Important Notice: The equipment must be
installed and operated in the manner for which
it was intended. Facility staff/user is responsible
for reading and understanding the product user
manual and contacting Joerns Healthcare, if
anything in this manual is unclear. Joerns will
not be held responsible for any injuries resulting
from failure to comply with the instructions and
precautions in this manual.
Warning: Joerns’ specialty support surfaces
are designed as mattress replacement systems.
The risk of entrapment may occur when the
equipment is placed on bed frames that leave
gaps of even a few inches between the mattress
and the headboard, footboard, and bed or side
rails. The equipment is NOT to be used when
such gaps are present.
Facility staff/user is responsible for ensuring
that all mattresses properly t the bed frames.
Joerns is not responsible for the placement of
its equipment on bed frames that leave gaps
between the mattress and the headboard,
footboard or bed or side rails which present a
risk of harm to residents.
Warning: The safe use of the equipment is
maximized when used in conjunction with bed
rails; there may be an increased risk of falls
when such bed rails are not present. Serious
injury or death can result from the use (potential
entrapment) or non-use (potential resident falls)
of side rails or other restraints. Local policies
regarding the use of side rails should be taken
into account. Whether and how to use side rails
is a decision that should be based on each
resident’s individual needs and should be made
by the physician, caregivers and responsible
parties.
When using the mattress system, always ensure
that the resident is positioned properly within the
connes of the bed. Do not let any extremities
protrude over the side or between the bed rails
when the mattress is being used.
Danger Explosion Hazard: Do not use in the
presence of ammable anesthetics. Do not use
in the presence of smoking materials or open
ame. Air owing through the air mattress will
support combustion.
Danger: To reduce the risk of shock, adhere to the
following instructions. Failure to do so could result in
personal injury or equipment damage.
• Immediately after using the DermaFloat® LR,
unplug it from its power source.
• Do not place or store the product where it can fall
or be pulled into a tub or sink.
• Do not place or drop the product into water or
other liquid.
• Do not remove the back of the control unit. Refer
servicing to Joerns.
Warning: To reduce the risk of burns, shock, re, or
personal injury, adhere to the following instructions.
Failure to do so could result in personal injury or
equipment damage.
1. Use this product only for its intended purpose as
described in this manual. Only use attachments
and/or accessories that are recommended by the
manufacturer.
2. If this product has a damaged power cord or plug,
is not working properly, has been dropped or
damaged, or has been dropped into water, do not
operate it. For examination and repair, return the
product to Joerns.
3. Keep the control unit and power cord away from
heated surfaces, e.g. space heaters.
4. Never block the air openings of the product. Do
not place the control unit on a surface, such as a
bed or couch, where the air opening and/or lter
compartment, located on the back of the control
unit, may be blocked. Keep the air openings free
of lint and hair.
5. Never drop or insert any object into any
opening or hose.
6. Do not spill food or liquids onto the control unit. If
a spillage does occur, turn off the unit, disconnect
it from its power supply and allow at least 24 hours
for drying.
7. Do not use the product outdoors, or where
aerosol-spray products are used.
8. Plug this product only into a properly grounded
outlet. Refer to “Grounding Instructions”.
9. Ensure nothing is placed on the power cord and
ensure it is not located where it can be stepped on
or tripped over.
10. Do not attempt to service the control unit. Please
call Joerns for any service requests.
11. The therapy pad (top cover) of this product is not
air permeable and may present a suffocation risk.
It is the responsibility of the caregiver to ensure
that the resident can use this product safely.
Although essential in the practice of long-term care,
bedside rails, in recent years, have also been a
subject of regulatory review and evolution in design
and use.
That focus includes not only the challenge of
achieving an appropriate balance between resident
security and unnecessary restraint, but also the
additional safety issue of entrapment.
The U.S. Food and Drug Administration (FDA),
working with our company and other industry
representatives has addressed the potential danger
of entrapment with new safety guidelines for medical
beds. These guidelines recommend dimensional
limits for critical gaps and spaces between bed
system components.
Entrapment zones involve the relationship of
components often directly assembled by the
healthcare facility rather than the manufacturer.
Therefore, compliance is the responsibility of the
facility.
Creating a Safer Care Environment
While the guidelines apply to all healthcare
settings, (hospitals, nursing homes and at
home), long-term care facilities have particular
exposure since serious entrapment events
typically involve frail, elderly or dementia
patients.
For More Information
To learn more about compliance options with
Joerns products, visit our website at www.
joerns.com, or contact our Customer Care reps
at 800-826-0270 and ask for free informational
publications.
To learn more about entrapment zones,
assessment methods and guidelines
concerning entrapment, contact Joerns
Healthcare at 800-826-0270 or consult the FDA
website: http://www.fda.gov/MedicalDevices/
ProductsandMedicalProcedures/
GeneralHospitalDevicesandSupplies/
HospitalBeds/default.htm.
As the leading manufacturer of long-term care beds
and a frontrunner in addressing this critical issue,
Joerns Healthcare can offer you the expertise,
assistance and products to bring your facility into
compliance.
Joerns® Compliance Solutions
Matching the right bed components in order to meet
regulatory guidelines can be complex.
That is why Joerns offers a wide array of compliance
options. We assist customers in selecting compliant
Important Precautions ................................................................................................................................ 2
Indications for Use ...................................................................................................................................... 5
Features ........................................................................................................................................................ 6-7
Therapy Control Unit Features
Therapy Mattress System Features
Therapy Pads
Additional Features ..................................................................................................................................... 9-10
CPR
Transport
Power Failure
Low Pressure Alarm
Optional Accessories
Recommended Linen
Changing the Therapy Pad
Patient Positioning and Comfort
Incontinence
Safety Information ...................................................................................................................................... 11
Storage and Care ......................................................................................................................................... 12
Control Unit
Support Surface
System Specifications ................................................................................................................................ 12
Call for Assistance ....................................................................................................................................... 12
The DermaFloat® LR, provided by Joerns Healthcare,
is a unique, continuous bilateral rotation mattress
replacement system. This system is designed to
assist in the effective management of residents at risk
from the complications of immobility while providing
low air loss therapy. Low air loss therapy has been
demonstrated to reduce the risk of pressure ulcers
as well as being a valuable aid in the treatment of
pressure ulcers.
Warning: The risk of entrapment can arise when
equipment is placed on bed frames that leave gaps
of even a few inches between the mattress and the
headboard, footboard, and bed or side rails. The
equipment is NOT to be used when such gaps are
present. See “Important Precautions” section of
this manual.
The DermaFloat LR mattress replacement system is
suitable for both the treatment of existing pressure
ulcers stage I through stage IV as well as those who
have been assessed at risk from the complications of
immobility. The DermaFloat LR is quiet, comfortable
and simple enough for single caregiver installation,
featuring rapid ination in just 15 minutes or less. The
user-friendly controls allow for easy adjustment of
resident comfort.
Additionally, low friction and low shear materials,
together with average interface pressures well below
capillary closure levels, means that the DermaFloat LR
meets the comfort and clinical requirements of your
residents up to 350 lbs.
Shear and Friction Reduction
Shearing occurs when the skin is stationary in relation
to the support surface, while the underlying tissues
and vessels are stretched and damaged. When a
resident’s skin rubs against another surface, the result
is friction. The top surface of the DermaFloat LR
therapy pad is constructed from a very smooth nylon
fabric with low friction and low shear properties to
protect the resident’s skin from these damaging forces.
Indications for Use
Note: The selection of a pressure-relieving surface
needs to be based on each individual resident’s clinical
condition, diagnosis and/or co-morbidities. The choice
and use of a support surface is one factor in a holistic
program of wound care
and treatment.
Spinal Cord Injury
The DermaFloat LR is not recommended for use by
residents with unstable spinal fractures. Advice should
be obtained from the appropriate physician before
using the DermaFloat LR system for these residents.
the four key areas in the treatment of compromised
skin: pressure relief, moisture control, and reduction in
both friction and shearing forces.
Moisture Control
Residents are at risk for skin maceration if excess
moisture is permitted to accumulate beneath the
resident. This may be due to perspiration, incontinence
or wound drainage.
On the DermaFloat LR, moisture is controlled via the
specially treated breathable, uid-proof, urethane
coated nylon therapy pad. The moisture vapor
permeable fabric of the therapy pad allows a sufcient
amount of air to circulate beneath the pad and wicks
away excess moisture.
AIDS Arthritis
Oncology
The DermaFloat® LR provides distribution of weight
over a wide surface area, which relieves pressure
against bony prominences and provides a soft, gentle
therapy surface to lie on. For residents experiencing
severe pain and discomfort due to pressure and/or
positioning limitations, consider the DermaFloat LR as
an adjunct to pain management interventions.
Note: Pressure relief and pain management are
conditions and diagnoses for which the DermaFloat LR
may be indicated. Occasionally, there are orthopedic
and neurological residents that require body
positioning to be maintained in specic alignment. The
use of the DermaFloat LR for these residents should
be considered on an individual basis and discussed
with the attending physician.
• True low air loss with up to 100 liters of airow per
minute.
• Three modes of operation – Autorm, Turn and
Static.
• Static mode (default) provides continuous low air
loss pressure relief.
• Autorm mode provides maximum air ination
designed to assist both residents and caregivers
during resident transfer
and treatment.
• Pause feature allows the caregiver to pause
the mattress rotation at any point for resident
emergencies or care.
• Automatic panel lock out to avoid unwanted or
accidental adjustments.
• Nine therapeutic comfort control settings to
maximize resident compliance and
promote healing.
• Closed loop pressure sensor control system
eliminates concerns of changes in mattress
interface pressure due to ambient temperature and
pressure changes.
• Quick disconnect hose feature allows for rapid
attach and CPR deate at the control unit.
• Compact lightweight control unit is quiet, robust
and powerful, with a reusable air lter.
• Crisp, easy to read graphics for intuitive set up and
therapy control.
• Integrated swing out hanging brackets for afxing
to most bed types.
• Quilted therapy pad is vapor permeable,
breathable, uid-proof, and minimizes friction
and shearing.
• Durable base tub is constructed from 100%
heavy weight 1680 denier nylon with a 1.5
oz. urethane coating and incorporates bed
attachment loops for stability.
• Supports up to 350 lbs. resident weight
• CPR deation
• Anti-kink, easy clean air supply hose set
Therapy Pads
Many healthcare facilities are facing the challenge of
infection control. Joerns quilted therapy covers are
treated with an antimicrobial to protect the therapy
pad itself from the growth of mold, mildew and odorcausing bacteria.
Key features and benets:
• Treated with a highly effective bacteriostat agent
to inhibit the growth of bacterial and fungus.
• Constructed from a very smooth nylon fabric
with low friction and low shear properties to
protect the resident’s skin from damaging
friction/shearing forces.
• Breathable, moisture vapor permeable fabric
allows air to circulate beneath the pad and wicks
away excess moisture. This keeps your resident
dry and helps to prevent skin maceration.
• Two-way stretch therapy pad is designed for
optimal comfort, moisture vapor transfer, stain
resistance and ease of laundering
Therapy Mattress System Features
• Twenty individual therapy cells help to evenly
distribute the resident’s weight and maximize
pressure relief. Modular cell design mattress for
ease of cleaning, re-assembly and cost
effective service.
• Lower two-inch therapy support cell provides
additional therapeutic support and remains inated
for up to 12 hours in the event of a power failure.
• Eight-inch deep therapy cells are constructed
of highly durable, polyurethane coated nylon to
provide adequate support and prevent “bottoming
out” for most residents within
weight limit.
• Ten-inch safety side bolsters provide comfort and
support for residents during the
rotation cycle.
Warning: Use a properly grounded, three-prong, 120V
AC outlet for this product. Failure to use a grounded
outlet could result in personal injury or damage to
equipment or house wiring, including risk of re. A
qualied electrician should be contacted to correct the
wiring and ensure a properly
grounded outlet.
Before installing this product, have the electrical
system checked to make sure the electrical circuits
and the electrical service are properly grounded.
Having a three-prong outlet does not necessarily
mean it is grounded. Sometimes two-prong outlets are
replaced with a three-prong type even though there is
no ground wire.
There is always a chance of a loose connection or
poor installation of a ground wire that causes the loss
of proper ground at the outlet. Inadequate grounding
at electrical outlets can occur even if there is a
ground wire. Wires can become loose over time at the
connection to the outlet.
Note: To install new wires on a circuit requires a
qualied electrician.
How to Determine if Your Outlet has the
Proper Grounding
Most hardware stores sell circuit testers (Figure 2) that
can be used to test an outlet for proper grounding.
The tester plugs into an outlet and by observing
the indicator lights you can determine if the outlet is
properly grounded. For a higher level of assurance, an
electrician should be requested to thoroughly test the
electrical system with more reliable equipment.
If repair or replacement of the cord or plug is
necessary, please contact Joerns Healthcare for
assistance.
Setup
Warning: For important precautions, please see
page two.
Caution: Do not place the
control unit on the oor.
Position the power cord to
keep personnel from tripping
over it.
Note: The DermaFloat LR
system must be installed
on bed frames that are
equipped with side rails.
Please raise side rails on
the bed and lock them in
position after the resident
is on the mattress. Never
leave resident unattended
on mattress system with bed
side rails in the
down position.
When the product is not in use, properly store
the power cord. Failure to do so could result in
personal injury.
• Remove the existing mattress from the bed.
• Place the DermaFloat LR mattress with the
hose connection at the foot end of the bed and
the therapy cells facing up. Secure the straps
on the mattress securely to the movable part
of the bed frame.
• If the therapy pad is not already on the
mattress, place it on the mattress. Attach the
elastic straps to the mattress buckles around
each corner of the mattress. Attach the six (6)
additional straps to the movable part of the
bed frame.
• Hang the control unit on the foot of the bed
facing away from the bed. Attach the hose
connector marked CPR to the control unit.
• Plug in the control unit and the press the
Power key. The control unit will start and the
Power light will illuminate.
• Allow up to a minute for full ination. Place
the resident on the mattress. Mattress can be
inated with resident on it, but will take longer,
depending on their weight/size. Note: Keep
the control unit on while the resident is on
the mattress.
• Position the resident’s head in the position that
the resident will be in for the largest portion of the
day. If the resident is lying at, please use three
(3) ngers for the hand check. If the resident will
be sitting up for the majority of the day, please use
two (2) ngers.
• Perform a hand check by placing ngers locally
under the resident’s buttocks between two
cushions. The resident should not bottom out. If
they do, increase the therapy control by one LED
(light), until they no longer
bottom out.
• The safety side bolsters cannot be inated unless
the CPR plug is installed. The plug is located
on the hose assembly between the mattress
and the control unit. Caution: The safety side
bolsters should always be inated when resident is
unattended on the mattress and/or the mattress is
in Turn mode.
Operation
Warning: For important precautions, please see
page two.
Caution: The resident’s head should be positioned
in the center of the top section of the mattress. When
using the mattress system always ensure that the
resident is positioned properly within the connes of
the bed. Do not let any extremities protrude over the
side or between the bed rails when the mattress is
being used.
Resident Comfort Controls and Monitoring
Power
The Power button is used to turn the power on and off.
Autorm
Autorm mode provides maximum air ination
designed to assist both residents and caregivers
during resident transfer and treatment. The unit will
automatically return to the mode it was in prior to
Autorm (either Turn or Static) in approximately
15 minutes.
Pause
Use this feature to immediately stop all rotation and
hold the current mattress position. The mattress will
hold this position until this feature is turned off by
pressing the Pause button a second time.
Figure 3
Lockout
This unit is designed to lock out all of the adjustment
controls after the resident has been positioned
correctly. Ten minutes after the last button has been
pushed, push the Power light begins to ash indicatingLockout is enabled. This feature is to prevent any
unauthorized changes to the resident settings. To
unlock and make adjustments to the settings, press
both up and down comfort arrows at the same time.
Lockout mode will return after ten minutes.
Note: The unit is designed to lock out all the
adjustment controls after the resident has been
positioned correctly. In approximately ten minutes after
the last button push the power on light begins to ash
indicating Lockout is enabled.
Comfort Adjust
The Comfort Adjust function is located across the
top of the control panel. The DermaFloat® LR can be
customized to meet individual resident needs within
a therapeutic window. Use the soft and rm keys to
simultaneously increase or decrease pressure in the
entire mattress in Turn and Static modes. The current
setting is shown in the LED display in the center of the
control panel.
Turn Mode
The Turn mode and sets the desired rotation type. The
resident can be rotated to either side individually or to
both sides in a continuous cycle.
Turn Time
Use the Turn Time feature to set the time cycle for
the selected Turn mode. The DermaFloat LR can be
placed into a 10, 20, or 60 minute cycle based on what
is determined to be the optimal therapy setting for the
individual resident.
The Turn Angle feature allows for the desired turn
angle to be selected. The mattress can make a half or
full turn based on the optimal therapy for
the resident.
Additional Features
Warning: A possible re hazard exists. This product is
suitable for use with oxygen administering equipment
of the nasal, mask, or half bed-length, tent-type only.
To prevent personal injury or equipment damage,
ensure that the oxygen tent does not extend below
the mattress.
CPR
The hose connection at the control unit is marked
CPR. Disconnect the hose from the control unit.
Deation times will vary based on resident weight and
prole. To resume therapy, reattach the hose to the
control unit.
Transport
To transport the resident in bed, turn the control unit
off. Unplug the power cord from the outlet. Do not
disconnect the hose connection at the control unit.
The lower support cell will prevent the resident from
bottoming out for up to 12 hours.
Power Failure
If power fails to the system an alarm will sound and
the Power Fail LED will turn on until power is restored.
When the power is restored, the control unit will return
to the previous comfort adjust setting in Static mode.
The Turn mode must be reselected manually.
Low Pressure Alarm
If the control unit senses a low pressure condition,
such as a disconnected hose or therapy cell, then it
will activate an audible alarm and will ash “L” and “P”
on the display. Once the low pressure condition has
been corrected, the alarm will cease and the control
unit will return to normal operation.
Optional Accessories
• Additional therapy pads – available for purchase
Troubleshooting
Therapy Surface is Not Inflating
• Ensure the hose connection from the therapy
mattress system (mattress) to the control unit is
securely connected.
• Ensure that the control unit is plugged into an AC
outlet.
• Ensure that the power is not on Standby. If on
Standby, press the Power button.
• Ensure that all air cells are connected to the
internal mattress manifold.
Unable to Change Therapy Mode or Adjust
Comfort Control
Make sure the Lockout function is disabled. To
disable, press the up and down Comfort Adjust arrows
simultaneously.
Troubleshooting Modes of Operation
The DermaFloat® LR is designed to offer continuous
bilateral or unilateral rotation.
Continuous bilateral rotation (Turning the resident
to both left and right)
1. Ensure the mattress is fully inated with the
resident held centrally on the support surface
and correctly supported. If necessary, make
adjustments using the comfort control buttons.
2. Check to make sure the control panel is not locked
out. If the Power LED is ashing, then press the Comfort Control arrows together to unlock the
system.
3. Verify the system is in Static mode (Turn mode
None LED will be lit).
4. Press the Turn mode button until the Both LED is
lit.
5. Press the Turn Angle button until the desired target
angle LED is illuminated.
6. Select the Turn Cycle time by pressing the Turn
Time button until the desired time period is
Caution: Before commencing rotation, it is important
to ensure that all lines/feeds to the resident are free
from obstruction and have sufcient length for the
angle of turn required.
Caution: Observe the resident through at least one full
rotation cycle to ensure resident safety and tolerance
of angles selected.
Continuous Unilateral Rotation (Turning the
resident to only the left or only the right)
1. Ensure the mattress is fully inated with the
resident held centrally on the support surface
and correctly supported. If necessary, make
adjustments using the comfort control buttons.
2. Ensure that the control unit is not locked. If the
Power LED is ashing, then press the Comfort
Control arrows together to unlock the system.
3. Verify the system is in Static mode (Turn mode
None LED will be lit).
4. Press the Turn mode button until either the Left or
Right LED is lit.
5. Press the Turn Angle button until the desired target
angle LED is illuminated.
6. Select the Turn Cycle time by pressing the Turn
Time button until the desired time period is
illuminated.
Nursing Procedures
Recommended Linen:
Special linens are not necessary for the DermaFloat
LR. There is no need for a bottom sheet as the therapy
pad should be covering the therapy cells at all times.
The resident should never be lying directly on the
therapy cells. Upon the resident’s specic needs, the
following linens may be utilized:
• Draw or slide sheet to aid in positioning and to
further minimize friction and shearing
• Incontinence barrier pad for residents incontinent
of urine and/or stool, and residents with heavily
draining wounds
• Add top sheet, blanket and/or bedspread as
needed for resident comfort
• Keep the amount of padding between the resident
and bed to a minimum for
optimum performance
Changing the Therapy Pad
• Place the therapy pad over the therapy cells, tting
the corner of the cushions into the corner of the
therapy pad. (Similar to a tted sheet)
• Zip the therapy pad along each side of the
mattress tub.
Resident Positioning and Comfort
General Repositioning
Residents should be turned and repositioned per
individual turning schedule or per facility policy. It may
be helpful to activate the Autorm mode to achieve
a rm surface for repositioning purposes. The unit
will automatically return to the mode it was in prior
to Autorm in approximately 15 minutes or you can
manually return to therapy mode once resident has
been repositioned.
Unless counter indicated, it is desirable to keep
the head of the bed in the low position to provide
optimal pressure relief and minimize the risk of
shearing injuries.
Elevating Resident into Sitting Position
The special properties of the DermaFloat® LR therapy
pad reduce the opportunity for shear and friction that
may occur when raising the head of other beds. As
with any surface, sliding can be expected, therefore
residents should be repositioned after elevation. The
knee gatch or foot of the bed may be elevated rst, to
help prevent the resident from sliding when the head
of the bed is elevated.
Incontinence
Moisture against the skin surface leads to maceration,
or softening of the tissues. To prevent maceration, it
is recommended that an incontinence barrier pad be
used to absorb excess moisture.
In the event of incontinence or excess drainage on
the therapy pad, wipe off the excess uid from the bed
Specialty bed products are designed to reduce/
relieve pressure and the shearing/friction forces on the
resident’s skin. The risk of gradual movement and/or
sinking into hazardous positions of entrapment and/
or inadvertent bed exit may be increased due to the
nature of these products.
Traction
With any traction or unstable fractures, maintain
physician-directed angle of articulation and guard
against risks of resident migration or inadvertent
deation of resident surface.
Skin Care
Monitor skin conditions regularly, particularly in areas
where incontinence and drainage occur or collect,
and consider adjunct or alternative therapies for high
acuity residents. Early intervention may be essential to
preventing serious skin breakdown.
Bed Height
To minimize the risks of falls or injury, the resident
surface should always be in the lowest practical
position when the resident is unattended. Make sure
areas under and around the frame are clear of objects,
persons and parts of body before adjusting height.
Control Unit
Wipe off dust. If necessary, clean the housing exterior
with a disinfectant solution or a mild detergent and a
damp cloth. Then wipe dry.
General Cleaning
If there is no visible soilage with possible body uids,
it is recommend that the mattress system be cleaned
with a mild detergent and warm water. If disinfection is
desired, a combination cleanser/disinfectant may be
used as explained in “Disinfecting” area.
• Resident care equipment that does not come into
contact with mucous membranes or
non-contact skin, requires low-level disinfection.
Wiping surfaces with a properly prepared
detergent or disinfectant carries out low level
disinfecting.
• Processing of dirty resident care equipment should
take place in a designated area away from clean
or sterile supplies and food preparation areas.
• Detergent/disinfectants should not be mixed with
other germicides or detergents. Using the proper
dilution insures the most effective killing power of
the disinfectant.
• Wash hands often and well, including after
removal of gloves.
• Resident care equipment that is used in isolation
areas should be disinfected in accordance with all
internal policies and procedures regarding such
equipment.
Disinfecting
Cleaning
Warning: Unplug the control unit from its power
source. Failure to do so could result in personal injury
or equipment damage.
Warning: Do not expose the unit to excessive
moisture that would allow for liquid pooling. Personal
injury or equipment damage could occur.
Caution: Do not use harsh cleansers/detergents, such
as scouring pads and heavy-duty grease removers, or
solvents, such as acetone. Equipment damage
could occur.
When there is visible soilage and between residents, it
is recommend that the unit and mattress be disinfected
with a tuberculocidal disinfectant. Disinfectant should
be registered with the Environmental Protection
Agency (EPA).
• Use rubber gloves and eye protection.
• Prepare detergent/disinfectant (registered by EPA
as hospital disinfectant) solution according to
instructions on label for correct use-dilution.
• With support surface deated, thoroughly wipe
down entire mattress, as air cells will lie at. Be
sure to reach all areas underneath and in-between
air cells. Allow to air dry.
• If dust or other soiling has accumulated along
air hoses, remove using swabs moistened with
detergent/disinfectant as necessary. Allow all
components to air dry. Wrap mattress in plastic
and return to storage area.
• Thoroughly wipe down outside of control unit and
allow to air dry. Cover with plastic and return to
storage area
The therapy pad can be wiped down with a disinfectant
solution or a mild detergent with a damp cloth. If
heavily soiled, the therapy pad can be laundered in a
washer and dryer with warm water (no more than 120º
Fahrenheit). A non-bleach detergent should be used
sparingly. Wipe dry or allow to air dry.
Steam Cleaning
Do not use any steam cleaning device on the unit.
Excessive moisture can damage mechanisms in
this unit.
Filter Cleaning
Check the air lter on the rear of the unit regularly for
buildup of dust/dirt. If buildup is visible, turn off the
control unit and disconnect the power cord from the
wall outlet. Remove the lter by grasping the lter
pulling outward. Replace with the second supplied
lter. Ensure the replaced lter covers the entire
lter region.
Hand-wash the removed lter in warm soapy water
and allow to air dry. When dry, store the lter in a safe
place for the next lter maintenance.
Support Surface
Gently roll up the support surface, expelling any
residual air, for temporary storage. The mattress
should be wrapped in plastic and/or a clean bag
for storage.
System Specifications
Standard Features
Weight
Control unit: ............................... 14 lbs (6 Kg)
Mattress: .................................... 22 lbs (11 Kg)
Maximum weight capacity*: ....... 350 lbs (159 Kg)
Dimensions
Control unit:
5.75" (15 cm) W x 10.5" (27 cm) H x 12" (30 cm) D
Mattress:
35" (89cm) W x 80" (203cm) L x 10" (25cm) D
Electrical Specifications
USA
120V AC, 60 Hz, 0.6A
Maintenance
Warning: Only facility-authorized personnel trained
by Joerns Healthcare should perform preventative
maintenance. Preventative maintenance performed by
unauthorized personnel could result in personal injury
or equipment damage.
Any maintenance done without Joerns’s authorization
will invalidate any warranties on
this product.
Storage and Care
Note: Clean the DermaFloat LR as described in the
previous section prior to storage.
Control Unit
The power cord may be wrapped around the unit for
convenience. Wrap the unit in a plastic bag for dust
resistance, then store the unit in an area appropriate
for an electronic medical device.
Environmental Conditions
Operating Conditions
Ambient Temperature: +10ºC to +40ºC
Relative Humidity: 30% to 75% Non-Condensing
Storage and Shipping Conditions
Ambient Temperature: +10ºC to +40ºC
Relative Humidity: 10% to 100%
Agency Approvals
• UL Classied Medical Equipment No. 60601.1
• UL 60601-1 Can/CSA C22.2
UL Classication refers to the power unit only, not the
complete mattress replacement system.
Call for Assistance
If you have any questions or require service on a
Joerns product, please call Joerns Healthcare at
LR mattress to be sold free from defects in
workmanship and materials, under normal and
proper use, for a period of two (2) years on
the mattress, and one (1) year on the cover
and electromechanical mattress components
(compressors, valves, printed circuit boards,
hoses, and couplers). Damages arising from
improper use will not be covered by this warranty.
Improper use is dened as, but not limited to those
caused by:
• Burns
• Use of improper chemical agents
• Needle punctures, cuts, or abrasions
• Excessive loads
• Staining
• Negligent or excessive usage
• Improper maintenance, handling and/or
cleaning
• Failure to use in the manner indicated in the
DermaFloat LR user manual
Any modication, repair or alteration done to the
DermaFloat LR that was not authorized in writing
by Joerns will void this warranty.
Parts
Joerns’s DermaFloat LR contains various parts that
wear from normal use. Joerns’ obligation under this
warranty is limited to supplying replacement parts,
servicing or replacing, at its option, any product which
is found by Joerns to be defective. When requested
by Joerns, parts must be returned for inspection at the
customer’s expense. Credit will be issued only after
inspection.
Service
Most service requests can be handled by the facility
Maintenance Department with assistance from the
Joerns Product Service Department.
Most parts requested can be shipped next day air at
the customer’s expense.
Should a technician be required, one will be provided
by Joerns, at our discretion. Only the Joerns Product
Service Department can dispatch authorized
technicians.
Damage caused by use in unsuitable
environmental conditions, abuse or failure to
maintain the product in accordance with user and
service instructions is not covered.
This warranty is extended to the original purchaser
of the equipment.
Post Acute, Acute, HomeCare