![](/html/2d/2d59/2d5948ffab25d59ebdfa07a01cf240a4a9150a05be20df03eafc5ab255ed320d/bg1.png)
84
|
CALIBRATION SOLUTIONS
100 to 1300°C
Outer Case
Fibre Insulation
Ceramic Enclosure
with heater windings
Insulation
Connector tube from
Target to case
Blackbody Target
Outer case with
Te rminal box
Fibre Insulation
Ceramic Enclosure
with heater windings
Height adjustment
Fibre Insulation
Control Thermocouple
(included)
Access for a
Reference
Thermocouple
(not included)
Fibre Insulation
Metal Outer Case
Optical Access
Optical access tube
fitted between metal case
and the blackbody. Travels
through heater ceramics.
Blackbody
Cyclops Assembly Diagram Plan View (shown in section)
g
Spherical Blackbody Source
g
Wide Temperature Range 100°C to 1300°C
g
Can be adapted for Thermocouple Calibration
The Cyclops Model 878 is a spherical blackbody source.
It consists of an inner black sphere that sits inside a
spherical furnace and is suitable for use as a radiation
source for infrared thermometers.
The inner sphere has a nominal diameter 230mm and is
accessed by an optical sighting tube. The furnace can be
supplied in one of two constructions, one providing an
aperture size of 17mm and one of 45 mm.
The furnace can be adapted for thermocouple operation
by replacing the inner sphere with an equalizing block and
using a different control sensor.
Blackbody Source
Cyclops
Model 878
Temperature Range 100°C to 1300°C
Emissivity Greater than 0.999
Stability ±0.1°C
Display resolution 0.1°C to 999.9;
1°C from 1000 to 1300
Time to temperature 90mins hour to 700°C
4 hours to 1300°C
PC Interface included
Power 3kW typical
Voltage 100-130 or 208-240 Vac
50/60Hz
Dimensions 425mm Diameter
Weight 25 kg
Options
Ceramic Equalising Block to accept up to 8 thermocouples
878-02-08.
How to Order
Model 878 Cyclops. Please state supply voltage required
Please state target diameter either 17mm standard or
45mm to special order