i SENS ISMART300A User Manual

Operator’s Manual
Software version 1.0.0.0
P/N 7983 Draft
July 2015
Copyright © 2015 i-SENS, Inc. All Rights Reserved
Operators should read the entire manual
before installing and operating the analyzer.
Contents
1 Introduction
Safety ··························································································· 1-1
Symbols and Terms ············································································· 1-3
Product Description ············································································ 1-5
2 Installation
Installation Requirements ······································································ 2-1
Power Cable Connection ······································································· 2-2
Power-On ······················································································· 2-3
Printer Paper Installation ······································································· 2-5
Cartridge Installation ··········································································· 2-7
3 Analyzer Settings
Contents
Screen Menus ··················································································· 3-1
Sample Setup ·················································································· 3-10
QC Setup ······················································································ 3-15
Analyzer Setup ················································································ 3-17
Interface Setup················································································· 3-21
4 Sample Analysis
Patient Samples ················································································· 4-1
QC Samples ···················································································· 4-8
5 Database
Patient Data ····················································································· 5-1
QC Data ························································································ 5-5
Calibration Data ················································································ 5-9
Cartridge Data ················································································· 5-12
Contents
6 Shutdown
Analyzer Shutdown ············································································· 6-1
Cartridge Restart ················································································ 6-3
7 Maintenance
Calibration ······················································································· 7-1
Cartridge Removal ·············································································· 7-4
System Information ············································································· 7-6 Analyzer Status ············································································· 7-7 Cleaning ···················································································· 7-8
8 Troubleshooting
Troubleshooting ················································································· 8-1
Error Code ······················································································· 8-6
9 Specification
Operating Specification ····································································· 9-1 Analyzer Specification ····································································· 9-2 Cartridge Specification ····································································· 9-3 Measuring Principles ······································································· 9-4
Appendices
Appendics A: Sample Collection and Handling ·············································· 10-1 Appendics B: Order Information ····························································· 10-2 Appendics C: Warranty ······································································· 10-3
1. Introduction
Safety ··················································································· 1-1
Symbols and Terms···································································· 1-3
Product Description ··································································· 1-5
1. Introduction
1. Introduction
Safety
General Warnings
General Caution
Use only the provided power cord and adapter.
Do not use damaged power cord, adapter, or loose outlet.
Never touch the power supply accessories with wet hands.
Keep dust away from the outlet and power cord.
Do not use or store any hot equipment or flammable material
near the analyzer or power supply accessories.
Do not allow water or foreign substances to get into the analyzer
or power supply accessories.
Do not directly look at the red light emitting from the barcode
scanner as direct eye exposure to this light can damage vision.
Power cord must be plugged into a grounded outlet.
The rear vent of the analyzer must be free of obstruction and not
covered by cloth or any other material.
Do not install or operate the analyzer in an area where
ferromagnetic fields are generated (e.g., the MRI room).
Before long term storage of the analyzer, remove the cartridge,
turn off the power, and clean any contamination from the analyzer.
There are no operator seviceable parts inside the product.
Do not disassemble, repair, or modify the product.
If electromechanical problems are suspected, call a service
engineer to report the problems.
When disposing the analyzer and provided electrical accessories,
contact your local distributor where you purchased them. You must not discard this electrical/electronic product in domestic household waste.
Changes or modifications not expressly approved by i-SENS, Inc.
could void the purchaser’s authority to operate the equipment.
1-1 i-Smart 300 Blood Gas Analyzer
Safety, continued
1. Introduction
Biohazard Caution
All materials used in collecting blood and/or other samples from
humans should be treated as biohazardous materials with the potential carrying infectious agents capable of producing disease.
All biohazardous materials should be handled and disposed of in
accordance with applicable rules and regulations of the hospital, laboratory, or other testing facility.
Wear appropriate personal protective clothing (lab coat, gloves,
goggles, etc.) before operating the analyzer.
Be careful not to let the sample and/or collecting tools directly
touch the mouth, eyes, mucus membranes, or any area with broken skin.
Wash hands after collecting the sample or using the analyzer.
Operator’s Manual 1-2
1. Introduction
Authorized representative in the European
This marking indicates that you must not discard
DC power port
Off (power)
On (power)
Symbols and Terms
Symbols
Symbol Description
CE mark (In Vitro Diagnostic Directive 98/79/EC)
In vitro diagnostics device
Consult operating instructions
community
Serial number
Caution: Attention, see operating instructions
Manufacturer
Temperature limitation
Catalogue number
Batch code
Contains sufficient for <n> tests
Use by
Biological risk
this electrical/electronic product in domestic household waste. Distributors of this product within the EU have taken the necessary step to comply with the Waste Electrical and Electronic Equipment (WEEE), Directive 2002/96/EC.
Printer
1-3 i-Smart 300 Blood Gas Analyzer
Symbols and Terms, continued
A statement that alerts the operator to
the
1. Introduction
Terms
Term Description
pH
pCO2
pO2
cNa+
cK+
cCa2+
cCl-
Hct
Cal
Cal 1
Cal 2
Negative log of the hydrogen ion activity
Partial pressure of carbon dioxide
Partial pressure of oxygen
Concentration of sodium ion
Concentration of potassium ion
Concentration of calcium ion (ionized calcium)
Concentration of chloride ion
Hematocrit
Calibration
1-Point calibration
2-Point calibration
QC
Warning
Caution
Quality control
A statement that alerts the operator to the possibility of injury, death, or other serious adverse reactions associated with the use or misuse of the device.
possibility of device malfunction, device failure, damage to the device or damage to other property associated with the use or misuse of the device.
Operator’s Manual 1-4
1. Introduction
cover
scanner
Handle
Product Description
Intended Use
Analyzer
i-Smart 300 is an automated system that measures pH, pCO2, pO2, cNa+, cK+, cCl-, cCa2+, and hematocrit in whole blood.
It is intended for use in a point-of-care setting or laboratory environment.
Touchscreen
Barcode
Sampler
Cartridge
door
Printer
VGA port
LAN port
Power switch
Power port
Serial port
Vent
USB ports
i-Smart 300 Analyzer
1-5 i-Smart 300 Blood Gas Analyzer
Product Description, continued
1. Introduction
Accessories of
analyzer
Cartridge
Power adapter Power code
Printer papers
Cartridge cover
i-Smart 300 Cartridge
Operator’s Manual 1-6
1. Introduction
This page is intentionally left blank.
1-7 i-Smart 300 Blood Gas Analyzer
2. Installation
Installation Requirements ··························································· 2-1
Power Cable Connection ···························································· 2-2
Power-On ············································································· 2-3
Printer Paper Installation ····························································· 2-5
Cartridge Installation ·································································· 2-7
2. Installation
2. Installation
Installation Requirements
Unpacking
Environmental Requirements
Inspect the shipping package of the product for any signs of
damage from shipment.
Ensure all listed components are included in the package.
Location : indoors
Temperature : 15 ~ 35°C
Humidity : 5 ~ 85%, relative humidity
Altitude : under 3,000 m
Note:
The analyzer should be installed on a horizontal, stable surface in
a well-ventilated area without direct sunlight or heat.
Allow enough space for the cartridge door of the analyzer to
open.
Allow at least 10 cm of space from the back of the analyzer for
adequate ventilation.
This equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Electrical
Volt AC : 100 ~ 240 Va.c., grounded electrical supply
Requirements
Amp : 1.5 A
Frequency : 50/60 Hz
Power cord : 3-wire, approved
Power adapter : +24 Vd.c., 2.7 A (Output)
2-1 i-Smart 300 Blood Gas Analyzer
Power Cable Connection
2. Installation
Caution
Connect Power Cable
Check that the power switch is in “ ” (OFF) position.
Always connect the power according to the steps described in
this manual.
Always use a grounded electrical outlet.
Use only the power cord and power adapter provided by the
manufacture.
Electrical plugs and sockets vary by countries. If necessary, use
appropriate power plug or power strips to connect the provided power cord to the outlet.
1. Plug the power cord into the socket of the power adapter.
2. Plug the power adapter into the power port on the back of the
analyzer.
3. Plug the power cord into an electrical outlet.
Operator’s Manual 2-2
2. Installation
Power-On
Caution
Turn Power Switch On
Remove any attached USB memory device or network cable
from the analyzer before turning the power switch on.
2. Facing the back of the analyzer, flip the black power switch on
the lower left side, marked “ ” to the “ ” position.
3. The screen will turn on and the analyzer will boot.
4. If the boot is completed, the message of the last shutdown
information will appear.
2-3 i-Smart 300 Blood Gas Analyzer
Power-On, continued
2. Installation
Turn Power Switch On, continued
5. Check that the power cord icon appears at the bottom
row of the screen.
Note:
If the power cord icon does not appear on the screen,
check for any loose power connections between the analyzer and the electrical outlet.
If no loose power connection is found and the power cord icon
still does not appear, call a service engineer immediately.
6. Check the date and time of the analyzer and, if necessary, change
date and/or time. Refer to 3. Analyzer Settings
Note:
Changing the date and time of the analyzer is only allowed
before cartridge is installed on the analyzer. Once the cartridge is installed, the date and time cannot be changed until the cartridge is removed.
Operator’s Manual 2-4
2. Installation
Printer Paper Installation
Install Printer Paper
1. Facing the back of the analyzer, push the PUSH button and flip
down the printer cover.
2. Place the paper roll into the paper compartment and place the end
of the paper in parallel with the printer slot.
2-5 i-Smart 300 Blood Gas Analyzer
Printer Paper Installation, continued
2. Installation
Install Printer Paper, continued
3. Close the printer cover.
Note:
The printer will cut the paper exposed to the outside
automatically.
Operator’s Manual 2-6
2. Installation
Cartridge Installation
Install Cartridge Note:
After opening the cartridge, it must be installed directly to the
analyzer.
1. Take out the cartridge from the cartridge box.
2. Press down the cover lock and remove the cartridge cover.
3. Press Install at the following screen
4. Slide down the lever on the cartridge door, open the cartridge
door.
2-7 i-Smart 300 Blood Gas Analyzer
Cartridge Installation, continued
2. Installation
Install Cartridge, continued
5. Insert the new cartridge into the analyzer, follow the instructions
on the screen.
6. Facing the cartridge label, insert the cartridge into the analyzer.
7. Close the cartridge door and slide up the lever on the cartridge
door. If the door is properly locked, it will make a clicking sound.
Operator’s Manual 2-8
2. Installation
Cartridge Installation, continued
Install Cartridge, continued
8. If the RFID tag information of the cartridge is valid, the analyzer
will change to the Cartridge Warming-Up screen.
9. The cartridge warming-up will take approximately 25 minutes.
Note:
Once the cartridge cover is removed, be careful with sharp edges
of the cartridge parts inside.
2-9 i-Smart 300 Blood Gas Analyzer
3. Analyzer Setting
Screen Menus ·················································································· 3-1
Sample Setup ·················································································· 3-10
QC Setup ······················································································· 3-15
Analyzer Setup ··············································································· 3-17
Interface Setup ················································································ 3-21
3. Analyzer Setting
3. Analyzer Setting
Screen Menus
Ready Screen
No.
Screen naming
Analyzer operation schedule
Remaining sample number and date of the installed cartridge
Current state of interface setup
Current state of power connection
Current state of and battery recharge level
Current date and time of the analyzer
Menu access
Current state of each sensor: Pressing each button twice will display the latest slope of the
sensor.
Description
3-1 i-Smart 300 Blood Gas Analyzer
Screen Menus, continued
3. Analyzer Setting
Menu options
Operator’s Manual 3-2
3. Analyzer Setting
Screen Menus, continued
Menu Access
1. Press Menu. The quick menus and the category menus will appear.
2. Press the current screen box at the top right corner to show the
current menu options.
Note:
The Menu button is disabled on the following screens:
- Install Cartridge, Remove Cartridge
- Introduce Sample, Sample Information, Sample Results
- Introduce QC Sample, QC Information, QC Results
3-3 i-Smart 300 Blood Gas Analyzer
Screen Menus, continued
3. Analyzer Setting
Icons and Buttons
Icons and Buttons Description
Print results or settings on
Transmit results to LIS/HIS /
screen
Transfer complete
Close current screen
Go to list screen
Go to search screen
View result screen
Operator’s Manual 3-4
Scroll to previous or next page
Go to previous or next result
screen
3. Analyzer Setting
Screen Menus, continued
Icons and Buttons, continued
Icons and Buttons Description
Delete entered data on current
Execute search
Run 1-point calibration or
2-point calibration
Run QC
Accept or discard QC results
Add or delete QC lot
information
tab
Okay or accept
Save
Go to previous screen
Copy cartridge data to USB
memory
3-5 i-Smart 300 Blood Gas Analyzer
Screen Menus, continued
3. Analyzer Setting
Icons and Buttons, continued
Icons and Buttons Description
Unselected item
Selected item
Item that were transmitted to
Show keyboard for data entry
Show keyboard for numeric
LIS/HIS
Selected item that were
transmitted to LIS/HIS
entry
Operator’s Manual 3-6
Interface was enabled
Analyzer is on AC power
Battery level
3. Analyzer Setting
Screen Menus, continued
Keyboards
The analyzer offers following keyboards for data entry.
Alphabetic Keyboard
Numeric & Special Character Keyboard
Date and Time Entry Keyboards
3-7 i-Smart 300 Blood Gas Analyzer
Loading...
+ 75 hidden pages