Janitza UMG 96-PA User guide

Power Quality Analyzer
UMG 96-PA
User Manual and Technical Data
Doc. no. 2.061.015.1b 07/2018
www.janitza.com
Janitza electronics GmbH Vor dem Polstück 6 D-35633 Lahnau Support phone number +49 6441 9642-22 Fax +49 6441 9642-30 E-mail: info@janitza.com Internet: http://www.janitza.com
UMG 96-PA www.janitza.de
Content
1. General 6
1. 1 Disclaimer 6
1. 2 Copyright notice 6
1. 3 Technical modifications 6
1. 4 Declaration of Conformity 6
1. 5 Comments on the Manual 6
2. Safety 8
2. 1 Safety Instructions 8
2. 2 Safety Measures 9
2. 3 Qualified Staff 9
3. Intended Use 10
3. 1 Input Check 10
3. 2 Intended Use 10
3. 3 Delivery Contents 11
3. 4 Deliverable accessories 11
4. Product Description 12
4. 1 Measurement Procedure 12
4. 2 Operating Design 12
4. 3 Network Analysis Software GridVis® 12
4. 4 Performance characteristics 13
4. 5 Product Overview 14
5. Installation 16
5. 1 Installation site 16
5. 2 Installation position 16
5. 3 Securing the device 16
6. Network systems 17
6. 1 Voltage measurement 18
6. 2 Rated voltages 19
6. 2. 1 Three-phase 4-wire network with earthed neutral conductor 19
2
www.janitza.de UMG 96-PA
7. Installation 20
7. 1 Connection to a PC 20
7. 2 Circuit breaker 21
7. 3 Supply voltage 21
7. 4 Measured voltage 22
7. 4. 1 Overvoltage 22
7. 4. 2 Frequency 22
7. 4. 3 Connection variants for voltage measurement 24
7. 5 Current measurement 24
7. 5. 1 Current measurement connection variants 25
7. 5. 2 Direction of the current 26
7. 5. 3 Total current measurement 26
7. 5. 4 Ammeter 26
8. Interface 27
8. 1 Screening 27
8. 2 Termination resistors 28
8. 3 Bus structure 29
9. Digital inputs and outputs 30
9. 1 Digital inputs 30
9. 1. 1 S0 pulse input 31
9. 2 Digital outputs 31
9. 3 LED status bar 32
10. Analog output 33
11. Operation 34
11. 1 Key assignment 34
11. 2 Measurement value display "Overview" 34
11. 3 Option menu 34
11. 4 Overview of menu displays 35
3
UMG 96-PA www.janitza.de
12. Configuration 36
12. 1 Language 36
12. 2 Communication 36
12. 3 Measurement 37
12. 3. 1 Rated current 37
12. 3. 2 Current and voltage transformers / Nominal current 39
12. 4 System 40
12. 4. 1 Firmware / Serial number 40
12. 4. 2 Time 40
12. 4. 3 Password 40
12. 4. 4 Reset 41
12. 5 Display 43
12. 6 Colors 44
12. 5. 1 Brightness 44
12. 5. 2 Standby 44
12. 5. 3 Brightness (standby) 44
13. Putting the device into service 45
13. 1 Supply voltage 45
13. 2 Measured voltage 45
13. 3 Frequency measurement 46
13. 4 Rotary field direction 46
13. 4. 1 Phasor diagram basics 47
13. 5 Measurement current 48
13. 6 Measurement range violation 49
13. 7 Checking the power measurement 49
13. 8 Checking communication 50
13. 9 Delete min./max. values individually 50
13. 10 Harmonics 51
13. 11 Communication in the bus system 52
13. 11. 1 RS485 52
13. 12 Digital inputs/outputs 53
13. 12. 1 Digital inputs 53
13. 12. 2 Digital outputs 55
13. 13 Analog output 60
13. 14 “Drag indicator” function 61
13. 14. 1 Internal synchronization 61
13. 14. 2 External synchronization 62
13. 14. 3 Synchronization priority 64
13. 15 Recordings 65
13. 16 Tariff switching 66
4
www.janitza.de UMG 96-PA
14. Overview of measurement value displays 68
15. Service and maintenance 74
15. 1 Repairs 74
15. 2 Front film 74
15. 3 Disposal 74
15. 4 Service 74
15. 5 Device calibration 74
15. 6 Re-calibration 74
15. 7 Firmware update 74
15. 8 Battery 75
15. 9 Procedure in the event of an error 76
16. Technical Data 77
16. 1 Function performance characteristics 80
16. 2 Modbus address list of frequently used measurement values: 82
16. 3 Number formats 83
16. 4 Dimension views 84
16. 5 Connection example 1 85
5
UMG 96-PA www.janitza.de

1. General

1. 1 Disclaimer

Compliance with the information products for the devices is a precondition for the safe operation and for achieving the stated perfor­mance and product characteristics. Janitza electronics GmbH does not accept any liability for personal injury, material damage or financial loss incurred as a result of not complying with the information products. Ensure that your information products are kept easily accessible to read.

1. 2 Copyright notice

© 2017 - Janitza electronics GmbH - Lahnau. All rights reserved. Any duplication, editing, distribution or other unauthorized utilization in whole or in part is prohibited. All trademarks and their corresponding rights are the property of the respective owners of those rights.
1. 3 Technical modifications

1. 5 Comments on the Manual

We welcome your comments. If anything in this manual seems unclear to you, please let us know by sending an e-mail to:
info@janitza.de
• Ensure that your device corresponds to the installation guide.
• First, read and understand the documents included with the product.
• Keep the product documents available for the entire life cycle of the product and provide them to any subsequent users.
• Learn about device revisions and the cor­responding updates to the product docu­mentation at www.janitza.de.

1. 4 Declaration of Conformity

The laws, norms and guidelines applied by Janitza electronics GmbH in the production of this product can be found in the declara­tion of conformity at www.janitza.de.
6
www.janitza.de UMG 96-PA
7
UMG 96-PA www.janitza.de

2. Safety

Please read this user manual as well as all other publications that must be referred to for working with this product. This applies especially to installation, operation and main­tenance.
Please follow all safety instructions and warn­ings. Not following the instructions may result in personal injury or damage to the product.
Any unauthorized modifications to or use of this device that exceeds its stated mechan­ical, electric or other operational limits may result in personal injury or damage to the product.
Any such unauthorized modification consti­tutes "misuse" and/or "negligence" in the meaning of the warranty for this product, and therefore invalidates the warranty for the coverage of possible damages occurring as a result.
The User Manual:

2. 1 Safety Instructions

Symbols used:
As an addition to the safety in-
c
C
Safety instructions are marked with a warning triangle and depicted as follows depending on the degree of hazard:
m
HAZARD!
structions, this symbol indicates an electrical hazard.
Together with the word "Instruc­tions," this symbol describes:
• Procedures that do not entail any risk of injury.
• Important information, proce­dures or actions.
Indicates an immediate hazard that results in serious or fatal injury.
• Read before operating the device.
• Keep available for the entire life cycle of the product and for reference.
Comply with the legal and safety guidelines required additionally for the specific instance of application when using the device.
m
WARNING!
m
CAUTION!
Indicates a possibly hazardous situation that may result in seri­ous or fatal injury.
Indicates a possibly hazardous situation that may result in mi­nor injury or material damage.
8
www.janitza.de UMG 96-PA

2. 2 Safety Measures

When electrical devices are in operation, certain parts of these devices are necessar­ily subject to dangerous levels of voltage. Serious bodily harm or material damage may therefore result if they are not handled properly:
c
WARNING!
Serious bodily harm or death may result from hazardous voltages.
For this reason, please note the follow­ing:
Before making connections, ground the device on the protective earth ter­minal, if available.
Hazardous voltages can occur in any of the circuits connected to the volt­age supply.
There may be hazardous voltages in­side the device even after disconnect­ing the voltage supply.
Outfit single-wire leads with ferrules.
Only connect screw terminals with the
same number of poles that are of the same make.
Before starting to work, de-energize the system.
Risk of injury from electric
voltage!
2. 3 Qualified Staff
This device is to be operated and maintained only by qualified staff.
Qualified staff are people who are able to recognize risks and avoid possible hazards that the operation or maintenance of the device may cause, based on their specialized training and their experience.
c
WARNING!
If the device is not operated in accordance with the documentation, safety is no longer ensured and the device may pose a risk.
Risk of Injury from
Improper Use
9
UMG 96-PA www.janitza.de

3. Intended Use

3. 1 Input Check

Proper transport and professional storage, installation and assembly, as well as careful control and maintenance, are all precondi­tions for the proper and safe operation of this device.
Perform unpacking and packing with the usu­al care and without the use of force, and only use suitable tools.
Perform a visual inspection of the devices to ensure a proper mechanical condition.
Check the delivery contents for complete­ness before starting to install the device.
If it is to be assumed that safe operation is no longer possible, the device is to be placed out of order immediately and pro­tected against unintended operation. It is to be assumed that safe operation is no longer possible if the device e.g.:

3. 2 Intended Use

The device is:
• intended to be built into control cabinets and small distribution boards.
• not intended to be built into vehicles! The use of the device in mobile equipment is considered an exceptional environment condition, and is only admissible with a separate agreement.
• not intended to be built into environments with hazardous oils, acids, gases, fumes, dust, radiation, etc.
C
ATTENTION!
All of the screw terminals included in the delivery contents are connected to the device.
• shows visible signs of damage,
• no longer functions despite an intact pow­er supply,
• has been exposed to longer periods of adverse conditions (e.g. storage not con­sistent with the admissible climate limits without an adjustment of the room climate, thawing, etc.) or transport stresses (e.g. fall from a great height even without any outward visible signs of damage, etc.).
C
ATTENTION!
All of the delivered options and design variants are described on the delivery note.
10
www.janitza.de UMG 96-PA

3. 3 Delivery Contents

Number Art. no. Name
1 52.32.xxx
1)
UMG 96-PA
1 33.03.360 Installation Guide
1 33.03.361 "GridVis Software" Quick Start Guide
1 10.01.896 Screw terminal, pluggable, 3-pole (auxiliary energy)
1 10.01.849 Screw terminal, pluggable, 4-pole (voltage measurement)
1 10.01.871 Screw terminal, pluggable, 6-pole (current measurement)
1 10.01.909 Screw terminal, pluggable, 3-pole (RS 485)
1 10.01.865
Screw terminal, pluggable, 10-pole (digital inputs/outputs, analog output)
1 52.22.251 Mounting Kit
1)
See delivery note for item number

3. 4 Deliverable accessories

Item no. Name
21.01.058
Lithium battery CR2032, 3 V (Authorization as per UL 1642)
29.01.065 Seal, 96 x 96
15.06.015 Interface converter RS485 <-> RS232
15.06.025 Interface converter RS485 <-> USB
11
UMG 96-PA www.janitza.de

4. Product Description

The device is intended for:
• the measurement and calculation of elec­trical quantities such as voltage, current, power, energy, harmonics in the building installation, on distributors, circuit breakers and busbar distributors.
• the measurement of measuring voltag­es and currents originating in the same network.
• measurements in low-voltage networks in which rating voltages of up to 600 V conductor to ground and surges of over­voltage category III may occur.
• Measurement in medium and high-voltage networks usually does not occur through current or voltage transformers.
• being built into non-mobile control cab­inets or small distribution boards. The position at which it is built in is arbitrary.
• measurement in medium- and high-voltage networks with current and voltage trans­formers.
• use in domestic and industrial settings.
• the measurement of current via external ../1 A or ../5 A current transformers.
Directly on the device using the 6 keys and the display.
• Via the programming software GridVis®.
• You can change and open data with the help of the Modbus address list via the Modbus protocol. You can find this list at www.janitza.de.
This user manual only describes the 6-key operation of the device. The network analysis software GridVis® has its own "Online Help Guide."

4. 3 Network Analysis Software GridVis®

You can program the device and read out the data with the network analysis software Grid­Vis® available at www.janitza.de. To do this, a PC must be connected to the device e.g. via a serial interface (RS485) or via a gateway connection.
With the network analysis software GridVis®, you can:
The measurement results can be displayed and read out via the device interface for fur­ther processing.

4. 1 Measurement Procedure

The device:
• continuously measures and calculates all effective values over a 200 ms interval.
• measures the real effective value (TRMS) of the voltages and currents connected to the measurement inputs.

4. 2 Operating Design

You can program the device in many ways and/or recall measurement values.
• program the device.
• configure and read out recordings.
• analyze the data you read out.
• save data in the database.
• display measurement values graphically.
• program client-specific applications.
Malfunction can result
c
CAUTION!
If the device is connected improperly, it may deliver measurement values with errors.
For this reason, please note the follow­ing:
Measurement voltages and measure­ment currents must come from the same network.
Do not use the device to measure direct current.
Earth the conducting patch boards.
from an improper
connection
12
www.janitza.de UMG 96-PA

4. 4 Performance characteristics

General
• Front board installation device measuring 96 x 96 mm
• Expansion by modules
• Connection via plug-in terminals with screw connections
• Color graphics display 320 x 240 px
• Control via 6 keys
• 3 voltage measurement inputs (600 V, CAT III)
• 3 current measurement inputs (via current transformer)
• 3 digital outputs
• 3 digital inputs (configured as impulse counters with simultaneous power calculation
• 1 analog output (0 - 20 mA)
• Data storage 4 MByte Flash
• RS485 interface (Modbus RTU, Slave, up to 115 kbps)
• Recording of more than 2,000 measurement values
• Clock and battery
• Operating temperature range -10 °C .. +55 °C
Measurement uncertainty
• Active energy, measurement uncertainty class 0.5S for ../5A transformers
• Active energy, measurement uncertainty class 1 for ../1A transformers
• Reactive energy class 1
Measurement
• Recording of more than 800 measurement values
• Measurement in TN and TT networks
• Measurement in networks with rated voltages of up to L-L 720 Vrms and L-N 417 Vrms (as per IEC)
• Measurement range for current 0.005 .. 6 Arms
• Real active value measurement (TRMS)
• continuous sampling of voltage and current measurement inputs
• Frequency range of the fundamental oscillation 45 Hz .. 65 Hz
• Measurement of the harmonics 1 to 25 for ULN and I
• ULN, ULL, I, P (drawn/delivered), Q (ind./cap.)
• 2 tariffs (switching via Modbus or digital input 1)
13
UMG 96-PA www.janitza.de

4. 5 Product Overview

1
2
Fig. Front view of UMG 96-PA
3 4
5
8 7
6
1 Device Type 2 Description of the function keys 3 Key 1: Configuration menu, Back (ESC) 4 Key 2: Select digit, set option field () 5 Key 3: Lower digit by 1, select menu item (), set option field () 6 Key 4: Increase digit by 1, select menu item (), set option field () 7 Key 5: Select digit, set option field () 8 Key 6: Open option menu, activate entry, confirm selection (Enter)
14
www.janitza.de UMG 96-PA
1
8
Fig. Back view of UMG 96-PA
2
3 4 5
6
7
1 Supply voltage 2 RS485 interface 3 Digital inputs 4 Digital outputs 5 Analog output 6 Module connector 7 Current measurement inputs I1 to I3 8 Voltage measurement inputs V1 to V3
15
UMG 96-PA www.janitza.de

5. Installation

5. 1 Installation site

The device is suitable for installation in non-mobile and weather-protected control panels in indoor environments.
Plan on an earth for conducting control pan­els.
Material damage may
m
CAUTION!
result if the installation in-
structions are not followed!
Not following the installation instructions may damage or destroy your device.
Comply with the details on the installation position in sections „Installation“ and „Technical Data“.

5. 2 Installation position

The cut-out dimensions of the control panel are 92
+0,8
mm x 92
+0,8
mm.
To achieve sufficient ventilation, take the following measures:
• Install the device vertically.
• Maintain a minimum clearance of 50 mm above and below.
• Maintain a minimum clearance of 20 mm on each side.

5. 3 Securing the device

The device is secured in the control panel with the side mounting brackets. The brack­ets are to be removed e.g. using a screw­driver as a horizontal level before the device is put in.
Fig. Side view of UMG 96-PA with mounting brackets. The brackets can be loosened by leveraging them hori­zontally with a screwdriver.
Then slide in and snap the brackets into place, and finally screw in the screws to se­cure the device.
• Turn the clamping screws until they just touch the assembly plate.
• Then tighten the clamping screws with two more rotations each (the mounting brack­ets may be destroyed if the screws are tightened too much).
16
Fig. Back view of the installation posi-
tion of the UMG 96-PA
Assembly plate
Mounting brackets
Clamping screw
Once the clamping screws touch the surface of the assembly plate: use two further rotations at most to secure the device
Screwdriver
www.janitza.de UMG 96-PA
L1
E
L1 L2
L3 E
N
R
L1
L2 L3 EE
L1
L2 L3 EE

6. Network systems

Network systems and maximum rated voltages as per DIN EN 61010-1/A1:
Three-phase four-wire systems
with earthed neutral conductor
L2
N
L3
IEC U
UL U
L-N
L-N
/ U
417 VLN / 720 VLL
L-L:
/ U
347 VLN / 600 VLL
L-L:
EE
The device can be used in
• TN and TT networks
• domestic and industrial areas.
c
WARNING!
Risk of injury from
electric voltage!
If the device is exposed to measurement voltage surges above the admissible over­voltage category, safety-related insulations in the device may be damaged, whereupon the safety of the product can no longer be guaranteed.
Only use the device in environments in which the admissible measurement surge voltage is not exceeded.
17
UMG 96-PA www.janitza.de

6. 1 Voltage measurement

You can use the UMG 96-PA to measure volt­age in TN and TT systems. Voltage measurement in the UMG 96-PA is designed for the overvoltage category 600V CATIII (measurement surge voltage 6 kV).
347V/600V 50/60Hz
L1
L2
L3
N
PE
L1
240V 50/60Hz
N
In systems without N, the measurement val­ues that require an N are based on a calcu­lated N.
c
WARNING!
Risk of injury from electric
voltage!
If voltage transformers are used, the con­nections on the secondary side may not be short-circuited!
For this reason, please note the follow­ing:
Check that the voltage transformers have been prop­erly installed. To do this, read the re­spective information about the voltage transformers.
V1 V3V2 VN
AC/DC
4M
4M
4M
Voltage measurement
UMG 96-PA
Fig. Principle circuit diagram - Measurement in three-
phase 4-wire systems.
4M
Auxiliary energy
DC
18
www.janitza.de UMG 96-PA

6. 2 Rated voltages

The following figures show lists of the net­works and the corresponding network rated voltages at which the device can be used.

6. 2. 1 Three-phase 4-wire network with earthed neutral conductor

U
/ U
L-N
L-L
66V / 115V 120V / 208V 127V / 220V 220V / 380V 230V / 400V 240V / 415V 260V / 440V 277V / 480V 347V / 600V 400V / 690V 417V / 720V
Maximum network voltage of the network as per UL
Maximum network voltage of the network
Fig. Network rated voltages as per EN 60664-1:2003
suited for measurement inputs
19
UMG 96-PA www.janitza.de

7. Installation

7. 1 Connection to a PC

There are various options for connecting the device to a PC:
1. Connection via an interface converter:
PC with GridVis®
UMG 96-PA
RS232
RS232
RS485
RS485
PC with GridVis®
UMG 96-PA
USB
USB
RS485
RS485
2. Use of the UMG 96-PA (slave) via a UMG (master) with gateway functionality (e.g. UMG 512)
PC with GridVis®
Ethernet
Ethernet
20
UMG 512-PRO
as gateway
UMG 96-PA
Slave 1
ModbusModbus Modbus
UMG 96-PA
Slave 2
UMG 103
Slave 3
www.janitza.de UMG 96-PA

7. 2 Circuit breaker

For building installation, plan for a suitable circuit breaker for the supply voltage to de-energise the device.
• The circuit breaker must be installed near the device and easily accessible to the user.
• The switch must be marked as the circuit breaker for this device.

7. 3 Supply voltage

The device needs a supply voltage to oper­ate. The type and amount of supply voltage required for your device can be found on the ratings plate.
The supply voltage is connected on the rear of the device via terminal blocks.
Before connecting the supply voltage, ensure that the voltage and frequency correspond to the details on the ratings plate.
Connect the supply voltage through a UL/IEC approved fuse.
After the supply voltage has been connect­ed, an indicator will appear on the display. If no indicator appears, check that the supply voltage is within the rated voltage range.
m
CAUTION!
result from failure to com-
ply with the connection
requirements.
Noncompliance with the connection re­quirements may result in your device being damaged or destroyed.
For this reason, please note the following:
Comply with the details on voltage and frequency on the ratings plate.
Connect the supply voltage through a fuse in accordance with the technical specifications.
Do not source the supply voltage on the voltage transformers.
Plan for a fuse for the neutral conduc­tor if the neutral conductor connection to the source is not earthed.
L
N
PE/FE
1.)
Material damage may
1.)
2.)
Fuse
(UL/IEC listed)
2.)
Disconnector
(Circuit breaker)
Functional earth
c
WARNING!
Risk of injury from electric
voltage!
Serious bodily harm or death may result from:
• Touching bare or stripped wires that are live.
• Inputs on the device that are dangerous to touch.
For this reason, please note the following:
The inputs for supply voltage are dangerous to touch.
De-energise your system before starting work!
Fig. Connection example for supply voltage
ATTENTION!
C
If the functional earth is not connected, the device displays a non-applied residual voltage.
21
UMG 96-PA www.janitza.de
Overcurrent protective device for the line protection of the supply voltage
Recommendation for the overcurrent pro­tective device for the line protection of the supply voltage, depending on the variants:
• Option 230 V: 6 - 16 A (Char. B)
• Option 24 V: 1 - 6 A (Char. B)
C
Recommendation for the maximum number of devices on one miniature circuit breaker, depending on the variants:
• Option 230 V:
• Option 24 V:
ATTENTION!
The circuit breaker serves only as line protection - it does not provide protection for the device!
For one miniature circuit breaker B6A, maximum of 4 devices. For one miniature circuit breaker B16A, maximum of 11 devices.
For one miniature circuit breaker B6A, maximum of 3 devices. For one miniature circuit breaker B16A, maximum of 9 devices.

7. 4 Measured voltage

The device has 3 voltage measurement in­puts (V1 to V3) on the back of the device.

7. 4. 1 Overvoltage

The voltage measurement inputs are suit­able for measurement in networks in which overvoltage of category 600 V CAT III (voltage measurement surge of 6 kV) may occur.

7. 4. 2 Frequency

The device:
• requires the network frequency to measure and calculate the measurement values.
• is suitable for measurement in networks in which the fundamental oscillation of the voltage is within the range of 45 to 65 Hz.
The network frequency is derived from the measured voltage in phase L1. The sampling rate of the voltage and current measurement inputs is derived from the network frequency.
If the measurements have severely distort­ed voltages, it will no longer be possible to precisely determine the frequency of the fundamental oscillation of the voltage. I.e. the corresponding network frequency should be stipulated for measured voltages that exhibit severe distortions. (Voltage distortions occur e.g. in measure­ments of consumers operated with a phase angle control). Current distortions do not affect the determination of the frequency.
22
More information can be found in Section 12 Configuration / Rated frequency"
www.janitza.de UMG 96-PA
1.)
Fuse
1) 1) 1)
2) 2) 2)
Fig. Connection example for voltage measurement.
(UL/IEC listed)
2.)
Disconnector
(Circuit breaker)
L1
L2
L3
N
c
WARNING!
Risk of injury from electric
voltage!
Serious bodily harm or death may result from not complying with the connection requirements for the voltage measurement inputs.
For this reason, please note the follow­ing:
De-energise your system before starting work! Check that it has been de-energised!
Connect voltages that are over the admissible network rated voltages through voltage transformers.
The voltage measurement inputs on the device are dangerous to touch!
Install a circuit breaker as described in Section „7. 2 Trennschalter“.
Use a UL/IEC approved overvoltage protective device with a rating value that is measured for short-circuit cur­rent at the point of connection.
c
CAUTION!
Malfunction can result from
an improper connection
If the device is connected improperly, it may deliver measurement values with errors.
For this reason, please note the follow­ing:
Measurement voltages and currents must come from the same network.
The device is not suitable for measur­ing direct current.
ATTENTION!
C
The device can only determine mea­surement values if there is a voltage L1-N greater than 20 Veff (4-wire measurement) or a voltage L1-L2 greater than 34 Veff (3-wire measure­ment) at the voltage measurement input V1.
As an overvoltage protective device for the voltage measurement, use a line protection (1 - 10 A) with IEC/UL approval.
23
UMG 96-PA www.janitza.de

7. 4. 3 Connection variants for voltage measurement

V1
V2
Fig. Voltage measurement in the three-phase
four-wire system
V1
V2
V3
V3
VN
VN

7. 5 Current measurement

The device:
• is designed to be connected to current transformers with secondary currents of ../1 A and ../5 A.
• is only approved for measuring current through current transformers.
• does not measure direct currents.
The factory preset current transformer ratio is 5/5 A, and may need to be adjusted to the current transformer in use.
Fig. Voltage measurement in the three-phase
four-wire system
V1
V2
Fig. Voltage measurement in the single-phase
three-wire system
V3
VN
L1
L2
L3
N
Fig. Connection example "Current measurement through
a current transformer."
S1 S2
S1 S2
S1 S2
24
www.janitza.de UMG 96-PA
7. 5. 1 Current measurement connection
m
WARNING!
tric voltage on the current
transformers!
Voltage peaks that are extremely dangerous to touch may occur on current transformers operated with an open secondary side that may result in serious bodily harm or death.
For this reason, please note the following:
De-energise your system before start­ing work! Check that it has been de-en­ergised!
Avoid open operation of the current transformers.
Short-circuit uncharged current trans­formers.
The secondary terminals of the current transformer must be short-circuited before the power supply lines are dis­connected.
If a test switch that automatically short-circuits the current transformer secondary leads is available, it is suffi­cient to put this into the "test" position provided the short-circuiters have been checked beforehand.
Only use current transformers that have a base insulation as per IEC 61010-1:2010.
Ensure that the attached screw termi­nal is sufficiently secured to the device with the two screws.
Even "open-safe" current transform­ers are dangerous to touch if they are operated openly.
Risk of injury from elec-
variants
Fig. Current measurement via current transfor­mer in three-phase four-conductor system
Fig. Current measurement in single-phase three-conduc­tor system
Fig. Current measurement via 2 current transformers in three-phase four-conductor system
m
WARNING!
Risk of injury from electric
voltage!
At high measurement currents, temperatures of up to 80 °C can develop at the terminals.
Therefore, use leads that are designed to have an operating temperature of at least 80 °C.
25
UMG 96-PA www.janitza.de

7. 5. 2 Direction of the current

The current direction can be individually cor­rected on the device or via the serial interfac­es for each phase.
In the case of incorrect connection, the cur­rent transformer does not need to be subse­quently reconnected.

7. 5. 3 Total current measurement

If the current measurement takes place via two current transformers, the total transform­er ratio of the current transformer must be programmed on the device. Setting the cur­rent transformer ratios is described in Section „Messwandler“.
Example:
The current measurement is performed via two current transformers. Both current trans­formers have a transformer ratio of 1000/5 A. The total measurement is then carried out with a total current transformer 5+5/5 A .

7. 5. 4 Ammeter

If you want to measure the current not only with the UMG but also with the ammeter, the ammeter must be connected in series to the UMG.
UMG
I
S2
1
S
A
Einspeisung Supply
(k)S
1 S2(l)
2(L)(K)P1
P
Fig. Circuit diagram with additional ammeter connected
in series
Verbraucher
Consumer
The device must then be set as follows:
Primary current: 1000 A + 1000 A = 2000 A Secondary current: 5 A
UMG
I
S
S2
1
Einspeisung 1 Supply 1
1P1
1S1
(K)
(k) (l)
(L)
1S
1P2
Verbraucher A Consumer A
2
P1
1S1 1S2 2S1 2S2
Fig. Example of voltage measurement via a total current
transformer
P2
Einspeisung 2
Supply 2
2S1 (k)
(l)
2S2
Verbraucher B
Consumer B
2P
(K) (L)
2P2
1
26
www.janitza.de UMG 96-PA

8. Interface

The RS485 interface in this device is a 3-pole plug contact that communicates via the Mod­bus RTU protocol.
Cable types:
• Recommendation: Unitronic Li2YCY(TP) 2x2x0.22 (Lapp cable)
Connection options for the connectors:
• 0.2 - 1.5 mm (see Section "Technical Data")
e.g.
Connection
to the
UMG 604
as gateway
(master)
A
B
Data GND
2
RS485 bus
A
B
A
B
e.g. Connection of further slave devices
C
CAT cables are not suitable for bus wiring. Use the cable types recom­mended for this purpose.

8. 1 Screening

Use a drilled and shielded cable for the con­nections via the interface, and observe the following points for the shielding:
• Earth the screens of all cables that lead to the cabinet where they enter the cabinet.
• Connect the screen over a larger area and in a manner that will conduct well, to a low-noise earth.
• Do NOT connect the screen to terminal C (GND)
• Gather the cables mechanically above the earthing clamp in order to avoid damage due to cable movement.
• Use suitable cable glands to feed the ca­bles into the cabinet - for example armored conduit couplings.
ATTENTION!
Fig. RS485 interface, 3-pole plug contact
RS485 bus A
B
Data GND
Fig. RS485 interface, 3-pole plug contact with termina-
tion resistor (item no. 52.00.008)
120 Ω
A
B
Cable Strain relief
Screen braid of the cable
Earthing clamp
Low-noise earth
Fig. Screening procedure at cabinet entry.
27
UMG 96-PA www.janitza.de
m
CAUTION!
Atmospheric discharge may cause transmis­sion errors and hazardous voltages on the device.
For this reason, please note the following:
Place the screening on the functional earth (PE) at least once.
For larger sources of interference, put the frequency inverter in the control cabinet and position the screening as close as possible on the device to the functional earth (PE).
Maintain a maximum cable length of 12000 m at a baud rate of 38.4 k.
Use screened cables.
Position interface leads so they are
spatially separated or additionally in­sulated from the components carrying network voltage.
Transmission errors and risk of injury result from
electrical interference.

8. 2 Termination resistors

The cable is terminated with resistors (120 Ohm, 1/4 W) at the beginning and at the end of a segment.
The device does not contain any termination resistors.
Correct
Incorrect
Terminal strip in the cabinet.
Device with RS485 interface. (without termination resistor)
28
Device with RS485 interface. (with termination resistor on the device)
www.janitza.de UMG 96-PA

8. 3 Bus structure

• All devices are connected in a bus struc­ture (line).
• Each device has its own address within the bus (see also Parameter programming).
• Up to 32 subscribers can be connected together in a single segment.
• The cable is terminated with resistors (bus termination 120 Ohm, 1/4 W) at the begin­ning and at the end of a segment.
• With more that 32 subscribers, repeaters (amplifiers) must be used to connect the individual segments.
• Devices for which the bus connection is switched on must be under current.
Master
Power supply necessary
Bus terminator on
T
T
Slave
• It is recommended that the master be placed at the end of a segment.
• If the master is replaced with a bus con­nection, the bus must be switched off.
• Replacing a slave with a bus connection that is either switched on or de-energised can destabilise the bus.
• Devices that are not connected to the bus can be replaced without destabilising the bus.
T
Slave Slave Repeater
T
Slave
Fig. Bus structure
Slave
T
Slave Slave
29
UMG 96-PA www.janitza.de

9. Digital inputs and outputs

The device has
• 3 digital inputs and
• 3 digital outputs.
External
17
18
Digital
Input 1
19
Digital
Input 2
auxiliary voltage
-
S1
S2

9. 1 Digital inputs

The UMG96-PA has three digital inputs to each of which you can connect one signal transducer. When a signal is present, the corresponding LED lights up green.
An input signal is detected on a digital input if
• a voltage of at least 18 V and maximum 28 V DC (typically at 4 mA) is applied.
• a current of at least 1 mA and maximum 6 mA flows.
UMG 96-PA
Digital inputs 1-3
LED
LEDLED
2k21
2k21
2k21
2k21
24V DC
+
C
Note the correct polarity of the sup­ply voltage.
ATTENTION!
+
-
Fig. Connection of digital inputs
Transmission errors
c
CAUTION!
and material damage
result from electrical
interference.
2k21
2k21
Fig. Example for the connection of external switch
contacts S1-S3 to digital inputs 1, 2 and 3.
20
Digital
Input 3
S3
30
For wiring over 30 m, there is an increased probability of transmission errors and damage to the device from atmospheric discharge.
Use screened wiring for connections to the digital inputs and outputs.
www.janitza.de UMG 96-PA

9. 1. 1 S0 pulse input

You can connect an S0 pulse transducer per DIN EN62053-31 to any digital input.
This requires an auxiliary voltage with an output voltage in the range 18 .. 28V DC and a resistor of 1.5 kOhm.

9. 2 Digital outputs

External
auxiliary voltage
-
UMG 96-PA
Digital inputs 1-3
LED
LEDLED
Fig. Example for the connection of an S0 pulse trans­ducer to digital input 1.
2k21
2k21
2k21
2k21
2k21
2k21
17
18
Digital
Input 1
19
Digital
Input 2
20
Digital
Input 3
1.5k
S0 pulse
transducer
24V DC
+
The device has three digital outputs which:
• are galvanically separated from the analy­sis electronics using optocouplers.
• have a joint reference.
• are not short-circuit proof.
• require an external auxiliary voltage.
• can be used as impulse outputs.
• can switch between AC and DC loads.
• can be controlled via Modbus.
• can display the results of comparators.
c
CAUTION!
Material damage from
connection errors
The digital outputs are not short-circuit proof! Connection errors may therefor result in damage to the connections.
Pay attention to the correct wiring when connecting to the outputs.
C
Functions for the digital outputs can be adjusted clearly in the GridVis® software (see www.janitza.de). Use of the GridVis® software requires a connection between the device and the PC via an interface.
ATTENTION!
m
CAUTION!
Measurement errors in use
as pulse output
When using the digital outputs as a pulse output, measurement errors may result from residual ripple.
So use a power adapter whose residual ripple is less than 5% for the supply volt­age (DC) to the digital inputs and outputs.
31
UMG 96-PA www.janitza.de
~
Fig. Connection of digital/pulse outputs
UMG 96-PA
21
LED
Digital Ouput 1
22
External
auxiliary voltage
24V DC
+
DC
K1

9. 3 LED status bar

The different statuses of the inputs and out­puts are displayed via the LED status bar on the back of the device.
Digital inputs
The LED assigned to a respective input lights up green when a signal of at least 1 mA flows on this interface.
Digital outputs
The LED assigned to a respective output lights up red when the output is set as enabled - regardless of whether there is a continuing connection to this interface.
-
LED status bar
Digital inputs
LED status bar Digital outputs
LEDLED
Digital Ouput 2
Digital Ouput 3
23
24
Fig. Example for the connection of two relays to the digital outputs
DC
K2
32
www.janitza.de UMG 96-PA

10. Analog output

The device has 1 passive analog output, which can emit current of 0 - 20 mA. An external power adapter (24 V DC) is required for operation.
The connected load impedance may not exceed a resistance of 300 Ohm. If the analog output is connected to great­er resistance, the output range (20 mA) is limited.
The measurement value, initial and final val­ues, and the output range 4 - 20 mA or 0 - 20 mA assigned to the analog output can be set using the GridVis® software (cf. Section 13 "Analog Output")
-
=
+
External
operating voltage
UMG 96-PA
Analog output 1
230 V AC
26
25
+
+
24 V DC
-
Analog inputs
Fig. Analog output connection
33
UMG 96-PA www.janitza.de

11. Operation

The device is operated using six function keys that are assigned to different functions, depending on the context:
• Selection of measurement value displays.
• Navigation in the menu.
• Control of device settings.
Display title
Measurement values
Labeling of the function keys
Function Keys
Fig. UMG 96-PA Measurement value display "Over­view"

11. 2 Measurement value display "Overview"

After returning to the network, the device starts with the measurement value display "Overview."
This measurement value display includes the device name and an overview of the import­ant measurement values. In its delivery state, the device name consists of the device type and the serial number of the device.
Press key 1 (Esc) repeatedly to open the measurement value display "Overview."

11. 1 Key assignment

Key
Function
• Option menu
• Leave menu, Back (Esc)
• Press multiple times: Back to the measurement val­ue display "Overview"
• Select digit
• Set option field ()
• Change (Digit -1)
• Set option field ()
• Select menu item ()
• Change (Digit +1)
• Set option field ()
• Select menu item ()
• Select digit
• Set option field ()
Fig. Measurement value display "Overview"

11. 3 Option menu

If you are in the measurement value display "Overview," open the main menu with key 1 (Esc).
Fig. UMG 96-PA Main menu
34
• Open option menu
• Activate entry
• Confirm selection (Enter)
Key 1 (Esc): Main menu
www.janitza.de UMG 96-PA

11. 4 Overview of menu displays

Main menu
Summary (start screen)
Voltage
Voltage L-N
Voltage L-L
Linewriter
Current
Current
THD-I
Linewriter
Power
Sum power
Active power
Reactive power
Apparent power
Linewriter active power
Linewriter reactive power
Linewriter apparent power
Energy
Active energy
Reactive energy
Apparent energy
Tariff
Consumption overview
Active energy/month
Reactive energy/month
Apparent energy/month
Active energy/day
Reactive energy/day
Apparent energy/day
Harmonics
Voltage L1
Voltage L2
Voltage L3
Current L1
Current L2
Current L3
Oscilloscope
Voltage L1
Voltage L2
Voltage L3
Voltage L1-3
Current L1
Current L2
Current L3
Current L1-3
Phasor diagram
System
Configuration
Language
Communication
Field bus: Device address
Field bus: Baud rate
Field bus: Framing
Measurement
Measurement
Current transformer
Voltage transformer
Nominal current
Nominal frequency
System
Version
Serial no.
Time
Password
Reset
Display
Colors
Overview COM ports
Comparator 1
Comparator 2
Select menu:
• Select the menu item with the keys 3 () and 4 ().
• Confirm this by pressing key 6 (Enter).
• You can exit the selection by pressing key 1 (Esc).
35
UMG 96-PA www.janitza.de
12. Configuration
The device must be connected to a power supply to configure it. To do so, proceed as described in „13. 1 Supply voltage“.
• If you are not in the measurement value display "Overview," switch to this view by repeatedly pressing key 1 (Esc).
• Open the main menu with key 1 (Esc). With keys 3 () and 4 (), select "Sys- tem" and confirm your selection with key 6 (Enter).
• Select "Configuration" and confirm by pressing key 6 (Enter).
Fig. Main menu, Selection "System"
• With keys 3 () and 4 (), select the lan­guage you want ("German," "English").
• Confirm your selection by pressing key 6 (Enter).
• Exit the configuration with key 1 (Esc).
• Open the main menu with key 1 (Esc).
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).

12. 2 Communication

Set the parameters for the RS485 interface of your device in the configuration menu.

12. 1 Language

In the configuration menu, you can set the language for the measurement value displays and menus in "Language."
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the item "Language."
• Confirm by pressing key 6 (Enter).
Fig. Configuration menu, Selection "Language"
Fig. Configuration menu, Selection "Communication"
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "Communication."
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the pa­rameter you want
- Device address,
- Baud rate
- Data framework.
• Confirm your selection by pressing key 6 (Enter).
• Change the parameter sizes with keys 3 () and 4 (). Device address: Set the digit position with keys 2 () and 5 ().
• Confirm by pressing key 6 (Enter).
• Exit the menu "Communication" with key 1 (Esc).
36
www.janitza.de UMG 96-PA
• Exit the configuration menu with key 1 (Esc).
• Open the main menu with key 1 (Esc).
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).
Fig. Menu Communication, Selection "Device address"
Settings:
• Device address: Select a device address for the device that the device will be addressed with in the bus. This address must be unique within the bus structure. Settings range: 1 - 250 Default: 1
• Baud rate: Select a uniform baud rate for all of the devices within the bus structure. Settings range: Auto, 9600, 19200, 38400, 57600, 115200 kbps Default: Auto
• Data framework: Select a uniform data framework for all of the devices within the bus structure. Settings range:
• "odd" (parity odd with 1 stop bit)
• "even" (parity even with 1 stop bit)
• "1 stop bit" (parity none with 1 stop bit)
• "2 stop bits" (parity none with 2 stop bits)
• Default: 1 stop bit (no parity)
Material damage may result
m
CAUTION!
Incorrect network settings may cause inter­ruptions in the IT network.
Find out about the correct network set­tings for your device from your network administrator.

12. 3 Measurement

In the menu "Measurement," you can set the ratio of the current and voltage transformers (primary-to-secondary side), the rated current and the rated frequency.

12. 3. 1 Rated current

The network frequency is required to mea­sure and calculate the measurement values. The device is suitable for measurements in networks that have a frequency range of 45 ­65 Hz.
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "Measurement" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the item "Rated frequency" and confirm your selec­tion with key 6 (Enter).
from incorrect network
settings.
Fig. Menu Measurement, Selection Rated frequency
37
UMG 96-PA www.janitza.de
• With keys 3 () and 4 () select the setting you want, and confirm the selec­tion with key 6 (Enter). You can abort this action by pressing key 1 (Esc).
• Exit the menu "Measurement" with key 1 (Esc).
• Exit the configuration menu with key 1 (Esc).
• Open the main menu with key 1 (Esc).
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).
Settings range:
• Auto (45-65 Hz)
• 60 Hz (fixed frequency)
• 50 Hz (fixed frequency)
Default
• Auto (45-65 Hz)
C
C
ATTENTION!
If the network frequency is outside of the range 45-65 Hz
• no error or warning notification will be sent.
• the corresponding setting will be used for the entry of a constant frequency (50 / 60 Hz).
• The last determined frequency in the range 45-65 Hz is used for the selection of the automatic frequency detection (auto).
ATTENTION!
It takes 10 seconds to determine the frequency. The frequency does not represent a 200 ms measure­ment value!
C
The device can only determine the network frequency if there is a voltage L1-N great­er than 20 Veff (4-wire measurement) or a voltage L1-L2 greater than 34 Veff (3-wire measurement) at the voltage measurement input V1.
ATTENTION!
Devices that use automatic fre­quency frequency detection will require about 5 seconds to determine the network frequency. During this time, the measurement values do not adhere to the guaranteed measure­ment uncertainty.
38
www.janitza.de UMG 96-PA

12. 3. 2 Current and voltage transformers / Nominal current

For the defined operation of the device it is necessary to set the correct current and voltage transformer ratios and the nominal current.
When connecting voltage transformers, observe the measured voltage on the ratings plate!
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "Measurement" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the sub­menu "Rated Measurement" and confirm your selection with key 6 (Enter).
• With keys 2 to 5 () and 4 () select the setting you want, and confirm the selection with key 6 (Enter).
• Set the digit position with keys 2 () and 5 ().
• With keys 3 () and 4 () change the digit (-1 / +1).
• Confirm by pressing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• Exit the view "Measurement L1..L3" with key 1 (Esc).
• Exit the view "Measurement" with key 1 (Esc).
• Exit the configuration menu with key 1 (Esc).
• Open the main menu with key 1 (Esc).
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).
C
Settings:
• Current transformer (primary):
• Current transformer (secondary):
• Voltage transformer (primary):
• Voltage transformer (secondary):
• Rated current:
ATTENTION!
The adjustable value of 0 for the primary current transformer doesn't make any sense and may not be used.
Settings range: 0 - 32000 A Default: 5 A
Settings range: 0 - 5 A Default: 5 A
Settings range: 0 - 32000 V Default: 400 V
Settings range: 0 - 999 V Default: 400 V
Settings range: 0 - 999999 A Default: 150 A
Fig. View "Measurement L1..L3," Configuration of the current and voltage transformer ratios and the rated voltage
39
UMG 96-PA www.janitza.de

12. 4 System

Display of device-specific system settings, password assignment and value reset func­tion.
1 2 3 4 5
Fig. View "System"
1 Firmware version 2 Device serial number 3 Device time / date 4 Password function 5 Reset settings
View of the display "System":
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "System" from the configuration menu.
• Confirm by pressing key 6 (Enter).

12. 4. 1 Firmware / Serial number

Use the firmware version and the device-spe­cific serial number for possible support requests and to register on the homepage (www.janitza.de).

12. 4. 2 Time

Shows the current device time. You can change the settings for Time Synchronization, Date/Time Zones and Clock Time with the GridVis® software or via the corresponding Modbus addresses.

12. 4. 3 Password

You can use a password to block access to the configuration. Then it is only possible to change the configuration directly on the device after entering the password.
The password consists of a 5-digit combina­tion of numbers.
Settings range:
• 1-99999 = with password
• 0 = without password
Default:
• 0 = without password
No password (0) is programmed as the facto­ry preset default setting.
40
Fig. View "Configuration," Selection "System"
You must know the current password to change a previously created password.
Remember and note the password.
Set password:
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "System" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the submenu "Password" and confirm your selection with key 6 (Enter).
www.janitza.de UMG 96-PA

12. 4. 4 Reset

This area makes it possible to delete and reset measurement values and device param­eters.
Energy
You can delete all of the energy counters in the device at once. It is not possible to select
Fig. View "System," Selection "Password"
• With the keys 2 to 5 (), set the password you would like to use.Use keys 2 () and 5 () to set the digit position. With keys 3 () and 4 () change the digit (-1 / +1).
• Confirm by pressing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• Exit the view "System" with key 1 (Esc).
• Exit the configuration menu with key 1 (Esc).
• Open the main menu with key 1 (Esc).
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).
certain energy counters.
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "System" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the sub­menu "Reset" and confirm your selection with key 6 (Enter).
• With keys 3 () and 4 (), select the func­tion you want.
C
ATTENTION!
Remember and note the password. No device settings can be changed if you do not know the password. If you lose the password, contact support!
Fig. View "Reset," Reset the energy counters
• Confirm the details by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the op­tion you want "Yes," or "No."
• Confirm by pressing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• Exit the view "Reset" with key 1 (Esc).
• Exit the view "System" with key 1 (Esc).
• Exit the configuration menu with key 1 (Esc).
• Open the main menu with key 1 (Esc).
41
UMG 96-PA www.janitza.de
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).
Min./Max. values
You can delete all of the minimum and max­imum values in the device at once. It is not possible to select certain energy counters.
ATTENTION!
C
Delete any production-related energy counter contents, minimum and max­imum values, and recordings prior to putting the device into service.
• Exit the configuration menu with key 1 (Esc).
• Open the main menu with key 1 (Esc).
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).
Factory settings
In "Factory settings" you can reset all set­tings, as well as the configuration and the recorded data back to the factory presets.
Fig. Menu "Reset," Delete min./max. values
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "System" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the sub­menu "Reset" and confirm your selection with key 6 (Enter).
• With keys 3 () and 4 (), select the func­tion "Min./Max. Values" and confirm your selection with key 6 (Enter).
• With keys 3 () and 4 (), select the op­tion you want "Yes," or "No."
• Confirm by pressing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• Exit the view "Reset" with key 1 (Esc).
• Exit the view "System" with key 1 (Esc).
Fig. Menu "Reset," Load factory settings
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "System" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the sub­menu "Reset" and confirm your selection with key 6 (Enter).
• With keys 3 () and 4 (), select the func­tion "Factory settings" and confirm your selection with key 6 (Enter).
• With keys 3 () and 4 (), select the op­tion you want "Yes," or "No."
• Confirm by pressing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• Confirm the warning notification by press­ing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• The device restarts.
42
www.janitza.de UMG 96-PA
Restart
To manually restart the device, proceed as follows:
Fig. Menu "Reset," Restart device
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "System" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the sub­menu "Reset" and confirm your selection with key 6 (Enter).
• With keys 3 () and 4 (), select the func­tion "Restart" and confirm your selection with key 6 (Enter).
• With keys 3 () and 4 (), select the op­tion you want "Yes," or "No."
• Confirm by pressing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• The device restarts.

12. 5 Display

Use this menu item to adjust the display set­tings on the device:
• Brightness
• Standby after
• Brightness (standby)
Fig. Menu "Display,"
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "Display" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the sub­menu you want and confirm your selection with key 6 (Enter).
• Use keys 2 to 5 () to set the dis­play value. Set the digit position with keys 2 () and 5 (). With keys 3 () and 4 () change the digit (-1 / +1).
• Confirm by pressing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• Exit the menu "Display" with key 1 (Esc).
• Exit the configuration menu with key 1 (Esc).
• Open the main menu with key 1 (Esc).
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).
43
UMG 96-PA www.janitza.de

12. 5. 1 Brightness

Set the display brightness of your device.
• Settings range: 30 - 100 Default: 70
with 30% = dark 100 % = very bright

12. 5. 2 Standby

Set the time in seconds after which the dis­play brightness switched to the set standby brightness.
• Settings range: 60 - 3600 Default: 900

12. 5. 3 Brightness (standby)

Set the display brightness that the device will switch to after the standby time has expired.
• Settings range: 20 - 60 Default: 30
with 20% = dark 60 % = very bright

12. 6 Colors

Set the colors that show the current and volt­age in the graphical displays.
Fig. Menu "Colors,"
• Open the configuration menu (see Section 12 "Configuration").
• With keys 3 () and 4 (), select the menu item "Colors" from the configuration menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the op­tion you want and confirm your selection with key 6 (Enter).
• Set the color with keys 3 () and 4 () .
• Confirm by pressing key 6 (Enter) or abort the action by pressing key 1 (Esc).
• Exit the menu "Color" with key 1 (Esc).
• Exit the configuration menu with key 1 (Esc).
• Open the main menu with key 1 (Esc).
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).
44
www.janitza.de UMG 96-PA

13. Putting the device into service

This section will explain everything you need to know about putting the device into service for the first time.

13. 1 Supply voltage

To connect the supply voltage, proceed as follows:
1. Connect the supply voltage to a terminal on the back of the device.
2. After connecting the supply voltage, the first measurement value display "Over­view" will appear on the display a few seconds later.
3. If no display appears, check whether the supply voltage is within the rated voltage range.
Material damage may
m
CAUTION!
Noncompliance with the connection re­quirements may result in your device being damaged or destroyed.
For this reason, please note the following:
Comply with the details on voltage and frequency on the ratings plate.
Do not use the device to measure di­rect voltage.
result from failure to
comply with the
connection requirements.

13. 2 Measured voltage

Voltage measurements in networks with rated voltages above the maximum stated rated voltage (cf. Section 6.2 Rated Voltages) are to be connected via voltage transformers.
To connect the measured voltage, proceed as follows:
1. Connect the supply voltage to a terminal on the back of the device.
2. After the measured voltage has been con­nected, the measurement values displayed by the device for the voltages L-N and L-L will have to match those at the measure­ment input.
3. Consider any utilized voltage transformer factors.
c
WARNING!
If the device is exposed to voltage surges above the admissible overvoltage category, safety-related insulations in the device may be damaged, whereupon the safety of the product can no longer be guaranteed.
Only use the device in environments in which the admissible surge voltage is not exceeded.
Risk of injury from electric voltage!
ATTENTION!
C
Delete any production-related energy counter contents, minimum and max­imum values, and recordings prior to putting the device into service.
45
UMG 96-PA www.janitza.de

13. 3 Frequency measurement

To perform this measurement, the device requires the network frequency, which can be either supplied by the user or automatically determined by the device (cf. Section "Con­figuration").
• The device can only determine the network frequency if there is a voltage L1-N great­er than 20 Veff (4-wire measurement) or a voltage L1-L2 greater than 34 Veff (3-wire measurement) at the voltage measurement input V1.
• The network frequency needs to be within the range of 45 Hz to 65 Hz.
• If the measurement voltage is not suffi­ciently high, the network frequency can­not be determined, and subsequently no measurement can be performed.
More information can be found in Section 12 Configuration / Rated frequency"
13. 4 Rotary field direction
To do this, open the menu display "Phasor diagram":
• If you are not in the measurement value display "Overview," switch to this view by repeatedly pressing key 1 (Esc).
• Open the main menu with key 1 (Esc).
• With keys 3 () and 4 (), select "Pointer diagram" and confirm your selection with key 6 (Enter).
Fig. Main menu, Selection "Phasor diagram"
Check the direction of the voltage rotary field on the measurement value display of the device.
• A "right" rotary field is usually available.
UL1-UL2-UL3 = right rotary field UL1-UL3-UL2 = left rotary field
Fig. Presentation of the phase sequence according to the rotary field direction.
46
www.janitza.de UMG 96-PA
u
i
u
i
u
i

13. 4. 1 Phasor diagram basics

The phasor diagram graphically describes the phase shift / phase angle between the voltage and the current. The pointer rotates around an origin point at a constant speed ­proportionally to the frequency of the voltage and the current. The pointer diagram there­fore describes the current status of the quantities in an alternating cur­rent circuit.
Representation of Ohmic resistance:
• Voltage and current have the same phase
u
i
Representation of inductance:
• The voltage rushes ahead of the current
• In an "ideal coil," the phase shift is 90°
Representation of capacity:
• The current rushes ahead of the voltage
• In an "ideal condenser," the phase shift is 90°
i
u
In a combination of states, the phase angle "current to voltage" values can range from
-90° to +90°.
Current: Display with
short pointers
Voltage: Display with
I
long pointers
U
u
i
Example of a phasor diagram (3-phase)
UL2
IL2
UL2
IL1
UL1
IL3
The voltage and the current are pushed against each other. The current rushes ahead of the voltage, i.e. the network has a capaci­tive load.
47
UMG 96-PA www.janitza.de
Voltage and current input monitoring in pointer diagrams:
The phasor diagram can be used to check for incorrect connections to the voltage and current inputs.
Example 1
Primarily Ohmic load.
Voltage and Current only have minor difference in the phase.
• The current measurement input is assigned to the correct voltage measurement input.
Example 2
Primarily Ohmic load.
Voltage and Current only have a difference of about 180° in the phase.

13. 5 Measurement current

The device:
• is designed to be connected to current transformers with secondary currents of ../1 A and ../5 A.
• does not measure direct currents.
• has current measurement inputs that can be loaded with 60 A (sinusoidal) for 1 second.
The factory preset current transformer ratio is 5/5 A, and may need to be adjusted to the current transformer in use.
1. Short-circuit all of the current transformer outputs except for one.
2. Compare the current indicated on the de­vice with the attached input current.
• The currents must match when the current transformer conversion ratio is accounted for.
• The device must show approx. zero am­peres in the short-circuited current mea­surement inputs.
48
• The measurement current input is assigned to the correct voltage measurement input.
• In the present current measurement, the connections k and I are switched or energy is being fed back into the supply network.
Fig. phasor diagram
C
ATTENTION!
Voltages and currents outside of the admissible measurement range can destroy the device.
www.janitza.de UMG 96-PA

13. 6 Measurement range violation

The device display shows the warning "Mea­surement range violation" in the event of a
measurement range violation and indicates the current/voltage path.
Measurement range violations are displayed for as long as they continue and cannot be acknowledged. There is a measurement range violation if at least one of the voltage or current measurement inputs is outside of its specified measurement range.
Thresholds for measurement range violation (200 ms effective values):
I = 6 A rms UL-N = 600 V rms

13. 7 Checking the power measurement

Short-circuit all current transformer outputs except one and check the displayed power outputs.
• The device may only display one power output in the phase with the current con­verter input that is not short-circuited.
• If this does not apply, check the con­nection of the measurement voltage and measurement current.
If the amount of effective power is correct but the sign of effective power is negative, this may have two causes:
1. The connections S1(k) and S2(l) are switched on the current transformer.
2. Effective power is returned to the network.
Fig. Warning with measurement range violations in the current and voltage paths I1-I3 and U1-U3
Fig. The pointer diagram displays voltages with long pointers and currents with shorter pointers
Open the pointer diagram with information on the power:
• If you are not in the measurement value display "Overview," switch to this view by repeatedly pressing key 1 (Esc).
• Open the main menu with key 1 (Esc).
• With keys 3 () and 4 (), select "Pointer diagram" and confirm your selection with key 6 (Enter).
49
UMG 96-PA www.janitza.de

13. 8 Checking communication

The device counts all received (RX), all sent (TX) and all faulty data packages.
In the ideal case, the number of errors shown in the error column is zero.
Fig. Menu "Overview" with a view of the communication parameters (Com. View)
• If you are not in the measurement value display "Overview," switch to this view by repeatedly pressing key 1 (Esc).
• Open the main menu with key 1 (Esc).
• With keys 3 () and 4 (), select the menu item "System" from the main menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 (), select the sub­menu "COMports Overview" and confirm your selection with key 6 (Enter).
Open the help menu:
• If you are in the view "COMports Over- view," open the submenu with key 1 (Esc).
• Open the main menu by pressing key 1 (Esc) again.
• Select the measurement value display you want with keys 3 () and 4 (). Confirm your selection by pressing key 6 (Enter) or jump directly to the measurement value display "Overview" by pressing key 1 (Esc).

13. 9 Delete min./max. values individually

In the measurement displays for voltage, current and power with
• voltage L-N and voltage L-L
• current and THD-I
• total power and
• effective, reactive, and apparent power the min./max. values for each can be deleted with key 6 (Enter).
Fig. Measurement display "Voltage L-N" with the menu "Delete/Abort"
• If you are not in the measurement value
display "Overview," switch to this view by repeatedly pressing key 1 (Esc).
• Open the main menu with key 1 (Esc).
• With keys 3 () and 4 (), select the menu
item "Voltage" or "Current" or "Power" from the main menu and confirm with key 6 (Enter).
• With keys 3 () and 4 (), select a mea-
surement display (see above) and confirm your selection with key 6 (Enter).
• In the measurement display you can delete
the corresponding minimum and maximum values by pressing key 6 (Enter).
• For this, confirm the "Delete/Abort" notifi-
cation with key 6 (Enter). You can abort the action by using the option "Abort" (select with key 3 or 4) and then confirming with key 6 (Enter).
50
www.janitza.de UMG 96-PA

13. 10 Harmonics

Harmonics are caused by equipment with nonlinear characteristics. These additional frequencies are integer multiples of a funda­mental oscillation, and show how the equip­ment affect the power grid.
The possible effects of harmonics include:
• additional heating of equipment.
• additional current on the neutral conductor
• an overload and reduced life cycle of elec­trical usage
Harmonic loads are the primary cause of invisible voltage quality problems associated with enormous repair costs and investment to replace defective devices.
The device measures the fundamental oscil­lation of the voltage in the range of 45 - 65 Hz. This fundamental oscillation is the basis for the calculation of the harmonics for the voltages and currents.
The device calculates harmonics up to 25 times the fundamental oscillation.
• If you are not in the measurement value display "Overview," switch to this view by repeatedly pressing key 1 (Esc).
• Open the main menu with key 1 (Esc).
• With keys 3 () and 4 (), select the menu item "Harmonic" and confirm your selec­tion with key 6 (Enter).
• With keys 3 () and 4 (), select the harmonics for the respective voltage or current and confirm your selection with key 6 (Enter).
Fig. Measurement value display "Harmonic"
51
UMG 96-PA www.janitza.de

13. 11 Communication in the bus system

13. 11. 1 RS485

The data can be accessed from the param­eter and measurement value list (cf. Section "Configuration / Communication“) via the MODBUS RTU protocol with the CRC check on the RS485 interface.
Modbus functions (Slave)
03 Read Holding Registers 04 Read Input Registers 06 Preset Single Register 16 (10Hex) Preset Multiple Registers 23 (17Hex) Read/Write 4X Registers
The byte order is highbyte before lowbyte (Motorola format).
Transfer parameters
Data bits: 8 Parity: odd even none (1 Stopbit) none (2 Stopbits)
Stopbits (UMG 96-PA): 1 / 2 Stopbits, external: 1 / 2
Number formats
short 16 bit (-215 .. 215 -1) float 32 bit (IEEE 754)
For further information about the interface settings, refer to Section 12 "Configuration / Communication."
Example: Read-out of voltage L1-N
Voltage L1-N is saved in the measurement value list at the address 19000. Voltage L1-N is available in FLOAT format. In this example, 01 will be used as the device address.
The "Query Message" will then appear as follows:
Name Hex Comment
Device address 01 Address=1
Function 03 "Read Holding Reg"
Starting ad­dress Hi
Starting ad­dress Lo
Number of val­ues Hi
Number of val­ues Lo
Error Check (CRC)
The "response" of the device may look as follows:
Name Hex Comment
Device address 01 Address=1
Function 03
Byte counter 06
Data 00 00hex=00dez
Data E6 E6hex=230dez
Error Check (CRC)
4A 19000dez = 4A38hex
38
00 2dez = 0002hex
02
-
-
52
The voltage L1-N read back from the address 19000 is 230 V.
www.janitza.de UMG 96-PA
21 22 23
24V DC
K1 K2
=
-
+
=
+
-
S1 S2
Digital Outputs
17 18 19
Digital Inputs
20
S3
24
K3

13. 12 Digital inputs/outputs

Your device has three digital inputs and three digital outputs.
Fig. Digital inputs and outputs
• You can configure the inputs and outputs via the GridVis® software.
• The GridVis® software can be found on the homepage (www.janitza.de) in the download area.

13. 12. 1 Digital inputs

It is possible to transmit information from oth­er devices that have a digital output directly to your device via the digital inputs.
It is still possible to configure the digital inputs as function inputs. In this mode, each digital input has a specific function. In this configuration it is not possible for these in­puts to function as pulse counters!
You can set the digital outputs in the "Periph- ery" area via the configuration window of the GridVis® software:
Function mode
• Which function should be assigned to the digital input.
Pulse counter
• Which value type the incoming signal has (e.g. electrical power, gas and water consumption, CO2 ...)
• Which pulse value should be used for the value.
• How long the transmission time is.
The statuses of the digital inputs are each on their own Modbus address. For each digital input, the last 16 switching operations (events) are logged with a time stamp.
Fig. Configuration of the digital inputs via the GridVis®
software
Function mode
Each digital input can be assigned a specific function with:
• Digital input 1 Configuration as tariff converter (HT/NT) possible.
• Digital input 2 Synchronization of the internal clock; choosing between minute or hour synchro­nization is possible. However, synchronization is also possible via a Modbus address.
• Digital input 3 Configuration as reset entry for the syn­chronization values of the drag indicator function. It is also possible to synchronize the drag indicator via the Modbus address.
53
UMG 96-PA www.janitza.de
Pulse counter
All digital inputs can be operated with a frequency of 25 Hz. Here, the pulse duration and the pulse interval must be greater than 20 ms. The typical pulse duration for S0 pulses is 30 ms.
Calculation of measurement value:
Measurement value = Pulse x Pulse value
Calculation of power value:
Pulse duration
> 20ms
Pulse pause
> 20ms
The maximum number of pulses per hour is based on the minimum pulse duration and the minimum pulse interval:
Pulse
length
20 ms 20 ms 90 000 pulses/h
30 ms 30 ms 60 000 pulses/h
50 ms 50 ms 36 000 pulses/h
100 ms 100 ms 18 000 pulses/h
500 ms 500 ms 3 600 pulses/h
1 s 1 s 1 800 pulses/h
10 s 10 s 180 pulses/h
Examples of the maximum number of pulses per hour.
Pulse inter-
val
Max. pulses/h
The pulse counters can be configured with a simultaneous measurement value or power calculation. The pulses are counted as 64-bit numbers and overflow after approx. 1.17 x 1010 years of continuous operation (25 Hz).
Pulse value: A pulse value can be assigned to each digital input. Use the impulse value to specify the measurement value or power value (e.g. ener­gy) to which a pulse should correspond.
Power value =
Pulse x Pulse value
Time [sec]
Since the pulse interval may be very large, continual calculation of the measurement or power values is not possible. For this reason, only the mean values are calculat­ed. For measurement value calculation, the mean values are calculated from the number of pulses per period multiplied by the pulse value. For power value calculation, this value must be divided by the amount of time that can be set.
The period is assigned to the digital input and can be set in a range of 1 to 60 minutes. After the period ends, the value can be called via the Modbus.
For each digital input, an external synchroni­zation can be triggered; in this case, a syn­chronisation pulse completes one period and starts a new period. A fixed capture period of 30 seconds is preset for external synchroni­zation. If no synchronization pulse exists after the period expires, a maximum of 30 seconds is waited and the software then carries out the synchronization. All further periods are synchronized by the software.
A period of 15 min is the factory setting.
54
C
ATTENTION!
The pulse interval is proportional to power within the selected settings.
The calculation result for the S0 power value is not available until the end of the cycle.
www.janitza.de UMG 96-PA
C

13. 12. 2 Digital outputs

Various functions can be assigned to the three digital outputs:
• Digital output 1
• Digital output 2
• Digital output 3
You can set the digital outputs in the "Periph- ery" area via the configuration window of the GridVis® software:
ATTENTION!
When programming with GridVis®, a selection of energy values that are derived from the power values is received.
- Pulse output for effective energy
- Output for the timer
- ModBus remote output
- Pulse output for reactive energy
- Output for comparator group 1
- Output for the timer
- ModBus remote output
- Output for comparator group 2
- Output for the timer
- ModBus remote output
Pulse output
Digital outputs 1 and 2 can be used for the output of pulses for counting effective energy and apparent energy. For this,a pulse is set at the output after a certain adjustable amount of energy has been reached.
To use a digital output can as a pulse output, you must apply two different settings using the GridVis® Software in the configuration menu:
• Mode for the digital input: S0 output
• Output polarity: Closer, Opener
• Pulse value
Pulse value
With the pulse value, you indicate how much energy (Wh or varh) one pulse should corre­spond to.
The pulse value is determined by the max­imum connected load and the maximum number of pulses per hour.
If the pulse value is specified with a:
• positive sign, pulses will only be issued if the measured value also has a positive sign.
• negative sign, pulses will only be issued if the measured value also has a negative sign.
Fig. Configuration of the digital outputs via the Grid-
Vis® software
C
C
ATTENTION!
Since the active energy meter works with a return stop, pulses are only issued during import of electrical energy.
ATTENTION!
Since the reactive energy meter works with a return stop, pulses are only issued under inductive load.
55
UMG 96-PA www.janitza.de
Determine pulse value
1. Set the pulse length according to the re­quirements of the connected pulse receiv­er. For a pulse length of e.g. 30 ms, the device can issue a maximum number of 60,000 pulses (see Table "Maximum Pulse Number") per hour.
2. Determining the maximum connected load.
Example: Current transformer = 150/5 A Voltage L-N = max. 300 V
Power per phase = 150 A x 300 V = 45 kW
Power in 3 phases = 45 kW x 3 Maximum connection power = 135 kW
m
CAUTION!
When using the digital outputs as a pulse output, measurement errors may result from residual ripple.
So use a power adapter whose residual ripple is less than 5% for the supply volt­age to the digital inputs and outputs.
Measurement errors in use
as pulse output
3. Calculate the pulse value:
Pulse value =
Maximum connected load
Max. number of pulses/h
[Pulses/Wh]
Pulse value = 135 kW / 60000 pulses/h Pulse value = 0.00225 pulses/kWh Pulse value = 2.25 pulses/Wh
External
230V AC
Data col-
lector
UMG 96-PA
Switch and pulse outputs
Digital Ouput 1
Digital Ouput 2
operating voltage
17
+24V=
18
19
24V DC
+
1.5k
-
56
Fig.: Fig.: Connection example for wiring the pulse output.
www.janitza.de UMG 96-PA
Timer output
In the device, 64 independent weekly timers can be set with:
• a resolution of 1 minute.
• a freely defined active period within one day. The active day within the week can be selected.
Example: 9:25 to 11:45 on Sunday, Monday and Friday
The weekly timers can be configured as
• tariff converters (1 and 2)
• Set the digital outputs 1 to 3
• as "inoperable
." The status can always be queried via the Modbus. The statuses of the timers are linked via an "OR" operator at the digital output.
The weekly timers can be configured using the GridVis® software in the "Timer" configu­ration area)
Output for Modbus remote
Enables switching of the outputs via a Mod­bus address.
This function can be configured via the Grid­Vis® software:
• Open the device configuration in GridVis®.
• Set the mode for the digital outputs under "Periphery" to "Modbus remote output."
• Determine the output polarity with:
- Current flow when active (closer)
- Current flow when inactive (opener)
Fig. Configuration of the weekly timer via the GridVis®
software
Fig. Configuration of the digital outputs as "Modbus
remote" in the GridVis® software
57
UMG 96-PA www.janitza.de
Output for the comparator group
Two comparator groups are available to monitor the thresholds (comparators 1 and
2), each with 3 comparators (A - C).
The results of the comparators A to C can be linked using the "AND" or "OR" operators.
The linkage result from comparator group 1 can be assigned to digital output 2 and the linkage result from comparator group 2 can be assigned to digital output 3.
The comparators can only be configured using the GridVis® software in the "Compar- ator" configuration area.
Comparator
Threshold
Actual value
Status
Comparator runtime
Fig. "Comparators" in the menu "Overview / Comparators"
Logic
58
Fig. Configuration of the comparators in the GridVis®
software
Read out comparator settings on the device:
• With keys 3 () and 4 (), select the menu item "Overview" from the main menu.
• Confirm by pressing key 6 (Enter).
• With keys 3 () and 4 () select the submenu "Comparator 1" for the compar­ator group 1 and "Comparator 2" for the comparator group 2.
• Confirm by pressing key 6 (Enter).
www.janitza.de UMG 96-PA
Comparator runtime
Comparator runtimes are time counters that create a total for a set comparator output. I.e. if the condition of the comparator is met and the lead time has expired, the counter increases by the respective amount of time ­the minimum switch-on time is not taken into account here!
Comparator with set threshold violation
• The set threshold is compared to the mea­surement value.
• If there is a threshold violation for at least the duration of the lead time, a change will be made to the comparator result.
• The result remains for at least the duration of the minimum switch-on time and at the most for the duration of the threshold violation. If there is no longer a threshold violation and the minimum switch-on time has expired, the result is reset.
Threshold
Threshold violation (e.g. exceeding the
threshold)
Lead time
Minimum
Switch-on time
Comparator result
Comparator runtime
Measurement
UMG 96-PA
Comparator group 1
Comparator A
Comparator B
Comparator C
Fig.: Block diagram: Use of the digital output 2 for monitoring the thresholds
Logic
Result
0/1
Inverter
- not inverted
- inverted
Display
status
0/1
Digital­output 1
21
23
59
UMG 96-PA www.janitza.de

13. 13 Analog output

The device has 1 passive analog output, which can output a maximum current of 0 ­20 mA. An external 24 V DC power adapter is re­quired for operation.
20 mA
4 mA
230 V
Initial value
Fig.: Principle of the analog output with voltage monitoring
250 V
Final value
The analog output can be configured in a user-friendly way via the GridVis® software. To do this, enter the assigned measurement value, the initial and final value and the output range in "Periphery."
Examples:
Assignment of effective power L1 (output range 4 - 20 mA)
20 mA
4 mA
500 W
Initial value
1000 W
Final value
• If the effective power is 500 W, a current of 4 mA is outputted via the analog output. At an effective power of 1000 W, the output is 20 mA. The measured effective power is directly proportional to the current of the analog output.
Assignment of the calculated effective power factor cos ϕ (math.) (output range 4 - 20 mA)
Fig. Configuration of the analog output in the Grid-
Vis® software
20 mA
20 mA
4 mA
cos ϕ = -1
Initial value
0
cos ϕ = 1
Final value
• Monitoring of the effective power factor cos ϕ (math.) with cos ϕ (math.) > 0 effective power, consumed cos ϕ (math.) < 0 effective power, delivered.
60
www.janitza.de UMG 96-PA

13. 14 “Drag indicator” function

The “drag indicator” function describes the three highest mean value of a measured val­ues over a defined period of time (time base).
• The calculated mean values can be re­trieved via the GridVis® software and via a parameter with a time stamp.
• The period duration (time base), synchro­nization and capture time can be set in the GridVis® software by using the corre­sponding parameters.
• The calculation of the mean values is de­rived from the measurement values of the following value types:
- Current L1
- Current L2
- Current L3
- Effective power L1
- Effective power L2
- Effective power L3
- Effective power, total (L1...L3)
- Apparent power L1
- Apparent power L2
- Apparent power L3
- Apparent power, total (L1...L3)
Capture time:
The individually adjustable Capture time de­scribes a time window in which an incoming pulse triggers a synchronization of the time point. If the device receives a pulse outside of the capture time, the calculated mean val­ues are deleted and the time is reset.
Attention: The setting for the capture time — e.g. in the GridVis® software — describes half of the time window of the entire capture time!
Timepoint Calculation of average value
Capture time Capture time Capture time Capture time
Fig.: Principle of synchronization
Timepoint Calculation of average value
Timepoint Calculation of average value
Time baseTime baseTime base
Timepoint Calculation of average value

13. 14. 1 Internal synchronization

Period duration (time base):
Individually adjustable period duration in sec­onds for the calculation of the mean values over this timeframe (duration of the mea­surement value recording). When an internal synchronization is selected, the mean values are recalculated once the set timeframe has expired.
Synchronization mode:
A synchronization defines a starting time point for the calculation periods of the mean values. A synchronization is triggered
• by the internal clock (internal synchroniza- tion),
• by setting a parameter (via Modbus) or
• optionally via the digital input 3 (external synchronization)
.
The means are calculated after the adjustable period duration (time base) has expired. The internal synchronization is performed on the full minute if that describes a multiple of the time base.
Time base
[min]
2 09:00:00 09:02:00 09:04:00 09:06:00
5 09:00:00 09:05:00 09:10:00 09:15:00
15 09:00:00 09:15:00 09:30:00 09:45:00
Fig. Examples of an internal synchronization with different time bases
Sync 1
(Clock time)
Sync 2
(Clock time)
Sync 3
(Clock time)
Sync 4
(Clock time)
ATTENTION!
C
The options synchronization via Modbus AND synchronization via the digital output 3 must be deac­tivated for an internal synchroniza­tion!
61
UMG 96-PA www.janitza.de

13. 14. 2 External synchronization

An external synchronization to calculate the 3 highest mean values can be carried out
• via the digital output 3 (e.g. via a pulse transducer) or
• via a Modbus command
Scenarios of external synchronization:
"No pulse despite setting"
If no pulse occurs via the digital output 3 or a Modbus command, the measurement values are saved as they are in an internal synchro­nization - but on the minute!
Capture time Capture time Capture time Capture time
Time base Time base Time base
Principle of synchronization
Pulse curve for the digital input 3
Fig.: Principle of synchronization with "No pulse despite setting"
"One pulse"
If the device receives a single pulse or a Modbus command outside the capture time, the measurement values that have been totaled up to that point for the calculation of the mean value and the time are reset. The time point is redefined as a relative null point and a recalculation is carried out!
Capture time Capture time Capture time Capture time
Time base Time base Time base
Principle of synchronization
Relative zero point
Time base Time baseTime base
Resynchronization
Pulse curve for the digital input 3
Fig.: Principle of synchronization with "a pulse outside the capture time"
Example Maximum value Value Time
stamp
Effective current L1 Drag indicator 1 3.51 A 09:13:07
Effective current L1 Drag indicator 2 2.52 A 09:08:07
Effective current L1 Drag indicator 3 1.52 A 09:03:07
Fig. Example of drag indicator saving with time stamp (with a set time base of 5 min)
Example:
Example Maximum value Value Time
Effective power L1
Effective power L1
Effective power L1
Fig. Example of drag indicator saving with time stamp (with a set time base of 5 min)
Drag indicator
Ref. 1
Drag indicator
Ref. 2
Drag indicator
Ref. 3
396.73 W 09:18:47
207.34 W 09:13:47
80.59 W 09:08:47
stamp
Power increases as the time runs. The values are set back to 0 by the pulse (09:06:47) outside the capture time A new totaling of the interim values starts at that time point. Because no further pulse arrives, the mean value is calculated after the set time (time base).
62
www.janitza.de UMG 96-PA
"Periodic pulses"
Different scenarios may result if the device receives periodic pulses via the digital output 3 or periodic Modbus commands.
Capture time
Relative zero point
Fig.: Principle of synchronization with "periodic pulses" on the digital input 3
Capture time Capture time Capture time
Time base Time base Time base
Principle of synchronization
Resynchronization
2 minutes
Pulse curve for the digital input 3
Scenario "pulse outside the capture time"
• Totaled interim values are set to 0
• The time is set to 0 (new relative zero point)
• No value calculation occurs
• Totaled interim values are set to 0
• The time is set to 0 (new relative zero point)
• No value calculation occurs
Pulse
Time base
stop
Scenario "pulse before the time base, inside the capture time"
• Now perform the value calculation
• The time is set to 0 (new relative zero point)
• Delete the totaled interim values
Pulse
Time base
stop
Pulse
Capture time
Time base
stop
Scenario "pulse after the time base, but inside the capture time"
Attention: The time is synchronized for each pulse in periodic synchronization!
63
UMG 96-PA www.janitza.de

13. 14. 3 Synchronization priority

An external synchronization is carried out according to various priorities:
• Priority 1: Modbus synchronization For this, use the Modbus tool to set the "Enable Flag" (address.: 822) or select the option "Synchronization via Modbus" in the GridVis® software in the configuration area for the drag indicator.
• Priority 2: Synchronization via the digital input 3
For this, set the Modbus parameter "FUNC_SYNC_RECORD" (address: 30048) to the value 4 or select the option "Drag Indicator Synchronization" in the GridVis® software in the configuration area for the periphery (digital input 3). Attention: Ensure that in the configuration for the drag indicator the option "Synchro­nization via Modbus" is NOT selected!
• Priority 3: Internal synchronization
Modbus Address
820
821 Time base in seconds 60 .. 65535
822 Enable flag for the Modbus trigger 0 .. 1
823 Capture time in seconds 0 .. 255
30048 Configuration of the inputs 0 .. 4*
* 0 = FUNC_NONE; 1 = FUNC_TARIF; 2 = FUNC_SYNC_CLOCK_MIN; 3 = FUNC_SYNC_CLOCK_H; 4 = FUNC_SYNC_RECORD
Modbus address table for a synchronization
Drag indicator synchronization
Function Settings
Set trigger flag for the
range
0 .. 1
64
Fig. Drag indicator synchronization in the GridVis®
software
Fig. Configuration "Synchronization via the
digital input 3 in the GridVis® software
www.janitza.de UMG 96-PA

13. 15 Recordings

2 recording profiles are preconfigured in the default factory setting of the device. Record­ings can be adjusted and extended via the GridVis® software.
• The minimum recording time base is 1 minute.
• A maximum of 4 recordings, each with 29 measured values, is possible. If minimum and maximum values are defined addi­tionally, the number is lowered to 19 or 14 values.
• Within the recording configuration, the measurement values are defined via a time base according to the types Mean Value, Sample, Maximum / Minimum.
• Type Mean Value: Arithmetic mean of the
measurement values over a predeter­mined duration of time.
• Type Maximum and Minimum: Maximum
/ Minimum values over a predetermined duration of time.
• Type Sample: Measurement value at the
end of a predetermined duration of time.
Attention: Recording of operational values can only be done with the type Sample.
• Effective current L1
• Effective current L2
• Effective current L3
• Effective current total L1..L3
• Effective power L1
• Effective power L2
• Effective power L3
• Effective power total L1..L3
• Apparent power L1
• Apparent power L2
• Apparent power L3
• Apparent power total L1..L3
• cos phi(math.) L1
• cos phi(math.) L2
• cos phi(math.) L3
• cos phi(math.) total L1..L3
• Reactive power fundamental oscillation L1
• Reactive power fundamental oscillation L2
• Reactive power fundamental oscillation L3
• Reactive power fundamental oscillation total L1..L3
Recording 2
The following measurement values are re­corded with the time base of 1 hour.
• Drawn effective energy L1
• Drawn effective energy L2
• Drawn effective energy L3
• Drawn effective energy total L1..L3
• Inductive apparent energy L1
• Inductive apparent energy L2
• Inductive apparent energy L3
• Inductive apparent energy total L1..L3
Fig. Recording configuration in the GridVis® software
Recording 1
The following measurement values are re­corded with the time base of 15 minutes.
• Effective voltage L1
• Effective voltage L2
• Effective voltage L3
65
UMG 96-PA www.janitza.de

13. 16 Tariff switching

Electrical energy values (effective, reactive and apparent energy) can be recorded via internal counters for two tariffs each.
Switching between the tariffs (HT/NT) is sup­ported by
• Modbus via the
• digital input 1 (see Section "Digital Inputs") or via the
• weekly timer (see Section "Timer Output")
.
Fig. Configuration of the digital input 1 as the tariff
control input in the GridVis® software
Device display of total (L1..L3) effective, reactive and apparent energy by tariffs
The tariff switching can be configured via the GridVis® software.
Fig. Timer configuration in the
GridVis® software
66
www.janitza.de UMG 96-PA
67
UMG 96-PA www.janitza.de

14. Overview of measurement value displays

Main menu (Overview)
Main menu (Voltage)
Network analysis (splash page)
Display of voltage L1, L2, L3-N; current L1, L2, L3; power L1, L2, L3 and effective/apparent energy L1-L3
Voltage L-N
Display of voltage L1-N, L2-N, L3-N and their minimum and maximum values
Voltage L-L
Display of voltage L1-L2, L2-L3, L1-L3 and their minimum and maximum values
Curve
Display of voltage curve for L1-N, L2-N, L3-N
68
www.janitza.de UMG 96-PA
Main menu (Current)
Current
Display of current L1, L2, L3 and their minimum and maximum values
Curve
Display of current curve for L1, L2, L3
THD-I
Display of distortion factors for the current (THD-I) L1, L2, L3 and their minimum and maximum values
Main menu (Power)
Total powers
Display of the total (L1..L3) of effective, reac­tive, apparent power and their minimum and maximum values
Effective / reactive / apparent power curves
Display of curves for effective, reactive or apparent power (total L1..L3)
Effective / reactive / apparent power
Display of power values (effective, reactive, apparent power) L1-N, L2-N, L3-N and their minimum and maximum values
69
UMG 96-PA www.janitza.de
Main menu (Energy)
Effective energy
Display of total (L1..L3) effective energy (total/ drawn/delivered)
Apparent energy
Display of total (L1..L3) apparent energy
Reactive energy
Display of total (L1..L3) reactive energy (total/ inductive/capacitive)
Tariff
Display of total (L1..L3) effective, reactive and apparent energy by tariffs
70
Main menu (Consumption Overview)
Main menu (Harmonics)
Effective / reactive / apparent power /
month
Display of effective, reactive or apparent energy per month (in the last three years)
Voltage L1 / L2 / L3
Display of effective, reactive or apparent energy per month (in the last three years)
Effective / reactive / apparent power /
day
Display of effective, reactive or apparent energy per day (for the current month)
Current L1 / L2 / L3
Display of effective, reactive or apparent energy per day (for the current month)
www.janitza.de UMG 96-PA
Main menu (Oscilloscope)
Voltage L1 / L2 / L3
Display of oscillogram for voltage L1, L2 or L3
Current L1 / L2 / L3
Display of oscillogram for current L1, L2 or L3
Voltage L1..L3
Display of oscillogram for voltages L1, L2 and L3
Current L1..L3
Display of oscillogram for currents L1, L2 and L3
Main menu (Pointer Diagram)
Pointer diagram
Display of voltages and currents in the pointer diagram
71
UMG 96-PA www.janitza.de
Main menu (System)
Attention: Further information on configuration can be found in the Section Operation and Configuration.
Configuration
Submenu Device Configuration
Comparator 1 / Comparator 2
Display of threshold, actual value, comparator runtime, logic and status
COMports Overview
Display of received (RX), sent (TX) and faulty data packages. Switching current analog output
72
www.janitza.de UMG 96-PA
Submenu (System / Configuration)
Attention! Further information on configuration can be found in the Section Operation and Configuration.
Language
Settings in the device language
Measurement (frequency)
Settings for the standard frequency
Communication
Settings for the parameters device ad­dress, baud rate and data framework
Measurement L1..L3
Settings for the current transformer and voltage transformer conversion ratios, and the rated current
System
Display of device data, password assignment and ways to reset the device
Display
System (Reset)
Display of device data, password assignment and ways to reset the device
Colors
Color settings of the graphsDisplay settings
73
UMG 96-PA www.janitza.de

15. Service and maintenance

The device underwent various safety checks before delivery and is marked with a seal. If a device is open, then the safety checks must be repeated. Warranty claims will only be accepted if the device is unopened.

15. 1 Repairs

Repair work can be carried out by the manu­facturer only.
15. 2 Front film
The front film can be cleaned with a soft cloth and standard household cleaning agents. Do not use acids and products containing acid for cleaning.

15. 3 Disposal

Please comply with national regulations! Dis­pose of individual components, depending on their constitution and the country-specific legal provisions, e.g. as:
• electronic scrap
• plastics
• metals
or commission a certified disposal operation with the scrapping of the device.

15. 5 Device calibration

The devices are calibrated by the manufac­turer at the factory. It is not necessary to recalibrate the device providing that the envi­ronmental conditions are complied with.

15. 6 Re-calibration

We recommend having a re-calibration car­ried out by the manufacturer or an accredited laboratory approximately every 5 years.

15. 7 Firmware update

To update the firmware, connect the device to a computer and request it via the GridVis® software.
Open the firmware assistant by clicking on "Update Device" in the "Extras" menu.
Select the corresponding update file and perform the update.
74

15. 4 Service

Should questions arise, which are not de­scribed in this manual, please contact the manufacturer directly. We will need the following information from you to answer any questions:
• Device name (see rating plate)
• Serial number (see rating plate)
• Software release (see system display)
• Measurement voltage and supply voltage
• Precise description of the error.
Fig. Updating the device firmware in the GridVis®
software
www.janitza.de UMG 96-PA

15. 8 Battery

The internal clock is fed from the supply volt­age. If the supply voltage fails then the clock is powered by the battery. The clock provides date and time information, e.g. for the record­ings and the minimum and maximum values.
The life expectancy of the battery is at least 5 years with a storage temperature of +45 °C. The typical life expectancy of the battery is 8 to 10 years.
The battery is replaced via the battery insert provided on the bottom of the device. Ensure that the correct type of battery is used here and observe correct polarity when changing it (plus pole facing the back of the device, and the minus pole towards the front)!
c
WARNING!
Serious bodily harm or death may result from hazardous voltages.
For this reason, please note the follow­ing:
De-energise the system before starting work.
C
C
ATTENTION!
Grease or dirt on the contact sur­faces form a transfer resistance that will shorten the life of the battery. Only touch the battery at the edges.
ATTENTION!
Make sure that the correct type of battery is used and observe correct polarity when changing it.
Risk of injury from
electric voltage!
75
UMG 96-PA www.janitza.de

15. 9 Procedure in the event of an error

Possible error Cause Remedy
No display External fuse for the supply voltage was
No current display. Measurement voltage not
Displayed current is too high or too low.
Displayed voltage is too high or too low.
Displayed voltage is too low. Measurement range violation. Use voltage transformer.
Phase shift ind./cap. Current path is assigned to the wrong voltage
Effective power drawn/delivered have been switched.
Effective power is too high or too low.
An input/output does not respond. The input/output was programmed incorrectly. Check programming and correct as needed.
Display "Measurement range violation"
No connection to device. RS485
Despite taking the above measures, the device does not work.
triggered.
connected.
Measurement current not connected. Connect measurement current.
Current measurement in wrong phase. Check connection and correct as needed.
Current transformer factor programmed incor­rectly.
The peak current value at the measurement input was exceeded by current harmonics.
The current at the measurement input was exceeded.
Measurement in wrong phase. Check connection and correct as needed.
Voltage transformer factor programmed incor­rectly.
The peak voltage value at the measurement input was exceeded by harmonics.
path.
At least one current transformer connection has been switched.
A current path is assigned to the wrong voltage path.
The programmed current transformer conver­sion ratio is wrong.
The current path is assigned to the wrong voltage path.
The programmed voltage transformer conver­sion ratio is wrong.
The input/output was connected incorrectly. Check connection and correct as needed.
There is a measurement range violation Check connection and correct as needed.
- Wrong device address.
- Different bus speeds (Baud rate) and/or data framework
- Wrong protocol.
- No termination.
Device is defective. Send device back to
Replace fuse.
Connect measurement voltage.
Read out and program the current transformer conversion ratio on the current transformer.
Install a current transformer with a larger current transformer conversion ratio.
Install a current transformer with a smaller current transformer conversion ratio.
Read out and program the voltage transformer conversion ratio on the voltage transformer.
Attention! Ensure that the measurement inputs are not overloaded.
Check connection and correct as needed.
Check connection and correct as needed.
Check connection and correct as needed.
Read out and program the current transformer conversion ratio on the current transformer
Check connection and correct as needed.
Read out and program the voltage transformer conversion ratio on the voltage transformer.
Correct current /voltage transformer conversion ratio.
- Correct device address.
- Correct speed (baud rate). Correct data framework.
- Correct protocol.
- Connect bus to a termination resistor.
manufacturer with an exact description of the error/fault.
76
www.janitza.de UMG 96-PA

16. Technical Data

General
Net weight (with connectors attached) approx. 250 g
Packaged weight (incl. accessories) approx. 500g
Battery Lithium type CR2032, 3V (approved as per UL 1642)
Life cycle of the backlighting 40000h (Backlighting diminishes over this
Transport and storage
The following details apply for devices that are transported and stored in the original packaging.
Free fall 1m
Temperature K55 (-25°C to +70°C)
Relative humidity 0 to 90% RH
Environmental conditions for operation
The device is suitable for weather-protected, non-mobile use. Protection class II as per IEC 60536 (VDE 0106, Part 1).
Measurement temperature range K55 (-10°C .. +55°C)
Relative humidity 0 to 75 % RH
Operating height 0 .. 2000m over NN
Pollution degree 2
Installation position vertical
Ventilation External ventilation is not required.
Protection from foreign objects and water
- Front
- Back
- Front with seal
time period to approx. 50%)
IP40 as per EN60529 IP20 as per EN60529 IP54 as per EN60529
Supply voltage
Option 230V Nominal range AC 90 V - 277 V (50/60 Hz) or DC 90 V - 250 V;
Power consumption max. 4.5 VA / 2 W
Option 24V Nominal range AC 24 V - 90 V (50/60Hz) or DC 24 V - 90 V;
Power consumption max. 4.5 VA / 2 W
Operating range +-10% of nominal range
Internal fuse not exchangeable
Recommended overvoltage protective device for the Line protection (UL approval)
Recommendation for the maximum number of devices on the line circuit breaker: Option 230V : Line circuit breaker B6A: max. 4 devices / Line circuit breaker B16A: max. 11 devices Option 24V : Line circuit breaker B6A: max. 3 devices / Line circuit breaker B16A: max. 9 devices
Type T1A / 250 V DC / 277 V AC as per IEC 60127
300 V CATIII
150 V CATIII
Option 230 V: 6 - 16 A (Char. B) Option 24 V: 1 - 6 A (Char. B)
77
UMG 96-PA www.janitza.de
Voltage measurement
Three-phase 4-wire systems with rated volatges up to 417 V / 720 V (+-10%) as per IEC
Overvoltage category 600 V CAT III
Rated surge voltage 6 kV
Fuse for voltage measurement 1 - 10 A (with IEC/UL approval)
Measurement range L-N 01) .. 600 Vrms (max. overvoltage 800 Vrms)
Measurement range L-L 01) .. 1040 Vrms (max. overvoltage 1350 Vrms)
Resolution 0.01V
Crest factor 2.45 (relative to the measurement range)
Impedance
Power consumption approx. 0.1 VA
Sampling rate 8.33 KHz
Frequency of the fundamental oscillation
- Resolution
1)
The device can only determine measurement values if there is at the voltage measurement input V1 a voltage L1-N greater than 20 Veff (4-wire measurement)
or a voltage L1-L2 greater than 34 Veff (3-wire measurement).
347 V / 600 V (+-10%) as per UL
3 M/Phase
45 Hz.. 65 Hz
0.01Hz
Current measurement I1 - I3
Rated current 5 A
Measurement range 0.005 .. 6 Arms
Crest factor (based on rated current) 2
Overvoltage category 300 V CAT II
Rated surge voltage 2 KV
Power consumption
Overload for 1 sec. 60 A (sinusoidal)
Sampling rate 8.33 KHz
Serial interface
RS485 - Modbus RTU/Slave 9.6 kbps, 19.2 kbps, 38.4 kbps, 57.6 kbps, 115.2 kbps
Digital outputs
3 digital outputs, semiconductor relay, not short-circuit proof.
Switching voltage max. 33 V AC, 60 V DC
Switching current max. 50 mAe ff AC/DC
Response time approx. 200 ms
Pulse output max. 50Hz (energy pulses)
Digital inputs
3 digital inputs, semiconductor relay, not short-circuit proof.
Maximum counter frequency 20 Hz
Input signal is on 18 V.. 28 V DC (typically 4 mA)
Input signal is not on 0 .. 5 V DC, current less than 0.5 mA
approx. 0.2 VA (Ri=5 m)
78
Conductor length (digital inputs/outputs)
up to 30m not screened
more than 30m screened
www.janitza.de UMG 96-PA
Analog output
External supply max. 33 V
Current 0 .. 20 mA
Update time 1 sec
Resolution 10 bit
Connectability of the terminals (supply voltage)
Connectable conductors. Only connect one conductor per terminal!
Single-wire., multiwire, fine-wire 0.08 - 4.0 mm2, AWG 28-12
Pin terminals, wire-end ferrules 0.2 - 2.5 mm
Tightening torque 0.4 - 0.5 Nm
Insulation length 7 mm
Connectability of the terminals (voltage measurement)
Connectable conductors. Only connect one conductor per terminal!
Single-wire., multiwire, fine-wire 0.08 - 4.0 mm2, AWG 28-12
Pin terminals, wire-end ferrules 0.2 - 2.5 mm
Tightening torque 0.4 - 0.5 Nm
Insulation length 7 mm
2
2
Connectability of the terminals (current measurement)
Connectable conductors. Only connect one conductor per terminal!
Single-wire., multiwire, fine-wire 0.2 - 2.5 mm2, AWG 26-12
Pin terminals, wire-end ferrules 0.2 - 2.5 mm
2
Tightening torque 0.4 - 0.5 Nm
Insulation length 7 mm
Connectability of the terminals (serial interface)
Single-wire., multiwire, fine-wire 0.2 - 1.5 mm2, AWG 28-16
Pin terminals, wire-end ferrules 0.2 - 1.5 mm
2
Tightening torque 0.2 - 0.25 Nm
Insulation length 7 mm
Connectability of the terminals (digital inputs/outputs, analog output)
Single-wire., multiwire, fine-wire 0.2 - 1.5 mm2, AWG 28-16
Pin terminals, wire-end ferrules 0.2 - 1.5 mm
2
Tightening torque 0.2 - 0.25 Nm
Insulation length 7 mm
79
UMG 96-PA www.janitza.de

16. 1 Function performance characteristics

Function Symbol Precision class Measurement range Display range
Total effective power P 0,55) (IEC61557-12) 0 W .. 12.6 kW 0 W .. 999 GW *
Total reactive power QA, Qv 1 (IEC61557-12) 0 var .. 16,6 kvar 0 var .. 999 Gvar *
Total apparent power SA, Sv 0,55) (IEC61557-12) 0 VA .. 12,6 kVA 0 VA .. 999 GVA *
Total effective energy Ea
Total reactive energy ErA, ErV 1 (IEC61557-12) 0 varh .. 999 Gvarh 0 varh .. 999 Gvarh *
Total apparent energy EapA, EapV
Frequency f 0.05 (IEC61557-12) 45 Hz.. 65 Hz 45.00 Hz.. 65.00 Hz
Phase current I 0.2 (IEC61557-12) 0 Arms.. 7 Arms 0 A.. 999 kA
Neutral conductor current, calculated
INc 1.0 (IEC61557-12) 0.03 A.. 25 A 0.03 A.. 999 kA
Voltage U L-N 0.2 (IEC61557-12) 10 Vrms.. 600 Vrms 0 V.. 999 kV
Voltage U L-L 0.2 (IEC61557-12) 18 Vrms .. 1040 Vrms 0 V.. 999 kV
Power factor PFA, PFV 0.5 (IEC61557-12) 0.00 .. 1.00 0.00 .. 1.00
Short-term flicker, long-term flicker Pst, Plt - - -
Voltage dips (L-N) Udip - - -
Voltage swells (L-N) Uswl - - -
Transient overvoltages Utr - - -
Voltage interruptions Uint - - -
Voltage symmetry (L-N)
Voltage symmetry (L-N)
1)
2)
Unba - - -
Unb - - -
Voltage harmonics Uh Cl. 1 (IEC61000-4-7) 1 .. 15 (only odd) 0 V.. 999 kV
Voltage THD
Voltage THD
3)
4)
THDu 1.0 (IEC61557-12) 0 % .. 999 % 0 % .. 999 %
THD-Ru - - -
Current harmonics Ih Kl. 1 (IEC61000-4-7) 1 .. 15 (only odd) 0 A.. 999 kA
Current THD
Current THD
3)
4)
THDi 1.0 (IEC61557-12) 0 % .. 999 % 0 % .. 999 %
THD-Ri - - -
Network signal voltage MSV - - -
1) Reference to amplitude.
2) Reference to phase and amplitude.
3) Reference to fundamental oscillation.
4) Reference to effective value.
5) Accuracy class 0.5 with ../5A transformer. Accuracy class 1 with ../1A transformer.
6) Accuracy class 0.5S as per IEC62053-22
5) 6)
0,5S
(IEC61557-12)
5) 6)
0.5
(IEC61557-12)
0 Wh .. 999 GWh 0 Wh .. 999 GWh *
0 VAh .. 999 GVAh 0 VAh .. 999 GVAh *
* When the maximum total energy levels have been reached, the device jumps back to 0 W .
80
www.janitza.de UMG 96-PA
81
UMG 96-PA www.janitza.de

16. 2 Modbus address list of frequently used measurement values:

Address Format RD/WR Variable Unit Comment
19000 float RD _ULN[0] V Voltage L1-N 19002 float RD _ULN[1] V Voltage L2-N 19004 float RD _ULN[2] V Voltage L3-N 19006 float RD _ULL[0] V Voltage L1-L2 19008 float RD _ULL[1] V Voltage L2-L3 19010 float RD _ULL[2] V Voltage L3-L1 19012 float RD _ILN[0] A Apparent current, L1 19014 float RD _ILN[1] A Apparent current, L2 19016 float RD _ILN[2] A Apparent current, L3 19018 float RD _I_SUM3 A Vector sum; IN=I1+I2+I3 19020 float RD _PLN[0] W Real power L1 19022 float RD _PLN[1] W Real power L2 19024 float RD _PLN[2] W Real power L3 19026 float RD _P_SUM3 W Sum; Psum3=P1+P2+P3 19028 float RD _SLN[0] VA Apparent power L1 19030 float RD _SLN[1] VA Apparent power L2 19032 float RD _SLN[2] VA Apparent power L3 19034 float RD _S_SUM3 VA Sum; Ssum3=S1+S2+S3 19036 float RD _QLN[0] var Reactive power (mains frequ.) L1 19038 float RD _QLN[1] var Reactive power (mains frequ.) L2 19040 float RD _QLN[2] var Reactive power (mains frequ.) L3 19042 float RD _Q_SUM3 var Sum; Qsum3=Q1+Q2+Q3 19044 float RD _COS_PHI[0] Fund.power factor, CosPhi; UL1 IL1 19046 float RD _COS_PHI[1] Fund.power factor, CosPhi; UL2 IL2 19048 float RD _COS_PHI[2] Fund.power factor, CosPhi; UL3 IL3 19050 float RD _FREQ Hz Measured frequency 19052 float RD _PHASE_SEQ Rotation field; 1=right, 0=none, -1=left
19054* float RD _WH_V[0] Wh Real energy L1, consumed 19056* float RD _WH_V[1] Wh Real energy L2, consumed 19058* float RD _WH_V[2] Wh Real energy L3, consumed
19060 float RD _WH_V_HT_SUML13 Wh Real energy L1..L3 19062 float RD _WH_V[0] Wh Real energy L1, consumed 19064 float RD _WH_V[1] Wh Real energy L2, consumed 19066 float RD _WH_V[2] Wh Real energy L3, consumed 19068 float RD _WH_V_HT_SUML13 Wh Real energy L1..L3, consumed, rate 1 19070 float RD _WH_Z[0] Wh Real energy L1, delivered 19072 float RD _WH_Z[1] Wh Real energy L2, delivered 19074 float RD _WH_Z[2] Wh Real energy L3, delivered 19076 float RD _WH_Z_SUML13 Wh Real energy L1..L3, delivered 19078 float RD _WH_S[0] VAh Apparent energy L1 19080 float RD _WH_S[1] VAh Apparent energy L2 19082 float RD _WH_S[2] VAh Apparent energy L3 19084 float RD _WH_S_SUML13 VAh Apparent energy L1..L3
19086* float RD _IQH[0] varh Reactive energy, inductive, L1 19088* float RD _IQH[1] varh Reactive energy, inductive, L2 19090* float RD _IQH[2] varh Reactive energy, inductive, L3
19092 float RD _IQH_SUML13 varh Reactive energy L1..L3
82
* The assignment of the marked device addresses does not correspond
to the assignment of other devices in the UMG series.
www.janitza.de UMG 96-PA
19094 float RD _IQH[0] varh Reactive energy, inductive, L1 19096 float RD _IQH[1] varh Reactive energy, inductive, L2 19098 float RD _IQH[2] varh Reactive energy, inductive, L3 19100 float RD _IQH_SUML13 varh Reactive energy L1..L3, ind. 19102 float RD _CQH[0] varh Reactive energy, capacitive, L1 19104 float RD _CQH[1] varh Reactive energy, capacitive, L2 19106 float RD _CQH[2] varh Reactive energy, capacitive, L3 19108 float RD _CQH_SUML13 varh Reactive energy L1..L3, cap. 19110 float RD _THD_ULN[0] % Harmonic, THD,U L1-N 19112 float RD _THD_ULN[1] % Harmonic, THD,U L2-N 19114 float RD _THD_ULN[2] % Harmonic, THD,U L3-N 19116 float RD _THD_ILN[0] % Harmonic, THD,I L1 19118 float RD _THD_ILN[1] % Harmonic, THD,I L2 19120 float RD _THD_ILN[2] % Harmonic, THD,I L3

16. 3 Number formats

Type Size Minimum Maximum
short 16 bit -2
ushort 16 bit 0 216 -1
int 32 bit -2
uint 32 bit 0 232 -1
float 32 bit IEEE 754 IEEE 754
15
31
215 -1
231 -1
C
saved immediately!
Configuration data are
83
UMG 96-PA www.janitza.de

16. 4 Dimension views

Fig. Front view
96
4,9
96
96
Fig. View from below
+0,8
92
Battery
87,2 6
84
87,2 6
Fig. Side view Fig. Cut-out dimensions
+0,8
92
www.janitza.de UMG 96-PA
Consumer
PE/FE

16. 5 Connection example 1

+
=
S1 S2 S3
-
17 18 19 20
Digital inputs
Supply
voltage
N/-L/+
11 12 13
N
L1
1 2 3
Functional earth
-
=
+
K3 K4
21 22 23 24 25 26
V+ 1 2 3
Digital outputs
K5
Analogue
UMG 96-PA
Voltage measurement
V1 V2 V3 VN
1 2 3 4
2)1) 2) 2)
PE
N
L1
L2
L3
230V/400V 50Hz
V+
output
-
=
+
A
B
Data GND
14
C
15 16
A B
RS485
Current measurement
I1 I2 I3
S1 S2 S1 S2 S1 S2
5 6 7 8 9 10
3) 3) 3)
S2
S1
S1S1S2
S2
1)
UL/IEC allowed overcurrent
circuit breaker
2)
UL/IEC allowed overvoltage
circuit breaker
3)
Shorting bridges (external)
85
Loading...