INFORMATION ON THE USER MANUAL47
MANUFACTURER48
DECLARATION OF CONFORMITY48
CLASSIFICATION49
INTENDED PURPOSE AND SCOPE OF USE49
TECHNICAL SPECIFICATIONS49
DEVICE DESCRIPTION AND CONTROLS52
LABELLING55
SOUND ENERGY EMITTED78
INTERFERENCE AND ELECTROMAGNETIC COMPATIBILITY TABLES78
Contents
IACER Srl 45 MNPG356-01
Page 4
Technical information
Information on the user manual
This user manual is addressed to:
- the machine user;
- the owner;
- supervisors;
- those in charge of moving it;
- installers;
- users;
- maintenance staff.
This document provides information for the installation and correct use of the
CR200 Tecar Therapy device.
This manual acts as indispensable reference guide for the user: before installing
and using the machines, it is essential to carefully read the contents of the manual
and keep it always at hand for quick reference.
Failure to comply, even partially, with the recommendations contained herein may
lead, in addition to malfunctions, also to damage to the equipment, and
invalidation of the warranty.
However, by closely following the instructions and recommendations provided by
the manufacturer, you will be assured of the best results as well as the availability
of a fast and efficient technical support service if needed.
The limits of this user manual are:
the user manual can never replace adequate user experience;
−
− the user manual, for particularly demanding tasks, can be only considered
a reminder of the main operations.
The user manual is to be considered part of the equipment and must be kept for
future reference until the final dismantling of the equipment. The user manual
must be available for consultation near the machine and stored correctly.
This user manual reflects the state of the art at the time of marketing and cannot
be considered inadequate only because it is subsequently updated on the basis of
new experience. The manufacturer reserves the right to update its products and
the related manuals without the obligation to update previous products and
manuals.
The company is relieved of any liability in the main cases of
- improper use of the machine;
- use contrary to the specific national regulations;
IACER Srl 47 MNPG356-01
:
Page 5
- incorrect installation;
- power supply faults;
- serious failure to comply with scheduled maintenance;
- unauthorised modifications or operations;
- use of spare parts or materials not specific to the model;
- total or partial failure to follow the instructions;
- exceptional events.
If any further information is needed, contact the manufacturer directly.
Manufacturer
I.A.C.E.R.S.r.l.
Via S. Pertini, 24/a • 30030 Martellago (VE)
Tel.: +39 041 5401356 • Fax: +39 041 5402684
IACER S.r.l. is an Italian manufacturer of medical devices (CE certificate no.
0068/QCO-DM/168-2020 issued by the Notified Body no. 0068 MTIC InterCert
S.r.l.).
Declaration of conformity
I.A.C.E.R.S.r.l
Via S.Pertini 24/A – 30030 Martellago (Ve), Italy
hereby declares under its own responsibility that the products
CR200
UMDNS code: 11244
Batch:
Serial No.:
are designed and built-in compliance with Directive 93/42/EEC concerning
medical devices (implemented in Italy with Legislative Decree46/97), as
amended by Directive 2007/47/EC (Legislative Decree37/2010) and
subsequent amendments/additions.
The devices are classified class IIb, according to Annex IX, rule 9 of Directive
93/42/EEC (and subsequent amendments/additions) and are marked
IACER Srl 48 MNPG356-01
Page 6
MASSIMO MARCON
Martellago, 15/05/2020
Mains power supply
110-240 V, 50-60 Hz, ±10%
Maximum power consumption:
260 W
The conformity of the products in question with the Directive 93/42/EEC
has been verified and certified by the Notified Body:
0068 - MTIC InterCert S.r.l.
Via Giacomo Leopardi 14, 20123 Milan, Italy
Certif. No.:
0068/QCO-DM/168-2020
according to the certification process provided for in Directive 93/42/EEC,
Annex II (excluding point 4).
_________________ _______________________
Place, date Legal Representative
Classification
CR200 devices assume the following classifications:
•class IIb device (Directive 93/42/EEC, Annex IX and subsequent
amendments);
• class I, with applied parts BF (EN 60601-1);
• active therapeutic device, non-invasive.
Intended purpose and scope of use
CR200 is an electromedical device that delivers Tecar therapy treatments, with the
aid of handpieces/applicators that allow the administering of the treatment.
The use of this device is reserved to operators who, by virtue of their professional
training, ensure adequate use and total safety for the patient.
In actual fact, the operator must be suitably qualified to be able to use these
machines, and must have passed an appropriate training course, or must operate
under the aegis of a doctor who is adequately qualified to use the machine under
safe conditions for the person subjected to treatment.
This machine can be used in a hospital environment, provided it is used by
personnel qualified in this regard and in compliance with what is stated in the user
manual.
Technical specifications
IACER Srl 49 MNPG356-01
Page 7
Maximum power supplied by the machine:
160 W
230 Vac
3.5 A-T - 5 x 20 mm
115 Vac
3.5 A-T - 5 x 20 mm
Backlit LCD touch-screen display for viewing,
checking operating parameters and setting
them for treatment purposes.
Programmable treatment time
Up to 60 minutes
Electrode holder handpiece
Emission frequency of the handpiece
Modulation
1÷100 Hz (in "Custom" mode)
Resistive, made of AISI 316L
Capacitive, made of
paint
35 mm diameter
60 mm diameter
Adjustable power
0-100%
Double time-delay safety fuse
on the mains (T):
Handpieces supplied
7 inch, 1024*600 resolution
of resistive type
Electrode holder handpiece
of capacitive type
Type of electrodes used
Electrode diameter
IACER Srl 50 MNPG356-01
500 kHz
steel
aluminium and coated in rilsan
Page 8
35 mm
diameter
60 mm
Classification according to directive 93/42/EEC
II B
Insulation class/parts applied according to EN
Ingress protection rating according to EN
60601-1
2 Independent
Stored protocols
34
External dimensions (Width x Height x Depth):
27 x 12 x 30.5 cm
Machine body weight:
3 kg
ambient
(+10 ÷ +40) °C
relative
(10 ÷ 80) % non-condensing
ambient
(-25 ÷ +70) °C
relative
(0 ÷ 93) % non-condensing
Electrode working area
9.61 cm2
60601-1
Output channels
diameter
28.26 cm2
I / BF
IP20
(1 output channel
for capacitive use,
1 output channel
for resistive use)
1 channel for bipolar handpiece
(to be used as an alternative to
the other 2)
1 channel for return plate
Conditions of use
Storage/transport conditions
IACER Srl 51 MNPG356-01
Page 9
Rear panel
Full-colour touch screen display
Encoder knob
Plate connector
Handpiece connectors
Atmospheric
pressure
The service life of the device is set at 3 years.
Device description and controls
(500 ÷ 1060) hPa
IACER Srl 52 MNPG356-01
Page 10
Cooling fan
ON/OFF switch
Power supply socket
Fuse Holder
Return Plate
Bipolar handpiece
Capacitive handpiece
Resistive handpiece
REAR PANEL
OUTPUT PANEL
ACCESSORIES
IACER Srl 53 MNPG356-01
Page 11
Description
Code
Supplied
Optional
Schuko plug power cable
CVAL REA
1
User manual
MNPG356
1
Capacitive handpiece
MAN-CAP REA
1
Kit 2 electrodes for capacitive
treatments (diameters 35, 60 cm)
ELCAP35 REA
ELCAP60 REA
Resistive handpiece
MAN-RES REA
1
Kit 2 electrodes for resistive
treatments (diameters 35, 60 cm)
ELRES35 REA
ELRES60 REA
CAVO+ELRIT/N
REA
CAVO+ELRIT/N
REA XL
Bottle of conductive cream
CREMA-TEC
1
Bipolar handpiece*
MAN-BIP REA
\
X
Handpieces with resistive and capacitive electrodes (see below)
Return Plate
The device is equipped with a mains power cable, and is compatible with the
following kit of accessories supplied:
Steel return plate 12x17 cm
Steel return plate 17x24 cm
IACER Srl 54 MNPG356-01
1
1
1
\ X
Page 12
Symbol
Meaning
*not included in the medical CE certification.
The assembly of the accessories is simple and intuitive: a colour is associated with
the corresponding connector + handpiece + electrode system (in the case of
resistive and capacitive handpiece). he division of colours is shown on the previous
page under "Output panel":
- White: return plate.
- Black: bipolar output.
- Grey: capacitive output.
- Green: resistive output.
In the event of installation problems or difficulties, contact IACER S.r.l.'s technical
support service.
Labelling
Product certification issued by notified body No. 0068.
Manufacturer data.
Warning, see the documents accompanying the product
Date of manufacture (MM-YYYY).
Consult the user manual.
IACER Srl 55 MNPG356-01
Page 13
Symbol
Meaning
SN
Device serial number
RF output signal (label placed near the handpiece
sensitive device (label located near the
RIT
BIP
CAP
RES
WEEE directive for the disposal of electronic and electrical
waste.
Check the contents of the pack. If any item is missing, contact your authorised
I.A.C.E.R. S.r.l. dealer immediately.
IACER Srl 56 MNPG356-01
Page 14
Notes
PRELIMINARY NOTES
The installation of the device is quick and easy.
USE
Interactions with the device take place via the touchscreen. During therapy it is
possible to increase or decrease the therapy using the encoder.
MAINTENANCE
For optimal use of the device and to ensure its maximum performance, it is
recommended to carry out maintenance correctly according to the timings and
ways recommended.
IACER Srl 57 MNPG356-01
Page 15
How to use the device
Introduction to the technology
General information
IACER Srl has recently developed a complete series of devices, accessories and
equipment, designed and built according to the highest quality standards,
adopting cutting-edge technologies in full compliance with the directives and
standards in force.
Particular attention has been paid to design, ease of operation, functionality and
safety. The result is a compact unit with which it is possible to interface via a large
display.
The multiple possibilities of therapeutic applications, together with guaranteed
patient and therapist safety (the unit complies with international standards), make
the machine a high-quality device.
These machines have been designed and manufactured so that their use, if it takes
place under the conditions and for the uses intended, does not compromise the
health and safety of patients, users and third parties, taking into account the
benefit provided to the patient.
These machines are not reserved for diagnosis, prevention, monitoring,
compensation for injury or handicap, replacement or modification of the anatomy,
control of conception, support of vital functions but allow you to treat particular
conditions and actually reduce the disease/condition.
No special intervention is required in case of failure of the medical device, only
normal maintenance/repair operations.
Endothermic therapy
The endothermic therapy system is a recently introduced therapeutic method
within physical therapy.It allows you to stimulate biological structures and natural
reparative and anti-inflammatory processes from the inside through the
application of energy, exploiting a form of interaction between electromagnetic
energy and the tissue, which refers to the electrical concept of the capacitor.This
device consists of 2 facing elements (called plates) separated by an insulating
material, connected to an electric generator which creates a potential difference
(p.d.) between the 2 plates.This causes electric charges to attract and repel each
other concentrating near the 2 elements.In this way there is a positive increase in
the charge density in one plate and negative in the other.
The mobile electrode and the fixed return plate must be positioned in such a way
as to create a circuit between them.
IACER Srl 58 MNPG356-01
Page 16
The current generator works in the field of long wave radio frequencies of 0,5 MHZ,
with a variable power up to a maximum of 160W.
In this way there is no external energy emission, but there is only an endogenous
or internal development at a biological tissue level produced by the movement of
ions and electrolytes, induced by the attraction and repulsion forces that are
generated between the 2 plates of the capacitor.
APPLICATION TECHNIQUES
Capacitive/resistive tecar therapy (acronym of Capacitive and Resistive Energy
Transfer) is a therapy that stimulates the body's natural repair processes,
shortening the recovery time.
Diathermy through the electromagnetic energy/tissue interaction produces an
increase in temperature which occurs inside the tissues in a uniform and controlled
way. This electromagnetic interaction gives rise to the appearance of an ionic flow
with a micro-hyperemia which ultimately favours the release of endogenous
"substances" (especially cortisol and endorphins) which are used to reduce pain,
oedema and inflammation.
CR200 therefore stimulates the increase in blood flow directly, thanks to the
increase in temperature and indirectly through the demand for oxygen by the
treated tissues; the increase in blood promotes the increase of normal immune
defences and stimulates tissue regeneration.
CR200 works in two modes:
- if you work in capacitive mode, there will be an increase in charge density
near the area below the mobile electrode and above all at the superficial
soft tissue level.
- if you work in resistive mode, the concentration of charges and therefore
the biological effect occurs in the tissues with the highest resistance that
are interposed between the mobile electrode and the return plate.
HOW IT WORKS
In order for the phenomenon of the increase in charge density to occur, the two
capacitor plates must be connected to an electric generator which has the task of
supplying the plates with charges.
A real current is thus established, which in the accumulation phase goes from the
generator to the capacitor. As the capacitor accumulates charges, the flow
decreases until it is zero when the capacitor is fully charged.
After this initial phase, if the polarity of the generator is inverted, there will be a
current in the opposite direction which will charge the capacitor with polarity
IACER Srl 59 MNPG356-01
Page 17
opposite to the previous one. If the generator cyclically reverses polarity there will
be a flow in both directions i.e. an alternating current.
The transfer by capacitive contact is carried out through a capacitive electrode
treated with an insulating coating which mobilises the ionic charges in the
subcutaneous tissues.
Rresistive transfer takes place by means of a non-insulated resistive electrode
which mobilises the charges making them concentrate in the areas of greater
depth and resistance.
The capacitive mode therefore acts specifically on soft tissues (superficial muscles,
vascular and lymphatic circulatory system, adipose tissue), whereas the resistive
mode acts on the tissues with greater resistance (bone, cartilage, tendons, deep
muscles).
The mobile electrode (capacitive or resistive) and the plate are positioned so that
the circuit created includes the area to be treated.
For a better transfer of energy to the tissues on the part to be treated, a cream is
used that prevents the interposition of air between the electrode and the body
surface and which favours a homogeneous interaction of the body with the applied
parts of the device.
CR200 is a device manufactured according to the MED 93/42/EEC directive (and
subsequent amendments) relating to medical devices.
Contraindications
Tecar therapy treatments cannot be administered in cases of people with:
− bleeding;
− tumors;
− phlebitis, thrombi and arteriopathies;
− decreased sensitivity in the area to be treated;
− metal prostheses;
− pacemaker;
− hearing aids;
− insulin pumps;
− intrauterine devices;
− fever or infections;
− ongoing pregnancies;
− epilepsy;
− or on children under 14 years.
IACER Srl 60 MNPG356-01
Page 18
Warnings
- The customer is responsible for damage resulting from inadequate
packaging.
case of return to the company.
- Do not use the device in places where it could get wet.
- Carefully check that the connections are in accordance with the instructions
provided before operating the machine.
To avoid any risk of electric shock, the device must only be connected to
-
mains with protective grounding systems.
- Do not use accessories other than the original ones supplied: these could
damage the machine and invalidate the warranty. In the event of installation
problems or difficulties, contact IACER S.r.l.'s technical support service.
- If you use an extension lead that the machine shares with other devices,
check that the total current consumption of the connected devices does not
exceed the maximum current allowed for that type of cable and that it is not
in any case greater than 15A.
- It is not possible to define a suggested number of sessions to evaluate the
effectiveness of the treatment, since they are linked to the power delivered
to the patient undergoing treatment, as well as to the minutes of therapy.The
doctor/healthcare professional must decide the number of therapeutic
sessions to which the patient is subjected according to the specific needs, in
order to guarantee the patient receives effective treatment over time that is
carried out under completely safe conditions.
Check the condition of the power supply cable and the connection cable to
-
the handpiece/applicator often: these must not be damaged or worn.
- It is a class A device in terms of emissions.The device can be used in a hospital
environment provided that it is duly taken into account that the device could
cause disturbance to electronic devices placed in the immediate vicinity.
- No modification of this device is permitted.
- The use of accessories, transducers and cables, other than those specified or
supplied by the manufacturer, could lead to higher electromagnetic
emissions or a decrease in the electromagnetic immunity level of the device,
with consequent incorrect operation.
- The device is not intended for outdoor use.
- The device is not intended for use in the domestic settings.
Keep the original packaging of the machine: it must be reused in
IACER Srl 61 MNPG356-01
Page 19
USE
- It is possible to provide a digital copy of the device's user manual upon
request.
- The device should not be used in close proximity to other equipment and, if
it is necessary to use it near other equipment, the electro-medical device
must be monitored to check the normal operation in the configuration in
which it is used.
- If the electro-medical device, interacting with another device, causes or
receives detected interferences, the user is encouraged to limit such
interference by adopting one or more of the following measures:
• reorient or relocate the receiving device;
• increase the distance between the devices;
• connect the device to a socket of a circuit different from the device(s)
that cause the interference;
• contact the manufacturer or your local technician for assistance.
- Portable and mobile radio communication equipment may affect the
operation of the device.
Use of device
- The perfect functionality of the device is guaranteed if you comply with the
installation and use standards indicated, and only use original accessories and
spare parts.
- In the event of installation problems or difficulties, contact IACER S.r.l.'s
technical support service.
- The correct transport position of the machine requires that the device is
handled only by holding the sides of the machine body with both hands.
: avoid handling the device by the handpiece/electrode holder tabs.
N.B.
- Before connecting the cable to the mains plug, check that the device has not
been damaged during transport and make sure that the characteristics of the
electricity supply on the available power socket meet the nameplate data
shown on the back of the machine.
- The device must only be connected to compliant systems.
- If extension leads are used, check the presence and integrity of the protective
earth conductor.
- Connect the equipment directly to the wall socket, possibly without using
extension leads. Failure to comply with this warning could result in hazardous
electric shocks to people and alter the operation of the machine.
IACER Srl 62 MNPG356-01
Page 20
- The manufacturer is only responsible for the fundamental safety, reliability
and performance of the device if:
•the electrical system of the premises complies with the appropriate
regulations;
•the device is used in accordance with the instructions for use.
GENERAL INFORMATION
- In order to guarantee the operation of the machine in conditions of absolute
safety for the patient, it is advisable to subject the machine to a cycle of routine
checks (at least 2 years) to be carried out by the manufacturer.
- It is recommended to leave the machine on standby for 5 minutes after each
cycle of therapy.
- It is absolutely forbidden to use the device in the presence of flammable
anaesthetic mixtures and oxygen-rich environments. In case of noncompliance with the indication provided, IACER Srl shall not be held
responsible for any accidents.
- It is absolutely forbidden to cover the rear vent: doing so may not allow the
machine to work safely. In case of non-compliance with the indication
provided, IACER Srl shall not be held responsible for any accidents.
- It is important to draw the operator's attention to the need to check the
correctness of the electrical installation of the equipment before operating the
mains switch.
- Before beginning treatment, the operator must make sure that the patient
removes any metal objects worn, so as to avoid triggering hazardous radio
frequency coupling phenomena.
- Tecar therapy treatments must be administered, under the strict control of the
operator, to "conscious" patients, capable of interacting with the operator with
regard to the electrical forces transmitted by the machine.
- It is advisable to suspend the therapeutic treatment if issues should appear
during its administering.
UNPACKING
The device is packed and prepared for shipping with its box, complete with filling
material, designed for safe storage and transport.
To unpack the machine, place the box on a flat, solid surface and remove the upper
polystyrene part.
Carefully remove the appliance.
IACER Srl 63 MNPG356-01
Page 21
INSTALLATION
The installation of the CR200 device is quick and easy.
The recommended environmental characteristics for installation are the following:
- ambient temperature: +10° to +40°C;
- Relative humidity 10 to 80% non-condensing;
- avoid direct exposure to sunlight, chemicals and vibrations.
CONNECTIONS
On the rear of the machine there is the built-in mains power supply module, which
includes the three-pole connector for the power cable, the removable fuse holder
with two fuses (see technical specifications) and the double-pole main switch.
Insert the three-pole female plug of the power cable into the built-in module,
checking that it is perfectly inserted inside the connector.
If extension leads are used, check the presence and integrity of the protective
earth conductor.
Failure to comply with this warning could result in hazardous electric shocks to
people and alter the operation of the machine.
To connect the applied parts, proceed as described: connect the steel plate to the
corresponding connector. Identify the desired mobile handpiece (resistive or
capacitive) for therapy and the associated electrode.The connector + handpiece +
electrode system can be identified by a dedicated colour.The description of the
corresponding colours and outputs is given in the "Output panel" section, in the
chapter relating to the description of the device.
If you need to use the bipolar handpiece, this is the only one that requires you to
disconnect the steel plate. Please note that this handpiece is only intended to be
used in "Custom" mode. This applicator does not fall under the medical CE
certification of the device, as it is suitable only for aesthetic purposes.
After carrying out the checks for correct installation and assembly, turn on the
main power switch, checking that the display comes on correctly.
N.B
: only connect the applied parts necessary for the treatment when using the
device.
OPERATION
The user-device interface is achieved by a large, clear touchscreen display: it
displays all the operational messages relevant for the operator, the operating
IACER Srl 64 MNPG356-01
Page 22
status of the machine during normal therapeutic activity, any error, visual and
acoustic.
The following sections describe how the device menu is divided.
Therapy selection
Within the "Therapy Selection" menu, you can choose from 2 types of pre-set
programmes (Rehabilitation, Sport) or decide to manually set the therapy
specifications (Custom).
Figure 7
PRE-SET PROGRAMMES
When you choose one of the first 2 submenus shown in Figure 1, a screen will open
displaying the list of available programmes. The programme list screen is shown
below (Figure 2):
IACER Srl 65 MNPG356-01
Page 23
Figure 8
Using the touchscreen, you can scroll through the available pre-set programmes.
For further information on the therapeutic protocols present, please refer to the
following chapter "List of programmes".
Once the desired programme has been selected, you can decide to start the
therapy or to return to the list of available programmes, as shown in the following
figure.
Figure 9
IACER Srl 66 MNPG356-01
Page 24
Once the programme has started, it is possible to pause the treatment at any
time.In the case of programmes that involve the use of both electrodes (resistive
and capacitive), the specific duration of the individual treatment phases are given,
in the predetermined order.
CUSTOM
When you press the button for the "Custom" programme, the following screen
appears.
Figure 10
Before starting the therapy, it is possible to choose the type of treatment
(capacitive/resistive/bipolar), as well as the duration, the carrier frequency and
the modulation.
WARNING
The selected therapy parameters can be stored by pressing the "Save" button. The
menu relating to free memory slots will then open, as shown below in Figure 5.
IACER Srl 67 MNPG356-01
: when setting the duration of the therapy, avoid 0:00.
Page 25
Figure 11
Once the desired memory slot has been selected, the device will ask you to enter
a name in order to save the Custom programme.
Figure 12
IACER Srl 68 MNPG356-01
Page 26
Time (min) and mode
1
Phantom limb
500
15 C + 15 R
2
Knee arthrosis
500
15 R + 10 C
3
Arthrosynovitis
500
10 R + 10 C
4
Bursitis
500
10 R + 20 C
5
Neck pain
500
15 R + 10 C
6
Whiplash syndrome
500
10 R + 20 C
7
Chondropathy
500
10 R + 10 C
8
Muscle contracture
500
5 R + 15 C
9
Contusion
500
20 R + 10 C
10
Muscular sprain
500
20 R + 10 C
11
Sprain
500
12 R + 12 C
12
Muscle distraction
500
10 R + 20 C
13
Joint pain
500
10 R + 10 C
14
Oedema
500
20 R + 10 C
15
Epicondylitis
500
10 R + 10 C
Each programme saved in free memory slots can be called up via the "Load" button
in the "Custom programme" menu (see Figure 4).
After starting the programme, you can adjust the power (initially set to 0) using
the encoder knob. To increase the power output, turn the knob clockwise.To
decrease the power, turn the knob in the opposite direction.
At the end of the therapy, press the "BACK" button until you reach the main menu,
then turn off the device using the switch on the rear panel. Finally unplug the
power plug from the socket.
The proposed preset programmes are the result of operational experience gained
over many years supporting expert professional users. The following section ("List
of programmes") contains the list of available protocols, by category.
List of programmes
REHABILITATION
No. Programme CarrierFrequency
(kHz)
(Capacitive/Resistive)
IACER Srl 69 MNPG356-01
Page 27
Time (min) and mode
16
Epitrochleitis
500
10 R + 10 C
Anterior cruciate ligament
(ACL)
18
Lymphoedema
500
15 C
19
Low back pain
450
10 C + 10 R
synovitis)
21
Muscle tear
500
10 R + 20 C
22
Acute tendinitis
500
10 R + 10 C
23
Chronic tendinitis
500
10 R + 10 C
24
Rotator cuff tendinopathy
500
25 C + 15 R
(kHz)
1
Muscle contracture
500
5 R + 15 C
2
Contusion
500
20 R + 10 C
3
Muscle recovery
500
10 C + 10 R + 5 C
4
Sprain
500
12 R + 12 C
5
Hematoma
500
20 R + 10 C
6
Tendon injury
500
10 R + 10 C
7
Superficial massage*
500
10 R + 10 C
8
Deep massage*
500
10 R + 10 C
10
Muscle tear
500
10 R + 20 C
No. Programme CarrierFrequency
17
20 Meniscopathy (meniscal
(kHz)
500 15 C + 15 R
500 10 R + 10 C
SPORT
No. Programme CarrierFreq.
(Capacitive/Resistive)
Time (min) and mode
(Capacitive/Resistive)
9 Muscle sprain 500 20 R + 10 C
*Treatment not covered by medical CE
IACER Srl 70 MNPG356-01
Page 28
Device care
Maintenance
CR200 Tecar therapy devices do not require particular maintenance operations,
except for routine maintenance and cleaning of the applicator handpieces, with
the aim of ensuring the best operating conditions, to guarantee the effectiveness
of the treatment and patient safety.
External cleaning of the equipment must only be done with a soft cloth moistened
with hot water, or using non-flammable cleaning liquids. The front control panel
can also be cleaned in the same way.
The handpieces/applicators, in particular the treatment head, must be periodically
cleaned with
Carefully put away the handpieces/applicators at the end of each treatment.
Contact IACER Srl authorised centres for information on original accessories and
spare parts.
Do not spray, nor pour liquids on the external container of the devices, nor on the
ventilation slots.
Do not immerse the machine in water.
After cleaning the outside of the box, make sure all parts are dry before putting
the device back into operation.
Under no circumstances must the device be disassembled for cleaning or checking
purposes: there is no need to clean the machines internally, and in any case this
operation must be done exclusively by IACER Srl specialised and authorised
technical personnel.
More specifically:
- handle the handpiece-applicator with care: rough handling can negatively
-
- Under no circumstances must the device be disassembled for cleaning or
- Do not use thinners, detergents, acid solutions, aggressive solutions or
water and denatured alcohol.
influence its performance and characteristics.
Under no circumstances are unauthorised technical personnel allowed to
open and/or disassemble the handpiece/applicator: such tampering, in
addition to damaging the characteristics of the handpiece, immediately
invalidates the warranty.
checking purposes: there is no need to clean the machine internally, and
in any case this operation must be done exclusively by IACER Srl
specialised and authorised technical personnel.
flammable liquids to clean the machine exterior and accessories. The use
of these substances, coupled with improper use of the accessories,
IACER Srl 71 MNPG356-01
Page 29
besides irreparably damaging the appliance, also invalidates the
warranty.
- For optimal use of the device and to ensure its maximum performance, it
is recommended to carry out maintenance correctly according to the
timings and ways recommended.
- To correctly replace the fuses on the machine, follow the instructions
below:
6. use a screwdriver to open the fuse box, taking care to insert the
screwdriver into the slot created on the fuse box and levering it
outwards
7. remove the fuse-holder structure by sliding it along the guide
8. remove the inserted fuses and replace them with new ones
9. insert the fuse-holder structure in the fuse box, sliding it towards
the guide
10. close the plastic door of the fuse box
− It is recommended to carry out routine maintenance every year,
checking:
• the intensity of any leakage currents;
• the continuity, and therefore the integrity, of the earth conductor;
• the correctness of the insulation resistance value
in order to guarantee the electrical safety of the device and to make sure that
it operates under the guaranteed safe conditions. For such operations, we
recommend contacting a qualified technical service centre or alternatively
IACER Srl or one of its authorised centres.
WARNING!
For safety reasons, before carrying out any maintenance and cleaning
-
operations on the device, it is NECESSARY to turn off the device using the
switch on the rear panel and disconnect the power cable from the power
socket.
- It is recommended to carefully clean the machine and accessories supplied
before using it in contact with the patient.
- It is useful to draw the operator's attention to the need for routine
maintenance of the handpieces/applicators, to be carried out by the
manufacturer.
- Cleaning and disinfection must always be done before carrying out the
therapeutic treatment on the patient.
IACER Srl 72 MNPG356-01
Page 30
PROBLEM
POSSIBLE CAUSE
SOLUTION
Mains plug not inserted correctly
Check the operation of the
Mains cable not correctly
device.
Insert the plug and cable
connector of the appliance.
Mains cable worn and broken.
Replace the mains cable.
The switch is off.
Operate the mains switch.
The LCD display
Replace the missing,
- Do not spray or pour liquids on the external container of the device, on the
ventilation slots, on the LCD display or on the fan grille.IACER Srl will not be
held responsible for any damage occurring if the machine has been used
without carrying the maintenance operations described above.
- Check the condition of the power supply cable and the connection cables of
the applicators/accessories applied to the patient often: these must not be
damaged or worn.
It is advisable to have the fuses replaced by personnel with adequate
-
technical expertise and training, in order to carry out the operation safely.
- Do not open the device: there are high electric voltages inside which can be
dangerous.
- Only technical personnel authorised by the manufacturer can access the
internal parts of the device. For repairs and further information it is necessary
to contact IACER Srl or its authorised service centres.
Troubleshooting
CR200 Tecar therapy machines have been designed and built adopting advanced
technological solutions, quality components, for continuous use that is always
efficient and reliable.
If, however, there is a problem with operation, it is recommended to consult the
following guide before contacting an authorised service centre.
The LCD display
on the front
panel does not
turn on: the
device does not
work.
on the front
IACER Srl 73 MNPG356-01
inserted in the connector of the
Defective or blown fuse or fuses.
in the power socket.
power socket.
correctly into the
defective or blown fuse (s).
Page 31
PROBLEM
POSSIBLE CAUSE
SOLUTION
panel does not
The LCD display
Some controls
Defective buttons or keys.
Carefully check the output
Handpiece-applicator cable
Replace the defective
handpiece-applicator which
shows clear signs of wear in
the treatment head and on
the cable.
Fault in the electronic circuit of
Contact an IACER Srl
The device works
Not perfectly efficient connection
Perform the maintenance
cable and the connector of
turn on: the
device does not
work.
on the front
panel does not
turn on.
on the front
control panel do
not work
properly.
The device does
not activate
during delivery.
Electronic control circuit failure.
Defective components on the
electronic control board.
Electronic control circuit failure.
Bad connections in the output
circuit applied to the patient
broken or incorrectly connected
Output cables are worn and/or
have loose contact.
Contact an IACER Srl
service centre.
Contact an IACER Srl
service centre.
Contact an IACER Srl
service centre.
connections are correct
and not damaged.
as normal, but
there is a
noticeable drop
in the
effectiveness of
the treatment.
IACER Srl 74 MNPG356-01
the power generator.
of the handpiece-applicator
output circuit.
service centre.
operations described.
Install and position the
device as described.
Check the condition of the
the handpiece-applicator.
Page 32
PROBLEM
POSSIBLE CAUSE
SOLUTION
Mechanical damage (due to falls
Check the perfect
Electronic circuit of the generator
Possible failure of the appliance's
or hard knocks) on the
handpiece-applicator, in
particular on the radiating head.
not perfectly calibrated.
power generator circuit.
When the conditions listed below occur, disconnect the device from the electrical
system and contact IACER Srl's technical support service:
• the cable or the built-in rear power supply module is worn or damaged;
• liquid has entered the device;
• the device has been exposed to rain.
-
Only technical personnel authorised by the manufacturer can access the
internal parts of the device.
- For repairs and further information it is necessary to contact I.A.C.E.R.Srl or
its authorised service centres.
WARNING!
− DO NOT OPEN the unit, there are HIGH ELECTRIC VOLTAGES inside which
can be DANGEROUS.
adherence of the parts
applied on the surface
involved in the treatment.
Contact an IACER Srl
service centre.
Disposal Information
CR200 Tecar therapy devices, in line with operating and safety requirements, have
been designed and built to have a minimal negative impact on the environment.
The criteria followed are those of minimising the amount of waste, toxic materials,
noise, unwanted radiation and energy consumption.
Careful research into optimising machine performance guarantees a significant
reduction in consumption, in accordance with the subject of energy saving.
This symbol indicates that this product should not be
disposed with other household waste.
IACER Srl 75 MNPG356-01
Page 33
Correct disposal of obsolete equipment, accessories and especially batteries, helps
to prevent possible negative consequences on human health and the
environment.
The user must dispose of the equipment to be scrapped by taking it to the
collection centre indicated for the subsequent recycling of electrical and electronic
equipment.
For more detailed information on the disposal of obsolete equipment, contact
your local council, waste disposal service or shop where you purchased the
product.
Warranty
I.A.C.E.R.Srl guarantees the quality of its devices, when used in accordance with
the instructions provided in this manual, for a period of 12 months from the date
of purchase.
During the warranty period, defective products will be repaired or replaced at the
company's discretion.
Replacement of the device itself is not part of the warranty.
The warranty does not cover malfunctions or damage resulting from
- inadequate placement and installation;
- incorrect use or that not in compliance with the instructions given in this
manual;
- improper or inadequate maintenance by the user;
- operation that does not comply with the environmental specifications
indicated for the product;
- unauthorised opening of external enclosures;
- tampering and/or unauthorised modifications;
- use of non-original accessories.
The warranty is provided ex I.A.C.E.R. srl's registeredoffice.
If a return is necessary, follow the packing instructions below and attach a copy of
the purchase receipt.
It is advisable to insure the shipment.
Before shipping the machine due to a suspected fault, it is recommended to
carefully consult the MAINTENANCE and TROUBLESHOOTING chapters: issues are
largely attributable to poor maintenance or minor technical problems that the user
can easily resolve.
A simple phone call to I.A.C.E.R Srl's Technical Service can be of great help in
solving any problem.
Instructions for packing and returning the device:
:
IACER Srl 76 MNPG356-01
Page 34
6. disconnect the power and connection cables with handpieces, applicator
devices, etc .;
thoroughly clean and disinfect all accessories and parts of the machine
7.
that have been in contact with the patient.
For obvious hygiene reasons, in order to guarantee adequate health
protection of technical personnel (workplace safety law, Legislative
Decree81/2008),
receiving them will not be checked;
8. disassemble the accessories and any mechanical supports;
9. reuse the original box and materials for packaging;
10. attach to the shipment the Support Request Form including the reasons
for the request for inspection/service, the type of failure or malfunction:
such information will help technicians and significantly reduce repair
times.
appliances deemed hygienically unsafe by the staff
Support
The manufacturer is the only point of contact for technical support regarding the
device.For all technical support matters, please contact:
I.A.C.E.R.S.r.l.
Via S. Pertini, 24/a • 30030 Martellago (VE)
Tel.: +39 041.5401356 • Fax: +39 041.5402684
Technical documentation concerning repairable parts may be provided, but only
with prior company authorisation and only after giving proper training to the
maintenance personnel.
Spare parts
Original spare parts for this device can be ordered at any time from the
manufacturer. To order them contact:
I.A.C.E.R.S.r.l.
Via S. Pertini, 24/a • 30030 Martellago (VE)
Tel.: +39 041.5401356 • Fax: +39 041.5402684
In order to preserve the warranty, guarantee operation and safety of the product,
it is recommended to only use original spare parts supplied by the manufacturer
(also see the Warnings paragraph).
IACER Srl 77 MNPG356-01
Page 35
Sound energy emitted
The sound energy emitted by the CR200 device during normal use (except for the
audible alarm signals) does not exceed the following levels:
-
80dBA for 24 h cumulative exposure over 24 h; to this value is added a deviation
of 3dBA when the total duration of exposure is halved over 24 h (for example
83dBA for 12 h over 24 h);
-
a sound pressure level of 140dBC (peak) for pulse energy or noise impact.
Interference and electromagnetic compatibility tables
The CR200 Tecar therapy device has been designed and built in compliance with
EN 60601-1-2:2015 and the current ELECTROMAGNETIC COMPATIBILITY
DIRECTIVE 2014/30/EC, with the aim of providing reasonable protection from
harmful interference in residential, civil and hospital settings.
All the necessary measurements and checks were carried out at I.A.C.E.R. srl's inhouse testing, measurement and inspection laband specialised external centres.
Upon request, customers may view the reports on EMC measurements at the
company.
Based on their operating principle, the CR200 tecar therapy devices do not
generate significant radio frequency energy and have an adequate level of
immunity to radiating electromagnetic fields. Under these conditions, harmful
interference cannot occur to radioelectric communications and to the operation
of electro-medical devices used for monitoring, diagnosis, therapy and surgery, to
the operation of electronic office devices such as computers, printers, copiers,
faxes, etc. and to any electrical or electronic appliance used in such environments,
provided that they comply with the ELECTROMAGNETIC COMPATIBILITY directive.
In any case, to prevent any problem with interference, it is recommended to
operate any therapy device at an appropriate distant from critical equipment for
monitoring patients' vital functions and to use caution in therapeutic applications
on patients with pacemakers.
CR200.All rights reserved. CR200 and the logo are the exclusive property
of I.A.C.E.R. Srl and are registered trademarks.
IACER Srl 78 MNPG356-01
Page 36
is recommended that
Campo magnetico a
30 A/m
30 A/m
I campi magnetici a
Nota_e UT è la tensione di rete in c.a. prima dell’applicazione del livello di prova
frequenza di rete
(50/60 Hz)
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
<5% UT
(>95% buco in_dip
)
in U
T
per_for 5 sec
<5% UT
(>95% buco in_dip
)
in U
T
per_for 5 sec
the CR200 be powered
from an uninterruptible
power supply or a
battery
frequenza di rete
dovrebbero avere livelli
caratteristici di una
località tipica in
ambiente commerciale
o ospedaliero.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment
UT is the a.c. mains voltage prior to application of the test level
IACER Srl 202 MNPG428-00
Page 37
GUIDA E DICHIARAZIONE DEL COSTRUTTORE – IMMUNITÀ ELETTROMAGNETICA –
La CR200 è prevista per funzionare nell’ambiente elettromagnetico sotto specificato. Il
Prova di immunità
Livello di prova
Livello di
Ambiente elettromagnetico –
Gli apparecchi di
PER GLI APPARECCHI ED I SISTEMI CHE NON SONO DI SOSTENTAMENTO DI FUNZIONI
VITALI
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
– FOR EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING
cliente o l’utilizzatore della CR200 deve garantire che esso venga usato in tale
ambiente.
The CR200 is intended for use in the electromagnetic environment specified below. The
customer or the user of the CR200 should assure that it is used in such an environment.
Immunity Test
IEC 60601
IEC 60601 test
level
conformità
Compliance
level
Guida
Electromagnetic environment -
Guidance
comunicazione a RF portatili e
mobili non dovrebbero essere
usati più vicino a nessuna parte
della CR200 compresi i cavi,
della distanza di separazione
raccomandata calcolata con
l’equazione applicabile alla
frequenza del trasmettitore
Portable and mobile RF
communications equipment
should be used no closet to any
part of the CR200, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the
transmitter.
IACER Srl 203 MNPG428-00
Page 38
Distanza di separazione
3 Veff_Vrms
P
V
d
=
1
5,3
P
V
d
=
1
12
P
E
d
=
1
12
raccomandata
Recommended separation
distance
RF condotta
Conducted RF
IEC 61000-4-6
da 150 kHz a
80 MHz
150 kHz to 80
MHz
6 Veff_Vrms
da 150 kHz a
80 MHz per
banda ISM
150 kHz to 80
MHz for ISM
band
3 Veff_Vrms
([V
] V)
1
6 Veff_Vrms
([V
] V)
1
for ISM band
RF irradiata
Radiated RF
IEC 61000-4-3
IACER Srl 204 MNPG428-00
3 V/m
da 80 MHz a
2,7 GHz
3V/m
[E
] V/m
1
da 80 MHz a
800 MHz
80 MHz to 800
MHz
Page 39
80MHz to 2,7
P
E
d
=
1
7
RF irradiate per
P
E
d
=
1
6
ove P è la potenza massima
dispositivi di
comunicazione
radio
Radiated RF to RF
wireless
communication
equipment
IEC 61000-4-3
GHz
3 V/m
da 80 MHz a 6
GHz
80 MHz to 6
GHz
3V/m
[E
] V/m
1
da 800 MHz a
2,7 GHz
800 MHz to 2,7
GHz
da 80 MHz a
6 GHz
80 MHz to 6 GHz
IACER Srl 205 MNPG428-00
nominale d’uscita del
trasmettitore in Watt (W)
secondo il costruttore del
trasmettitore e d è la distanza
di separazione raccomandata in
metri (m).
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
Page 40
Note_s:
Le intensità di campo dei
trasmettitori a RF fissi, come
determinato da un’indagine
elettromagnetica
potrebbe essere minore del
livello di conformità in ciascun
intervallo di frequenza
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
should be less than the
compliance level in each
frequency range
Si può verificare interferenza in
prossimità di apparecchi
contrassegnati dal seguente
simbolo:
a
del sito
b
.
b
a
,
Interference may occur in the
vicinity of equipment marked
with the following symbol:
IACER Srl 206 MNPG428-00
Page 41
(1) A 80 MHz e 800 MHz; si applica l’intervallo di frequenza più alto.
a Le intensità di campo per trasmettitori fissi come le stazioni base per
At 80 MHz and 800 MHz, the higher frequency range applies.
Queste linee guida potrebbero non applicarsi in tutte le situazioni. La propagazione
elettromagnetica è influenzata dall’assorbimento e dalla riflessione di strutture,
oggetti e persone.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
radiotelefoni (cellulari e cordless) e radiomobili terrestri, apparecchi di radioamatori,
trasmettitori radio in AM e FM e trasmettitori TV non possono essere previste
teoricamente e con precisione. Per valutare un ambiente elettromagnetico causato da
trasmettitori RF fissi, si dovrebbe considerare un’indagine elettromagnetica del sito. Se
l’intensità di campo misurata nel luogo in cui si usa un CR200, supera il livello di
conformità applicabile di cui sopra, si dovrebbe porre sotto osservazione il
funzionamento normale della CR200. Se si notano prestazioni anormali, possono
essere necessarie misure aggiuntive come un diverso orientamento o posizione della
CR200.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
CR200 is used exceeds the applicable RF compliance level above, the CR200 should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the CR200.
b L’intensità di campo nell’intervallo di frequenza da 150 kHz a 80 MHz
dovrebbe essere minore di [V1] V/m
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than [V1] V/m.
IACER Srl 207 MNPG428-00
Page 42
DISTANZE DI SEPARAZIONE RACCOMANDATE TRA APPARECCHI DI
La CR200 è prevista per funzionare in un ambiente elettromagnetico in cui sono sotto
Potenza di uscita
Distanza di separazione alla frequenza del trasmettitore (m)
0,01
0,12
0,2
0,12
0,23 – 0,1
0,38
0,63
0,38
0,73 – 0,2 – – – –
0,9 1 1,20
2,0
1,20
2,30 – 1,8 – – – –
2,7
RADIOCOMUNICAZIONE PORTATILI E MOBILI PER LA CR200 CHE NON SONO DI
SOSTENTAMENTO DELLE FUNZIONI VITALI
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT THE CR200 THAT ARE NOT LIFE-SUPPORTING
controllo i disturbi irradiati RF. Il cliente o l’operatore della CR200 possono contribuire a
prevenire interferenze elettromagnetiche assicurando una distanza minima fra gli
apparecchi di comunicazione mobili e portatili a RF (trasmettitori) e la CR200 come sotto
raccomandato, in relazione alla potenza di uscita massima degli apparecchi di
radiocomunicazione.
The CR200 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the CR200 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the CR200 as recommended
below, according to the maximum output power of the communications equipment.
massima del
trasmettitore
specificata
Rated maximum
output power of
transmitter
W
IACER Srl 208 MNPG428-00
Separation distance according to frequency of transmitter (m)
150 kHz
to
80 MHz
150 kHz
to
80 MHz
(ISM
band)
Da 80
MHz to
800
MHz
800
MHz to
2,7 GHz
Da 80 MHz to 6 GHz
(to RF wireless radio
communication
equipment)
Page 43
2 – – – –
2,8
10
3,80
6,3
3,80
7,30 – 100
12,00
20
12,00
23,00
–
Per i trasmettitori specificati per una potenza massima di uscita non riportata sopra, la
Nota_e
distanza di separazione raccomandata d in metri (m) può essere calcolata usando
l’equazione applicabile alla frequenza del trasmettitore, ove P è la potenza massima
d’uscita del trasmettitore in Watt (W) secondo il costruttore del trasmettitore
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
(1) A 80 MHz e 800 MHz, si applica l’intevallo della frequenza più alto.
At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
(2) Queste linee guida potrebbero non applicarsi in tutte le situazioni. La
propagazione elettromagnetica è influenzata dall’assorbimento e dalla riflessione di
strutture, oggetti e persone.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
IACER Srl 209 MNPG428-00
Page 44
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.