I-tech CR200 User guide

Page 1
MNPG428-00 Edition 01/03/2022
CR200
Page 2

USER MANUAL

MNPG356 Rev.1 of 15/01/2022

CR200
IACER Srl 43 MNPG356-01
Page 3
INFORMATION ON THE USER MANUAL 47 MANUFACTURER 48 DECLARATION OF CONFORMITY 48 CLASSIFICATION 49 INTENDED PURPOSE AND SCOPE OF USE 49 TECHNICAL SPECIFICATIONS 49 DEVICE DESCRIPTION AND CONTROLS 52 LABELLING 55
Pack contents 56
NOTES 57 INTRODUCTION TO THE TECHNOLOGY 58
General information 58 Endothermic therapy 58
CONTRAINDICATIONS 60 WARNINGS 61 USE OF DEVICE 62
List of programmes 69
MAINTENANCE 71 TROUBLESHOOTING 73 DISPOSAL INFORMATION 75 WARRANTY 76
Support 77 Spare parts 77
SOUND ENERGY EMITTED 78 INTERFERENCE AND ELECTROMAGNETIC COMPATIBILITY TABLES 78
Contents
IACER Srl 45 MNPG356-01
Page 4
Technical information

Information on the user manual

This user manual is addressed to:
- the machine user;
- the owner;
- supervisors;
- those in charge of moving it;
- installers;
- users;
- maintenance staff.
This document provides information for the installation and correct use of the CR200 Tecar Therapy device. This manual acts as indispensable reference guide for the user: before installing and using the machines, it is essential to carefully read the contents of the manual and keep it always at hand for quick reference. Failure to comply, even partially, with the recommendations contained herein may lead, in addition to malfunctions, also to damage to the equipment, and invalidation of the warranty. However, by closely following the instructions and recommendations provided by the manufacturer, you will be assured of the best results as well as the availability of a fast and efficient technical support service if needed. The limits of this user manual are:
the user manual can never replace adequate user experience;
the user manual, for particularly demanding tasks, can be only considered
a reminder of the main operations. The user manual is to be considered part of the equipment and must be kept for future reference until the final dismantling of the equipment. The user manual must be available for consultation near the machine and stored correctly. This user manual reflects the state of the art at the time of marketing and cannot be considered inadequate only because it is subsequently updated on the basis of new experience. The manufacturer reserves the right to update its products and the related manuals without the obligation to update previous products and manuals.
The company is relieved of any liability in the main cases of
- improper use of the machine;
- use contrary to the specific national regulations;
IACER Srl 47 MNPG356-01
:
Page 5
- incorrect installation;
- power supply faults;
- serious failure to comply with scheduled maintenance;
- unauthorised modifications or operations;
- use of spare parts or materials not specific to the model;
- total or partial failure to follow the instructions;
- exceptional events.
If any further information is needed, contact the manufacturer directly.

Manufacturer

I.A.C.E.R.S.r.l.
Via S. Pertini, 24/a • 30030 Martellago (VE)
Tel.: +39 041 5401356 • Fax: +39 041 5402684
IACER S.r.l. is an Italian manufacturer of medical devices (CE certificate no. 0068/QCO-DM/168-2020 issued by the Notified Body no. 0068 MTIC InterCert S.r.l.).

Declaration of conformity

I.A.C.E.R.S.r.l
Via S.Pertini 24/A – 30030 Martellago (Ve), Italy
hereby declares under its own responsibility that the products
CR200
UMDNS code: 11244
Batch:
Serial No.: are designed and built-in compliance with Directive 93/42/EEC concerning medical devices (implemented in Italy with Legislative Decree46/97), as amended by Directive 2007/47/EC (Legislative Decree37/2010) and subsequent amendments/additions. The devices are classified class IIb, according to Annex IX, rule 9 of Directive 93/42/EEC (and subsequent amendments/additions) and are marked
IACER Srl 48 MNPG356-01
Page 6
MASSIMO MARCON
Martellago, 15/05/2020
Mains power supply
110-240 V, 50-60 Hz, ±10%
Maximum power consumption:
260 W
The conformity of the products in question with the Directive 93/42/EEC has been verified and certified by the Notified Body:
0068 - MTIC InterCert S.r.l.
Via Giacomo Leopardi 14, 20123 Milan, Italy
Certif. No.:
0068/QCO-DM/168-2020
according to the certification process provided for in Directive 93/42/EEC, Annex II (excluding point 4).
_________________ _______________________
Place, date Legal Representative

Classification

CR200 devices assume the following classifications:
class IIb device (Directive 93/42/EEC, Annex IX and subsequent
amendments);
class I, with applied parts BF (EN 60601-1);
active therapeutic device, non-invasive.

Intended purpose and scope of use

CR200 is an electromedical device that delivers Tecar therapy treatments, with the aid of handpieces/applicators that allow the administering of the treatment. The use of this device is reserved to operators who, by virtue of their professional training, ensure adequate use and total safety for the patient. In actual fact, the operator must be suitably qualified to be able to use these machines, and must have passed an appropriate training course, or must operate under the aegis of a doctor who is adequately qualified to use the machine under safe conditions for the person subjected to treatment. This machine can be used in a hospital environment, provided it is used by personnel qualified in this regard and in compliance with what is stated in the user manual.

Technical specifications

IACER Srl 49 MNPG356-01
Page 7
Maximum power supplied by the machine:
160 W
230 Vac
3.5 A-T - 5 x 20 mm
115 Vac
3.5 A-T - 5 x 20 mm
Backlit LCD touch-screen display for viewing, checking operating parameters and setting them for treatment purposes.
Programmable treatment time
Up to 60 minutes
Electrode holder handpiece
Emission frequency of the handpiece Modulation
1÷100 Hz (in "Custom" mode)
Resistive, made of AISI 316L
Capacitive, made of
paint
35 mm diameter
60 mm diameter
Adjustable power
0-100%
Double time-delay safety fuse on the mains (T):
Handpieces supplied
7 inch, 1024*600 resolution
of resistive type
Electrode holder handpiece of capacitive type
Type of electrodes used
Electrode diameter
IACER Srl 50 MNPG356-01
500 kHz
steel
aluminium and coated in rilsan
Page 8
35 mm diameter
60 mm Classification according to directive 93/42/EEC
II B
Insulation class/parts applied according to EN
Ingress protection rating according to EN 60601-1
2 Independent
Stored protocols
34
External dimensions (Width x Height x Depth):
27 x 12 x 30.5 cm
Machine body weight:
3 kg
ambient
(+10 ÷ +40) °C
relative
(10 ÷ 80) % non-condensing
ambient
(-25 ÷ +70) °C
relative
(0 ÷ 93) % non-condensing
Electrode working area
9.61 cm2
60601-1
Output channels
diameter
28.26 cm2
I / BF
IP20
(1 output channel for capacitive use, 1 output channel for resistive use) 1 channel for bipolar handpiece (to be used as an alternative to the other 2) 1 channel for return plate
Conditions of use
Storage/transport conditions
IACER Srl 51 MNPG356-01
Page 9
Rear panel
Full-colour touch screen display
Encoder knob
Plate connector
Handpiece connectors
Atmospheric pressure
The service life of the device is set at 3 years.

Device description and controls

(500 ÷ 1060) hPa
IACER Srl 52 MNPG356-01
Page 10
Cooling fan
ON/OFF switch
Power supply socket
Fuse Holder
Return Plate
Bipolar handpiece
Capacitive handpiece
Resistive handpiece
REAR PANEL
OUTPUT PANEL
ACCESSORIES
IACER Srl 53 MNPG356-01
Page 11
Description
Code
Supplied
Optional
Schuko plug power cable
CVAL REA
1
User manual
MNPG356
1
Capacitive handpiece
MAN-CAP REA
1
Kit 2 electrodes for capacitive treatments (diameters 35, 60 cm)
ELCAP35 REA ELCAP60 REA
Resistive handpiece
MAN-RES REA
1
Kit 2 electrodes for resistive treatments (diameters 35, 60 cm)
ELRES35 REA ELRES60 REA
CAVO+ELRIT/N
REA
CAVO+ELRIT/N
REA XL
Bottle of conductive cream
CREMA-TEC
1
Bipolar handpiece*
MAN-BIP REA
\
X
Handpieces with resistive and capacitive electrodes (see below)
Return Plate
The device is equipped with a mains power cable, and is compatible with the following kit of accessories supplied:
Steel return plate 12x17 cm
Steel return plate 17x24 cm
IACER Srl 54 MNPG356-01
1
1
1
\ X
Page 12
Symbol
Meaning
*not included in the medical CE certification.
The assembly of the accessories is simple and intuitive: a colour is associated with the corresponding connector + handpiece + electrode system (in the case of resistive and capacitive handpiece). he division of colours is shown on the previous page under "Output panel":
- White: return plate.
- Black: bipolar output.
- Grey: capacitive output.
- Green: resistive output.
In the event of installation problems or difficulties, contact IACER S.r.l.'s technical support service.

Labelling

Product certification issued by notified body No. 0068.
Manufacturer data.
Warning, see the documents accompanying the product
Date of manufacture (MM-YYYY).
Consult the user manual.
IACER Srl 55 MNPG356-01
Page 13
Symbol
Meaning
SN
Device serial number
RF output signal (label placed near the handpiece
sensitive device (label located near the
RIT
BIP
CAP
RES
WEEE directive for the disposal of electronic and electrical waste.
Type BF applied part according to IEC 60601-1.
Fuses used on the machine
connectors)
Electrostatic­handpiece connectors)
Return plate output
Bipolar handpiece output
Capacitive handpiece output
Resistive handpiece output

Pack contents

The pack contains:
User manual;
1 mains power cable;
1 resistive handpiece;
resistive electrodes kit (diameter 35mm, 60mm);
1 capacitive handpiece;
capacitive electrodes kit (diameter 35mm, 60mm);
1000ml bottle of conductive cream;
steel plate.
Check the contents of the pack. If any item is missing, contact your authorised I.A.C.E.R. S.r.l. dealer immediately.
IACER Srl 56 MNPG356-01
Page 14

Notes

PRELIMINARY NOTES
The installation of the device is quick and easy.
USE
Interactions with the device take place via the touchscreen. During therapy it is possible to increase or decrease the therapy using the encoder.
MAINTENANCE
For optimal use of the device and to ensure its maximum performance, it is recommended to carry out maintenance correctly according to the timings and ways recommended.
IACER Srl 57 MNPG356-01
Page 15
How to use the device

Introduction to the technology

General information

IACER Srl has recently developed a complete series of devices, accessories and equipment, designed and built according to the highest quality standards, adopting cutting-edge technologies in full compliance with the directives and standards in force. Particular attention has been paid to design, ease of operation, functionality and safety. The result is a compact unit with which it is possible to interface via a large display. The multiple possibilities of therapeutic applications, together with guaranteed patient and therapist safety (the unit complies with international standards), make the machine a high-quality device. These machines have been designed and manufactured so that their use, if it takes place under the conditions and for the uses intended, does not compromise the health and safety of patients, users and third parties, taking into account the benefit provided to the patient. These machines are not reserved for diagnosis, prevention, monitoring, compensation for injury or handicap, replacement or modification of the anatomy, control of conception, support of vital functions but allow you to treat particular conditions and actually reduce the disease/condition. No special intervention is required in case of failure of the medical device, only normal maintenance/repair operations.

Endothermic therapy

The endothermic therapy system is a recently introduced therapeutic method within physical therapy.It allows you to stimulate biological structures and natural reparative and anti-inflammatory processes from the inside through the application of energy, exploiting a form of interaction between electromagnetic energy and the tissue, which refers to the electrical concept of the capacitor.This device consists of 2 facing elements (called plates) separated by an insulating material, connected to an electric generator which creates a potential difference (p.d.) between the 2 plates.This causes electric charges to attract and repel each other concentrating near the 2 elements.In this way there is a positive increase in the charge density in one plate and negative in the other. The mobile electrode and the fixed return plate must be positioned in such a way as to create a circuit between them.
IACER Srl 58 MNPG356-01
Page 16
The current generator works in the field of long wave radio frequencies of 0,5 MHZ, with a variable power up to a maximum of 160W. In this way there is no external energy emission, but there is only an endogenous or internal development at a biological tissue level produced by the movement of ions and electrolytes, induced by the attraction and repulsion forces that are generated between the 2 plates of the capacitor.
APPLICATION TECHNIQUES
Capacitive/resistive tecar therapy (acronym of Capacitive and Resistive Energy Transfer) is a therapy that stimulates the body's natural repair processes, shortening the recovery time. Diathermy through the electromagnetic energy/tissue interaction produces an increase in temperature which occurs inside the tissues in a uniform and controlled way. This electromagnetic interaction gives rise to the appearance of an ionic flow with a micro-hyperemia which ultimately favours the release of endogenous "substances" (especially cortisol and endorphins) which are used to reduce pain, oedema and inflammation. CR200 therefore stimulates the increase in blood flow directly, thanks to the increase in temperature and indirectly through the demand for oxygen by the treated tissues; the increase in blood promotes the increase of normal immune defences and stimulates tissue regeneration. CR200 works in two modes:
- if you work in capacitive mode, there will be an increase in charge density
near the area below the mobile electrode and above all at the superficial soft tissue level.
- if you work in resistive mode, the concentration of charges and therefore
the biological effect occurs in the tissues with the highest resistance that are interposed between the mobile electrode and the return plate.
HOW IT WORKS
In order for the phenomenon of the increase in charge density to occur, the two capacitor plates must be connected to an electric generator which has the task of supplying the plates with charges. A real current is thus established, which in the accumulation phase goes from the generator to the capacitor. As the capacitor accumulates charges, the flow decreases until it is zero when the capacitor is fully charged. After this initial phase, if the polarity of the generator is inverted, there will be a current in the opposite direction which will charge the capacitor with polarity
IACER Srl 59 MNPG356-01
Page 17
opposite to the previous one. If the generator cyclically reverses polarity there will be a flow in both directions i.e. an alternating current. The transfer by capacitive contact is carried out through a capacitive electrode treated with an insulating coating which mobilises the ionic charges in the subcutaneous tissues. Rresistive transfer takes place by means of a non-insulated resistive electrode which mobilises the charges making them concentrate in the areas of greater depth and resistance. The capacitive mode therefore acts specifically on soft tissues (superficial muscles, vascular and lymphatic circulatory system, adipose tissue), whereas the resistive mode acts on the tissues with greater resistance (bone, cartilage, tendons, deep muscles). The mobile electrode (capacitive or resistive) and the plate are positioned so that the circuit created includes the area to be treated. For a better transfer of energy to the tissues on the part to be treated, a cream is used that prevents the interposition of air between the electrode and the body surface and which favours a homogeneous interaction of the body with the applied parts of the device. CR200 is a device manufactured according to the MED 93/42/EEC directive (and subsequent amendments) relating to medical devices.

Contraindications

Tecar therapy treatments cannot be administered in cases of people with:
bleeding;
tumors;
phlebitis, thrombi and arteriopathies;
decreased sensitivity in the area to be treated;
metal prostheses;
pacemaker;
hearing aids;
insulin pumps;
intrauterine devices;
fever or infections;
ongoing pregnancies;
epilepsy;
or on children under 14 years.
IACER Srl 60 MNPG356-01
Page 18

Warnings

- The customer is responsible for damage resulting from inadequate
packaging. case of return to the company.
- Do not use the device in places where it could get wet.
- Carefully check that the connections are in accordance with the instructions
provided before operating the machine. To avoid any risk of electric shock, the device must only be connected to
-
mains with protective grounding systems.
- Do not use accessories other than the original ones supplied: these could
damage the machine and invalidate the warranty. In the event of installation problems or difficulties, contact IACER S.r.l.'s technical support service.
- If you use an extension lead that the machine shares with other devices,
check that the total current consumption of the connected devices does not exceed the maximum current allowed for that type of cable and that it is not in any case greater than 15A.
- It is not possible to define a suggested number of sessions to evaluate the
effectiveness of the treatment, since they are linked to the power delivered to the patient undergoing treatment, as well as to the minutes of therapy.The doctor/healthcare professional must decide the number of therapeutic sessions to which the patient is subjected according to the specific needs, in order to guarantee the patient receives effective treatment over time that is carried out under completely safe conditions. Check the condition of the power supply cable and the connection cable to
-
the handpiece/applicator often: these must not be damaged or worn.
- It is a class A device in terms of emissions.The device can be used in a hospital
environment provided that it is duly taken into account that the device could cause disturbance to electronic devices placed in the immediate vicinity.
- No modification of this device is permitted.
- The use of accessories, transducers and cables, other than those specified or
supplied by the manufacturer, could lead to higher electromagnetic emissions or a decrease in the electromagnetic immunity level of the device, with consequent incorrect operation.
- The device is not intended for outdoor use.
- The device is not intended for use in the domestic settings.
Keep the original packaging of the machine: it must be reused in
IACER Srl 61 MNPG356-01
Page 19
USE
- It is possible to provide a digital copy of the device's user manual upon
request.
- The device should not be used in close proximity to other equipment and, if
it is necessary to use it near other equipment, the electro-medical device must be monitored to check the normal operation in the configuration in which it is used.
- If the electro-medical device, interacting with another device, causes or
receives detected interferences, the user is encouraged to limit such interference by adopting one or more of the following measures:
reorient or relocate the receiving device;
increase the distance between the devices;
connect the device to a socket of a circuit different from the device(s)
that cause the interference;
contact the manufacturer or your local technician for assistance.
- Portable and mobile radio communication equipment may affect the
operation of the device.

Use of device

- The perfect functionality of the device is guaranteed if you comply with the
installation and use standards indicated, and only use original accessories and spare parts.
- In the event of installation problems or difficulties, contact IACER S.r.l.'s
technical support service.
- The correct transport position of the machine requires that the device is
handled only by holding the sides of the machine body with both hands.
: avoid handling the device by the handpiece/electrode holder tabs.
N.B.
- Before connecting the cable to the mains plug, check that the device has not
been damaged during transport and make sure that the characteristics of the electricity supply on the available power socket meet the nameplate data shown on the back of the machine.
- The device must only be connected to compliant systems.
- If extension leads are used, check the presence and integrity of the protective
earth conductor.
- Connect the equipment directly to the wall socket, possibly without using
extension leads. Failure to comply with this warning could result in hazardous electric shocks to people and alter the operation of the machine.
IACER Srl 62 MNPG356-01
Page 20
- The manufacturer is only responsible for the fundamental safety, reliability
and performance of the device if:
the electrical system of the premises complies with the appropriate
regulations;
the device is used in accordance with the instructions for use.
GENERAL INFORMATION
- In order to guarantee the operation of the machine in conditions of absolute
safety for the patient, it is advisable to subject the machine to a cycle of routine checks (at least 2 years) to be carried out by the manufacturer.
- It is recommended to leave the machine on standby for 5 minutes after each
cycle of therapy.
- It is absolutely forbidden to use the device in the presence of flammable
anaesthetic mixtures and oxygen-rich environments. In case of non­compliance with the indication provided, IACER Srl shall not be held responsible for any accidents.
- It is absolutely forbidden to cover the rear vent: doing so may not allow the
machine to work safely. In case of non-compliance with the indication provided, IACER Srl shall not be held responsible for any accidents.
- It is important to draw the operator's attention to the need to check the
correctness of the electrical installation of the equipment before operating the mains switch.
- Before beginning treatment, the operator must make sure that the patient
removes any metal objects worn, so as to avoid triggering hazardous radio frequency coupling phenomena.
- Tecar therapy treatments must be administered, under the strict control of the
operator, to "conscious" patients, capable of interacting with the operator with regard to the electrical forces transmitted by the machine.
- It is advisable to suspend the therapeutic treatment if issues should appear
during its administering.
UNPACKING
The device is packed and prepared for shipping with its box, complete with filling material, designed for safe storage and transport. To unpack the machine, place the box on a flat, solid surface and remove the upper polystyrene part. Carefully remove the appliance.
IACER Srl 63 MNPG356-01
Page 21
INSTALLATION
The installation of the CR200 device is quick and easy. The recommended environmental characteristics for installation are the following:
- ambient temperature: +10° to +40°C;
- Relative humidity 10 to 80% non-condensing;
- avoid direct exposure to sunlight, chemicals and vibrations.
CONNECTIONS
On the rear of the machine there is the built-in mains power supply module, which includes the three-pole connector for the power cable, the removable fuse holder with two fuses (see technical specifications) and the double-pole main switch. Insert the three-pole female plug of the power cable into the built-in module, checking that it is perfectly inserted inside the connector. If extension leads are used, check the presence and integrity of the protective earth conductor. Failure to comply with this warning could result in hazardous electric shocks to people and alter the operation of the machine. To connect the applied parts, proceed as described: connect the steel plate to the corresponding connector. Identify the desired mobile handpiece (resistive or capacitive) for therapy and the associated electrode.The connector + handpiece + electrode system can be identified by a dedicated colour.The description of the corresponding colours and outputs is given in the "Output panel" section, in the chapter relating to the description of the device. If you need to use the bipolar handpiece, this is the only one that requires you to disconnect the steel plate. Please note that this handpiece is only intended to be used in "Custom" mode. This applicator does not fall under the medical CE certification of the device, as it is suitable only for aesthetic purposes. After carrying out the checks for correct installation and assembly, turn on the main power switch, checking that the display comes on correctly. N.B
: only connect the applied parts necessary for the treatment when using the
device.
OPERATION
The user-device interface is achieved by a large, clear touchscreen display: it displays all the operational messages relevant for the operator, the operating
IACER Srl 64 MNPG356-01
Page 22
status of the machine during normal therapeutic activity, any error, visual and acoustic. The following sections describe how the device menu is divided.
Therapy selection
Within the "Therapy Selection" menu, you can choose from 2 types of pre-set programmes (Rehabilitation, Sport) or decide to manually set the therapy specifications (Custom).
Figure 7
PRE-SET PROGRAMMES
When you choose one of the first 2 submenus shown in Figure 1, a screen will open displaying the list of available programmes. The programme list screen is shown below (Figure 2):
IACER Srl 65 MNPG356-01
Page 23
Figure 8
Using the touchscreen, you can scroll through the available pre-set programmes. For further information on the therapeutic protocols present, please refer to the following chapter "List of programmes". Once the desired programme has been selected, you can decide to start the therapy or to return to the list of available programmes, as shown in the following figure.
Figure 9
IACER Srl 66 MNPG356-01
Page 24
Once the programme has started, it is possible to pause the treatment at any time.In the case of programmes that involve the use of both electrodes (resistive and capacitive), the specific duration of the individual treatment phases are given, in the predetermined order.
CUSTOM
When you press the button for the "Custom" programme, the following screen appears.
Figure 10
Before starting the therapy, it is possible to choose the type of treatment (capacitive/resistive/bipolar), as well as the duration, the carrier frequency and the modulation.
WARNING
The selected therapy parameters can be stored by pressing the "Save" button. The menu relating to free memory slots will then open, as shown below in Figure 5.
IACER Srl 67 MNPG356-01
: when setting the duration of the therapy, avoid 0:00.
Page 25
Figure 11
Once the desired memory slot has been selected, the device will ask you to enter a name in order to save the Custom programme.
Figure 12
IACER Srl 68 MNPG356-01
Page 26
Time (min) and mode
1
Phantom limb
500
15 C + 15 R
2
Knee arthrosis
500
15 R + 10 C
3
Arthrosynovitis
500
10 R + 10 C
4
Bursitis
500
10 R + 20 C
5
Neck pain
500
15 R + 10 C
6
Whiplash syndrome
500
10 R + 20 C
7
Chondropathy
500
10 R + 10 C
8
Muscle contracture
500
5 R + 15 C
9
Contusion
500
20 R + 10 C
10
Muscular sprain
500
20 R + 10 C
11
Sprain
500
12 R + 12 C
12
Muscle distraction
500
10 R + 20 C
13
Joint pain
500
10 R + 10 C
14
Oedema
500
20 R + 10 C
15
Epicondylitis
500
10 R + 10 C
Each programme saved in free memory slots can be called up via the "Load" button in the "Custom programme" menu (see Figure 4).
After starting the programme, you can adjust the power (initially set to 0) using the encoder knob. To increase the power output, turn the knob clockwise.To decrease the power, turn the knob in the opposite direction. At the end of the therapy, press the "BACK" button until you reach the main menu, then turn off the device using the switch on the rear panel. Finally unplug the power plug from the socket.
The proposed preset programmes are the result of operational experience gained over many years supporting expert professional users. The following section ("List of programmes") contains the list of available protocols, by category.

List of programmes

REHABILITATION
No. Programme CarrierFrequency
(kHz)
(Capacitive/Resistive)
IACER Srl 69 MNPG356-01
Page 27
Time (min) and mode
16
Epitrochleitis
500
10 R + 10 C
Anterior cruciate ligament (ACL)
18
Lymphoedema
500
15 C
19
Low back pain
450
10 C + 10 R
synovitis)
21
Muscle tear
500
10 R + 20 C
22
Acute tendinitis
500
10 R + 10 C
23
Chronic tendinitis
500
10 R + 10 C
24
Rotator cuff tendinopathy
500
25 C + 15 R
(kHz)
1
Muscle contracture
500
5 R + 15 C
2
Contusion
500
20 R + 10 C
3
Muscle recovery
500
10 C + 10 R + 5 C
4
Sprain
500
12 R + 12 C
5
Hematoma
500
20 R + 10 C
6
Tendon injury
500
10 R + 10 C
7
Superficial massage*
500
10 R + 10 C
8
Deep massage*
500
10 R + 10 C
10
Muscle tear
500
10 R + 20 C
No. Programme CarrierFrequency
17
20 Meniscopathy (meniscal
(kHz)
500 15 C + 15 R
500 10 R + 10 C
SPORT
No. Programme CarrierFreq.
(Capacitive/Resistive)
Time (min) and mode (Capacitive/Resistive)
9 Muscle sprain 500 20 R + 10 C
*Treatment not covered by medical CE
IACER Srl 70 MNPG356-01
Page 28
Device care

Maintenance

CR200 Tecar therapy devices do not require particular maintenance operations, except for routine maintenance and cleaning of the applicator handpieces, with the aim of ensuring the best operating conditions, to guarantee the effectiveness of the treatment and patient safety. External cleaning of the equipment must only be done with a soft cloth moistened with hot water, or using non-flammable cleaning liquids. The front control panel can also be cleaned in the same way. The handpieces/applicators, in particular the treatment head, must be periodically cleaned with Carefully put away the handpieces/applicators at the end of each treatment. Contact IACER Srl authorised centres for information on original accessories and spare parts. Do not spray, nor pour liquids on the external container of the devices, nor on the ventilation slots. Do not immerse the machine in water. After cleaning the outside of the box, make sure all parts are dry before putting the device back into operation. Under no circumstances must the device be disassembled for cleaning or checking purposes: there is no need to clean the machines internally, and in any case this operation must be done exclusively by IACER Srl specialised and authorised technical personnel. More specifically:
- handle the handpiece-applicator with care: rough handling can negatively
-
- Under no circumstances must the device be disassembled for cleaning or
- Do not use thinners, detergents, acid solutions, aggressive solutions or
water and denatured alcohol.
influence its performance and characteristics. Under no circumstances are unauthorised technical personnel allowed to open and/or disassemble the handpiece/applicator: such tampering, in addition to damaging the characteristics of the handpiece, immediately invalidates the warranty.
checking purposes: there is no need to clean the machine internally, and in any case this operation must be done exclusively by IACER Srl specialised and authorised technical personnel.
flammable liquids to clean the machine exterior and accessories. The use of these substances, coupled with improper use of the accessories,
IACER Srl 71 MNPG356-01
Page 29
besides irreparably damaging the appliance, also invalidates the warranty.
- For optimal use of the device and to ensure its maximum performance, it
is recommended to carry out maintenance correctly according to the timings and ways recommended.
- To correctly replace the fuses on the machine, follow the instructions
below:
6. use a screwdriver to open the fuse box, taking care to insert the
screwdriver into the slot created on the fuse box and levering it outwards
7. remove the fuse-holder structure by sliding it along the guide
8. remove the inserted fuses and replace them with new ones
9. insert the fuse-holder structure in the fuse box, sliding it towards
the guide
10. close the plastic door of the fuse box
It is recommended to carry out routine maintenance every year,
checking:
the intensity of any leakage currents;
the continuity, and therefore the integrity, of the earth conductor;
the correctness of the insulation resistance value
in order to guarantee the electrical safety of the device and to make sure that it operates under the guaranteed safe conditions. For such operations, we recommend contacting a qualified technical service centre or alternatively IACER Srl or one of its authorised centres.
WARNING!
For safety reasons, before carrying out any maintenance and cleaning
-
operations on the device, it is NECESSARY to turn off the device using the switch on the rear panel and disconnect the power cable from the power socket.
- It is recommended to carefully clean the machine and accessories supplied
before using it in contact with the patient.
- It is useful to draw the operator's attention to the need for routine
maintenance of the handpieces/applicators, to be carried out by the manufacturer.
- Cleaning and disinfection must always be done before carrying out the
therapeutic treatment on the patient.
IACER Srl 72 MNPG356-01
Page 30
PROBLEM
POSSIBLE CAUSE
SOLUTION
Mains plug not inserted correctly
Check the operation of the
Mains cable not correctly
device.
Insert the plug and cable
connector of the appliance.
Mains cable worn and broken.
Replace the mains cable.
The switch is off.
Operate the mains switch.
The LCD display
Replace the missing,
- Do not spray or pour liquids on the external container of the device, on the
ventilation slots, on the LCD display or on the fan grille.IACER Srl will not be held responsible for any damage occurring if the machine has been used without carrying the maintenance operations described above.
- Check the condition of the power supply cable and the connection cables of
the applicators/accessories applied to the patient often: these must not be damaged or worn. It is advisable to have the fuses replaced by personnel with adequate
-
technical expertise and training, in order to carry out the operation safely.
- Do not open the device: there are high electric voltages inside which can be
dangerous.
- Only technical personnel authorised by the manufacturer can access the
internal parts of the device. For repairs and further information it is necessary to contact IACER Srl or its authorised service centres.

Troubleshooting

CR200 Tecar therapy machines have been designed and built adopting advanced technological solutions, quality components, for continuous use that is always efficient and reliable. If, however, there is a problem with operation, it is recommended to consult the following guide before contacting an authorised service centre.
The LCD display
on the front
panel does not
turn on: the
device does not
work.
on the front
IACER Srl 73 MNPG356-01
inserted in the connector of the
Defective or blown fuse or fuses.
in the power socket.
power socket.
correctly into the
defective or blown fuse (s).
Page 31
PROBLEM
POSSIBLE CAUSE
SOLUTION
panel does not
The LCD display
Some controls
Defective buttons or keys.
Carefully check the output
Handpiece-applicator cable
Replace the defective handpiece-applicator which shows clear signs of wear in
the treatment head and on
the cable.
Fault in the electronic circuit of
Contact an IACER Srl
The device works
Not perfectly efficient connection
Perform the maintenance
cable and the connector of
turn on: the
device does not
work.
on the front
panel does not
turn on.
on the front
control panel do
not work properly.
The device does
not activate
during delivery.
Electronic control circuit failure.
Defective components on the
electronic control board.
Electronic control circuit failure.
Bad connections in the output
circuit applied to the patient
broken or incorrectly connected
Output cables are worn and/or
have loose contact.
Contact an IACER Srl
service centre.
Contact an IACER Srl
service centre.
Contact an IACER Srl
service centre.
connections are correct
and not damaged.
as normal, but
there is a
noticeable drop
in the
effectiveness of
the treatment.
IACER Srl 74 MNPG356-01
the power generator.
of the handpiece-applicator
output circuit.
service centre.
operations described.
Install and position the
device as described.
Check the condition of the
the handpiece-applicator.
Page 32
PROBLEM
POSSIBLE CAUSE
SOLUTION
Mechanical damage (due to falls
Check the perfect
Electronic circuit of the generator
Possible failure of the appliance's
or hard knocks) on the
handpiece-applicator, in
particular on the radiating head.
not perfectly calibrated.
power generator circuit.
When the conditions listed below occur, disconnect the device from the electrical system and contact IACER Srl's technical support service:
the cable or the built-in rear power supply module is worn or damaged;
• liquid has entered the device;
the device has been exposed to rain.
-
Only technical personnel authorised by the manufacturer can access the
internal parts of the device.
- For repairs and further information it is necessary to contact I.A.C.E.R.Srl or
its authorised service centres.
WARNING!
DO NOT OPEN the unit, there are HIGH ELECTRIC VOLTAGES inside which
can be DANGEROUS.
adherence of the parts
applied on the surface
involved in the treatment.
Contact an IACER Srl
service centre.

Disposal Information

CR200 Tecar therapy devices, in line with operating and safety requirements, have been designed and built to have a minimal negative impact on the environment. The criteria followed are those of minimising the amount of waste, toxic materials, noise, unwanted radiation and energy consumption. Careful research into optimising machine performance guarantees a significant reduction in consumption, in accordance with the subject of energy saving.
This symbol indicates that this product should not be disposed with other household waste.
IACER Srl 75 MNPG356-01
Page 33
Correct disposal of obsolete equipment, accessories and especially batteries, helps to prevent possible negative consequences on human health and the environment. The user must dispose of the equipment to be scrapped by taking it to the collection centre indicated for the subsequent recycling of electrical and electronic equipment. For more detailed information on the disposal of obsolete equipment, contact your local council, waste disposal service or shop where you purchased the product.

Warranty

I.A.C.E.R.Srl guarantees the quality of its devices, when used in accordance with the instructions provided in this manual, for a period of 12 months from the date of purchase. During the warranty period, defective products will be repaired or replaced at the company's discretion. Replacement of the device itself is not part of the warranty. The warranty does not cover malfunctions or damage resulting from
- inadequate placement and installation;
- incorrect use or that not in compliance with the instructions given in this
manual;
- improper or inadequate maintenance by the user;
- operation that does not comply with the environmental specifications
indicated for the product;
- unauthorised opening of external enclosures;
- tampering and/or unauthorised modifications;
- use of non-original accessories.
The warranty is provided ex I.A.C.E.R. srl's registeredoffice. If a return is necessary, follow the packing instructions below and attach a copy of the purchase receipt. It is advisable to insure the shipment. Before shipping the machine due to a suspected fault, it is recommended to carefully consult the MAINTENANCE and TROUBLESHOOTING chapters: issues are largely attributable to poor maintenance or minor technical problems that the user can easily resolve. A simple phone call to I.A.C.E.R Srl's Technical Service can be of great help in solving any problem. Instructions for packing and returning the device:
:
IACER Srl 76 MNPG356-01
Page 34
6. disconnect the power and connection cables with handpieces, applicator
devices, etc .; thoroughly clean and disinfect all accessories and parts of the machine
7.
that have been in contact with the patient. For obvious hygiene reasons, in order to guarantee adequate health protection of technical personnel (workplace safety law, Legislative Decree81/2008), receiving them will not be checked;
8. disassemble the accessories and any mechanical supports;
9. reuse the original box and materials for packaging;
10. attach to the shipment the Support Request Form including the reasons
for the request for inspection/service, the type of failure or malfunction: such information will help technicians and significantly reduce repair times.
appliances deemed hygienically unsafe by the staff

Support

The manufacturer is the only point of contact for technical support regarding the device.For all technical support matters, please contact:
I.A.C.E.R.S.r.l.
Via S. Pertini, 24/a • 30030 Martellago (VE)
Tel.: +39 041.5401356 • Fax: +39 041.5402684
Technical documentation concerning repairable parts may be provided, but only with prior company authorisation and only after giving proper training to the maintenance personnel.

Spare parts

Original spare parts for this device can be ordered at any time from the manufacturer. To order them contact:
I.A.C.E.R.S.r.l.
Via S. Pertini, 24/a • 30030 Martellago (VE)
Tel.: +39 041.5401356 • Fax: +39 041.5402684
In order to preserve the warranty, guarantee operation and safety of the product, it is recommended to only use original spare parts supplied by the manufacturer (also see the Warnings paragraph).
IACER Srl 77 MNPG356-01
Page 35

Sound energy emitted

The sound energy emitted by the CR200 device during normal use (except for the audible alarm signals) does not exceed the following levels:
-
80dBA for 24 h cumulative exposure over 24 h; to this value is added a deviation of 3dBA when the total duration of exposure is halved over 24 h (for example 83dBA for 12 h over 24 h);
-
a sound pressure level of 140dBC (peak) for pulse energy or noise impact.

Interference and electromagnetic compatibility tables

The CR200 Tecar therapy device has been designed and built in compliance with EN 60601-1-2:2015 and the current ELECTROMAGNETIC COMPATIBILITY DIRECTIVE 2014/30/EC, with the aim of providing reasonable protection from harmful interference in residential, civil and hospital settings. All the necessary measurements and checks were carried out at I.A.C.E.R. srl's in­house testing, measurement and inspection laband specialised external centres. Upon request, customers may view the reports on EMC measurements at the company. Based on their operating principle, the CR200 tecar therapy devices do not generate significant radio frequency energy and have an adequate level of immunity to radiating electromagnetic fields. Under these conditions, harmful interference cannot occur to radioelectric communications and to the operation of electro-medical devices used for monitoring, diagnosis, therapy and surgery, to the operation of electronic office devices such as computers, printers, copiers, faxes, etc. and to any electrical or electronic appliance used in such environments, provided that they comply with the ELECTROMAGNETIC COMPATIBILITY directive. In any case, to prevent any problem with interference, it is recommended to operate any therapy device at an appropriate distant from critical equipment for monitoring patients' vital functions and to use caution in therapeutic applications on patients with pacemakers.
CR200.All rights reserved. CR200 and the logo are the exclusive property of I.A.C.E.R. Srl and are registered trademarks.
IACER Srl 78 MNPG356-01
Page 36
is recommended that
Campo magnetico a
30 A/m
30 A/m
I campi magnetici a
Nota_e UT è la tensione di rete in c.a. prima dell’applicazione del livello di prova
frequenza di rete (50/60 Hz)
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
<5% UT
(>95% buco in_dip
)
in U
T
per_for 5 sec
<5% UT
(>95% buco in_dip
)
in U
T
per_for 5 sec
the CR200 be powered from an uninterruptible power supply or a battery
frequenza di rete dovrebbero avere livelli caratteristici di una località tipica in ambiente commerciale o ospedaliero.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment
UT is the a.c. mains voltage prior to application of the test level
IACER Srl 202 MNPG428-00
Page 37
GUIDA E DICHIARAZIONE DEL COSTRUTTORE – IMMUNITÀ ELETTROMAGNETICA –
La CR200 è prevista per funzionare nell’ambiente elettromagnetico sotto specificato. Il
Prova di immunità
Livello di prova
Livello di
Ambiente elettromagnetico –
Gli apparecchi di
PER GLI APPARECCHI ED I SISTEMI CHE NON SONO DI SOSTENTAMENTO DI FUNZIONI VITALI
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY – FOR EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING
cliente o l’utilizzatore della CR200 deve garantire che esso venga usato in tale ambiente.
The CR200 is intended for use in the electromagnetic environment specified below. The customer or the user of the CR200 should assure that it is used in such an environment.
Immunity Test
IEC 60601
IEC 60601 test
level
conformità
Compliance
level
Guida
Electromagnetic environment -
Guidance
comunicazione a RF portatili e mobili non dovrebbero essere usati più vicino a nessuna parte della CR200 compresi i cavi, della distanza di separazione raccomandata calcolata con l’equazione applicabile alla frequenza del trasmettitore
Portable and mobile RF communications equipment should be used no closet to any part of the CR200, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
IACER Srl 203 MNPG428-00
Page 38
Distanza di separazione
3 Veff_Vrms
P
V
d
 
 
=
1
5,3
P
V
d
 
 
=
1
12
P
E
d
 
 
=
1
12
raccomandata
Recommended separation distance
RF condotta
Conducted RF
IEC 61000-4-6
da 150 kHz a 80 MHz
150 kHz to 80 MHz
6 Veff_Vrms
da 150 kHz a 80 MHz per banda ISM
150 kHz to 80 MHz for ISM band
3 Veff_Vrms
([V
] V)
1
6 Veff_Vrms
([V
] V)
1
for ISM band
RF irradiata
Radiated RF
IEC 61000-4-3
IACER Srl 204 MNPG428-00
3 V/m
da 80 MHz a 2,7 GHz
3V/m
[E
] V/m
1
da 80 MHz a
800 MHz
80 MHz to 800 MHz
Page 39
80MHz to 2,7
P
E
d
 
 
=
1
7
RF irradiate per
P
E
d
 
 
=
1
6
ove P è la potenza massima
dispositivi di comunicazione radio
Radiated RF to RF wireless communication equipment
IEC 61000-4-3
GHz
3 V/m
da 80 MHz a 6 GHz
80 MHz to 6 GHz
3V/m
[E
] V/m
1
da 800 MHz a
2,7 GHz
800 MHz to 2,7 GHz
da 80 MHz a
6 GHz
80 MHz to 6 GHz
IACER Srl 205 MNPG428-00
nominale d’uscita del trasmettitore in Watt (W) secondo il costruttore del trasmettitore e d è la distanza di separazione raccomandata in metri (m).
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Page 40
Note_s:
Le intensità di campo dei trasmettitori a RF fissi, come determinato da un’indagine elettromagnetica potrebbe essere minore del livello di conformità in ciascun intervallo di frequenza
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey should be less than the compliance level in each frequency range
Si può verificare interferenza in prossimità di apparecchi contrassegnati dal seguente simbolo:
a
del sito
b
.
b
a
,
Interference may occur in the vicinity of equipment marked with the following symbol:
IACER Srl 206 MNPG428-00
Page 41
(1) A 80 MHz e 800 MHz; si applica l’intervallo di frequenza più alto.
a Le intensità di campo per trasmettitori fissi come le stazioni base per
At 80 MHz and 800 MHz, the higher frequency range applies.
Queste linee guida potrebbero non applicarsi in tutte le situazioni. La propagazione elettromagnetica è influenzata dall’assorbimento e dalla riflessione di strutture, oggetti e persone.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
radiotelefoni (cellulari e cordless) e radiomobili terrestri, apparecchi di radioamatori, trasmettitori radio in AM e FM e trasmettitori TV non possono essere previste teoricamente e con precisione. Per valutare un ambiente elettromagnetico causato da trasmettitori RF fissi, si dovrebbe considerare un’indagine elettromagnetica del sito. Se l’intensità di campo misurata nel luogo in cui si usa un CR200, supera il livello di conformità applicabile di cui sopra, si dovrebbe porre sotto osservazione il funzionamento normale della CR200. Se si notano prestazioni anormali, possono essere necessarie misure aggiuntive come un diverso orientamento o posizione della CR200.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CR200 is used exceeds the applicable RF compliance level above, the CR200 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the CR200.
b L’intensità di campo nell’intervallo di frequenza da 150 kHz a 80 MHz dovrebbe essere minore di [V1] V/m
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
IACER Srl 207 MNPG428-00
Page 42
DISTANZE DI SEPARAZIONE RACCOMANDATE TRA APPARECCHI DI
La CR200 è prevista per funzionare in un ambiente elettromagnetico in cui sono sotto
Potenza di uscita
Distanza di separazione alla frequenza del trasmettitore (m)
0,01
0,12
0,2
0,12
0,23 – 0,1
0,38
0,63
0,38
0,73 – 0,2 – – – –
0,9 1 1,20
2,0
1,20
2,30 – 1,8 – – – –
2,7
RADIOCOMUNICAZIONE PORTATILI E MOBILI PER LA CR200 CHE NON SONO DI SOSTENTAMENTO DELLE FUNZIONI VITALI
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT THE CR200 THAT ARE NOT LIFE-SUPPORTING
controllo i disturbi irradiati RF. Il cliente o l’operatore della CR200 possono contribuire a prevenire interferenze elettromagnetiche assicurando una distanza minima fra gli apparecchi di comunicazione mobili e portatili a RF (trasmettitori) e la CR200 come sotto raccomandato, in relazione alla potenza di uscita massima degli apparecchi di radiocomunicazione.
The CR200 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the CR200 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CR200 as recommended below, according to the maximum output power of the communications equipment.
massima del trasmettitore specificata
Rated maximum output power of transmitter
W
IACER Srl 208 MNPG428-00
Separation distance according to frequency of transmitter (m)
150 kHz to
80 MHz
150 kHz to
80 MHz (ISM band)
Da 80 MHz to 800 MHz
800 MHz to 2,7 GHz
Da 80 MHz to 6 GHz
(to RF wireless radio communication equipment)
Page 43
2 – – – –
2,8
10
3,80
6,3
3,80
7,30 – 100
12,00
20
12,00
23,00
Per i trasmettitori specificati per una potenza massima di uscita non riportata sopra, la
Nota_e
distanza di separazione raccomandata d in metri (m) può essere calcolata usando l’equazione applicabile alla frequenza del trasmettitore, ove P è la potenza massima d’uscita del trasmettitore in Watt (W) secondo il costruttore del trasmettitore
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
(1) A 80 MHz e 800 MHz, si applica l’intevallo della frequenza più alto.
At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
(2) Queste linee guida potrebbero non applicarsi in tutte le situazioni. La propagazione elettromagnetica è influenzata dall’assorbimento e dalla riflessione di strutture, oggetti e persone.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
IACER Srl 209 MNPG428-00
Page 44
Loading...