JENCO VisionPlus pH618N Operation Manual

pH618
Operation Manual
MODEL pH 618
Microcomputer Based
pH&Temperature
Pocket Meter
CONTENTS
INITIAL INSPECTION AND ASSEMBLY ................... 2
BEFORE YOUR FIRST USE ...................................... 2
A. Replace the Batteries ................................................... 2
B. Soak the Electrode........................................................ 2
C. Setup and Calibrate the Electrode and Meter.............. 2
VISIONPLUS pH 618 OVERVIEW.............................. 3
A. Meter Description.......................................................... 3
B. VisionPlus Graphical Display ...................................... 3
OPERATION MODES AND KEYPAD OPERATIONS 4
A. Operation Modes........................................................... 4
B. Keypad Operations....................................................... 4
USING VISIONPLUS pH 618...................................... 5
A. Power ON/OFF .............................................................. 5
B. Set Date and Time......................................................... 5
C. Select Buffer.................................................................. 6
D. Calibrate pH................................................................... 6
E. Measure ......................................................................... 7
F. Hold Data ....................................................................... 7
G. Save Data ...................................................................... 7
H. Recall Data .................................................................... 7
ONGOING MAINTENANCE........................................ 8
A. Replace Electrode......................................................... 8
B. Replace Calendar Battery ............................................ 8
ERROR DISPLAYS AND TROUBLESHOOTING ...... 9
SPECIFICATIONS .................................................... 10
WARRANTY ............................................................. 10
INITIAL INSPECTION AND ASSEMBLY
Carefully unpack the instrument and accessories. Inspect for damages made in shipment. If any damage is found, notify your Jenco representative immediately. All packing materials should be saved until satisfactory operation is confirmed.
BEFORE YOUR FIRST USE
A. Replace the Batteries
1. Remove the screws as shown in the right figure and gently take off the cover. A small screwdriver is provided to easily remove the screws.
2. Remove all of the old batteries and insert a new set of batteries ensuring the polarities are correct.
B. Soak the Electrode
1. Remove the electrode cap covering the VisionPlus pH meter.
2. Soak the electrode in a pH 4 solution for 10 minutes before first use or after storage.
C. Setup and Calibrate the Electrode and Meter
VisionPlus pH 618 must be setup and calibrated before your first use. Please follow the instructions detailed in section
USING VISIONPLUS pH 618.
VISIONPLUS pH 618 OVERVIEW
A. Meter Description
1. Battery cap
2. LCD screen
3. Keypad
4. Electrode collar
5. Electrode & ATC assembly (Electrode cap is not shown.)
B. VisionPlus Graphical Display
1. DATE (MM-DD-YY)
2. HOLD mode indicator
3. MEASURE mode indicator
4. CALIBRATION buffer(s)
5. TIME (HH:MM – 24H format)
6. pH reading
7. TEMPERATURE reading
8. NUMBER of calibration point
1
2
3
4
5
OPERATION MODES AND KEYPAD OPERATIONS
A. Operation Modes
VisionPlus pH meter has 6 operation modes:
1. Measure Mode. Measure Mode is used to make all pH and temperature measurements.
2. Calibration Mode. Calibration Mode is used to perform 1, 2 or 3 point calibration.
3. Hold Mode. Hold Mode is used to display held measured values for increased ease of use.
4. Recall Data Mode. Recall Data mode is used to display measured data which have been stored in memory.
5. Date and Time Set Mode. Date and Time Set Mode is used to change date and time.
6. Buffer Select Mode. Buffer Select Mode is used to
select the buffer set, which can either be
7.00(7.00/4.01/10.01) or 6.86(6.86/4.00/9.18).
B. Keypad Operations
Key
Operation Mode
Duration Function
All, except Hold
2 seconds or more
Turns meter on/off.
Measure 0 second
Holds current measurement reading. Press again to resume measuring.
Hold 0 second
Returns to Measure Mode.
HOLD
Date and Time Set
0 second
Move cursor to next set of data.
Measure 2 seconds
Enters Calibration Mode.
Measure
10 seconds or more
Enters Date and Time Set Mode.
Calibration 0 seconds
Leaves Calibration Mode.
CAL
Date and Time Set
2 seconds or more
Leaves Date and Time Set Mode.
Key
Operation Mode
Duration Function
Measure 2 seconds
Stores current pH reading, date and
time into memory. Date and Time Set
0 second
Increases current
number.
Recall Data
2 seconds or more
Leaves Recall Data
Mode.
STORE
Buffer Select
2 seconds or more
Leaves Buffer
Select Mode.
Measure 2 seconds
Enters Recall Data
Mode.
Measure
5 seconds or more
Enters Buffer
Select Mode.
Recall Data 0 second
Scrolls through
stored data. Date and Time Set
0 second
Decreases current
number.
SCROLL
Buffer Select
0 second
Selects new buffer
set.
USING VISIONPLUS pH 618
A. Power ON/OFF
Press “HOLD” key for 2 seconds or more. The unit will turn on and enter the “Measure Mode”. Repeat the process to turn off the unit. The unit will also automatically turn off after 10 minutes of no key activity.
B. Set Date and Time
1. Press “CAL” key for 10 seconds or more to enter “Date and Time Set Mode”.
2. Press “HOLD” key to select calendar or time parameters.
3. Press “STORE” and “SCROLL” keys to increase and decrease values respectively.
4. Press “CAL” key for 2 seconds to exit and return to “Measure Mode”.
C. Select Buffer
1. Press “SCROLL” key for 5 seconds or more to enter “Buffer Select Mode”.
2. Press “SCROLL” key again to select either 7.00 (7.00/4.01/10.01) or 6.86 (6.86/4.00/9.18) buffer sets.
3. Press “STORE” key for 2 seconds to exit and return to “Measure Mode”.
Note: There is no need to repeat this procedure every time unless one decides to change the buffer settings.
D. Calibrate pH
1. Press “CAL” key for 2 to 3 seconds to enter “Calibration” mode.
2. Clean and dip meter into either pH 7.00 or 6.86 buffer solution. A ‘WAIT” icon will flash.
3. VisionPlus pH meter will perform end point sensing to determine when the calibration reading is stable. Meter is calibrated when “WAIT” icon disappears, and the unit will automatically enter the next point calibration 5 seconds later. One point/first point calibration is now completed. Press “CAL” key to exit or continue to next step for two points/second point calibration.
4. Repeat Steps 2 and 3 for two points/second point calibration by dipping the meter in pH 4.01/4.00 or
10.01/9.18 buffers.
5. Continue to next step for three points/third point calibration or press “CAL” to exit two points calibration mode.
6. Repeat Steps 2 and 3 for three points/third point calibration.
7. When “WAIT” icon disappears, the unit will automatically return to Measure Mode.
Note: For accurate measurements, it is recommended that pH calibration is preformed once a week and after replacing the electrode.
E. Measure
Dip the meter into the test solution in the “Measure Mode”.
F. Hold Data
1. When the pH readings is stable, press “HOLD” key once to lock the reading.
2. Press “HOLD” key again to unlock reading and the unit will return to “Measure Mode”. The unit is now ready for another measurement.
Note: The unit cannot be power off in “Hold Mode”.
G. Save Data
Either in the “Hold Mode” or the “Measure Mode”, press “STORE” key for 2-3 seconds. The “DATA SAVE” and “M-XX” icons will appear indicating the reading has been saved and stored in memory location XX. The unit will automatically returns to “Measure Mode” and ready to perform other task.
Note: The Model pH 618 is equipped with a non-volatile memory that can store up to 50 different sets of readings (pH, temperature, Time and date). Non-volatile memory will be retained even if power is lost. When all 50 memory locations are used up, the next set of data saved will overwrite the data stored in the first location and so on.
H. Recall Data
1. In “Measure Mode”, press “SCROLL” key 2-3 seconds, the last set of saved data will appear.
2. By pressing the “SCROLL” key, previous saved data will appear.
3. Press “STORE” key for 2 seconds to exit and return to “Measure Mode”.
ONGOING MAINTENANCE
A. Replace Electrode
1. Unscrew the electrode collar to remove the electrode & ATC assembly as shown in the right figure.
2. Remove the old electrode from the electrode collar.
3. Insert a new electrode, make sure the electrode fit back into the meter correctly.
4. Screw back the electrode collar.
B. Replace Calendar Battery
1. Unscrew the electrode & ATC assembly as shown in the right figure.
2. Remove the two screws at the lower part of the unit before detaching the upper part of the unit from the case as shown in the right figure.
3. Remove the old battery and
insert a new battery ensuring that the polarity is correct.
Note: In normal usage, the battery should last for more than 5 years.
ERROR DISPLAYS AND TROUBLESHOOTING
pH LCD Display
ATC Display
DISPLAY Mode
Possible cause(s) [Action(s)]
Any Value
"OVER"
Measure
a. Temperature over 0 to
99.9°C range. [Bring solution to a lower/higher temperature.] [Replace temperature probe.] b. No temperature
sensor. [Use a temperature probe.]
"OVER" 0.0~ 99.9°C Measure
pH value over 0.00 to14.00pH range . [Over Range or Recalibrate.]
"OVER" 0.0 ~ 60.0°C
pH CAL
a.pH-
STAND
b.pH-
SLOPE
a. Offset @ 7.00pH:
mV>90mV
Offset@6.86pH:
mV>98.3mV b. New slope>ideal slope by 30% [Use a new buffer solution.] [Replace electrode.]
"OVER" 0.0 ~ 60.0°C
pH CAL
a.pH-
STAND
b.pH-
SLOPE
a. Offset @ 7.00pH:
mV<-90mV Offset @ 6.86pH: mV< -81.7 mV
b. New Slope<ideal
slope by 30% [Use a new buffer solution.] [Replace electrode.]
Any Value
"OVER"
pH CAL
a.pH-
STAND
b.pH-
SLOPE
Buffer temperature over 0 to 60 °C range. [Bring buffer temperature within 0 to 60°C range .]
10
SPECIFICATIONS
pH
Range Resolution Accuracy
0.00 to 14.00 pH 0.01 pH ±0.02 pH ± 1 LSD
Temperature
Range Resolution Accuracy
0.0 to 99.9 °C 0.1 °C ±0.3 °C
32.0 to 211.8 °F 0.2 °F ±0.6 °F
pH pH buffer recognition pH 7.00, 4.01, 10.01 or
pH 6.86, 4.00, 9.18
pH Temperature compensation
AUTO 0.0 to 99.9 °C
pH Buffer Temperature range
0.0 to 60.0°C
pH Electrode Offset recognition
±90 mV at pH 7.00
+98.3 mV / -81.7 mV at pH 6.86
pH Electrode Slope recognition
±30% at pH 4.00, 4.01, 9.18 and 10.01
Input impedance >10
12
ǂ
Temperature Temperature sensor Thermistor, 10 k at 25°C Temperature unit Default in factory
General Power: Meter : LR44 x 4
Timer : CR2032 Battery life Meter: 30~40 Hours Timer: 5 years
Ambient temperature range 0.0 to 50.0 °C Display: 98x64 graphic LCD Case IP67 water-tight case Weight 105 g
WARRANTY
Jenco Instruments, Ltd. warrants this product to be free from significant deviations in material and workmanship for a period of 2 years for the meter and 6 months for the electrode from date of purchase. If repair or adjustment is necessary and has
11
not been the result of abuse or misuse, within the warranty period, please return-freight-prepaid and the correction of the defect will be made free of charge. If you purchased the item from our Jenco distributors and it is under warranty, please contact them to notify us of the situation. Jenco Service Department alone will determine if the product problem is due to deviations or customer misuse.
Out-of-warranty products will be repaired on a charge basis.
RETURN OF ITEMS
Authorization must be obtained from one of our representatives before returning items for any reason. When applying for authorization, have the model and serial number handy, including data regarding the reason for return. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Jenco will not be responsible for damage resulting from careless or insufficient packing. A fee will be charged on all authorized returns.
NOTE: Jenco reserves the right to make improvements in design, construction and appearance of our products without notice.
(Version: 07302004 )
JENCO INSTRUMENTS, INC.
7968 ARJONS DR., SUITE C SAN DIEGO, CA 92126 TEL: (858)578-2828 FAX: (858)578-2886 E-Mail:jencoi@ix.netcom.com
JENCO ELECTRONICS, LTD.
P.O. BOX LINKOU-117 TAIPEI HSIEN, TAIWAN,R.O.C. TEL: (886-2)2601-6191 FAX: (886-2)2601-7206 E-Mail:jencoe@ms2.hinet.net
SHANGHAI JENCO ELECTRONICS, LTD.
18 WANG DONG ZHONG ROAD,SIJING, SONGJIANG,SHANGHAI, CHINA TEL: 86-21-5761-9599 FAX: 86-21-5761-9598 E-Mail:jenco@public.sta.net.cn Website:www.jenco.com.cn
Loading...