The information contained herein is proprietary and confidential and is being provided
subject to the condition that (i) it be held in confidence except to the extent required
otherwise by law and (ii) it will be used only for the purposes described herein. Any third
party that is given access to this information shall be similarly bound in writing.
Trademark Notice
Itron is a registered trademark of Itron, Inc.
All other product names and logos in this documentation are used for identification purposes
only and may be trademarks or registered trademarks of their respective companies.
Suggestions
For more information about Itron or Itron products, see www.itron.com.
If you have questions or comments about the software or hardware product, contact Itron
Technical Support Services.
Contact
• Email: support@itron.com
• Itron Access: https://support.itron.com
• Telephone Itron Technical Support North America: 1-877-487-6602
For technical support contact information by region, go to www.itron.com and select your
country and language.
Contents
Chapter 1 Important Safety and Compliance Information.............................. 1
USA, FCC Part 15 spectrum compliance............................................................................................................. 1
Do not drop...........................................................................................................................................................4
Chapter 2 Intelis Gas Meter................................................................................6
Related documents...............................................................................................................................................6
Intelis Gas Meter models...................................................................................................................................... 7
Intelis Gas Meter components..............................................................................................................................7
Intelis Gas Meter specications............................................................................................................................8
Intelis Gas Meter dimensions .............................................................................................................................10
Chapter 3 Installing the Intelis Gas Meter.......................................................11
Prior to start-up...................................................................................................................................................12
Removing or installing the pipe plug.................................................................................................................. 13
Removing the pipe plug........................................................................................................................... 13
Installing the pipe plug............................................................................................................................. 14
Chapter 4 Intelis Gas Meter LCD Operation................................................... 15
The Intelis Gas Meter LCD..................................................................................................................................15
Intelis Gas Meter LCD display alarms......................................................................................................17
Intelis Gas Meter LCD display icons........................................................................................................ 19
Performing an LCD test............................................................................................................................20
Chapter 5 Proving the Intelis Gas Meter......................................................... 21
Intelis Gas Meter Test Mode............................................................................................................................... 22
Entering Test Mode.................................................................................................................................. 22
Recording Test Mode consumption......................................................................................................... 22
Exiting Test Mode.....................................................................................................................................23
Intelis Gas Meter Installation Guide TDC-1782-002iii
Proprietary and Confidential
Proving the Intelis Gas Meter using the Honeywell SNAP proving system............................................. 23
Tempering the meter by exercising...................................................................................................24
Setting up the prover........................................................................................................................ 25
Proving the Intelis Gas Meter............................................................................................................29
Proving the Intelis Gas Meter using the Measurement Systems proving system....................................32
Temper the meter by exercising the meter....................................................................................... 33
Setting up the Measurement Systems prover.................................................................................. 34
Proving the Intelis Gas Meter............................................................................................................42
Proving the Intelis Gas Meter using the Energy Economics (EEI) proving system...................................44
Temper the meter by exercising the meter....................................................................................... 44
Setting up the EEI prover..................................................................................................................46
Proving the Intelis Gas Meter............................................................................................................50
Chapter 6 Adjusting the Intelis Gas Meter Calibration.................................. 52
To adjust the calibration of an Intelis Gas Meter.................................................................................................52
To seal an Intelis gas meter.................................................................................................................................58
To unseal an Intelis gas meter's UMU................................................................................................................ 60
Chapter 7 Intelis Gas Meter Index Replacement............................................62
Replacing the Intelis Gas Meter index assembly................................................................................................62
Intelis Gas Meter Installation Guide TDC-1782-002iv
Proprietary and Confidential
Chapter 1 Important Safety and Compliance
Information
This section provides important information for your safety and product compliance.
USA, FCC Part 15 spectrum compliance
This device complies with Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation.
Operation is subject to the following two conditions:
• This device may not cause harmful interference.
• This device must accept any interference that may cause undesirable operation.
This device must be installed to provide a separation distance of at least 20 centimeters (7.9
inches) from all persons to be compliant with regulatory RF exposure.
USA, FCC Class B-Part 15
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver
is connected.
• Consult the dealer or an experienced radio or TV technician for help.
Modifications and Repairs
To ensure system performance, this device and antenna shall not be changed or modified
without the express approval of Itron. Per FCC rules, unapproved modifications or operation
beyond or in conflict with these instructions for use could void the user's authority to operate
the equipment.
Intelis Gas Meter Installation Guide TDC-1782-0021
Proprietary and Confidential
Important Safety and Compliance Information
Canada, ISED spectrum compliance
Compliance Statement CanadaDéclaration de Conformité
This device complies with Innovation, Science
and Economic Development Canada (ISED)
license-exempt RSS standard(s). Operation is
subject to the following two conditions: (1) this
device may not cause interference, (2) this
device must accept any interference, including
interference that may cause undesired operation
of the device.
Under Innovation, Science and Economic
Development Canada (ISED) regulations, this
radio transmitter may only operate using an
antenna of a type and maximum (or lesser) gain
approved for the transmitter by Industry Canada.
To reduce potential radio interference to other
users, the antenna type and its gain should be so
chosen that the equivalent isotropically radiated
power (e.i.r.p.) is not more than that necessary
for successful communication.
Le présent appareil est conforme aux CNR
d'Industrie Canada applicables aux appareils
radio exempts de licence. L'exploitation est
autorisée aux deux conditions suivantes: (1)
l'appareil ne doit pas produire de brouillage, (2)
l'utilisateur de l'appareil doit accepter tout
brouillage radioélectrique subi, même si le
brouillage est susceptible d'en compromettre le
fonctionnement.
Conformément à la réglementation d'Industrie
Canada, le présent émetteur radio peut
fonctionner avec une antenne d'un type et d'un
gain maximal (ou inférieur) approuvé pour
l'émetteur par Industrie Canada. Dans le but de
réduire les risques de brouillage radioélectrique à
l'intention des autres utilisateurs, il faut choisir le
type d'antenne et son gain de sorte que la
puissance isotrope rayonnée équivalente (p.i.r.e.)
ne dépasse pas l'intensité nécessaire à
l'établissement d'une communication
satisfaisante.
RF Exposure (FCC/ISED)
This equipment complies with radiation exposure
limits set forth for an uncontrolled environment.
This equipment should be installed and operated
with minimum distance 20 cm between the
radiator and your body
be co-located or operating in conjunction with
any other antenna or transmitter.
. This transmitter must not
Cet équipement est conforme aux limites
d'exposition aux radiations dans un
environnement non contrôlé. Cet équipement do
it être installé et utilisé à distance minimum de 20
cm entre le radiateur et votre corps. Cet émetteur
ne doit pas être co-localisées ou opérant en
conjonction avec tout autre antenne ou
transmetteur.
Australia, ACMA spectrum compliance
When this device is sold and shipped to Australia, it is configured and labeled accordingly to
be compliant with ACMA Standards for the Radio, EMC and RF Exposure. This includes
standard AS/NZS 4268 RF spectrum standard for frequency and power out.
Intelis Gas Meter Installation Guide TDC-1782-0022
Proprietary and Confidential
Electromagnetic compatibility
Caution: Use only approved accessories with this equipment. All cables must be
high quality, shielded, and correctly terminated. Unapproved modifications or
operation beyond or in conflict with these use instructions may void the authority's
authorization to operate the equipment.
Intrinsic safety
Warning: Substitution of components may impair intrinsic safety.
Lithium battery
Warning: Follow these procedures to avoid injury to yourself or others.
• The lithium battery may cause a fire or chemical burn if it is not disposed of
properly.
• Do not recharge, disassemble, heat above 212°Fahrenheit (100°C Celsius),
crush, expose to water, or incinerate the lithium battery. Fire, explosion, and
severe burn hazard.
• The battery used in this device may present a risk of fire or chemical burn if
mistreated.
• Keep the lithium battery away from children.
Important Safety and Compliance Information
Transportation classification
The Federal Aviation Administration prohibits operating transmitters and receivers on all
commercial aircraft. When powered and not in Factory ship mode, the Itron device is
considered an operating transmitter and receiver and cannot be shipped by air. All product
returns must be shipped by ground transportation.
Electrostatic discharge
Warning: Internal circuit components can be sensitive to electrostatic discharge.
Before installation, discharge electrostatic buildup by touching a metal pipe or other
earth-grounded metal object prior to touching the meter body
Itron device.
Intelis Gas Meter Installation Guide TDC-1782-0023
Proprietary and Confidential
, register housing, or
Important Safety and Compliance Information
Electrostatic ignition hazard
Warning: Verify the area is not hazardous when installing, servicing, cleaning, or
touching the Itron device.
Device cleaning
Warning: Clean only with a damp cloth.
Do not drop
Warning: While Itron meters are designed to withstand a drop, dropping the meter
may damage the device, impact the meter accuracy, and void the warranty.
Intelis Gas Meter Installation Guide TDC-1782-0024
Proprietary and Confidential
Product notification
Warning: These instructions are not intended to replace any utility or companyestablished meter installation procedures. These instructions are provided for
additional information when the Intelis Gas Meter is installed. The meter installation
must comply with all country, state, and local building and safety regulations as well
as federal regulations including Section 192.353 of Title 49 of the Code of Federal
Regulations. Two pertinent paragraphs of the code are:
• Each meter and service regulator, whether inside or outside of a building must be
in a readily accessible location and be protected from corrosion and other
damage.
• Each meter installed within a building must be located in a ventilated place and
not less than 3 feet from any source of ignition or any source of heat which might
damage the meter.
The Intelis Gas Meter is rated for the following operating and storage temperature
ranges. Use of the meter outside of the listed temperature ranges is not
recommended.
• Operating ranges:
◦ Measurement -30° F (-34° C) to +131° F (55° C)
◦ Valve -13° F (-25° C) to +131° F (55° C)
◦ 500G -40° F (-40° C) to +158° F (70° C)
• Storage range:
◦ -40° F (-40° C) to +158° F (70° C)
Important Safety and Compliance Information
Direct customer inquiries as to the selection and application of gas meters to your
local Itron sales representative or Itron Support.
• Itron does not endorse or warrant the completeness or accuracy of any third-party
meter installation procedures or practices, unless otherwise provided in writing by
Itron.
• Follow your company's standard operating procedures regarding the use of
personal protection equipment (PPE).
• Adhere to guidelines issued by your company in addition to those given in this
document when installing or repairing meters.
• This product, as of the date of manufacture, is designed and tested to conform to
all governmental and industry safety standards as they may apply to the
manufacturer.
• The purchaser and user of this product must comply with all fire control, building
codes, and other safety regulations governing the application, installation,
operation, and general use of this meter to avoid leaking gas hazards resulting
from improper installation, start-up, or use of this product.
• To ensure safe and efficient operation of this product, Itron strongly recommends
installation by a qualified professional.
Intelis Gas Meter Installation Guide TDC-1782-0025
Proprietary and Confidential
Chapter 2 Intelis Gas Meter
The Intelis Gas Meter is a solid state ultrasonic meter featuring an internal safety shut-off
valve, built-in temperature conversion, and an integrated 500G communications module.
The meter supports the listed functionality.
• High flow alarm with optional automatic valve shutoff.
• High temperature alarm and optional automatic valve shutoff.
• Air detection.
• Reverse flow detection.
The Intelis Gas Meter can be configured for Mobile Mode or OpenWay Riva Network Mode.
This compact meter weighs only 4-1/2 pounds and mounts well within existing meter
footprints for easy meter installation and replacement of existing gas meters.
Note: OpenWay Riva Network Mode functionality will be available in a future Riva release.
Related documents
For more information about the Intelis Gas Meter and compatible system architecture
documentation, see your product's documentation.
DocumentItron part number
Intelis Gas Meter Technical Reference GuideTDC-1805-XXX
Intelis Gas Meter Ordering GuidePUB-1309-000
Intelis Gas Meter Specification Sheet101584-SP-XX
Intrinsic Safety Control Drawing Index Module Replacement
Intelis Gas Meter
OpenWay® Operations Center Collection Manager Device
Interface Guide
Itron Security Manager User GuideTDC-1380-009
Field Deployment Manager Tools Configuration GuideTDC-1711-XXX
Intelis Gas Meter Installation Guide TDC-1782-0026
TDC-0908-009
TDC-7083-XXX
Proprietary and Confidential
DocumentItron part number
Field Deployment Manager Tools Application GuideTDC-1713-XXX
Field Deployment Manager Field Representative's GuideTDC-1714-XXX
Intelis Gas Meter models
The Intelis Gas Meter is available in the listed types.
• Temperature compensated
• Non-temperature compensated
Intelis Gas Meter components
The Intelis Gas Meter features these components in a compact solid state meter.
• Intelis Gas Meter index
◦ The index contains the integrated 500G communications module, battery pack, and
the liquid crystal display (LCD). The LCD provides a visual display of consumption,
alarm messages, and status information. For information about LCD operation, see
The Intelis Gas Meter LCD on page 15.
• An internal shutoff valve
• Ultrasonic measuring unit (UMU)
• Inlet and outlet tube
• Dust trap and filter
• Die cast aluminum case
• Steel hubs
Intelis Gas Meter
Internal safety shutoff valve
Caution:
valve and ensure safe conditions before the flow of gas is resumed after a valve has
been closed.
The Intelis Gas Meter shutoff valve is located in the outlet of the meter. The automatic
shutof
automatic shut off for high flow or high temperature events. The valve can also be shut off
through an RF command over the Mobile or OpenWay Riva Network.
f valve is a safety feature automatically triggered if the meter is configured for
Automatic shutoff configuration
The Intelis Gas Meter can be configured for an automatic shut off for the following event
types.
• High flow. Configuration for a high flow event enables an automatic shut off if the high
flow threshold is reached or exceeded in a configurable window of time.
Intelis Gas Meter Installation Guide TDC-1782-0027
It is mandatory that a technician be on site at the meter to re-open the
Proprietary and Confidential
Intelis Gas Meter
• High temperature. Configuration for a high temperature event enables an automatic shut
off if the high temperature threshold is reached or exceeded.
Note: For more information about high flow and high temperature event configuration, see
the FDM documentation. For document information, see Related documents on page 6.
Integrated 500G communications module
The integrated 500G RF communications module is designed to operate in either Mobile or
OpenWay Riva Network Mode. When the Intelis Gas Meter is operating in Mobile Mode, it
may be read by ChoiceConnect handheld readers, mobile collection, or the Itron Mobile
Radio connected to a user-supplied computer or Bluetooth device. When the meter is
operating in OpenWay Riva Network Mode, it is read by the OpenWay Riva Network. For
compatible Itron reading and programming devices, see the Intelis Gas Meter Technical
Reference Guide (TDC-1805-XXX).
Note: The communication functionality is dependent on the module's firmware version.
Intelis Gas Meter specifications
This section lists the Intelis Gas Meter product specifications.
SpecificationSpecification value
Meter capacity
Measurement principleUltrasonic time of flight
AccuracyClass 1 and ±0.5% at room temperature
Meter typeTemperature compensated (TC) or non-temperature
214.4 mm161.9 mm152.4 mm138.3 mm78.6 mm230.6 mm48.8 mm
Note: B dimension can vary with dif
ferent hubs options. (20LT is shown).
Intelis Gas Meter Installation Guide TDC-1782-00210
Proprietary and Confidential
Chapter 3 Installing the Intelis Gas Meter
This section provides the Intelis Gas Meter installation information.
Warning: This product, as of the date of its manufacture, is designed and tested to
conform to all governmental or industry safety standards then existing as they may
apply to the manufacturer. The purchaser and user of this product are warned that
compliance with all country, state, and local codes required in order to avoid the
hazards of leaking gas resulting from improper installation, start-up, or use of this
product, and further, that all fire control, building codes, or other safety regulations
established under public laws which regulate or govern the application, installation,
operation, or general use of this product, should be complied with. In order to ensure
the safe and proper operation of this product, the manufacturer recommends that a
qualified technician install this product.
The Intelis Gas Meter must be installed with the inlet to the left, the outlet to
the right, the meter in a horizontal position and with the LCD facing out. INLET
and a flow arrow are marked on the Intelis Gas Meter casing.
Note: These instructions assume the gas is turned of
removed (when the Intelis Gas Meter is replacing an existing installation), piping
tests are completed, and all safety requirements are met.
Intelis Gas Meter Installation Guide TDC-1782-00211
f, the previous gas meter is
Proprietary and Confidential
Installing the Intelis Gas Meter
Prior to start-up
Note: The Intelis Gas Meter is shipped from the factory with the shutoff valve in the open
position. Verify the valve is open prior to installing the meter. Confirm the open valve icon
is displayed on the LCD display and that the valve closed icon is not displaying .
Perform a visual check of the inside of the meter outlet to also verify the shutoff valve is
open.
Before the Intelis Gas Meter start-up:
• Read the meter badge data regarding maximum allowable operating pressure (MAOP)
and capacity flow rate to be sure the meter meets the installation requirements.
• Sight across both swivel flanges or test with a suitable level to determine that the flanges
are nominally in the same plane and will not cause excessive strain on the meter or
piping when the connection nuts are tightened.
Start-up
1. Blowout the service lines before the meter is installed to ensure no dirt, debris, or liquids
of any kind will be carried into the meter when the gas is flowing in the line.
2. Place a new connection nut washer (gasket) on each connect nut or meter hub.
3. Support the meter so that both hubs are against the connection washers and run the
connection nuts down hand tight.
4. In an alternating fashion, tighten the nuts to the appropriate torque for the connection
size.
5. Check the meter's downstream system to ensure that all connections are leak free or that
the downstream valve, if one is present, is closed.
6. Open the upstream and downstream valves very slowly to prevent any pressure surges
into or out of the meter.
Intelis Gas Meter Installation Guide TDC-1782-00212
Proprietary and Confidential
Installing the Intelis Gas Meter
Caution: Avoid high differential pressure across the meter. Abnormal differential
pressures can damage internal meter components.
7. Alternate cracking open the upstream and downstream valves for a few seconds and
then slowly-over a period of 10 seconds (¼ PSIG systems)-turn the valve to the ¼ open
position and then to the full open position.
8. After the meter is pressurized, apply a soap solution or other good leak detecting liquid to
the connections to check for leaks.
9. Test the meter to verify the meter is registering the gas flow.
1. Press the button to activate the LCD.
2. Observe the LCD display and look for usage to increase.
Note: If the LCD does not display any expected consumption, investigate the
installation. If necessary, change out the meter.
10. Perform a shadow or lock-in test following company or utility procedures.
11. After the Intelis Gas Meter installation testing is complete, purge all air from the gas line
including the piping section.
Warning: Never allow gas to discharge into a confined space or area where there
are ignition sources unless precautions are in place to eliminate potential
hazardous conditions.
It is possible to get air detection alarms during Intelis Gas Meter installation due to
installation setups, gas pressure, flow rates, pipe length, and general installation
practices. Air detection alarms during installation should be ignored.
A rapid flow rate change can cause a temporary flow measurement hard failure alarm.
For example, a flow measurement hard fail alarm could occur during installation, T
Mode (proving the meter), external leak test (water tank test), or other rapid flow rate
change situations. The alarm will display on the LCD and as long as it is temporary (and
not continuously in the flow measurement hard fail alarm state), is acceptable behavior. If
the flow measurement hard fail alarm remains active and does not clear, remove the
meter and return it to Itron for evaluation.
Removing or installing the pipe plug
This section provides the steps to remove or re-install the pipe plug.
Removing the pipe plug
• Using a 3/16" hand-driven Allen wrench, remove the pipe plug.
est
Intelis Gas Meter Installation Guide TDC-1782-00213
Proprietary and Confidential
Installing the Intelis Gas Meter
Installing the pipe plug
1. Clean any debris from the pipe plug and inspect it for any damage or deformation.
Replace the pipe plug if any defects or wear are found.
2. Apply new sealant to the leading 3 to 4 threads of the pipe plug.
Note: Itron recommends RectorSeal #5 sealant.
3. Using a 3/16" hand-driven Allen wrench, begin tightening the pipe plug into the threads,
using care not to cross-thread or remove the sealant during the process.
4. Tighten the pipe plug. Do not exceed 50 inch-pounds torque (5.6 Newton Meters).
5. Leak test following your utility's leak test procedure.
Intelis Gas Meter Installation Guide TDC-1782-00214
Proprietary and Confidential
Chapter 4 Intelis Gas Meter LCD Operation
This section provides information about the Intelis Gas Meter LCD.
The Intelis Gas Meter LCD
This section describes the Intelis Gas Meter LCD menus that are displayed in a pre-defined
order (1-8).
Note: The Intelis Gas Meter ships in a sealed (locked) state.
Additionally, there are two screens that display in response to an overflow or test condition.
Note: The two additional menu pages provide overflow condition information and an
interface to perform an LCD test.
The LCD test is initiated by a long press. These additional menu pages are not a part of the
normal sequence. The LCD normal mode is of
menu page is active for 6 seconds before the LCD advances to the next menu page. When
the last menu page is reached, the process loops to the first menu page until the configured
timeout is reached.
f. A button press will activate the LCD. Each
Note: A button press will also advance the LCD to the next menu page.
LCD
menu
1.
LCD displayDescription
• TC meter display in Imperial units.
• NTC meter display in Imperial units.
• TC meter display in Metric units.
Note: All menu displays can be
configured for Imperial or Metric units.
The remaining displays are shown in
Imperial units.
The LCD displays with 5 digits to the left of
the decimal and 3 to the right of the decimal
(in cubic feet x 100 [Imperial] or m3 [Metric]
dependent on customer configuration). The
primary index (Menu 1) will be either TC or
NTC measurement based on the meter
configuration.
Intelis Gas Meter Installation Guide TDC-1782-00215
Proprietary and Confidential
Intelis Gas Meter LCD Operation
LCD
LCD displayDescription
menu
2.Displays current index volume measured
under an alarm condition.
3.Displays active alarms.
Note: In the illustration, 001 indicates the
number of the current displayed alarm while
003 indicates the total number of alarms.
Before the display moves to the next menu
page, all active alarms are displayed. If
there are no alarms, NO ALARM
displays.
For more information about the Intelis Gas
Meter display events and alarms, see Intelis
Gas Meter LCD display alarms on page 17.
4.Local date and time. The local date is
formatted as configured. The date and time
are adjusted by the local time zone and DST
configuration.
5.Current gas temperature and conversion
factor
. This is the gas temperature from the
temperature sensor in the UMU.
6.
Test mode. Flow rate and volume.
7.Pressure compensation.
8.
• R LNF: Register Legal Non-Fixed
Firmware; R LF: Register Legal Fixed
Firmware
Firmware versions. The display cycles
through three screens to display the six
Intelis Gas Meter firmware versions. The
firmware version types are displayed on the
three LCD display menus.
• E LNF: Encoder Legal Non-Fixed
Firmware; E NL: Encoder Non-Legal
Firmware
Intelis Gas Meter Installation Guide TDC-1782-00216
Proprietary and Confidential
Intelis Gas Meter LCD Operation
LCD
menu
LCD displayDescription
• E LF: Encoder Legal Fixed Firmware;
UMU: Ultrasonic Measurement Unit CRC
Intelis Gas Meter LCD display alarms
The Intelis Gas Meter reports the following alarms on the LCD. Active alarms are logged.
Events that create alarms are logged with a start and an end time. Users may cross check
alarms over the network or with a device loaded with FDM T
Cross checking Events with FDM.
LCD event or alarmDescription
Reverse flow detect. The meter is currently detecting
reverse flow; gas flowing from out to in.
The reverse flow event is an alarm for utilities that alert
them about possible tampering. If a reverse flow is
detected, the meter may be installed backwards.
ools. For more information, see
Ext. high flow detect. The meter detected flow that
exceeded the customer configured or Itron configured
high flow threshold of 500 CFH over 5 counts (about 10
seconds).
High temp detect. The meter detected that the
temperature exceeded the customer configurable high
temperature threshold.
Air in pipe detect. The meter detected air in the meter.
The air in pipe alarm will be set to ‘ready to detect’ state
after measuring more than 1.4 CFH (40 L/H). The speed
of sound of the gas being measured is periodically
compared to the speed of sound of air to determine if an
air in pipe alarm/event is triggered. The minimum flow
rate was chosen to reduce nuisance air in pipe alarms
upon initial installation.
Intelis is rated as Class 1 (+/-1% accuracy) for Natural
Gas T
ype H, E, L for reference gases G20, G21,G23,
G25, G26 and G27 which includes gas compositions
with up to 13% propane. It should be noted that during
Intelis Gas Meter Installation Guide TDC-1782-00217
Proprietary and Confidential
Intelis Gas Meter LCD Operation
LCD event or alarmDescription
propane peak shaving, for example, higher
concentrations in propane or nitrogen can occur which
are outside these reference gases. This can result in the
speed of sound for the gas composition being closer to
that of the speed of sound of air, which could result in a
false indication of air detection. For this reason, air
detect alarms/events should be compared to incidences
of propane peak shaving along with typical homeowner
gas usage and reverse flow detection before
considering further action for suspect tampering.
Time sync. The meter detected that its time significantly
fers (5 minutes or greater difference) from the network
dif
time.
Flow meas. hard fail. The meter detected that the flow
measurement sensor failed. The preconfigured meter
Qlmax of 500 CFH (14.2 m3/h) was exceeded. This
condition is a recoverable error with a start and end
event recorded.
Important:
temporary flow measurement hard failure alarm. For
example, a Flow measurement hard fail alarm could
occur during installation, Test Mode (proving the meter),
external leak test (water tank test), or other rapid flow
rate change situations. The alarm will display on the
LCD and as long as it is temporary (and not
continuously in the Flow measurement hard fail alarm
state), is acceptable behavior. If the Flow measurement
hard fail alarm remains active and does not clear,
remove the meter and return it to Itron for evaluation.
A rapid flow rate change can cause a
Intelis Gas Meter Installation Guide TDC-1782-00218
Flow meas. OOR detect. The meter detected flow that is
outside the physical range of the measurement sensor
Gas temp OOR detect. The meter detected that the
maximum (133° F/56° C) or minimum (-31° F/-35° C)
temperature range was exceeded. This condition is a
recoverable error with a start and end recorded.
Flow sens. meas. error. The meter detected that a flow
sensor measurement error occurred. This is an
unrecoverable error
Important: If the Flow Sens. Meas. Error remains active
and does not clear, remove the meter and return it to
Itron for evaluation.
Proprietary and Confidential
.
.
Intelis Gas Meter LCD Operation
LCD event or alarmDescription
Transducer error. The meter detected that its ultrasonic
transducers experienced an error. This is an
unrecoverable error.
Important: If the Transducer Error remains active and
does not clear, remove the meter and return it to Itron
for evaluation.
Additional LCD screen displays
The LCD displays two additional screens in an overflow or test event.
• Overflow. This screen displays if the meter reaches an overflow condition.
• Display test. This screen displays after a long button press of 4.8 seconds. The test
provides a visual confirmation that the LCD is functioning correctly
.
Intelis Gas Meter LCD display icons
The Intelis Gas Meter LCD display provides the following menu icons as a visual indications
of status information.
Display iconIcon definition
Meter status icons
Battery level
• Battery is greater than or equal to 80%
• Battery is greater than 40% and less than 80%
• Battery is greater 10% and less than 40%
•
Battery is less than 10%
Note: Battery life is checked every 12 hours and the icon is
updated accordingly
reaches 5% remaining power.
The meter has an active alarm.
. The battery icon flashes when the battery
Intelis Gas Meter Installation Guide TDC-1782-00219
Proprietary and Confidential
Intelis Gas Meter LCD Operation
Display iconIcon definition
The meter is in Factory Ship Mode.
Note: The truck icon clears when the Intelis Gas Meter is moved
to an operational mode.
Closed padlock icon. The meter is sealed. The lock icon clears
when the meter is unsealed.
Shutoff valve status icons
The shutoff valve is open.
The shutoff valve is closed.
Shutoff valve is in armed state. The shutoff valve opening is
actuated by pressing the button for 2 seconds and releasing it.
Important:
long, the LCD Test screen appears.
A technician should ensure the installation premise is safe by
ensuring there is no flow downstream of the meter prior to
opening the valve. Prior to pressing the button to open the valve,
the technician must check the LCD to verify there are no alarms.
If any alarms are present, investigate and remediate them prior to
opening the valve.
If the Intelis Gas Meter LCD button is pressed too
Performing an LCD test
The Intelis Gas Meter LCD T
and that there are no non-displaying pixels that could lead to misinterpretation of a value.
• Hold down the meter button longer than 4.8 seconds (long press).
The long press initiates the LCD test. After the long press, the display turns black for
600ms. After 600ms, the screen displays LCD TEST again.
Intelis Gas Meter Installation Guide TDC-1782-00220
est is used to confirm the LCD display is functioning properly
Proprietary and Confidential
Chapter 5 Proving the Intelis Gas Meter
This section provides information about proving the Intelis Gas Meter. Meteorological
proving allows users to test the Intelis Gas Meter accuracy. Temper meters unpackaged in
the proving area for 24 hours prior to the testing.
Warning: Follow your company's standard operating procedures regarding the use
of personal protection equipment (PPE). Adhere to guidelines issued by your
company in addition to those contained in this document when proving meters.
The Intelis Gas Meter Test Mode and a prover station are used to confirm the accuracy of
the meter.
Note: The Intelis Gas ultrasonic meter is highly accurate and sensitive to air flow. The meter
displays registration to 0.001 CCF. The meter sensitivity makes it possible to see the index
incrementing even with the hub caps in place.
The prover station captures and counts pulses. Each pulse reflects a specific volume (pulse
weight) of gas. Pulse width is a length of time for the pulse. Both pulse width and pulse
weight are configurable parameters using FDM. The Intelis Gas Meter pulse weight must
match the prover setup as described in this section's proving instructions.
Note: The Intelis Gas Meter cannot be programmed (configured) or calibrated while the
meter is in Test Mode. You must exit Test Mode to program or configure the meter.
Attempting to program or configure parameters while the meter is in Test Mode will result in
a temporary error. If the valve icon is blinking, the meter is unable to go into Test Mode. Wait
a few minutes and try again.
Test Mode is used in a meter shop or by a regulatory body. Some meter configuration and
prover setup are required for the prover to properly interpret the meter pulses. Meter
configuration is completed using FDM. Proving the Intelis Gas Meter requires attaching the
pulse output prover cable between the meter and the proving station. After the test is
complete, the proving cable is removed and the meter returns to normal operation.
Important: A rapid flow rate change can cause a temporary Flow measurement hard failure
alarm. For example, a Flow measurement hard fail alarm could occur during installation,
Test Mode (proving the meter), external leak test (water tank test), or other rapid flow rate
change situations. The alarm will display on the LCD and as long as it is temporary (and not
continuously in the Flow measurement hard fail alarm state), is acceptable behavior. If the
Flow measurement hard fail alarm remains active and does not clear, remove the meter and
return it to Itron for evaluation.
Intelis Gas Meter Installation Guide TDC-1782-00221
Proprietary and Confidential
Proving the Intelis Gas Meter
Proving accessoryItron part number
Itron BPG SONICAL SN G6 pulse prover cableCFG-7100-700
Intelis Gas Meter proving standFIX-7100-001
Intelis Gas Meter Test Mode
You must exit Test Mode to program Test Mode parameters. Attempting to program Test
Mode parameters while the meter is in Test Mode will result in a temporary error. Test Mode
configurable parameters include:
• Pulse width. Default pulse width is 150 ms.
• Pulse weight. Default pulse weight is 0.25 cf/pulse.
• Maximum Test Mode time. Default is 1 hour.
Important: The default Test Mode time is 1 hour. The Intelis Gas Meter will automatically
time out after 1 hour in Test Mode whether there is activity or not.
Test Mode configuration can be completed at the factory at the time the meter is
manufactured. Parameters are configured in one of three ways:
• Through a custom programming file.
• Through programmed default configuration.
• In the field using FDM Tools.
Note: Refer to the instructions for your prover type in the following sections for the
recommended Test Mode pulse width and pulse weight parameters.
Entering Test Mode
Entering Test Mode is accomplished using either the FDM mobile client or through the
magnet in the prover connection end of the prover cable. After the Intelis enters Test Mode,
the following steps occur.
1. The Intelis Gas Meter logs an Entering Test Mode event.
2. The Intelis Gas Meter enters Test Mode.
3. The Intelis Gas Meter Test Mode index is reset to a zero reading.
4. The Intelis Gas Meter starts an optical pulse output.
Caution:
constantly pulsing (such as a small pulse weight and a long pulse width with a
high flow rate.) Adjust the pulse width, the pulse weight, or flow rate as needed.
5. The Intelis Gas Meter LCD moves to Test Display Mode.
There are some parameters that could result in the Intelis Gas Meter
Recording Test Mode consumption
While the Intelis Gas Meter is in Test Mode, the listed operations occur.
Intelis Gas Meter Installation Guide TDC-1782-00222
Proprietary and Confidential
Proving the Intelis Gas Meter
1. The Intelis Gas Meter records the Test Mode consumption in the Test Mode index
register.
2. The Intelis Gas Meter continues to output an optical pulse.
Note: While the Intelis Gas Meter is in Test Mode, the LCD refresh rate is more frequent
than when the meter is operating in Normal Mode.
3. The Intelis Gas Meter continues to update the normal volume registers.
Exiting Test Mode
The Intelis Gas Meter exits Test Mode after it receives a command from the FDM mobile
client, after the proving cable is removed from the meter, or when the Test Mode timeout is
reached. The following operations occur when the Intelis Gas Meter exits Test Mode.
1. The Intelis Gas Meter logs an Exiting Test Mode event.
2. The Intelis Gas Meter ends the optical pulse output.
3. The Intelis Gas Meter LCD returns to normal display mode.
Note: Check the LCD display to verify the Intelis Gas Meter is no longer displaying Menu
6: Test Mode.
Prover types
The Intelis Gas Meter compatible prover types are listed. See the procedure for your prover
type. For questions about your specific prover type, see the manufacturer's prover
documentation.
Prover type
Honeywell SNAP
Measurement Systems
Energy Economics (EEI)
Can-Tronics
Itron BPG SONICAL SNG6
Proving the Intelis Gas Meter using the Honeywell SNAP
pr
oving system
This section provides the information to set up an Intelis Gas Meter using a Honeywell
SNAP proving system.
Note: Proving the Intelis Gas Meter requires Honeywell SNAP prover software .NET X.Y.
Temper meters unpackaged in the proving area for 24 hours prior to testing. If tempering is
not possible, complete steps 1-9 of Tempering the meter by exercising on page 24. If the
meters are tempered, continue to Setting up the prover on page 25.
Intelis Gas Meter Installation Guide TDC-1782-00223
Proprietary and Confidential
Proving the Intelis Gas Meter
Warning: Follow your company’s standard operating procedures regarding the use
of personal protection equipment (PPE). Adhere to guidelines issued by your
company in addition to those contained in this document when proving meters.
Adjust the proving table to accommodate the height of the Intelis Gas Meter. If necessary,
use the Intelis Gas Meter prover stand (Itron part number FIX-7100-001). Place the prover
stand on the prover table and set the meter on the stand.
Itron recommends these Intelis Gas Meter Test Mode parameter settings for the Honeywell
SNAP proving system:
• Test Mode pulse weight: 0.25 CF
• Test Mode pulse width: 150 ms
Tempering the meter by exercising
1. Log on to the prover as a Supervisor.
2. Select Special Functions.
3. Select Diagnostics (local).
4. After the Diagnostics screen opens, clamp the meter in place by pressing Clamp in the
lower right screen corner.
5. Press the clamp buttons on both sides of the SNAP prover.
6. After the Intelis Gas Meter is clamped to the prover, turn on the appropriate nozzles in the
center of the screen to exercise the meter.
Note: Typically, Nozzle Number 2-245 CFH is used to exercise the Intelis Meter.
Exercise the meter for 10 to 15 minutes prior to proving the meter.
Select Close Nozzles.
7.
8. Unclamp the meter.
9. Exit Special Functions.
Intelis Gas Meter Installation Guide TDC-1782-00224
Proprietary and Confidential
Proving the Intelis Gas Meter
Setting up the prover
1. If you have not previously done so, log on to the prover.
Note: If you are proving an Intelis Gas Meter for the first time on the prover or if you are
changing proving parameters for the Intelis Gas Meter, you must log on as a Supervisor.
To set up Itron Intelis Gas Meter and ITRON-RTY for the first time, follow steps 2 through
10.
2. Select Special Functions.
3. Select Edit Config T
4. Select Edit Setup V
ables.
alues.
Intelis Gas Meter Installation Guide TDC-1782-00225
Proprietary and Confidential
Proving the Intelis Gas Meter
5. Select the MeterMfg_T
able checkbox.
6. Click in the next available (open) row in the top left table and enter ITRON-RTY for the
meter manufacturer. Click off of the table and select Update Table.
Intelis Gas Meter Installation Guide TDC-1782-00226
Proprietary and Confidential
Proving the Intelis Gas Meter
Important: If Itron, Actaris, Sprague, or other is listed, it is still necessary to add a new
manufacturer name specifically for the Intelis Gas Meter. Itron suggests using ITRONRTY following the steps listed below. The MFR Type must have the RTY suffix to allow
setting and saving the revs desired for any given test and flow rate. Without the RTY
suffix, the settings will not be saved.
7. Select Meter_Table checkbox.
This step allows you to copy an existing meter setup and modify it to create the new
Intelis Gas Meter (for example, choose I-250, Itron). Select a similar meter's row
. Follow
the Meter-Table instructions: Click on the next open/empty row. To complete the addition,
click off of the table and select Add Chosen Meter. The following illustration shows the
successful copy of the I-250 meter to create the Intelis 250 MeterName, ITRON-RTY
MeterMfr.
8. Update the parameters in the table.
MeterName: Intelis 250
•
• MeterMfg: ITRON-RTY
• Open proof: 100
• Check proof: 100
• Other proof: 100
• Open flow: 250 (or the setting required by your utility)
• Check flow: 50 (or the setting required by your utility)
• Other flow: 100
• Open revs: 6
• Check revs: 2
Intelis Gas Meter Installation Guide TDC-1782-00227
Proprietary and Confidential
Proving the Intelis Gas Meter
• Other revs: 1
• Exercise revs: 1
• Revs unit volume: 4
• Jog rate: 50
• Max press diff: 0.5
• Miscellaneous customer-specific parameters as required (covered in Step 9)
9. Verify or enter the T
OLERANCE values for POS, NEG, Slope, Repair, and Meter Size.
These settings (values) are customer-specific.
10. Click off the table and select Update T
11. Select Exit, select Exit.
able.
Intelis Gas Meter Installation Guide TDC-1782-00228
Proprietary and Confidential
Proving the Intelis Gas Meter
1. Select Prove Meter.
2. Select the meter manufacturer (ITRON-RTY).
3. Select Intelis 250.
Proving the Intelis Gas Meter
4.
Select the correct mode for your meter (TC or NTC).
Note: REG is equivalent to NTC.
5. Select the meter results mode (% error, % accuracy, % proof).
Intelis Gas Meter Installation Guide TDC-1782-00229
Proprietary and Confidential
Proving the Intelis Gas Meter
6. Select the test mode (intest, outtest).
7. Select the optic test method (optic, mag).
8. Select the correct Index (volume per pulse) setting (Imperial, ¼ CF; Metric, 10 dm3).
Intelis Gas Meter Installation Guide TDC-1782-00230
Proprietary and Confidential
Proving the Intelis Gas Meter
Note: T
o reduce cycle time, contact the prover manufacturer to request a 0.10 CF per
pulse option added. Testing has shown acceptable results with settings as low as 4
pulses on open and 1 on check but Itron recommends more pulses to obtain better
results.
9. Click Select Meter to advance to the next screen.
10. Connect the SNAP pulse prover cable end (Itron part number CFG-7100-100) to the J4
(1) optical sensor port typically on the top of the SNAP prover electronic enclosure, next
to the meter temperature probe port.
Intelis Gas Meter Installation Guide TDC-1782-00231
Proprietary and Confidential
Proving the Intelis Gas Meter
11. Connect the Intelis Gas Meter end of the SNAP pulse prover cable to the Intelis Gas
Meter optical port by firmly pressing the aluminum round threads of the cable into the
port. Begin by positioning the cable at the 4:30 o'clock position. T
6:30 o'clock position.
Note: The Intelis Gas Meter will automatically go into T
prover output cable is connected.
12. Start the proving test by clicking RUN on the SNAP prover.
13. Verify the LED blinks (pulses) when the air begins to flow through the meter.
est Mode (Menu 6) when the pulse
ighten the cable to the
Proving the Intelis Gas Meter using the Measurement
Systems proving system
This section provides the information to set up an Intelis Gas Meter using a Measurement
Systems proving system.
Intelis Gas Meter Installation Guide TDC-1782-00232
Proprietary and Confidential
Proving the Intelis Gas Meter
Note: Proving the Intelis Gas Meter requires Measurement Systems prover software. See
the Measurement Systems prover documentation for version requirements.
Temper meters unpackaged and uncapped in the proving area for 24 hours prior to testing.
If tempering is not possible, complete steps 1-8 of Temper the meter by exercising the meter
on page 33. If the meters are tempered, continue to Setting up the Measurement Systems
prover on page 34.
Warning: Follow your company's standard operating procedures regarding the use
of personal protection equipment (PPE). Adhere to guidelines issued by your
company in addition to those contained in this document when proving meters.
Adjust the proving table to accommodate the height of the Intelis Gas Meter. If necessary,
use the Intelis Gas Meter prover stand (Itron part number FIX-7100-001). Place the prover
stand on the prover table and set the meter on the stand.
Note: The Measurement Systems application software on the Measurement Systems
proving system provides touch screen functionality. You may prove the Intelis Gas Meter
using the Measurement Systems Prover screen or a traditional mouse-click to complete the
steps.
Itron recommends these Intelis Gas Meter Test Mode parameter settings for the
Measurement Systems proving system:
• Test Mode pulse weight: 0.10 CF
• Test Mode pulse width: 150 ms
Temper the meter by exercising the meter
1. Open the Measurement Systems Proving System application software.
2. Select Meter Setup to exercise the meter.
3. Select the Global
Intelis Gas Meter Installation Guide TDC-1782-00233
tab > Operate Valves and View Sensor Readings.
Proprietary and Confidential
Proving the Intelis Gas Meter
4. From the Operate Valves and Observe Sensor Readings screen, press the 1 key or
Chuck V
alve to clamp the meter in place.
5. After the Intelis Gas Meter is clamped to the prover, turn on the appropriate nozzle in the
center of the screen to exercise the meter
.
Note: Nozzle 6-250 is typically used for the Intelis Gas Meter.
Exercise the meter for 10 to 15 minutes before proving the meter.
Close the nozzle.
6.
7. Unclamp the meter by pressing the Chuck Valve button.
8. Exit Operate Valves and Observe Sensor Readings.
Setting up the Measurement Systems prover
1. Open the MMS prover application.
2. Select Meter Setup Menu.
Intelis Gas Meter Installation Guide TDC-1782-00234
Proprietary and Confidential
3. Select the Meter Model-1 tab.
Proving the Intelis Gas Meter
4. Select the meter model. If the Intelis 250 or Intelis 250 TC are set up in the MMS prover,
go to step 10. If the Intelis 250 or Intelis 250 TC is not listed, continue to step 5.
5.
Select the Prover tab.
Note: To access the Prover tab, you will be asked for a password. Contact Measurement
Systems if you do not have a password.
6. Select Enter New Meter Model.
Intelis Gas Meter Installation Guide TDC-1782-00235
Proprietary and Confidential
Proving the Intelis Gas Meter
7. Select an existing meter model similar to the new meter as a template. Itron recommends
using the S-250 for an Intelis 250 gas meter and an S-250 TC for the Intelis 250 TC gas
meter
. Select the green bar on the bottom of the screen to use the selected meter as the
starting point.
Note: Measurement Systems uses the lack of TC at the end of the meter name to
designate a non-TC meter, for example; an S-250 is an Itron I-250 CFH non-TC meter
and an Intelis 250 would designate an Intelis 250 CFH non-TC meter.
8. Enter the new meter name in the New Meter Model Name text box (1) and click OK (2).
Intelis Gas Meter Installation Guide TDC-1782-00236
Proprietary and Confidential
Proving the Intelis Gas Meter
9. Press Enter or click the blue bar at the bottom of the screen to go back to the meter
setup menu.
10. Select the Exercise Cycle Duration. The Intelis Gas Meter Exercise Cycle Duration is
typically 15 seconds.
11. Set Leak T
Intelis Gas Meter Installation Guide TDC-1782-00237
est to Yes.
Proprietary and Confidential
Proving the Intelis Gas Meter
12. Verify or enter Test Revs: Index Open, Check, Other settings. Itron recommended
settings are listed.
• Open: 6
• Check: 2
• Other: 0
Note: T
esting has shown acceptable results with settings as low as 4 pulses on open and
1 on check but Itron recommends more pulses to obtain better results.
13. Select the Volume: Index setting. The Volume: Index setting is the recommended
volume per pulse. Itron recommended settings are listed.
• Imperial setting: 0.10 cu ft/rev
• Metric setting: 0.01 m3/rev
Note: Intelis Gas Meter volume per pulse is in dm3; 10dm3 = 0.01 m3.
14. Verify or enter Flow rates (Air Rate): Open, Check, Other settings. Itron recommended
typical settings are listed. For metric flow rates: choose from the flow rate dropdown list
options. These convert the Imperial (CFH) flow rates to Metric (CMH). The Other flow
rate is not required if the Test Revs: Index-Other is set to 0.
• Open: 250
Intelis Gas Meter Installation Guide TDC-1782-00238
Proprietary and Confidential
• Check: 50
• Other: 125
Proving the Intelis Gas Meter
Important: Required Open and Check flow rates may be dif
ferent than those shown
here.
15. Select the Proving Method: Index. The options are Index or Magnetic.
16. Select the mode for your meter. Available options are TC or NON-TC.
17. Select the Meter Model-2 tab.
Intelis Gas Meter Installation Guide TDC-1782-00239
Proprietary and Confidential
Proving the Intelis Gas Meter
18. Verify or enter the InTest and OutTest limit values for Open Limits, Check Limits, OtherLimits, and Delta Limit. The Open, Check, Other
, and Delta Limit settings (values) are
customer-specific.
19. Verify or enter the limit values for Check Dif Limits, Monitor Peaks?, and Max Peak.
The Check Dif Limits, Monitor Peaks? and Max Peak settings (values) are customerspecific.
Intelis Gas Meter Installation Guide TDC-1782-00240
Proprietary and Confidential
Proving the Intelis Gas Meter
20. Verify or enter Tangent Adjustment: Tangent Type.
Note: The Tangent Adjustment: Tangent Type setting can be any non-blank value. The
Tangent Adjustment: Tangent Type value is not used to prove the Intelis Gas Meter.
21. Select the Global
options are listed.
• Error
• Accuracy
• Proof
22. Select InT
est/OutTest. The options are InTest or OutTest.
tab and Display Method for your meter results. Display Method
Intelis Gas Meter Installation Guide TDC-1782-00241
Proprietary and Confidential
Proving the Intelis Gas Meter
23. Press or click the green bar at the bottom of the screen to exit the Meter Setup Menu.
Proving the Intelis Gas Meter
1.
Connect the Measurement Systems pulse prover cable (Itron part number
CFG-7100-500) to the port on the left side of the Measurement Systems prover.
Note: If your Measurement Systems prover does not look like the illustration, contact
Measurement Systems for an adapter cable to connect the Measurement Systems pulse
prover cable to your prover.
Intelis Gas Meter Installation Guide TDC-1782-00242
Proprietary and Confidential
Proving the Intelis Gas Meter
2. Connect the Intelis Gas Meter end of the Measurement Systems pulse prover cable to
the Intelis Gas Meter by firmly pressing the aluminum round threads of the cable into the
port. Begin by positioning the cable at the 4:30 o'clock position. T
ighten the cable to the
6:30 o'clock position.
Note: The Intelis Gas Meter automatically goes into Test Mode (Menu 6) after the cable is
properly connected to the optical port.
3. Enter the Intelis Gas Meter serial number.
Intelis Gas Meter Installation Guide TDC-1782-00243
Proprietary and Confidential
Proving the Intelis Gas Meter
4. Press or click the green bar on the bottom of the screen to begin the Measurement
Systems proving test.
5. Verify the LED blinks (pulses) when the air begins to flow through the meter.
Pr
oving the Intelis Gas Meter using the Energy Economics
(EEI) proving system
This section provides the information to set up an Intelis Gas Meter using an EEI proving
system.
Note: Proving the Intelis Gas Meter requires Energy Economics prover software.
Temper meters unpackaged in the proving area for 24 hours prior to testing. If tempering is
not possible, complete steps 1-5 of Temper the meter by exercising the meter on page 44.
If the meters are tempered, continue to Setting up the EEI prover on page 46.
Warning: Follow your company’
of personal protection equipment (PPE). Adhere to guidelines issued by your
company in addition to those contained in this document when proving meters.
Adjust the proving table to accommodate the height of the Intelis Gas Meter. If necessary,
use the Intelis Gas Meter prover stand (Itron part number FIX-7100-001). Place the prover
stand on the prover table and set the meter on the stand.
Itron recommends these Intelis Gas Meter Test Mode parameter settings for the EEI proving
system:
• Test Mode pulse weight: 0.10 CF
• Test Mode pulse width: 150 ms
s standard operating procedures regarding the use
Temper the meter by exercising the meter
1. Log on to the EEI prover application.
Intelis Gas Meter Installation Guide TDC-1782-00244
Proprietary and Confidential
Proving the Intelis Gas Meter
Note: Logging on to the EEI prover application requires maintenance level security
.
2. Select Maintain > Sensor Maintenance.
3. Clamp the Intelis Gas Meter to the prover and turn on appropriate nozzles in the center
column of the Sensor Maintenance window to exercise the meter
.
Note: Nozzle 192 CFH is typically used to exercise the Intelis Gas Meter.
4. Exercise the meter for 10 to 60 minutes until the Out T
than 0.2° F within a minute.
5. Close the nozzle.
Intelis Gas Meter Installation Guide TDC-1782-00245
emp value does not vary by more
Proprietary and Confidential
Proving the Intelis Gas Meter
6. Unclamp the meter.
7. Exit the Sensor Maintenance window by clicking Done in the lower right corner of the
window.
Setting up the EEI prover
1. Log on to the EEI prover application.
2. Clamp the Intelis Gas Meter to the EEI prover.
3.
Select the Meter tab and choose ITRON from the Manuf drop-down list.
Important: If ITRON is not in the Manuf. drop-down list, contact EEI Prover Systems for
help in adding ITRON to your EEI Systems Prover.
4. Select Intelis 250, Intelis 250 TC, or the name for the Intelis Gas Meter designated by
your company from the T
ype drop-down list.
Important: If the correct Intelis Gas Meter type is not in the Type drop-down list, contact
EEI Prover Systems for help adding the correct meter type to your EEI Systems Prover.
Intelis Gas Meter Installation Guide TDC-1782-00246
Proprietary and Confidential
5. Click the Options tab.
Proving the Intelis Gas Meter
The Meter Information > T
6. Select Eye
for the Test Options Sensor.
est Options window opens.
Intelis Gas Meter Installation Guide TDC-1782-00247
Proprietary and Confidential
Proving the Intelis Gas Meter
7. Select Open & Check for the Run T
8. Enter 100 for Flow
> Open %.
ype Test Options.
Intelis Gas Meter Installation Guide TDC-1782-00248
Proprietary and Confidential
Proving the Intelis Gas Meter
Important: Required Open flow rates may be dif
9. Enter 20 for Flow > Check %.
Important: Required Check flow rates may be dif
10. Select 1/10 cuft for Hand Size.
ferent than those shown here.
ferent than those shown here.
Intelis Gas Meter Installation Guide TDC-1782-00249
Proprietary and Confidential
Proving the Intelis Gas Meter
11. Click Done to complete the Meter Information > T
Note: All other settings (T
angents, Ultra Intest Lock, Ultra Adjust Lock and Station) on the
Test Options window are not relevant for the Intelis meter
Proving the Intelis Gas Meter
est Options and close the window.
1. Connect the EEI pulse prover cable (Itron part number CFG-7100-400) to the EEI prover
index eye connector.
Intelis Gas Meter Installation Guide TDC-1782-00250
Proprietary and Confidential
Proving the Intelis Gas Meter
2. Connect the Intelis Gas Meter end of the EEI pulse prover cable to the Intelis Gas Meter
by firmly pressing the aluminum round threads of the cable into the port. Begin by
positioning the cable at the 4:30 o'clock position. T
ighten the cable to the 6:30 o'clock
position.
Note: The Intelis Gas Meter automatically enters Test Mode when the pulse output cable
is connected to the meter.
3. Press the EEI prover's green button to begin the proving test.
4. Verify the LED blinks (pulses) when the air begins to flow through the meter.
Intelis Gas Meter Installation Guide TDC-1782-00251
Proprietary and Confidential
Chapter 6 Adjusting the Intelis Gas Meter
Calibration
Intelis Gas Meters are calibrated at the factory prior to shipment. FDM is used if it is
necessary to recalibrate the meter. For more information about calibrating the Intelis Gas
Meter using FDM, see the FDM Field Service Representatives Guide. For documentation
information, see Related documents on page 6. Sending new calibration values using FDM
initiates the following calibration operations.
1. Run the meter on a prover.
2. Exit Test Mode.
3. Enter the prover station results values into FDM.
Note: The Intelis Gas Meter must be unsealed to adjust the meter calibration.
4. Wait for the meter to complete calibration.
5. Enter Test Mode.
6. Re-prove the meter to verify the calibration.
7. Exit Test Mode.
To adjust the calibration of an Intelis Gas Meter
Before you perform the Adjust Calibration command, you are required to run the meter on a
prover station. This determines the meter's accuracy, and the results are used during the
Adjust Calibration workflow. Before you start the Adjust Calibration command, review all
information in the sections that follow.
Note: The meter must be out of test mode to perform the Adjust Calibration command. If the
meter is in test mode, perform the Exit Test Mode command before you start this procedure.
Part of this Adjust Calibration workflow includes unsealing the meter. After you complete this
workflow, you must use the separate Seal Meter command to re-seal the meter.
If the meter is operating in mobile mode, you must know the meter's Unlock ID to complete
the Adjust Calibration command.
1. From the Tools menu, select OWR Intelis Gas.
2. Select Test Mode Operations.
3. Select Adjust Calibration.
A warning message appears indicating that Adjusting Calibration will unseal the meter.
Intelis Gas Meter Installation Guide TDC-1782-00252
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
4. Select Y
es.
5. (Mobile endpoint mode only) If you are prompted to enter the Unlock ID, enter the Unlock
ID.
Intelis Gas Meter Installation Guide TDC-1782-00253
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
6. Select Next.
The Parameters screen appears.
Intelis Gas Meter Installation Guide TDC-1782-00254
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
To demonstrate an example of adjusting the calibration, a sample prover station test
results is displayed below and is provided as a reference.
Use the following settings to adjust the meter calibration as it relates to your test results:
• Prover station mode. Use the drop-down menu to select the % in which your prover
station results are expressed:
◦ (Example selection) % Error
% Accuracy
◦
◦ % Proof
• Prover flow rate unit. From the drop-down menu, select the flow rate unit:
◦ (Example selection) Imperial (cubic feet per hour)
◦ Metric (cubic meters per hour)
• Prover flow rate check. Enter the prover station check flow rate. This is typically
20-35% of the meter capacity. (Example entry 50 CFH)
• Prover percentage check. Enter the value from the test mode screen for the meter's
check result. (Example entry +0.7).
• Prover flow rate open. Enter the prover station open flow rate. This is typically
80-100% of the capacity. (Example entry 250 CFH).
• Prover percentage open. Enter the value from the prover screen for the meter's open
result. (Example entry -0.3).
7. Select Next.
The Adjust Calibration Summary appears so that you may review your parameter
changes.
Intelis Gas Meter Installation Guide TDC-1782-00255
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
8. Select Next.
A progress bar indicates that the meter's calibration is being adjusted.
Intelis Gas Meter Installation Guide TDC-1782-00256
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
The meter uses the entered values to calculate the required calibration parameters to a
target of one of the following, depending on your selected prover station mode:
• % Error. 0% open, 0% check
• % Accuracy. 100% open, 100% check
• % Proof. 100% open, 100% check
Click Quit or Finish to return to the Tools screen.
9.
Intelis Gas Meter Installation Guide TDC-1782-00257
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
10. (Recommended) Rerun the meter on the prover station to verify the meter calibration. It is
expected that you see the results that match the following example:
If the accuracy results are not as expected, repeat the calibration steps.
Note: Prover station accuracy is estimated to be +/- 0.15-0.25%.
T
o seal an Intelis gas meter
1. From the Tools menu, select OWR Intelis Gas Riva.
2. Select Test Mode Operations.
3. Select Seal Meter.
A progress bar indicates that the meter is sealing.
Intelis Gas Meter Installation Guide TDC-1782-00258
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
The Seal Meter Operation Successful screen indicates that the meter is sealed.
Intelis Gas Meter Installation Guide TDC-1782-00259
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
4. Select Finish
to return to the Tools screen.
To unseal an Intelis gas meter's UMU
1. From the Tools menu, select OWR Intelis Gas Riva.
2. Enter the Endpoint ID.
3. Select the Endpoint Mode.
• Mobile
4. Select Next.
5. Select Test Mode Operations.
6. Select the Utility ID.
7. Click Next.
8. Select Adjust Calibration for a device with a Seal State of Sealed.
A warning message appears indicating that Adjusting Calibration will unseal the meter.
9. Select Yes.
10. (Moblie endpoint mode only) Enter the Unlock ID.
Intelis Gas Meter Installation Guide TDC-1782-00260
Proprietary and Confidential
Adjusting the Intelis Gas Meter Calibration
a) Select Next.
The Adjust Calibration screen appears.
1
1. Click Quit or Finish to return to the Tools screen.
Intelis Gas Meter Installation Guide TDC-1782-00261
Proprietary and Confidential
Chapter 7 Intelis Gas Meter Index Replacement
Warning: The Intelis Gas Meter index assembly is a static sensitive device. Avoid
contact with the feed through connector pins on the index assembly. It is important to
take ESD precautions when the index assembly is replaced. To ensure the Intelis
Gas Meter retains consumption data, switch the Intelis Gas Meter to Factory Ship
Mode prior to index replacement.
Index replacement instructions are included with each shipment of replacement indexes
(Intrinsic Safety Control Drawing Index Module Replacement Intelis Gas Meter
TDC-0908-009).
Caution: After an index assembly is replaced, you must perform a valve operation to
place the automatic shut off valve in the desired state using FDM Tools. If the
desired valve state is open, send an FDM Tools Reconnect command. Conversely,
if the desired state is closed, send an FDM Tools Disconnect command. The Intelis
Gas Meter index may only be replaced using the UL classified replacement index kit
(Itron part number ERG-7100-105).
Replacing the Intelis Gas Meter index assembly
Important: Program the integrated 500G communications prior to replacing the index. The
illustrations in these instructions do not show the gas pipes connected to the meter. Perform
an LCD Test prior to replacing the index assembly with the new index to verify the display is
working properly. For more information, see Performing an LCD test on page 20.
Pictures and definitions
Component or toolItron part
number
Original indexERG-7100-005
Intelis Gas Meter index replacement kitERG-7100-105
Intelis meterMTR-7100-005
Badge and UL classification
Note: There are two badge options: red TC
or blue NTC
Required Tools:
Intelis Gas Meter Installation Guide TDC-1782-00262
Proprietary and Confidential
Intelis Gas Meter Index Replacement
Component or toolItron part
number
Phillips (size 1 or 2) slot-head (flat bladed sized 6-10 screwdriver)
(Optional) Awl
Hammer or mallet
(Optional) Diagonal cutters
Torque screwdriver (6-15 in-lbs) Phillips sizes 1-2 or slot-head sizes 6-10
Pre-Check
Warning: Do not attempt to remove meter from gas pipes or adjust/modify/damage
gas pipes/gas pipe connections in any way. Step replacement pictures do not show
the gas pipes connected to the meter. Ensure area is not hazardous (i.e. free of
potentially ignitable atmospheres) before replacing the index, or installing, servicing,
cleaning, or touching the index/meter. On the new index, be careful not to bend or
disturb the two red and two black wires that connect the batteries to the PCB. If the
wires are physically moved/disturbed on the new index, it cannot be installed on the
meter.
Locate and verify the meter assembly and replacement packaging are marked as follows:
On the metal gas meter, find the badge that contains the UL mark, and verify the UL mark in
the MTR-7100-005 badge & UL Classification pictures above are correct (cULus mark with
File number E323571 as above), and verify the following required text: “Intelis for use in
Hazardous Locations only as to Intrinsic Safety”. On the original index (ERG-7100-005),
verify the required text marked on the housing: “Model Intelis Gas Meter”. On the packaging
of the new index (ERG-7100-105), find and verify the following required information: same
cULus mark as in the "MTR-7100-005 badge & UL Classification pictures above" (not
including E323571 beside the mark), and text “Intelis Replacement Index Kit for use in
Hazardous Locations only as to Intrinsic Safety E323571; Model ERG-7100-105”.
1. If your meter installation has one or more security wires included between the original
index and the meter on one or both sides’ seal wire locations, cut the wire(s) with a
diagonal cutter.
Note: the security wire may appear physically dif
ferent from the image; security wire
types are dependent on the utility installation.
2. On the original index, break the (red) tamper seals and remove them. A hammer/mallet
and a screwdriver/awl may be used to accomplish this. Insert the screwdriver or awl until
it touches the tamper seal, then apply pressure into the tamper seal and twist the tool to
break the tamper seal enough to remove it, without damaging the enclosure or other
Intelis Gas Meter Installation Guide TDC-1782-00263
Proprietary and Confidential
Intelis Gas Meter Index Replacement
parts of the original index/meter/gas pipes. Do not use a power tool. Carefully pull the
tamper seal out of the hole, hooking it with the tool if necessary.
3. There are two screws holding the original index onto the meter located behind the tamper
seals just removed. Unscrew both screws while holding the original index. Do not discard
the screws.
4. Firmly and slowly pull the original index straight away from the meter.
5. Align the new index’s screw holes with the screw holes of the meter (oriented as shown in
the following image). Slowly align the new index with the meter until it contacts the meter
Gently push the new index onto the meter so that the 6-pin connectors from the meter
and the new index fully engage. The new index should be flush with the meter.
6. Use the screws from the original index to attach the new index to the meter. Tighten the
screws by hand with a Phillips or slot head screwdriver to a torque between 6 and 15
inch-pounds.
Ensure the head of the screws secure to the index housing. If torque monitoring is
7.
available, torque the screws to 6-15 inch-pounds.
8. Install the two new tamper seals (included with the new index) in the two tamper seal
holes on the new index's front face, pushing them firmly into each hole. On each tamper
seal, there is a raised letter “S” that is part of the plastic. Make sure this “S” is pointing out
of the screw hole, so that it is visible after installation. The replacement procedure is
complete.
.
Intelis Gas Meter Installation Guide TDC-1782-00264
Proprietary and Confidential
Intelis Gas Meter Index Replacement
Intelis Gas Meter Installation Guide TDC-1782-00265
Proprietary and Confidential
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.