Electronic Scale
1
Weighing Pan Fixing Steps
1. When installing the weighing pan, tighten the fixing screw in the centre of the plastic weighing pan.
2. Unscrew and remove the shipping protection screw on the side of the scale.
3. After removing the shipping protection screw on the side of the scale, fasten the shipping protection buckle.
Protection Device Disassembly Steps
1. In the upright position, locate the shipping protection screw device on the left side of the scale.
2. Before using the scale, unscrew counterclockwise and remove the shipping protection screw.
3. Before transporting the scale, hold the shipping protection screw down inwards and tighten it clockwise.
Precautions
1. Please pay attention to the environment when performing weight calibration on the scale. Avoid wind blow and vibration.
2. Use of wrong batteries or wrong wiring may cause danger.
3. 10 minutes warm up till the RED light off before activating the scale.
Electronic Scale
2
ON/OFF:
The on/off key for switching on/off the equipment.
ZERO:
The zero key for resetting the equipment, functional only when weight is within 4% of the weighing capacity.
TARE:
The tare key for deducting the weight of container on the weighing pan.
MODE:
The switching key to select weighing, counting and percentage modes in turn.
UNIT:
Weighing units switching key. (13 kinds of units)
SMPL:
In the weighing mode, the key is for switching on/off backlight; in the counting and percentage modes, the key is for sampling, by pressing which, you may select sample size of 20, 50, 100, 200 or 500 in turn.
Descriptions of Keys
Functional Operation of Keys
Weighing:
PressON/OFFkey, the screen will have full display and start counting down. When the screen displays
0.00 X , you can start weighing. If you need backlight, pressSMPLkey.
Counting:
PressMODEkey, the screen will show XXXX pcs, then press SMLP key, and the screen will show S= 20pcs. You may select sample size of 20, 50, 100, 200 or 500 in turn by pressingSMLPin succession. After selection, put the sample on the weighing pan and start counting operation when you hear a beep sound.
(Automatic averaging function is provided for simple counting)
Percentage:
PressMODEkey, the screen will show XXXX % , then press SMLPkey, and the screen will show S= 20 % . You may select sample size of 20, 50, 100, 200 or 500 in turn by pressingSMLPin succession. After selection, put the sample on the weighing pan and start using the percentage function when you hear a
beep sound. Notes:
1. Min. unit weight of sample >=1 readability (e.g.: unit weight of sample of 300g is at least bigger than 0.01g)
2. When weight is unstable, the STABLE indicator will go off.
3. After warming up the machine, the WARMING indicator will go off.
Electronic Scale
3
1
Level
6
Stable indicator
2
Warming up indicator
7
Tare key
3
Sampling/Backlight key
8
Zero key
4
Function selection key
9
Weight display window
5
Unit conversion key
Single Point Calibration of
Weight
Step 1: Power off while holding down MODEkey, and then press 「ON/OFF」to switch on the scale. The screen will show the following:
Count
Step 2: Press the MODEkey, the screen will show the following:
CAL
Step 3: Press the UNITkey to start zero point calibration. Make sure there is nothing on
the weighing pan. The screen will show the following:
ZEro
Please wait for the screen to show:
Panel Illustration
Electronic Scale
4
CAL-S
The calibration can be done at the point of 1/3 load, 2/3 load or full load. Choose the one you need and place the corresponding weight onto the pan.
If place 1/3 of load onto the pan, the
screen will show:
on1
If place 2/3 of load onto the pan, the
screen will show:
on2
If place full load onto the pan, the
screen will show:
on3
Step 4: Place weights of equivalent weight onto the pan, wait for a beep sound and the
screen will show:
PASS
Step 5: Press MODEand then press SMPL, the screen will show:
busy
Step 6: After countdown, the single point calibration is completed and the scale returns to
the normal weighing mode.
Linearity Calibration of Weight
Step 1: Power off while holding down 「ZERO」and 「TARE」keys, and then press 「ON/OFF」
to switch on the scale. The screen will show the following:
L-CAL
Step 2: Clear the weighing pan, then press TAREand the screen will show the following:
(zero point calibration)
on0
Step 3: Wait for a beep sound and the screen will show the following:
on1
Step 4: Place 1/3 of load on the weighing pan in ten seconds ,Waiting for a beep sound and the screen will show the following:
on2
Step 5: Place 2/3 of load on the weighing pan in ten seconds ,Waiting for a beep sound and the screen will show the following:
on3
Step 6: Place full load on the weighing pan in ten seconds, Waiting for a beep sound and the screen will show the following:
PASS
Step 7: Clear the weighing pan, then press TAREand the screen will show the following:
busy
Electronic Scale
5
Step 8: After countdown, the linearity calibration is completed and the scale returns to the normal weighing mode.
LCD Display
Function
Selection
Selection
Page
MODEkey
UNIT key
SMPLkey
Count
Internal value
display
Item 1, Page 6
CAL
Single point auto
calibration
Page 3
unit
Weight unit
selection
g、ct、lb、oz、dr、GN、ozt、dwt、MM、tl.j、tl.T(g/m2) tl.Ht(oz/yd2) ()for plaiting purpose
ON/OFF
Item 2, Page 6
InitU
Initial weight unit
selection
g、ct、lb、oz、dr、GN、ozt、dwt、MM、tl.j、tl.T(g/m2) tl.Ht(oz/yd2) ()for plaiting purpose
Item 3, Page 6
Auto
Auto power-off
selection
Au-no/5/15/30/60
Item 4, Page 6
bAud
Baud rate
selection
2400/4800/9600
Item 5, Page 6
ZEro
Auto zero range
selection
d0/1/2/3/4/5
Item 6, Page 6
Fil
Filter selection
Fil1/2/4/8
Item 7, Page 6
LigH
Backlight selection
onoFF/oFF/ON
Item 8, Page 6
Print
Printing mode
selection
Pr-St/Co
Item 9, Page 7
Parameter Setting
Hold down 「MODE」 key, then press 「ON/OFF」 key to enter the setting mode. The screen will show Count :
MODEkey: To select functions; UNITkey: To enter or exit function setting selection; SMPLkey: To end setting.
Electronic Scale
6
Pr Fu
External device
PC/ SH-24
Item 10, Page 7
diV-Z
Display range
Zero0/1/2/3/4/5
Item 11, Page 7
1. When the LCD display shows Count, press theUNITkey, the current internal value will be displayed. Press UNITto return to LCD display Count.
2. When the LCD display shows unit , pressUNIT」 to enter into unit selection mode, the screen will show on g or oFF g , on means to use the current unit displayed, oFF means not to use the current unit displayed. PressSMPL」 to switch between ON and OFF [to use the unit displayed select on, if you do not require the unit displayed select oFF]. PressUNITto select from 13 kinds of weighing units, after selecting the unit required, pressMODEto return to LCD display unit .
3. When the LCD display shows InitU , pressUNIT to enter initial unit selection mode. LCD display InitU xx X means the initial unit is selected. PressUNIT to select the unit required, then PressMODEto return to LCD display InitU.
4. When the LCD display shows Auto , pressUNIT to enter the Auto Power-Off selection. LCD display XXX X means Auto Power-off selection. Press UNIT to select the auto power-off time, then pressMODE」to return to LCD display Auto .
Au-no- means No auto power-off, 5- means 5 minutes, 15-means 15 minutes,
30-means 30 minutes, 60-means 60 minutes(in case of instability, the auto power-off
function will not work)
5. When the LCD display shows bAud , press UNIT to enter the RS-232 Baud Rate selection. LCD display XXX X represents the RS-232 baud rate. Press UNITto select the desired RS-232 baud rate, then pressMODE to return to LCD display bAud . 『2400, 4800, 9600』
6. When the LCD display shows ZEro , pressUNIT to enter auto zero range selection. LCD display d X X represents the auto zero range. PressUNITto select the desired auto zero range, then pressMODE to return to LCD display ZEro 『d0, 1, 2, 3, 4, 5』 (the higher the value is, the wider auto zero range is)
7. When the LCD display shows FiL , pressUNIT to enter filter setting mode. LCD display FiL X X means the filter level. PressUNIT to select the desired filter level and pressMODE to return to LCD display FiL . 『FiL1, 2, 4, 8』 (the higher the value is, the higher the filter level is)
8. When the LCD display shows LigH , pressUNIT」to enter backlight setting mode.
Electronic Scale
7
The LCD display XXXX X represents the backlight setting. PressUNIT」to select
1
ct
[MET.CARAT]
=0.1999694 g
1
lb
[AVORIRDUPOIS POUND]
=453.59237 g
1
oz
[AVORIRDUPOIS OUNCE]
=28.349523125 g
1
dr
[AVOIRDUPOIS DRAM]
=1.7718451 g
1
GN
[GRAIN](U.K)
=0.06479891 g
1
ozt
[TROY OUNCE]
=31.1034768 g
1
dwt
[PENNY WEIGHT]
=1.55517384 g
1
MM
[MOMME] (JPN)
=3.749996 g
1
tl.j
[HONG KONG JEWELRY TAEL]
=37.4290018 g
1
tl.T
[TAEL](TWN)
=37.49995 g
1
tl.H
[HONG KONG TAEL]
=37.799375 g
1
t
[TOLA] (INDIA)
=11.6638038 g
1
oz/yd2
=0.33898306 g
1
g/m2
=0.01 g
the desired backlight setting and pressMODEto return to LCD display LigH
onoFF- means auto-backlight, oFF-means backlight off, ON-means backlight on
9. When the LCD display shows Print , pressUNIT to enter printing mode selection. LCD display Pr-XX X represents the printing mode. PressUNITto select the desired printing mode and pressMODE to return to LCD display Print . 『Pr-Co-means continuous output, Pr-St –means stable output
10. When the LCD display shows Pr Fu , pressUNIT to enter external device selection mode. LCD display XXXX X represents external device. PressUNIT to select the desired external device, then pressMODE to return to LCD display Pr Fu . 『-PC external PC, SH-24 – external dot matrix mini printer
11. When the LCD display shows diV-Z , press UNIT to enter external device selection mode. LCD display ZEroX X represents display range. PressUNIT」to select the desired external device, then pressMODEto return to LCD display diV-Z 『Zero0, 1, 2, 3, 4, 5』(means weight displayed when the weight exceeds the capacity)
12. Before entering function setting, press SMPLto end function setting and return to normal weighing mode. Press theMODEkey continuously to select function setting modes.
Units Conversion Table
Electronic Scale
8
CAP
150
300
600
1500
3000
Gram
150X0.005
300×0.01
600×0.02
1500×0.05
3000×0.1
Carat
750.34X0.02
1500.70X0.05
3001.36X0.1
7503.4X0.2
15007.0X0.5
Pound(Avoir)
0.33080X0.00002
0.66160X0.00005
1.323X0.0001
3.3080X0.0002
6.6160X0.0005
Ounce(Avoir)
5.2926X0.0002
10.5855X0.0005
21.171X0.001
52.926X0.002
105.855X0.005
Dram(Avoir)
84.685X0.005
169.37X0.01
338.72X0.02
846.85X0.05
1693.7X0.1
Grain
2315.5X0.1
4631.0X0.2
9262.0X0.5
23155X1
46310X2
Ounce(Troy)
4.8240X0.0002
9.6480X0.0005
19.296X0.001
48.240X0.002
96.480X0.005
Penny weight
96.480X0.005
192.96X0.01
385.92X0.02
964.80X0.05
1929.6X0.1
Momme
40.012X0.002
80.025X0.005
160.05X0.01
400.12X0.02
800.25X0.05
Tael (HK Jewelry)
4.0088X0.0002
8.0175X0.0005
16.035X0.001
40.088X0.002
80.175X0.005
Tael (Taiwan)
4.0012X0.0002
8.0025X0.0005
16.005X0.001
40.012X0.002
80.025X0.005
Tael (Hong Kong)
3.9696X0.0002
7.9390X0.0005
15.878X0.001
39.696X0.002
79.390X0.005
Tola (India)
12.865X0.0005
25.728X0.001
51.456X0.002
128.640X0.005
257.28X0.01
oz/yd
2
442.49X0.01
884.99X0.01
1770.27 X0.01
g/m
2
15000.5 X0.5
30000 X1
60000 X1
Cross-reference Table of Full Loads
Note: Full Load = Capacity + (Readabilityx9)
RS-232 Serial Interface
The 9PIN connector at the right back is a RS-232 standard interface. Pin 3 is for output, Pin 5 is for grounding and the rest have no functions.
BAUD RATE2400 OR 4800 OR 9600 bps DATA BIT 8 PARITY BITNNONE STOP BIT1 CODEASCII
Electronic Scale
9
BIT FORMAT
0 1 2 3 4 5 6 7
S/U T G W +/-
LF
CR
SP
SP
WEIGHT
SP
UNIT
LSB MSB
START BIT DATA BIT STOP BIT
DATA FORMAT
ST:STABLE UT:UNSTABLE GW:GROSS WEIGHT SP:SPACE
Examples: ST GW + 150.00 g UT NW - 80.00 ct …………..
PC Program: 10 OPENCOM1:9600,N,8,1,RS,DS,LFAS #1
(BASICA) 20 INPUT #1,A$ 30 PRINT A$ 40 GOTO 20 50 END
Electronic Scale