36
|
SOLUTIONS FOR PRIMARY & SECONDARY LABORATORIES - Isotech North America - www.isotechna.com - ph: +802-863-8050 - email: sales@isotechna.com
0 to 1600°C
g
Type R and Type S
g
Gas Tight Assembly
g
Premium Grade Wire
The Isothermal range of Thermocouple Standards are the
result of many years development. The type R and S
standards will cover the range from 0°C to 1600°C.
The measuring assembly comprises a 7mm x 300mm or
600mm gas tight 99.7% recrystallized alumina sheath
inside which is a 2.5mm diameter twin bore tube holding
the thermocouple.
The inner 2.5mm assembly is removable since some
calibration laboratories will only accept fine bore tubed
thermocouples and some applications require fine bore
tubing.
Thermocouple
Standard
Type R & S Standard Thermocouple, Model 1600,
Premium grade wire, gas tight assembly,
No intermediate junctions.
The covered noble metal thermocouple wire connects the
measuring sheath to the reference sheath which is a
4.5mm x 250mm stainless steel sheath suitable for
referencing in a 0°C reference system. Two thermo
electrically free multistrand copper wires (teflon coated)
connect the thermocouple to the voltage measuring device.
The thermocouple material is continuous from the hot or
measuring junction to the cold, or referencing junction.
Calibration
The 1600 is supplied with a certificate giving the error
between the ideal value and the actual emf of the thermocouple at the gold point. For types R and S thermocouples,
manufacturing tolerances are small and, therefore, the use
of a standard reference table is particularly apt. A few
calibration points, only, are required to determine the small
differences between the characteristics of an individual
thermocouple and the standard reference table. As an
example of consistency, 48 Isotech thermocouples
calibrated at NPL, had a standard deviation of the
differences from the reference table value at the gold
point (11, 364µV) of only 7µV, equivalent to about 0.5°C.
Thermocouple characteristics are sufficiently smooth to
allow interpolation of deviations from the reference table
to be carried out over fairly wide temperature spans
without introducing unacceptable errors. Isotech can
offer a 4-point UKAS calibration for temperatures up to
1100°C (supplied as standard), a 6 point UKAS calibration
up to 1300°C with the option of a table of millivolts to
degrees Celsius in 10°C steps or, alternatively, arrange
for an NPL calibration for temperatures up to 1600°C.
Please contact Isotech to obtain current prices for
calibration.
Also available without the physical cold junction - Specify No Cold Junction (NCJ).
Model 1600
Hot Sheath
Temperature Range 0°C to 1600°C (R or S)
Emf Vs Temperature According to relevant document
Response Time 5 mintues
Hot Junction see diagram
Dimensions
Connecting Cable see diagram
Cold Junction 250mm long x 4.5 diameter
Copper Extension Wires 2000mm
Immersion 100mm min.
Case Dimensions Height 65mm
Width 710mm
Depth 165mm
Gross Weight 900g
Feature Removeable inner assembly
The standard thermocouple described can be supplied
in the following noble metal combinations
TYPE R: Platinum vs Platinum 13% Rhodium
TYPE S: Platinum vs Platinum 10% Rhodium
How to order
Model 1600 Type R/300
Model 1600 Type R/600
Model 1600 Type S/300
Model 1600 Type S/600
If cold junction not required, specify NCJ.
UKAS calibration is included