Janitza UMG96L User guide

Universal Measuring Device
L1-N / L1-L2 L2-N / L2-L3 L3-N / L3-L1
UMG96L
Operating Instructions
Brief instructions see last page
Mean value
Programming mode Sum measurement Phase to phase
Password Voltage transformer Current transformer
Peak value Minimum value
Dok Nr. 1.031.014.0
Supply
Janitza electronics GmbH
Vor dem Polstück 1 D-35633 Lahnau Support Tel. (0 64 41) 9642-22 Fax (0 64 41) 9642-30 e-mail: info@janitza.de Internet: http://www.janitza.de
Table of contents
The meaning of symbols 3 Hints for the user 3 Receipt control 3
Contents of delivery 3
Hints for maintenance 4
Service 4
Product description 5
Intended use 5 Functional description 5
Installation 6
Mounting place 6 Measuring and operating voltage 6 Current measurement 7 Sum current measurement 7 Connecting options 8
Putting into service 10
Installation 10 Attach measuring and operating voltage 10 Connect measuring current 11 Check phase assignment 11 Check current direction 11 Check measurement 11 Check phase power 11 Check sum power 11
Elimination of errors 12 Error messages 13
Usage and indication 14
Indication mode 14 Programming mode 14 Key functions 15 Table, measured value indication 16 Current transformer 18 Voltage transformer 19 Averaging times 20 Measured value indication 21
Measured value rotation time 21
Measured value selection 22 Delete minimum and maximum values 23 Delete energy 23 Working hours counter 24 LCD contrast 24 Software Release 25 User password 25
Uncertainty of measurement 26 Configuration data 28 Declaration of conformity 29
Safety guidelines 29 Test voltage 29 EMC requirements 29
Technical data 30
Ambient conditions 30 Measurement 30 Connectable cables 30 Dimensions 31 Connection example 31
Brief instructions 32
Change current transformer ratio 32 Call up measured values 32
All rights reserved. No part of this manual may be reproduced or duplicated without the written permission of the author. Any contraventions are punishable and will be prosecuted with all legal means. No liability can be taken for the faultless condi­tion of the manual or damage caused by the use of it. As failures cannot be avoided com­pletely, we shall be very grateful for any advice. We are anxious to remove all failures as soon as possible. The mentioned software and hard­ware descriptions are registered trademarks in the most cases and are subject to the regula­tions by law. All registered trademarks are prop­erty of the corresponding companies and are fully recognized by us.
Page 2
Issue Note
13.01.2004 First Edition.
The meaning of symbols
Receipt control
The symbols, which are used in this manual, have the following meaning:
Beware of dangerous electrical voltage.
Ꮨ 
Hints for the user
This device may be inserted and used by quali­fied personnel only according to the safety regulations. Please follow the legal and safety regulations for the corresponding application while using the device. Qualified personnel are persons, who are famil­iar with installation, mounting, putting into serv­ice and operation of the product and have quali­fications according to their occupation, for ex­ample:
- Education or instruction resp. the right to switch on or off, ground or characterize current circuits or devices according to the standards of safety techniques.
- Education or instruction in care and use of safety equipment according to the standard safety techniques.
This symbol shall warn you about possi­ble danger, which can occur during instal­lation, putting into service and usage.
In order to ensure a perfect and safe use of the device, a proper transport, expert storage, erec­tion and mounting and careful usage and main­tenance is required. In case that a safe opera­tion can no longer be granted, the device has to be put out of service and has to be protected against unintentional putting into service. A safe operation can no longer be assumed, if the device
• shows visible damage,
• does not work in spite of intact net supply,
• has been exposed to disadvantageous condi­tions for a longer time (e.g. storage beyond al­lowed climate without adaption to the room cli­mate, dew etc.) or transport use (e.g. falling from great height, even without visible dam­age).
Please test the contents of delivery for comple­tion, before starting the installation of the de­vice.
Contents of delivery
1 pc. UMG96L, 2 pcs. fixing brackets, 1 pc. manual. As an option, a seal with part no. 29.01.907 can be delivered.
All delivered options and versions are listed on the delivery papers.
Attention!
If the device is not used according to this manual, a safe use cannot be granted, and the instrument might cause danger.
Attention!
If a UMG 96L is installed in panels of metal, the panel must be earthed.
This manual also describes options, which have not been delivered and do not belong to the contents of delivery.
Page 3
Hints for maintenance
Before delivery the device is tested in various safety checks and marked with a seal. If the device is opened, these checks must be re­peated. For instruments, which are opened outside the manufacturing works, no warranty is granted whatsoever.
Repairing and calibration
Repairing and calibration work can be carried out in the manufacturing works only.
Front foil
The cleaning of the front foil has to be effected with a soft cloth using a common cleansing agent. Acid or acidic agents are not allowed for cleaning.
Service
If there are questions, which are not described in this manual, please contact us directly. For an efficient handling we need the following information:
- Description of device (see type plate),
- Serial digit (see type plate),
- Software Release,
- Measuring and operating voltage and
- detailed description of error.
You can reach us: Monday until Thursday
between 07:00 and 15:00
and Friday
between 07:00 and 12:00
Waste management
The UMG96L can be disposed and recycled as electronical waste according to the legal regu­lations.
Janitza electronics GmbH Vor dem Polstück 1 D-35633 Lahnau Support: Tel. (0 64 41) 9642-22 Fax (0 64 41) 9642-30 e-mail: info@janitza.de Internet: http://www.janitza.de
Page 4
Product description
Intended use
The UMG96L is suitable for fix mounting and the measurement of voltage, current, power etc. in low voltage switchgear. The measure­ment is designed for 3 phase systems with neu­tral conductor (TN and TT-mains). The UMG96L is suitable for mounting in fix and weather protected panels. Conducting panels must be earthed. Due to the high resistance against interference, the UMG 96L is suitable for continous and un­supervised operation.
As the UMG96L receives its power from phase L1 of the measuring voltage, at least phase L1 and the neutral conductor N must be con­nected. The attached voltage must be within the range of the measuring and operation voltage shown on type plate. Either ../5A or ../1A current transformers can be connected to the current measuring inputs. The connection of the measuring and operating voltage is carried out on the back side of the UMG96L via all-insulated spring power clamps.
Functional description
The electronical three phase measurement sys­tem determines and digitalizes the effective val­ues of voltages and currents in 50/60 Hz net­works. The operating voltage for operation of UMG96L is generated by the measuring voltage L1-N. A random test measurement is carried out each second at all current and voltage inputs. Meas­uring signal interruptions, which last longer than one second are safely recognized. For each random test one period is scanned. From those sampled values the microprocessor calculates the electric quantities. These meas­ured values are indicated within the program­mable display. The energy as well as the mini­mum and maximum values are stored every 15 minutes whereas the programmed data are stored immedately is a none volatile memory (EEPROM). The scanning frequency for all measuring in­puts is calculated from the net frequency of phase 1. At a net frequency of 50Hz the scan­ning frequency is 2,5kHz and for 60Hz it is 3,0kHz.
The measuring and operating voltages must be connected to the UMG 96L via a separation (switch or power switch) and an overcurrent protection fuse (2-10A) in the building installa­tion. The separation (switch or power switch) must be near the UMG 96L and easily accessi­ble.
L1
L2
L3
N
PE
3M
3M
Voltage measurement
Generation of oper-
ating voltage
3M
UMG96L
Diagr. Generation of the operating voltage from measurement voltage.
Page 5
Installation
Mounting place
The UMG96L is suitable for a fix insertion into low and medium voltage switchgear. Any mounting position is possible.
Measuring and operating voltage
The operating voltage of the UMG96L is gener­ated by the measuring voltage. The measure­ment is designed for three phase systems with neutral conductor (TN and TT mains). The measuring and operating voltages must be con­nected to the UMG 96L via a separation (switch or power switch) and an overcurrent protection (2-10A) within the building installation.The con­nection of the measuring and operating voltages is carried out at the back side of the UMG 96L via shock protected spring clamps.
230V/400V Standard version
Phase L1 and the neutral conductor N must be connected, and the attached voltage must be within the range of the measuring and operat­ing voltage.
- The connection wires for the operating volta­ge leading to the UMG96L must be suitable for voltages up to 300V against ground.
- The measuring and operating voltage has to be protected by a fuse, which must be in the range of 2A to 10A.
- A switch or power switch for the operating vol­tage must be provided within the building instal­lation.
- The switch has to be near the instrument and easily accessible.
- The switch must be marked as separation for this instrument.
120V/220V Special version
Phase L1 and the neutral conductor N must be connected, and the attached voltage must be within the range of the measuring and operat­ing voltage.
60V/110V Special version
Phase L1 and the neutral conductor N must be connected, and the attached voltage must be within the range of the measuring and operat­ing voltage.
Attention!
The limits indicated in the technical data may not be exceeded, not even while checking and putting into service of the UMG 96L.
Attention!
Before the device is connected to volta­ge for the first time, it should be depos­ited the installation room for at least 2 hours, to create a temperature assimila­tion and avoid condensation and dew.
Page 6
Current measurement
The current measurement is carried out via ../ 5A or ../1A current transformers. In case that the current has to be measured by an Ampere­meter in addition to the UMG96L, the Ampere­meter has to be connected in series to the UMG96L. In mains with voltage up to 300VAC against ground, currents up to 5A can be connected and measured directly to the UMG 96L.
Verbraucher
Consumer
L l
K k
A
Einspeisung
Supply
Diagr. Connection example: UMG96L with Amperemeter in series.
l
UMG96L
k
Sum current measurement
If the current measurement is carried out via two current transformers, the total transformer ratio must be set to the UMG96L.
Example: Sum current transformer A current measurement is carried out via one current transformer with a ratio of 1000/5A and another one with a ratio of 200/5A. The sum measurement is carried out with a sum trans­former 5+5/5A. The UMG96L has to be programmed as fol­lows:
Primary current: 1000A + 200A = 1200A Secondary current: 5A
Verbraucher 1
Consumer 1
L l
Verbraucher 1
Consumer 1
L l
Inaccuracy
The inaccuracy of the current measuring input is +- 1% of the measuring range (5A). There­fore, the inaccuracy of the current measure­ment is +-50mA. Example: Inaccuracy With a current transformer 200/5A, the measur­ing range is 200A. The inaccuracy is +-1% of 200A = +-2A.
Resolution
The maximum resolution of the current meas­uring inputs is 10mA. The indication changes in
0.01A steps. Example: Resolution current transformer 200/ 5A With a current transformer ratio of 200/5A, a resolution of 10mA*40 = 400mA is effected. The display changes in 0.4A steps.
Small currents
The minimum working current is 20mA. At short circuited or open current measurement input, the UMG 96L can indicate a small current. This current is within the range of the allowed meas­urement inaccuracy.
K k
Einspeisung 1
Supply 1
AK AL
K l
k l
UMG96L
BK Bl
K k
Einspeisung 1
Supply 1
Attention!
Current transformers may not be handled in open condition of the secondary, as the secondary clamps can lead live vol­tage.
Attention!
Prior to connecting or exchanging the UMG 96L, the secondary clamps of the external current transformers have to be short circuited.
Attention!
The program allows current and voltage transformer ratio setting only, if the maxi­mum phase power can reach up to
50.0MW and the sum power can reach a maximum value of 150.0MW.
Page 7
Connecting options
UMG96L
Voltage Measure-
ment
see type plate
1234 5678910
L1 L2 L3 N
L1 L2 L3 PEN
Diagr.: Connection example 1 Four wire measurement with three current transformers
Current
measurement
0,02 .. 5A
1k 1l 2k 2l 3k 3l
kl
k
../5(1)A
../5(1)A
l
../5(1)A
k
l
Consumer
Verbraucher
UMG96L
Voltage Measure-
ment
see type plate
1234 5678910
L1 L2 L3 N
L1 L2 L3 PEN
Diagr: Connection example 2. Four wire meas­urement with two current transformers.
Current
measurement
0,02 .. 5A
1k 1l 2k 2l 3k 3l
kl
../5(1)A
../5(1)A
k
l
Consumer
Verbraucher
UMG96L
Voltage Measure-
ment
see type plate
1234 5678910
L1 L2 L3 N
uu
u
xx
x X
XX
L1 L2 L3 PEN
UUU
Current
measurement
0,02 .. 5A
1k 1l 2k 2l 3k 3l
kl
k
../5(1)A
../5(1)A
l
../5(1)A
UMG96L
Voltage Measure-
ment
see type plate
1234 5678910
L1 L2 L3 N
uu
u
xx
x
X
XX
L1 L2
k
l
Verbraucher
Consumer
L3 PEN
UUU
Current
measurement
0,02 .. 5A
1k 1l 2k 2l 3k 3l
kl
../5(1)A
../5(1)A
k
l
Consumer
Verbraucher
Diagr.: Connection example 3 Measurement with three voltage transformers and three current transformers.
Page 8
Diagr.: Connection example 4 Measurement with three voltage transformers and two current transformers.
UMG96L
UMG96L
Voltage Measure-
ment
see type plate
1234 5678910
L1 L2 L3 N
uu
u
x X
XX
L1 L2 L3
UUU
xx
Current
measurement
0,02 .. 5A
1k 1l 2k 2l 3k 3l
kl
../5(1)A
k
../5(1)A
../5(1)A
l
k
Consumer
Verbraucher
Diagr.: Connection example 5 Medium voltage measurement with three volta­ge transformers and three current transformers.
Voltage Measure-
ment
see type plate
1234 5678910
L1 L2 L3 N
uu
u
x X
XX
L1 L2 L3
UUU
xx
Current
measurement
0,02 .. 5A
1k 1l 2k 2l 3k 3l
kl
../5(1)A
../5(1)A
kl
Consumer
Verbraucher
Diagr.: Connection example 6 Medium voltage measurement with three volta­ge transformers and two current transformers.
UMG96L
Voltage Measure-
ment
see type plate
1234 5678910
L1 L2 L3 N 1k 1l 2k 2l 3k 3l
L1 L2 L3 N
PE
Current
measurement
0,02 .. 5A
kl
../5(1)A
k
../5(1)A
../5(1)A
UMG96L
Voltage Measure-
ment
see type plate
1234 5678910
L1 L2 L3 N
uu
u
x
xx
X
XX
l
k
l
Consumer
Verbraucher
L1 L2 L3
PE
UUU
Current
measurement
0,02 .. 5A
1k 1l 2k 2l 3k 3l
kl
k
../5(1)A
../5(1)A
l
../5(1)A
k
l
Consumer
Verbraucher
Diagr.: Connection example 7 Measurement in IT networks via three current transformers.
Diagr.: Connection example 8 Measurement in IT networks with three volta- ge transformers and three current transformers.
Page 9
Putting into service
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
The putting into service of the UMG 96L should be carried out as follows:
Installation
The UMG96L is designed for mounting in low voltage distributions, which contain overvoltage in measurement category III. The UMG96L is suitable for installation in fixed and weather proof panels. Conducting panels must be earthed. Any mounting position is possible. For mounting on front panels or doors, the delivered fixing brackets have to be used.
Panel max. 6
Fixing brack-
ets
90
42
Diagr. Side view
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
23
96
6
Attach measuring and operating voltage
The allowed measuring and operating voltage of UMG 96L is indicated on the type plate.
Attention!
Mesuring and operating voltage, which do not correspond to the indication on the type plate, can lead to malfunction and damage of the instrument.
The connection wires for measuring voltage leading to the UMG96L must be suitable for voltages up to 300V against ground and 520V phase to phase. After connecting the stated measuring and op­erating voltage to the UMG 96L, all segments of the display appear. After approx. two sec­onds, the UMG 96L switches to the first meas­ured value indication. In case that no indication appears, please check, if the attached operating voltage is within the rated voltage range.
S V J M A D E P
UMG96L
45-65Hz IP20
2,5VA
Made in Germany
4
N
Spannungsmessung
Voltage Measurement
1
2
3
L3
L1
L2
2-10A
Strommessung/Current Measurement
k l
L2 L3L1
k l
k l
57 10896
Program current and voltage transformers
When the device leaves the manufacturer, a current transformer of 5/5A is set. The set voltage transformer ratio has to be changed only, if a voltage transformer is con­nected. When connecting voltage transformers, please observe the allowed measuring and operating voltage stated on type plate!
Page 10
Loading...
+ 22 hidden pages