Invacare CG8000 User Manual

Invacare 8000 Alternating Pressure Mattress
Assembly, Installation and Operating Instructions
Model No. CG8000
NOTE: The above model number contains one (1) Alternating Pressure Mattress and one (1) Power Unit.
NOTE: Check ALL parts for shipping damage. In case of shipping damage, DO NOT use. Contact Carrier/Dealer for further instruction.
SAFETY SUMMARY
To ensure the safe installation of the Invacare 8000 Alternating Pressure Mattress and Power Unit, these instructions MUST be followed:
WARNING
DO NOT use the mattress near an open flame or while smoking. Inspect power cord, plug and air hoses of the Invacare 8000 Alternating
Pressure Mattress/Power Unit for breaks, tears, or kinks. If damage is found, DO NOT use.
DO NOT, under any circumstances, cut or remove the round grounding prong from any plug used with or for Invacare products. Devices are equipped with three-prong (grounding) plugs for protection against possible shock hazards. Where a two-prong wall receptacle is encountered, it is the personal responsibility and obligation of the customer to contact a qualified electrician and have the two-prong receptacle replaced with a properly grounded three-prong wall receptacle in accordance with the National Electrical Code. If you must use an extension cord, use ONLY a three-wire extension cord having the same or higher electrical rating as the device being connected.
DO NOT place hot objects (for example: heating pads, electric blankets, etc.) on top of the mattress. This may weaken the mattress material.
Use extreme care when handling sharp objects such as needles near the mattress.
Do NOT inflate mattress of an OCCUPIED bed. The Alternating Pressure Mattress has a weight limitation of 250 lbs. (114 kg.). The Invacare 8000 APM MUST be installed on medical bed frames or medical-
grade home care beds with side rails. After placing the patient on the mattress, verify the side rails provide adequate protection for the patient. The side rails MUST be in the raised position, whenever a patient is on the bed.
INVACARE 8000 PRESSURE MATTRESS Part No. 1078996 Rev. B
1
INSTALLING THE INVACARE 8000 (FIGURES 1 AND 2)
NOTE: Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet.
1. Do one (1) of the following: A. Place the Invacare 8000 Power Unit on
a flat surface near the foot end of the bed.
Hooks
(2 Each)
Power Unit
Power Unit
Air Hose Quick
Disconnect
B. Open the hooks on the back of the Invacare 8000 Power Unit and install on the outside of the bed frame at the foot end of the bed as shown in FIGURE 1.
2. Remove existing mattress from bed.
Foot
End
Mattress Air
Hose Quick
Disconnect
3. Place Invacare 8000 mattress on top of bed with the cushions facing up and air hoses at the foot end of the bed.
4. Insert mattress air hose quick disconnects
FIGURE 1- INSTALLING
THE INVACARE 8000
Outside Bed Frame
into the power unit air hose quick disconnects as shown in FIGURE 1.
5. Slide the transport mat inbetween the cushions and the mattress base as shown in FIGURE 2.
NOTE: The transport mat will ensure the patient has support should the cushions deflate.
6. Inflate the mattress. Refer to INFLATING THE MATTRESS in this instruction sheet.
INFLATING THE MATTRESS (FIGURE 2)
NOTE: Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet. NOTE: Before inflating, ensure that ALL cushion fittings, both air hose quick disconnects and both
CPR plugs are secured, otherwise, inflation will not occur.
1. Plug the power cord on the power unit into wall outlet.
2. Turn the power switch to the ON position.
NOTE: The green light on the power unit will be illuminated when in the ON position.
3. Turn the knob CLOCKWISE on the power unit to the +MAX setting.
NOTE: It will take approximately 30 - 45 minutes for the mattress to be fully inflated at the maximum p.s.i.
NOTE: Keep the hose quick disconnects attached to the mattress and power unit, otherwise, deflation will occur.
4. Install the therapy pad to the mattress. Refer to INSTALLING/REMOVING THERAPY PAD in this instruction sheet.
INVACARE 8000 PRESSURE MATTRESS Part No. 1078996 Rev. B
2
Power
Unit
Power Switch
Cushion
-MIN
Setting
Cushion
Fittings
Transport Mat
Power
Unit
Mattress Air Hose Quick Disconnect
Knob
+MAX
Setting
Power Unit Air
Hose Quick
Disconnect
CPR Plugs
FIGURE 2- INFLATING THE MATTRESS
INSTALLING/REMOVING THERAPY PAD (FIGURE 3)
INSTALLING
1. Position therapy pad over mattress.
2. Insert one (1) of the ten (10) straps into open latch as shown in FIGURE 3.
3. Pull strap until therapy pad fits tight around mattress.
4. Push down on latch to secure.
5. Repeat STEPS 2 - 4 for remaining nine (9) straps.
6. Secure mattress base to the bed frame. Refer to SECURING MATTRESS BASE TO BED FRAME in this instruction sheet.
REMOVING
1. Remove patient from bed.
2. Pull up on one (1) of the ten (10) latches that secure the therapy pad to the mattress.
Strap
Latch
FIGURE 3 - INSTALLING/REMOVING
THERAPY PAD
3. Remove strap from latch.
4. Repeat STEPS 2 - 3 for remaining nine (9) latches.
5. Remove therapy pad from mattress.
Part No. 1078996 Rev. B INVACARE 8000 PRESSURE MATTRESS
3