The B&K Precision Models 885 & 886 Synthesized In-Circuit
LCR/ESR Meter is a high accuracy hand held portable test
instrument used for measuring inductors, ca paci tor s and resistors
with a basic accuracy of 0.2%. It i s the most advanced handheld
AC/DC impedance measurement instrument to date. The 885 or 886
can help engineers and student s to understand the character istic of
electronics components as well as being an essential tool on any
service bench.
The instrument is auto or manual ranging. Test frequencies of
100Hz, 120Hz, 1KHz 10KHz or 100KHz (886) may be selecte d on
all applicable ranges. The test voltages of 50mVrms, 0.25Vrms,
1Vrms or 1VDC (DCR only) may also be selected on all appli ca ble
ranges. The dual display feature permits simultaneous
measurements.
Components can be measured in the series or parallel mode as
desired; the more standard method is automatically selected first but
can be overridden.
The Model 885 and 886 offers three useful modes for sorting
components.
The highly versatile Mode ls can perf orm virtua lly all the functions
of most bench type LCR bri dges. With a basic accuracy of 0.2%,
this economical LCR meter may be adequately substituted for a
1
www..com
information@itm.com1.800.561.8187
more expensive LCR bridge in many situations. The meter is
powered from two AA Batteries and is supplie d with an AC to DC
charging adapter and two AA Ni -Mh Rechargeable Batterie s.
The instrument has applications in electronic engineering labs,
production facili ties, service shops, and schools. It can be used to
check ESR v alues of capacitors , sort values, selec t precision values,
measure unmarked and unknown i nductors, capacit ors or resistor s,
and to measure capacitance, inductance, or resistance of cables,
switches, circuit boar d foils, etc.
The key features are as following:
Te st condition:
Frequency : 100Hz / 120Hz / 1KHz / 10KHz /
1
100KHz (886)
Level : 1Vrms / 0.25Vrms / 50mVrms /
2.
1VDC (DCR only)
Measurement Parameters : Z, Ls, Lp , C s, C p , D C R ,
ESR, D, Q and
Basic Accu racy: 0.2%
Dual Liquid Crystal Display
Fast/Slow Measurement
Auto Range or Range Hold
Open/Short Calibrati on
Primary Parameters Display:
Z : AC Impedance
DCR : DC Resistance
Ls : Serial Inductance
Lp : Parallel Inductance
θ
2
www..com
information@itm.com1.800.561.8187
Cs : Serial Capacitance
Cp : Parallel Capacitance
Second Parameter Display:
θ
: Phase Angle
ESR : Equivalence Serial Resistance
D : Dissipati on Fact or
Q : Quality Factor
Combinations of Display:
θ
Serial Mode : Z –
Parallel M od e : Cp – D, Cp – Q, Lp – D, Lp – Q
1.2 Impedanc e Parameters
Due to the different test ing signal s on t he impedance mea surement
instrument, there are DC impedance and AC impedance. The
common digital multi -meter can only measure the DC impedance,
but the Model 885 can do both. It is a very important issue to
understand the impedance parameters of the electronic component.
When we analysis the impedance by the impedance measurement
plane (Figure 1.1). It can be visualized by the rea l element on the
X-axis and the imaginary element on the y-axis. T his impedance
measuremen t plan e can also be seen as th e pola r coordin at es. Th e Z
is the magnitude and the
θ
, Cs – D, Cs – Q, Cs – ESR, Ls –
D, Ls – Q, Ls – ESR
is the phase of the impedance.
3
www..com
information@itm.com1.800.561.8187
( )
()
()
()
()
Ohm
Reactance
Resistance
Impedance
1
22
=Ω
=
=
=
==
+==
Ω∠=+=
−
S
S
s
s
s
sss
ss
X
R
Z
R
X
TanSinZX
XRZCosZR
ZjXRZ
θθ
θ
θ
s
X
s
R
()
sX,RZ
s
Z
θ
Imaginary Axis
Real Axis
Figure 1.1
fCC
X
fLLX
C
L
πω
πω
2
11
2====
There are two different types of reactance: Inductive (X
Capacitive (X
). It can be defined as f ollows:
C
) and
L
L = Inductance (H)
C = Capacitance (F)
f = Frequency (Hz)
Also, there are quality factor (Q) and the di ssi pation factor (D) that
need to be discussed. For component, the quality fa ctor serve s as a
measure of the reacta nce purity. In the re al world, t here is always
4
www..com
information@itm.com1.800.561.8187
some associated resistance that dissipates power, decreasing the
pp
p
p
p
p
sss
s
s
s
RC
L
R
X
R
G
B
RCR
L
R
X
D
Q
ω
ω
ω
ω
δ
===
=
===
==
1
tan
11
amount of energy that can be rec overed. The quality fac tor can be
defined as the ratio of the stored energy (reactance) and the
dissipated energy (resistance). Q is generally used for inductors and
D for capacito rs .
There ar e two types of the cir cuit mode. One is series mode, the
other is parallel mode. See Fi gure 1.2 t o find out t he rela tion of t he
series and parallel mode.
5
www..com
information@itm.com1.800.561.8187
Parameter
Range
Z
L
0.000 µH to 9999 H
C
0.000 pF to 9999 F
DCR
0.000 Ω to 9999 MΩ
ESR
D
0.000 to 9999
Q
0.000 to 9999
-180.0 ° to 180.0 °
Real and imaginary components are serial
ss
jXRZ+=
Real and imaginary components are Parallel
jB=1/jX
jBGY+=
jX
R
P
jX
1
P
R
1
Y+=
Rs jX
s
p
p
Figure 1.2
G=1/R
p
p
1.3 Specification
LCD Display Range:
0.000 Ω to 9999 MΩ
0.000 Ω to 9999 Ω
θ
6
www..com
information@itm.com1.800.561.8187
|Zx|
20M ~
10M ~
1M ~
100K ~
10 ~ 1
1 ~ 0.1
DCR
100Hz
120Hz
1KHz
10KHz
100KHz
NA
5%±1
2%±1
0.4% ±1
2%±1
5%±1
2
1 Dx+
Accuracy (Ae):
Z Accuracy:
Freq.
(886)
Note : 1.The accuracy applies when th e test level is set to 1V rm s .
: Ae is not specified if th e test level is set to 50mV.
10M
2% ±1
5% ±1
2.Ae multiplies 1. 25 when the test level is set to 250mVrms.
3.Ae multiplies 1.50 when the test level is set to 50mVrms.
4.When measuring L and C, multiply Ae by
Dx
>
(Ω)
0.1.
1M
100K
(Ω)
1%
2% ±1
(Ω)
±
1 0.5% ±1 0.2% ±1 0.5% ±1 1% ±1
10
(Ω)
Ω
)
Ω
(
(
if the
)
7
www..com
information@itm.com1.800.561.8187
C Accuracy :
79.57
pF
159.1
nF
1.591
nF
15.91
uF
159.1
uF
1591
mF
2% ± 1
1% ± 1
0.5%
1
0.2%
1
0.5%
1
1% ± 1
66.31
pF
132.6
nF
1.326
nF
13.26
uF
132.6
uF
1326
mF
2% ± 1
1% ± 1
0.5%
1
0.2%
1
0.5%
1
1% ± 1
7.957
pF
15.91
pF
159.1
nF
1.591
uF
15.91
uF
159.1
mF
2% ± 1
1% ± 1
0.5%
1
0.2%
1
0.5%
1
1% ± 1
0.795
pF
1.591
pF
15.91
pF
159.1
uF
1.591
uF
15.91
uF
5% ± 1
2% ± 1
0.5%
1
0.2%
1
0.5%
1
1% ± 1
NA
0.159
pF
1.591
pF
15.91
nF
159.1
uF
1.591
uF
NA
5% ± 1
2%± 1
0.4%
1
2%± 1
5% ± 1
100Hz
159.1
pF
|
pF
1.591
nF
|
15.91
nF
|
159.1
uF
|
1591
uF
|
|
15.91
pF
|
120Hz
1KHz
10KHz
100KHz
(886)
132.6
15.91
1.591
pF
|
pF
|
1.326
159.1
15.91
1.591
pF
|
pF
|
pF
|
pF
|
±
±
±
nF
|
pF
|
pF
|
pF
|
nF
132.6
±
nF
15.91
±
pF
1.591
±
pF
159.1
±
13.26
±
1.591
±
159.1
±
15.91
8
uF
|
1326
±
uF
|
159.1
±
uF
|
15.91
±
nF
|
1.591
uF
|
|
13.26
uF
|
|
1.591
uF
|
|
159.1
uF
|
|
15.91
www..com
information@itm.com1.800.561.8187
L Accuracy :
31.83
KH
15.91
H
1591
H
159.1
mH
15.91
mH
1.591
uH
2% ± 1
1% ± 1
0.5%
1
0.2%
1
0.5%
1
1% ± 1
26.52
KH
13.26
H
1326
H
132.6
mH
13.26
mH
1.326
uH
2% ± 1
1% ± 1
0.5%
± 1
0.2%
± 1
0.5%
± 1
1% ± 1
31.83
KH
1.591
H
159.1
H
15.91
mH
1.591
uH
159.1
uH
2% ± 1
1% ± 1
0.5%
± 1
0.2%
± 1
0.5%
± 1
1% ± 1
318.3
H
159.1
H
15.91
H
1.591
uH
159.1
uH
15.91
uH
5% ± 1
2% ± 1
0.5%
1
0.2%
1
0.5%
1
1% ± 1
31.83
H
15.91
H
1.591
mH
159.1
uH
15.91
uH
1.591
uH
NA
5% ± 1
2%± 1
0.4%
2%± 1
5% ± 1
KH
100Hz
120Hz
1KHz
10KHz
100KHz
(886)
15.91
KH
13.26
KH
1.591
159.1
15.91
H
H
KH
|
1591
KH
|
1326
KH
|
159.1
|
15.91
|
1.591
H
H
|
|
159.1
15.91
±
H
|
|
132.6
13.26
H
|
|
15.91
1.591
H
H
|
|
1.591
159.1
±
H
H
|
|
159.1
15.91
|
±
H
|
H
|
H
|
±
mH
|
mH
1.591
mH
1.326
mH
159.1
15.91
1.591
±
uH
±
uH
mH
|
|
159.1
mH
|
|
132.6
uH
|
|
15.91
uH
|
|
1.591
uH
|
|
0.159
9
www..com
± 1
information@itm.com1.800.561.8187
D Accuracy :
|Zx|
Freq.
20M ~
(Ω)
10M ~
(Ω)
1M ~
(Ω)
100K ~
(Ω)
10 ~ 1
(Ω)
1 ~ 0.1
(Ω)
0.020
0.010
0.005 ±0.002 ±0.005 ±0.010
0.050
0.020
100KHz
(886)
NA ±0.050
0.020 ±0.004 ±0.020 ±0.050
|Zx|
Freq.
20M ~
(Ω)
10M ~
(Ω)
1M ~
(Ω)
100K ~
(Ω)
10 ~ 1
(Ω)
1 ~ 0.1
(Ω)
1.046
0.523
0.261 ±0.105 ±0.261 ±0.523
2.615
1.046
100KHz
(886)
NA ±2.615
1.046 ±0.209 ±1.046 ±2.615
100Hz
120Hz
1KHz
10KHz
10M
±
±
1M
100K
±
±
±
±
10
θ Accuracy :
100Hz
120Hz
1KHz
10KHz
10M
±
±
1M
100K
±
±
±
±
10
10
www..com
information@itm.com1.800.561.8187
Z Accuracy:
CxfZx⋅⋅⋅
=
π
2
1
2
1 Dx+
Ω=
⋅⋅⋅⋅
=
⋅⋅⋅
=
−
1590
10100102
1
2
1
93
π
π
Cxf
Zx
As shown in table 1.
C Accuracy:
= Ae of |Zx|
C
Ae
f : Test Frequency (Hz)
Cx : Measured Capacitance Value (F)
|Zx| : Measured Impedance Value (
Accuracy applies when Dx (measured D value)
Ω
)
≦
0.1
When Dx > 0.1, multiply C
Example:
Tes t Condition:
Frequency : 1KHz
Level : 1Vrms
Speed : Slow
DUT : 100nF
Then
Refer to the accuracy table, get C
11
Ae
by
=±0.2%
Ae
www..com
information@itm.com1.800.561.8187
L Accuracy:
LxfZx⋅⋅⋅=
π
2
2
1 Dx+
100
Ae
XxESR
Ae
⋅±=
Cxf
LxfXx
⋅⋅⋅
=⋅⋅⋅=
π
π
2
1
2
= Ae of |Zx|
L
Ae
f : Test Frequency (Hz)
Lx : Measured Inductance Value (H)
|Zx| : Measured Impedance V alue (
Accuracy applies when Dx (measured D value)
Ω
)
≦
0.1
When Dx > 0.1, multiply L
Ae
by
Example:
Tes t Condition:
Frequency : 1KHz
Level : 1Vrms
Speed : Slow
DUT : 1mH
Then
2
π
⋅⋅⋅=
π
LxfZx
−
33
Ω=⋅⋅⋅=
283.610102
Refer to the accuracy table, get L
=±0.5%
Ae
ESR Accuracy :
12
www..com
information@itm.com1.800.561.8187
ESRAe = Ae of |Zx|
Ω=
⋅⋅⋅⋅
=
⋅⋅⋅
=
−
1590
10100102
1
2
1
93
π
π
Cxf
Zx
Ω±=⋅±=18.3
100
Ae
XxESR
Ae
100
Ae
D
Ae
±=
f : Test Frequency (Hz)
Ω
Xx : Measured Reactance Value (
)
Lx : Measured Inductance Value (H)
Cx : Measured Capacitance Value (F)
Accuracy applies when Dx (measured D value)
Example:
Tes t Condition:
Frequency : 1KHz
Level : 1Vrms
Speed : Slow
DUT : 100nF
Then
≦
0.1
Refer to the accuracy table, get
=±0.2%,
C
Ae
D Accuracy:
13
www..com
information@itm.com1.800.561.8187
DAe = Ae of |Zx|
Ω=
⋅⋅⋅⋅
=
⋅⋅⋅
=
−
1590
10100102
1
2
1
93
π
π
Cxf
Zx
002.0
100
±=⋅±=
Ae
D
Ae
DeQx
DeQx
Ae
Q
⋅
⋅
±=
1
2
Accuracy applies when Dx (measured D value)
When Dx > 0.1, multiply Dx by (1+Dx)
Example:
Tes t Condition:
Frequency : 1KHz
Level : 1Vrms
Speed : Slow
DUT : 100nF
Then
≦
0.1
Refer to the accuracy table, get
=±0.2%,
C
Ae
Q Accuracy:
Q
= Ae of |Zx|
Ae
Qx : Meas ured Quality Factor Value
De : Relative D Accuracy
14
www..com
information@itm.com1.800.561.8187
Accuracy applies when
1<⋅ DeQx
005.0
100
±=⋅±=
Ae
De
1.01
θ
100
Ae
π
180
e⋅=θ
Example:
Tes t Condition:
Frequency : 1KHz
Level : 1Vrms
Speed : Slow
DUT : 1mH
Then
π
LxfZx
2
π
⋅⋅⋅=
33
−
Ω=⋅⋅⋅=
283.610102
Refer to the accuracy table, get
=±0.5%,
L
Ae
If measur ed Qx = 20
Then
2
⋅
1
±=⋅
DeQx
DeQx
±=
Q
Ae
2
Accuracy:
15
www..com
information@itm.com1.800.561.8187
Example:
Ω=
⋅⋅⋅⋅
=
⋅⋅⋅
=
−
1590
10100102
1
2
1
93
π
π
Cxf
Zx
deg115.0
100
2.0180
100
Ae180
Ae
±=⋅
π
±=
⋅
π
±=θ
Tes t Condition:
Frequency : 1KHz
Level : 1Vrms
Speed : Slow
DUT : 100nF
Then
Measuring Speed:
Fast : 4.5 meas. / sec.
Slow : 2.5 meas. / sec.
16
www..com
information@itm.com1.800.561.8187
General:
°
Temperature : 0
-20
Relative Humidity : Up to 85%
Battery Type : 2 AA size Ni-Mh or Alkaline
Battery Ch arge : Consta nt current 150mA
Battery Operatin g Time : 2.5 Hours typical
AC Operation : 110/220V A C, 60/50Hz with proper
Low Power Warning : under 2.2V
Dimensions : 174mm x 86mm x 48mm (L x W x H)
Weight : 470g
Considerations
Tes t Frequency. The test frequency is user selectable and can be
changed. G ene rally, a 1 KHz test signal or higher is used to measure
capacitors that are 0.01uF or smaller and a 120Hz test signal is used
for capacitors that are 10uF or larger. T ypically a 1 kHz test signal or
higher is used to measure induc tors that are used in audio and RF
(radio frequency) circuits. This is because these components operate
at higher frequencies and require that they be measure d at a hi gher
frequency of 1 KHz. Generally, inductors below 2mH should be
measured at 1 kHz and inductors above 200H should be measured at
120Hz.
It is best to check with the componen t manufacturers’ data sheet to
determine the best test fre quency for the device.
C to 70°C (Operating)
°
C to 70°C (Storage)
approximately
adapter
6.9” x 3.4” x 1.9”
17
www..com
information@itm.com1.800.561.8187
Charged Capacitors
Always discharge any capacitor prior to making a
measurement since a charged ca pacitor may serious ly damage
the meter.
Effect Of High D on Accuracy
A low D (Dissipation Factor) reading is desirable. Electrolytic
capacitor s i nherently have a higher dissipation factor due to their
normally high internal leakage characteristics. If the D (Dissipation
Factor) is ex c es sive, the capacitance measurement accuracy may be
degraded.
It is best to check with the componen t manufacturers’ data sheet to
determine the desirabl e D val ue of a good component.
Measuring Capacitance of Cables, Switches or Other Parts
Measuring the capacitance of coaxial cables is very useful in
determining the actual length of the cable. Most manufacturer
specifications list the amount of capacitance per foot of cable and
therefore the length of the cable can be determined by measuring the
capacitance of that cable.
For examp le: A manufa cturers , specificat ion calls out a certa in cable,
to have a capa citance of 10 pF per foot, After measuring the cable a
capacitance reading of 1.000 nF is displayed. Dividing 1000pF
(1.000 nF) by 10 pF per foot yields the length of the cable to be
approximately 100 feet.
18
www..com
information@itm.com1.800.561.8187
Even if the manufacturers’ specification is not known, the
capacitance of a measured length of cable (such as 10 feet) can be
used to determine the capacitance per foot; do not use too short a
length such as one foot, because any error becomes magnified in the
total leng th calculations.
Sometimes , the cap aci tan ce of switches, inter co n n ect cables, circuit
board foils, or other parts, affecting stray capacitance can be critical
to circuit design, or must be repeatable from one unit to anot her.
Series Vs P a ra llel Measurem en t (for Inductors)
The seri es mode dis plays the more accurate me asurement in most
cases. The series equivalent mode is essential for obtaining an
accurate Q reading of l ow Q inductors. Where ohmic losses are
most significant, the series equivalent mode is preferre d. However,
there are cases where the parallel equivalent mode may be more
appropriate. For iron core induct or s operating at higher fre quenci es
where hysteresis and eddy currents become significant,
measurement in the parallel equivalent mode is preferred.
1.4 Accessories
Operating Manual 1 pc
2 AA Size Ni-Mh Rechargeable Batteries 2 pc
Shorting Bar 1 pc
AC to DC Adapter 1 pc
TL885A SMD Test Probe 1 pc
TL885B 4-Wire Test Clip (Optional)
TL08C Kelvin Clip (Optiona l)
19
www..com
information@itm.com1.800.561.8187
Carrying Case (Optional)
20
www..com
information@itm.com1.800.561.8187
2. Operation
1. NA
2. Primary Parameter Display
5. Model Number
6. Power Switch
7. Relative Key
8. Measurement Level Key
9. Open/Short Calibration Key
10. Measurement Frequency Key
13. Range Hold Key
14. L/ C/ Z /DCR Function Key
15. Bat tery Charge I ndicator
16. DC Adapter Input Jack
17. Guard Termina l
18. HPOT/HCUR Terminal
19. LPOT/LCUR Terminal
20. Battery Compartment
2.1 Physical Description
G
H
POT
CUR
UARD
L
POT
LH
CUR
G
UARD
3. Secondary Parameter Display 4. Low Battery Ind icator
12. D/Q/
θ
11. Display Update Speed Key
/ESR Function Key
21
www..com
information@itm.com1.800.561.8187
2.2 Making Measurement
1
Screws
Battery Compartment
4
Norm/Ni-Mh Switch
5
Back Case
6
Tilt Stand
2.2.1 Battery Replacement
When the LOW BATTERY INDICATOR lights up duri ng normal
operation, the batteries in the Models 885 & 886 should be replaced
or recharged to maintain proper operation. Please perform the
following steps to change the batteries:
1. Remove the battery hatch by unscrewing the screw of the
battery compartment.
2. Take out the old batteries and insert the new batteries int o the
battery compartment. Please watch out for battery polarity
when installing new batteries.
3. Replace the battery hatch by rever sing the procedure use d to
remove it.
2
Hatch
3 Batteries
Battery Replacement
22
www..com
information@itm.com1.800.561.8187
23
2.2.2 Battery Recharging/AC operation
!
rechargeable batteries.
Caution
Only the Models 885 or 886 st andar d acces sory AC t o DC
adapter can be used with Model 885. Other battery elimina tor
or charger may result in damage to Modes 885 or 886.
The Models 885 & 886 works on external AC power or internal
batteries. To power the Mode l 885 wi th A C source, make s ure t hat
the Models 885 or 886 is off, then plug one end of the AC to DC
adapter into the DC jack on the right si de of the instr ument and the
other end into an AC outlet.
There is a small slide switch inside the battery compartment
called Battery Select Switch. If the Ni-Mh or Ni-Cd rechargeable
batteries are installed in Models 885 or 886, set the Batter y Select
Switch to "Ni -Mh" posi tion. The Ni-Mh or Ni-Cd batteries can be
recharged when the instrument i s oper at ed by AC source. The LED
for indicating battery charging will light on. If the non-rechargeable
batteries (such as alkaline bat teries) are i nstalled in Models 885 or
886, set the Battery Select Switch to "NORM" position for
disconnecting the charging circuit to the batterie s.
Warning
The Battery Select Switch must be set in the "NORM"
position when using nonNon-rechargeable batt erie s may e xplode if the AC a dapter is
used with non-rechargea ble batteries. Warranty is voided if
this happened.
www..com
information@itm.com1.800.561.8187
24
2.2.3 Open and Short Calibrati on
The Models 885 & 886 provides open/short calibration
capability so the user can get better accuracy in measuring high and
low impedance. We recommend that the user performs open/shor t
calibratio n if the test level or frequency has been change d.
Open Calibration
First, remain ing the m easurem ent terminals with the open status,
then press the CAL key shortl y (no m ore t han tw o second) , the
LCD will display:
This calibration ta kes about 10 se con ds. After it is finished, the
Model 885 will beep to show that the calibration is done.
Short Calibration
To pe rform the short calibration, insert the Shorting Bar into the
measurement terminals. Press the CAL key for more than two
second, the LCD will display:
This calibration ta kes about 10 se con ds . After it i s fi nishe d, t he
Model 885 will beep to show that the calibration is done.
www..com
information@itm.com1.800.561.8187
25
2.2.4 Display Speed
The Models 885 & 886 provides two different display speeds
(Fast/Slow). It is controlled by the Speed key. When the speed is set
to fast, the display will update 4.5 readings every second. When the
speed is set to slow, it’s only 2.5 readings per second.
2.2.5 Relative Mod e
The relative mode lets t he user to make quick sort of a bunch of
components. First, insert the standard value compone nt to get the
standard value reading. (A pproximate ly 5 seconds in Fast Mode to
get a stable reading.) Then, press the Relative key, the primary
display will reset to zero. Remove the standard value component
and insert the unknown component, the LCD will show the value
that is the difference between the standard value and unknown
value.
2.2.6 Range Hold
To set the range hold, insert a standard component in that
measurement range. (A pproximately 5 seconds in Fa st Mode to get
a stable reading.) Then, by pressing the Range Hold key it will hold
the range within 0.5 to 2 times of the curr ent measurement range.
When the Range Hold is press the LCD disp lay:
www..com
information@itm.com1.800.561.8187
26
2.2.7 DC Resistance Measurement
The DC resistance measurement measures the resistance of an
unknown component by 1VDC. Select the L/C/Z/DCR key to make
the DCR measurement. The LCD display:
2.2.8
AC Impedan ce Mea s u r em ent
The AC impedance measurement measures the Z of an unknow n
device. Sele ct th e L/C/Z/DCR key to mak e th e Z m eas urem en t. Th e
LCD display:
The testing level and fre quency can by selected by pressing the
Level key and Frequency key, respectively.
2.2.9 Capacitance Measurement
To measure the capacitance o f a com ponent, select the L/C/Z/DCR
key to Cs or Cp mode. Due to the circuit structure, t here are two
modes can by selected (Serial Mode – Cs and Parallel Mode – Cp).
If the serial mode (Cs) is selected, the D, Q and ESR ca n be shown
on the secondary display. If the parallel mode (Cp) is selected, only
the D and Q can be shown on the secondary display. The following
www..com
information@itm.com1.800.561.8187
27
shows some examples of capacitance measurement:
The testing level and fre quency can by selected by pressing the
Level key and Frequency key, respectively.
2.2.10 Inductance Measurement
Select the L/C/Z/DCR key to Ls or Lp mode for measuring the
inductance in serial mode or parallel mode. If the serial mode (Ls) is
selected, the D, Q and ESR can be shown on the seconda ry display.
If the parallel mode (Lp) is selected, only the D and Q can be shown
on the secondary display. The following shows some examples of
capacitance measurement:
The testing level and fre quency can by selected by pressing the
Level key and Frequency key, respectively.
www..com
information@itm.com1.800.561.8187
28
2.3 Accessory Operation
Follow the figures below to attach the test probes for making
measurement.
Shorting Bar
TL885A SMD Test Probe
www..com
information@itm.com1.800.561.8187
29
H
H
P
C
C
L
L
P
TL885B 4-Wire Test Clip
TL08C Kelvin Clip
www..com
information@itm.com1.800.561.8187
30
4. Application
R
H
CUR
H
POT
DUT
(b) BLOCK DIAGRAM
DUT
V
A
Co
o
L
o
RoL
o
(a) CONNECTION
(c) TYPICAL IMPEDANCE MEASUREM ENT RANGE(£[)
2T
1m 10m 100m 1 101K 10K 100K 1M10010M
L
POT
L
CUR
4.1 Test Leads Connection
Auto balancing bridge has four terminals (H
) to connect to the device under test (DUT). It i s important to
L
POT
CUR
, H
POT
, L
CUR
and
understand what connection method will affect the measurement
accuracy.
2-T e rminal (2T)
2-T erminal is the easiest way to connect the DUT , but it contents
many errors which are the inductor and resi stor as well as the
parasitic capacitor of the test leads (Figure 3.1). Due to these
errors in measurement, the effective impedance measurement
Ω
range will be limited at 100
to 10KΩ.
Figure 3.1
3-T e rminal (3T)
www..com
information@itm.com1.800.561.8187
31
3-Terminal uses coaxial cable to reduce the effect of the
DUT
V
A
(d) 2T CONNECTION WITH SHILDING
H
CUR
H
POT
DUT
(b) BLOCK DIAGRAM
DUT
V
A
Co
Ro
Lo
RoL
o
Co doesn't
effect
measurement
result
(a) CONNECTION
(c) TYPICAL IMPEDANCE MEASUREMENT RANGE(£[)
3T
1m 10m 100m 1 101K 10K 100K 1M10010M
L
POT
L
CUR
parasitic capacitor (Figure 3.2). The shield of the coaxi al cable
should connect to guard of the instrument to increase the
Ω
measurement range up to 10M
.
Figure 3.2
4-T e rminal (4T)
4-Terminal connection reduces the effect of the test lead
www..com
information@itm.com1.800.561.8187
32
resistance (Figure 3.3). This connection can improve the
H
CUR
H
POT
DUT
(b) BLOCK DIAGRAM
DUT
V
A
(a) CONNECTION
(c) TYPICAL IMPEDANCE MEASUREM ENT RANGE (£[)
4T
1m 10m 100m 1 101K 10K 100K 1M10010M
L
POT
L
CUR
Ω
measurement range down t o 10m
. However, the effect of the
test lead inductance can’t be eliminated.
Figure 3.3
5-T e rminal (5T)
5-Ter mi nal connection is the combination of 3T and 4T (Figure
3.4). It has four coaxial cables. Due to the advanta ge of the 3T
and 4T, this connection can widely increase the measurement
Ω
range for 10m
to 10MΩ.
www..com
information@itm.com1.800.561.8187
33
(d) WRONG 4T CONNECTION
H
POT
DUT
(b) BLOCK DIAGRAM
(a) CONNECTION
(c) TYPICAL IMPEDANCE MEASUREM ENT RANGE (£[)
5T
1m 10m 100m 1 101K 10K 100K1M10010M
H
CUR
DUT
V
A
DUT
V
A
L
POT
L
CUR
Figure 3.4
4-T e rminal Path (4TP)
4-Ter minal Path connection solves the proble m that caused by
the test lead inductance. 4TP uses four coaxial cables to isolate
the current path and the vol tage sense cable (Figure 3.5). The
return current will flow through t he coaxial ca ble as well as the
shield. Theref ore, the magnetic flux that generated by i nternal
conductor will cancel out the magnetic flux generated by
external conductor ( shield). The 4TP connection increase s the
www..com
information@itm.com1.800.561.8187
34
measuremen t range from 1mΩ to 10MΩ.
(b) BLOCK DIAGRAM
(a) CONNECTION
DUT
V
A
(c) TYPICAL IMPEDANCE
MEASUREMENT RANGE(£[)
4T
1m 10m100m 1 101K 10K 100K1M10010M
H
POT
DUT
H
CUR
L
CUR
L
POT
H
POT
DUT
H
CUR
L
CUR
L
POT
(d) 4T CONNECTION W ITH SHILDING
Figure 3.5
Eliminating the Effect of the Parasitic Capacitor
When measuring the high impedance component (i.e. low
capacitor), the par asitic capacitor becomes an important issue
(Figure 3.6). In figure 3.6(a), the parasitic capacitor Cd is
paralleled to DUT as well as the Ci and Ch. To correct this
problem, add a guard pla ne (F igure 3.6(b)) in be twee n H and L
terminals to break the Cd. If the guard plane is connected to
instrument guard, t he effect of Ci a nd Ch will be removed.
www..com
information@itm.com1.800.561.8187
35
HCURHPOTLPOTLCUR
HCURHPOTLPOTLCUR
Cd
DUT
ChCl
(a) Parastic Effect
Guard
Plant
Connection
Point
Ground
(b) Guard Plant reduces
Parastic Effect
Figure 3.6
4.2 Open/Short Compensation
For those precision impeda nce mea suri ng instr ument, the ope n and
short compensation nee d to be used to reduce the parasitic effect of
the test fixture. The parasitic effect of the test fixture can be treated
like the simple passive components in figure 3.7(a). When the DUT
ω
is open, the instrument gets the conductance Yp = Gp + j
Cp
(Figure 3.7(b)). When the DUT is short, the instrument gets the
ω
impedance Z s = Rs + j
Ls (Figure 3.7(c)). After the open and short
compensation, Yp and Zs are for calculating the real Zdut ( Figure
3.7(d)).
www..com
information@itm.com1.800.561.8187
36
H
CUR
H
POT
L
CUR
L
POT
Zdut
C
o
RsL
s
G
o
Z
m
Redundant
Impedance
(Z
s
)
Parastic
Conductance
(Y
o
)
Parastic of the Test Fixture
(a) Parastic Effect of the Test Fixture
H
CUR
H
POT
L
CUR
L
POT
C
o
R
s
L
s
G
o
(b) OPEN Measurement
Y
o
OPEN
Y
o
= Go + j£sC
o
1
(R
s
+ j£s<< )
G
o
+j£sC
o
H
CUR
H
POT
L
CUR
L
POT
C
o
R
s
L
s
G
o
(c) SHORT Measurement
Z
s
SHORT
Z
s
= Rs + j£sL
s
Figure 3.7
www..com
information@itm.com1.800.561.8187
37
Zs
Zm
YoZdut
(d) Compensation Equation
Zm - Zs
Zdut =
1-(Z
m-Zs)Yo
Figure 3.7 (Continued)
4.3 Selecting t h e Series or Parallel Mode
According to different measuring requirement , there are series
and parallel modes to describe the measurement result. It is
depending on the hi gh or low i mpeda nce va lue t o deci de what
mode to be used.
Capacitor
The impedance and capacitance in the capacitor are negatively
proportional. Therefore, the large capacitor means the low
impedance; the small capacitor means the high impedance.
Figure 3.8 shows the equivalent circuit of capacitor. If the
capacitor is small, the Rp is more i mportant than the Rs. If the
capacitor is large, the Rs shouldn’t be avoided. Hence, uses
parallel mode to measure low capacitor and series mode to
measure high capacitor.
www..com
information@itm.com1.800.561.8187
38
R
C
R
Effect
No Effect
Large capacitor
R
P
C
R
Effect
Figure 3.8
Small capacitor
(High impedance)
(Low impedance)
P
S
Inductor
No Effect
S
The impedance and inductive in the inductor are positively
proportional. Therefore, the large induct or equals to the high
impedance and vice versa. Figure 3.9 shows the equivalent
circuit of inductor. If the inductor is small, the Rs is more
important than the Rp. I f the inductor is large, the Rp should be
taking care of. So, uses series mode to measure low inductor and
parallel mode to measure hi gh inductor.
www..com
information@itm.com1.800.561.8187
39
R
P
L
R
(High impedance)
No Effect
R
P
L
R
Large inductor
Effect
S
Small inductor
(Low impedance)
No Effect
S
Effect
Figure 3.9
www..com
information@itm.com1.800.561.8187
40
5. Limited Three-Year Warranty
B&K Precision Corp. warrants to the original
purchaser that its prod ucts and the component parts
thereof, will be free from defects in workmanship and
materials for a period of three years
purchase.
B&K Precision Corp. will, without c harge, repair or
replace, at its option, defective product or component
parts. Returned product must be accompanied by
proof of the purchase date in the form of a sales
receipt.
lusions: This warranty does not apply in the event of misuse or
Exc
abuse of the product or as a result of unauthor ized alternations or
repairs. It is void if the serial number is alternated, defaced or
removed.
B&K Precision Corp. shall not be liable for any consequential
damages, including wit hout limitati on damages resul ting from loss
of use. Some states do not allow limitation of incidental or
consequential da mages, so the above limitation or exclusion may
not apply to you.
This warranty gives you specific rights and you may have other
rights, which vary from state-to-state.
from date of
www..com
information@itm.com1.800.561.8187
41
Service Information
Warranty Service:
packaging with proof of purchase to the below address. Clearly
state in writing the performance problem and return any leads,
connectors and accessories that you are using with the device.
Non-Warranty Service:
packaging to the below address. Clearly state in writing the
performance problem and return any leads, connectors and
accessories that you a re using with the device. Cust omers not on
open account must include payment in the form of a money order or
credit card. For the most c urrent repair c harges contact the fact ory
before shipping the product.
Return all merchandise to B&K Precision Corp. with pre-paid
shipping. The flat-rate repair charge includes return shipping to
locations in North America. For overnight shipments and non-North
America shipping fees c onta ct B&K Precision Corp..
I
nclude with the instrument your complete return shipping address,
contact name, phone number and description of problem.
Please return the product in the original
Return the product in the original
www..com
information@itm.com1.800.561.8187
42
6. Safety Precaution
!
SAFETY CONSIDERATIONS
The Models 885 & 886 LCR Meter has been designed and tes ted
according to Class 1A 1B or 2 according to IEC479-1 and IEC
721-3-3, Safety requirement for Electronic Measuring Apparatus.
SAFETY PRECAUTIONS
SAFETY NOTES
The following general safety preca utions must be observed during
all phases of operation, service, and repair of this instrument.
Failure to comply with these precauti ons or with specific war nings
elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the instr ume nt.
The manufacturer assumes no l iability for the customer‘s failure to
comply with these requirements.
BEFORE APPL YING POWER
Verify that the product is set t o match the available line voltage is
installed.
www..com
information@itm.com1.800.561.8187
43
SAFETY SYMBOLS
!
Caution, risk of electric shock
Earth ground symbol
Equipment protected throughout by double
insulation or reinforced insulation
Caution (refer to accompanying documents)
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not
install substitute part s or perform any una ut horize d modificat ion to
the instrument. Return the instrument to a qualified dealer for
service and r ep ai r to en su r e th at s a fe ty f eatures are maintained.
INSTRUMENTS WHICH APPEAR DAMAGED OR
DEFECTIVE SHOULD NOT BE USED! PLEASE CONTACT
B&K PRECISION FOR INFORMATION O N REPAIRS.
www..com
information@itm.com1.800.561.8187