JDS Uniphase OLP-35, 230201, OLP-34, OLP-38 User Manual

COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS
SmartPocket™ Optical Power Meters
OLP-34/35/38
Applications
• Measure optical power levels and link insertion loss
• Enterprise LAN networks with 850/1300 nm multimode wavelengths capability
• Test access and metro (LAN/WAN) networks in multimode and single­mode
Key Features
The JDSU SmartPocket OLP-3x series next-generation small, rugged optical power meters are ideal for quickly, easily, and conveniently measuring optical power level and loss in fiber networks from the field. The instruments’ rugged ergonomic design and large, sharp display show relevant results and settings at the same time. Their intuitive one-button operation and automatic wavelength recognition is why technicians choose SmartPocket OLPs when they test fiber. SmartPocket instruments’ unprecedented data storage capacity for saving up to 100 results and optional micro USB port for downloading results to a PC fulfills all of your testing and reporting needs.
• Cost-effective, rugged high-performance solution
• 3-year recalibration period
• 1 nm incremental universal wavelength settings
• Universal optical interface supports all 2.5mm with an option to support 1.25mm connectors
• Auto λ detection and TwinTest mode
• Internal data storage and optional PC download capability
• Quick start operation eliminates warm-up
• Standard and high-power level tests for use in telecom, CATV, and military applications
WEBSITE: www.jdsu.com
2
SMARTPOCKET OLP-3x SERiES OPTiCAL POwER METERS
A small tool with big testing capabilities
Large, sharp screen displays:
• Battery level status
• Power mode (PERM, ECON)
• Measurement results in dBm/W/dB
• Wavelength setting
• Auto wavelength detection
• Tone detection
Rugged, shock resistant and splash proof design
Straightforward 4-button operation
3-way power
• AA dry
• AA rechargeable
• AC operation (optional)
Pocket-sized power meters for various applications
For multimode enterprise networks and high power applications, the SmartPocket series power meters offer a range of optical fiber testing and troubleshooting capabilities for the field.
• The OLP-34 is optimized for use in LAN/WAN access and enterprise networks and multimode or single-mode applications.
• The OLP-35 is optimized for use in standard telecommunication network applications.
• The OLP-38 is a higher power (+26 dBm) version for use in CATV (with analog RF transmission) or amplified DWDM systems applications.
Test and report field measurements
JDSU OFS-355 Reporting Software lets users quickly and efficiently download test results data from the power meter’s memory with just a few clicks. After it’s downloaded, the software reporting functions let users generate and customize professional certification reports.
SMARTPOCKET OLP3X SERIES OPTICAL POWER METERS
3
Specifications
Technical OLP-34 OLP-35 OLP-38
Detector type Germanium InGaAs Filtered InGaAs
Optical interface/connectors Universal 2.5 mm/1.25 mm
Wavelength range 780 to 1600 nm 780 to 1650 nm 780 to 1650 nm
Wavelength settings 780 to 1600 nm, in 1 nm steps 780 to 1650 nm, in 1 nm steps 780 to 1650 nm, in 1 nm steps
Programmable wavelengths 5 presets (customized) 5 presets (customized) 5 presets (customized)
Calibrated wavelengths 850, 980, 1300, 1310,
1490, 1550 nm
Power range –60 to +5 dBm –65 to +10 dBm –50 to +26 dBm
Display range –60 to +10 dBm –65 to +13 dBm –50 to +26 dBm
Max. input power +13 dBm +16 dBm +27 dBm
Measurement units dB/dBm/W dB/dBm/W dB/dBm/W
Absolute uncertainty
3
Linearity
2
± 0.2 dB (±5%) ± 0.2 dB (±5%) ± 0.2 dB (±5%)
± 0.06 dB (–50 to +5 dBm) ± 0.06 dB (–50 to +5 dBm) ± 0.06 dB (–32 to +20 dBm)
Tone detection 270 Hz, 1 kHz, 2 kHz 270 Hz, 1 kHz, 2 kHz 270 Hz, 1 kHz, 2 kHz
Auto λ mode
Multi λ mode
4
4
Yes Yes Yes
TwinTest/serial TripleTest TwinTest/serial TripleTest TwinTest/serial TripleTest
Notes:
1. Optional 1.25 mm UPP accessory adapter is available
2. Under these reference conditions: 20 dBm (CW), 1310 nm ±1 nm, 23°C ±3K, 5 to 75% relative humidity 9 to 50 μm test fiber with ceramic DIN connector
3. –5 to +45°C
4. With JDSU light sources
1
Universal 2.5 mm/1.25 mm
850, 980, 1300, 1310, 1490, 1550, 1625 nm
1
Universal 2.5 mm/1.25 mm
1
850, 980, 1300, 1310, 1490, 1550, 1625 nm
General
Storage and interface
Data storage 100 results
Data download capability micro USB interface for PC transfer (option)
Power supply
Dry batteries 2x Mignon (AA) 1.5 V
Rechargeable batteries 2x Mignon (AA) NiMH 1.2 V
AC operation (optional) via micro USB and SNT-505
Battery life ≥ 200 hr Automatic power o after 20 minutes
Environmental conditions
EMI/ESD CE compliant
Recommended calibration interval 3 years
Operating temperature –10 to +55°C (+14 to +131°F)
Storage temperature –40 to +70°C (–40 to +158°F)
Dimensions (H x W x D) 30 x 80 x 150 mm (1.2 x 3.1 x 5.9 in)
Weight 200 g (0.45 lb)
SMARTPOCKET OLP3X SERIES OPTICAL POWER METERS
Ordering Information
Power Meters
Product Number Description
2302/01 OLP-34 Optical Power Meter
Germanium, +5 dBm with data storage
2302/02 OLP-35 Optical Power Meter
InGaAs, +10 dBm with data storage
2302/03 OLP-38 Optical Power Meter
Filtered InGaAs, +26 dBm with data storage
2302/11 OLP-34 Optical Power Meter*
Germanium, +5 dBm with data storage and USB download capability
2302/12 OLP-35 Optical Power Meter*
InGaAs, +10 dBm with data storage and USB download capability
2302/13 OLP-38 Optical Power Meter*
Filtered InGaAs, +26 dBm with data storage and USB download capability
Each optical power meter includes a 2.5 mm UPP adapter, AA alkaline batteries, operating manual,
neck strap, and a belt bag.
* Includes downloading capability in addition to USB connection cable and a USB memory stick con-
taining reporting software and product documentation.
Optional Accessories
Product Number Description
FCR-CLN-01 Fiber Cleaning Consumables kit
2237/90.02 NiMH rechargeable batteries, Mignon AA, 1.2 V (2 batteries required)
2302/90.01 SNT-505; universal AC power adapter, micro USB connector
2256/90.03 UPP adapter for LC, MU (1.25 mm)
K 807 USB connection cable
OFS-355 Reporting Software (free download from jdsu.com)
Test & Measurement Regional Sales
no rth a mer ica
TOLL FREE: 1 855 ASK-JDSU 1 855 275-5378
Product specifications and descriptions in this document subject to change without notice. © 2013 JDS Uniphase Corporation 30162955.002.0213 OLP-34-35-38.DS.FOP.TM.AE February 2013
la tin a mer ica
: +1 954 688 56 60
te l fa x: + 1 95 4 34 5 4 668
as ia paci fi c
: +8 52 2 892 0990
te l fa x: + 852 2892 077 0
em ea
: +4 9 71 21 8 6 2 222
te l fa x: + 49 7 121 86 122 2
www.jdsu.com/test
Loading...