Janitza UMG 96 RM-M User guide

Power Analyser
Art. Nr. 33.03.162
Dok. Nr. 1.040.071.1.m
www.janitza.com
Janitza electronics GmbH Vor dem Polstück 1 D-35633 Lahnau Support tel. 0049 6441 9642-22 Fax 0049 6441 9642-30 E-mail: info@janitza.com Internet: http://www.janitza.com
UMG 96 RM-M
Operating instructions and technical data
Power Analyser
UMG 96RM-M
Contents
General 4 Incoming goods inspection 6
Scope of delivery of the UMG 96RM-M 7 Available accessories 7
Product description 8
Intended use 8 Characteristics of the UMG 96RM-M 9 Measuring method 10 Netzanalysesoftware GridVis 11 Connection options 11
Assembly 12 Installation 14
Supply voltage 14 Voltage metering 16 Current measurement 22 M-Bus interface 29 Digital outputs 32
Operation 34
Display mode 34 Programming mode 34 Parameters and measured values 36
Configuration 38
Applying the supply voltage 38 Current and voltage transformers 38 Programming current transformers 39 Programming voltage transformers 40 Programming parameters 41
2
Commissioning 54
Applying the supply voltage 54 Applying the measured voltage 54 Applying the measured current 54 Rotation field direction 55 Checking the phase assignment 55 Checking the power measurement 55 Checking the measurement 55 Checking the individual power ratings 55 Check the sum power ratings 56
M-Bus interface 57
Number of data points 57 Measurement signal level 58 Structure of the RSP_UD2 telegram 58 List of data points 59 Telegramm 61 M-Bus test 63 Analysis via M-Bus Scanners (Excerpt) 64 Work values within the software GridVis 65 Control of the values 65
Digital outputs 66
Pulse output 68 Comparator 74 Parameter list comparator and digital outputs 77
Service and maintenance 80
Device calibration 80 Calibration intervals 80
Error messages 82 Technical data 88
Parameters of functions 94 Table 1 - Parameter list 96 Dimensional drawings 102
Overview of measured value displays 104 Declaration of conformity 110 Anschlussbeispiel 111 Brief instructions 112
UMG 96RM-M
3
UMG 96RM-M

General

Copyright
This manual is subject to the laws of copyright protection and may not be mechanically or electronically photocopied, reprinted, reproduced or otherwise reproduced or published in part or as a whole, without the legally binding, written consent of
Janitza electronics GmbH, Vor dem Polstück 1, D 35633 Lahnau, Germany.
Trademarks
All trademarks and the rights resulting from them remain the property of the trademark holder of these rights.
Disclaimer
Janitza electronics GmbH assumes no responsibility for errors or omissions in this manual and assumes no obligation to keep the contents of this manual up to date.
4
Comments about the manual
Your comments are welcome. If anything in this manual is unclear, please let us know and send us an e-mail at: info@janitza.com
Meaning of the symbols
The following pictograms are used in this manual:
Dangerous voltage!
c
m
Risk of death or serious injury. Disconnect the power before working on the system and device.
Attention!
Please refer to the documentation. This symbol will warn you of possible dangers that could occur during assembly, commissioning and operation.
Note!
C
Application notes
UMG 96RM-M
Please read these operating instructions and all other publications that must be consulted in order to work with this product (particularly for installation, operation or maintenance).
Please observe all safety regulations and warnings. Non­compliance with the instructions can lead to personal injury and/or damage to the product.
Any unauthorised alteration or use of this device which exceeds the specified mechanical, electrical or other operational limits can cause personal injury and/or damage to the product.
Any such unauthorised alterations are grounds for "abuse" and/or "negligence" in terms of the product's guarantee and thus excludes the warranty for covering any possible resulting damages.
This device must only be operated and maintained by qualified personnel.
Qualified personnel are persons who, due to their respective training and experience, are able to recognise risks and avoid potential hazards that can be caused by operation or maintenance of the device.
When using the device, the legal and safety regulations required for the respective application must also be observed.
Safety is no longer guaranteed and the
c
m
m
device may be dangerous if the device is not operated according to the operating instructions.
Conductors consisting of single wires must be provided with ferrules.
Only screw terminals with the same number of poles and the same type may be plugged together.
5
UMG 96RM-M
About these operating instructions
These operating instructions are part of the product.
• Read the operating instructions prior to using the device.
• Keep the operating instructions at hand throughout the entire service life of the product and keep ready for referencing.
• Hand over the operating instructions to each subsequent owner or user of the product.
C
6
All supplied screw terminals are attached to the device.

Incoming goods inspection

The proper and safe operation of this device requires appropriate transport, proper storage, installation and assembly as well as careful operation and aintenance. When it is assumed that safe operation is no longer possible, the device must immediately be taken out of operation and secured against accidental start-up. Unpacking and packing must be carried out with the usual care, without the use of force and only with the use of suitable tools. The devices must be visually inspected for proper mechanical condition. It can be assumed that safe operation is no longer possible if the device, e.g.
• shows visible damage,
• does not work despite intact power supply,
• and was exposed to unfavourable conditions (e.g. storage outside of the permissible climatic limits without adaptation to the ambient climate, condensation, etc.) or transport stresses (e.g. falling from a great height even without exterior visible damage, etc.) for prolonged periods.
• Please check that the delivery is complete before you begin with installation of the device.

Scope of delivery of the UMG 96RM-M

Quantity Item no. Designation
1 52.22.039 UMG 96RM-M
2 29.01.036 Mounting brackets
1 33.03.162 Operating instructions
1 51.00.116 CD with the following contents
- GridVis programming software
- GridVis functional description
1 10.01.855 Screw terminal, pluggable, 2-pin (auxiliary energy)
1 10.01.849 Screw terminal, pluggable, 4-pin (voltage measurement)
1 10.01.871 Screw terminal, pluggable, 6-pin (current measurement)
1 10.01.857 Screw terminal, pluggable, 2-pin (M-Bus)
1 10.01.859 Screw terminal, pluggable, 3-pin (digital/pulse output)

Available accessories

Item no. Designation
29.01.907 Seal, 96 x 96
15.06.048 M-Bus signal converter PW60
UMG 96RM-M
7
UMG 96RM-M

Product description

Intended use

The UMG 96RM-M is provided for the measurement and calculation of electrical parameters such as voltage, current, power, energy, harmonics, etc. for building installations, to distributors, circuit breakers and busbar trunking systems. The UMG 96RM-M is suitable for installation in permanent, weatherproof switchboards. Conducting switchboards must be earthed. Measurement voltages and measurement currents must originate from the same grid. The measurement results can be displayed and can be read and processed over the M-Bus interface.
The voltage measurement inputs are designed for measuring in low voltage grids in which nominal voltages up to 300V phase can occur in countercurrent with ground and overvoltages of overvoltage category III. The UMG 96RM-M current measurement inputs are connected via external ../1A or ../5A current transformers.
Measurements in medium and high voltage systems generally use current and voltage transformers.
The UMG 96RM-M can be used in residential and industrial areas.
Device characteristics
• Installation depth: 45 mm
• Supply voltage: 20V - 250V (45..65Hz) or DC 20V - 300V
• Frequency range: 45-65 Hz
Device functions
• 3 voltage measurements, 300 V
• 3 current measurements (via current transformer)
• M-Bus interface
• 2 digital outputs
8

Characteristics of the UMG 96RM-M

UMG 96RM-M
• General
• Front panel-mounted with the dimensions 96x96 mm
• Connection via screw-type terminals
• LC display with backlighting
• Operation via 2 buttons
• 3 voltage measurements inputs (300V CATIII)
• 3 current measurement inputs for current transformer
• M-Bus interface
• 2 digital outputs
• Working temperature range -10°C .. +55°C
• Storage of minimum and maximum values (without time stamp)
• Measurement uncertainty
• Active energy, measuring uncertainty class
0.5 for ../5 A transformer
• Active energy, measuring uncertainty class 1 for ../1 A transformer.
• Reactive energy, class 2
• Measurement
• Measurement in IT, TN and TT networks
• Measurement in networks with nominal voltages up to L-L 480 V and L-N 277 V
• Current metering range 0 .. 5 Aeff
• True root mean square measurement (TRMS)
• Continuous scanning of voltage and current measurement inputs
• Frequency range of the mains frequency 45 Hz .. 65 Hz
• Measurement of harmonics 1 to 40 for ULN and I
• Uln, I, P (import/delivery), Q (ind./cap.)
• Fourier analyses 1 to 40. Harmonic for U and I
• 7 power meter for
Active energy (import) Active energy (export) Active energy (without a backstop) Reactive energy (ind.) Reactive energy (capacitive) Reactive energy (without a backstop) Apparent energy each for L1, L2, L3 and total
9
UMG 96RM-M

Measuring method

The UMG 96RM-M measures uninterrupted and calculates all root mean squares over a 10/12-period interval. The UMG 96RM-M measures the true root mean square (TRMS) of the voltages and currents applied to the measuring inputs.
Operating concept
The UMG 96RM-M can be programmed directly on the device via the 2 buttons. In addition, measurement val­ues can be called up via the M-Bus interface - e.g. with the GridVis read-out software.
The programming software of the GridVis has its own “online help”.
10
UMG 96RM-M

Netzanalysesoftware GridVis

The UMG 96RM-M can be programmed and read with the GridVis network analysis software which is part of the scope of delivery. For this a PC must be connected via a serial interface (RS232 / USB) for example via an M-Bus Master (level converter) to the M-Bus interface of the UMG 96RM-M.
The configuration of the UMG96RM-M is implemented exclusively via the two buttons on the device - the Grid­Vis software does not support this function!
It is not possible to read out M-Bus devices provided by other manufacturers using the GridVis software!
Characteristics of GridVis
• Reading of online measurement values
• Grafische Darstellung der Messwerte

Connection options

Connection of a UMG 96RM-M to a PC via a M-Bus signal converter (RS232):
PC
GridVis
RS232 M-Bus
Signal
converter
Item-No. 15.06.048
M-Bus
UMG 96RM-M
UMG 96RM-M
Connection of a UMG 96RM-M to a PC via a M-Bus signal converter (USB):
PC
GridVis
USB M-Bus
Signal
converter
M-Bus
UMG 96RM-M
UMG 96RM-M
11
UMG 96RM-M

Assembly

Installation location
The UMG 96RM-M is suitable for installation in perma­nent, weatherproof switchboards. Conducting switch­boards must be earthed.
Installation position
The UMG 96RM-M must be installed vertically in order to achieve sufficient ventilation. The clearance to the top and bottom must be at least 50 mm and 20 mm at the sides.
Front panel cutout
Cutout dimensions:
+0.8
x 92
+0.8
mm.
92
Fig. UMG 96RM-M installation location (rear view)
12
Mounting
The UMG 96RM-M is mounted on the switchboard by the side mounting brackets. These must be removed before using the device. Mounting is carried out by inserting and engaging the brackets.
Fig. UMG 96RM-M mounting bracket (side view)
Failure to comply with the minimum
m
spacing can destroy the UMG 96RM-M at high ambient temperatures!
UMG 96RM-M
13
UMG 96RM-M

Installation

Supply voltage

A supply voltage is required to operate the UMG 96RM-M.
The voltage supply is connected via plug-in terminals on the back of the device.
Before applying the supply voltage, ensure that the voltage and frequency correspond with the details on the nameplate!
The supply voltage must be connected via a UL/IEC approved fuse (6 A, type C).
L
N
Fuse
Separator
Fig. Connection example of the supply voltage to the UMG 96RM-M
14
m
UMG 96RM-M
• In building installations, the supply voltage must be provided with a disconnect switch or circuit breaker.
• The disconnect switch must be attached near the device and must be easily accessible by the user.
• The switch must be labelled as a separator for this device.
• Voltages that exceed the permissible voltage range can destroy the device.
15
UMG 96RM-M

Voltage metering

The UMG 96RM-M can be used for voltage measurement in TN, TT and IT systems. Voltage measurement in the UMG 96RM-M is designed for the 300 V overvoltage category CATIII (4 kV rated pulse voltage).
L1
L2
277V/480V 50/60Hz
L3
N
PE
V1 V3V2 VN
AC/DC
4M
4M
4M
4M
DC
Measuring voltage
UMG 96RM
Auxiliary energy
Fig. Principle circuit diagram - Measurement in three-phase 4-wire systems.
16
In systems without a neutral, measured values that require a neutral refer to a calculated neutral.
L1
L2
480V 50/60Hz
L3
Impedanz
V3V2
VN
AC/DC
4M
4M
DC
System earthing
V1
4M
4M
Measuring voltage
UMG 96RM
Fig. Principle circuit diagram - Measurement in three-phase
Auxiliary energy
3-wire systems.
Rated mains voltage
Lists of the networks and their rated mains voltage in which the UMG 96RM-M can be used.
UMG 96RM-M
Three-phase 4-wire systems with earthed neutral conductor.
U
/ U
L-N
L-L
66 V/115 V 120 V/208 V 127 V/220 V 220 V/380 V 230 V/400 V 240 V/415 V 260 V/440 V 277 V/480 V
Fig. Table of the rated mains voltages suitable for the voltage measuring inputs according to EN60664-1:2003.
Maximum rated voltage of the network
Unearthed three-phase, 3-wire systems.
U
L-L
66 V 120 V 127 V 220 V 230 V 240 V 260 V 277 V 347 V 380 V 400 V 415 V 440 V 480 V
Fig. Table of the rated mains voltages suitable for the voltage measuring inputs according to EN60664-1:2003.
Maximum rated voltage of the network
17
UMG 96RM-M
Voltage measurement inputs
The UMG 96RM-M has three voltage measurement inputs (V1, V2, V3).
Overvoltage The voltage measurement inputs are suitable for measurement in networks in which overvoltages of overvoltage category 300V CATIII (4 kV rated pulse voltage) can occur.
Frequency The UMG 96RM-M requires the mains frequency for the measurement and calculation of measured values. The UMG 96RM-M is suitable for measurements in the frequency range of 45 to 65 Hz.
18
L1
L2
L3
N
Fuse
Separator
Fig. Connection example for the voltage measurement
UMG 96RM-M
When connecting the voltage measurement, the following must be observed:
• A suitable separator must be provided in order to switch off the power to the UMG 96RM-M.
• The separator must be placed near the UMG 96RM­M, marked for the user and easily accessible.
• Use a fuse protected, UL/IEC approved 10A circuit breaker (type C) as an overcurrent protection device and separator.
• The overcurrent protection device must have a nominal value that is designed for the short circuit current on the connection point.
• Measurement voltages and measurement currents must originate from the same grid
c
c
c
Attention!
Voltages that exceed the permitted ratedmains voltages must be connected via voltage transformers.
Attention!
The UMG 96RM-M is not suitable for the measurement of DC voltages.
Attention!
The voltage measurement inputs on the UMG 96RM-M are dangerous to touch!
19
UMG 96RM-M
Connection diagram, voltage measurement
• 3p 4w (addr. 509= 0), factory setting
L1
L2
L3
N
V1 V2 V3 VN
Fig. System with three-phase conductors and a neutral conductor.
• 3p 4u (addr. 509 = 2)
L1
L2
L3
V1 V2 V3 V N
Fig. System with three-phase conductors and no neutral conductor. Measured values that re­quire a neutral refer to a calculated neutral.
20
• 3p 4wu (addr. 509 = 1)
L1
L2
L3
N
V1 V2 V3 V N
Fig. System with three-phase conductors and a neutral conductor. Measurement via voltage transformer.
• 3p 2u (addr. 509 = 5)
L1
L2
L3
V1 V2 V3 V N
Fig. System with three-phase conductors and no neutral conductor. Measurement via voltage transformer. Measured values that require a neutral refer to a calculated neutral.
UMG 96RM-M
• 1p 2w1 (addr. 509 = 4)
L1
N
V1 V2 V3 V N
Fig. Measured values derived from the V2 and V3 voltage measurement inputs are assumed to be zero and not calculated.
• 1p 2w (addr. 509 = 6)
L1
L2
V1 V2 V3 V N
Fig. TN-C system with single-phase, three-wire connection. Measured values derived from the V3 voltage measurement input Zero are assu­med to be zero and not calculated.
• 2p 4w (addr. 509 = 3)
L1
L2
L3
N
V1 V2 V3 V N
Fig. System with uniform phase loading. The measured values for the V2 voltage measure­ment input are calculated.
• 3p 1w (addr. 509 = 7)
L1
L2 L3
L1
L2 L3
L1
L2 L3 N
V1 V2 V3 V N
Fig. Three systems with uniform phase loading. The measurement values L2/L3 resp. L1/L3 resp. L1/L2 of the respective system are calculated.
21
UMG 96RM-M

Current measurement

The UMG 96RM-M is designed for connecting cur­rent transformers with secondary currents of ../1A and ../5A. The factory set current transformer ratio is 5/5 A and may need to be adapted to the current transformers. It is not possible to perform a direct measurement with­out a current transformer with the UMG 96RM-M. Only AC currents (and not DC currents) can be meas­ured.
c m
c
22
Attention!
The current measurement inputs are dangerous to touch.
Attention!
The UMG 96RM-M is not suitable for the measurement of DC voltages.
Earthing current transformers!
If a connection is provided for earthing the secondary winding, it must be connected to the earth.
Load
Fig. Current measurement via current transformer (connection example)
The attached screw terminal has to be
m
fixed sufficiently with two screws on the device!
L1
L2
L3
N
Direction of the current
If incorrectly connected, a subsequent re-connection of the current transformer is required.
UMG 96RM-M
c
Current transformer terminals!
The secondary terminals of the current transformer must be short-circuited to this before the power supply lines to the UMG 96RM-M are disconnected!
If a test switch which automatically short­circuits the current transformer secondary leads is available, it is sufficient to put this into the “test” position provided the short­circuiters have been checked beforehand.
c
Open current transformer!
High voltage peaks that are dangerous to touch can occur on current transformers that are operated in an open state at the secondary terminals. In “open-safe current transformers”, the winding insulation is measured so that the current transformers can operate in an open state. However, these current transformers are also dangerous to touch if they are operated in an open state.
23
UMG 96RM-M
Connection diagram, current measurement
• 3p 4w (addr. 510= 0), factory setting • 3p 2i (addr. 510 = 1)
L1
L2
L3
N
L1
L2
L3
N
I1 I2 I3
Fig. Measurement in a three-phase net-work with an unbalanced load.
• 3p 2i0 (addr. 510 = 2)
L1
L2
L3
I1 I2 I3
Fig. The measured values for the I2 current measurementinput are calculated.
24
I1 I2 I3
Fig. System with uniform phase loading. The measured values for the I2 current measurement input are measured.
• 3p 3w3 (addr. 510 = 3)
L1
L2
L3
I1 I2 I3
Fig. Measurement in a three-phase net-work with an unbalanced load.
UMG 96RM-M
• 3p 3w (addr. 510 = 4)
L1
L2
L3
N
I1 I2 I3
Fig. System with uniform phase loading. The measured values for the I2 and I3 current measurement inputs are calculated.
• 1p 2i (addr. 510 = 6)
L1
L2
I1 I2 I3
Fig. Measured values derived from the I3 current measurement input are assumed to be zero and not calculated.
• 2p 4w (addr. 510 = 5)
L1
L2
L3
N
I1 I2 I3
Fig. System with uniform phase loading. The measured values for the I2 current measurement input are calculated.
• 1p 2w (addr. 510 = 7)
L1
N
I1 I2 I3
Fig. Measured values derived from the I2 and I3 current measurement inputs are assumed to be zero and not calculated.
25
UMG 96RM-M
Connection diagram, current measurement
• 3p 1w (addr. 510 = 8)
L1
L2 L3
L1
L2 L3
L1
L2 L3
I1 I2 I3
Fig. Three systems with uniform phase load­ing. The current measurement values of the phases of the respective system where are no CTs connected are calculated (I2/I3 resp. I1/I3 resp. I1/I2).
26
c
Caution!
The UMG96RM-M is only approved for a current measurement using the current transformer.
Total current measurement
UMG 96RM-M
If the current measurement takes place via two current transformers, the total transformer ratio of the current transformer must be programmed in the UMG 96RM-M.
UMG
I
S
S2
1
Einspeisung 1 Supply 1
1P1 (K)
(L) 1P2
Verbraucher A Consumer A
1S1
1S
P1
1S1 1S2 2S1 2S2
(k) (l)
2
P2
Einspeisung 2
Supply 2
2S1 (k)
(l) 2S2
Verbraucher B
Consumer B
2P
(K) (L)
2P2
1
Fig. Current measurement via a total current transformer (example).
Example: The current measurement takes place via two current transformers. Both current transformers have a transformer ratio of 1000/5 A. The total measurement is performed with a 5+5/5 A total current transformer.
The UMG 96RM must then be set as follows: Primary current: 1000 A + 1000 A = 2000 A Secondary current: 5 A
27
UMG 96RM-M
Ammeter
If you want to measure the current not only with the UMG 96RM-M but also with the ammeter, the ammeter must be connected in series with the UMG 96RM-M.
UMG
I
S2
1
S
A
Einspeisung Supply
(k)S
1 S2(l)
2(L)(K)P1
P
Verbraucher
Fig. Current measurement with an additional ammeter (example).
28
Consumer
UMG 96RM-M

M-Bus interface

The M-Bus interface is designed with the UMG 96RM-M as a 2-pole plug contact and communicates via the M­Bus protocol.
The UMG 96RM-M loads the M-Bus with an M-Bus de­vice load of 1.5 mA.
M+ M-
M-Bus interface, 2-pole plug contact
2-pin connector with cable connection ( cable type: 2 x
2
0.75 mm
) via twin ferrules
Cable connections
Twisted screened cable should be used for connections via the M-Bus interface.
• Cable paths should be designed to be as short as possible.
• Maintain as much distance as possible to power ca­bles and to consumers (e.g. electrical motors, neon tubes, transformers).
• In order to prevent cross currents in the bus, there should be no ground coupling, or a maximum of one instance of ground coupling.
• Gather the cables mechanically above the earthing clamp in order to avoid damage due to cable move­ments.
• Use suitable cable glands to feed the cables into the cabinet - for example armoured conduit couplings.
29
UMG 96RM-M
Cable type
The cable used must be suitable for an ambient temperature of at least 80 °C.
Use 2-core, twisted, screened cable wherever possible for optimum data transmission.
Recommended cable types: Unitronic LIYCY 4x0.75
30
Bus structure
• All devices are connected in a star, line or tree struc­ture, whereby each device has its own address within the bus (see also Parameter programming).
• A subdivision of the network structure into individual segments is implemented via repeaters (line amplifi­ers).
• Up to 250 subscribers can be connected together in a single segment. However, the characteristics of the Master device are the defining factors here.
• If the master is replaced, the bus is out of service.
• Devices can be replaced without the bus being un­stable.
Star structure
• Each measurement device is linked directly to the M-Bus Master. Faults in the bus system are localised faster by switching the individual devices on and off.
Line structure
• The connection of the measurement devices is se­quential, in a line. With this possible faults in the bus system may arise due to the voltage drop. Faults within the system are harder to localise in this cheaper structure.
Tree structure
• This topology combines the star and line structures. Repeaters generally divide the branches into individu­al segments. Thus in the event of a fault only a specific branch is affected and so a fault in the bus system can be quickly localised.
Slave
(device)
Slave
(device)
Master
(control)
Slave
(device)
Slave
(device)
UMG 96RM-M
Slave
(device)
Slave
(device)
Slave
(device)
Master
(Zentrale)
Illustration of bus type: Tree structure
Master
(control)
Slave
(device)
Slave
(device)
Slave
(device)
Illustration of bus type: Star structure
Slave
(device)
Slave
(device)
Slave
(device)
Slave
(device)
Slave
(device)
Slave
(device)
Illustration of bus type: Line structure
Repeater
31
UMG 96RM-M

Digital outputs

The UMG 96RM-M has 2 digital outputs. These outputs are electrically isolated from the evaluation electronics by optocouplers. The digital outputs have a common reference.
• The digital outputs can switch DC and AC loads.
• The digital outputs are not short circuit protected.
• Connected cables longer than 30 m must be shielded.
• An external auxiliary voltage is required.
• The digital outputs can be used as pulse outputs.
• The digital outputs can output results from comparators.
Fig. Connection of digital/pulse outputs
~
32
External
auxiliary voltage
UMG 96RM-M
Digital outputs 1-2 Digital outputs 1-2
13
Digital output 1
Digital output 2
Abb. Anschluss von zwei Relais an die digitalen Aus­gänge 14 und 15.
C
When using the digital outputs as a pulse output, the auxiliary voltage (DC) must only have a maximum residual ripple of 5%.
14
15
AC/DC
AC/DC
24 V
AC/DC
~~
K1 K2
UMG 96RM-M
33
UMG 96RM-M

Operation

The UMG 96RM-M is operated using buttons 1 and
2. Measured values and programming data appears
on a liquid crystal display.
A distinction is made between display mode and pro- gramming mode. The accidental changing of program­ming data is prevented by the entry of a password.

Display mode

In the display mode, you can scroll between the programmed measured value displays using buttons 1 and 2. All factory-set measured value displays listed in section 1 can be called up. Up to three measured values are displayed per measured value display. The measured value relaying allows select measured value displays to be shown alternately after a settable changeover time.

Programming mode

In the programming mode, the settings required for operating the UMG 96RM-M can be displayed and changed. Pressing buttons 1 and 2 simultaneously for about one second calls up the programming mode after the password prompt. If no user password was
34
programmed, the user arrives directly in the first programming menu. Programming mode is indicated by the text “PRG” on the display. Button 2 can now be used to switch between the following programming menus:
- current transformer,
- voltage transformer,
- parameter list.
If the device is in programming mode and no button has been pressed for approximately 60 seconds or if buttons 1 and 2 are pressed simultaneously for approx. one sec­ond, the UMG 96RM-M returns to display mode.
Export
Mean value
Programming mode
Sum measurement
Phase conductor­Phase conductor
Password
CT: Current transformer VT: Voltage transformer
K1: Output 1 K2: Output 2
Button 2
Button 1
UMG 96RM-M
35
UMG 96RM-M

Parameters and measured values

All parameters necessary for operating the UMG 96RM­M, e.g. the current transformer data, and a selection of frequently required measured values are stored in the table. The contents of most addresses can be accessed via the serial interface and the buttons on the UMG 96RM-M.
Only the first 3 significant digits of a value can be entered on the device. The device always only displays the first 3 significant digits of a value.
Selected measured values are summarised in measured value display profiles and can be shown in display mode using buttons 1 and 2.
36
Example of the parameter display
On the UMG 96RM-M display the value “001” is shown as the content of address “000”. This parameter reflects the device address (here “001”) of the UMG 96RM on a bus in list form.
Example of the measured value display
In this example, the UMG 96RM-M display shows the voltages L to N with 230 V each. The K1 and K2 transistor out­puts are conductive and cur­rent can flow.
Button functions
UMG 96RM-M
Display mode
Change mode
simultaneous
Browse
short
long
Measured
values 1a
Measured
values 2a
long short
Measured
values 2b
Password
Programming
menu 1
(flashes)
Programming mode
short
long
(flashes)
Change mode
simultaneous
Browse
Programming
menu 1
Programming
menu 2
Programming
menu 3
Programming
Confirm selection
Short: digit +1
Long: digit -1
Short: value x 10
(decimal to the right)
Long: Value /10
(decimal to the left)
37
UMG 96RM-M
Configuration

Applying the supply voltage

To configure the UMG 96RM-M, the supply voltage must be connected.
The level of supply voltage for the UMG 96RM-M can be found on the nameplate.
If no display appears, check the operating voltage to determine whether it is within the rated voltage range.

Current and voltage transformers

A current transformer is set to 5/5 A in the factory. The pre-programmed voltage transformer ratio only needs to be changed if voltage transformers are connected.
When connecting voltage transformers, the measure­ment voltage on the UMG 96RM-M nameplate must be observed!
38
c
C
m
Attention!
Supply voltages that do not correspond to the nameplate information can lead to device malfunction or destruction.
The adjustable value 0 for the primary current transformer does not produce any useful energy values and must not be used.
Devices, which are programmed to au­tomatic frequency detection, need ap­proximately 20 seconds to detect grid frequency. During this period, the mea­sured values do not keep the confirmed measuring accuracy.
UMG 96RM-M

Programming current transformers

Switching to programming mode
• Simultaneously press buttons 1 and 2 in order to switch to programming mode. If a user password was programmed, the password request will appear with "000". The first digit of the user password flashes and can be changed with button 2. The next digit is selected by pressing button 2 and will begin flashing. If the correct combination was entered or if no user password was programmed, the device will enter pro­gramming mode.
• The symbols for the programming mode (PRG) and for the current transformer (CT) appear.
• Confirm the selection with button 1.
• The first digit of the input area for the primary current starts flashing.
Current transformer primary current input
• Change the flashing digit with button 2.
• Select the next digit to be changed with button 1. The selected digit to be changed starts flashing. If the entire number is flashing, the decimal point can be moved with button 2.
Current transformer secondary current input
• Only 1 A or 5 A can be set as the secondary current.
• Select the secondary current with button 1.
• Change the flashing digit with button 2.
Leaving programming mode
• Simultaneously press buttons 1 and 2 to exit the pro­gramming mode.
39
UMG 96RM-M

Programming voltage transformers

• Switch to the programming mode as described. The symbols for the programming mode (PRG) and for the current transformer (CT) appear.
• Use button 2 to switch to the voltage transformer setting.
• Confirm the selection with button 1.
• The first digit of the input area for the primary current starts flashing. The ratio of primary to secondary voltage of the voltage transformer can be set in the same way as the assignment of the current transformer ratio of primary to secondary current.
40
Current transformer, primary
Programming mode
Units display
Current transformer, secondary
Current transformer symbol
Voltage transformer, primary
Programming mode
Units display
Voltage transformer, secondary
Voltage transformer, symbol
UMG 96RM-M

Programming parameters

Switching to programming mode
• Switch to the programming mode as described. The symbols for the programming mode (PRG) and for the current transformer (CT) appear.
• Use button 2 to switch to the voltage transformer setting. The first parameter of the parameter list is shown by repeatedly pressing button 2.
Changing parameters
• Confirm the selection with button 1.
• The most recently selected address is displayed with the associated value.
• The first digit of the address flashes and can be changed using button 2. Button 1 provides a selection of digits that, in turn, can be changed with button 2.
Changing the value
• Once the desired address is set, a digit of the value is selected with button 1 and changed with button 2.
Leaving programming mode
• Simultaneously press buttons 1 and 2 to exit the programming mode.
Fig. Password request
If a password was set, it can be entered using buttons 1 and 2.
Fig. Current transformer programming mode
The primary and secondary currents can be changed using buttons 1 and 2 (cf. page 39).
Fig. Programming mode Voltage transformer
The primary and secondary currents can be changed using buttons 1 and 2 (cf. page 40).
Fig. Programming mode Parameter display
The individual parameters can be changed using buttons 1 and 2 (cf. page 36).
41
UMG 96RM-M
Device address (addr. 000)
If several devices are connected to one another via the M-Bus interface, a master device can only differentiate between these devices by means of their device addresses. Therefore, each device in a network must have a different device address. Addresses can be set in the range from 1 to 250.
The adjustable range of the device
C
Secondary device address (addr. 081-084)
The secondary address provides - in addition to the pri­mary address - a further opportunity to speak directly to the device within the bus system.
The composition of the secondary address is broken down into a device-specific section and an extended section:
• The secondary address is comprised of 8 Bytes and is coded as BCD.
• The extended section of the secondary address is
42
address is between 0 and 255. The values 0 and 251 to 255 are reserved and may not be used.
pre-assigned with the device serial number. This sec­tion can be changed by the customer (addr. 081- 084).
• The device-specific section of the secondary address cannot be changed.
Extended section, 8 digits
Device-specific section,
XX XX XX XX 2E 28 09 02
8 digits
Version, 2 digits Internal release, 2 digits Manufacturer ID, 4 digits Extended section, 4th part Extended section, 3rd part Extended section, 2nd part Extended section, 1st part
UMG 96RM-M
Baud rate (addr. 001)
A common baud rate is adjustable for the M-Bus inter­faces. The baud rate must be chosen to be a uniform value in the network. The parameter data bits (8), parity (even) and stop bits (1) are permanently set.
Setting Baud rate
0 300 Baud 1 600 Baud 2 1200 Baud 3 2400 Baud 4 4800 Baud 5 9600 Baud 6 19200 Baud 7 38400 Baud
Mean value
Mean values are formed over an adjustable period for the current, voltage and power measured values. The mean values are identified with a bar above the measured value. The averaging time can be selected from a list of nine fixed averaging times.
Current averaging time (addr. 040) Power averaging time (addr. 041) Voltage averaging time (addr. 042)
Setting Averaging time/sec.
0 5 1 10 2 15 3 30 4 60 5 300 6 480 (factory setting) 7 600 8 900
43
UMG 96RM-M
Averaging method
After the set averaging time, the exponential averaging method used achieves at least 95% of the measured value.
44
Minimum and maximum values
All measured values are measured and calculated every 10/12 periods. Minimum and maximum values are deter­mined for most of the measured values. The minimum value is the smallest measured value that has been determined since the last reset. The maxi­mum value is the largest measured value that has been determined since the last clearance. All minimum and maximum values are compared with the corresponding measured values and are overwritten if they are undercut or exceeded. The minimum and maximum values are stored in an EE­PROM every 5 minutes, without the date and time. This means that if the operating voltage fails, only the mini­mum and maximum values of the last 5 minutes are lost.
Clearing minimum and maximum values (addr. 506)
If "001" is written to the address 506, all minimum and maximum values are simultaneously cleared. The maximum value of the current mean value is an exception. The maximum value of the current mean value can also be cleared directly in the display menu by pressing and holding button 2.
Mains frequency (addr. 034)
UMG 96RM-M
For automatic ascertainment of the mains frequency, an L1-N voltage larger than 10Veff must be applied to the voltage measurement input V1.
The mains frequency is then used to calculate the sampling rate for the current and voltage inputs.
If there is no measurement voltage, the mains frequency cannot be determined and thus no sampling rate can be calculated. The acknowledgeable error message “500” appears. The voltage, current and all other resulting values are calculated based on the previous frequency measurement and possible cable-connecting sockets and continue to be displayed. However, these derived measured values are no longer subject to the specified accuracy.
If it is possible to re-measure the frequency, then the error message will disappear automatically after a period of approx. 5 seconds once the voltage has been restored.
The error is not displayed if a fixed frequency has been configured.
Adjustment range: 0, 45 .. 65
0 = automatic frequency determination. The mains frequency is determined from the measurement voltage.
45..65 = fixed frequency The mains frequency is preselected.
45
UMG 96RM-M
Energy meter
The UMG 96RM-M has energy meters for active energy, reactive energy and apparent energy.
46
Reading the active energy
Total active energy
The active energy in this example is: 12 345 678 kWh
The active energy in this example is: 134 178 kWh
UMG 96RM-M
TH
fund
TH
fund
Harmonics
Harmonics are the integer multiple of a mains frequency. The voltage mains frequency for the UMG 96RM-M must be in the range between 45 and 65 Hz. The calculated voltage and current harmonics refer to this mains frequency. Harmonics up to 40x the mains frequency are recorded.
The harmonics for currents are given in amperes and the harmonics for voltages are given in volts.
Number of the harmonic
Phase L3
Current harmonic
Value
Fig. Display of the 15th harmonic of the current in the L3 phase (example).
Harmonics are not displayed in the factory default setting.
C
Total Harmonic Distortion (THD)
THD is the ratio of the root mean square value of harmonics to the root mean square value of the mains frequency.
Total Harmonic Distortion of the current (THDI):
M
1
D
=
I
I
n
2
I
.
nHarm
2
=
Total Harmonic Distortion of the voltage (THDU):
M
1
D
=
U
U
n
=
2
U
.
nHarm
2
Phase L3
Voltage
Value
Fig. Display of the total harmonic distortion of the voltage from the L3 phase (example).
47
UMG 96RM-M
Measured value relay
All measured values are calculated every 10/12 periods and can be recalled once per second on the measured value displays. Two methods are available for retrieving the measured value displays:
• The automatically changing display of selected measured values, referred to here as measured value relaying.
• Selection of a measured value display using buttons 1 and 2 from a preselected display profile.
Both methods are simultaneously available. Measured value relaying is active if at least one measured value display is programmed with a changeover time greater than 0 seconds. If a button is pressed, the measured value displays of the selected display profile can be browsed. If no button is pressed for about 60 seconds, the device switches to the measured value relay and the measured values from the selected display change profile of the programmed measured value displays are shown one after the other.
48
Changeover time (addr. 039)
Adjustment range: 0 .. 60 seconds If 0 seconds are set, no changeover takes place between the measured value displays selected for the measured value relay. The changeover time applies for all display change profiles.
Display change profile (addr. 038)
Adjustment range: 0 .. 3 0 - Display changeover profile 1, by default. 1 - Display changeover profile 2, by default. 2 - Display changeover profile 3, by default.
UMG 96RM-M
Measured value displays
After return of the power supply, the UMG 96RM-M shows the first measured value panel from the current display profile. In order to keep the selection of meas­ured values to be displayed arranged in a clear manner, only one part of the available measured values is pre­programmed for recall in the measured value display by default. A different display profile can be selected if other measured values are required to be shown on the UMG 96RM-M display.
Display profile (addr. 037)
Adjustment range: 0 .. 3 0 - Display profile 1, default setting. 1 - Display profile 2, default setting. 2 - Display profile 3, default setting.
User password (addr. 050)
A user password can be programmed in order to impede any accidental change to programming data. A switch to the next programming menu can only be made after entering the correct user password. No user password is specified in the factory. In this case, the password menu is skipped and the current transformer menu is reached directly.
If a user password was programmed, the password menu will appear with the display “000”. The first digit of the user password flashes and can be changed with button 2. The next digit is selected by pressing button 1 and will begin flashing. The programming menu for the current transformer can only be accessed after entering the correct number combination.
49
UMG 96RM-M
Clear energy meter (addr. 507)
The active, apparent and reactive energy meters can only be cleared together.
Address 507 must be written with "001" in order to clear the contents of the energy meters.
Clearing the energy meters means this
C
50
data in the device is gone. In order to avoid possible data loss, read and save the measured values with the GridVis software before clearing.
Rotation field direction
The rotation field direction of the voltages and the frequency of phase L1 are shown on the display. The rotation field direction indicates the phase sequence in three-phase systems. Usually there is a "clockwise spinning rotation field". The phase sequence at the voltage measurement inputs is checked and displayed in the UMG 96RM-M. A movement of the character string in the clockwise direction means a "right rotation" and a counter­clockwise movement indicates a "left rotation". The rotation field direction is determined only if the measurement and operating voltage inputs are fully connected. If one phase is missing or two of the same phases are connected, the rotation field direction will not be determined and the character string does not appear on the display.
Fig. Display of the mains frequency (50.0) and the rotation field direction
Fig. No rotation field direction detectable.
UMG 96RM-M
LCD contrast (addr. 035)
The preferred direction of viewing for the LCD is from "below". The user can adjust the LCD contrast of the LCD screen. It is possible to set the contrast in the range from 0 to 9 in steps of 1.
0 = characters are very light 9 = characters are very dark
Factory default setting: 5
Backlight
The LCD backlight allows the display to be read easily even in poor light. The brightness can be controlled by the user in stages from 0 to 9.
The UMG 96RM has two different types of backlight:
- the operation backlight
- the standby backlight
Operation backlight (addr. 036) The operation backlight is activated by pushing the ap­propriate button, or with a restart.
Standby backlight (addr. 747) This backlight is activated after an adjustable period of time (addr. 746). If no button is pressed within this peri­od, then the device switches to the standby backlight. If buttons 1 - 3 are pressed, the device switches to the operation backlight and the defined period of time begins again.
If the brightness settings for the two backlights are set to the same value, then no change is discernible between the operation and standby backlights.
Addr. Description Setting
036 Brightness for
operation backlight
746 Period of time after
which the backlight will switch to standby
747 Brightness for
standby backlight
0 = min. brightness, 9 = max. brightness
range
0 .. 9 6
60 .. 9999 Sek.
0 .. 9 0
Default setting
900 Sek.
51
UMG 96RM-M
Time recording
The UMG 96RM-M records the operating hours and the total running time of each comparator
• where the time of operating hours is measured with a resolution of 0.1 h and is displayed in hours or
• the total running time of the comparator is represented in seconds (when 999999 seconds is reached, the display changes to hours).
For the querying of measured value displays, the times are marked with the numbers 1 to 6:
none = operating hours meter 1 = total running time, comparator 1A 2 = total running time, comparator 2A 3 = total running time, comparator 1B 4 = total running time, comparator 2B 5 = total running time, comparator 1C 6 = total running time, comparator 2C
A maximum of 99999.9 h (= 11.4 years) can be shown on the measured value display.
52
Operating hours meter
The operating hours meter measures the time for which the UMG 96RM-M records and displays measured values. The time of operating hours is measured with a resolution of 0.1 h and is displayed in hours. The operating hours meter cannot be reset.
Total running time of the comparator
The total running time of a comparator is the sum of all time for which there is a limit value violation in the comparator result. The total running time of the comparator can only be reset via the GridVis software. The reset is carried out for all total running times.
Fig. Operating hours meter of the measured value display The UMG 96RM-M shows the number 140.8 h in the ope­rating hours meter. This cor­responds to 140 hours and 80 industrial minutes. 100 in­dustrial minutes correspond to 60 minutes. In this example, 80 industrial minutes therefore represent 48 minutes.
Serial number (addr. 754)
The serial number shown by UMG 96RM-M has 6 digits and is part of the serial number displayed on the name­plate. The serial number cannot be changed.
Serial number display
Serial number information on the nameplate: XX00-0000
Software release (addr. 750)
The software for UMG 96RM is continuously improved and expanded. The software version in the device is marked with a 3-digit number, the software release. The user cannot change the software release.
UMG 96RM-M
53
UMG 96RM-M

Commissioning

Applying the supply voltage

• The level of supply voltage for the UMG 96RM-M can be found on the nameplate.
• After applying the supply voltage, the UMG 96RM-M switches to the first measured value display.
• If no display appears, the supply voltage must be checked to determine whether it is in the rated voltage range.

Applying the measured voltage

• Voltage measurements in networks with rated voltages above 300V AC to ground must be connected to a voltage transformer.
• After the measured voltages are connected, the measured values for the L-N and L-L voltages displayed by the UMG 96RM-M must match those at the voltage measurement input.
Attention!
m
54
Voltages and currents outside the permis­sible metering range can result in personal injury and damage to the device.

Applying the measured current

The UMG 96RM-M is designed for connecting ../1 A and ../5 A current transformers. Only AC currents and not DC currents can be measured via the current measurement inputs. Short circuit all current transformer outputs except for one. Compare the currents displayed on the UMG 96RM-M with the applied current. The current displayed by the UMG 96RM-M must match the input current, taking the current transformer ratio into consideration. In the short circuit current measurement inputs, the UMG-M 96RM must show approx. zero amperes.
The factory-set current transformer ratio is 5/5 A and may need to be adapted to the current transformer used.
Attention!
m
m
Supply voltages that do not correspond to the nameplate information can lead to device malfunction or destruction.
Attention!
The UMG 96RM-M is not suitable for the measurement of DC voltages.
UMG 96RM-M
Rotation field direction
Check the direction of the voltage rotation field on the measured value display of the UMG 96RM-M. Usually there is a "clockwise" spinning rotation field.

Checking the phase assignment

The assignment of the phase conductor to the current transformer is correct if a current transformer is short circuited at the secondary terminals and the current shown by the UMG 96RM-M in the corresponding phase sinks to 0A.

Checking the power measurement

Short circuit all current transformer outputs except for one and check the displayed power. The UMG 96RM-M must only show one rating in the phase with the non-short-circuited current transformer input. If this does not apply, check the measured voltage connection and the measured current connection.
If the magnitude of the real power is correct but the sign of the real power is negative, this can be due to two causes:
• The connections S1 (k) and S2 (I) on the current transformer are inverted.
• Active energy is being returned to the network.

Checking the measurement

If all voltage and current measurement inputs are correctly connected, the individual and sum power ratings are accurately calculated and displayed.

Checking the individual power ratings

If the current transformer is assigned to the wrong phase conductor, the associated power rating will be incorrectly measured and displayed. The assignment of the phase conductor to the current transformer on the UMG 96RM-M is correct if there is no voltage between the phase conductor and the associated current transformer (primary). In order to ensure that a phase conductor on the voltage measurement input is assigned to the correct current transformer, the respective current transformer can be short-circuited at the secondary terminals. The apparent power shown by the UMG 96RM-M must then be zero in this phase.
If the apparent power is correctly displayed but the real power is shown with a "-" sign, the current transformer terminals are inverted or power is being fed to the power company.
55
UMG 96RM-M

Check the sum power ratings

If all voltages, currents and power ratings for the respective phase conductor are correctly displayed, the sum power ratings measured by the UMG 96RM-M must also be correct. For confirmation, the sum power ratings measured by the UMG 96RM-M should be compared with the energy of the active and reactive power meters at the power feed.
56
UMG 96RM-M

M-Bus interface

The data of the parameter list and measurement values list can be accessed via the M-Bus interface with the help of the primary or secondary address. Changing these values is not possible via the M-Bus.
The primary device address is factory preset to „1“. The extended section of the 8-Byte long secondary address is factory preset to contain the device serial number and can be individually changed via the corre­sponding parameters. The device-specific section of the secondary address cannot be adjusted (see page 42).
M-Bus device features
• Addressing possible via primary address and sec­ondary address (0 .. 250)
• Freely selectable number of data points (0 .. 27)
• Supports protocol types: SND_NKE/$E5 and REQ_UD2/RSP_UD2
• Slave search: Search on M-Bus
The UMG 96RM-M loads the M-Bus with
C
an M-Bus device load of 1.5 mA.

Number of data points

The number of data points to be transmitted for the RSP_UD2 telegram is defined via this address.
Address: 080 Meaning: Number of data points for RSP_UD2 Setting range: 0 .. 27 Default setting: 0 (0 = All data points)
In order to call up all data points (0), a telegram must be sent.
Example: Read out data points 1 to 6
Set the parameter of the address to 6. With each request all data points up to and including data point 6 will be transmitted.
Example: Read only data point 10
Set the parameter of the address to 10. With each re­quest all data points up to and including data point 10 will be transmitted. Use only the data point required and ignore those that are not required.
57
UMG 96RM-M

Measurement signal level

The data transfer in the M-Bus network is implemented through modulation of the supply voltage, whereby the voltage for a high signal is 36 V and 24 V for a low signal. The slave device answers the master via the modulation of its current draw, whereby the high signal is 1.5 mA and the low signal is 11-20 mA.
Signal Voltage Reply current
High-Signal 36 V 1.5 mA
Low-Signal 24 V 11-20 mA
58

Structure of the RSP_UD2 telegram

Byte 1 2 3 4 5 6
Name Start Length Length Start C A
Cont. 68 68 8
Byte 7 8 9 10 11 12
Name CI ID1 ID2 ID3 ID4 MAN1
Cont. 72 46
Byte 13 14 15 16 17 18
Name MAN2 GEN MED TC Status SIG1
Cont. 40 8 2 0 0
Byte 19 20 ... ... N-1 N
Name SIG2 DIF Data Data SC Stop
Cont. 0 16
UMG 96RM-M

List of data points

Data
Discription Unit Resolut. Device Format
point
1 Real energy, without backstop dev. Wh 10 0 6
2 Real energy, obtained Wh 10 0 6
3 Real energy, supplied Wh 10 0 6
4 Reactive energy, inductive varh 10 1 6
5 Reactive energy, capacitive varh 10 1 6
6 Reac. energy, without backst. dev. varh 10 1 6
7 Apparent energy VAh 10 2 6
8 Runtime comparator 1a sek 1 1 4
9 Runtime comparator 1b sek 1 2 4
10 Runtime comparator 1c sek 1 3 4
11 Runtime comparator 2a sek 1 4 4
12 Runtime comparator 2b sek 1 5 4
13 Runtime comparator 2c sek 1 6 4
14 Operating hours counter sek 1 0 4
15 I_sum mA 1 4 4
16 P_sum W 1 5 4
17 Q_sum, mains frequency var 1 6 4
18 S_sum VA 1 7 4
Byte
59
UMG 96RM-M
List of data points
Data
Discription Unit Resolut. Device Format
point
19 Uln - Phase L1 mV 100 1 4
20 Uln - Phase L2 mV 100 2 4
21 Uln - Phase L3 mV 100 3 4
22 I - Phase L1 mA 1 1 4
23 I - Phase L2 mA 1 2 4
24 I - Phase L3 mA 1 3 4
25 P - Phase L1 W 1 1 4
26 P - Phase L2 W 1 2 4
27 P - Phase L3 W 1 3 4
Byte
60

Telegramm

UMG 96RM-M
Data point
Discription DIF DIFE DIFE DIFE VIF VIFE
1 Real energy,
without backstop dev.
2 Real energy, obtained 0x86 0x10 X X 0x04 X
3 Real energy, supplied 0x86 0x20 X X 0x04 X
4 Reactive energy, inductive 0x86 0x40 X X 0x04 X
5 Reactive energy, capacitive 0x86 0x50 X X 0x04 X
6 Reac. energy,
without backst. dev
7 Apparent energy 0x86 0x80 0x40 X 0x04 X
8 Runtime comparator 1a 0x84 0x40 X X 0x24 X
9 Runtime comparator 1b 0x84 0x80 0x40 X 0x24 X
10 Runtime comparator 1c 0x84 0xC0 0x40 X 0x24 X
11 Runtime comparator 2a 0x84 0x80 0x80 0x40 0x24 X
12 Runtime comparator 2b 0x84 0xC0 0x80 0x40 0x24 X
13 Runtime comparator 2c 0x84 0x80 0xC0 0x40 0x24 X
14 Operating hours counter 0x04 X X X 0x24 X
15 I_sum 0x84 0x80 0x80 0x40 0xFD 0x59
16 P_sum 0x84 0xC0 0x80 0x40 0x2B X
17 Q_sum
mains frequency
18 S_sum 0x84 0xC0 0xC0 0x40 0x2B X
0x06 X X X 0x04 X
0x86 0x60 X X 0x04 X
0x84 0x80 0xC0 0x40 0x2B X
61
UMG 96RM-M
Telegram
Data point
(X - no value available)
Discription DIF DIFE DIFE DIFE VIF VIFE
19 Uln - Phase L1 0x84 0x40 X X 0xFD 0x48
20 Uln - Phase L2 0x84 0x80 0x40 X 0xFD 0x48
21 Uln - Phase L3 0x84 0xC0 0x40 X 0xFD 0x48
22 I - Phase L1 0x84 0x40 X X 0xFD 0x59
23 I - Phase L2 0x84 0x80 0x40 X 0xFD 0x59
24 I - Phase L3 0x84 0xC0 0x40 X 0xFD 0x59
25 P - Phase L1 0x84 0x40 X X 0x2B X
26 P - Phase L2 0x84 0x80 0x40 X 0x2B X
27 P - Phase L3 0x84 0xC0 0x40 X 0x2B X
62

M-Bus test

Data string M-Bus
$68$F7$F7$68$08$01$72$37$21$10$57$2E$28$09$02$02$00 $00$00$06$04$7E$18$00$00$00$00$86$10$04$7E$18$00$00 $00$00$86$20$04$00$00$00$00$00$00$86$40$04$28$00$00 $00$00$00$86$50$04$00$00$00$00$00$00$86$60$04$28$00 $00$00$00$00$86$80$40$04$92$18$00$00$00$00$84$40$24 $00$00$00$00$84$80$40$24$00$00$00$00$84$C0$40$24$00 $00$00$00$84$80$80$40$24$00$00$00$00$84$C0$80$40$24 $00$00$00$00$84$80$C0$40$24$00$00$00$00$04$24$FA$4F $00$00$84$80$80$40$FD$59$00$00$00$00$84$C0$80$40$2 B$00$00$00$00$84$80$C0$40$2B$00$00$00$00$84$C0$C0$ 40$2B$00$00$00$00$84$40$FD$48$C8$08$00$00$84$80$40 $FD$48$ED$03$00$00$84$C0$40$FD$48$EC$03$00$00$84$ 40$FD$59$00$00$00$00$84$80$40$FD$59$00$00$00$00$84 $C0$40$FD$59$00$00$00$00$84$40$2B$00$00$00$00$84$8 0$40$2B$00$00$00$00$84$C0$40$2B$00$00$00$00$0F$25$ 16
UMG 96RM-M
63
UMG 96RM-M

Analysis via M-Bus Scanners (Excerpt)

Data points 1 to 6
64
Note: The procedure for the verifica­tion of the M-Bus occurred with a M-Bus-Scanner of Wachen­dorff GmbH / Geisenheim. The figure shows a part of the soft­ware and is subject of the copy right of Wachendorff GmbH.
Work values within the software GridVis

Control of the values

UMG 96RM-M
$187E = 6270 * 10 (resolution) = 62700 Wh
65
UMG 96RM-M

Digital outputs

The UMG 96RM-M has 2 digital outputs. The following functions can be optionally assigned to the digital outputs:
Digital output 1 Address 200 = 0 Result of the comparator group 1 Address 200 = 1 Pulse output
Digital output 2 Address 202 = 0 Result of the comparator group 2 Address 202 = 1 Pulse output
Comparator group 1
Comparator A
Comparator B
Comparator C
Pulse output
Addr. 100 = Measured value address Addr. 106 = Minimum pulse length Addr. 102 = Pulse value
66
Logic
Result
Addr. 616
0/1
0/1
Source selection
Addr. 200 =0
Addr. 200 =1
State of digital output 1
Addr. 608 =0
Inverter
Addr. 201=0 (not inverted)
0/1
Addr. 201=1 (inverted)
Fig.: Overall block diagram for digital output 1
UMG 96RM-M
UMG 96RM
Digital output 1
0/1
13
14
Digital outputs - status indicators
UMG 96RM-M
The status of the switching outputs is represented in the UMG 96RM-M display by circular symbols.
State of digital output 1 State of digital output 2
States of the digital output
A current of <1 mA can flow. Digital output 1: Address 608 = 0 Digital output 2: Address 609 = 0
A current of <50 mA can flow. Digital output 1: Address 608 = 1 Digital output 2: Address 609 = 1
C
Since the display is only updated once per second, faster changes of the output states cannot be displayed.
67
UMG 96RM-M

Pulse output

Among other things, the digital outputs can also be used for the output of pulses to meter the energy consumption.
Measured value selection (addr. 100, 101)
Enter the power value here that is to be issued as an
energy pulse. See Table 2. After reaching a certain adjustable amount of energy, a pulse of defined length is applied to the output. Various adjustments must be made in order to use a digital output as a pulse output.
• Digital output
• Source selection
• Measured value selection
Source selection (addr. 200, 202)
Enter the source that delivers the measured value
to be issued at the digital output.
Selectable sources:
• Comparator group
• Pulse
• Pulse length
• Pulse value
Source
Pulse
Addr. 100 = 874 (Address of Psum3) Addr. 106 = 50 (ms) Addr. 102 = 1000 (Wh/Impuls)
0/1
selection
Addr. 200 =1
Fig.: Block diagram; Example of digital output 1 as a pulse output.
68
Addr. 608 =0
0/1
State of digital output 1
Inverter
Addr. 201=0 (not inverted) Addr. 201=1 (inverted)
UMG 96RM-M
UMG 96RM
Digital output 1
0/1
13
14
Pulse length (addr. 106)
UMG 96RM-M
The pulse length applies for both pulse outputs and is permanently fixed via parameter address 106.
Adjustment range: 1 .. 1000 1 = 10ms Default: 5 = 50ms
The typical pulse length for S0 pulses is 30 ms.
Pulse pause
The pulse pause is at least as long as the selected pulse length. The pulse pause depends on the measured energy, for example, and can be hours or days.
Pulse length 10 ms .. 10 s
C
Pulse pause
>10 ms
Pulse spacing
The pulse spacing is proportional to the power within the selected setting.
Due to the minimum pulse length and minimum pulse pause, the values in the table are for the maximum number of pulses per hour.
Pulse length Pulse pause Maximum pulses/hour
10 ms 10 ms 180,000 pulses/hour
30 ms 30 ms 60,000 pulses/hour
50 ms 50 ms 36,000 pulses/hour
100 ms 100 ms 18,000 pulses/hour
500 ms 500 ms 3,600 pulses/hour
1 s 1 s 1,800 pulses/hour
10 s 10 s 180 pulses/hour
Examples for the maximum possible number of pulses per hour.
69
UMG 96RM-M
Pulse value (addr. 102, 104)
The pulse value specifies how much energy (Wh or varh) should correspond to a pulse. The pulse value is determined by the maximum connected load and the maximum number of pulses per hour.
If the pulse value is specified with a positive sign, pulses will only be issued if the measured value also has a positive sign.
If the pulse value is specified with a negative sign, pulses will only be issued if the measured value also has a negative sign.
Pulse value =
C C
70
maximum connection power
maximum number of pulses per hour
Since the active energy meter works with a return stop, pulses are only issued during import of electrical energy.
Since the reactive energy meter works with a return stop, pulses are only issued under inductive load.
[pulse/Wh]
UMG 96RM-M
Determining the pulse value
Setting the pulse length Set the pulse length according to the requirements of the connected pulse receiver. For a pulse length of 30 ms, for example, the UMG 96RM can issue a maximum number of 60,000 pulses (see Table "Maximum Pulse Number") per hour.
Determining the maximum connected load Example:
Current transformer = 150/5 A L-N voltage = max. 300 V
Power per phase = 150 A x 300 V = 45 kW Power for 3 phases = 45 kW x 3 Maximum connected load = 135 kW
Calculating the pulse value
Pulse value =
maximum connection power
maximum number of pulses per hour
[pulse/Wh]
Pulse value = 135 kW / 60000 pulses/h Pulse value = 0.00225 pulses/kWh Pulse value = 2.25 pulses/Wh
operating voltage
UMG 96RM
Switching and pulse outputs
+24V=
Fig.: Connection example for wiring the pulse output.
When using the digital outputs as a pulse
C
output, the auxiliary voltage (DC) must only have a maximum residual ripple of 5%.
External
13
14
15
230 V AC
24 V DC
+ -
Data logger
1.5 k
71
UMG 96RM-M
Limit value monitoring
Two comparator groups are available for monitoring a limit value.
Comparator group 1 is assigned to digital output 1 and comparator group 2 is assigned to digital output 2.
Source
Comparator group 1
Comparator A
Comparator B
Comparator C
Logic
Result
Addr. 616
0/1
selection
Addr. 200 =0
Block diagram: Use of digital output 1 for limit value monitoring.
72
State of digital output 1
Addr. 608 =0
Inverter
0/1
Adr. 201=0 (not inverted) Adr. 201=1 (inverted)
UMG 96RM-M
UMG 96RM
Digital output 1
0/1
13
14
Example: Current monitoring in the neutral line
UMG 96RM-M
If the current in the neutral line is greater than 100 A for 60 seconds, the digital output 1 should trip for at least 2 minutes.
The following must be programmed:
1. Comparator group 1 Select comparator group 1 for the limit value monitoring. The comparator group acts only on digital output 1. Since only one limit value is monitored, select comparator A and program it as follows:
The address of the measured value to be monitored by comparator A: Address 110 = 866 (address of the current in the neutral line)
The measured values for the B and C comparators are set to 0. Address 116 = 0 (the comparator is inactive) Address 122 = 0 (the comparator is inactive)
The limit value to be observed. Address 108 = 100 (100 A)
For a minimum exposure time of 2 minutes, digital output 1 should remain switched if the limit value is exceeded. Address 111 = 120 seconds
For the lead time of 60 seconds, any exceeding should be minimised. Address 112 = 60 seconds
The operator for comparison between the measured value and the limit value. Address 113 = 0 (corresponds >=)
2. Source selection Select comparator group 1 as the source. Address 200 = 0 (comparator group 1)
3. Inverter The result from comparator group 1 can also be inverted here. The result is not inverted. Address 201 = 0 (not inverted)
4. Linking comparators The B and C comparators have not been set and are equal to zero. The result of comparator A is issued as a comparator result through the OR link of comparators A, B and C. Address 107 = 0 (OR link)
Result Digital output 1 is tripped for at least 2 minutes if the current in the neutral line is greater than 100 A for more than 60 seconds. Digital output 1 is conductive. Current can flow.
73
UMG 96RM-M

Comparator

Two comparator groups, each with 3 comparators, are available for monitoring limit values. The results from comparators A, B and C can be AND or OR linked.
The linkage result from comparator group 1 can be assigned to digital output 1 and the linkage result from comparator group 2 is assigned to digital output 2.
74
Comparator group 1
Comparator A
Measured value (addr. 110) Limit value (addr. 108) Minimum turn-on time (addr. 111) Lead time (addr. 112) Operator ">=", "<" (addr. 113)
Comparator result (addr. 610) Comparator result (addr. 611) Comparator result (addr. 612)
Total running time
(addr. 5898)
Link the results from comparators A, B and C
Link the results from comparators A, B and C as AND or OR (addr. 107).
Comparator B
Measured value (addr. 116) Limit value (addr. 114) Minimum turn-on time (addr. 117) Lead time (addr. 118) Operator ">=", "<" (addr. 119)
Total running time
(addr. 5900)
Linkage result (addr. 616)
Comparator C
Measured value (addr. 122) Limit value (addr. 120) Minimum turn-on time (addr. 123) Lead time (addr. 124) Operator ">=", "<" (addr. 125)
Total running time
(addr. 5902)
UMG 96RM-M
• Measured value (addr. 110,116,122,129,135,141)
The address of the measured value to be monitored is in the measured value. If measured value = 0, the comparator is inactive.
• Limit value (addr. 108,114,120,127,133,139)
Write the value in the limit that is to be compared with the measured value.
• Minimum turn-on time (addr. 111,117,123,130,136,142)
The linkage result (e.g. address 610) is maintained for the duration of the minimum turn-on time. Adjustment range: 1 to 32,000 seconds
• Lead time (addr. 112,118,124,131,137,143)
If a limit value violation is present for at least the duration of the lead time, the comparator result is changed. Times in the range from 1 to 32,000 seconds can
be assigned to the lead time.
• Operator (addr. 113,119,125,132,138,144)
Two operators are available for comparing the measured value and the limit value. Operator = corresponds to 0 greater than or equal to (>=) Operator = corresponds to 1 less than (<)
• Comparator result (addr. 610,611,612,613,614,615)
The result from the comparison between the measured value and the limit value is in the comparator result. Therefore: 0 = there is no limit value violation. 1 = there is a limit value violation.
• Total running time
The sum of all times for which there was a limit value violation in the comparator result.
• Linkage (addr. 107, 126)
Link the results from comparators A, B and C as AND or OR.
• Linkage (addr. 107, 126)
Link the results from comparators A, B and C as AND or OR.
• Total linkage result (addr. 616,617)
The linked comparator results from comparators A, B and C are in the total linkage result.
75
UMG 96RM-M
Limit value
Exceedance
Measured value
Lead time
Minimum
turn-on time
Comparator result
76
2 seconds
2 seconds
UMG 96RM-M

Parameter list comparator and digital outputs

Address Format RD/WR Unit Note Adjustment Range Default
100 SHORT RD/WR - Address of the measured value, Digital output 1 0..32000 0 101 SHORT RD/WR - Address of the measured value, Digital output 2 0..32000 0 102 FLOAT RD/WR Wh Pulse value, Digital output 1 -1000000..+1000000 0 104 FLOAT RD/WR Wh Pulse value, Digital output 2 -1000000..+1000000 0 106 SHORT RD/WR 10ms Minimum pulse length (1=10 ms) Digital output 1/2 1..1000 5 (=50ms) 107 SHORT RD/WR - Result from comparator group 1; 0,1 0 Link A, B, C (1=and, 0=or) 108 FLOAT RD/WR - Comparator 1A, Limit value -1012-1..+1012-1 0 110 SHORT RD/WR - Comparator 1A, Address of the measured value 0..32000 0 111 SHORT RD/WR s Comparator 1A, Minimum turn-on time 0..32000 0 112 SHORT RD/WR s Comparator 1A, Lead time 0..32000 0 113 SHORT RD/WR - Comparator 1A, Operator 0,1 0 „>=“=0, „<“=1 114 FLOAT RD/WR - Comparator 1B, Limit value -1012-1..+1012-1 0 116 SHORT RD/WR - Comparator 1B,, Address of the measured value 0..32000 0 117 SHORT RD/WR s Comparator 1B, Minimum turn-on time 0..32000 0 118 SHORT RD/WR s Comparator 1B, Lead time 0..32000 0 119 SHORT RD/WR - Comparator 1B, Operator 0,1 0 „>=“=0 „<“=1 120 FLOAT RD/WR - Comparator 1C, Limit value -1012-1..+1012-1 0
77
UMG 96RM-M
Address Format RD/WR Unit Note Adjustment Range Default
122 SHORT RD/WR - Comparator 1C, Address of the measured value 0..32000 0 123 SHORT RD/WR s Comparator 1C, Minimum turn-on time 0..32000 0 124 SHORT RD/WR s Comparator 1C, Lead time 0..32000 0 125 SHORT RD/WR - Comparator 1C, Operator 0,1 0 „>=“=0 „<“=1 126 SHORT RD/WR - Result from comparator group 2; 0,1 0 Link A, B, C (1=and, 0=or) 127 FLOAT RD/WR - Comparator 2A, Limit value -1012-1..+1012-1 0 129 SHORT RD/WR - Comparator 2A, Address of the measured value 0..32000 0 130 SHORT RD/WR s Comparator 2A, Minimum turn-on time 0..32000 0 131 SHORT RD/WR s Comparator 2A, Lead time 0..32000 0 132 SHORT RD/WR - Comparator 2A, Operator 0,1 0 „>=“=0 „<“=1 133 FLOAT RD/WR - Comparator 2B, Limit value -1012-1..+1012-1 0 135 SHORT RD/WR - Comparator 2B, Address of the measured value 0..32000 0 136 SHORT RD/WR s Comparator 2B, Minimum turn-on time 0..32000 0 137 SHORT RD/WR s Comparator 2B, Lead time 0..32000 0 138 SHORT RD/WR - Comparator 2B, Operator 0,1 0 „>=“=0 „<“=1 139 FLOAT RD/WR - Comparator 2C, Limit value -1012-1..+1012-1 0 141 SHORT RD/WR - Comparator 2C, Address of the measured value 0..32000 0 142 SHORT RD/WR s Comparator 2C, Minimum turn-on time 0..32000 0
78
Address Format RD/WR Unit Note Adjustment Range Default
143 SHORT RD/WR s Comparator 2C, Lead time 0..32000 0 144 SHORT RD/WR - Comparator 2C, Operator 0,1 0 „>=“ = 0 „<“ = 1
200 SHORT RD/WR - Select the source for Digital output 1 0..4 201 SHORT RD/WR - Digital output 1 inverter 0..1 202 SHORT RD/WR - Select the source for Digital output 2 0..4 203 SHORT RD/WR - Digital output 2 inverter 0..1
602 SHORT RD/WR - Value for output 1 0, 1 605 SHORT RD/WR - Value for output 2 0, 1 608 SHORT RD - State of output 1 609 SHORT RD - State of output 2 610 SHORT RD - Comparator result 1 Output A 611 SHORT RD - Comparator result 1 Output B 612 SHORT RD - Comparator result 1 Output C 613 SHORT RD - Comparator result 2 Output A 614 SHORT RD - Comparator result 2 Output B 615 SHORT RD - Comparator result 2 Output C 616 SHORT RD - Linkage result of comparator group 1 617 SHORT RD - Linkage result of comparator group 2
*1
0
*2
0
*1
0
*2
0
*1 0=comparator group, 1=pulse output, 2=reserved, 3=reserved, 4=reserved *2 0= not inverted, 1=inverted
UMG 96RM-M
79
UMG 96RM-M

Service and maintenance

The device is subject to various safety tests prior to delivery and is marked with a seal. If a device is opened, the safety tests must be repeated. A warranty is only given for unopened devices.
Repair and calibration
Repairs and calibration can only be carried out by the manufacturer.
Front membrane
The front membrane can be cleaned with a soft cloth and common household cleaning agents. Acids and acidic agents must not be used for cleaning.
Disposal
The UMG 96RM-M can be disposed of as electronic scrap in accordance with the statutory recycling provisions. The lithium battery must be disposed of separately.
80
Service
If questions arise that are not described in this manual, please contact the manufacturer directly.
We require the following information from you in order to deal with questions:
- device designation (see nameplate),
- serial number (see nameplate),
- software release (see measured value display),
- measured voltage and supply voltage,
- precise description of the error.

Device calibration

The devices are calibrated by the manufacturer at the factory - it is not necessary to recalibrate the de­vice providing that the environmental conditions are complied with.

Calibration intervals

It is recommended to have a new calibration carried out by the manufacturer or an accredited laboratory every 5 years approximately.
UMG 96RM-M
81
UMG 96RM-M

Error messages

The UMG 96RM-M shows three different error messages on the display:
- warnings,
- serious error and
- metering range exceedances.
If there are warnings and serious errors, the error message is indicated by the symbol "EEE" followed by an error number.
Symbol for an error message
Error number
The three-digit error number is composed of the error description and (if detectable by the UMG 96RM-M) one or more error causes.
82
Symbol for an error message
Error cause
Description of the error
Example of error message 911:
The error number is composed of serious error 910 and internal error cause 0x01.
In this example, an er­ror occurred when read­ing the calibration from the EEPROM. The device must be sent to the manu­facturer for inspection.
UMG 96RM-M
Warnings
Warnings are less serious errors and can be acknowledged with buttons 1 or 2. The measured values continue to be recorded and displayed. This error is re­displayed after each voltage recovery.
Error Description of the error EEE 500
Serious errors
The device must be sent to the manufacturer for inspection.
Error Description of the error EEE 910
The mains frequency cannot be determined. Possible causes:
The voltage on L1 is too small. The mains frequency is not in the range from 45 to 65Hz.
Error when reading the calibration.
Internal causes of the error
The UMG 96RM-M can usually determine the cause of an internal error and then report it with the following error code. The device must be sent to the manufacturer for inspection.
Error Description of the error 0x01 EEPROM does not answer. 0x02 Address range exceeded. 0x04 Checksum error. 0x08 Error in the internal I2C bus.
83
UMG 96RM-M
Metering range exceedance
Metering range exceedances are displayed for as long as they are present and cannot be acknowledged. A metering range is exceeded if at least one of the three voltage or current measuring inputs is outside of its specified metering range. The phase in which the metering range exceedance occurred is indicated with the "up" arrow. The "V" and "A" symbols show whether the metering range exceedance occurred in the current or voltage circuit.
A = current circuit V = voltage circuit
Display of the phase (L1/L2/ L3) with the metering range exceedance.
Limit values for metering range exceedance:
I = 7 Aeff UL-N = 520 VL-N
84
Examples
A = current circuit
Fig.: Display of the metering range exceedance in the current circuit of the 2nd phase (I2).
V = voltage circuit
Fig.: Display of the metering range exceedance in the voltage circuit L3.
Parameters of the metering range exceedance
A continuative error description is stored encoded in the parameters of the metering range exceedance (addr. 600) in the following format:
UMG 96RM-M
FFFFFFFF
Phase 1:
Phase 2:
Phase 3:
Example: Error in phase 2 in the current circuit:
0xF2FFFFFF
Example: Error in phase 3 in the voltage circuit UL-N:
0xFFF4FFFF
0x
1
1
2
2
4
4
Current:
U L-N
85
UMG 96RM-M
Procedure in case of error
Possibility of error Cause Help
No display External fuse for the power supply has
No current display Measurement voltage not
The displayed current is too large or too small.
The displayed voltage is too small or too large.
The displayed voltage is too small. Metering range exceedance. Use voltage transformer.
tripped.
connected.
Measurement current not connected. Connect measurement current.
Current measurement in the wrong phase. Check and correct the connection if
Voltage transformer factor incorrectly programmed.
The peak current value at the measurement input was exceeded by current harmonics.
The current at the measurement input was exceeded.
Measurement in the wrong phase. Check and correct the connection if
Voltage transformer incorrectly programmed.
The peak voltage value at the measurement input was overwritten by harmonics.
Replace fuse.
Connect measurement voltage.
necessary.
Read and program the current transformer ratio on the current transformer.
Install current transformer with a higher current transformer ratio.
Install current transformer with a lower current transformer ratio.
necessary.
Read and program the voltage transformer ratio on the voltage transformer.
Attention! It must be ensured that the measurement inputs are not overloaded.
86
Possibility of error Cause Help
Ind./cap. phase shift The current circuit is assigned to the
Real power is too small or too large. The programmed current transformer ratio
wrong voltage circuit.
is incorrect.
The current circuit is assigned to the wrong voltage circuit.
The programmed voltage transformer ratio is incorrect.
Check and correct the connection if necessary.
Reading and programming the current transformer ratio on the current transformer
Check and correct the connection if necessary.
Read and program the voltage transformer ratio on the voltage transformer.
UMG 96RM-M
The active energy import/export is inverted.
An output is not reacting.
"EEE" on the display See error messages.
No connection to the device. Incorrect device address Correct the device address.
Despite the aforementioned measures the device does not work.
At least one current transformer connection is inverted.
A current circuit is assigned to the wrong voltage circuit.
The output was incorrectly programmed. Check the programming and correct if
The output was incorrectly connected. Check and correct the connection if
Different bus speeds (baud rate) Correct the speed (baud rate).
Device is defective. Send the device to the manufacturer
Check and correct the connection if necessary.
Check and correct the connection if necessary.
necessary.
necessary.
for inspection and include a detailed description of the error.
87
UMG 96RM-M

Technical data

General
Net weight (with attached connectors) 300g
Packaging weight (including accessories) 625g
Device dimensions ca. l = 42mm, w = 97mm, h = 100mm
Service life of the backlight 40,000 hours (50% of initial brightness)
Transport and storage
The following information applies for devices that are transported or stored in their original packaging.
Free fall 1m
Temperature K55 (-25 °C to +70 °C)
Relative humidity 0 to 90% RH
Ambient conditions during operation
The UMG 96RM-M is intended for use in weather-protected, fixed locations. Protection class II according to IEC 60563 (VDE 0106, part 1).
Rated temperature range K55 (-10 °C .. +55 °C)
Relative humidity 0 to 75 % RH
Operational altitude 0 .. 2000 m above sea level
Degree of pollution 2
Installation position vertical
Ventilation Forced ventilation is not required.
Foreign body and water protection
- Front
- Back
- Front with seal
88
IP40 according to EN60529 IP20 according to EN60529 IP42 according to EN60529
Supply voltage
Installation overvoltage category 300V CAT II
Protection of the power supply (fuse) 6 A, type C (approved by UL/IEC)
Nominal range 20V - 250V (45..65Hz) oder DC 20V - 300V
Working area +-10% from the nominal range
Power consumption max. 4.5VA / 2W
Connection capacity of the terminals (power supply)
Connectable conductor. Only one conductor may be connected per contact point!
Single-wire, multi-wire, finely stranded conductor 0.2 - 2.5mm2, AWG 26 - 12
Pin terminals, ferrules 0.2 - 2.5mm
Tightening torque 0.4 - 0.5Nm
Stripping length 7mm
Outputs
2 digital outputs, semi-conductor relay, not short circuit protected.
Switching voltage max. 33V AC, 60V DC
Switching current max. 50mAeff AC/DC
Reaction time 10/12 periods + 10ms *
Pulse output (energy pulses) max. 50Hz
* Reaction time at 50 Hz, for example: 200 ms + 10 ms = 210 ms
2
UMG 96RM-M
89
UMG 96RM-M
Connection capacity of the terminals (outputs)
Rigid/flexible 0.14 - 1.5mm2, AWG 28-16
Flexible with ferrules without plastic sleeve 0.20 - 1.5mm
Flexible with ferrules with plastic sleeve 0.20 - 1.5mm
Tightening torque 0.20 - 0.25Nm
Stripping length 7mm
Voltage metering
Three-phase, 4-wire systems with nominal voltages up to 277V/480V (+-10%)
Three-phase, 3-wire systems, unearthed, with nominal voltages up to
IT 480V (+-10%)
Overvoltage category 300V CAT III
Rated surge voltage 4kV
Metering range L-N 01) .. 300 Vrms
Metering range L-L 01) .. 520Vrms
(max. overvoltage 520 Vrms )
(max. overvoltage 900Vrms )
Resolution 0.01V
Crest factor 2.45 (relative to the metering range)
Impedance 4MOhm/phase
Power consumption approx. 0.1 VA
Sampling rate 21.33 kHz (50 Hz), 25.6 kHz (60 Hz) per measuring channel
Mains frequency
- Resolution
1)
The UMG 96RM can only determine measured values if a L-N voltage greater than 10 Veff or a L-L voltage larger than 18 Veff is applied
to at least one voltage measurement input.
45 Hz .. 65 Hz
0.01 Hz
2
2
90
Current measurement
Rated current
Metering range
Crest factor
Resolution
Overvoltage category
Rated surge voltage
Power consumption
Overload for 1 sec.
Sampling rate
Connection capacity of the terminals (voltage and current measurement)
Connectable conductor. Only one conductor may be connected per contact point!
5A
0 .. 6Arms
1.98
0.1 mA (display 0.01 A)
300V CAT II
2kV
ca. 0.2 VA (Ri=5mOhm)
120 A (sinusoidal)
21.33 kHz (50 Hz), 25.6 kHz (60 Hz) per measuring channel
Current Voltage
Single-wire, multi-wire, finely stranded conductor
Pin terminals, ferrules
Tightening torque
Stripping length
0.2 - 2.5mm2, AWG 26-12 0.08 - 4.0mm2, AWG 28-12
0.2 - 2.5mm
0.4 - 0.5Nm 0.4 - 0.5Nm
7mm 7mm
UMG 96RM-M
2
0.2 - 2.5mm
2
91
UMG 96RM-M
Serial interface
M-Bus 300, 600, 1200, 2400, 4800, 9600, 19200, 38400 baud
M-Bus device load max. 20 mA
Stripping length 7mm
Connection capacity of the terminals (M-Bus)
Single-wire, multi-wire, finely stranded conductor
Pin terminals, ferrules
Tightening torque
Stripping length
92
0.20 - 1.5mm
0.20 - 1.5mm
0.20 - 0.25Nm
7mm
2
2
UMG 96RM-M
93
UMG 96RM-M

Parameters of functions

Function Symbol Accuracy class Metering range Display range
Total real power P 0.55) (IEC61557-12) 0 .. 5.4 kW 0 W .. 999 GW * Total reactive power QA, Qv 1 (IEC61557-12) 0 .. 5.4 kvar 0 varh .. 999 Gvar * Total apparent power SA, Sv 0.55) (IEC61557-12) 0 .. 5.4 kVA 0 VA .. 999 GVA * Total active energy Ea 0.55) (IEC61557-12) 0 .. 5.4 kWh 0 Wh .. 999 GWh * Total reactive energy ErA, ErV 1 (IEC61557-12) 0 .. 5.4 kvarh 0 varh .. 999 Gvarh * Total apparent energy EapA, EapV 0.55) (IEC61557-12) 0 .. 5.4 kVAh 0 VAh .. 999 GVAh * Frequency f 0.05 (IEC61557-12) 45 .. 65 Hz 45.00 Hz .. 65.00 Hz Phase current I 0.5 (IEC61557-12) 0 .. 6 Arms 0 A .. 999 kA Measured neutral conductor current Calculated neutral conductor current Voltage U L-N 0.2 (IEC61557-12) 10 .. 300 Vrms 0 V .. 999 kV Voltage U L-L 0.2 (IEC61557-12) 18 .. 520 Vrms 0 V .. 999 kV Displacement factor PFA, PFV 0.5 (IEC61557-12) 0.00 .. 1.00 0.00 .. 1.00 Short-term flicker, long-term flicker Pst, Plt - - ­Voltage dips (L-N) Udip - - ­Voltage surges (L-N) Uswl - - ­Transient overvoltages Utr - - ­Voltage interruptions Uint - - ­Voltage unbalance (L-N) Voltage unbalance (L-N)
1)
2)
Voltage harmonics Uh Class 1 (IEC61000-4-7) up to 2.5 kHz 0 V .. 999 kV THD of the voltage THD of the voltage
3)
4)
IN - - ­INc 1.0 (IEC61557-12) 0.03 .. 25 A 0.03 A .. 999 kA
Unba - - ­Unb - - -
THDu 1.0 (IEC61557-12) up to 2.5 kHz 0 % .. 999 % THD-Ru - - -
94
UMG 96RM-M
Function Symbol Accuracy class Metering range Display range
Current harmonics Ih Class 1 (IEC61000-4-7) up to 2.5 kHz 0 A .. 999 kA THD of the current THD of the current Mains signal voltage MSV - - -
1) Referred to amplitude.
2) Referred to phase and amplitude.
3)
4)
THDi 1.0 (IEC61557-12) up to 2.5 kHz 0 % .. 999 % THD-Ri - - -
* The display returns to 0 W when the maximum total energy values are reached.
3) Referred to mains frequency.
4) Referred to root mean square value.
5) Accuracy class 0.5 with ../5 A transformer. Accuracy class 1 with ../1 A transformer.
95
UMG 96RM-M
Parameter and Modbus address list
The following excerpt from the parameter list contains settings that are necessary for proper operation of the UMG 96RM, such as current transformers and device addresses. The values in the parameter list can be written and read.
In the excerpt, the measured value list files the measured and calculated measured values, output status data and recorded values so that they can be read.

Table 1 - Parameter list

Adress Format RD/WR Unit Note Adjustment Range Default
0 SHORT RD/WR - Device address 0..255 1 SHORT RD/WR kbps Baud rate (0=300, 1=600, 2=1200, 0..7 5 3= 2400, 4=4800, 5=9600. 6=19200, 7=38400 Baud 3 SHORT RD/WR Only for internal use 10 FLOAT RD/WR A Current transformer I1, primary 0..1000000 12 FLOAT RD/WR A Current transformer I1, sec. 1..5 5 14 FLOAT RD/WR V Voltage transformer V1, prim. 0..1000000 16 FLOAT RD/WR V Voltage transformer V1, sec. 100, 400 400 18 FLOAT RD/WR A Current transformer I2, primary 0..1000000 20 FLOAT RD/WR A Current transformer I2, sec. 1..5 5 22 FLOAT RD/WR V Voltage transformer V2, prim 0..1000000 400
(*1)
The values 0 and 248 to 255 are reserved and must not be used.
(*2)
The adjustable value 0 does not produce any sensible energy values and must not be used.
96
(*1)
1
(*2)
5
(*2)
400
(*2)
5
UMG 96RM-M
Adress Format RD/WR Unit Note Adjustment Range Default
24 FLOAT RD/WR V Voltage transformer V2, sec. 100, 400 400 26 FLOAT RD/WR A Current transformer I3, primary 0..1000000 5 28 FLOAT RD/WR A Current transformer I3, sec. 1..5 5 30 FLOAT RD/WR V Voltage transformer V3, prim. 0..1000000 400 32 FLOAT RD/WR V Voltage transformer V3, sec. 100, 400 400 34 SHORT RD/WR Hz Frequency determination 0, 45 .. 65 0 0=Auto, 45 .. 65=Hz 35 SHORT RD/WR - Display contrast 0 .. 9 5 0 (low), 9 (high) 36 SHORT RD/WR - Backlight 0 .. 9 6 0 (low), 9 (high)
37 SHORT RD/WR - Display profile 0 .. 3 0 0=default display profile 1=default display profile 2=default display profile 3=only for internal use 38 SHORT RD/WR - Display change profile 0 .. 3 0
0..2=default display change profiles 39 SHORT RD/WR s Changeover time 0 .. 60 0
40 SHORT RD/WR - Averaging time, I 0 .. 8* 6 41 SHORT RD/WR - Averaging time, P 0 .. 8* 6 42 SHORT RD/WR - Averaging time, U 0 .. 8* 6
45 USHORT RD/WR mA Response threshold of 0 .. 50 5 I1 .. I3
50 SHORT RD/WR - Password 0 .. 999 0 (no password)
* 0 = 5sec.; 1 = 10sec.; 2 = 15sec.; 3 = 30sec.; 4 = 1min.; 5 = 5min.; 6 = 8min.; 7 = 10min.; 8 = 15min.
97
UMG 96RM-M
Adress Format RD/WR Unit Note Adjustment Range Default
81 SHORT RD/WR - Secondary address, 0..99 extended section 1 82 SHORT RD/WR - Secondary address, 0..99 extended section 2 83 SHORT RD/WR - Secondary address, 0..99 extended section 3 84 SHORT RD/WR - Secondary address, 0..99 extended section 4 500 SHORT RD/WR - Terminal assignment, I L1 -3..0..+3 +1 501 SHORT RD/WR - Terminal assignment, I L2 -3..0..+3 +2 502 SHORT RD/WR - Terminal assignment, I L3 -3..0..+3 +3 503 SHORT RD/WR - Terminal assignment, U L1 0..3 1 504 SHORT RD/WR - Terminal assignment, U L2 0..3 2 505 SHORT RD/WR - Terminal assignment, U L3 0..3 3 506 SHORT RD/WR - Clear min. and max. values 0..1 0 507 SHORT RD/WR - Clear energy meter 0..1 0 508 SHORT RD/WR - Force write EEPROM 0..1 0
Note: Energy values and minimum and maximum values are written to the EEPROM every 5 minutes.
509 SHORT RD/WR - Voltage connection diagram 0..7 0 510 SHORT RD/WR - Current connection diagram 0..8 0 511 SHORT RD/WR - Relative voltage for THD and FFT 0, 1 0
The voltages for THD and FFT can be shown on the display as L-N or L-L values. 0=LN, 1=LL
600 UINT RD/WR - Metering range exceedance 0..0xFFFFFFFF
746 SHORT RD/WR s Period of time after which the backlight will switch to standby 60 .. 9999 900 747 SHORT RD/WR s Brightness of the standby backlight 0 .. 9 0
98
Only the first three positions (###) of a value are shown on the display. Values larger than
C
1,000 are marked with „k”. Example: 003k = 3000
Tabelle 2 - Adress list (frequently used measured values)
Addresse Format RD/WR Unit Note
750 SHORT RD - Software release
754 SERNR RD - Serial number 756 SERNR RD - Production number
800 FLOAT RD Hz Frequency 802 FLOAT RD - Voltage, positive sequence 804 FLOAT RD - Voltage, negative sequence 806 FLOAT RD - Voltage, zero sequence 808 FLOAT RD V Voltage L1-N 810 FLOAT RD V Voltage L2-N 812 FLOAT RD V Voltage L3-N 814 FLOAT RD V Voltage L1-L2 816 FLOAT RD V Voltage L2-L3 818 FLOAT RD V Voltage L1-L3 820 FLOAT RD - Fund. power factor, CosPhi; U L1-N IL1 822 FLOAT RD - Fund. power factor,, CosPhi; U L2-N IL2 824 FLOAT RD - Fund. power factor,, CosPhi; U L3-N IL3 826 FLOAT RD - Sum; CosPhi sum3=POsum3/Ssum3 828 FLOAT RD - Power factor; U L1-N IL1 830 FLOAT RD - Power factor; U L2-N IL2 832 FLOAT RD - Power factor; U L3-N IL3 834 FLOAT RD - Sum; Power factor sum3=Psum3/Ssum3 836 FLOAT RD % THD, U L1N, based on U0 L1 838 FLOAT RD % THD, U L2N, based on U0 L2
UMG 96RM-M
99
UMG 96RM-M
Addresse Format RD/WR Unit Note
840 FLOAT RD % THD, U L3N, based on U0 L3 842 FLOAT RD % THD, U L1L2, based on U0 L1L2 844 FLOAT RD % THD, U L2L3, based on U0 L2L3 846 FLOAT RD % THD, U L1L3, based on U0 L1L3 848 FLOAT RD V Voltage, real part U1 L1N 850 FLOAT RD V Voltage, real part U2 L2N 852 FLOAT RD V Voltage, real part U3 L3N 854 FLOAT RD V Voltage, imaginary part U L1N 856 FLOAT RD V Voltage, imaginary part U L2N 858 FLOAT RD V Voltage, imaginary part U L3N 860 FLOAT RD A Current I1 L1 862 FLOAT RD A Current I2 L2 864 FLOAT RD A Current I3 L3 866 FLOAT RD A Vector sum; IN=I1+I2+I3 868 FLOAT RD W Real power P1 L1N 870 FLOAT RD W Real power P2 L2N 872 FLOAT RD W Real power P3 L3N 874 FLOAT RD W Sum; Psum3=P1+P2+P3 876 FLOAT RD var Fund. reactive power Q1 L1N 878 FLOAT RD var Fund. reactive power Q2 L2N 880 FLOAT RD var Fund. reactive power Q3 L3N 882 FLOAT RD var Sum; Qsum3=Q1+Q2+Q3 884 FLOAT RD VA Apparent power S1 L1N 886 FLOAT RD VA Apparent power S2 L2N 888 FLOAT RD VA Apparent power S3 L3N 890 FLOAT RD VA Sum; Ssum3=S1+S2+S3 892 FLOAT RD W Fund. real power P01 L1N 894 FLOAT RD W Fund. real power P02 L2N 896 FLOAT RD W Fund. real power P03 L3N 898 FLOAT RD W Sum; P0sum3=P01+P02+P03 900 FLOAT RD var Harmonic distortion power D1 L1N 902 FLOAT RD var Harmonic distortion power D2 L2N
100
Loading...