The hardwareand/orsoftwaredescribed herein are furnished underalicenseand maybeused orcopied onlyin accordancewith thetermsofsuchlicense.
SafetyNotices
ACAUTIONsigndenotes ahazard.Itcallsattention to anoperating procedure orpracticethat,ifnotcorrectly performedoradhered to,could resultindamage to the productorloss ofimportantdata.Do notproceedbeyond aCAUTIONsign untilthe indicatedconditionsare fullyunderstood andmet.
AWARNINGsigndenotesahazard.
Itcalls attention to anoperating procedure orpracticethat,ifnotcorrectly performedoradhered to,could resultinpersonalinjury ordeath.Do notproceed beyond a WARNINGsign untilthe indicatedconditions are fully understoodand met.
Restricted RightsLegend
Restricted permissions ofthe U.S.government.Permissionsforsoftware and technicaldata whichare authorized to the U.S. Government onlyincludethose forcustomprovisionto end users.ITECHfollowsFAR12.211 (technicaldata),12.212 (computersoftware).DFARS252.227-7015 (technicaldata--commercialproducts)fornationaldefense and DFARS227.7202-3(permissions forcommercialcomputersoftware orcomputersoftware documents)while providing thecustomized businesslicensesofsoftware and technicaldata.
NOTE
ANOTEsigndenotesimportant hint. It calls attentionto tipsorsupplementary information that is essential forusers to referto.
We certifythatseriesIT8500+electronicload meetsallthe published specificationsattime ofshipmentfromthe factory.
Warranty
ITECHwarrantsthatthe productwillbe free fromdefectsin materialand workmanship undernormaluse fora period ofone (1)yearfromthe date ofdelivery(exceptthose described in the Limitation ofWarrantybelow).
Forwarrantyservice orrepair,the productmustbe returned to a service centerdesignated byITECH.
The productreturned to ITECHforwarrantyservice mustbe shipped
PREPAID. And ITECH will pay forreturn of the product to customer.
Ifthe productisreturned to ITECHforwarrantyservice fromoverseas,all
the freights,dutiesand othertaxesshallbe on the accountofcustomer.
Limitation of Warranty
ThisWarrantywillbe rendered invalid in case ofthe following: Damage caused bycircuitinstalled bycustomerorusing customerown
The following safetyprecautionsmustbe observed during allphasesofoperation ofthisinstrument.Failure to complywith these precautionsorspecificwarningselsewhere inthismanualwillconstitute a defaultundersafetystandardsofdesign,manufacture and intended use ofthe instrument.ITECHassumesno liabilityforthe customer’sfailure to complywith these precautions.
SeriesIT8500+electronicload supports110V/220VACinputand need to
switch the inputvoltage before operation.
Do notuse the instrumentifitisdamaged.Before operation,checkthe
casing to see whetheritcracks.Do notoperate the instrumentin the presence ofinflammable gasses,vaporsordusts.
Theelectronicload isprovided with a three-core powerline during delivery
and should be connected to a three-core junction box.Before operation,besure thatthe instrumentiswellgrounded.
Make sure to use the powercord supplied byITECH.Checkallmarkson the instrumentbefore connecting the instrumentto
powersupply.
Use electricwiresofappropriate load.Allloading wiresshould be capable
ofbearing maximumshort-circuitcurrentofelectronicload withoutoverheating.Ifthere are multiple electronicloads,each pairofthe powercord mustbe capable ofbearing the full-loaded rated short-circuitoutputcurrent
Ensure the voltage fluctuation ofmainssupplyislessthan 10%ofthe
working voltage range in orderto reduce risksoffire and electricshock.
Do notinstallalternative partson the instrumentorperformany
unauthorized modification.
Do notuse the instrumentifthe detachable coverisremoved orloosen.To preventthe possibilityofaccidentalinjuries,be sure to use the power
adaptersupplied bythe manufactureronly.
Neveruse theinstrumentwith a life-supportsystemoranyotherequipment
subjectto safetyrequirements.
Failure to use the instrumentasdirected bythe manufacturermayrender
itsprotective featuresvoid.
Alwaysclean the casing with a drycloth.Do notclean the internals.Make sure the venthole isalwaysunblocked.
The CEmarkindicatesthatthe productcomplieswith allthe relevantEuropean
galdirectives.The specificyear(ifany)affixed refersto the yearwhenthe design wasapproved.
The instrumentcomplieswith the WEEEDirective (2002/96/EC)marking
productlabel
indicatesthatyou mustnotdiscard the
trical/electronicproductin domestic
Thissymbolindicatesthe time period during which no hazardousortoxicsubstancesare expected to leakordeteriorate during normaluse.The
roductcan be used safely
yearEnvironmentFriendlyUse Period (EFUP).Upon expiration ofthe EFUP,the productmustbeimmediatelyrecycled.
(WEEE)Directive
Thisproductcomplieswith the WEEEDirective (2002/96/EC)marking requirement.Thisaffixproductlabelindicatesthatyou
household waste.
ProductCategory
Environmental Conditions
The instrumentisdesigned forindooruse and an areawith lowcondensation.The table belowshowsthe generalenvironmentalrequirementsfortheinstrument.The speed offan willchange intelligentlybythe temperature ofradiator.When the temperature isup to 40°C,the fan willbeon and adjustintelligentlywhen temperature changes.
SPECIFICATIONSOFRED AND BLACK TESTLINES.............................................................................................................. 55
Itis the testreportofthe instrumentbefore delivery.
Chapter1Inspection and Installation
1.1 Verifying the Shipment
Unpackthe boxand checkthe contentsbefore operating the instrument.Ifwrong itemshave been delivered,ifitemsare missing,orifthere isa defectwith the appearance ofthe items,contactthe dealerfromwhich you purchasedthe instrumentimmediately.The package contentsinclude:
The instrumentshould be installed atwell-ventilated andrational-sized space.Please selectappropriate space forinstallation based on the electronicload size.
IT8500+serieselectronicload differentmodelsare notthe same size,the detailsize ofthe electronicload are shown asbelow.
powercordthatmatchesthe specificationsofpowersocketused inthe area.See the SectionConnecting thePower
IT8511+/IT8511A+/IT8511B+/IT8512+/IT8512A+/IT8512B+/IT8512C+/IT8512H/IT8513A+/IT8513C+seriesloadsare equipped with ahandle forusertoeasilycarryand place it.
Inspection and Installation
The load handle maybe adjusted based on three methods(asshown in iconsbelow).Be sure thatappropriate force isapplied to adjustthe load handle to appropriate position.
1.4 Disassembly of Load Handle
Please disassemble the load handle before installing equipmenton the support.Disassemblysteps:
1.Adjustthe handle to the position asshown in the figure below.
NOTE
Toeasilydisassemblehandle,alignthelockingmouth and lockingdevice,whichisbetweenthe handleandtheinstrument.
2.Align the locking mouth,and pulloutthe handle towardstwo sides.
NOTE
Donotusetoomuchforceand mind yourhandsduring disassemblyofloadhandle.
IT8511+/IT8511A+/IT8511B+/IT8512+/IT8512A+/IT8512B+/IT8512C+/IT8512H+/IT8513A+/IT8513C+loadscan be installed on standard 19-inch rack.ITECH providesuserwith IT-E151/IT-E151Arack,asan optionalmountkit.The detailed operation please referto the UserManualofyourmountkit.
IT8514B+/IT8514C+/IT8516C+need notmounton rack,theycan installed on cabinetdirectlybyscrew.
1.6 Connecting the Power Cord
Connectthe powercord afterchecking thatthe powerswitch ofthe instrumentisturned OFF.Onlyuse the powercord supplied asa standard accessory.
Selectfromthe following Schedule ofPowerCord Specificationsan appropriate powercord thatmatchesthe voltage forthe area inwhich you use the instrument.Ifthe powercord included in the instrumentyou purchased doesnotmatch the voltage,contactthe dealerormanufacturerforchange.
Thischapterintroducesthe frontpanel,the rearpanel,keyfunctionsand VFDdisplayfunction ofthe electronicload,make sure thatyou can quicklyknowtheappearance,instruction and the keyfunction before you operate the load,Help you make betteruse ofthisseriesofelectronicload.
2.1 Brief Introduction
IT8500plusseriesDCelectronicloadsare single channelprogrammable electronicload which can provide multiple solutionsaccording to therequirementsofyourdesign and test.Thisserieshave internationaladvanced functionsand features.
High-visibilityvacuumfluorescentdisplay(VFD)Measurementresolution:0.1mV,0.1mAVoltage and currentMeasurementspeed:up to 40KHZFouroperation modes:CV(ConstantVoltage),CC,CR,CWBattery test functionOCP test , OVP test Auto test function: ShortcircuitfunctionRemote Sense functionMemorycapacityto save/recallsetting parameters:100 registersIntelligentfansBuild-in BuzzerasalarmsignalPoweroffmemoryfunction Rotaryknob,making the operation more convenientMeasure testfunction,testtherising/dropping time ofthe voltage Listmodes
Thischapterelaborateson the functionsandcharacteristicsofelectronicloads.Contentsfollowing sections:
Switching oflocal/remote operation modesConstant-statusoperation modeInputOn/Offfunction Keyboard locking function
Short-circuitanalog functionSystemsetup function
Triggering function ListmodeTestfunction Save/RecallVON functionFullprotection function,OCP,OVP,OTP,OPP,Reverse voltageRemote Sense functionCurrentmonitoring function
Ripple function
3.1 Local Mode/Remote Mode
There are two typesofcontrolmodesforIT8500+seriesproducts:Localmodeand Remotemode.
In remote mode,you can operate the electronicloadsthrough PCvia communication cable(optional).WhileAfterpoweron electronicloads,itdefaultsin localmode and allbuttonsare avaiable in thismode.In remote controlmode,the keyson the frontpannelcan notworkexceptlocalkey.Customerscould through [ Local ] key to switch the control mode.
3.2 Operation Mode
There are fouroperation modesofIT8500+seriesproducts:
In constantpowermode,the DCload willcause a constantpowerto bedissipated in the load.Asshown below,the load currentisdecreasing with the rising ofinputvoltage,while poweralwaysmaintain the setting value.
Functionsand Characteristics
POWER-ON
Power on state of instrument
RST(default)
Do notrememberstateinSAVE0.Customercansavethe DCload nexttime.
SAV0
Rememberstate inSAVE0
BUZZER
ON(default)
Enable audible beep whenkeyispressed
OFF
No sound whenkeyispressed
KNOB
Diagram3-4 I-Vcurve in CPmode
3.3 Input On/Off Control
[ On/Off ]button on the frontpanelisused to manuallytoggle the instrument
between itssetmode and an infinite impedance state,i.econtrolinputon andoff.[ On/Off ] button lightedindicatesthe load inputison,meanwhile the[OFF ]indicatorwilldisappear.
3.4 Short-circuit Analog Function
Shortcircuitsimulation and shortcircuitcurrentmeasurement:you maypress[ Shift ] +[ 1]botton to emulate a shortstate.Itcan be used to checkwhetherthe tested instrument’sshortprotection isavaiable.
In shortmode,the DCload willdrawmaximumcurrentfromthe DCsupplyinanyofthe fouroperation modes(CC,CV,CWorCR).In CC,CV,orCRmode,you maypress[ Shift ] +[1]to stop short.The DCload willreturn to itspreviousoperation.However,in CWmode,the shortcurrentwillcontinue to be drawn.To stop the short,you mustpressthe[On/Off]keyafteryou press[Shift ] + [ 1 ].
When emulating a Shortin CC,CWorCRmode,the maximumallowable shortcurrentisequalto the 110%ofcurrentrange.UnderCVmode,shortcircuitcurrentisequivalentto thatconstantvoltage value ofload is0 V.
A-LIMITSet software current protecting stateA-LIMIT
ONEnable software overcurrentprotectionfunctionA-LIMPOIN=30.000ASet the software OCP levelA-LIMDELAY=3SSet the OCP delay timeOFFDisable the software OCPfuntion
P-LIMITSetsoftware powerprotectingstate.P-LIMPOIN=150.00WSet the software OPP level.
P-LIMDELAY=3SSet the OPP delay time.TIMERSet load on timer
The transienttestallowsswitching between two differentload values.Acommon application isto testthe dynamiccharacteristicsofDCsource.
There are three differenttypesoftransientoperation:continuous,pulse,
toggled.
3.8.1 Continuous Mode
In continuoustransientoperation,theload is continuouslyswitched between two load values.An example isshown in the following figure:
3.8.2 Pulse Mode
In pulse operatrion,the load operatesatthe Avalue thathasbeen entered untila triggerisreceived.Atthe trigger,the load switchesto the Bvalue and staysatthatlevelforthe Btiming value.Then the load switchesbackto the Avalue and staysthere untilanothertriggerisreceived.
In toggled transientoperation,the load startsatthe stored parametersforthemode.When a triggerisreceived,the load switchesto Bvalue.When anothertriggerisreceived,the load switchesto the Alevel.Itstaysatthe Avalue untilanothertriggerisreceived,atwhich pointitswitchesto the Bvalue.Here’sanexample:
3.9 LIST Operation
Listmode allowsyou to generate a complexcurrentsequence.Moreover,themode.
change can be synchronized with an internalorexternalsignal,to accomplish dynamicand precise testAlistfile includesfollowing parameters:filename,
stepcounts (range 2-84),time widthofsingle step(0.00005s~3600s),stepvalue andslope.The edited listfile can be saved in nonvolatile memory,can be
recalled easily.The DCload provides7 nonvolatile registersto save listfilesforrecalllater.
In listmode,the DCload startto run the listfile once receiving a triggersignal,continue to run once receiving anothertriggersignal.To illustrate the use ofa list,we’llcreate a listthatrunsthe following constantcurrentprofile on a powersupply:
3.10 Saving and Recalling Settings
Wecan save some often-used parametersin the non volatile memory,includingworking mode,voltage/currentvalue and so on.IT8500plusseriesprovide 100non-volatileregisters.
Theyare divided into 10 Memorygroups:Group0-9.You can setitin the system
Saving operationwilloverwriteanyvaluespreviouslysaved inthatregister.RecallingoperationwilllighttheEnter.Youshould press[ESC]toescape therecalling state
beforesettingotherparameters.
3.11 VON Function
The DCload can be setto onlyturn on ifthe voltage isabovea setvalue(VONset)underconfigure menu bypressing [ Shift ]+[9].There are two typesofVONfunction:LivingandLatch.The following willhave detailed description forthe two types.
NOTE
VONsetisusedtoensureanelectronicsystemundertestwillnothavepowerappliedunless thesupply voltage is above a certain value.If you have nosuch testing request, donotsetthisvaluearbitrarily.Ifyourinstrumentcannotworknormally,forexample,setCC=1A,afterturnontheinputwhilethecurrentisstill0Ainsteadofsettingvalue1A,thenyou should check VON set firstly.If VON set is not 0V, then please modify to 0V.
VON LIVING MODEIn Living mode,when powerisapplied to the DCload,thevoltage mustrise
above VONsetting before the load drawscurrentfromthe source.Ifthe voltge belowVONsetting on the load’sterminals,theload willturn offinput.
After above steps,press[](Trigger)button to run ocp testfile.
Recall OCPFile:
1.Press[ Shift]+[ Enter]button to selectprograme file,the paneldisplays“CALLOCPFILE=1.Enterthe file name(1-10),press[ Enter]button to confirm.
2.According to the following stepsto escape OCPmode:press[ Shift ]+[ 8](system)------“RUNMODE”-----[Enter]-------select“NORMAL”mode----[ Enter].
3.13 OPP Operation
OPPtestprocess:When the inputvoltage hasreached VONpoint,powerwill
begin to workaftera delaytime.The powervalue willincrease bya step size atregularintervals.Simultaneously,the DCload willjudge wetherthe inputvoltageislowerthan OPPvoltage(you need to set).Ifitis,then the presentcurrentvaluewillbe compared to see ifitisin the currentrange you’ve set,in thisrange,theOPPtestwillPassorfail.On the contrary,the powerwillcontinue to increasewithin the cut-offcurrentrange.And then compare OPPvoltage withinputvoltage too.
RUN <OCP_TEST
STOP0.000A
OPPTEST
To start a OPP test, press[ Shift ]+[](Trigger).Press [ Shift ]+[CW](OPP)to enterOPPtestoperation.
Afterabove steps,press[](Trigger)button to run OPPtestfile.
Recall OPPFile
1.Press[ Shift]+[ Enter] button to selectprograme file,the paneldisplays“CALLOPPFILE=1.Enterthe file name(1-10),press[ Enter] button to confirm.
2.According to the following stepsto escape OPPmode:press[ Shift ]+[8](system)------”RUNMODE”----[ Enter]------select“NORMAL”mode----[ Ente
r].
3.14 BatteryTest
IT8500plusseriesproductstestthe batterycapabilityin CC/CW/CRmode.The testmode should be setfirst,and then the discharge stop conditions.
There are three discharge stop conditionsto be setforIT8500 plusseriesproducts.When useronlyneed to dobatterytesting in one ortwo stopconditions,the otherconditionsshould be setto the specified value (STOPVOLT:0V;STOPCAP:999.999AH;STOPTIMER:99999S).When the systemchecksthe discharging time orbatteryvoltage orcapacityisequalto the setting stop value orunderaninsecuritystate,the batterytestwillstop,and the E-Loadwillturn off.The batteryvoltage,discharge current,discharge time and discharged capabilityare displayed on the VFDwhile testing.
[Shift ]+[ 8 ](system)---”RUNMODE”---[ Enter ]----select“NORMAL”mode---[ Enter ]
3.15 CR-LED Test Function
With adding ofdiode break-overvoltage setting in the IT8500+serieselectronicload underconventionalCRmode,the electronicload onlyworkswhen voltage applied atitsboth endsishigherthan the diode break-overvoltage to give a realsimulation ofdiode working principle,i.e.,the ripple currentatrealLED test.
The IT8500+serieselectronicload isprovided with specialvoltage rise/droptime measurementfunction.Thisfunction givesa simple analog ofvoltage rise/drop speed ofoscilloscope testpower.
6.Press[]keyto select“On”.Starttimerfunction and press[Enter]key.
7.Press[ESC] to exit setting.
8.VFDsecond line willdisplaytime 0.0000Sbetween powervalue and setvalue.
0.0001V0.0002A
0.00W0.0000SCC=0.000A
MeasurementofRise Time
9.ConnectDCpowerto be tested to the inputterminalofthe electronicload.The powerissetwith a value thatishigherthan the setfinalvoltage value.Keep poweroutputin OFFstatus.
10.Seta constantcurrentvalue on the load and open the load input.
11.Open poweroutput.
12.The electronicload timerstartstiming.Afterending,time willkeep stable,which isrise time ofvoltage.
13.Close the poweroutput.The electronicload VFDwilldisplayvoltage droptime.
When the DCload temperature dropsto the protecting point,pressanykeyon the frontpanel,the OTP)displayed on the frontpanelwilldisappear,the DCload exitsOTPprotection state.
3.17.5Reverse Voltage Protection(LRV)
Thisfeature protectsthe DCload in case the DCinputterminalsare connectedto a powersource with reversed polarity.Ifa reverse voltage condition isdetected,the buzzerwillsound andwillbe displayed on the VFD.
Remotesensing isusedto counteractthe effectoflead resistance.Forexample,ifyou connecta powersupplyto the DCLoad,the voltage atthepowersupply'sterminalswillnotbe the same asthe voltage atthe DCLoad'sterminalsifthere isa currentflowing because ofthe finite resistance fromthewires.Using remote sensing,you can sense the voltage atthe powersupply'sterminals,effectivelyremoving the effectofthe voltage drop in the connection wire.
When using remote sensing,the powerdisplayed bythe instrumentincludesboth the powerdissipated inside the instrumentandthe powerdissipated in the
leadsfromthe powersupplyto the DCLoad'sinputterminals.Stepsto enable remote sensing in the menu:
1.Press[Shift] +[9]keyinto the menu
2.VFDdisplays>,press[Enter] key to confirm
3.Press[]to choose > , press[Enter] key to confirm
4.Press[]to choose >,press[Enter]keyto confirm,thenremotesense function hasbeen set,and VFDdisplayindicator.
Remote Sensing:SENSE(+)and SENSE (–)are the remote sensing inputs. Byeliminating the effectofthe inevitable voltage drop in the DCload leads,remote sensing providesgreateraccuracybyallowing the DCload to regulate directlyatthe source'soutputterminals.
Wiring DiagramforRemote Sensing:
3.19.2 External Triggering
EXTERNAL:An externaltriggerisa TTLlowsignalapplied to the Trigger
connection on the backpanel.ThisTTLsignalmustlastformore than5 ms.Atriggerapplied to thisinputcan beused to change settings(voltage,current,resistance),toggle between settingsin transient-toggle mode,orgenerate a pulse in pulse mode.
Operation to selectthe triggersource asexternal:
[Shift] +[8](system)to enterthe menu,use []to select TRIGGER,press[Enter],and then selectEXTERNAL.Press [ESC] to exit the menu.
3.19.3CurrentMonitoring (IMonitor)
Currentmonitoring terminalwilloutput0-10Vanalog signalto corresponding to 0 tofullrange ofinputcurrent.You can connectan externalvoltmeteroran oscilloscope to displaythe inputcurrent’schanging
IT8500plusseriesDCelectronicloadshave testripple function.You can readripple voltage and ripple currentbysending instructions.See in IT8500+programming guide.
(Seta value from0 to upperlimitofpower)There are three waysto setthe powervalue:
1.InCWmode,rotate Rotarykonb.
2.In CWmode,inputvalue through numberkeyboardsdirectly,press[Enter]
to confirm
3.In CWmode,movethe stepping cursorbypressing [],and then
adjustthe powerbypressing [].
To setpowerrange,followthe steps:
[Shift] +[CV]
4.4 Constant Resistance Operation
(Allowed setting range is0.05Ωto 7500Ω)There are three waysto setthe resistance value:
1.In CRmode,rotate Rotaryknob.
2.In CRmode,inputvalue through numberkeyboardsdirectly,press[Enter]
to confirm
3.In CRmode,move the stepping cursorbypressing[],and then
adjustthe resistance bypressing []
To setresistance range,followthe steps:
Press [CR],then [Shift] +[CV]
4.5 Transient Test Operation
Transientoperation enablesthe DCload to periodicallyswitch between twolevels.
To edita transienttestfile,related parametersneed to be set:Alevel,Blevel,time width (onlyin pulse mode),frequency,dutyand running mode(Continuous/Pulse/Toggled).Ifin CCdynamicmode,usercan setcurrentrising and falling slope additionally.
Following isan example to illustrate the three transientoperations.
Press [Shift]+[2](Tran)to enterthe transientoperation setup menu
[] keyto selecton.,
press
to confirm.
TOGGLE(the
CONTINUOUS>
function,press[Enter]to
TRANand Trig
0.00W TRAN. 0
[Shift]+[] (Trigger)totriggertransienttest,ifyou wantto continue the testagain,please
repeat1-9 steps.
4.6 List Operation
Before run a listfile,you should editthe listfile firstlyand save itin a non-volatilememory.The following exampleswillhelpyou understand the function well.In the example,the outputvoltage and currentare 10Vand 3A,and the DCload is
Testfilesare a generalization oflists—theyletyou generate a sequence oftestsusing differentmodes,mode parameters,and durations.Theyare usefulforexecuting a setoftestson a device,then displaying whetherthe testspassed orfailed.We willillustrate howto use testfilesbya shortexample.
You can editup to 10groupsoftesting files,each file has10 steps,itcan editup to 100 stepswhich can be saved in EEPROM(address).
Suppose we have a smallACto DCpowersupply(a "wall-wart")and we wantto setup an acceptance testfora numberofthese devices.Ourtestwillconsistof two steps:
1.Setthe DCload to constantcurrentmodeto drawthe rated currentof1.2A
fromthedevice.The outputvoltage ofthe device atthe rated currentmustbe between 4.4Vand 4.6V.
2.Setthe DCload to constantvoltage 3V.The outputcurrentofthe device is
between 2Aand 3A.
Selectthe listfile.Press
to
3.When the device operatesinto a short,the supplied currentmustbe larger
than 3.0 A.
[Shift]+[6]
pauseduring the test.
=NNNNNNN32Y
When itis paused,press[]
press [3].Andpress [Enter]
=NNNNNNNY21
wantto load on 2S,press[2],and then press [Enter]toconfirm.
SetTon forthe second step,ifyou wantto load 2S,press
confirm
SEQ02 ON =2S
9
SetToffofthe second step,ifthenpress[Enter]to confirm.
SEQ02 OFF =2S
10
Settesting delaytime ofthesecondstep,e.g.1S,press
[1]
SEQ02 P/F =1S
11
SetTon forthe third step,if
confirm
SEQ03ON =3S
12
Set Toffofthe thirdstep,ifyou
[Enter]to confirm.
SEQ03 OFF =2S
13
Settesting delaytime ofthethird step, e.g. 2S,press [2].
SEQ03P/F =2S
14
setstartvoltage.Please refer
function ofauto start
voltage”.
AUTOSTART=0.000V
15
Setstop condition
STOP COMPFAILURE
[2],then press[Enter]to
you need 2S,press [2],and
you wantto load on 3S,press[3],then press [Enter]to
need 2S,press[2],then press
to “
Function ofAuto StartVoltage:
1.Auto start=0V.
Auto testfile startto run when receive a triggersignalbypressing [shift]+[trigge]rorproviding a externaltriggersignal.
2.Auto startisnotequalto 0V(Take 2Vasan example)
In this condition,useronlyneed to connectthe chargerto inputtermianlsofE-load.The unitcan auto startto run testfilewhen detecta rising edge from0-2V.Auto startvoltage isnotsuggested to be a big value.2Vissuitable.
DB9 inthe rearpanelofthe DCload could connectwithRS-232 throughonTTLconnector.Thefollowing description may helpyou to knowhow to control the output of the DC loadthrough PC.
WARNING:Don’tconnecttheDCload’sDB9connectortoastandardRS232instrument;doingsomay damage the instrument.
5.1Communication Modules Intruduction
IT-E121 Communication Module
The DB9 interfaceconnectoron the rearpanelofthe DCload isTTLvoltage level;you can use the communication module IT-E121 and an a standardRS232 extension cable to connectthe DB9 interface connectorofthe DCloadand the RS-232 interface connectorofcomputerforthe communication.
IT-E121ACommunication Module
The DB9 interfaceconnectoron the rearpanelofthe DCload isTTLvoltage level;you can use the communication module IT-E121Aand an a standardRS232 extension cable to connectthe DB9 interface connectorofthe DCloadand the RS-232 interface connectorofcomputerforthe communication.
IT-E121Aisderived on the basisofIT-E121,the main difference between themisthatthe DB9 interfaceconnectorofthe RS232changesfromfemale to male ,so thatcan be directlyconnected tothe standard LANinterface.
IT-E121Acommunicationcable
IT-E122 Communication Module
TheDB9 interfaceconnectoron the rearpanelofthe DCload isTTLvoltage level;
IT-E122 hasa USBinterface on one end,you can use IT-E122 and an a standard USBextension cable to connectthe DB9 interface connectoroftheDCload and the USBinterface connectorofcomputerforthe communication.
IT-E123 Communication Module
The DB9 interfaceconnectoron the rearpanelofthe DCload isTTLvoltage level;
The interface on both side portofIT-E123 are DB9 interface and RS485 interface,you
can use the communication module IT-E123 and an a standard RS485-RS232conversion cable to connectthe DB9 interface connectorofthe DCload andthe RS-232 interface connectorofcomputerforthe communication.
RS485 pins
5.2 Communication with PC
Before using the remote operation mode,please make sure thatthe baud rateand communication addressin the DCload are the same asin the computersoftware,otherwise,the communication willfail,you can change the baud rateand communication addressfromthe frontpanelorfromcomputer.
IT8514B+/IT8514C+/IT8516C+serieselectronicload have a DB9 interface on rearpanel.ConnectE-load and computerbycable ofCOMends(DB9).Composite key[Shift]+[8]on frontboard can be used to entersystemmenu foractivation.
RS-232 InterfaceIn RS-232 interface,allSCPIcommandscan be usedforprogramming.If
RS-232 interface isselected,in accordance with internalconnection ofdataterminalequipment(DTE)and data communication equipment(DCE)asdefined in EIARS-232,the load isconnected to anotherDTE(e.g.,PCCOMinterface)with direct-connected Modemcable.
to setthe baudrate.Data bit:8Stop bit: 1Paritybit:(none,even,odd)
EVEN8 data bits have evenparityODD8 data bits have oddparityNONE8 data bits have noparity
Native machine address:(0 ~31,factorydefaultis0)
Start Bit 8 Data BitsParity=NoneStop Bit
5.3USB Interface
Connectthe load and the computerusing a cable with two USBinterfaces(each end).Allfunctionsofthe load can be programmed via USB.
Afterconnecting the load andcomputerbyUSB,you need toinstallIT-E122 driverorIT-E132 driver(see in ITECHCDorcontactITECHagent).The devicemanagerofPCwilldisplay‘ProlificUSB-to-SerialCOMPort’afterinstalling.
Notes:OnlyIT8514B+/8514C+/IT8516C+models have the USBcommucation interface.JustuseUSBline to connectin commucation.(Please don’tconnectDB9 interface atthe same time).Don’tneed to setthe menu.
Fansrunning speed isdetermined byradiatortemperature.Whentemperature reaches40℃,fans start to work and intelligently adjust itsspeed with temperature variation.
ITECHprovidesyou with optionalred and blacktestlines,which individualsalesand you can selectfortest.ForspecificationsofITECHtestlinesand maximumcurrentvalues,referto the table below.