Jadever JWI 700C Instruction Manual

JWI-700C 0
Table of Contents
………………………………………………………………………………
1
2. Precaution…………………………………………………….......................................
1
3. Product Introduction
3-1 Specifications & Features………………………………………………………………....
3-2 Front Panel 3-2-1 Display …………………………………………………………………………………… 3-2-2 Keyboard…………………………………………………………………………………. 3-3 Rear Panel………………………………………………………………………………….
3-4 Power Supply……………………………………………………………………………….
1
2 3 4 5
4. Installation
4-1 Load Cell Connection………………………………………………………………………
4-2 Assembly Description of Upright Pole……………………………………………………
5 5
5. Setting Mode
5-1 Maximum Weighing Capacity Setting…………………………………………………… 5-2 Resolution Setting ………………………………………….………………………………
5-3 Function Setting …………………………………………………………………………..
7 7 8
6. Calibration Mode
6-1 Linear calibration…………………………………………………………………………..
6-2 Single point calibration…………………………………………………………………….
10 10
7. Operation
7-1 Basic Weighing 7-1-1 Weighing…………………………………………………………………………………. 7-1-2 Tare & Preset T are………………………………………………………………………. 7-1-3 Unit Switch Operation …………………………………………………………………. 7-2 Basic Counting 7-2-1 Upper & Lower Quantity Limit Checking ……………………………………………... 7-2-2 Entering a Known Unit Weight …………….…………………………………………..
7-2-3 Sample Counting & ACAI……………………………………………..........................
7-2-4 Accumulation, Accumul ation Display and Accumulation Clear……………………..
11 11 12
12 12 13
13
8. Serial Interface
8-1 RS-232 Connector…………………………………………………………………………. 8-2 Single Option………………………………………………………………………………. 8-3 RS-232 Output Format…………………………………………………………………….
14 15 15
9. Troubleshooting and Error Message…………………………………………...
16
JWI-700C 1
1. Introduction
Thank you for deciding to purchase a JWI-700C Counting indicator from Jadever. This goods has the excellent performance and splendid properties under severe quality management .It
is recommended to read this manual in full before using it for good fu nct i on appl ication.
2. Precautions
Place the indicator on a flat and stable surface . Verify that the input voltage and the plug type matches the local AC power supply . See
3-4 power supply.
Keep the indicator away from EMI noise , strong wind and vibration ,which might cause
incorrect reading.
Avoid sudden temperature changes (suitable operating temperature is between-5~
40℃.)
Disconnect the power supply when cleaning the indicator . Do not immerse the indicator in water or other liquids .
3. Product Introduction 3-1 Specifications & Features Specifications
Model JWI 700C
Input sensitivity 0.13uV/D
Input voltage range -0.5mV to 16.5mV Load cell excitation DC 5V ,Up to 8 ×350 ohm load cells
System linearity 0.003% of full scale Input impedance 10M ohm or more AD conversion mode -Σ
A/D internal resolution 700,000 count AD conversion speed 8 times/second External display resolution. 15,000 count Displays LCD ( 6 Digits 5 Digits6 Digits) Power supply AC 110V/220V (AC±
10%) or Rechargeable
battery 6V/4A
JWI-700C 2
Features
LCD display 6/5/6, LED backlight Kg and Lb switchable Up to 1/30000 resolution Adjustable capacities, resolutions and p ar a met er s Adjustable stand for bench scale Enclosed with PVC dustproof cover Able to adjust filter range and zero band Low battery and charging status indicator Automatic average weighing unit function Single point and linear calibration available HI/OK/LO indicator, HI/OK/LO alarm function Full range tare, pre-tare, auto-zero tracking Rechargeable battery or AC power Automatic counting accuracy improvement (ACAI) Accumulation of weight and quantity (max 99 pieces) Auto power off selection, save battery life Suitable for a wide range of bases and load cells
3-2 Front Panel 3-2-1Display
1) Low battery indication
2) Tare or Preset T are indication
JWI-700C 3
3) Center of Zero Indication, The zeroing range is ±2of weighing capacity.
4) Charge Lamp
5) HI Lamp (ON) –Indicating parameter item is selected when setting buzz sounds. (See 5. Function Setting, setting buzz sounds)
OK Lamp (ON) -- Indicating parameter item or is selected when setting buzz sounds.
LOW Lamp (ON)--Indicating parameter item is selected when setting buzz sounds.
6) “PRESET TARE” Symbol “ ” points at PRESET TARE” when Tare value entered via
numeric keys.
7) “NET Net weight--Gross weight minus Tare. Symbol “ ” points at NET when TARE or
PRESET TARE action is done.
8) “STABLEStable indication, Symbol ” appears at the top of STABLEwhen
scale is in stable condition.
9) “ACAISymbol “ ” points at “ACAI” when Automatic counting accuracy improvement is
functioning.
10) M+ Symbol “ ” points at “M+” when scale is in accumulation mode.
11) “UW” Under entering a known unit we ight mode, Symbol “ ” points at “UW” when
entered unit weight is lower than 4/5 of scale division. Unit weight is too small for ensuring accurate quantity calculations.
12) “SMPLUnder the sample counting mode, Symbol “ ” points at “SMPL” when
calculated unit weight is lower than 4/5 of scale division or the quantity of sample is less than 10 pieces.
13) “UNIT WEIGHT” Unit Weight Window --Displays the average unit weight value, or
number of weighments.
14) “TOTAL COUNT” Total count window –Displays the calculated number of pieces on the
weighing pan, or accumulated piece count value.
15) “WEIGHT” Weight Window--Displays weight of object on weighing pan, or accumulated
weight value.
3-2-2 Keyboard
1) 0 ~9 numeric keys Enters specific values for tare; unit weight and other data entries.
JWI-700C 4
2) key Decimal point ; toggles betw een up per limit and lower limit sett ing when HI-LO-OK function is enabled .(see 7-2-1)
3) PRINT key Outputs data when Prt.Pr (print stably) is selected as the print mode.
4) ZERO key Zeros the display (within 2% of max.capacity) or cancels Tare action.
5) TARE key Inputs the weight of the object on the weighing pan as a Tare value ; Inputs the indicated value entered via the numeric keys as a pre-set Tare value ; cancel tare
action.
6) CE key Clears the indicated input values.
7) MC/CK key Short press deletes accumulation records; long press to initial Upper
& Lower Quantity Limit setting.
8) MR key Recalls and displays the total Accumulation data (weight, total count and
number of weightments ) and the first 10 records in details.
9) UWS key Inputs the indicated unit weight value entered via numeric keys
10) M+ key Adds the indicated weight or piece count value into Accumulation
memory.
11) SMPL key Short press enters the sampling mode.
3-3 Rear Panel
1) Port for connecting load cell.
2) power socket
3) power ON/OFF switch
4) Two-stage switch(110V or 220V)
5) RS-232 port: Serial inter face por t (computer, printer and Light Tower)
JWI-700C 5
3-4 Power supply
Please verify the local AC power source and switch the two-stage switch to the proper place before plugging into the power outlet.
Alternative power supply
1) AC 110V/220V (AC±10%)
2) (6V/4A) Internal Rechargeable Battery
Power Consumption
About 300mW, 80hrs (without backlight) About 380 mW, 65hrs (with backlight)
Low battery warning
When “ ” appears in the upper left corner of the weight window, the battery power requires recharg i ng . The charge lam p t ur ns green from red when the recharging is completed
(which takes about 8 hours). Disconnect the scale from power supply when it is fully charged. Note: Battery is to replaced only by an authorized service dealer .Risk of explosion can occur if replaced with the wrong type or connected improperly.
4. Installation
4-1Load Cell Connection
CONN
1 2
3
4
5
E+ E­S+ S-
LOAD-CELL
Shield
LOAD CELL
CONNECTION
PIN
SIGNAL
1
E+ 2 E- 3 S+ 4 S-
5
SHIELD
4-2 Assembly Description of Upright Pole
(1) Rod seat (2) Upright pole (3) Bracket (4) Indicator (5) Screw (for fixing the upright pole) (6) Screw (for fixing the Bracket) (7) Knob pole (8) Bracket slot (9) Load cell wire
JWI-700C 6
Step 1: Thread the wire of the Load Cell (9)on the rod seat (1) through the upright pole (2), insert the upright pole into the rod seat and then lock it with two screws (5).
Step 2: After threading the Load Cell wire
through the bracket (3), attach the bracket to the upright pole and then lock it with the screw (6).
Note: if the load cell connector is too big to thread throug h the bracket, separate the
bracket by removing the Knob pole (7), see the following pictures.
Step 3: Install the Indicator (4) on the bracket,
Step 4: After connecting load cell connector to load cell port, the installation is completed.
Note: Use the knob pole (7) to adjust the inclination angle of indicator and the screw (6) to adjust direction of the indicator. After adjusting the indicator to an optimal position and lock the screw.
JWI-700C 7
5. Setting Mode
5-1 Maximum Weighing Capacity Setting
1) Turn on the power while pressing key ZERO.
2) Press key to cycle through weight capacity
options. Avai lable settings are 15,30,60,75,150,300,600, 750,1500,3000,6000 and7500 (kg).Here we choose (75kg).
3Press key SMPL to save and return to the Setting Menu.
4) Press down numeric key 1 to check the current
weighing capacity and division.
5) Press key ZERO to go back to the weighing mode.
5-2 Resolution Setting
1) Turn on the power while pressing key SMPL.
2) Key in 1123 with the numeric keys.
3) Enter the resolution selection mode by pressing
key SMPL. Numeric Key 1 is to circle through resolutions options. Available setti ng are 1(1/6000 or 1/7500 ),21/15000and3(1/30000).
4) Press to key SMPL save and return to the setting
menu.
5) Press key ZERO to return to weighing mode.
Note: Division changes with the changing of resolution. when weight unit Lb is selected, it won’t be possible to shift to resolution 1.
JWI-700C 8
5-3 Function Setting
1) Turn on the power while pressing key ZERO.
Note: The following Steps (2) ~ (11) do not require to operate in order.
2) Press numeric key 0 to shift backlight modes.
Options are On, OFF and OnOFF.
on the pan.
= No backlight
= Backlight
3) Press numeric key 2 to set the level in which
the stable indication turns on (filtering) .The lower the setting, the faster stabilization time. Options are 1 and 2(level).
4Press numeric key3 to set the period of inactivity before the scale automatically turns off. Options
are OFF(Non power-off) 51030 and 60 (minutes).
5) Press numeric key 4 to set the range in which the
Zero indication turns on .Options are d0, d1, d2, d3, d4 and d5. (d= scale division)
6Press numeric key 5 to select serial transmission
rates. Options are 9600, 4800 and 2400 .
7) Press numeric key 7 to set buzz sounds. Options
are UnInnoLo and nbEEP.
=There will be a warning sound when the quantity of the articles exceeds the upper limit.
=There will be a warning sound when the quantity of the articles is between the upper and
lower l limit (including the upper and lower limits).
= There will be a warning sound when the material quantity exceeds the upper and lower
limits, and the weight of the material is more than 20 divisions.
JWI-700C 9
= There will be a warning sound w hen t he material quantity is less than the lower limit, and
the weight of the material is more than 20 divisions.
= No sound alarm.
8) Press numeric key 8 to set whether to save the
upper & lower limit of quantity checking.
. = Previously set quantity checking values are not retained when the unit is turned on.
. = Previously set quantity checking values are retained when the unit is turned on.
9) Press numeric key 9 to set print modes . Opti ons are
Prt.Pr, Prt.Co and Prt.St. If Prt.Co is chosen as the print mode, PC will be automatically selected as the external devices.
. = manual print
. = continuous print
. = Stable printing (the weight of the weighted articles should be more than 9 divisions.
Weighted articles should be removed and the scale goes back to zero before print out the next record.)
10) Press key MC/CK to select external devices.
Options are PCAXTP,Godex ,SH,
ET,TDP and ZEBRA.
11) Press Key MR to switch On o r O FF RTC
function.
12) When setting is completed, press key ZERO to save and return to weighing mode.
6. Calibration Mode
1) If good linearity of the load cell was not guaranteed, linear calibration is required. After the
linear calibration is made, there is no need to make single point calibration.
2) If there is limit in weights, single point calibration is needed .it would be more accurate to
calibrate the scale with 1/3 or more of the full load.
3) Here we take 75kg weighing capacity as an example.
JWI-700C 10
6-1 Linear calibration
1Turn on the power while pressing key TARE to initiate linear calibration.
2) Again press key TARE to enter zero calibration
mode, with “ ”flashing on the weight window.
3) Wait till “ ” appears put weights of 1/3
of full load on the weighing pa n and pr ess down key
TARE. (E.g. for 75kg weighing capacity, 1/3 of full
load is 25kg)
4) Wait till appears, then put weights
of 2/3 of full load50kgon and press down key TARE.
5) Wait till “ ” appears, then put weights
of full load(75kg) on and press key TARE.
6) The calibration Procedure is completed with a
symbol of “ ”flashing, then take away the weights.
7Press key TARE to return to weighing mode.
6-2 Single point calibration
1) Turn on the power while pressing key SMPL to
initiate single point calibration.
2) Input 11 with numeric keys
3) Again press key SMPL to enter the zero point
calibration mode, with “ ”flashing.
JWI-700C 11
4) Wait till “” is stable, key in the calibration value
25(1/3 full load) and put the corresponding weights On, then press key UWS. “25” is flashing.
5) The calibration procedure is completed with a“
flashing on the weight screen. Now, remove all the weights.
6) Press key SMPL to save, and then press key ZERO to return to normal weighing mode.
7. Operation
7-1 Basic Weighing 7-1-1 Weighing
Place item to be weighed on the scal e. The Weight
window shown is 2.000kg, Gross Weight.
7-1-2 Tare & Preset Tare Tare
When weighing a sample that must be held in a container, taring stores the container weight into memory.
1) Under the weighing mode, place the container on the scale, weight
window shown is 0.500kg. Press the Key TARE to complete tare action.
The symbol appears and with symbol “ ” pointing at “NET.
2To clear tare, with an empty pan, Press down the key TARE or key ZERO.
Preset Tare
1) Under the unload condition , Enter tare
value using the numeric keys together with key , then press key TARE. The weight
window displays the tare value.
2) Put the load on the container, the scale will automatically deduct the value of the container
from the total value. Note: during the process of Tare or Preset Tare, if the load and container are not removed, press zero to check the gross weight.
JWI-700C 12
7-1-3 Unit Switch Operation
1) Press key SMPL while powering on the scale.
2) Key in 1132 via numeric keys.
3) Again press key SMPL to enter unit selection
mode. Numeric key 1 is to toggle between kg
and Lb units.
4) Press key SMPL to save, and then press key ZERO to return to the weighing mode.
7-2 Basic Counting 7-2-1 Upper & Lower Quantity Limit Check ing
1) Under the weighing mode, press and
hold key MC/CK to initial upper/lower limits
setting. (“ ” is blinking.)
2) Use numeric keys to set the upper/lower
Limits. Key is to save upper limit value and advance to the lower limit setting.
3) When setting is completed, again press key MC/CK to return to weighing and checking
mode.
4)To cancel checking action , long press key MC/CK, then press key “·” twice .Checking
action is cancelled, with “ ”displays momentarily before the screen go back to zero mode.
7-2-2 Entering a Known Unit Weight
1) Under the weighing mode, use the numeric
keys together with key “·” to input unit weight value, and then press UWS.
2Put the article on the weighing pan , the
scale starts counting.
JWI-700C 13
Note: Symbol “ ” points at “UW” when entered unit weight is lower than 4/5 of scale division.
7-2-3 Sample Counting & ACAI Sample Counting
1) Place samples onto the weighing pan
(Or into a tared container) and input the quantity with numeric keys, then press key SMPL.
The unit weight is calculated, with Symbol “ ” Pointing
at “ACAI”
.
Note:
●The larger the sample size, the more accurate unit weight.
Symbol “ ” points at “SMPL” when calculated unit weight is lower than 4/5 of scale
division or the quantity of sample is less than 10 pieces.
2) Remove the samples and put the load on, the scale begins to count.
3) Press key CE to return to the weighing mode.
ACAI Automatic Counting Accuracy Improvement (ACAI) results in a more accurate count by increasing the reference weight without the need to count additional parts. A higher reference weight is im port ant w hen there is a risk o f inc onsi ste nt pi ece w ei ght s or i f t he r e ference w eig ht is close to the minimum. ACAI uses an initial averaging unit weight to count additional pieces that are placed on the scale .after a few seconds, the scale gives a beep as the new higher reference weight is used to recalculate the averaging unit weight. The process can be repeated as long as the additional weight is less than the previous reference weight . Once this limit is exceeded ACAI is turned off.
7-2-4 Accumulation, Accumulation Display and Accumulation clear
1) Enter the indicated unit weight, press key UWS
and put first piece of load on the weighing pan (or into a tared container). (Refer to 7-2-2)
2) Press key M+, the first accumulation event
is displayed momentarily before Symbol “ ”
pointing at
“M+”. The display reverts to normal weighing mode in a second .Remove the fir st piece of load.
JWI-700C 14
3) Put the second piece of load on, then press key M+ to add the seco n d accumulation even t into memory. Repeat step 2-3 till accumulation actions are finished.
Note: Maximum is 99 pieces.
Accumulation display
Press key MR to displays the total accumulation data (weight, count and number of weighments) and the first 10 accumulation events in detail.
Accumulation clear
To clear the total accumulation da t a or the fir st 10 accu mula t ion events, Press key MC/CK while the data is displayed. Symbol “ ” disappears when the total accumulation data is
deleted.
8. Serial Interface
If external interface is needed, please select the proper two-in-one board first, which integrates RTC (time display), RS-232 and relay (weight checking) functional module onto one circuit board. Only when this board is adopted, the three functions can be enabled.
8-1 RS-232 connector
JWI-700C 15
8-2 Single Option
1) RS232+RTC+Relay+ TDP / SH-24TP/ ZEBRA / GODEX printer
2) RS232+RTC+Relay+ LED Light Tower (Applicable to the quality control of the factory product quantity or weight and that of the total production line.)
3) RS232+RTC+Relay+Computer
8-4 RS-232 Output Format
Baud Rate : 2400、4800、9600 Data Bit : 8 Parity : N ( None ) Stop Bit : 1 Code : ASCI Bit Format
LSB
MSB
0 1 2 3 4 5 6 7 8
Start
Parity
Stop
Data Format Kg
G/N . W . :
+/- k g CR
LF
weight
U . W . : g / p c s CR
LF
Unit weight
T o t a l : p c s CR
LF
pcs
Example: G.W. + 2.2352 kg U.W. + 0.5352 g/pcs Total : 4176 pcs lb
G/
N
. W . : +/- l b CR
LF
weight
U . W . : l b / p c s
CR
LF
Unit weight
T o t a l : p c s CR
LF
pcs
Example: G.W. + 2.2352 lb
U.W. + 0.5352 lb/pcs Total : 4pcs G = GROSS N = NET
JWI-700C 16
9. Troubleshooting and Error message
Error message Problems Solutions
Initial zero point exceeds /-
30% (take 30% as
reference basis).
1. To check whether there a
re other alien
articles on the scale pan, remove those
articles.
2. LOAD CELL failure, which requires to be
changed or to contact our Service.
Higher or lower than A/D
resolution range.
1. Check whether it is A/D failure, if yes,
please replace AD.
2. LOAD CELL failure, replacement is
required or contact our Service.
EEPROM Check sum failure Re-sold EEPROM or contact our Service.
The weighed articles are
overload .
Do not load the item exceeds the maximum
tolerance.
The calibration value exceeds the full load capacity while processing
single point Calibration.
Adjust the calibration value.
The accumulated number of weighments, total count or weight exceeds display
range.
No more accumulations.
(total count
window)
Under unit weight entering or sampling mode, the quantity of weighed articles
exceeds display range.
-----------------------------
(unit
weight window)
The calculate
unit weight
exceeds the display range
during sampling.
----------------------------
Low battery
Recharge the battery, the indicator can be used while it is charging.
JWI-700C 17
JWI-700C 18
JWI-700C 19