Invacare APM User Manual

Invacare
®
Alternating Pressure Mattress
Invacare
Alternating Pressure Mattress APM
www.invacare.com
USA
One Invacare Way Elyria, Ohio 44035-4190 (800) 333-6900
Canada
5970 Chedworth Way Mississauga, Ontario L5R 3T9 Canada (800) 668-5324
© 2001 Invacare Corporation
Form No. 01-011
Invacare is a registered trademark and Yes, you can. is a trademark of Invacare Corporation. Specifications are subject to change without notification.
T
he Invacare®APM is a full-depth (7") powered air mattress replacement system which reduces interface pressure for the treatment of Stage I through Stage IV pressure ulcers.
With either 5-minute alternating pressure cycles or static operating mode, the caregiver can optimize the patient’s therapy and comfort level through the APM’s 22 adjacent air cells. The APM’s powerful, lightweight, and easy-to-transport power unit inflates the mattress in less than 12 minutes. Caregiver-friendly controls and several valuable user functions — including Auto­Firm, CPR deflate, simplified patient weight setting, and automatic back-up flotation in the event of a power failure — further enhance the APM’s advanced therapeutic performance.
Alternating Mode
In alternating mode, the Invacare APM’s 22 adjacent air cells alternately inflate and deflate at 5-minutes intervals, which redistributes the pressure against the skin to promote capillary circulation.
FEATURES
Alternating and Static Flotation modes
12-hour Automatic Backup flotation in the event
of power failure
Single point, rapid CPR deflate for 20-second
mattress deflation
Easy-to-understand, caregiver-friendly, water-
resistant membrane control panel
Auto-Firm control inflates mattress to firm, with
30-minute timer for nursing procedures
Quick mattress inflate fills mattress in under 12
minutes
Specifications
Model no. BB9653000
Electrical
Power 115vac, 50/60Hz, 2A
Dimensions
Mattress 82" L x 35" W x 7" H Power unit 20.5" W x 10.8" H x 5.6" D
Weight
Mattress 27 lb.
Power unit 23 lb. Weight capacity 350 lb. Safety code approval UL544, CA106, CA117 (cover)
Limited warranty
Power unit 1 year
Mattress 6 months
CAPILLARY CLOSURE - 32 mmHg
13
20
24
(6.2)
(7.4)
(8.9)
Average pressure per pressure point for all subjects over 10 cycle
( ) Standard deviation * 30º laterally inclined - all readings ** Head raised 50º - all readings
Testing completed in an independent test lab
Testing conducted on ten (10) subjects of various heights and weights
Scapula
Sacral Prominence
Trochanter *
23
(10.0)
Coccyx **
User-Friendly Controls
Power Button with LED indicator informs the
operator if the unit is in standby or operating mode.
Weight select increases or decreases the
mattress pressure setting to accommodate patient weight
Alternating pressure mode provides an air
pressure cycle every 5 minutes; static flotation mode cancels alternating pressure mode for those patients who can’t tolerate alternating pressure
Auto-Firm with LED indicator rapidly inflates the
mattress to maximum firmness to facilitate patient handling. Auto-Firm mode automatically cancels if left unattended for 30 minutes
Pressure Relief
Loading...