SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Declaration of conformity
Name und Adresse des Herstellers / Manufacture´s name and address /
ISM Deutschland GmbH, Laubach 30,
Die ISM Deutschland GmbH bescheinigt die Konformität für das Produkt the product / ISM Deutschland
GmbH dérclare la conformite du produit
Bezeichnung / Product name :
SF6-Raumluftmonitor / SF6-Roommonitor
Typ / Type / Type:
SF6 AREACHECK P2
Mit den folgenden Bestimmungen / with applicable regulations / avec les directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG
EMC Drective 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
Prüfschärfe /Level / Niveau = 2
EN 61000-4-4: 1995 / IEC (CEI) 1000
Prüfschärfe /Level / Niveau = 3
EN 50081-1 : 1992 / EN 55011 : 1991 / CISPR11
Gruppe / group / groupe = 1, Klasse / Class / Classe = B
Datum / Date / Date
01.08.2005
Unterschrift / Signature / Signatur
Hr. Becker – Technical Director
2 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Contents
Declaration of conformity ............................................................................................................................... 2
Auto-zero at startup ........................................................................................................................................ 7
Lamp test and manual reset .......................................................................................................................... 8
Configuration via the PC interface ................................................................................................................. 9
Connecting to the host computer ................................................................................................................... 9
User parameters ............................................................................................................................................ 9
Changing the sensor unit ............................................................................................................................. 10
Sensor Type ................................................................................................................................................. 11
User Interface ............................................................................................................................................... 11
Technical data ................................................................................................................................................ 12
Service ......................................................................................................................................................... 14
SF6 AREACHECK P2 is designed for indoor wall mounting. Since SF6 tends to sink towards the ground
in undisturbed air, we suggest to place it approx. 50 … 60 cm above ground level.
4 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
RS 485 daisy chain PIN B
RS 485 daisy chain PIN A
RS 485 daisy chain PIN GND
Alarm Relay PIN “Normally Open”
Alarm Relay PIN “Common”
Alarm Relay PIN “Normally Closed”
RS 232 Connector
Supply Voltage “EARTH”
Supply Voltage “L1”
Supply Voltage “N”
Installation
Wiring
Mains power connects to the orange clamp block, according to the designation printed on the PCB
near each clamp.
For an optional audible and/or visible alarm device, a voltage free switchover relay output is supplied.
It is available on the blue 3 position clamp block in the middle. Contacts are, from left to right: N.O. /
COM / N.C.
During normal operation, COM connects to N.O. During an alarm or error condition, COM disconnects
from N.O. and switches over to N.C. The same applies in case of power failure, and whenever the
instrument is taken into setup mode via the host computer interface via the RS232 connector.
The leftmost 6 position clamp block is provided for network operation. Physically, the network follows
RS485 standards and hence has three wires named A, B, and GND. Their positions are marked on the
PCB next to the clamps. They must be wired to the corresponding pins on the SF6
NETWORKCONTROLLER P2NC. Double clamps are provided for each wire for easy “daisy chaining” of
several instrument.
5 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
GND
A
B
Installation
Network diagram
Wiring details
6 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Operation
Auto-zero at startup
To precisely determine SF6 concentration, SF6 AREACHECK P2 requires a reference value representing
clean air. This reference value is taken and stored whenever the instrument is powered up or reset. You
should therefore make sure that power up takes place under clean air conditions.
A fresh reference value is also taken after manual reset via the key switch. Therefore, after each
manual reset, the instrument will temporarily only respond to SF6 concentrations exceeding the level
that was present at the time of manual reset. This will avoid the next measuring cycle to trigger an
undesired, new alarm. The reference value will auto-adapt to falling SF6 concentrations, so after once
meeting a clean air condition, SF6 AREACHECK P2 will automatically return to full sensitivity.
Since a reset is also performed when the instrument is taken into setup mode via the host computer
interface, the above also applies in this case.
After Power up or reset, the first measuring cycle starts immediately, beginning with the fan run time.
Measuring interval
Since SF6 AREACHECK P2 is designed to determine slowly changing concentrations, it samples the air
at a fixed time interval rather than measuring continuously. This feature minimizes the amount of air
per time that passes the instrument’s intake system, thus extending its service life.
Via the measuring time interval, it is possible to perfectly adapt the instrument to the actual
application. For monitoring generally unattended spaces, a long time interval should be selected, while
in the working area a shorter interval is suggested.
SF6 AREACHECK P2 comes with a factory set measuring interval of 30 minutes, which is the suggested
starting point for most applications.
Indicator lights
6 LEDs on the instrument’s front panel represent its state as follows:
Placement of LEDs:
RUN
This light acts as a life sign to confirm the SF6 AREACHECK P2 is
in operation. As long as no internal
error condition is met, it will blink at all times. During the actual SF6
measurement (normally, this phase
is very short), it will blink at a faster rate.
Should this lamp stop blinking, an internal error has occurred.
ALARM
This light will be permanently on if the last measuring cycle has determined a SF6 concentration above
the selected alarm threshold.
In case of an internal error state, it will blink quickly.
The instrument’s internal audible alarm works in parallel with this lamp.
500 / 1000 / 1500 / 2000 [ppm]
These 4 yellow lamps reflect the last sampling period’s result in terms of alarm levels as designated.
On instruments equipped with the optional oxygen sensor, these lights have the following additional
functions:
• The two leftmost LEDs (500+1000ppm) will blink when an oxygen concentration below the
lower alarm threshold of 16% is present.
• The two rightmost LEDs (1500+2000ppm) will blink when an oxygen concentration above
the upper alarm threshold of 25% is present.
7 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Operation
Lamp test and manual reset
When the key switch is turned to the spring return "TEST" position, all lamps, the internal audible
alarm and the optional external alarm device will turn on until the key switch is released.
When released, a pending alarm condition is cleared.
NOTE: An error condition will NOT be cleared!
When the key switch is turned and released twice within 2 seconds, a system reset is performed
which also clears any pending alarm and error conditions, and logs a fresh reference value.
8 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Configuration via the PC interface
Connecting to the host computer
Parameter modification in stand-alone operation is
done exclusively through the host computer interface.
This has the clear advantage that any unintended or
unauthorised fiddling is discouraged.
The user parameters described in the following are
relevant for stand alone operation only. In network
operation, parameters received from the network
base unit will take precedence.
The host computer interface follows RS232
standards. A special interface cable connects to the 3
position plug, which is found between the two blue
clamp blocks.
The lamp next to the plug indicates the presence of valid RS232 levels from the host. It will be lit when the
host is on and correctly connected.
On the host computer, a standard terminal program is used to talk to the SF6 AREACHECK P2. A suitable
program called "Hyper Terminal" is a standard Microsoft Windows component and by default present in most
installations. Optionally, any other terminal software will work.
Set up the terminal program for the following parameters: 9600 baud, 8 data bits, no parity, 1 stop bit. This is
a standard configuration also known as "9600-8-N-1". Further, select "none" for the handshake protocol, and
"TTY" as emulation if available. The option "append CR" for received lines should be checked for a more
readable screen output. For your convenience, we will be pleased to send you a predefined configuration file
for "Hyper Terminal" which will perform all relevant settings for you.
Note:
While the PC interface is connected, the instrument will talk to the host computer only and no longer
communicate over the network. When present, an SF6 NETWORK CONTROLLER P2NC will see this
station as "OFFLINE" and issue an alarm accordingly.
User parameters
User adjustable parameters are the time interval between each two samples ("Scan period") and the alarm
level. To activate the setup, from the terminal first send a line feed, then the instruction "TtS", followed by
another carriage return. A choice menu with the following options will then be presented:
<1> Scan period
<2> Alarm level
<3> Exit and resume operation
Just type the corresponding number to invoke the desired option. Any other inputs will be discarded.
After selecting "<1> Scan period", enter the desired time interval in minutes, followed by a carriage return.
The allowed range is 1-99 minutes.
Upon selecting "<2> Alarm level", enter the number corresponding to the desired alarm level.
Option <3> will finish the setup and return the instrument to normal operation. Since it will now perform a
restart.
9 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Changing the sensor unit
For changing the sensor unit, carry out the following Steps:
• Power down the P2 (remove the fuse, observing the hints below)
• Remove screws
• Take off SmartSensor as indicated below
• Pull the connector off the SmartSensor
• Fit the replacement SmartSensor
• Power up the P2 (insert fuse observing the hints below)
Note:
To precisely determine SF6 concentration, SF6 AREACHECK P2 requires a reference value representing
clean air. This reference value is taken and stored whenever the instrument is powered up or reset.
You should therefore make sure that power up takes place under clean air conditions.
ATTENTION: The fuse (T1A) is at mains voltage. All works on it are to be carried out by qualified
personnel only, observing appropriate precautions!
10 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
O2 sensor
Functionality
The optionally available O2 sensor continuously measures the O2 content in the ambient air, and reports it to
the network controller if present.
An alarm condition on both the measuring station and network controller is raised whenever the oxygen level
is outside the allowed range of 16…25%.
Physical location
The O2 sensor is part of the SmartSensor, it is found protruding downwards from it.
Sensor Type
The O2 sensor is of galvanic cell type. It exhibits low cross sensitivity to CO2 and other gases, good stability
and linearity over its range.
Sensor Lifetime
With a reasonable safety margin, sensor life is fixed at one year, no matter if shelfed or used – this is due to
the nature of galvanic cells, in which the chemical reaction goes on at all times as long as oxygen is present.
When lifetime is over, the SmartSensor must be replaced against a fresh, readily calibrated one.
User Interface
When a low oxygen condition is encountered, the measuring station will issue an alarm, involving the built in
alarm buzzer, the red alarm lamp, and the alarm relay. The low oxygen condition will be reflected by the
yellow "500" and "1000" ppm lamps flashing.
In case of a high oxygen condition, the same applies, except that the "1500" and "2000" ppm lamps
will flash.
On the SF6 NETWORK CONTROLLER P2NC, the O2 status will be marked as "OK", "HI", or "LO",
respectively, for each measuring station.
Safety
Should the O2 sensor fail, be missing or spent, a low oxygen alarm will be issued.
Standard Warranty can be extended to up to 2 years on the SF6 P1 & P1:p when registering
your instrument via our website: www.ionscience.com/instrument-registration
To receive your Extended Warranty, you need to register within one month of purchase (Terms
and Conditions apply). You will then receive a confirmation email that your Extended Warranty
Period has been activated and processed.
Full details, along with a copy of our Warranty Statement can be found by visiting:
www.ionscience.com/instrument-registration
Service
Ion Science is pleased to offer a number of service options on our SF6 P1 & P1:p product range
that allow you to choose the instrument cover that best suits your needs.
At Ion Science we recommend that all of our gas detection instruments be returned for service
and factory calibration once every 12 months.
Contact Ion Science or your local distributor for service options in your area.
Find your local distributor by visiting: www.ionscience.com
Contact details
14 of 15
SF6 P1 & P1:p MANUAL Ion Science Ltd
Unrivalled Detection. www.ionscience.com
Manual Version
Amendment
Date updated
Instrument Firmware
PC Software
SF6 P2 V1.1
Layout and font update
28/01/13
Manual log
15 of 15