Italray MOEBS03 OPERATOR MANUAL

OPERATOR MANUAL
Code MOEBS03
VIA DEL PARLAMENTO EUROPEO, 9D – I-50010 SCANDICCI (FI) - TEL. ++39-0557228511 - FAX ++39-0557228512
ELECTROMEDICAL RADIOLOGICAL EQUIPMENT
www.italray.it info@italray.it
BS 45 MOEBS03
REVISIONE REVISION MOEBS03 - VALID STARTING FROM S/N 12-244-03
PREPARATO PREPARED BETTACCINI LEANDRO
VERIFICATO CHECKED RICCARDO SARTI
APPROVATO APPROVED STEFANO BALDINI
DATA DI EMISSIONE ISSUE DATE 13/10/2003
REVISION INDEX
MOIBS01
24/04/2003 MOIBS02 – Change of telephone number Italray, general revision
13/10/2003 MOIBS03 – General revision of the device.
FOLLOWING PRODUCT EQUIPMENT DOCUMENT
Part Number:
S/N.
DOCUMENTATION ATTACHED TO THE MANUAL
Cassette holder manual KL
CODE / PART NUMBER REVISION
OPERATOR MANUAL
Page 2 of 8
BS 45 MOEBS03
CAUTION !
Carefully read the contents of this manual before starting the equipment.
Although this equipment is designed and constructed respecting the most up-to-date safety standards, the X-ray source always constitutes a danger when the operator is not qualified or informed. Excessive exposure to X-rays causes bodily damage.
Therefore all precautions must be taken to avoid unauthorized and unqualified persons using this apparatus so constituting a danger to themselves and others.
Before carrying out any operation, the persons qualified and authorized to use this equip­ment must be informed about the protection measures established by the International Commission on Radiological Protection and the relevant national standards.
The X-ray emission control point must be in a zone screened from radiation.
If it is necessary to move into the immediate equipment area during an exposure, put on X-ray protection garments.
The company refuses all liability for consequences following interventions made by unau­thorized personnel.
The equipment must not be used for any purposes other than those intended.
The company refuses all liability if authorized personnel do not adhere to the requirements contained in this manual during the performance of their tasks.
If abnormalities or malfunctions occur that cannot be resolved as indicated in this manual the authorized person is to consult the equipment manufacturer for the necessary relevant in­structions:
ITALRAY s.r.l.
Via del Parlamento Europeo 9/D – 50010 - Scandicci – Florence - ITALY Telephone +39 055 7228511 Fax +39 055 7228512 E-mail info@italray.it Website www.italray.it
OPERATOR MANUAL
Page 3 of 8
BS 45 MOEBS03
Index
1 INTRODUCTION ..............................................................................................................................4
2 DESCRIPTION .................................................................................................................................4
3 TECHNICAL DATA...........................................................................................................................4
4 INSTRUCTIONS FOR USE ..............................................................................................................6
5 CLEANING, ROUTINE AND NON-ROUTINE MAINTENANCE ........................................................7
6 EQUIPMENT FEATURES.................................................................................................................8
1 INTRODUCTION
This manual provides the user with instructions for correct equipment use.
The equipment must not be used for purposes other than those intended.
The user is responsible for fulfilling legal obligations regarding the installation and the
functioning of the equipment itself.
The manufacturer cannot be considered liable for any breakages, malfunction or bodily damage if the equipment is not operated correctly, or is not suitably maintained.
2 DESCRIPTION
The BS 45 is a compact vertical Bucky complete with mobile grid and long vertical travel, de­signed for radiographic examinations of patients in the upright position and on patient trol­leys.
3 TECHNICAL DATA
The equipment meets the following directives :
Directive 93/42/EEC (Class I) Directive 89/336/EEC and later amendments.
and the following technical standards:
IEC EN 60601-1 (Class I type B) and later amendments. IEC EN 60601-1-2 IEC EN 60601-1-3 IEC EN 60601-2-32
OPERATOR MANUAL
Page 4 of 8
BS 45 MOEBS03
Transport and storage conditions Temperature : -10° to +70° C Humidity : 20 to 80%
Operating environment conditions Temperature : +10° to +40° C Humidity : 30 to 75%
ELECTRICAL SPECIFICATIONS Power input : 24V DC
BUCKY
- Motorized grid movement (with variable speed in factory)
- Grid ratio 1:10 D.F.F. 150. Inch scales 103 (standard grid)
- Cassette holder, manual self-centring.
IDENTIFICATION An identification data plate is positioned above left Fig. 1 indicating equipment code, series, number and other necessary information.
When requesting maintenance or replacement parts it is always necessary to give this infor­mation.
ACCESSORIES ON REQUEST
- Ionization chamber for automatic exposure meter.
- Grid to customer's specifications.
- “Potentiometer” kit for automatic Bucky tracking.
- Patient armrest.
OPERATOR MANUAL
Page 5 of 8
BS 45 MOEBS03
1
DISPOSAL OF MATERIALS
- Packing:
Wood and bubble sheet 100% recyclable
- Assembly materials :
Carbon steel 100% recyclable Elastomers and plastic materials 100% recyclable Cables and other electrical materials
The equipment contains no toxic or noxious substances.
Disposal must conform with the local laws in force.
4 INSTRUCTIONS FOR USE
Release the Bucky's vertical movement by operating the
release button, positioned on the top of the handle Pos.
Position the patient and the Bucky, keeping the release
button pressed.
Halt the Bucky by releasing the button Pos. c.
Centre the X-ray tube
Pull out the cassette holder Pos.
Position the securing spring in the special seating for the
radiographic cassette format used.
Fig. 1
Close the clamps.
Lock the clamps with the securing device e Fig. 2
Insert the cassette holder c Fig. 2 into the Bucky.
Ensure that the BS45 accessory has been correctly inserted on the X-Ray generator
Open the clamps. d Fig. 2
Insert the radiographic cassette
c
Fig. 2
c
Perform the X-ray exposure
Pull out the cassette holder
Turn the locking lever Pos.
Remove the radiographic cassette.
Push the cassette holder c Fig. 2 into the Bucky.
OPERATOR MANUAL
Fig. 2 , open the clamps Pos. d Fig. 2
e
Page 6 of 8
BS 45 MOEBS03
FIG. 2
2
1 – X-RAY CASSETTE HOLDER
1
2
3
4
2 – CASSETTE SUPPORT CLAMP
3 - CASSETTE SECURING DEVICE
4 – LOWER X-RAY CASSETTE ARREST
5 CLEANING, ROUTINE AND NON-ROUTINE MAINTENANCE
Only clean the painted surfaces using a slightly dampened cloth and neutral detergent, dry with a dry cloth, do not use solvents (benzine, alcohol, acetone, etc.), corrosive and abra­sive agents. Do not use acids to clean the floor.
The routine and non-routine maintenance must be carried out only by the manufacturer's
personnel or personnel authorized by them, and following the maintenance manual. Rou­tine maintenance is to be made annually.
The company refuses all liability following interventions by persons other than those de­scribed above.
OPERATOR MANUAL
Page 7 of 8
BS 45 MOEBS03
2
4
6
7
6 EQUIPMENT FEATURES
1
5
3
FIG. 3
1 –TELERADIOGRAPH BS 45 COLUMN 2 – BUCKY 3 – ANTI-DIFFUSER GRID 4 – IONIZATION CHAMBER (OPTIONAL) 5 – RADIOGRAPHIC CASSETTE HOLDER 6 – ELECTROMECHANICAL BRAKE 7 – PATIENTS' HANDLE (OPTIONAL)
OPERATOR MANUAL
Page 8 of 8