Joerns Healthcare 6110064-RevB-Final-DermaFloat-LR-Manual-v1 User-Service Manual Joerns Support Surface DermaFloat® LR Model

Page 1
User-Service Manual
Joerns® Support Surface
DermaFloat® LR Model
To avoid injury, read user’s manual before using.
Page 2
Joerns Support Surface DermaFloat LR Model
Important Precautions
Important Notice: The equipment must be installed and operated in the manner for which it was intended. Facility staff/user is responsible for reading and understanding the product user manual and contacting Joerns Healthcare, if anything in this manual is unclear. Joerns will not be held responsible for any injuries resulting from failure to comply with the instructions and precautions in this manual.
Warning: Joerns’ specialty support surfaces are designed as mattress replacement systems. The risk of entrapment may occur when the equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the headboard, footboard, and bed or side rails. The equipment is NOT to be used when such gaps are present.
Facility staff/user is responsible for ensuring
that all mattresses properly t the bed frames.
Joerns is not responsible for the placement of its equipment on bed frames that leave gaps between the mattress and the headboard, footboard or bed or side rails which present a risk of harm to residents.
Warning: The safe use of the equipment is maximized when used in conjunction with bed rails; there may be an increased risk of falls when such bed rails are not present. Serious injury or death can result from the use (potential entrapment) or non-use (potential resident falls) of side rails or other restraints. Local policies regarding the use of side rails should be taken into account. Whether and how to use side rails is a decision that should be based on each resident’s individual needs and should be made by the physician, caregivers and responsible parties.
When using the mattress system, always ensure that the resident is positioned properly within the
connes of the bed. Do not let any extremities
protrude over the side or between the bed rails when the mattress is being used.
Danger Explosion Hazard: Do not use in the presence of ammable anesthetics. Do not use
in the presence of smoking materials or open ame. Air owing through the air mattress will support combustion.
Danger: To reduce the risk of shock, adhere to the following instructions. Failure to do so could result in personal injury or equipment damage.
• Immediately after using the DermaFloat® LR,
unplug it from its power source.
• Do not place or store the product where it can fall
or be pulled into a tub or sink.
• Do not place or drop the product into water or
other liquid.
• Do not remove the back of the control unit. Refer
servicing to Joerns.
Warning: To reduce the risk of burns, shock, re, or
personal injury, adhere to the following instructions. Failure to do so could result in personal injury or equipment damage.
1. Use this product only for its intended purpose as described in this manual. Only use attachments and/or accessories that are recommended by the manufacturer.
2. If this product has a damaged power cord or plug, is not working properly, has been dropped or damaged, or has been dropped into water, do not operate it. For examination and repair, return the product to Joerns.
3. Keep the control unit and power cord away from heated surfaces, e.g. space heaters.
4. Never block the air openings of the product. Do not place the control unit on a surface, such as a
bed or couch, where the air opening and/or lter
compartment, located on the back of the control unit, may be blocked. Keep the air openings free of lint and hair.
5. Never drop or insert any object into any opening or hose.
6. Do not spill food or liquids onto the control unit. If a spillage does occur, turn off the unit, disconnect it from its power supply and allow at least 24 hours for drying.
7. Do not use the product outdoors, or where aerosol-spray products are used.
8. Plug this product only into a properly grounded outlet. Refer to “Grounding Instructions”.
9. Ensure nothing is placed on the power cord and ensure it is not located where it can be stepped on or tripped over.
10. Do not attempt to service the control unit. Please call Joerns for any service requests.
11. The therapy pad (top cover) of this product is not air permeable and may present a suffocation risk. It is the responsibility of the caregiver to ensure that the resident can use this product safely.
2
2
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
Page 3
Joerns Support Surface
Save These Instructions for tFuture Reference
Bed System Entrapment Information
Although essential in the practice of long-term care,
bedside rails, in recent years, have also been a subject of regulatory review and evolution in design and use.
That focus includes not only the challenge of achieving an appropriate balance between resident security and unnecessary restraint, but also the additional safety issue of entrapment.
The U.S. Food and Drug Administration (FDA),
working with our company and other industry representatives has addressed the potential danger of entrapment with new safety guidelines for medical beds. These guidelines recommend dimensional limits for critical gaps and spaces between bed system components.
Entrapment zones involve the relationship of components often directly assembled by the healthcare facility rather than the manufacturer. Therefore, compliance is the responsibility of the facility.
Creating a Safer Care Environment
While the guidelines apply to all healthcare settings, (hospitals, nursing homes and at home), long-term care facilities have particular exposure since serious entrapment events typically involve frail, elderly or dementia patients.
For More Information
To learn more about compliance options with Joerns products, visit our website at www. joerns.com, or contact our Customer Care reps at 800-826-0270 and ask for free informational publications.
To learn more about entrapment zones, assessment methods and guidelines concerning entrapment, contact Joerns
Healthcare at 800-826-0270 or consult the FDA website: http://www.fda.gov/MedicalDevices/
ProductsandMedicalProcedures/
GeneralHospitalDevicesandSupplies/
HospitalBeds/default.htm.
As the leading manufacturer of long-term care beds
and a frontrunner in addressing this critical issue, Joerns Healthcare can offer you the expertise, assistance and products to bring your facility into compliance.
Joerns® Compliance Solutions
Matching the right bed components in order to meet regulatory guidelines can be complex.
That is why Joerns offers a wide array of compliance options. We assist customers in selecting compliant
accessories recommended for their specic bed
model.
3
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
3
Page 4
Joerns Support Surface DermaFloat LR Model
Table of Contents
Important Precautions ................................................................................................................................ 2
Introduction ................................................................................................................................................. 5
Moisture Control Shear and Friction Reduction
Indications for Use ...................................................................................................................................... 5
Spinal Cord Injury Pressure Relief Pain Management
Features ........................................................................................................................................................ 6-7
Therapy Control Unit Features Therapy Mattress System Features Therapy Pads
Grounding Instructions ............................................................................................................................... 7
Setup ............................................................................................................................................................ 7
Operation ..................................................................................................................................................... 8
Patient Comfort Controls and Monitoring
Additional Features ..................................................................................................................................... 9-10
CPR Transport Power Failure Low Pressure Alarm Optional Accessories
Troubleshooting ........................................................................................................................................... 9-10
Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Troubleshooting Modes of Operation
Nursing Procedures ..................................................................................................................................... 10-11
Recommended Linen Changing the Therapy Pad Patient Positioning and Comfort Incontinence
Safety Information ...................................................................................................................................... 11
Patient Migration Traction Skin Care Bed Height
Cleaning ........................................................................................................................................................ 11-12
Control Unit General Cleaning Disinfecting Therapy Pad Steam Cleaning Filter Cleaning
Maintenance ................................................................................................................................................. 12
Storage and Care ......................................................................................................................................... 12
Control Unit Support Surface
System Specifications ................................................................................................................................ 12
Electrical Specifications Environmental Conditions Agency Approvals
Call for Assistance ....................................................................................................................................... 12
Warranty ....................................................................................................................................................... 14
4
4
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
Page 5
Joerns Support Surface
Introduction
The DermaFloat® LR, provided by Joerns Healthcare,
is a unique, continuous bilateral rotation mattress replacement system. This system is designed to assist in the effective management of residents at risk from the complications of immobility while providing low air loss therapy. Low air loss therapy has been demonstrated to reduce the risk of pressure ulcers as well as being a valuable aid in the treatment of pressure ulcers.
Warning: The risk of entrapment can arise when equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the headboard, footboard, and bed or side rails. The equipment is NOT to be used when such gaps are present. See “Important Precautions” section of this manual.
The DermaFloat LR mattress replacement system is
suitable for both the treatment of existing pressure ulcers stage I through stage IV as well as those who have been assessed at risk from the complications of
immobility. The DermaFloat LR is quiet, comfortable
and simple enough for single caregiver installation,
featuring rapid ination in just 15 minutes or less. The
user-friendly controls allow for easy adjustment of resident comfort.
Additionally, low friction and low shear materials,
together with average interface pressures well below
capillary closure levels, means that the DermaFloat LR
meets the comfort and clinical requirements of your residents up to 350 lbs.
Shear and Friction Reduction
Shearing occurs when the skin is stationary in relation to the support surface, while the underlying tissues and vessels are stretched and damaged. When a resident’s skin rubs against another surface, the result
is friction. The top surface of the DermaFloat LR
therapy pad is constructed from a very smooth nylon fabric with low friction and low shear properties to protect the resident’s skin from these damaging forces.
Indications for Use
Note: The selection of a pressure-relieving surface needs to be based on each individual resident’s clinical condition, diagnosis and/or co-morbidities. The choice and use of a support surface is one factor in a holistic program of wound care and treatment.
Spinal Cord Injury
The DermaFloat LR is not recommended for use by residents with unstable spinal fractures. Advice should
be obtained from the appropriate physician before
using the DermaFloat LR system for these residents.
Pressure Relief
Pressure Ulcers Rehabilitation Respiratory Conditions Neurology
Dermatology Burns Amputations
Pain Management
We have ensured that the DermaFloat LR addresses
the four key areas in the treatment of compromised skin: pressure relief, moisture control, and reduction in both friction and shearing forces.
Moisture Control
Residents are at risk for skin maceration if excess moisture is permitted to accumulate beneath the resident. This may be due to perspiration, incontinence or wound drainage.
On the DermaFloat LR, moisture is controlled via the specially treated breathable, uid-proof, urethane
coated nylon therapy pad. The moisture vapor
permeable fabric of the therapy pad allows a sufcient
amount of air to circulate beneath the pad and wicks away excess moisture.
AIDS Arthritis
Oncology
The DermaFloat® LR provides distribution of weight
over a wide surface area, which relieves pressure against bony prominences and provides a soft, gentle therapy surface to lie on. For residents experiencing severe pain and discomfort due to pressure and/or
positioning limitations, consider the DermaFloat LR as
an adjunct to pain management interventions.
Note: Pressure relief and pain management are
conditions and diagnoses for which the DermaFloat LR
may be indicated. Occasionally, there are orthopedic and neurological residents that require body
positioning to be maintained in specic alignment. The use of the DermaFloat LR for these residents should
be considered on an individual basis and discussed with the attending physician.
5
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
5
Page 6
Joerns Support Surface DermaFloat LR Model
Features
The DermaFloat LR is comprised of two components:
Therapy control unit
Therapy mattress system
Therapy Control Unit Features
• True low air loss with up to 100 liters of airow per
minute.
Three modes of operation – Autorm, Turn and
Static.
Static mode (default) provides continuous low air loss pressure relief.
• Autorm mode provides maximum air ination
designed to assist both residents and caregivers during resident transfer and treatment.
Pause feature allows the caregiver to pause the mattress rotation at any point for resident emergencies or care.
• Automatic panel lock out to avoid unwanted or
accidental adjustments.
Nine therapeutic comfort control settings to maximize resident compliance and promote healing.
Closed loop pressure sensor control system eliminates concerns of changes in mattress interface pressure due to ambient temperature and pressure changes.
Quick disconnect hose feature allows for rapid attach and CPR deate at the control unit.
Compact lightweight control unit is quiet, robust and powerful, with a reusable air lter.
Crisp, easy to read graphics for intuitive set up and
therapy control.
Integrated swing out hanging brackets for afxing
to most bed types.
Quilted therapy pad is vapor permeable, breathable, uid-proof, and minimizes friction
and shearing.
• Durable base tub is constructed from 100%
heavy weight 1680 denier nylon with a 1.5 oz. urethane coating and incorporates bed attachment loops for stability.
Supports up to 350 lbs. resident weight
• CPR deation
• Anti-kink, easy clean air supply hose set
Therapy Pads
Many healthcare facilities are facing the challenge of infection control. Joerns quilted therapy covers are treated with an antimicrobial to protect the therapy pad itself from the growth of mold, mildew and odor­causing bacteria.
Key features and benets:
Treated with a highly effective bacteriostat agent to inhibit the growth of bacterial and fungus.
Constructed from a very smooth nylon fabric with low friction and low shear properties to protect the resident’s skin from damaging friction/shearing forces.
Breathable, moisture vapor permeable fabric allows air to circulate beneath the pad and wicks away excess moisture. This keeps your resident dry and helps to prevent skin maceration.
Two-way stretch therapy pad is designed for optimal comfort, moisture vapor transfer, stain resistance and ease of laundering
Therapy Mattress System Features
Twenty individual therapy cells help to evenly
distribute the resident’s weight and maximize pressure relief. Modular cell design mattress for ease of cleaning, re-assembly and cost effective service.
Lower two-inch therapy support cell provides additional therapeutic support and remains inated
for up to 12 hours in the event of a power failure.
Eight-inch deep therapy cells are constructed of highly durable, polyurethane coated nylon to provide adequate support and prevent “bottoming out” for most residents within weight limit.
Ten-inch safety side bolsters provide comfort and support for residents during the rotation cycle.
6
6
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
Figure 1
Page 7
Joerns Support Surface
Grounding Instructions
Warning: Use a properly grounded, three-prong, 120V
AC outlet for this product. Failure to use a grounded
outlet could result in personal injury or damage to
equipment or house wiring, including risk of re. A qualied electrician should be contacted to correct the
wiring and ensure a properly grounded outlet.
Before installing this product, have the electrical system checked to make sure the electrical circuits and the electrical service are properly grounded.
Having a three-prong outlet does not necessarily mean it is grounded. Sometimes two-prong outlets are replaced with a three-prong type even though there is no ground wire.
There is always a chance of a loose connection or poor installation of a ground wire that causes the loss of proper ground at the outlet. Inadequate grounding at electrical outlets can occur even if there is a ground wire. Wires can become loose over time at the connection to the outlet.
Note: To install new wires on a circuit requires a
qualied electrician.
How to Determine if Your Outlet has the Proper Grounding
Most hardware stores sell circuit testers (Figure 2) that can be used to test an outlet for proper grounding. The tester plugs into an outlet and by observing the indicator lights you can determine if the outlet is properly grounded. For a higher level of assurance, an electrician should be requested to thoroughly test the electrical system with more reliable equipment.
If repair or replacement of the cord or plug is necessary, please contact Joerns Healthcare for assistance.
Setup
Warning: For important precautions, please see page two.
Caution: Do not place the control unit on the oor.
Position the power cord to keep personnel from tripping over it.
Note: The DermaFloat LR
system must be installed on bed frames that are equipped with side rails. Please raise side rails on the bed and lock them in position after the resident is on the mattress. Never leave resident unattended on mattress system with bed side rails in the down position.
When the product is not in use, properly store the power cord. Failure to do so could result in personal injury.
Remove the existing mattress from the bed.
• Place the DermaFloat LR mattress with the
hose connection at the foot end of the bed and the therapy cells facing up. Secure the straps on the mattress securely to the movable part of the bed frame.
If the therapy pad is not already on the mattress, place it on the mattress. Attach the
elastic straps to the mattress buckles around
each corner of the mattress. Attach the six (6)
additional straps to the movable part of the bed frame.
Hang the control unit on the foot of the bed facing away from the bed. Attach the hose
connector marked CPR to the control unit.
Plug in the control unit and the press the
Power key. The control unit will start and the Power light will illuminate.
• Allow up to a minute for full ination. Place
the resident on the mattress. Mattress can be
inated with resident on it, but will take longer,
depending on their weight/size. Note: Keep the control unit on while the resident is on the mattress.
Figure 2
7
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
7
Page 8
Joerns Support Surface DermaFloat LR Model
Position the resident’s head in the position that
the resident will be in for the largest portion of the
day. If the resident is lying at, please use three (3) ngers for the hand check. If the resident will
be sitting up for the majority of the day, please use
two (2) ngers.
• Perform a hand check by placing ngers locally
under the resident’s buttocks between two cushions. The resident should not bottom out. If
they do, increase the therapy control by one LED
(light), until they no longer bottom out.
• The safety side bolsters cannot be inated unless
the CPR plug is installed. The plug is located on the hose assembly between the mattress and the control unit. Caution: The safety side
bolsters should always be inated when resident is
unattended on the mattress and/or the mattress is in Turn mode.
Operation
Warning: For important precautions, please see page two.
Caution: The resident’s head should be positioned in the center of the top section of the mattress. When using the mattress system always ensure that the
resident is positioned properly within the connes of the bed. Do not let any extremities protrude over the
side or between the bed rails when the mattress is being used.
Resident Comfort Controls and Monitoring
Power
The Power button is used to turn the power on and off.
Autorm
Autorm mode provides maximum air ination
designed to assist both residents and caregivers during resident transfer and treatment. The unit will automatically return to the mode it was in prior to Autorm (either Turn or Static) in approximately 15 minutes.
Pause
Use this feature to immediately stop all rotation and hold the current mattress position. The mattress will hold this position until this feature is turned off by pressing the Pause button a second time.
Figure 3
Lockout
This unit is designed to lock out all of the adjustment controls after the resident has been positioned correctly. Ten minutes after the last button has been pushed, push the Power light begins to ash indicating Lockout is enabled. This feature is to prevent any unauthorized changes to the resident settings. To unlock and make adjustments to the settings, press both up and down comfort arrows at the same time. Lockout mode will return after ten minutes.
Note: The unit is designed to lock out all the adjustment controls after the resident has been positioned correctly. In approximately ten minutes after
the last button push the power on light begins to ash
indicating Lockout is enabled.
Comfort Adjust
The Comfort Adjust function is located across the
top of the control panel. The DermaFloat® LR can be
customized to meet individual resident needs within
a therapeutic window. Use the soft and rm keys to
simultaneously increase or decrease pressure in the entire mattress in Turn and Static modes. The current
setting is shown in the LED display in the center of the
control panel.
Turn Mode
The Turn mode and sets the desired rotation type. The resident can be rotated to either side individually or to both sides in a continuous cycle.
Turn Time
Use the Turn Time feature to set the time cycle for the selected Turn mode. The DermaFloat LR can be placed into a 10, 20, or 60 minute cycle based on what is determined to be the optimal therapy setting for the individual resident.
8
8
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
Page 9
Joerns Support Surface
Turn Angle
The Turn Angle feature allows for the desired turn angle to be selected. The mattress can make a half or full turn based on the optimal therapy for the resident.
Additional Features
Warning: A possible re hazard exists. This product is
suitable for use with oxygen administering equipment of the nasal, mask, or half bed-length, tent-type only. To prevent personal injury or equipment damage, ensure that the oxygen tent does not extend below the mattress.
CPR
The hose connection at the control unit is marked
CPR. Disconnect the hose from the control unit. Deation times will vary based on resident weight and prole. To resume therapy, reattach the hose to the
control unit.
Transport
To transport the resident in bed, turn the control unit
off. Unplug the power cord from the outlet. Do not
disconnect the hose connection at the control unit. The lower support cell will prevent the resident from bottoming out for up to 12 hours.
Power Failure
If power fails to the system an alarm will sound and the Power Fail LED will turn on until power is restored. When the power is restored, the control unit will return to the previous comfort adjust setting in Static mode. The Turn mode must be reselected manually.
Low Pressure Alarm
If the control unit senses a low pressure condition, such as a disconnected hose or therapy cell, then it will activate an audible alarm and will ash “L” and “P” on the display. Once the low pressure condition has been corrected, the alarm will cease and the control unit will return to normal operation.
Optional Accessories
• Additional therapy pads – available for purchase
Troubleshooting
Therapy Surface is Not Inflating
Ensure the hose connection from the therapy mattress system (mattress) to the control unit is securely connected.
• Ensure that the control unit is plugged into an AC
outlet.
Ensure that the power is not on Standby. If on Standby, press the Power button.
Ensure that all air cells are connected to the internal mattress manifold.
Unable to Change Therapy Mode or Adjust Comfort Control
Make sure the Lockout function is disabled. To disable, press the up and down Comfort Adjust arrows simultaneously.
Troubleshooting Modes of Operation
The DermaFloat® LR is designed to offer continuous
bilateral or unilateral rotation.
Continuous bilateral rotation (Turning the resident to both left and right)
1. Ensure the mattress is fully inated with the
resident held centrally on the support surface and correctly supported. If necessary, make adjustments using the comfort control buttons.
2. Check to make sure the control panel is not locked
out. If the Power LED is ashing, then press the Comfort Control arrows together to unlock the system.
3. Verify the system is in Static mode (Turn mode
None LED will be lit).
4. Press the Turn mode button until the Both LED is
lit.
5. Press the Turn Angle button until the desired target
angle LED is illuminated.
6. Select the Turn Cycle time by pressing the Turn Time button until the desired time period is
illuminated.
9
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
9
Page 10
Joerns Support Surface DermaFloat LR Model
Caution: Before commencing rotation, it is important to ensure that all lines/feeds to the resident are free
from obstruction and have sufcient length for the
angle of turn required.
Caution: Observe the resident through at least one full rotation cycle to ensure resident safety and tolerance of angles selected.
Continuous Unilateral Rotation (Turning the resident to only the left or only the right)
1. Ensure the mattress is fully inated with the
resident held centrally on the support surface and correctly supported. If necessary, make adjustments using the comfort control buttons.
2. Ensure that the control unit is not locked. If the
Power LED is ashing, then press the Comfort Control arrows together to unlock the system.
3. Verify the system is in Static mode (Turn mode
None LED will be lit).
4. Press the Turn mode button until either the Left or
Right LED is lit.
5. Press the Turn Angle button until the desired target
angle LED is illuminated.
6. Select the Turn Cycle time by pressing the Turn Time button until the desired time period is
illuminated.
Nursing Procedures
Recommended Linen:
Special linens are not necessary for the DermaFloat
LR. There is no need for a bottom sheet as the therapy pad should be covering the therapy cells at all times. The resident should never be lying directly on the
therapy cells. Upon the resident’s specic needs, the
following linens may be utilized:
• Draw or slide sheet to aid in positioning and to
further minimize friction and shearing
Incontinence barrier pad for residents incontinent
of urine and/or stool, and residents with heavily draining wounds
• Add top sheet, blanket and/or bedspread as
needed for resident comfort
Keep the amount of padding between the resident
and bed to a minimum for optimum performance
Changing the Therapy Pad
• Place the therapy pad over the therapy cells, tting
the corner of the cushions into the corner of the
therapy pad. (Similar to a tted sheet)
Zip the therapy pad along each side of the
mattress tub.
Resident Positioning and Comfort
General Repositioning
Residents should be turned and repositioned per individual turning schedule or per facility policy. It may be helpful to activate the Autorm mode to achieve
a rm surface for repositioning purposes. The unit
will automatically return to the mode it was in prior to Autorm in approximately 15 minutes or you can manually return to therapy mode once resident has been repositioned.
Unless counter indicated, it is desirable to keep the head of the bed in the low position to provide optimal pressure relief and minimize the risk of shearing injuries.
Elevating Resident into Sitting Position
The special properties of the DermaFloat® LR therapy
pad reduce the opportunity for shear and friction that
may occur when raising the head of other beds. As
with any surface, sliding can be expected, therefore residents should be repositioned after elevation. The
knee gatch or foot of the bed may be elevated rst, to
help prevent the resident from sliding when the head of the bed is elevated.
Incontinence
Moisture against the skin surface leads to maceration, or softening of the tissues. To prevent maceration, it is recommended that an incontinence barrier pad be used to absorb excess moisture.
In the event of incontinence or excess drainage on
the therapy pad, wipe off the excess uid from the bed
surface.
10
10
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
Page 11
Joerns Support Surface
Safety Information
Resident Migration
Specialty bed products are designed to reduce/ relieve pressure and the shearing/friction forces on the resident’s skin. The risk of gradual movement and/or sinking into hazardous positions of entrapment and/ or inadvertent bed exit may be increased due to the nature of these products.
Traction
With any traction or unstable fractures, maintain physician-directed angle of articulation and guard against risks of resident migration or inadvertent
deation of resident surface.
Skin Care
Monitor skin conditions regularly, particularly in areas where incontinence and drainage occur or collect, and consider adjunct or alternative therapies for high acuity residents. Early intervention may be essential to preventing serious skin breakdown.
Bed Height
To minimize the risks of falls or injury, the resident surface should always be in the lowest practical position when the resident is unattended. Make sure areas under and around the frame are clear of objects, persons and parts of body before adjusting height.
Control Unit
Wipe off dust. If necessary, clean the housing exterior with a disinfectant solution or a mild detergent and a damp cloth. Then wipe dry.
General Cleaning
If there is no visible soilage with possible body uids,
it is recommend that the mattress system be cleaned with a mild detergent and warm water. If disinfection is desired, a combination cleanser/disinfectant may be
used as explained in “Disinfecting” area.
Resident care equipment that does not come into
contact with mucous membranes or non-contact skin, requires low-level disinfection. Wiping surfaces with a properly prepared detergent or disinfectant carries out low level disinfecting.
Processing of dirty resident care equipment should
take place in a designated area away from clean or sterile supplies and food preparation areas.
• Detergent/disinfectants should not be mixed with
other germicides or detergents. Using the proper dilution insures the most effective killing power of the disinfectant.
Wash hands often and well, including after
removal of gloves.
Resident care equipment that is used in isolation
areas should be disinfected in accordance with all internal policies and procedures regarding such equipment.
Disinfecting
Cleaning
Warning: Unplug the control unit from its power source. Failure to do so could result in personal injury or equipment damage.
Warning: Do not expose the unit to excessive
moisture that would allow for liquid pooling. Personal injury or equipment damage could occur.
Caution: Do not use harsh cleansers/detergents, such
as scouring pads and heavy-duty grease removers, or solvents, such as acetone. Equipment damage could occur.
When there is visible soilage and between residents, it is recommend that the unit and mattress be disinfected
with a tuberculocidal disinfectant. Disinfectant should
be registered with the Environmental Protection
Agency (EPA).
Use rubber gloves and eye protection.
• Prepare detergent/disinfectant (registered by EPA
as hospital disinfectant) solution according to instructions on label for correct use-dilution.
• With support surface deated, thoroughly wipe down entire mattress, as air cells will lie at. Be
sure to reach all areas underneath and in-between
air cells. Allow to air dry.
If dust or other soiling has accumulated along
air hoses, remove using swabs moistened with
detergent/disinfectant as necessary. Allow all
components to air dry. Wrap mattress in plastic and return to storage area.
Thoroughly wipe down outside of control unit and allow to air dry. Cover with plastic and return to storage area
Remove gloves and dispose; wash hands.
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
11
11
Page 12
Joerns Support Surface DermaFloat LR Model
Therapy Pad
The therapy pad can be wiped down with a disinfectant solution or a mild detergent with a damp cloth. If heavily soiled, the therapy pad can be laundered in a washer and dryer with warm water (no more than 120º
Fahrenheit). A non-bleach detergent should be used
sparingly. Wipe dry or allow to air dry.
Steam Cleaning
Do not use any steam cleaning device on the unit.
Excessive moisture can damage mechanisms in this unit.
Filter Cleaning
Check the air lter on the rear of the unit regularly for
buildup of dust/dirt. If buildup is visible, turn off the control unit and disconnect the power cord from the
wall outlet. Remove the lter by grasping the lter
pulling outward. Replace with the second supplied
lter. Ensure the replaced lter covers the entire lter region.
Hand-wash the removed lter in warm soapy water and allow to air dry. When dry, store the lter in a safe place for the next lter maintenance.
Support Surface
Gently roll up the support surface, expelling any residual air, for temporary storage. The mattress should be wrapped in plastic and/or a clean bag for storage.
System Specifications
Standard Features
Weight
Control unit: ............................... 14 lbs (6 Kg)
Mattress: .................................... 22 lbs (11 Kg)
Maximum weight capacity*: ....... 350 lbs (159 Kg)
Dimensions
Control unit:
5.75" (15 cm) W x 10.5" (27 cm) H x 12" (30 cm) D
Mattress:
35" (89cm) W x 80" (203cm) L x 10" (25cm) D
Electrical Specifications
USA
120V AC, 60 Hz, 0.6A
Maintenance
Warning: Only facility-authorized personnel trained by Joerns Healthcare should perform preventative maintenance. Preventative maintenance performed by unauthorized personnel could result in personal injury or equipment damage.
Any maintenance done without Joerns’s authorization
will invalidate any warranties on this product.
Storage and Care
Note: Clean the DermaFloat LR as described in the
previous section prior to storage.
Control Unit
The power cord may be wrapped around the unit for convenience. Wrap the unit in a plastic bag for dust resistance, then store the unit in an area appropriate for an electronic medical device.
Environmental Conditions
Operating Conditions
Ambient Temperature: +10ºC to +40ºC Relative Humidity: 30% to 75% Non-Condensing
Storage and Shipping Conditions
Ambient Temperature: +10ºC to +40ºC Relative Humidity: 10% to 100%
Agency Approvals
• UL Classied Medical Equipment No. 60601.1
• UL 60601-1 Can/CSA C22.2
UL Classication refers to the power unit only, not the
complete mattress replacement system.
Call for Assistance
If you have any questions or require service on a Joerns product, please call Joerns Healthcare at
800.826.0270.
12
12
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
* Mattress weight capacity only; total weight must
not exceed bed frame manufacturers’ specied load capacity.
Page 13
Notes:
Joerns Support Surface
13
13
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003
Page 14
Joerns Healthcare Warranty Program
for all Joerns DermaFloat LR Models
Joerns Healthcare, warrants the DermaFloat
LR mattress to be sold free from defects in workmanship and materials, under normal and proper use, for a period of two (2) years on the mattress, and one (1) year on the cover and electromechanical mattress components (compressors, valves, printed circuit boards,
hoses, and couplers). Damages arising from
improper use will not be covered by this warranty.
Improper use is dened as, but not limited to those
caused by:
Burns
Use of improper chemical agents
Needle punctures, cuts, or abrasions
Excessive loads
Staining
Negligent or excessive usage
Improper maintenance, handling and/or
cleaning
Failure to use in the manner indicated in the DermaFloat LR user manual
Any modication, repair or alteration done to the DermaFloat LR that was not authorized in writing
by Joerns will void this warranty.
Parts
Joerns’s DermaFloat LR contains various parts that
wear from normal use. Joerns’ obligation under this warranty is limited to supplying replacement parts, servicing or replacing, at its option, any product which is found by Joerns to be defective. When requested by Joerns, parts must be returned for inspection at the customer’s expense. Credit will be issued only after inspection.
Service
Most service requests can be handled by the facility
Maintenance Department with assistance from the Joerns Product Service Department.
Most parts requested can be shipped next day air at the customer’s expense.
Should a technician be required, one will be provided by Joerns, at our discretion. Only the Joerns Product
Service Department can dispatch authorized
technicians.
Damage caused by use in unsuitable
environmental conditions, abuse or failure to maintain the product in accordance with user and service instructions is not covered.
This warranty is extended to the original purchaser of the equipment.
Post Acute, Acute, HomeCare
2430 Whitehall Park Dr. Ste 100 Charlotte, NC 28273 (P) 800.826.0270 (F) 800.457.8827
www.Joerns.com • www.RecoverCare.com email: info@joerns.com
VA/Government
19748 Dearborn Street Chatsworth, CA 91311 (P) 800.966.6662 (F) 800.232.9796
Manufactured by:
Joerns Healthcare, LLC
2100 Design Rd. Ste 100 Arlington, TX 76014
Canadian Ofce
1000 Clarke Rd. Ste 6 London, ON Canada N5V 3A9 (P) 866.546.1151 (F) 519.451.8662
United Kingdom and Other Countries
+44 (0)844 811 1156 +44 (0)844 811 1157
© 2015 Joerns Healthcare • 6110064 RevB • 15-2003