible Injury. Keep bed in lowest
position except when providing care (bathing, clothing changes, etc.). Bed should be at lowest convenient height for entry or exit. Failure to do so could result in injury.
Warning: Possible Injury Or Death. Resident/
patients may become entangled in pendant cord. Resident/patients with reduced mental acuity should
not be allowed access to pendant. Unsupervised use
of pendant could result in injury or death.
Warning: Possible Injury Or Death. Bed safe working load is 500 pounds. This is total weight
counting resident/patient, mattress, bedding accessories and any other equipment or persons likely
to be on bed. Do not exceed 500-pound safe working load. Exceeding the safe working load could result in
property damage, injury or death.
Warning: Possible Injury Or Death. Do not roll bed over abrupt thresholds with resident/patient in bed. Bed is not designed for transporting resident/patient. Transporting resident/patient in the bed could result in
injury or death.
Warning: Possible Shock Hazard. Injury may result
from improper routing of the power cord. Always follow the proper factory-installed routing conguration. Failure to do so could result in injury or death.
Warning: Possible Injury Or Death. Use a mattress
that is properly sized to t mattress deck, that will
remain centered on mattress deck relative to State and Federal guidelines. Joerns Healthcare recommends the use of a mattress with minimum dimensions of 35 inches wide and 6 inches deep. Length should match
mattress support platform. Use of an improperly tted mattress could result in injury or death. If using the CareWide accessory or the bed is equipped with Care­Fit integrated width expansion, Joerns Healthcare
recommends the use of a mattress with a minimum width of 39 inches for a 39 inch setting and 42 inches for 42 inch setting.
Warning: Possible Injury Or Death. If a resident/
patient’s mental or physical condition could lead to resident/patient entrapment, the mattress support platform should be left in the at position when unattended. Failure to do so could result in injury or
death.
Warning: An optimal bed system assessment should
be conducted on each resident by a qualied clinician or medical provider to ensure maximum safety of
the resident. The assessment should be conducted
within the context of, and in compliance with, the
state and federal guidelines related to the use of
restraints and bed system entrapment guidance,
including the Clinical Guidance for the Assessment and Implementation of Side Rails published by the
Hospital Bed Safety Workgroup of the U.S. Food
and Drug Administration.
Warning: Possible re hazard when used with oxygen
administrating equipment other than nasal, mask or ½ bed length tent type. Not for use with full bed length tent or hyperbaric oxygen chamber. Oxygen tent
should not extend below mattress support platform.
Lock hand control at foot of bed when using oxygen administering equipment.
Warning: Possible Shock Hazard. Unplug
power cord from wall outlet before performing any maintenance, cleaning or service to the bed. Failure to
do so could result in injury or death. Note: In case of
unexpected movement, unplug bed from any outlet.
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Warning: Possible Injury Or Death. Use a properly
sized mattress in order to minimize the gap between
the side of mattress and assist device. This gap must be small enough to prevent resident/patient from getting his/her head or neck caught in this location.
Make sure that raising or lowering bed, or contouring the sleep surface, does not create any hazardous gaps. Excessive gaps may result in injury or death.
Warning: Possible Injury Or Death. Locate mounting
brackets for assist devices strictly according to
instructions. Gap between head/foot panel and assist device must be small enough or large enough to prevent resident/patient from getting his/her head or
neck caught in this location (see specic Assist Device
Installation Instructions for more information).
Page 3
Joerns® Bed Frames
EasyCare®
If multiple assist devices are used, position each so
that gap between them is large enough for trunk and
hips to pass through easily. Make sure raising or lowering bed, or contouring sleep surface does not create any hazardous gaps. Failure to do so could
result in injury or death. Important Note: Powered air
mattress surfaces may pose a risk of entrapment. Prior to use, ensure the therapeutic benets outweigh the
risk of entrapment. Warning: Possible Injury Or Death. Do not use
an assist device if any openings within assist device body allow resident/patient to get his/her head or neck
lodged within these openings. Use of assist devices
with large openings may result in injury or death.
Warning: Possible Injury Or Death. Do not use any
Joerns
®
assist device until you verify it is designed
specically for EasyCare® and it is locked in place.
Failure to lock assist devices may result in injury or
death.
Warning: Possible Injury Or Death. Do not use any
replacement parts not manufactured, marketed or provided by Joerns Healthcare on any Joerns bed. Use of unapproved replacement parts may result in injury or death.
Warning: Possible Injury Or Death. If using
accessories not manufactured, marketed or provided by Joerns Healthcare for Joerns beds, consult with the manufacturer for compatibility and limitations prior to use. Failure to do so may result in injury or death.
Warning: Possible Shock Hazard. If the bed power
cord has a hospital-grade, 3-prong grounding plug, grounding reliability can only be achieved when equipment is connected to an equivalent receptacle
marked hospital-grade. Warning: Possible Injury. Sitting on or sliding over
the Mattress Side Stop can cause skin injury if the bed
mattress compresses to less than 1.5 inches under user’s weight.
Warning: Possible Injury or Death. Floor locks
increase bed stability and resident safety. Make
sure the bed is in a locked position with respect to
the oor by ensuring the casters are off the oor or if equipped, the Care-Lock is in the locked position before attempting any resident transfers. Failure to do so could result in injury or death.
Warning: Possible Injury. Use caution when activating the Care-Lock function. This feature was
designed to be activated by your foot. Using your hand to activate the Care-Lock system could result in injury.
Caution: In the absence of line voltage, all bed
adjustments can be achieved via an externally available backup battery (Joerns P/N N511). Joerns recommends a minimum of one battery for every twenty ve beds per facility.
Caution: This device complies with EMC requirements
of IEC 60601-1-2. Radio transmitting equipment, cellular phones or similar electronic devices, used in proximity of bed, may affect bed performance.
Caution: Particular precaution must be considered during exposure to strong emission sources such
as High Frequency surgical equipment and similar devices. Do not route High Frequency cables on or near device. If in doubt, contact a qualied technician.
Caution: The Joerns’ EasyCare bed is intended for
use within an institutional healthcare environment
(ie: skilled nursing, transitional care, rehabilitational care, assisted living). Joerns Healthcare recommends compliance to the regulations and guidelines specied to your locality.
This product is only one part of your healthcare bed system. Proper combinations of mattress, head/foot panels and assist devices are needed to minimize the risk of entrapment.
Save These Instructions
Warning: Possible Injury. Risk of entanglement or
injury may occur if mattress used with Mattress Side Stop does not ll the entire width between stops or
which compresses to less than 1.5 inches under user’s weight.
Thank you and congratulations for choosing Joerns EasyCare bed. Joerns beds are developed with safety,
comfort and convenience of both residents/patients and caregivers in mind.
Advanced ergonomic design will provide years of reliable service in your facility.
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 4
Joerns® Bed Frames EasyCare®
Table of Contents
Important Precautions ...................................................................................................................................................2
Entrapment Information .................................................................................................................................................4
Features & Specications ..............................................................................................................................................5
Assembly .......................................................................................................................................................................6
Operation ......................................................................................................................................................................8
Bed Transport...............................................................................................................................................................10
Care-Fit Integrated Width (option)................................................................................................................................11
Servicing ......................................................................................................................................................................12
Troubleshooting ...........................................................................................................................................................14
Maintenance ................................................................................................................................................................14
Preventative Maintenance ...........................................................................................................................................14
Cleaning ......................................................................................................................................................................14
Accessories .................................................................................................................................................................15
Warranty ......................................................................................................................................................................18
Bed System Entrapment Information
Although essential in the practice of long-term care, bedside rails, in recent years, have also been a subject of regulatory review and evolution in design
and use.
That focus includes not only the challenge of achieving an appropriate balance between resident security and unnecessary restraint, but also the additional safety
issue of entrapment.
The U.S. Food and Drug Administration (FDA), working with our company and other industry
representatives has addressed the potential danger
of entrapment with new safety guidelines for medical
beds. These guidelines recommend dimensional limits for critical gaps and spaces between bed
system components.
Entrapment zones involve the relationship of components often directly assembled by the healthcare facility rather than the manufacturer.
Therefore, compliance is the responsibility of
the facility.
Joerns® Compliance Solutions
Matching the right bed components in order to meet
regulatory guidelines can be complex.
That is why Joerns offers a wide array of compliance
options. We assist customers in selecting compliant accessories recommended for their specic bed model.
Creating a Safer Care Environment
While the guidelines apply to all healthcare settings, (hospitals, nursing homes and at home), long-term
care facilities have particular exposure since serious
entrapment events typically involve frail, elderly or
dementia patients.
As the leading manufacturer of long-term care beds
and a frontrunner in addressing this critical issue, Joerns Healthcare can offer you the expertise,
assistance and products to bring your facility
into compliance.
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 5
Joerns® Bed Frames
EasyCare®
Features and Specifications
Standard Features
76" or 80" Sleep Surface
Travel Range: 7" to 30"
Locking System:
Roll-in-Low Lock System
Care-Lock Roll at any Height Mobility Pendant Jacks on both left and right side Whisper Quiet DC Motors Welded Tubular Frame Grid Sleep Surface SoftTone Frame Color Baseboard Bumper Wide Stable Base
Heavy-Duty Casters
Mattress Stops
Patented Lifting Mechanism (US Patent 6473922) EasyLift Leg Positioner Battery Pack Cable
Electrical Specifications
Power: 120 VAC Frequency: 60 Hz Current Rating: 4.0 Amps (120V)
Classication: Class I, Type B or Class II, Type B
Mode of Operation: 10% Maximum Duty Cycle
2 minutes on/18 minutes off
Circuit Protection: LA27 Actuators: Auto Reset
Current Sensor
Accessories And Options
76", 80" and 84" Joerns
Joerns Bed Panels
F18 SoftCare half rail
F14SC Assist Device Set F025 Assist Handle F026 Assist Device
F027 Pivot Assist Handle F028 Deluxe Assist Device
Staff Control IV Rod Floor Mat
Emergency Battery Backup System
Battery Backup 24V
Battery Charger 24V
• Wall Mounting Bracket for Charger
Trapeze and Trapeze Adapter
Drainage Bag Holder
®
Mattresses
Bed Identification Numbers
When ordering parts or when contacting our
Customer Care Department, please include the bed’s model, product and serial numbers, found on the identication labels. The identication
labels are located on bed frame under the sleep deck foot section.
Environmental Conditions
Operating Conditions:
Ambient Temperature: +10°C to +40°C
Relative Humidity: 30% to 75% Non-Condensing Atmospheric Pressure: 700 hPa to 1060 hPa
Storage and Shipping Conditions:
Ambient Temperature: -40°C to +70°C Relative Humidity: 10% to 100%
Atmospheric Pressure: 500 hPa to 1060 hPa
I.D. Label Symbol Definition
Attention, Consult accompanying documents
Safe Working Load
Type B Applied Parts
Class II Equipment
Protective Earth (ground)
Mechanical Specifications
Description: Dimensions:
76" ........................................80.75"L x 35.0"W x 7.0"H
80" ......................................84.25"L x 35.0"W x 7.65"H
Weight
Roll-in-Low ...................................................220 lbs.
Roll at any Height ......................................... 230 lbs.
Care-Fit Integrated Width.................................+40 lbs.
Maximum Back Angle .........................65° to horizontal
Maximum Knee Angle ........................35° to horizontal
Maximum Knee To Leg Angle ................................. 45°
Minimum Height Mattress Support Surface
Roll-in-Low ........................................................7.00"
Roll at any Height ..............................................7.65"
Maximum Height Mattress Support Surface
Roll-in-Low ......................................................30.00"
Roll at any Height ............................................30.65"
Safe Working Load .......................................... 500 lbs.
Hi-Lo Speed
Unloaded .................................................50 seconds
With 200 lbs ............................................52 seconds
Tolerance ± 3 seconds
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 6
Joerns® Bed Frames EasyCare®
Assembly
Installation Of Head and Foot Panels
Refer to Bed Panel Installation Sheet for more
Tools required: #2 Phillips screwdriver
Head Panel Installation:
1. Locate the head panel; the head panel has a label on the back.
2. Lay panel on a at soft surface, face down, with
the threaded t-nuts facing away from the bed
occupant.
3. Head panel brackets have rubber bumpers mounted on them.
4. The panel brackets are right-left oriented. Items welded to the side of the panel mounting tube
should be placed to the inside (Figure 1). Install
head panel brackets with four screws provided. Do not tighten at this time.
5. Insert the brackets into the bed frame sockets
(Figure 2); maintain support of the bed panel. Insert the safety pins on each side and latch them (Figure 3).
6. Press the mounting bracket against the bed frame and tighten the four panel mounting screws; right and left side.
Figure 2
Figure 3
Foot Panel Installation:
1. Locate the Foot Panel; the foot panel does not have a label on the back.
2. Lay panel on a at soft surface, face down, with the threaded t-nuts facing towards the bed occupant.
3. The panel brackets are right-left oriented. Items welded to the side of the panel mounting tube
should be placed to the inside (Figure 4). Install
foot panel brackets with four screws provided. Do not tighten at this time.
4. Insert the brackets into the bed sockets (Figure 5);
maintain support of the bed panel. Insert the safety pins on each side and latch them (Figure 6).
5. Press the mounting bracket against the bed frame and tighten the four panel mounting screws; right and left side.
Figure 4
Figure 5
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Figure 1
Figure 6
Page 7
Caution: Carefully remove the packaging straps
(orange zip ties) before attempting to operate the bed.
Installation Of Baseboard Bumper
Recommended Tool: Rubber mallet
Joerns® Bed Frames
EasyCare®
1. Place the baseboard bumper as shown (Figure 7).
2. Tap into place with a rubber mallet or press rmly
with the sole of your shoe until it locks into place as shown (Figure 8).
Installation Of Baseboard Bumper (Advanced Sta Control)
Recommended Tool: Rubber mallet
The bed will ship with the standard bumper installed.
If the 12° Trendelenburg Mode is enabled (optional) it will be necessary to remove the installed bumper and
install the 12° bumper. Note: The 12° bumper can be
identied by a silver label with “12°” printed on it. It is
longer than the standard bumper.
1. Remove the shorter bumper by tapping the tangs
on either side of the leg with a rubber mallet.
2. Place the 12° baseboard bumper.
3. Tap into place with a rubber mallet or press rmly
with the sole of your shoe until it locks into place.
Installation of Mattress Stop
Figure 7
Figure 8
Mattress Stops are designed to help keep the mattress
from sliding laterally and longitudinally on the mattress
support platform.
Installation Steps
Required tool - #2 Phillips screwdriver
1. Locate the screws at each corner of the inside
tube at the foot end of the mattress support
platform; one per side (Figure 9).
2. Remove the screws using the #2 Phillips
screwdriver; do not discard the screws.
3. Position and insert the mattress stops in the
holes provided (Figure 10). Reinstall the screws
removed in step 2.
4. Repeat at each corner of the head end of the
mattress support platform.
Figure 9
Figure 10
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 8
Joerns® Bed Frames EasyCare®
IV Sockets
Joerns
®
EasyCare standard equipment. There are two sockets at head end of bed and two at foot end (Figures 11 and 12).
The IV sockets are compatible with Joerns IV poles
and most ½" IV poles.
®
bed has four IV sockets as
Installation of Sta Control Panel
(Optional)
1. Mount the foot panel on the bed as indicated in
previous instructions.
2. Align and insert the staff control cable plug into the
cable jack, located on the bed (Figure 13).
3. Ensure that the cable plug is fully inserted into the
cable jack.
4. To secure tighten the union nut on both cables
(Figure 14).
Operation
Pendant Operation
Your EasyCare bed comes standard with one of the
pendants shown in Figures 19 and 20 on page 9. The pendant plugs into a jack under either side of the seat pan.
IV Socket
Figure 12
Figure 13
1. Unscrew the union nut (Figure 15) and insert
the cable plug, with the arrow facing up into jack (Figure 16).
2. Ensure that the cable plug is fully inserted into the
jack (Figure 17, page 9).
3. Screw on the union nut from the cable to secure
cable (Figure 18, page 9).
4. The pendant has raised, high-contrast buttons and
graphics.
IV Socket
Figure 11
Figure 14
Figure 15
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Figure 16
Page 9
Figure 17
Joerns® Bed Frames
EasyCare®
Sta Control Panel Operation (Optional)
(Figures 21 and 22)
The Staff Control Panel allows the caregivers to operate the bed from the foot panel without walking to
the bedside. The caregiver can also individually lock out all functions, both Pendant and Staff Control.
1. To lock out a function, press and hold the “Lock” graphic on the panel and push the function to lock out. Illuminated amber LED indicates that the function is now locked out.
2. To unlock a function, press and hold the “Lock” graphic on the panel and push the function to
unlock. Amber LED is no longer illuminated,
signaling that the function is now operable.
Figure 18
Standard Pendant Operation
(Figure 19)
Head Operation
1. Push button to raise Back
2. Push button to lower Back
Knee Operation
3. Push button to raise Knee
4. Push button to lower Knee
Bed Height Operation
5. Push button to raise Bed
6. Push button to lower Bed
Advanced Pendant Operation (Figure 20) Head Operation
1. Push button to raise Back
2. Push button to lower Back
Knee Operation
3. Push button to raise Knee
4. Push button to lower Knee
Bed Height Operation
5. Push button to raise Bed
6. Push button to lower Bed
Comfort Chair Operation
7. Push button for Comfort Chair
8. Push button to lower both head and knee to at and place bed
at transfer height
31
2 4
5 6
Figure 19
2 4
5 6
7 8
Figure 20
1
53
7
4 6
2
Standard Staff Control (Figure 21)
Head Operation
1. Push to raise Back
2. Push to lower Back
Knee Operation
3. Push to raise Knee
4. Push to lower Knee
1
Bed Height Operation
5. Push to raise Bed
6. Push to lower Bed
Lock Operations
7. Push and hold to change lockout status in combination with another function button.
8
12 7
2
Advanced Staff Control (Figure 22)
Head Operation
31
1. Push to raise Back
2. Push to lower Back
Knee Operation
3. Push to raise Knee
4. Push to lower Knee
Bed Height Operation
5. Push to raise Bed
6. Push to lower Bed
4 65391011
Lock Operations
7. Push and hold to change lockout status in combination with another function button.
Additional Operations
8. Push for Comfort Chair
9. Push to lower both head
and knee to at and place
bed at transfer height
10. Push for Trendelenburg
11. Push for reverse Trendelenburg
12. Push button to toggle under bed light on and off
(optional)
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 10
Joerns® Bed Frames EasyCare®
Setting 12 Degree Trendelenburg Mode
(Advanced Sta Control)
The bed will ship set to 4° Gravity Assist mode. This
mode limits the head down angle to 4°. The software can be changed to 12° Trendelenburg mode to enable
Trendelenburg functionality.
1. Install Advanced Staff Control Panel described.
2. Simultaneously, press and hold Trendelenburg
(Button 10, Figure 22 on page 10) and reverse Trendelenburg (Button 11, Figure 22, page 10)
buttons for 5 seconds on Advanced Staff Control Panel.
3. Trendelenburg/Reverse Trendelenburg lock icon
will begin ashing indicating that the system is in
programming mode.
4. Press bed up (Figure 22, page 10) button to select full 12° Trendelenburg mode.
5. Bed Hi-Low lock out icon will ash twice conrming software is set to 12° Trendelenburg mode. Note: Be sure correct baseboard bumper is installed. Failure to install the correct baseboard bumper can result in damage to the bed or walls.
The 12° bumper is identied by a silver label with “12°” printed on it.
6. To return bed to 4° Gravity Assist, repeat steps 2
and 3 then press bed down (Figure 22, page 10). Bed Hi-Low lock icon will blink once conrming software is now in 4° Gravity Assist mode. Install
correct baseboard bumper.
Care-Lock Feature (Optional)
The bed features a rolling base that allows it to be
rolled or locked at any bed height.
To stop motion of the bed, enable the Care-Lock
feature:
1. Engage the steering locks, located on the dual
wheel casters at the head of the bed (Figure 24).
2. Engage the Care-Lock at the foot end of the bed
by pressing down on the red or right side pedal (Figure 25).
To resume bed movement, disable the Care-Lock
feature:
1. Disengage by pressing down on the green or left
pedal (Figure 26). The steering locks may remain in place for desired maneuverability.
Warning: Possible Injury. Before leaving the bed
unattended, check that the Care-Lock feature is in the locked position. Never leave an unlocked bed
unattended. Warning: Possible Injury. Check that the Care-Lock
feature is in the locked position before allowing the resident to enter or exit the bed.
For best practices, the Care-Lock feature should
be locked at all times, except when the bed is
being moved.
Bed Transport
Roll in Low Lock System
1. To keep the bed from rolling when in the lowest
position, press down on the locking lever to lock casters (Figure 23).
2. The standard caster set comes with two locking casters placed at opposing corners of bed.
Mobile Position
Roll in Low Lock
System
10
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Low only Any Height
Locked Position
(casters must be locked in
Low)
Figure 23
Figure 24
Swivel Casters
Head & Foot (in Low
position only)
Figure 25
Figure 26
Page 11
Care-Fit Integrated Width Expansion (optional)
The EasyCare Care-Fit Integrated Width Expansion option allows the mattress support platform (MSP) to be set at a bed width of 35", 39", or 42". If your Joerns bed has been congured with the Care-Fit option the following steps will provide adequate instruction on the
operation of this feature.
1. If applicable, remove the sleep surface.
2. Locate the width expansion spring pin with tether
on any of the six MSP extensions (Figure 27).
3. Pull the tether in the direction towards the oor to
allow the MSP extension to move freely.
4. Slide the MSP extension to the desired width. Note
the width indicator found on the head and foot
extensions (Figure 28 and 29).
5. Release the tether to engage the detent
(Figure 27).
6. Repeat the above steps on the remaining
MSP extensions.
Joerns® Bed Frames
EasyCare®
Figure 27
Warning: Possible Injury Or Death. Use a mattress
that is properly sized to t mattress deck, that will
remain centered on mattress deck relative to State and Federal guidelines. Joerns Healthcare recommends the use of a mattress with minimum dimensions of 35 inches wide and 6 inches deep. Length should
match mattress support platform. Use of an improperly tted mattress could result in injury or death. If using CareWide or the bed is equipped with Care-Fit, Joerns
Healthcare recommends the use of a mattress with a minimum width of 39 inches for a 39 inch setting and 42 inches for 42 inch setting.
Warning: Possible Injury or Death. Do not apply
any width expansion accessory kit to a Joerns bed equipped with the Care-Fit integrated width expansion feature. Failure to comply may cause the bed to become unstable, leading to a possible injury or equipment failure.
Figure 28
Figure 29
11
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 12
Joerns® Bed Frames EasyCare®
EasyLi Leg Positioner Operation
1. To raise: Grasp the bed’s foot deck and raise to
the desired position (Figure 30).
2. To lower: Grasp the foot end of the bed’s deck and
raise ½-inch to ¾-inch above present notch, then lower gently;
3. Or, grasp the foot end of the bed deck, raise to the
highest position, and lower gently.
Servicing
Power Cord and Grounding Connectors
Inspection of power cord and grounding connectors
(if present) should be conducted on a yearly basis to ensure efcient bed operation.
1. Check the power cord to make sure it has not been damaged. If the power cord needs
replacement due to damage, remove the bed
from use/service until the power cord has been replaced.
2. Check the grounding connectors (if present)
to ensure they remain intact and are securely
fastened to their intended connection point.
Actuator and Control Box Information
Warning: Possible Shock Hazard. Unplug the
power cord from the wall outlet before performing any
maintenance/service to the control box.
Note: Overload protection is achieved via current
sensing technology, which will reset automatically if any bed function fails to operate. Ensure that the bed
is not overloaded or constrained from movement.
1
Figure 30
2
Figure 31
6
3
4
HB
7
Actuator And Control Box Identication
Cable and Port Identication (Figures 31 and 32)
1. Foot Leg Hi/Low Actuator (red)
2. Back Actuator (yellow)
3. Head Leg Hi/Low Actuator (orange)
4. Knee Actuator (blue)
5. Control Harness (green)
6. Emergency Battery Pack
7. AC Power
Note: The Actuators and Control Box contain no
serviceable parts. Overload protection is achieved via current sensing technology, which will automatically reset. If any bed function fails to operate, check to
see if the bed is overloaded or constrained from movement.
12
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Figure 32
Page 13
Joerns® Bed Frames
EasyCare®
To Remove Control Box
1. Plugs and outlets are numbered and color-coded
so they can be easily reinstalled later (See Actuator And Control Box Identication).
2. Unplug from wall outlet.
3. Unplug all actuator cords and pendant/staff
control plug.
4. Remove the control box retaining clip (Figure 33).
5. Slide the Control Box off the actuator assembly.
6. To replace the Control Box, reverse
these instructions.
T o Remove Actuator
Recommended Tools: Screwdriver, Hammer, Masking Tape, Zip Ties
1. Identify the actuator to be replaced.
2. Tip the bed on its side to replace high/low
actuator(s).
3. Unplug the actuator cord.
4. The actuator is held in place by two pins. Use a
screwdriver to remove the lock ring on each pin.
5. Slide the pins out of holes. If necessary, use the
hammer to tap out the pins.
6. You may now replace the actuator(s).
7. To reassemble the bed, reverse the previous
steps, making note of the following two important
items:
• Assemble the pin and lock rings as
originally installed.
Zip-ties should be replaced, with cords returned to their original position and routing direction to the control box.
To Remove AC Power Cable
Recommended Tools: Small Screwdriver, Masking Tape, Zip Ties, 5/16" Nut Driver
1. Raise the bed so that there is about 27 inches
of clearance between the oor and the mattress
support platform of bed.
2. Unplug the power cable from the outlet.
3. Cut the zip-ties that hold the power cable to the frame. Tape the remaining cables in their original
position. It is very important to keep the actuator
cables routed along the frame in their original
factory-installed positions. If this step is not followed, damage to cables or electric shock may
result.
4. Detach the ground wire from the bed frame
if equipped.
Warning: Possible Shock Hazard. Injury may result
from improper power cable routing. Always follow proper factory-installed routing conguration. Be sure
to check condition of the power cable on a regular
basis. Do not use damaged or frayed power cables.
5. Using a small screwdriver, alternately depress the small red latch tabs above and below AC
receptacle on the control box. Simultaneously, pull gently outward on the power cable where it connects to the control box, until the plug comes
out.
6. Plug the replacement cable into the control box until it latches. Attach the ground wire to the bed frame.
7. Re-route the power cable along the frame in the same position as the original routing.
8. Replace zip-ties along with cables in their original
position and routing to the control box. If this step
is not followed, damage to the cables or electric shock may result.
Figure 33
13
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 14
Joerns® Bed Frames EasyCare®
Troubleshooting
Any loss of electrical function should be investigated
with the following steps:
Event Possible Remedies
Bed not responding to pendant control
Bed not responding to Staff Control
Note: Electrical damage can occur if a non-approved
pendant is connected to the electrical system.
1. Check lockout status.
2. Check connection is fully inserted with cable secured.
3. Check the bed is plugged into power source and control box green light is illuminated
4. Thermal overload condition. This should self-correct in 2 minutes
or less. Do not surpass 10% duty cycle (2 min on/ 18 min off).
1. Check lockout status
2. Check connection is fully inserted with cable secured.
3. Check the bed is plugged into power source and control box green light is illuminated.
To Maximize Service Life
1. Keep the bed clean.
2. Observe bed operation and report any discrepancies or problems to Joerns Healthcare
3. After initial week of use, check all threaded
fasteners for looseness, and make sure all pins are in their normal location and fastened securely. Check monthly for loose bolts, nuts, pins and other retaining hardware. Tighten any loose hardware, and contact Joerns Healthcare to order any
appropriate service parts.
4. Make sure each inspection includes the underside of the bed frame and mattress support platform.
5. Visually inspect the bed frame and accessories
for any cracking, bending, or hole enlargement. remove the bed from service immediately,
and replace the affected parts.
6. Check wiring for proper connections and damage
(fraying, kinking, or deterioration).
7. Check actuators for correct mounting at
attachment points and ensure all related pins are
mounted securely and properly to the bed frame. Actuators are not serviceable, but are replaceable if required.
8. Lubricate pivot point, pins and bolts as required.
The recommended lubricant is Joerns-approved
grease #12500013
Maintenance
Maintenance/Inspection Information:
Visually inspect the bed and accessories for broken
welds or cracks and check for loose hardware on a
monthly basis.
If any broken welds or cracks are found, remove
bed from service immediately and replace affected part(s).
Preventative Maintenance
To ensure maximum life of your product, follow
all warnings and cautions in the User Manual and
maintain your bed with care, as outlined below. The maintenance required will be dictated by your bed’s
usage and care – a thorough inspection should be
conducted monthly.
14
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
If any discrepancies are noted during inspection, they must be corrected before continuing bed frame
use. Major service and repairs must be performed
by authorized Joerns Healthcare
Cleaning
Caution: Possible damage to electrical parts, and/or
shock hazard.
1. Clean all the electrical components (motors, control
box, pendant and cords) on your Joerns damp, well-wrung out cloth. Excess water in or on electrical parts, could damage these components. Do not submerge the pendant in liquid.
2. Your Joerns bed features a durable, baked-on
epoxy nish that is easily cleaned with a mild
solution of antiseptic cleaner and warm water.
Read and follow all label instructions on your
cleaning products.
3. Wood veneer surfaces should be maintained by cleaning with furniture cleaner or polish.
®
bed with
Page 15
Joerns® Bed Frames
EasyCare®
Accessories
All Accessories
Warning: Possible Injury Or Death. Injury or death
may result from improper use of bed accessories. Consult accessory manufacturer for compatibility,
proper installation and use instructions.
Trapeze And Trapeze Adapter
Warning: Possible Injury Or Death. Improper
installation of the Trapeze Adapter could alter or dislodge the head panel position, increasing the risk of injury or suffocation.
Warning: Possible Injury Or Death. Follow the
manufacturer’s directions when installing the Trapeze Bar. Improper installation of the Trapeze Bar on the Trapeze Adapter could cause the Trapeze Bar to disconnect/collapse or otherwise fail, resulting in injury
or death. Warning: Possible Injury Or Death. A minimum of
1-inch clearance between the bottom of the Trapeze Bar and the oor is required when the bed is in its
lowest position. Failure to do so could damage the
oor and/or cause the bed to become unstable, resulting in injury or death.
Trapeze (Figure 34)
1. We recommend using the Joerns
the Joerns Trapeze Adapter.
2. Install the Trapeze assembly according to
manufacturer’s directions included with
the Trapeze.
Figure 34
®
Trapeze with
Drainage Bag Holder
Warning: Possible BioHazard. Prior to use, consult
the Drain Bag manufacturer’s installation and use instructions. Failure to do so could result in a bag
puncture, resulting in a potential biohazard.
Warning: Possible Injury Or Death. Many items
that attach to the Holder are gravity fed and must be
below the bed user to function. Do not allow items attached to the Holder to rest or be positioned on the
oor where they might be stepped on and adversely
affected or damaged. Exercise particular caution when the bed is moved to its lowest position as items
attached to the Holder at that height are especially at risk. Failure to do so could result in injury or death.
Warning: Possible Injury Or Death. Make sure items hung on this holder have proper clearance and are not pinched or caught during bed adjustment. Failure to do
so could result in injury or death.
15
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 16
Joerns® Bed Frames EasyCare®
Trapeze Adapter (Figure 35)
1. The Trapeze Adapter is designed to allow
attachment of the Trapeze system to the EasyCare® Bed. Prior to use, please read
important warnings related to this option and follow all instructions below.
2. Install the Trapeze Adapter according to the
supplied instructions.
3. Install the trapeze assembly onto the Trapeze
Adapter according to the trapeze manufacturer’s
instructions.
4. Operate the bed from its highest through lowest positions to ensure proper clearance and
operation. Some trapeze systems may not be suitable for use with the EasyCare Bed, due to its
low operating height.
5. Adjustment and use of a Trapeze should be in accordance with instructions from a healthcare
professional (doctor or therapist).
6. Do not exceed the maximum load condition listed on installation instructions.
Figure 35
Drainage Bag Holder (Figure 36)
1. The Drainage Bag Holder is designed as an attachment point for items such as urine collection bags. Make sure items attached to holder are
safely secured. Prior to use, please read important
warnings related to the Drainage Bag Holder and follow all instructions below.
2. Install the Drainage Bag Holder and attach items to the Holder according to supplied instructions.
3. It is important to operate all bed functions through their entire ranges to ensure that items attached to the Holder have proper clearance and are not pinched or caught.
4. Attach the Holder only to the point and position
outlined in installation instructions. Operate all
bed functions through their full ranges to ensure that items attached to the Holder have proper clearance and are not pinched or caught.
5. Many items that attach to the Holder are gravity fed and must be below bed user to function. Do not allow items suspended from the Holder to
rest on the oor, particularly when the bed is in its lowest position. Otherwise, a safety condition may occur or the function of attached items may
be impacted.
Figure 36
16
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 17
Battery Back-up
Battery Backup Includes:
Charger with Power Cord (with Cord Clip)
Battery with Cable
Wall Mount Bracket (with two screws)
• User Instructions
Battery Charger
1. Simply mount the Battery Charger in a central
location convenient to caregiver staff. Plug the
Battery Charger into any standard wall socket (110V; 50-60 Hz.).
2. Insert the Battery in the Battery Charger.
• Green light -- indicates main power is on
• Yellow light -- indicates Battery is charging
3. After about four hours of charging, the yellow
light goes off and the Battery will be fully charged.
Note: We recommend that users charge up the
Battery immediately upon receipt.
4. The Battery has a useful life of about 3.5 years,
depending upon usage. It is recommended to
discharge the Battery as little as possible and to recharge quite frequently (by keeping the Battery connected to the Battery Charger when not in use), to provide the longest Battery life.
5. When the Battery needs replacing, dispose of the
old Battery in the same manner you would dispose of a car (lead acid) battery.
Joerns® Bed Frames
EasyCare®
Figure 37
Powering the EasyCare Bed with Battery Backup:
Attach the battery to the Battery Pack Patch Cord Socket at the bed foot end (Figure 37).
Depending on the Battery condition and load on the
bed, one battery can typically raise and lower the
bed, constituting one cycle, 20 to 25 times on single,
full charge.
17
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537
Page 18
Joerns Healthcare Warranty Program
for Joerns
®
EasyCare® bed, Model ECS
Joerns EasyCare bed is guaranteed for a period of four years from the date of delivery, against defects in materials and workmanship, under normal use and
service.
This four-year warranty includes all mechanical and
electrical components. Steel structural components on beds are covered
under warranty for a period of 15 years from the date of delivery.
Welds are covered under warranty for the lifetime
1
of
the product.
Damage caused by use in unsuitable environmental conditions, abuse or failure to maintain the product in
accordance with user and service instructions is not covered.
Any alteration, modication, or repair unless performed by, or authorized in writing by, Joerns Healthcare, will void this warranty.
Parts
Joerns beds contain various parts that wear from
normal use. These parts, such as DC batteries, are not covered under the four-year warranty but are covered for 90 days after date of delivery.
Joerns Healthcare’s obligation under this warranty is limited to supplying replacement parts, servicing or replacing, at its option, any product which is found by
Joerns Healthcare to be defective.
Warranty replacement parts are covered by the terms of this warranty until the product’s original four-year warranty period expires.
When requested by Joerns Healthcare, parts must
be returned for inspection at the customer’s expense.
Credit will be issued only after inspection.
Service
Most service requests can be handled by the facility
Maintenance Department with assistance from the Joerns Healthcare Product Service Department.
Most parts requested can be shipped next day air at
the customer’s expense.
Should a technician be required, one will be provided by Joerns Healthcare, at our discretion. Only the
Joerns Healthcare Product Service Department can
dispatch authorized technicians.
This warranty is extended to the original purchaser of the equipment.
1
Weld lifetime dened as 20 years.
© 2015 Joerns Healthcare • 6110158 RevC • 15-2537