Register your instrument online for extended warranty
Thank you for purchasing your Ion Science instrument.
The standard warranty of your instrument can be extended to up to five years
on PhoCheck Tiger and two years on other Ion Science instruments.
To receive your extended warranty, you must register your instrument online
within one month of purchase (terms and conditions apply.)
Visit www.ionscience.com/instrument-registration
Part number: 849215
Page 3
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Declaration of conformity
Manufacturer: Ion Science Ltd, The Way, Fowlmere, Cambridge, SG8 7UJ, UK
Product: TVOC
Product Description: Anintrinsically safe fixed continuous monitor comprising of a photo-ionisation
detector for detecting and measuring volatile organic compounds with a 4-20mA
output
Directive 94/9/EC ATEX
Identification: II 2G EEx ia IIC T4 (-20
o
C ≤ Ta ≤ +50
Notified Body: Baseefa Ltd, 1180, Buxton, UK
Identification: II 3G Ex nA IIC T4 (-40oC < Ta < +50oC)
Ion Science Ltd Self certified EC Type Examination Certificate(s)
Baseefa05ATEX0277 latest supplement Baseefa05ATEX0277/2 issued 24
Ionscience09849 Examination certificate issued 4th September 2009
Ref Ion Science Cert Report 849237, 849242
International Standards
IECEx BAS 06.0057 latest revision no.1 issued 20th November 2008
Ref Baseefa Cert Report 05(c)0423, 06(c)0614, 08(c)0578
Ex ia IIC T4 (-20oC Ta + +50oC)
Standards
BS EN 60079-0:2006 Electrical Apparatus for Potentially Explosive Atmospheres – General Requirement
BS EN 60079-11:2007 Explosive Atmospheres - Equipment Protection by Intrinsic Safety ‘i’
BS EN 60079-15:2007 Explosive Atmospheres - Equipment Protection by other means ‘n’
BS EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control and
EN61000-6-3:2007 Class 1 (light industrial and domestic) Electromagnetic Compatibility (EMC) Generic
standards. Emission standard for residential, commercial and light industrial
environments
EN50270:2006 Electromagnetic compatibility. Electrical apparatus for the detection and
measurement of combustible gases, toxic gases or oxygen.
Other Standards
BS EN ISO 9001:2008 Quality Management Systems – Requirements
BS EN 13980:2002 Potentially Explosive Atmospheres – Application of Quality Systems
On behalf of Ion Science Ltd, I declare that, on the date this product accompanied by this declaration is
placed on the market, the product conforms to all technical and regulatory requirements of the above listed
directives.
Name: Mark Stockdale Position: Technical Director
Signature: Date: 20th November 2009
o
C)
th
November 2008
Page 3 of 32
Page 4
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Contents
Declaration of conformity ............................................................................................................................... 3
Responsibility for use ..................................................................................................................................... 5
Introduction to TVOC ...................................................................................................................................... 6
Packing list ....................................................................................................................................................... 8
Standard TVOC instrument ............................................................................................................................ 8
Cable and gland requirements ..................................................................................................................... 11
Dimensions for mounting ............................................................................................................................. 12
Power requirements ..................................................................................................................................... 13
Start Up ........................................................................................................................................................ 16
Adjusting the Duty Cycle .............................................................................................................................. 16
Setting the ZERO ......................................................................................................................................... 18
Setting the gas concentration (The actual value that the calibration has been supplied at) ....................... 19
Setting the SPAN ......................................................................................................................................... 19
System recommendations ............................................................................................................................ 24
Remote area monitoring............................................................................................................................... 24
Gas sample systems .................................................................................................................................... 24
Instrument warranty and service ................................................................................................................. 26
Service ......................................................................................................................................................... 26
TVOC detects a large range of gases, which are potentially dangerous from both a poisoning and/or an
explosive perspective. TVOC has a number of selectable features allowing the detector to be used in a
variety of applications. Ion Science Ltd can accept no responsibility for the incorrect adjustment of features
that cause harm or damage to persons or property. It is the users’ responsibility to respond appropriately to
the readings delivered.
Inadequate performance of the gas detection equipment described in this manual may not necessarily be
self-evident and consequently equipment must be regularly inspected and maintained. Ion Science
recommends that personnel responsible for equipment use institute a regime of regular checks to ensure it
performs within calibration limits, and that a record be maintained which logs calibration check data. The
equipment should be used in accordance with this manual, and in compliance with local safety standards.
Warning
1. Substitution of components may impair intrinsic safety and result in unsafe conditions.
2. For reasons of safety, TVOC must only be operated and serviced by qualified personnel.
3. Please read and understand this user manual fully before installing, operating or servicing TVOC.
Quality assurance
TVOC has been manufactured in compliance with ISO9001:2008, which ensures that the equipment
supplied to our customers has been designed and assembled reproducibly, from traceable components, and
leaves Ion Science calibrated to stated standards.
Disposal
Dispose of TVOC and its components in accordance with all local and national safety and environmental
requirements. This includes the European WEEE (Waste Electrical and Electronic Equipment) directive. Ion
Science Ltd offers a take back service. Please contact us for more information.
Legal Notice
Whilst every attempt is made to ensure the accuracy of the information contained in this manual, Ion Science
accepts no liability for errors or omissions, or any consequences deriving from the use of information
contained herein. It is provided "as is" and without any representation, term, condition or warranty of any
kind, either express or implied. To the extent permitted by law, Ion Science shall not be liable to any person
or entity for any loss or damage which may arise from the use of this manual. We reserve the right at any
time and without any notice to remove, amend or vary any of the content which appears herein.
Page 5 of 32
Page 6
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Introduction to TVOC
TVOC is a fixed continuous monitor for the detection and measurement of total volatile organic compounds,
which can be dangerous from both a poisoning and explosive perspective. Total VOC’s are detected using
PID (Photo ionisation detection) technology.
TVOC has a number of user selectable features which define its operation and will be application dependent.
The user can define the detection ranges or 0.01 - 10ppm, 0.1 – 100 ppm or 1 – 1,000 ppm (Default);
detection units, either ppm or mg/m3; and the duty cycle time i.e. the frequency that the instrument takes a
measurement. Duty cycles are explained in more detail under TVOC set up, and Operation.
IMPORTANT: TVOC can be calibrated on alternative gases to Isobutylene however to ensure TVOC
remains within specification, the correct concentration for the alternative gas must be chosen.
Example: Benzene gas has a response factor of 0.5 which means only half as much concentration is
required to produce the equivalent to Isobutylene.
The ideal gas / concentrations used to set the calibration SPAN are shown below: Lower limit Upper limit
10 ppm range 8 ppm 12 ppm Isobutylene equivalent
100 ppm range 90 ppm 110 ppm Isobutylene equivalent
1,000 ppm range 90 ppm 110 ppm Isobutylene equivalent
How to calculate alternative concentration of gases for gases other than Isobutylene:-
Calibration Ideal response Response factor Ideal calibration gas
Gas for Isobutylene concentration
Benzene 100 x 0.5 = 50 ppm
Isoprene 100 x 0.7 = 70 ppm
Isobutylene 100 x 1.0 = 100 ppm
Carbon disulfide 100 x 1.4 = 140 ppm
Ethyl acrylate 100 x 2.0 = 200 ppm
For other gas response factors please refer to the Ion Science web site or contact Ion Science Ltd.
http://www.ionscience.com/GasSearch/tabid/87/Default.aspx
WARNING: Ignoring these guide lines may result in calibration failure, loss of resolution or loss of range.
The default settings of TVOC are:
Detection range – 1 – 1000 ppm
Units - ppm
Signal update - 60 seconds
TVOC gives a continuous 4-20 mA output which can be integrated into a DCS control system to give an
indication of VOC levels in the operating environment. Note the 4-20 mA output must be externally powered
with 8-35 V.
In addition to the 4-20 mA output TVOC has an LCD display showing gas concentration and 4 colour LED’s.
LED functions are as follows:
Green - is an indicator of TVOC’s working status
Red - is the fault indicator
Yellow x 2 - are calibration status indicators
Please see Calibration for more information on LED status.
Page 6 of 32
Page 7
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Introduction to TVOC
For installation requirements please refer to Installation and Technical Specification of this manual. Before
attempting an installation, please fully read and understand this user manual, and for hazardous area (IS)
installations please also refer to the TVOC Intrinsically safe certificate for further details.
The TVOC safety rating permits its deployment in all hazardous areas of the quoted (or less demanding)
rating. For detail please refer to the marking on your instrument (found on the front main label). As of 1st of
October units will be manufactured with two protection concepts applied by separate certifications. Intrinsic
safety (ia) and Non sparking (nA). All units manufactured up to that date will only have intrinsic safety.
a) Intrinsic Safety (ia) permits the deployment in areas where explosive gases (of group IIA, IIB and IIC) are
intermittently present (zone 1) with an ambient temperature range of -20 to 50°C. Intrinsically safe
installations will require the use of safety barriers and appropriate wiring.
b) Non sparking (nA) permits deployment in areas where explosive gases (of group IIA, IIB, IIC) are unlikely
but possible (Zone 2) with ambient temp range of -20°C to 50°C. Non sparking installations do not require
safety barriers however the maximum working voltage should be strictly adhered to as a matter of safety. As
no safety barriers are required there is complete flexibility on the implementation of a three wire system.
TVOC does require regular calibration onsite please see the Calibration section for more information. TVOC
calibration mode is accessed via a magnetic on site switch system.
Page 7 of 32
Page 8
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
3. Calibration
1. Calibration
6. Allen keys X 2
4. Aspirator
2. Calibration
5. Carbon
Diagram 1
Packing list
All items of equipment shipped by Ion Science Ltd are packed in suitable containers and enclosed in a shock
absorbing filling which affords a considerable degree of protection against physical damage.
Contents should be carefully removed and checked against the packing list. Any discrepancies between the
contents and the packing list must be reported to Ion Science Ltd within ten days of receipt of shipment. Ion
Science cannot be held responsible for shortages not reported with in the period.
TVOC has a number of settings that can be selected by the
user via a row of four selector pins mounted on the reverse side
of the main PCB. Diagram 2 shows the location of the functional
selector pins labelled - A, B, C & D.
The selector pins absence or presence determines the chosen
setting.
TVOC is shipped with all four selector pins fitted as shown in
diagram 3.
Diagram 4 shows the removal of 1 selector pin.
The following table shows the selector pin combinations and the corresponding function.
Selector pin ‘B’ and ‘C’ selects the ranges - 10 ppm, 100 ppm or 1000 ppm
Selector pin ‘D’ selects the units - ppm’ or ‘mg/m3
Selector pin ‘A’ - Fault output range
NOTE: During an Alarm condition (F1 or F2) the output will drop to either 3.5 mA or 2.0 mA.
Jumper ‘A’ fitted: 3.5 mA during an alarm condition
Jumper ‘A’ removed: 2.0 mA during an alarm condition.
As default, the TVOC has a 100 ppm factory calibration
Page 9 of 32
Page 10
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
TVOC setup
NOTE: * TVOC is shipped with ALL selector pins fitted as standard
* TVOC only reads the selector pin settings when power is connected.
Always ensure power is disconnected from TVOC before changing
setting or carrying out maintenance.
* Never place selector pins on the programming port connector.
* Ensure static build up is discharged before touching components.
* If the jumper selection is not a recognised option then error F4 will occur. To recover switch
the instrument off perform the correct jumper selection and switch the instrument back on.
Duty Cycle
The life of a photoionization lamp is used up when the lamp is illuminated, the lamp window can also
contaminate resulting in a reduced level of detection. The TVOC pulses the lamp on and off which extends
lamp life and reduces lamp contamination. This in turn can increase the time between servicing.
The TVOC switches on the lamp, takes a reading and then
updates the 4-20 mA output and display screen. The default Duty
Cycle is the frequency between test cycles, the factory default is
set to 1 minute however this time can be adjusted from 0 to 5
minutes.
Please note: When bump testing after a calibration the response
may be delayed depending on when in the gas is applied during
the test cycle.
Fast response mode
The TVOC can be set to Fast Response Mode, when selected the TVOC will update the output every one
second. Fast response mode can be selected by setting both the seconds and minutes to zero.
Please note: The warrantee of Ion Science PID lamps is based upon a duty cycle of 1 minute. Lamps used
with Duty Cycle below one minute will not be covered under the standard warrantee terms.
Page 10 of 32
Page 11
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Installation
Location
There are many variables involved in defining the optimum location for a gas detector. Obvious though it may
sound, the most important rule is:
A detector will not detect gas unless gas actually reaches the detector.
TVOC should be mounted in the location most likely to detect gas.
TVOC must be mounted vertically with the sensor underneath the case. This avoids dust and debris
blocking the access to the detector cell.
If possible, mount TVOC near the ceiling to target VOC gases that are lighter than air or just above
floor level to detect VOC gases that are heavier than air.
Mount TVOC in an area that has good air circulation. Restricting natural air current may result in
delayed detection.
Never mount TVOC in direct sunlight or over a heat source such as a radiator. This may cause TVOC
to exceed its certified working temperature range.
Do not mount TVOC in areas likely to flood.
Mount TVOC’s in areas that are easily accessible for servicing.
Percentage variations in the composition of air (78% Nitrogen, 21% Oxygen and 1% Argon) may affect
the detected signal.
Please refer to the technical specifications of the instrument under Technical Specifications of this
manual before beginning a TVOC installation.
Cable and gland requirements
The Screw terminal sockets for connecting wires in the TVOC accept wires of 0.5mm2 to 2.5mm
C.S.A.
Screened cables for both power and signal out are necessary to achieve EMC compliance. Cable
screens must be terminated at both ends for them to be effective in achieving electromagnetic
compatibility (EMC). The screens should be terminated to the TVOC enclosure using EMC
compatible cable glands. It is important for an EMC compatible gland to make electrical contact to the
enclosure. This is usually done using a serrated washer or EMC lock nut that bites through the paint
making electrical contact. EMC lock nuts have been provided. These locknuts are orientation specific
the points should be facing against the case in order to cut through the paint during tightening.
Two M20 cable glands and one blanking plug have been supplied as standard to ease installation in
most circumstances. These parts will not be appropriate for all applications so they should be used at
the discretion of the installation engineer. It is recommended that the manufacturer’s instructions are
followed for the gland and blanking plug installation. The M20 blanking plug has been included to
enable installations using a single entry cable (for example a three wire system) to be implemented.
While Ion Science can not recommend the cable gland suitable for every application the following
information may be useful to the installation engineer for the selection of the appropriate cable gland.
For intrinsically safe operation using safety barriers, the only requirements on the cable
glands are that they are better than IP20 rated. This may seem extraordinary however
where the power is safely limited to the TVOC it is intrinsically safe so does not rely on the
prevention of dirt or moisture into the enclosure.
For intrinsically safe operation in Zone 2 without safety barriers, cable glands must be ATEX
or IECX rated (Ex e, Ex n or Ex d) and at least IP54.
Installation of the TVOC depends on preferences of the installer, intrinsically safe operating legislation and
the application.
Ion Science suggests connecting the screens from cables via a gland as shown on the diagrams and under
Installation Section of this manual. There may however be technical circumstances where connection to the
PCB is necessary.
2
Page 11 of 32
Page 12
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Diagram 5
Installation
Dimensions for mounting
Page 12 of 32
Page 13
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Diagram 6
Diagram 7
Installation
NOTE: The TVOC case can be used as a template when marking out fixing holes but do not drill
through the fixing holes.
Power requirements
Non Intrinsically Safe Operation applications:
Input power 5-28 Vdc. 80 mA max. (0.5 to 2.5 mm2 C.S.A.)
Output power (4-20mA) 8-35 Vdc. 80 mA max. (0.5 to 2.5 mm2 C.S.A.)
Note: 4-20mA loop must be externally powered.
4 Wire System
(Non IS)
3 Wire System
(Non IS)
Page 13 of 32
Page 14
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Installation – Hazardous location
Warning!
The label on the enclosure details the IS and nA certifications. The appropriate section of the Intrinsically
Safe label should be covered up or blocked out to reflect the type of installation. For Non-intrinsically Safe
applications the label should be covered completely. For Intrinsically Safe (IS) Zone 2 applications with out
safety barriers the ia certification details should be blanked out. For Zone 1 or Zone 2 IS applications with
safety barriers the nA certification details should be covered. This will prevent safety discrepancies and or
mis-use in the future should the equipment be moved or the site be re-defined as a hazardous area.
Entry parameters for intrinsically safe Zone 2, nA (non-sparking) areas only:
Input power 5-24 Vdc, 80 mA max
Output power (4-20 mA) 8-35 Vdc, 22 mA
For information only: please see Intrinsically Safe operation certificate before installation.
The wiring of Intrinsically Safe systems is identical to the wiring showing for non Intrinsically Safe
applications. The only difference is the requirements to ensure that the power supplied used provides 24 V or
less in normal operation.
Intrinsically Safe (IS) applications: Entry parameters
Input power Ui = 18V Ii = 800mA Pi = 1.2W Ci = 0F Li = 0mH
Output power(4-20mA) Ui = 30V Ii = 200mA Pi = 1.2W Ci = 0F Li = 0mH
For information only please see Intrinsically Safe certificate before installation
Ion Science Ltd suggest using the following zener barriers for IS approved applications. Use competent
installation engineers and ask them for installation and application advice.
Note
All cable lengths are estimated for Zone 1 IIC installation of the equipment
Using 1 mm2 C.S.A conductor S.T.P cables with capacitance per meter of 250 pF/m, L/R ratio of 25
muH/ conductor resistance at 20C of 18.4 /km.
Calculation assumes cables are at 50C.
Warning! Intrinsically Safe (IS) applications
Units should not be powered using non IS power supplies prior to installation in an IS application.
If non IS power is supplied the unit will require inspection by Ion Science Ltd or an Ion Science Approved
Service Centre prior to installation in an IS application.
TVOC fuses may not be replaced in the field.
If a fuse is blown TVOC will require inspection by Ion Science Ltd or and Ion Science Approved Service
Centre before it is used in an IS application.
Page 14 of 32
Page 15
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Diagram 8
Installation – Hazardous location
4-wire Control equipment sensing 4 – 20 mA on return line
Notes:
For intrinsically safe installations, the enclosure may be opened while the system is live,
provided the work is undertaken by competent personal and suitable safety precautions are
taken. This overrides the warning on the enclosure label.
Initial calibration
TVOC is calibrated at Ion Science before dispatch using 100 ppm Isobutylene. However, if you wish to
calibrate your TVOC after installation, Ion Science recommends that the TVOC is left to run on its chosen
settings (see ‘TVOC Setup Section’) for 24 hours before an initial calibration is carried out, to allow the
instrument to stabilise.
Page 15 of 32
Page 16
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Operation
Start Up
After electrical power is connected, TVOC runs through a ‘Start up’ routine, which lasts for approximately 1
minute. During this ‘Start up’ routine TVOC demonstrates the following characteristics:
* The LCD screen displays the software version number
* The GREEN LED flashes
* The 4-20 mA output is set to 4 mA (0.0 ppm)
NOTE: If the 0-10 ppm range is selected the TVOC will require calibration using 10ppm Isobutylene before
use (See the calibration section of this manual)
During this ‘Start-up’ time the user can adjust the Duty Cycle time as referred to in ‘TVOC Setup’ under the
Duty Cycle section. This time can be adjusted from 0 to 5 minutes in 10 second increments.
Adjusting the Duty Cycle
NOTE: Please read this entire procedure before attempting to adjust the lamp OFF time.
There are two stages to adjusting the Duty Cycle time. The first stage adjusts the seconds 0 to 59 seconds,
the second stage adjusts from 0 to 5 minutes. Please see the description of the Duty Cycle on page 10.
1. During the 1 minute ‘Start up’ time place the calibration magnet over the top
right hand corner of the Ion Science logo. (See diagram 9) Two numbers on the
right of the LCD screen will flash.
Diagram 9
2. While the magnet is held in this position the number on screen will roll over from 00
through to 50 in 10 second steps. Quickly remove the magnet when the desired time
appears.
3. After 5 seconds the TVOC will move on to the next stage. (Reapply the magnet
within 5 seconds if further adjustment is necessary)
4. The TVOC will move to the next stage indicated by a digit appearing on the left of
the screen. Holding the magnet in place will cause this number to roll over from 0 to 5
minutes. Quickly remove the magnet when the desired time appears.
5. After 5 seconds the TVOC will display both the minutes and seconds selected.
(Reapply the magnet within 5 seconds if further adjustment is necessary)
6. TVOC will restart the start-up sequence and then enter running mode. The Duty
cycle is then saved to memory. The duty cycle will be remembered through a power
cycle.
Fast response mode
If the Duty Cycle time is adjusted lower than 0:10 (10 seconds) the instrument will update the output every
one second. In this mode of operation the F2 fault condition is disabled. This means that removal of the
sensor may go undetected.
Page 16 of 32
Page 17
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Diagram 10
Operation
LED’s
see Diagram 12
The green indicates TVOC’s working status:
ON Indicates the PID lamp is OFF
Flashing Indicates the PID lamp is ON
A red indicates fault status:
ON Indicates TVOC cannot function correctly. The 4-20mA output drops
to 3.5mA.
URGENT attention is required.
Flashing Indicates the TVOC cannot function but a reading is being
attempted
A yellow Zero LED indicates during normal operation
ON Last calibration zero level was outside expected limits
(The instrument will operate normally with the previous calibration)
A yellow Span LED indicates during normal operation
ON Last calibration Span level was outside expected limits
(The instrument will operate normally with the previous calibration)
last calibration carried out was outside expected limits. The instrument will continue
working but service is required.
Please see Section 7 Calibration for YELLOW LED status.
Selector pins
TVOC has a range of options that are selected via the selector pins on the back of the main PCB (please see
the ‘TVOC Set up’ Section).
The options are shown below with the defaults highlighted in bold:
1-1000ppm or 0.01 – 10 ppm or 0.1 – 100 ppm
ppm units or mg/m3 units
Calibration mode
TVOC calibration mode can be accessed using the calibration
magnet supplied in the TVOC calibration tool kit (Part number: A-
845214).
To enter calibration place the magnet over the top right of the Ion
Science Logo until the state changes (i.e. an LED illuminates and/or
the display changes).
Ensure the magnet is withdrawn away from the magnetic switch by
at least 1 cm once a change has occurred.
Calibration gases
TVOC has three (3) ranges that can be selected however the 010ppm range requires the instrument is calibrated by the user before
use. The 0-100ppm and the 0-1000 ppm ranges can initially run
using the Factory calibration carried out during the instruments
manufacture, however because the 0-10 ppm range is more
sensitive and more susceptible to environmental and system
tolerances the instrument must be calibrated before use.
If the 0-10 ppm range is selected TVOC will display the following
screen to indicate that a calibration is required before it can be used.
Follow the calibration instructions before use.
Page 17 of 32
Page 18
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
STATUS LED
ZERO LED
SPAN LED
FAULT LED
Calibration adaptor
Carbon filter
Hand
Diagram 11
Diagram 12: LED’s
STATUS, ZERO, SPAN, FAULT
Diagram 13
Calibration
Calibration routine
Ion Science recommends calibrating TVOC after any maintenance or lamp cleaning is carried out and on a 3
monthly basis to ensure TVOC is working to specification.
NOTE: Please read this entire calibration procedure before attempting a calibration.
TVOC calibration is undertaken in four steps:
1. Setting the ZERO (Using clean gas via carbon filter)
2. Setting the gas tolerance (1 – 200 ppm Isobutylene on 0-100 ppm and 0-1000 ppm ranges)
(0.1 – 20.0 ppm Isobutylene on the 0-10ppm range)
3. Setting the SPAN (Using the SPAN gas)
Setting the ZERO
1 – Place the calibration magnet over the top right hand corner of the Ion Science logo to enter
calibration mode. The ‘ZERO LED’ will illuminate and the ‘STATUS LED’ will extinguish.
2 – Remove the calibration magnet
3 – Insert the calibration adaptor into the sensor cap, attach the carbon filter to the calibration adaptor.
The carbon filter should be attached to the hand aspirator.
4 – Again place the calibration magnet over the logo. The ‘ZERO LED’ will flash, during this stage
TVOC displays a direct millivolt (mV) output from the PID sensor.
aspirator
Page 18 of 32
Page 19
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Calibration adaptor
SPAN gas
Calibration connector
Diagram 14
Calibration
5 - Slowly and repeatedly squeeze the hand aspirator to push clean air through the carbon filter and past
the PID sensor. The ‘ZERO LED’ will stop flashing after approximately 2 minutes.
6 – Now remove the equipment except for the calibration adaptor.
NB During the above process the 'mV' reading on the display must fall below 30 mV to pass the Calibration
stage.
An acceptable ZERO calibration level will result in the illumination of the ‘STATUS LED’.
An unacceptable ZERO calibration level will result in the illumination of the ‘FAULT LE’D.Should this occur
the TVOC will not proceed to the gas tolerance setting and return to normal operation the previous
calibration levels will be used. The Yellow ‘ZERO LED’ will be illuminated to indicate the failure to set a zero
calibration level.
Setting the gas concentration (The actual value that the calibration has been supplied at)
1 Now, again place the calibration magnet over the corner of the logo –
The GREEN or FAULT LED will be extinguished and the numeric
display will start to flash.
2 If the 0-10ppm range is selected you will be able to adjust between 1.0 ppm and 100ppm, for both
the 100 ppm and 1000 ppm ranges a concentration of between 1 and 200 ppm can be selected.
Leaving the magnet in position will cause the displayed number to cycle through the concentration,
remove the magnet when the displayed value matches that stated on the Calibration gas bottle.
Should you miss the desired reading quickly reapply the magnet to continue scrolling through the
numbers and remove the magnet at the desired figure.
The selected number will be recorded 5 seconds after the removal of the magnet when the display
stops flashing. The display will initially show 90, should this be the desired figure remove the magnet
straight away.
Setting the SPAN
1 Connect the SPAN gas to the calibration connector then the connector to the adaptor. Do not over
tighten the Luer fitting; only a one quarter turn is required.
Switch on the gas supply and then place the magnet over the logo.
The SPAN LED will flash and TVOC will display the direct millivolt (mV) output from the PID sensor.
Now remove the magnet.
Page 19 of 32
Page 20
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
2 The SPAN LED will stop flashing after approximately 2 minutes.
3 Turn off the gas and disconnect the equipment
Calibration
NB during the span measurement process the mV reading of the span must be greater than the zero level to
pass the ZERO calibration stage.
An acceptable SPAN calibration level will result in the illumination of the green STATUS LED.
An unacceptable SPAN calibration level will result in the illumination of the red FAULT LED. Should this
occur the previous calibration levels will be used when the instrument returns to normal operation. The
Yellow span LED will be illuminated to indicate the failure to calibrate the span.
Now place the magnet over the logo corner to return to normal monitoring routine.
The GREEN LED will start flashing as TVOC starts its cycle with a lamp ON cycle.
If either the ZERO and/or SPAN levels were outside acceptable levels the RED LED will flash until an
acceptable calibration has been carried out.
Unacceptable calibration levels will not be adopted; the instrument will default to the previously stored
acceptable calibration factor.
NOTE:
* Always check the accuracy of your calibration by checking readings from TVOC when it is running
normally using the ZERO filter and SPAN gas.
* Accuracy of calibration is the responsibility of the person carrying out the calibration. If in doubt seek
advice.
* TVOC must be calibrated after lamp/cell cleaning or general maintenance.
NOTE: Causes of contamination and error in signal:
* Changes in air pressure when detecting ppm concentration
* Variation in oxygen, argon beyond ambient levels.
* Variation in ambient moisture content.
Page 20 of 32
Page 21
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Sensor cap
Locking screw
PID sensor
Sensor housing
Diagram 15
Diagram 16
Maintenance
Cleaning / replacing the lamp
TVOC has been designed to ensure servicing is quick and easy:
1. Before servicing TVOC,
disconnect the electrical power
supply.
2. Remove the locking screw from
the metal sensor cap using the
Allen key supplied within the
Calibration tool kit. (part number
A-849214)
Warning:
Failing to remove the locking
screw before unscrewing the
sensor cap will damage the
sensor holder.
3. Unscrew the sensor cap to
access the PID sensor.
4. Remove the PID Sensor by
withdrawing it from the sensor housing. Only light force is
required
Warning:
Do not twist the PID Sensor while
it is within the sensor housing.
Page 21 of 32
Page 22
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Electrode
PID lamp
Metal spring
Alumina powder
Cotton bud / Alumina
powder
Diagram 17
Diagram 18
Pellet
Maintenance
5. The PID sensor’s metal spring
cover can be pulled off using
moderate force. The electrode stack
and PID lamp can then be removed
by holding the PID sensor upside
down.
Warning:
Ensure the electrode stack and
PID Lamp falls onto a soft surface
such as a piece of tissue. This will
avoid damaging the parts as they
fall out and avoids finger contact
with the PID lamp window.
6. Clean the PID Lamp using
Alumina powder loaded onto a
cotton bud. Rub the cotton bud /
Alumina power on the PID lamp
window using a circular motion
until a light audible squeak can be
heard.
NOTE:
Contamination of the PIDLamp window can considerably reduce
the detection capability of the PID cell, even when the contamination
is not visible.
Cleaning of the lamp should be carried out on a regular
basis depending on the duty cycle of the PID lamp
and the environment.
stack
cover
Page 22 of 32
Page 23
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Electronic pins
Diagram 19
Maintenance
The humidity of the air and contaminants may affect the time required between servicing.
Warning:
7. The Electrode Stack should be inspected for visible signs of contamination, if contamination can be
seen the Stack it must be replaced. ( part number 1/EA-02)
Reassembly:
8. Place the clean PID Lamp into the PID Cell avoiding finger contact with the Window.
9. Refit the electrode stack with the electrical pin holes and electrode contacts facing down.
10. Refit the metal spring cover.
11. Ensure the electronic
pins of the PID sensor are
at the 12 o’clock position
before inserting the PID
sensor.
The PID sensor should
insert into the connectors
easily, if significant
resistance is felt, remove
the cell and check
alignment before
reinserting.
Warning:
Irreparable damage will be caused by forcing the PID into the sensor housing if not correctly aligned.
NOTE: Always calibrate TVOC after servicing is carried out.
Fuse rupture and replacement
TVOC has a 125 mA BASEEFA certified fuse to provide intrinsically safe protection when the unit is installed
in hazardous areas.
This fuse may rupture for example due to over voltage or a current surge, if this is the case replacement is
different depending upon the application being intrinsically safe (IS) or not.
For IS applications the unit must be inspected and have the fuse replaced by Ion Science Ltd or an Ion
Science Approved Service Centre. The intrinsically safe rating is not maintained if the fuse is simply
replaced.
For non-IS) applications the fuse may be replaced by an equivalent rated fuse and operation tested by
qualified engineering personnel.
Lamp Warranty Statement
Please note: The standard warranty period for a 10.6 PID lamp is one year, providing that the default duty
cycle for lamp is set to 1 minute of greater.
Page 23 of 32
Page 24
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
5/JHS-01
Diagram 20
Diagram 21
Diagram 22
Diagram 23
Diagram 24
System recommendations
TVOC is generally used to measure gas concentrations in the ambient atmosphere. The sensor is open to
the atmosphere and any gas that diffuses or moves under convection into the locality of the TVOC sensor
will be detected. Generally, the ambient environment in which TVOC is located is readily accessed, but there
are applications listed below that require special consideration.
Remote area monitoring
TVOC may be placed in a roof or other space that is hard to access but where the TVOC requires regular
bump tests. In this event, TVOC can be fitted with a Fixed calibration adaptor, Diagram 22. A flow adaptor is
also available for remote testing.
Fixed calibration
adaptor
Gas sample systems
It is sometimes necessary to pump or draw a gas sample past the TVOC. For this a “Flow Adaptor " can be
fitted. The flow adaptor has an inlet and an outlet port so that gas may be pushed or drawn across the
sensor. See diagram 2 above.
TVOC Flow Adaptor
Page 24 of 32
Page 25
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
System recommendations
For a gas sample system we have the following recommendations:
1. Where possible use the Flow Adaptor (as described above) that is sold by ISL. This has an integral
o-ring to seal the sensor housing and ports for connection of inlet and out let sample tubes. Please refer
to Spare parts section for part number.
2. The pressure difference of a pumped system relative to atmosphere should be minimised to avoid the
effects of gas law.
3. The maximum pressure that can be applied to the TVOC sensor housing is 300 mbar. However this is
not a recommended working pressure. Ideally working pressure should be +/- 30 mBar relative to
ambient pressure.
4. In line flow restrictions must be minimized. Flow restrictions cause differential pressures which can
directly affect the TVOC ready. If flow restrictions are unavoidable the flow rate should be lowered to
minimize the pressure effects, this however will increase the response time.
5. A flow between 250 and 500 ml/min is recommended for calibration. This will ensure a full response to
applied gas in a sensible time.
6. The working flow should be very similar to that use to calibrate the instrument or errors in output will
occur (see point 2).
7. The response time of the system is defined by the sensor response rate and sample flow rate will in
combination with the tube length and diameter as well as any dead volumes.
Page 25 of 32
Page 26
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
USA Office
Ion Science LLC
4153 Bluebonnet Drive
Stafford
TX 77477
USA
Standard Warranty can be extended to up to 2 years on the TVOC when registering your
instrument via our website: www.ionscience.com/instrument-registration
To receive your Extended Warranty, you need to register within one month of purchase (Terms
and Conditions apply). You will then receive a confirmation email that your Extended Warranty
Period has been activated and processed.
Full details, along with a copy of our Warranty Statement can be found by visiting:
www.ionscience.com/instrument-registration
Service
Ion Science is pleased to offer a number of service options on our TVOC product range that
allow you to choose the instrument cover that best suits your needs.
At Ion Science we recommend that all of our gas detection instruments be returned for service
and factory calibration once every 12 months.
Contact Ion Science or your local distributor for service options in your area.
Find your local distributor by visiting: www.ionscience.com
Contact details
Page 26 of 32
Page 27
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Diagram 25
Diagram 26
Diagram 27
Diagnostics
Below are two conditions your TVOC can be in when it is working correctly:
The two conditions shown below are of the instrument in an error state with potential checks/cures
for these faults:
If an F1 error occurs when the instrument is first switched on there may not be an issue. The instrument
needs to be left on for a few cycles to see if the lamp strikes on its own. If an F1 error, as shown below, is
displayed when the instrument is still on screen after 5 minutes then look at the information below
If the F1 error persists for more than 5 minutes, firstly check the red cable between the sensor PCB and the
main PCB is securely connected. The red cable is sufficiently rigid to be gripped inside the case and pushed
into the sensor housing. Instruments wit serial numbers exceeding 10-1790 include an improved
connection and diagnostic fault F3 which indicates if this connection has failed.
Page 27 of 32
Page 28
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Diagram 28
Diagram 29
Diagram 30
Diagnostics
If the F1 diagnostic persists, replace the sensor electrode stack, part number 1/EA-02, re-power the
instrument and wait 5 minutes. If the F1 error continues, replace the sensor lamp, part number 833229.
Pellet and lamp spares are identified in.
If the instrument continues to exhibit the F1 failure after replacement of parts please contact your Ion
Science distributor.
The F2 error occurs when the lamp appears to be illuminated yet power is removed from the PID sensor,
TVOC checks the PID lamp status using a light sensor. Running TVOC with the sensor or the sensor cap
removed will result in an F2 error.
Please note: The F2 error is disabled when the duty cycle is set below 10 seconds.
A light guard is fitted within the PID sensor housing. This ensures the light sensor is unaffected by ambient
light.
IMPORTANT:
The F3 error occurs when the sensor is disconnected from the power supply. Instruments with serial
numbers earlier than 10-01790 will not indicate an F3 error if the cable. Instead, an F1 error will be
displayed instead and the diagnostics associated with F1 errors, as above, should be performed.
If an F3 error occurs then check the cable is properly connected to the main PCB. If not push it fully home.
F4 errors
The F4 error occurs when an incorrect jumper cable selection is used, see TVOC Set up.
0- 1,000 ppm / 0 – 2280 mg/m3 (1.0 resolution)
Non IS applications Input power 5 – 28 Vdc. 65 mA (0.5mm2 to 2.5mm2 C.S.A.)
4 – 20 mA power 8 – 35 Vdc. 22 mA (0.5mm2 to 2.5mm2 C.S.A.)
4-20mA loop must be externally powered
ATEX approval II 2G EEx ia IIC T4 (-20°C ≤Ta ≤+50°C)
IECEx approval Ex ia IIC T4 (-20°C Ta +50°C)
Certificate number Baseefa05ATEX0277
IS applications Entry parameters
Input power Ui = 18V Ii = 800mA Pi = 1.2W Ci = 0F Li = 0mH
4-20mA power Ui = 30V Ii = 200mA Pi = 1.2W Ci = 0F Li = 0mH
For information only please see ATEX certificate before installation
Zener Barriers: Required Ask your installation engineers for installation/application advice.
(Non sparking) approval II 3G Ex nA IIC T4 (-40oC < Ta < +50oC) (nA for zone 2 only)
Dimensions Weight
Height 188 mm (7.40”)Instrument 1.3kg (2.9 lb)
Width 126 mm (4.96”)Packed 1.47kg (3.2lb)
Depth 78 mm (3.07”)
TVOC output update: Variable 1 second to 10 minutes
Accuracy 0 to 100 ppm: +/- 1 ppm (whichever is greater) 100 to 1000 ppm: +/- 10% or +/- 10 ppm (whichever is greater)
Temperature Operation: -20 oC to +50 oC (-4 oF to 122 oF)
Humidity 0 – 95% RH (non-condensing)
EMC Screened cables are required to archive the light industrial immunity
levels.
NB: All specifications are against an isobutylene calibration at 20 °C, 90% RH and up to 100 ppm.
Page 30 of 32
Page 31
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Manual Version
Amendment
Date updated
Instrument
Firmware
PC
Software
TVOC Manual
V2.1
Diagram 5.2.1 updated on
page 11
31/3/08
V3.09
N/A
TVOC Manual
V2.2
Diagram updated on page
15
21/4/08
V3.09
N/A
TVOC Manual
V2.3
Declaration on conformity
change only
27/11/08
V3.09
N/A
TVOC Manual
V2.4
Log added to back of manual
09/01/09
V3.09
N/A
TVOC Manual
V2.5
Diagnostics added and P&F
barrier changed
20/02/09
V3.09
NA
TVOC Manual
V2.6
TVOC firmware modified to
allow customers to calibrate
using an alternative
concentration. Aldo
improved the description of
the F2 fault condition. (P6,
P17, P24)
22/07/09
V3.10
N/A
TVOC Manual
V2.7
TVOC ATEX certification to
allow zone 2 installation
without safety barriers. (P12,
P13, P26)
1.09.09
V3.10
N/A
TVOC Manual
V2.9
Inclusion of Cable gland in
kit, so addition of details
regarding cable gland
selection and use
Change to EMC details as
result of testing
23.3.10
V3.10
N/A
TVOC Manual
V3.0
Warranty Cover updated
Declaration of Conformity
updated
Contents page updated
Instrument Warranty and
Service updated
22/07/10
V3.10
N/A
TVOC Manual
V3.1
Addition of fault output level
on page 9, Calibration fault
LED indication changed on
Page 20, 22 ,23
07/08/10
V3.25
N/A
TVOC Manual
V3.1
Addition of F3 and F4 fault to
diagnostics section pages
27, 28, & 29. Additional
notes in TVOC setup,
selector pin section page 11
08/09/10
V3.25
N/A
TVOC Manual
V3.1
Addition wire C.S.A areas
accepted on pages 14 & 30
Mounting diagram 5.2.1 on
page changed to general
assembly drawing
06/10/10
V3.25
N/A
TVOC Manual
V3.2
Page 9. Text amended to
emphasise when TVOC is
un-calibrated.
06/11/10
V3.25
N/A
TVOC Manual
V3.4
Page 14. Barrier part
numbers removed
20/12/10
V3.25
N/A
Manual log
Page 31 of 32
Page 32
TVOC MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
TVOC Manual
V3.5
Barrier details removed for
technical specification, notes
added that 4-20mA requires
external power
14/01/2011
V3.25
N/A
TVOC Manual
V3.6
Pages 3&4 updated to
correct Quality Management
System
Page 5 Responsibility of use
updated an legal notice
added
Page 18 Setting the gas
tolerance changed to ‘0.1 –
20.0’ to reflect instrument
function.
Page 24 & 25 System
Recommendations added.
Page 26 contact details
moved here
Page 28 details of Sensor
Light Guard added.
Page 29 Parts notes
updated.
Pages 7,11,13 & 14 ATEX
comments changed.
23/03/11
V3.27
N/A
TVOC Manual
V3.7
Page 30 – Accuracy and
Linearity specifications
updated
15/03/12
V3.27
N/A
TVOC Manual
V3.8
IS wiring diagrams and
photos updated.
24/05/12
V.3.27
N/A
TVOC Manual
V3.9
Slight modification to
installation instructions.
Page 11, Point 7
Page 30, IP clarification
26/05/12
V3.27
N/A
TVOC Manual
V4.0
Page 30, Accuracy
specification only
29/10/2012
V3.27
N/A
TVOC Manual
V4.1
Manual format and layout
updated
07/01/2013
V3.27
N/A
TVOC Manual
V4.2
Change to duty cycle. Pages
9, 10, 16 and 28
28/02/2013
V4.09
N/A
V4.3
Firmware now suitable for
old and new product
09/07/2013
V4.11
N/A
V4.4
Certification updated Page 3
and various minor updates
made through out to text,
capitalisation, italics etc…
15/05/2014
V4.11
N/A
Manual log
Page 32 of 32
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.