Invacare®Alegio™NG
enBed
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,thismanualMUSTbereadandsaved
forfuturereference.

©2023InvacareCorporation
Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibitedwithoutpriorwritten
permissionfromInvacare.Trademarksareidentiedby™and®.AlltrademarksareownedbyorlicensedtoInvacare
Corporationoritssubsidiariesunlessotherwisenoted.

Contents
1General.........................................4
1.1Introduction..................................4
1.1.1SymbolsinthisDocument.....................4
1.2IntendedUse.................................4
1.2.1IntendedOperator...........................4
1.3ServiceLife...................................5
1.4Compliance...................................5
1.4.1Product-specicstandards.....................5
1.5WarrantyInformation...........................5
1.6LimitationofLiability............................5
2Safety..........................................6
2.1Generalsafetyinformation.......................6
2.2Mattresses...................................7
2.3Safetyinformationonelectromagneticinterference.....7
2.4Labelsandsymbolsontheproduct.................8
2.4.1Identicationlabel...........................8
2.4.2OtherLabels...............................8
3Setup...........................................10
3.1GeneralSafetyInformation.......................10
3.2Mainpartsofthebed...........................10
3.3Assemblyofthebed............................10
3.4ControlBox...................................11
3.5Wiring......................................12
3.6Installingbedends.............................12
3.6.1Bedendbrackets............................12
3.6.2BedendswithU-bracketstyletting.............13
3.6.3BedendswithSannestyletting................13
3.6.4SabrineandCamilabedends...................13
3.7Installingsiderails..............................13
3.7.1VersoIIsiderail.............................13
3.7.2Scala2siderail.............................14
3.7.3BrittVandLinesiderail......................14
3.7.4Lisasiderail................................15
3.8Installingthemattresssupportextension.............16
3.9InstallRastox.................................16
3.10Dismantlethebed.............................16
4Operatingthebed.................................18
4.1GeneralSafetyInformation.......................18
4.1.1Mattressretainers...........................18
4.2Handcontrol.................................18
4.3Lockingfunction...............................18
4.4Operatingsiderails.............................18
4.4.1VersoIIsiderails............................19
4.4.2OperatingScala2siderail.....................19
4.4.3BrittVandLinesiderails......................19
4.4.4Lisasiderail................................20
4.5Castorsandbrakes.............................20
4.5.1Castorbrake...............................20
4.5.2Centralbrakesystem.........................20
4.6Liftingpole...................................20
4.6.1Placetheliftingpole.........................20
4.6.2Adjustingthehandleheight....................21
4.7Adjustingthelegsection.........................21
4.8Emergencypositions............................21
4.8.1Emergencyreleaseofamattresssupportsection....21
4.9TransportationandStorage.......................21
5Options.........................................22
5.1Listofavailableoptions..........................22
5.2Transporttrolley...............................22
6Maintenance.....................................25
6.1Generalmaintenanceinformation..................25
6.2Dailyinspections...............................25
6.3Generalservicinginformation.....................25
6.4Checklist-Maintenance.........................25
6.5Inspectionafterrelocation-Preparefornewuser......25
6.6Checklist-Afterrelocation.......................25
6.7CleaningandDisinfection........................26
6.7.1Generalsafetyinformation.....................26
6.7.2CleaningIntervals............................26
6.7.3CleaningbyHand............................26
6.7.4DisinfectionInstructions.......................26
6.8Lubrication...................................26
7AfterUse........................................27
7.1Disposal.....................................27
7.2Reconditioning................................27
8Troubleshooting...................................28
8.1Troubleshootingelectricalsystem...................28
9T echnicalData....................................29
9.1Materials....................................29
9.2Dimensions...................................29
9.3Weights.....................................29
9.4MattressDimensions............................29
9.5SideRails....................................30
9.6Environmentalconditions........................30
9.7Electricalsystem...............................30
10Electromagneticcompatibility(EMC)..................31
10.1GeneralEMCinformation.......................31
10.2Electromagneticemission........................31
10.3ElectromagneticImmunity.......................31
10.3.1TestspecicationsforimmunitytoRFwireless
communicationsequipment...................32

Invacare®Alegio™NG
1General
1.1Introduction
Thisusermanualcontainsimportantinformationaboutthe
handlingoftheproduct.Toensuresafetywhenusingthe
product,readtheusermanualcarefullyandfollowthe
safetyinstructions.
Onlyusethisproductifyouhavereadandunderstood
thismanual.Seekadditionaladvicefromahealthcare
professionalwhoisfamiliarwithyourmedicalconditionand
clarifyanyquestionsregardingthecorrectuseandnecessary
adjustmentwiththehealthcareprofessional.
Notethattheremaybesectionsinthisdocument,whichare
notrelevanttoyourproduct,sincethisdocumentapplies
toallavailablemodels(onthedateofprinting).Ifnot
otherwisestated,eachsectioninthisdocumentreferstoall
modelsoftheproduct.
Themodelsandcongurationsavailableinyourcountrycan
befoundinthecountry-specicsalesdocuments.
Invacarereservestherighttoalterproductspecications
withoutfurthernotice.
Beforereadingthisdocument,makesureyouhavethe
latestversion.Y oundthelatestversionasaPDFonthe
Invacarewebsite.
Ifyoundthatthefontsizeintheprinteddocumentis
difculttoread,youcandownloadthePDFversionfromthe
website.ThePDFcanthenbescaledonscreentoafont
sizethatismorecomfortableforyou.
Formoreinformationabouttheproduct,forexampleproduct
safetynoticesandproductrecalls,contactyourInvacare
distributor .Seeaddressesattheendofthisdocument.
Incaseofaseriousincidentwiththeproduct,youshould
informthemanufacturerandthecompetentauthorityin
yourcountry.
1.1.1SymbolsinthisDocument
Symbolsandsignalwordsareusedinthisdocumentand
applytohazardsorunsafepracticeswhichcouldresultin
personalinjuryorpropertydamage.Seetheinformation
belowfordenitionsofthesignalwords.
WARNING
Indicatesahazardoussituationthatcouldresult
inseriousinjuryordeathifitisnotavoided.
CAUTION
Indicatesahazardoussituationthatcouldresult
inminororslightinjuryifitisnotavoided.
NOTICE
Indicatesahazardoussituationthatcouldresult
indamagetopropertyifitisnotavoided.
TipsandRecommendations
Givesusefultips,recommendationsand
informationforefcient,trouble-freeuse.
OtherSymbols
(Notapplicableforallmanuals)
UKResponsiblePerson
Indicatesifaproductisnotmanufacturedin
theUK.
Triman
Indicatesrecyclingandsortingrules(only
relevantforFrance).
1.2IntendedUse
Themedicalbediselectricallyoperated,adjustableinheight
andfeaturesaprolingsupportsurfaceintendedtobeused
incombinationwithamattress.Medicalbedsaredesigned
forsupportofthepatientandusedforthemitigation,
relief,andrecoveryfromcertainmedicalconditions.The
medicalbedalsosupportsandeasesthecaregivers’working
environment.
TheMedicalbedisintendedtobeusedon:
•Applicationenvironment3;Long-termcareina
medicalareawheremedicalsupervisionisrequired
andmonitoringisprovidedifnecessaryandmedical
electricalequipmentusedinmedicalproceduresmay
beprovidedtohelpmaintainorimprovethecondition
ofthepatient.
•Applicationenvironment4;Careprovidedinadomestic
areawheremedicalelectricalequipmentisusedto
alleviateorcompensateforaninjury,disabilityor
disease.
IntendedUsers
Apersonwithamedicalconditionthatneedstobecared
forinbed.
Indications
Thebedisintendedforadultusers,havingaphysicalsize
equaltoormorethan146cm,aweightequaltoormore
than40kgandabodymassindex(BMI)equaltoormore
than17.
Thebedisintendedforindooruseonly .
Maximumuserweightis135kgandsafeworkingloadis
170kg.
Contra-indications
Thebedisnotintendedfortransportationofusers.Itis
mobilewithinaroomwithauserinit.Castorsarelockable.
Thebedisnotintendedforpersonswithapsychosocial
disability.
WARNING!
Anyotherorincorrectusecouldleadtohazardous
situations.
Invacareacceptsnoliabilityforanyuse,change
orassemblyoftheproduct,otherthanstatedin
thisusermanual.
1.2.1IntendedOperator
Ahealthcareprofessionalorprivatepersonwhohasreceived
propertrainingistheintendedoperatorofthisproduct.
Thehandcontrolcanalsobeoperatedbytheoccupantof
thebed.
4
1563578-H

General
1.3ServiceLife
Theexpectedservicelifeofthisproductisveyearswhen
useddailyandinaccordancewiththesafetyinstructions,
maintenanceintervalsandcorrectuse,statedinthismanual.
Theeffectiveservicelifecanvaryaccordingtofrequencyand
intensityofuse.
1.4Compliance
Qualityisfundamentaltothecompany’soperation,working
withinthedisciplinesofISO13485.
ThisproductfeaturestheCEmark,incompliancewiththe
MedicalDeviceRegulation2017/745ClassI.
ThisproductfeaturestheUKCAmark,incompliancewith
PartIIUKMDR2002(asamended)ClassI.
Wearecontinuouslyworkingtowardsensuringthatthe
company’simpactontheenvironment,locallyandglobally,
isreducedtoaminimum.
WeonlyuseREACHcompliantmaterialsandcomponents.
WecomplywiththecurrentenvironmentallegislationsWEEE
andRoHS.
Forfurtherinformationaboutlocalstandardsandregulations,
contactyourlocalInvacarerepresentative.Seeaddresses
attheendofthisdocument.
1.5WarrantyInformation
Weprovideamanufacturer’swarrantyfortheproduct
inaccordancewithourGeneralTermsandConditionsof
Businessintherespectivecountries.
Warrantyclaimscanonlybemadethroughtheprovider
fromwhomtheproductwasobtained.
1.6LimitationofLiability
Invacareacceptsnoliabilityfordamagearisingfrom:
•Non-compliancewiththeusermanual
•Incorrectuse
•Naturalwearandtear
•Incorrectassemblyorset-upbythepurchaserorathird
party
•T echnicalmodications
•Unauthorisedmodicationsand/oruseofunsuitable
spareparts
1.4.1Product-specicstandards
TheproducthasbeentestedandconformstoIEC60601-2-52
(Particularrequirementsforthebasicsafetyandessential
performanceofmedicalbeds)andallrelatedstandards.
1563578-H5

Invacare®Alegio™NG
2Safety
2.1Generalsafetyinformation
Thissectionofthemanualcontainsgeneralsafety
informationaboutyourproduct.Forspecicsafety
information,refertotheappropriatesectionofthemanual
andprocedureswithinthatsection.
WARNING!
Riskofinjuryordamage
–Donotusethisproductoranyavailableoptional
equipmentwithoutrstcompletelyreading
andunderstandingtheseinstructionsandany
additionalinstructionalmaterialsuchasuser
manualsorinstructionsheetssuppliedwith
thisproductoroptionalequipment.Ifyouare
unabletounderstandthewarnings,cautionsor
instructions,contactahealthcareprofessional,
Invacareproviderorqualiedtechnicianbefore
attemptingtousethisproduct.
–Donotmakeanyunauthorizedalterationsor
modicationstotheproduct.
WARNING!
Riskofentrapment/suffocation
There’sariskofentrapment/suffocation
betweenmattresssupport,siderailandbedend
orbetweenmovingpartsandobjectsplaced
nearbythebed.
–Thebedmustnotbeusedbypersonsunder12
yearsofage,orbypersonswithabodysize
equivalenttoanaverage12yearsoldorsmaller .
–Thebed,incombinationwithsiderailsmust
notbeusedbypersonshavingaphysicalsize
lessthan146cm,aweightlessthan40kgora
bodymassindex(BMI)lessthan17.
–Duetomattresscompression,anincreased
riskmayoccurovertime.Periodicallymonitor
gapsbetweenthebed,mattressand/orside
rail.Replacemattressifthegapsmayleadto
entrapment.
WARNING!
Riskofslippingthroughtheopenings
Thebedfullsallrequirementsregarding
maximumdistances.However,itispossible
thatpersonswithsmallbodydimensionsslip
throughtheopeningsbetweenthesiderailsor
throughtheopeningbetweenthesiderailand
themattresssupport.
–Payspecialattention,ifthebedisusedforthe
careofpersonswithsmallbodydimensions.
WARNING!
Riskofinjuryordamage
Inappropriatehandlingofcables,incorrect
connectionsanduseofunauthorizedequipment
cancauseelectricalshockandproductfailure.
–Donotkink,shearorotherwisedamagethe
cablesoftheproduct.
–Makesurethatnocablesarejammedor
damaged,whentheproductisused.
–Disconnecttheplugfromthemainsbefore
movingthebed.
–Makesurethewiringiscorrectandconnections
areproper.
–Donotuseunauthorizedequipment.
WARNING!
Riskoftripping,entanglementorstrangulation
Improperroutingofcablesmaycausetripping,
entanglementorstrangulationhazard.
–Ensureallcablesareroutedandsecured
properly.
–Ensuretherearenoloopsofexcesscable
extendingawayfromtheproduct.
WARNING!
Riskofinjuryordamage
Ignitionsourcescancauseburnsorre.
–Bedmustbeplacedwithasafetyspacefrom
possibleignitionsources(heater ,replace,etc.)
–Thepatientandassistantsmustnotsmoke
whileusingoroperatingthebed.
WARNING!
Riskofinjuryordamage
Toavoidinjuryordamagewhenoperatingthe
product:
–Closesupervisionisnecessarywhentheproduct
isusednearchildrenorpets.
–Donotletchildrenplaywiththeproduct.
CAUTION!
–Forapersonenteringorexitingthebed,always
lowerthebedtoanappropriateheight.For
enteringorexitingthebed,thebackrestsection
ofthemattresssupportcanbeusedasan
assistance.Makesurethepatientislaiddown
atthecentreofthebackrestsection.When
thebackrestsectionisbeingraisedtheallowed
maximumweightforthebackrestsectionis
45%ofmaximumsafeworkingload.
–Alwayslowerthebedtothelowestposition
beforeleavingapersoninthebedunattended.
–Makesurethatthereisnothingunder ,over ,
ornearthebedthatcanobstructtheheight
adjustment,likeforexamplefurniture,liftsor
windowframes.
61563578-H

CAUTION!
Non-originalorwrongaccessoriesmayaffectthe
functionandsafetyoftheproduct.
–Onlyuseoriginalaccessoriesfortheproduct
inuse.
–Duetoregionaldifferences,refertoyour
localInvacarecatalogorwebsiteforavailable
accessoriesorcontactyourlocalInvacare
representative.Seeaddressesattheendof
thisdocument.
CAUTION!
Riskofinjuryordamage
Theproductcangethotwhenexposedtosunlight
orotherheatsources.
–Donotexposetheproducttodirectsunlightfor
prolongedperiods.
–Keeptheproductawayfromheatsources.
WARNING!
Riskoffalling
Theusercanfallovertheedgeandgetseriously
injured,iftheverticaldistanceAbetweenthe
topofthemattressandtheedgeoftheside
rail/bedend,istooshort.Seeimageabove.
–AlwayskeepaminimumdistanceAof22cm.
–Followthemaximummattressheightin
combinationwiththesiderailasstatedinthe
mattresstableinchapter9T echnicalData,
page29
.
Safety
2.2Mattresses
CAUTION!
Thereisariskofentrapmentofngersinthe
bedmovingparts.
–Payattentiontoyourngers.
NOTICE!
Thebedhasnoisolator(mainswitch).Ifthebed
needstobeelectricallydisconnected,ithastobe
unpluggedfromthemainssocket.
–Alwaysplacethebedinawaysothemains
socketiseasilyaccessibletounplugthebed.
NOTICE!
Accumulationoflint,dustandotherdirtcan
impairtheproduct.
–Keeptheproductclean.
WARNING!
Safetyaspectsregardingcombinationofside
railsandmattresses:
Togetthehighestpossiblesafetylevel,when
usingsiderailsonthebed,theminimumand
maximummeasuresformattresses,mustbe
respected.
–Forcorrectmattressmeasuresseemattress
tableinchapter9TechnicalData,page29.
2.3Safetyinformationon
electromagneticinterference
WARNING!
Riskofmalfunctionduetoelectromagnetic
interference
Electromagneticinterferencebetweenthis
productandotherelectricalequipmentcanoccur
anddisturbtheelectricaladjustmentfunctions
ofthisproduct.Toprevent,reduceoreliminate
suchelectromagneticinterference:
–Onlyuseoriginalcables,accessoriesandspare
parts,tonotincreaseelectromagneticemission
orreduceelectromagneticimmunityofthis
product.
–Donotuseportableradiofrequency(RF)
communicationsequipmentcloserthan30cm
toanypartofthisproduct(includingcables).
–Donotusethisproductnearactive
high-frequencysurgicalequipmentandthe
RFshieldedroomofasystemformagnetic
resonanceimaging,wheretheintensityof
electromagneticdisturbancesishigh.
–Ifdisturbancesoccur,increasethedistance
betweenthisproductandtheotherequipment
orswitchitoff.
–Refertothedetailedinformationandfollow
theguidanceinchapter10Electromagnetic
compatibility(EMC),page31
.
WARNING!
Riskofentrapmentand/orsuffocation
–Theusercouldgettrappedand/orsuffocate,if
thehorizontalspace,betweenthemattressside
andtheinsideofthesiderail,istoobig.Follow
theminimumwidth(andlength)ofmattresses
incombinationwithasiderail,asstatedin
themattresstableinchapter9T echnicalData,
page29.
–Beawarethatusingverythickorsoft
mattresses(lowdensity),oracombinationof
these,increasestherisk.
1563578-H
7

Invacare®Alegio™NG
YYYY-MM
Invacare Portugal, Lda
Rua Estrada Velha 949
4465-784 Leça do Balio, Portugal
= xxx kg
XXXXXXXXXXXXX
XXX XXX XX XX XX X
XXX XXX XX XX XX X
XXX XXX XX XX XX X
XXX XXX XX XX XX X
XXX XXX XX XX XX X
XX
XX
XX
XX
XX
XX
XXX
XXX
= XXX
WARNING!
Riskofmalfunctionduetoelectromagnetic
interference
Donotusethisbedadjacenttoorstackedwith
otherelectricalequipment,thanspeciedinthe
following,asitcouldresultinimproperoperation.
Ifsuchuseisnecessary,thebedandtheother
equipmentmustbecloselyobservedtoverify
thattheyareoperatingnormally.
ThisbedcanbeusedtogetherwithInvacare
approvedaccessoriesandmedicalelectrical
equipmentconnectedtotheheart(intracardially)
orbloodvessels(intravascularly)providedthat
followingpointsarerespected:
–Medicalelectricalequipmentshouldnotbe
xedonthebed’smetallicaccessoriessuchas
siderails,liftingpole,driprod,bedends,etc.
–Themedicalelectricalequipmentpowersupply
cordshouldbekeptclearoftheaccessoriesor
anyothermovingpartsofthebed.
TypeBappliedpart
WEEEconform
EuropeanConformity
UKConformityAssessed
Abbreviationsfortechnicaldata:
•Iin=IncomingCurrent
•Uin=IncomingVoltage
•Int.=Intermittence
•AC=AlternatingCurrent
•Max=maximum
•min=minutes
Formoreinformationabouttechnicaldata,referto9
TechnicalData,page29.
2.4Labelsandsymbolsontheproduct
2.4.1Identicationlabel
Theidenticationlabelisplacedontheframeofthebedand
containsthemainproductinformation,includingtechnical
data.
Serialnumber
Referencenumber
Manufacturer
Dateofmanufacture
2.4.2OtherLabels
Readtheusermanualbeforeusingthisproductandfollow
allinstructionsforsafetyanduse.
Denitionofmin.weight,min.heightand
min.bodymassindexofanadultuser .
SeeIntendedUse.
Refertouserdocumentationforthe
correctmattressmeasures.
See9TechnicalData,page29.
Totalweightoftheproductwiththe
Medicaldevice
maximumsafeworkingloadapplied
Temperaturelimit
Max.userweight
Humiditylimitation
Max.safeworkingload
Atmosphericpressurelimitation
CLASSIIequipment
81563578-H

Safety
Transportandstorageconditions
Operationconditions
Marksoutthelocationofconnectionformeansof
potentialequalization.
(Notpresentonallversionsofthisproduct)
Labelon3/4lengthsiderails
Indicatesthemaximumspacebetweensiderailandthe
headbedend.
Seethemountinginstructioninthismanualorthe
instructiondeliveredwiththesiderail.
1563578-H9

Invacare®Alegio™NG
3Setup
3.1GeneralSafetyInformation
WARNING!
Riskofinjuryordamage
Damagedpartsmayaffectthesafetyofthe
product.
–Checkallpartsforshippingdamagebeforeuse.
–Incaseofdamage,donotusetheproduct
andcontactyourInvacareproviderforfurther
instructions.
Mattresssupport,headsection
A
Mattressretainers,4pcs
B
Motor,mattresssupportbackrestsection
C
D
Mattresssupport,thigh/legsection
Liftmotor
E
Castor,4pcs
F
ControlBox
G
Sheararms
H
Mattresssupportframe
I
Handcontrol(notshowninimage)
WARNING!
Riskofinjuryordamagetoproperty
Theassemblyofthebedandinstallationof
accessoriesmustonlybeperformedbyaqualied
technicianorapersonwhohasreceivedproper
training.
–Followinstructionscarefully.Ifyouhaveany
questionregardingtheassembly,contactyour
Invacareprovider .
–Theelectricalequipmentofthebedmustnot
bedismantledorcombinedwithotherelectrical
equipment.
–Aftereachassembly,checkthatallttingsare
properlytightenedandthatallpartshavethe
correctfunction.
NOTICE!
–T opreventcondensation,thebedshould
notbeuseduntilithasreachedoperating
temperature.See9T echnicalData,page29
3.2Mainpartsofthebed
Siderailsandbedendsarenotdeliveredinthe
standardconguration.Formoreinformationabout
siderails,bedendsandotheraccessories,contact
yourInvacarerepresentative.
WARNING!
Riskofinjury
Theusercanfallovertheedgeandgetseriously
injuredifthebedisusedwithoutbedendsand
siderails
3.3Assemblyofthebed
Mattresssupport
1.
Placebaseandsheararmassemblyontheoorand
lockallcastors.
2.
a.Placethemattresssupportfootsectionattheendof
thebedwhichhastheliftingactuator .
b.AligntheplasticsnaplockttingsAonthemattress
supportwiththepivotpinsBonthesheararm.
c.PressdownrmlytoengagethepivotpinsBinthe
snaplockttingA.RotatethelockingpinConthe
snaplockttingtopreventdisengagementofthe
footsection.
101563578-H

Setup
3.
a.Placetheheadsectiononthesheararmsatthe
oppositeendofthebed.
b.Alignthesquareplasticguideswiththechannelon
theheadsection.
c.Liftthelowerendofthefootsectiontoalignthe
bracketswiththeframetubesoftheheadsection.
4.
8.
Attachthecablettingtotheholeofthedeckbracket
withtheclevispinprovidedandmakesuretheheadof
theclevispinisonthesamesideastherastoxdeck
brackets.
9.
a.Pushthefootsectiontowardstheheadsectionuntil
theframetubesaretouching.
b.Securetheheadandfootsectiononeachsidewith
thengerscrewA.
Ifthebedisprovidedwithawireactuatedfoot
section,followsteps5–10.
5.Unfastenthecablefromtheheadsection.
6.
Fixtheclevispinwiththehitchpin.
10.Lowerthekneepivotpointandensurethecableand
pinaremountedproperly.
3.4ControlBox
Thecontrolboxislatchedtotheliftmotor.
Liftthefootsectionandlaycableovertubeandguide
pulley.
7.Lowerthefootsectionandraisethekneepivotpoint.
1563578-H
ThecontrolboxAisprovidedwithalabelwithsymbols
showingwheretoconnectthemotorplugs:
•Thighsection’smotorB.
•LiftmotorC.
•BackrestmotorD.
•HandcontrolE.
•notusedF.
ThecontrolboxisequippedwithagreenLEDG
whichindicatesitisconnectedtomainspower.
WhenitisconnectedtomainspowerthegreenLED
Gison.
11

Invacare®Alegio™NG
3.5Wiring
NOTICE!
–Thecablesmustbemountedinsuchaway
thattheyarekeptclearofftheooranddo
notblockthecastors.
Inordertopreventthecablesfrombeingdamagedwhen
activatingthemotors,followtheinstructionsbelow.
1.
3.
a.Routetheheadsectionmotorcableovertheshear
armassemblycrosstubeandintothemotorpipepin
andhookonthemattresssupportcrosstube.
b.Plugcableintothemotorandsecurewithlockingclip.
2.
a.Routethethighsectionmotorcableovertheshear
armassemblycrosstube.
b.Plugcableintomotorandsecurewithlockingclip.
Routethepowercablethroughthesheararmassembly
hooks.
4.Routethehandcontrolcableunderthebaseframe.
3.6Installingbedends
Allenkey
3.6.1Bedendbrackets
Dependingonthebedendtobemounted,itmightbe
requiredtopre-mountorchangethebedendbrackets.
Mounting
1.Insertthebedendbracketsintothesidetubesofthe
mattresssupportplatform.
2.Fixthebedendbracketsonthesidetubeswiththe
screwsfrombelowandtightenrmly.
Dismounting
1.Removethescrewsandpullthebedendbracketsoutof
thesidetubes.
12
1563578-H

Setup
3.6.2BedendswithU-bracketstyletting
Mounting
1.LowerthebedendintotheU-brackettting.
2.Pressdownrmlyandensurethebedendisfullyseated
atthebottomoftheU-brackettting.
Dismounting
1.Holdthebedendrmlyoneachsideandpullup
verticallyonbothsidesatthesametime.
3.6.3BedendswithSannestyletting
CAUTION!
Riskofsqueezing
Ifthebedendisnotcorrectlylockeditcanfall
downandcauseinjurytouser .
–Makesurethelockingpinsareengaged.
Mounting
3.6.4SabrineandCamilabedends
Mounting
1.Fixthebedendbracketsonthebedendpanelwith
thescrews.
2.Insertthebedendbracketsintothesidetubesofthe
mattresssupportplatform.
3.ForSabrinebedends,xthebedendonthesidetubes
withthengerscrewsinsertedinthesecondhole.In
somecongurationsyouhavetoinsertthengerscrews
intherstholetoxthefootend.
ForCamilabedends,xthebedendonthesidetubes
withthescrews.
Dismounting
1.Unscrewthengerscrewsorscrews,thatxthebed
endonthesidetubes.
2.Pulloutthebedendpanelwiththebedendbracketsof
thesidetubesofthemattresssupportplatform.
1.Aligntheslotsinthebedendwiththetabsonthe
Sannetting.
2.Placethebedendontothetabs.
3.Pushdownrmlyandensurethatthebedendisfully
seatedinthetabs.
4.Rotatethelockingpinonbothsidestoallowitto
engagewiththebedend.
Dismounting
1.Pulloutthelockingpinandrotatetotemporarily
disengagefromthebedendonbothsides.
2.Holdthebedendrmlyoneachsideandpullup
verticallyonbothsidesatthesametime.
3.7Installingsiderails
WARNING!
Riskofpersonaldamage
Thereisariskofentrapmentorsqueezing,while
assemblingordisassemblingthesiderail.
–Payattentiontoyourngers.
–Followinstructionscarefully.
–Aftereachassembly,checkthatallttingsare
properlytightenedandthatthesiderailhas
thecorrectfunction.
Removable3/4lengthsiderails
WARNING!
Riskofentrapment
There’sariskofentrapmentorsuffocation
betweenmattresssupport,siderailandbedend.
–Whenusingremovablesiderails,alwaysmake
surethatthedistancebetweenthebedend
andthesiderail’shandleupperedgeisless
than6cmintheheadendandmorethan32
cminthefootendofthebed.
3.7.1VersoIIsiderail
Mounting
1563578-H13

Invacare®Alegio™NG
Themountingofthesiderailmustbedonewiththe
siderailinlockedposition.
1.Placebothbracketsofthesiderailontheframeofthe
bed,withthelockingbuttonAfacingtowardsthefoot
endofthebed.
2.Ensurethatthebracketneartheheadendofthebedis
placedencompassingtheliftingpolesupportasshown
indetailB.
3.ForbedsequippedwithbedendswithSannestyle
tting,itisbeforehandnecessarytotemporarily
disengagethelockingpinCfromtheheadbedend
toplacethebracket.
a.PullthelockingpinCandrotate1/4turntolockit
inthedisengagedposition.
b.Positionthebracketunderthelockingpinas
describedinstep2.
c.Rotatethelockingpinandallowittoengagewith
thebedend.
4.Afterthebracketsarepositionedcorrectlyontheframe
ofthebed,xthesiderailbyrmlytighteningthe
ngerscrewsD.
Dismounting
1.Untightenthengerscrewsandliftthesiderailofthe
bedframe.
3.7.2Scala2siderail
Mountingsiderail
1.
Placethesiderailoverthetopframewiththerelease
systemAattheheadendofthebed.
Theforklinksonthesiderailmustbemounted
accordingtotheinstructiononthesiderail.
2.
TightenthengerscrewsBtoattachthesiderailon
thetopframe.Makesureitissteadyandfastened.
Dismountingsiderail
1.LoosenthetwongerscrewsBandremovetheside
rail.
3.7.3BrittVandLinesiderail
WARNING!
Riskofentrapment/suffocation
Theusercangettrappedorfalloutofthebedif
thesiderailisnotcorrectlymountedordamaged.
–Makesurethatallglidingshoesarecorrectly
guidedintothebedendguideways.
–Makesurethatallsiderailbarsareabovethe
lockingpinandproperlylocked.
–Makesurethatthestrapsbetweenthebarsare
notdamagedorloose.
Mounting
1.Lifttheupperbarinoneendofthesiderail.The
releasebuttonsCmustfaceup/outwards.
2.PressinthelockingpinAintheendofthesiderail.
14
1563578-H

Setup
3.GuidethethreeglidingshoesB(intheendoftheside
railbars)upintothebedendguideways,untiltheyhave
allpassedthelockingpinD.
4.Repeatsteps2and3tomounttheothersideofthe
siderail.
Dismounting
1.Lowerthesiderail.
2.Liftthelowerbar,inoneend,tomakethelockingpin
Dvisibleandpressitinwithascrewdriver.
3.Lowerallthreebars,untiltheyareoutoftheguideways.
4.Repeatsteps2to3todismounttheotherendofthe
siderail.
3.7.4Lisasiderail
Installingthesiderail
Allenkey
WARNING!
Riskofentrapment/suffocation
Theusercangettrappedorfalloutofthebedif
thesiderailisnotcorrectlymountedordamaged.
–Makesurethatallglidingshoesarecorrectly
guidedintothebedendguideways.
–Makesurethatthelockingscrewandtab
areproperlyinstalledatthebottomofthe
guideways.
Attachingtheglidingsystemandinstallingthe
siderail
1.Raisethebedto1/3offullheight.
2.
Removethescrewsofthebedendbracketsonboth
sidesofonebedend.
3.
Assemblingtheglidingsystem
1.
PushthepawlAfromthebackside,throughtheglider
Bandmakesureitlockswithanaudibleclick.
2.Dothesameforallfourpawlsineachglider.
Loosenthescrewandtabatthebottomofthe
guidewaysonthebedendsandononesidepushthe
gliderhalfwayupintotheguidewayuntilitattaches
withanaudibleclick.
4.
Pushtheuppersiderailontothetwoupperpawls.
5.
Intheotherendofthesiderailbar ,pushthetwoupper
pawlsintotheendofthebar.
1563578-H15

Invacare®Alegio™NG
6.
Pushtheattachmentintotheguidewayuntilitattaches
withanaudibleclick(=properlyengagedinthelower
position).
7.Mountthelowersiderailbarinbothendsasinstep4.
8.Pushthesiderailinbothends,allthewayupuntilthe
upperbarlocksintheupperposition.
9.
4.Reinstallthefootbedendandxittothebracketswith
thetappingscrews.
NOTICE!
Onanextendedbed,thefootbedendmust
besecuredbyxingittothebracketswith
tappingscrews.
5.
Attachmattressdeckextendertothefootendofthe
mattressdeck.
3.9InstallRastox
(optional)
1.PressbottomofRastoxatthefootendofthebed.
Retightenthescrewandtabunderthelowersiderail
bar,ineachcornerofthebedends.
10.Reinsertandtightenthescrewsofthebedendbrackets
removedinstep2.
3.8Installingthemattresssupport
extension
(optional)
1.Removethefootbedend.
2.
Loosenthescrewsthatsecurethestandardbedend
bracketsandremove
3.Inserttheextendedlengthbedendbracketsintothe
frametubesandsecurewithscrews.
2.SqueezeupperendofRastoxtogether.
3.RotateRastoxbetweentabsandrelease.
4.Slidetheplasticcapandlockthebottomoftherastox.
3.10Dismantlethebed
1.Lockallcastorsandremoveallaccessories,endboards
andmattress.
2.Bringthebedtoitslowestpositionandadjustall
mattresssupportsectionstohorizontalposition.
3.Disconnectthepowercablefromthemainssocket.Roll
thecableontothecontrolboxandliftmotor.
Ifthebedisprovidedwithacableactuatedfoot
section,followstep4.Otherwiseproceedwithstep
5.
161563578-H

Setup
4.Unfastencablefromfootsectiondeckbracket,being
suretoretaintheclevispinandhitchpin.
5.Unscrewthethreadedknobsfromthemiddleofboth
sidesofthebedframe.
6.Pulltheheadsectionawayfromthefootsectionto
disengagebracketsandplasticglides.
7.Onthefootsection,rotatethelockingpinsandthe
snap-lockttings.Pullthefootsectionupandaway
fromthesheararmpivotpins.
1563578-H
17

Invacare®Alegio™NG
4Operatingthebed
4.1GeneralSafetyInformation
WARNING!
Riskofpersonalinjuryanddamagetoproperty .
–Thebedmustbeplacedsothattheheight
adjustmentisnotobstructedby,forexample,
liftsorfurniture.
–T akecarethatnobodypartsarebeingsqueezed
betweenxedparts(suchassiderails,bed
endsetc)andmovingparts.
–Thehandcontrolmustnotbeusedbychildren.
WARNING!
Riskofinjury
Ifthebedisusedbyconfusedorrestlessusers
oruserswithspasms:
–either ,whenthebedisequippedwitha
lockablehandcontrol,lockthehandcontrol
functions,
–ormakesurethatthehandcontrolisoutof
reachfortheuser .
4.1.1Mattressretainers
WARNING!
Riskofinjury
Withoutproperlyinstalledmattressretainers,the
mattresscanslipsidewaysandcausetheuserto
falloutofthebed.
–Neverusethebedwithoutmattressretainers.
–Alwaysmakesurethemattressretainersare
properlyinstalledandfacingupwards.
–Onlyusetheoriginalmattressretainers.
4.2Handcontrol
Thehandcontrolcanbeequippedwitheithertwoorfour
buttons.
Sittingposition
1.Up:pressleftsideofthebutton(p).
2.Down:pressrightsideofthebutton(q).
4.3Lockingfunction
Thelockingfunctionpreventsusageofcertainfunction
buttons.Regardlessofhowmanybuttonsyourhandcontrol
has,theycanallbelockedindividually.
1.InsertkeyAinthekeyholelocatedjustabovethe
desiredfunction.
2.Tolock,turnthekeyclockwise.
3.Tounlock,turnthekeyanti-clockwise.
4.4Operatingsiderails
WARNING!
Riskofentrapmentorsuffocation
There’sariskofentrapmentorsuffocation
betweenmattresssupport,siderailandbedend.
–Alwaysensurecorrectttingandtighteningof
thesiderails.
WARNING!
Riskoffalling
Ifithasbeendecidedthattheuserneedsside
rails:
–Neverleavetheuserunattendedinthebed
withthesideraildown.
–Makesurethesiderailisinitshighestposition
andproperlylockedwhileleavingtheuser
unattended.
CAUTION!
RiskofInjury
Ifthesiderailisnotlockedproperlyitcanfall
down.
–Pull/pushthetopbarofthesiderailtoensure
thatthelockingsystemisproperlyengaged.
Backrestsection
1.Up:pressleftsideofthebutton(p).
2.Down:pressrightsideofthebutton(q).
Thighsection
1.Up:pressleftsideofthebutton(p).
2.Down:pressrightsideofthebutton(q).
CAUTION!
RiskofInjury
Thereisariskofentrapmentorsqueezingwhile
operatingthesiderail.
–Payattentiontoyourngersandbodyparts
ofthepatient.
–Neverforceordropthesiderailwhilehandling
it.
Heightadjustment
1.Up:pressleftsideofthebutton(p).
2.Down:pressrightsideofthebutton(q).
181563578-H

Operatingthebed
4.4.1VersoIIsiderails
Foldingdowntoopenthesiderail
1.HoldthetopbarBwithonehandandpullthelocking
buttonAwiththeotherhand.
2.Pullthetopbarsidewardsandreleasethelocking
button.
3.Foldthesideraildowntoitslowestposition.
Foldinguptoclosethesiderail
1.PullthetopbarBandraisethesiderailuntilitis
properlyengagedinthelockingsystem.
4.4.3BrittVandLinesiderails
Lock/Upperposition
Pullupthetopwoodenside
railbar ,untilthelockingpin
lockswithanaudibleclick
Release
Liftthetopwoodenside
railbarandpressthetwo
lockingringstogether.
Down/Lowerposition
Lowerthesiderail.
Usingasiderailheightextender
4.4.2OperatingScala2siderail
NOTICE!
–Whenmanoeuvringthesiderail,placeyour
handsonlyintheareasmarkedwiththegrip
label.
1.2.
1.Up:LiftandpullthetopbarBofthesiderailtowards
theendwiththelockingmechanism(releasebutton)A
.Ensure,itislockedinplace.
2.Down:PressthereleasebuttonAandpushthetopbar
Bofthesiderailawayfromthelockingmechanism.
LineandBrittVsiderailscanbesupplementedbyaheight
extender .Asiderailheightextenderincreasestheallowed
mattressthicknessby15cm.
WARNING!
Riskofinjuryordeath
Amisappliedsiderailheightextendercouldfall
off.There’sanincreasedriskfortheusertofall
outofthebed.
–Makesurethatthesiderailheightextenderis
properlyattachedtothesiderail.Refertothe
usermanualofthesiderailheightextender .
–Makesuretousethecorrectheightextender
foryoursiderail(theshapemustmatchto
theshapeofthesiderail.Alabelonthe
heightextenderdeterminestowhichsiderailit
belongs.Seeimagebelow.
HeightextenderforLINE
siderail:
HeightextenderforBRITTV
siderail:
1563578-H19

Invacare®Alegio™NG
4.4.4Lisasiderail
Operatingthesiderail
Raisethesiderail
1.Usethetwohandlesontheupperbartoraisetheside
railuntilitlocksintheupperpositionwithanaudible
click.
Lowerthesiderail
1.
PresstheglidingsystemlatchAononeendofthebed
andlowerthesiderailtothelowestposition.
2.Repeattheprocedureattheotherendofthebed.
4.5.2Centralbrakesystem
Allthewheelsarelockedbylockingoneofthepedals.
Alwaysstandinthemiddleofthebed'slongside
whenlockingthebrakes.
1.Lockthebrakes:
StepontheredpedalB.
4.5Castorsandbrakes
CAUTION!
Riskoftrapping/squeezing
Allbrakesarefoot-operated.
–Donotreleasethebrakewiththengers.
CAUTION!
Riskofinjury
Usercanfall,whileenteringorexitingthebed,if
brakesarenotlocked.
–Alwayslockthebrakesbeforetheuseris
movinginoroutofbedorwhenattendingto
theuser.
Castorsmightmakemarksondifferenttypesof
absorbingoorsurfaces-suchasuntreatedor
badlytreatedoors.T opreventmarks,Invacare®
recommendstoplaceasuitablekindofprotection
betweenthecastorsandtheoor.
4.5.1Castorbrake
1.Unlockthebrakes:
Steponthegreenpedal
Auntilthebrakeisin
neutralposition=Both
pedalsonthesame
level.
4.6Liftingpole
WARNING!
Riskofinjury
Thebedcantipifthehandleisused,whilethe
liftingpoleisturnedawayfromthebed.
–Theliftingpolealwayshastobepositioned
withthehandlehangingoverthebedarea.
–Donotexceedthemaximumloadofthelifting
pole;80kg.
4.6.1Placetheliftingpole
Theliftingpolecanbeplacedeitherontheleftortheright
sideoftheheadendofthebed.
1.Inserttheliftingpoleintotheliftingpoletubeandxit
withthengerscrew.
It’snotrequiredtotightenthengerscrew,incase
youwanttheliftingpoletoswingawaytotheside
ofthebed.
Lockingthebrake-StepontheouterpartBofthebrake
pedal
Unlockingthebrake-StepontheinnerpartAofthebrake
pedal
201563578-H

Operatingthebed
4.6.2Adjustingthehandleheight
Thehandleheightshouldalwaysbeadjustedtotheuser’s
need.
1.2.
1.HoldthecordoutwhileslidingtheplasticcordlockA
upwardsordownwardsuntilthehandlereachesthe
preferredheight.
2.Lockthehandlebypressingbackthecordinthelock
Aandpullthehandledownwards.
NOTICE!
Afteradjustingthehandleheight:
–Checkthatthetwocordsabovethecordlock
areparallelandinsidethecordlock.
–Makesurethecordisproperlylockedbypulling
thehandlehard.
•incaseofpowerfailureorurgentneedforaction,use
themanualemergencyreleaseandlowerthemattress
supportsection.
Referto4.8.1Emergencyreleaseofamattresssupport
section,page21
.
4.8.1Emergencyreleaseofamattresssupport
section
Incaseofapower-ormotorfailure,anemergencyrelease
oftheback,thighorlegsectioncouldbenecessary.An
emergencyreleaseoftheheightadjustmentisNOTpossible.
CAUTION!
Riskofinjury
–Aminimumoftwopersonsisrequiredfor
anemergencyreleaseofamattresssupport
section.
–Whenreleasingamattresssupportsection,
itmightlowerfast.Donotreachunderthe
mattresssupportwhileloweringit.
NOTICE!
–Beforeanemergencyreleaseofthemattress
support,removetheplugfromthemains
socket.
4.7Adjustingthelegsection
Operatethelegsectionbyliftingthemattresshandle:
1.Up:Liftthemattresshandleonthelegsection.
2.Down:Liftthemattresshandleonthelegsectionright
upandthenlowerthelegsection.
4.8Emergencypositions
Incaseofamedicalemergency,itmightberequiredto
bringallorcertainmattresssupportsectionsintoaattened
position-e.g.aattenedbacksectionforcardiopulmonary
resuscitation(CPR).
Tobringamattresssupportsectionintoaattenedposition,
•usethecorrespondingfunctiononthehandcontrol
1.Bothpersonsholdthemattresssection.
2.Oneofthemlocatesthemotorinquestionandpulls
outthesafetypin.
3.Bothpersonsslowlylowerthemattresssectionuntilit
iscompletelydown.
4.9TransportationandStorage
Thebedcanbetransportedorstoredinassembledor
disassembledstate.
See3.10Dismantlethebed,page16fordetailedinstructions
onhowtodisassemblethebed.
Totransportorstorethebedinitsassembledstate:
1.Disconnectthepowercablefromthemainssocket.
2.Coilupthepowercableandplaceitaroundtheside
tubeofthebed,soitiskeptclearoftheoorand
damageduringtransportorstorageisprevented.
OR
1563578-H
21

Invacare®Alegio™NG
5Options
5.1Listofavailableoptions
Duetoregionaldifferences,refertoyourlocal
Invacarewebsiteorcatalogfordetailsaboutavailable
optionsorcontactyourInvacareprovider .
•Siderails
–VersoII
3/4lengthcollapsiblesteelsiderailmountedonthe
sidetube.Releasebuttoninthefootsection
–ScalaBasic2
3/4lengthcollapsiblesteelsiderail,mountedonthe
sidetube.Releasebuttonintheheadsection
–ScalaBasicPlus2
3/4lengthcollapsiblesteelsiderail,mountedonthe
sidetube.Releasebuttonintheheadsection
–ScalaMedium2
3/4lengthcollapsiblesteelsiderailmountedonthe
sidetube.Releasebuttonintheheadsection
–BrittV
Fulllengthcollapsiblewoodsiderailmountedonthe
bedendguideways.Releasebuttonsinthemiddle
–Linestandard
Fulllengthcollapsiblealuminiumsiderailmounted
onthebedendguideways.Releasebuttonsinthe
middle
–Lineextendable
Fulllengthcollapsiblealuminumsiderailwith
telescopicextensionmountedonthebedend
guideways.Releasebuttonsinthemiddle
–Lisa
Fulllengthcollapsiblewoodsiderailmountedonthe
bedendguideways.Releasebuttonsinbothends
–HeightextendersforBrittVandLinesiderails
•Mattresssupportextension(footend)–15cm
•Liftingpole
•IVdriprod
•Fixedsupporthandles–40cmx30cmand40cm
x50cm
•Handcontrols
–withbackandheightadjustment
–withsittingpositionandback,thighandheight
adjustment
–withsittingpositionandback,thighandheight
adjustment(lockable)
•Rastox
•Transporttrolley
NOTICE!
–Useonlyoriginaloptionandspareparts.Spare
partslistsandextrausermanualscanbe
orderedfromInvacare.
5.2Transporttrolley
Thetransporttrolleyisnotcompatiblewithbeds
equippedwithcentralbrake.
22
1563578-H

Options
Mountingthebedonthetransporttrolley
1.Locktrolleycastors.
2.
MountthefootsectionontothetubesAwiththemotor
facingawayfromthetrolleycross-channelE.Secure
thefootsectiononthetrolleywiththengerscrews.
3.
5.
a.Lockthebrakesofthe2bottomcastorsofthebed
baseandpositiontheminwardstoavoidthatthey
touchthegroundduringthemanipulation.
b.Alignsquareglidebushingsofbaseassemblywith
bracketsDandpushthebedbaseontothetrolleyas
faraspossible.
6.
PlacetheliftingpoleontoitslocationB.
4.
MounttheheadsectionontothetubesCwiththe
motorfacingawayfromthetrolleycross-channelE.
Securethefootsectiononthetrolleywiththenger
screws.
Liftendofbaseassemblyintoaverticalposition,
allowingthesquareglidebushingstoengageintothe
bracketsD.Securebasetotrolleywithspringpins.
1563578-H23

Invacare®Alegio™NG
7.
PlacetheVersoIIsiderails(ifapplicable)into
cross-channelofthetrolleyE.
8.
Placebedends(ifapplicable)ontotrolleyhooksF.
24
1563578-H

Maintenance
6Maintenance
6.1Generalmaintenanceinformation
CAUTION!
Riskofinjuryordamage
–Donotperformanymaintenanceorservice
procedureswhiletheproductisinuse.
Followthemaintenanceproceduresdescribedinthismanual
tokeepyourproductincontinuousoperation.
6.2Dailyinspections
WARNING!
Riskofinjuryordamage
Damagedorwornpartscanaffectthesafetyof
theproduct.
–Theproductshouldbecheckedeachtimeit
isused.
–Donotusetheproductifdamageisfound
oryouquestionthesafetyofanypartof
theproduct.ContactyourInvacareprovider
immediatelyandensuretheproductisnotused
untilrepairsaremade.
DailyInspectionChecklist
qVisuallyinspectthebed.Checkallpartsforexternal
damageorwear .
qCheckallhardware,attachmentpoints,stressedparts
andanypivotpointsforsignsofwear,cracking,fraying,
deformationordeterioration.
qCheckthehandcontrolandallactuatorsforsmooth
operation.
Everyyear
•Werecommendasafetytestcomprisingtheactuators’
performanceandmechanicalstate.
Everysecondyear
•Westronglyrecommendaserviceaccordingtothe
belowchecklistaftertwoyearsofnormaluseandthen
everysecondyear .
Actuators,handcontrolandcontrolunitareserviced
byexchangingthefaultypart.
6.4Checklist-Maintenance
Checkpoints
qCirclips,cotterpinsandplasticxingring-properly
lockedandintact.
qScrews-tightened.
qWeldings-intact.
qSideraillockingandmovingsystem-properlylocking
andrunningsmoothly .
qCastorttings-tightened.
qCastorbrakes-lockingproperly .
qHeightadjustmentmotor-runningproperly.
qBackrestmotor-runningproperly.
qThighsectionmotor-runningproperly.
qWires-correctlywiredandundamaged.
qElectricplugs-undamaged.
qDamagedcoating-repaired.
qLinesiderailstraps-notfrayedorcracked.
qAccessories-correctlyassembledandcorrectfunction.
qRastoxtting-checkfunction.
qMotors,controlunitandhandcontrol-keepcleanfrom
dustanddirt.Checkfordamage.
qAnchorpoints,cables,pistonrod,casingandplugs-
Checkfordamage.
6.3Generalservicinginformation
NOTICE!
–Themattresssupportmustbesupported
duringserviceinspections,topreventaccidental
lowering.
–Onlypersonnelwhohavereceivedthe
necessaryinstructionsortrainingmayperform
serviceandmaintenanceofthebed.
–Afterreconditioningthebed,orifbedfunctions
change,servicemustbecarriedoutaccording
tothechecklist.
Aservicecontractcanbemadeinthecountries,where
Invacarehasitsownsalescompany.Incertaincountries
Invacareofferscoursesinserviceandmaintenanceofthe
bed.Sparepartslistsandadditionalusermanualsare
availablefromInvacare.
BeforeUse
•Ensurethatallmanualandelectricalpartsfunctions
correctlyandareinasecurestate.
•Check,byraisingandloweringthebed,thattheshear
armsrunsmoothlyintheguideways.
Afterthreemonths
•Ensurethatallmanualandelectricalpartsare
functioning,andtightenbolts,screws,nuts,etc.
6.5Inspectionafterrelocation-Prepare
fornewuser
NOTICE!
Whenthebedhasbeenrelocated;beforegiven
toanewuserithastobethoroughlyinspected.
–Inspectionmustbedonebyatrained
professional.
–Forregularmaintenanceseemaintenancechart.
6.6Checklist-Afterrelocation
Checkpoints
qCheckthettingsbetweensheararmandmattress
support(screwstightenedproperly).
qCheckthelockingofthemotors(pipepinscorrectly
mounted).
qChecktheelectronicwiringforthemotors(wiresnot
squeezed).
qCheckthattheenclosingofthemotorsareintact(no
cracksthatenablesuidstopierce).
qCheckthesectionsofthemattresssupportbyusing
thehandcontroltoactivateallfunctionsofthemoving
parts.
qCheckthefunctionofthebrakes.
1563578-H25

Invacare®Alegio™NG
qCheckthelockingfunction(s)ofthesiderails.
6.7CleaningandDisinfection
6.7.1Generalsafetyinformation
CAUTION!
Riskofcontamination
–T akeprecautionsforyourselfanduse
appropriateprotectiveequipment.
CAUTION!
Riskofelectricshockandproductdamage
–Switchoffthedeviceanddisconnectfrom
mains,ifapplicable.
–Whencleaningelectroniccomponentsconsider
theirprotectionclassregardingwateringress.
–Makesurethatnowatersplashestotheplug
orthewalloutlet.
–Donottouchthepowersocketwithwethands.
NOTICE!
Wronguidsormethodscanharmordamage
theproduct.
–Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand
mustprotectthematerialstheyareusedto
clean.
–Neverusecorrosiveuids(alkalines,acidetc.)
orabrasivecleaningagents.Werecommend
anordinaryhouseholdcleaningagentsuchas
dishwashingliquid,ifnotspeciedotherwisein
thecleaninginstructions.
–Neveruseasolvent(cellulosethinner ,acetone
etc.)thatchangesthestructureoftheplastic
ordissolvestheattachedlabels.
–Alwaysmakesurethattheproductiscompletely
driedbeforetakingintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term
careenvironments,followyourin-houseprocedures.
•Cleaningagent:Werecommendusingamilddetergent
withneutralornearneutralpH(5-9).
Commerciallyavailableconsumerproductslike
handdishwashingliquidorall-purposecleaners
canbeused.Readtheinstructionsonthe
labelofyourcleaningagentanduseitatthe
indicatedconcentration.
•Max.watertemperature:40°C
1.Wipesurfacesthoroughlywithasoftclothslightly
moistenedwithdetergentsolutionuntilallvisiblesoil
isremoved.
2.Removealldetergentresiduewithacleandampcloth.
3.Wipedryusingacleansoftcloth.
Removabletextiles(includingupholsteryandmattresses)
•Seeattachedlabeloneachproduct.
6.7.4DisinfectionInstructions
InDomesticCare
•Disinfectant:Werecommendusinganalcohol-based
surfacedisinfectant(with70-90%alcohol).
Readtheinstructionsonthelabelofyour
disinfectant.Itprovidesinformationonthe
activityspectrum(bacteria,fungiand/orviruses),
materialcompatibilityandthecorrectexposure
time.
1.Ensurethesurfacesarecleanedbeforedisinfection.
2.Moistenasoftclothandwipe-disinfectallaccessible
surfacesandkeepthemmoistenedfortheexposure
timestatedonthelabelofthedisinfectant.
3.Allowtheproducttoair-dry.
InInstitutionalCare
Followyourin-housedisinfectionproceduresandonlyuse
thedisinfectantsandmethodsspeciedtherein.
6.8Lubrication
6.7.2CleaningIntervals
NOTICE!
Regularcleaninganddisinfectionenhancesmooth
operation,increasestheservicelifeandprevents
contamination.
Cleananddisinfecttheproduct:
–regularlywhileinuse,
–beforeandafteranyserviceprocedure,
–whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser .
6.7.3CleaningbyHand
Allcomponents(excludingremovabletextiles)
Werecommendlubricatingthebedaccordingtothe
followingtable:
PartofbedLubricationmethod
Pointsofrotationinmattress
supportandbaseframe
Motorattachmentpointsto
mattressplatform
Upperattachmentpointof
Rastoxtting
Scissorpivotboltsand
attachmentbolts
Fororderingofcorrectoilandgrease,pleasecontact
yourInvacaredealer.
Oil(medicallyclean)
Oil(medicallyclean)
Oil(medicallyclean)
Oil(medicallyclean)
261563578-H

AfterUse
7AfterUse
7.1Disposal
Beenvironmentallyresponsibleandrecyclethisproduct
throughyourrecyclingfacilityatitsendoflife.
Disassembletheproductanditscomponents,sothedifferent
materialscanbeseparatedandrecycledindividually .
Thedisposalandrecyclingofusedproductsandpackaging
mustcomplywiththelawsandregulationsforwaste
handlingineachcountry.Contactyourlocalwaste
managementcompanyforinformation.
7.2Reconditioning
Thisproductissuitableforreuse.Toreconditiontheproduct
foranewuser,carryoutthefollowingactions:
•Inspection
•Cleaninganddisinfection
Fordetailedinformation,see6Maintenance,page25.
Makesurethattheusermanualishandedoverwiththe
product.
Ifanydamageormalfunctionisdetected,donotreuse
theproduct.
1563578-H
27

Invacare®Alegio™NG
8Troubleshooting
8.1Troubleshootingelectricalsystem
Symptom
Mainsindicatorlightsup,butthemotor
isnotrunning.Therelayinthecontrol
unitmakesaclickingnoise.
Mainsindicatorlightsup,butthemotor
isnotrunning.Norelaysoundisheard
fromthecontrolunit.
Controlunitisinorderexceptforone
directionononechannel.
Motorisrunning,butthepistonrod
doesnotmove.
Themotorcannotliftfullload.
Motornoise,butnomovementof
pistonrod.
Pistonrodoperatesinwardsandnot
outwards.
*Serviceandmaintenanceofthebedmustonlybeperformedbypersonnelwhohavereceivedthenecessaryinstructionor
training.
PossiblecauseRemedy
Mainsarenotconnected
Fuseinthecontrolunitisblown
Controlunitisdefective
Motorplugisnotfullyinsertedintothe
controlunit.
Themotorisdefective.
Motorcableisdamaged.
Controlunitisdefective
Controlunitisdefective
Handcontrolisdefective
Controlunitisdefective
Handcontrolisdefective
Motorisdamaged
Connectmains
*Replacethecontrolunit Mainsindicatordoesnotlightup
*Replacethecontrolunit
Insertthemotorplugproperlyintothe
controlunit
*Replacethemotor
*Replacethecable
*Replacethecontrolunit
*Replacethecontrolunit
*Replacethehandcontrol
*Replacethecontrolunit
*Replacethehandcontrol
*Replacethemotor
Riskofpersonalinjuryanddamagetotheproduct.
–Thebedmustbeunpluggedfromthemainpowersourcebeforeopeningorrepairingelectricalparts.
WARNING!
Riskofpersonalinjuryanddamagetotheproduct.
–Thebedmustbeunpluggedfromthemainpowersourcebeforeopeningorrepairingelectricalparts.
281563578-H

TechnicalData
9TechnicalData
9.1Materials
MattressPlatform
BedEnds
SideRails
LiftingPole
Actuatorshousing,
Handcontrol,
Controlbox,Castors
andotherPlastic
parts
Boltsandnuts
Matressretainers
9.2Dimensions
Steel(powder-coated)
Steel(powder-coated)andMDF
Steel(powder-coated),woodor
aluminium
Steel(powder-coated),PPandPOM
Materialaccordingtomarking(PA,PP ,
PE,ABS)
Steel(Zinc-plated)
ABS
Dimensions[cm]
TotallengthH
Totalheight(withlifting
pole)I
Mattresssupporttooor
heightJ
BasetooorheightK
Alldimensionsarestatedwithouttolerancesfora
non-extendedbedwithcastorsof125mmdiameter .
153-205
28-80
9.3Weights
MaximumLoads
Max.patientweight
(providedthattheweightofthe
mattressandtheaccessoriesdonot
exceed35kg)
Max.safeworkingload
(patient+accessories)
211
16
135kg
170kg
MattresssupportwidthA
TotalwidthB
(dependingonbedend)
MattresssupportlengthC
BacksectionlengthD
SeatsectionlengthE
ThighsectionlengthF
LegsectionlengthG
Dimensions[cm]
90
Emma,Vibeke,Piggy:90
Sabrine:95
Anita,Susanne,Sophie,
Victoria:100
Camila:102
200
80
23
30
67
BedPartsandAccessories
AlegioNGstandardbed,complete,excluding
accessories
Mattresssupport,standardheadend14.0kg
Mattresssupportwithauto-regression,head
end
Mattresssupport,footend,15.5kg
Mattresssupport,footend,footmotor17.0kg
Mattresssupport,footendthighmotor17.0kg
Baseandsheararm35.0kg
Base
Sheararm19.5kg
BedEnd,Vibeke–1piece6.0kg
BedEnd,Emma–1piece7.0kg
BedEndSusanne/Sophie–1piece
BedEnd,Camila–1piece11.0kg
BedEnd,Sabrine–1piece6.8kg
BedEnd,Victoria–1piece10.0kg
BedEnd,Piggy–1piece6.0kg
BedEnd,Anita–1piece9.0kg
Liftingpole4.2kg
Rastox0.3kg
Mattresssupportextension(15cm)
65.0kg
17.0kg
15.5kg
8.5kg
1.8kg
9.4MattressDimensions
Allowedmattresssizes,dependingonthesiderail
Mattressmeasures(incm)
SiderailHeightWidthLength*
Versoll10–1885,5–95195–206
Linesiderail12–2085,5–95195–206
1563578-H29

Invacare®Alegio™NG
Lisasiderail12–1785,5–95195–206
BrittVside
rail12–2085,5–95195–206
ScalaBasic210–1785,5–95195–206
ScalaBasic
Plus210–1785,5–95195–206
ScalaMedium
210–2285,5–95195–206
*Forbedsequippedwithamattresssupportextension,add
15cm.
Min.density:38kg/m
9.5SideRails
SideRail
VersoII
3/4lengthcollapsiblesteel
siderail
ScalaBasic2
3/4lengthcollapsiblesteel
siderail
ScalaBasicPlus2
3/4lengthcollapsiblesteel
siderail
ScalaMedium2
3/4lengthcollapsiblesteel
siderail
BrittV
Fulllengthcollapsible
woodsiderail
Linestandard
Fulllengthcollapsible
aluminiumsiderail
Lineextendable
Fulllengthcollapsible
aluminumsiderailwith
telescopicextension
Lisa
Fulllengthcollapsible
woodsiderail
Measures
[cm]
150x407.0
168x407.4
168x407.4
165x468.9
205x40
206x404.5
206-226x406.5
205x407.0
Weights
[kg/pc]
7.7
9.6Environmentalconditions
Storageand
Transportation
Temperature
Relativehumidity
Atmospheric
pressure
Allowtheproducttoreachoperationtemperature
beforeusage:
•Warmingupfromminimumstoragetemperature
mighttakeupto24hours.
•Coolingdownfrommaximumstorage
temperaturemighttakeupto24hours.
-10°Cto+50°C+5°Cto+40°C
20%to80%-notcondensing
700hPato1060hPa
9.7Electricalsystem
Voltage
supply
Maximum
currentinput
Intermittent
(periodic
motor
operation)
Degreeof
protection
Insulation
class
TypeB
AppliedPart
Soundlevel
Thebedhasnoisolator(mainswitch).Ifthebed
needstobeelectricallydisconnected,ithastobe
unpluggedfromthemainssocket.
Uin=230Volt,AC,50/60Hz(AC=
Alternatingcurrent)
Iinmax.=1.5A
Int=10%,max.2min/18min
IPX6
Thecontrolunitandmotorsareprotected
accordingtoIPX6.
IPX6–Theelectricalsystemisprotected
againstwaterprojectedinpowerfuljets
fromanydirection.
ClassIIequipment
AppliedPartcomplyingwiththespecied
requirementsforprotectionagainstelectrical
shockaccordingtoIEC60601-1.
45to50dB(A)
Operation
301563578-H

Electromagneticcompatibility(EMC)
10Electromagneticcompatibility(EMC)
10.1GeneralEMCinformation
MedicalElectricalEquipmentneedstobeinstalledandusedaccordingtotheEMCinformationinthismanual.
ThisproducthasbeentestedandfoundtocomplywithEMClimitsspeciedbyIEC/EN60601-1-2forClassBequipment.
PortableandmobileRFcommunicationsequipmentcanaffecttheoperationofthisproduct.
Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagneticemissionspermittedbytheabove
standard.T odetermineiftheemissionfromthisproductiscausingtheinterference,runandstoprunningthisproduct.If
theinterferencewiththeotherdeviceoperationstops,thenthisproductiscausingtheinterference.Insuchrarecases,
interferencemaybereducedorcorrectedbythefollowing:
•Reposition,relocate,orincreasetheseparationbetweenthedevices.
10.2Electromagneticemission
Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthis
productshouldassurethatitisusedinsuchanenvironment.
Emissionstest
RFemissions
CISPR11
RFemissions
CISPR11
Harmonic
emissions
IEC61000-3-2
Voltageuctuations
/ickeremissions
IEC61000-3-3
ComplianceElectromagneticenvironment-guidance
ThisproductusesRFenergyonlyforitsinternalfunction.Therefore,itsRF
GroupI
ClassB
ClassA
Complies
emissionsareverylowandarenotlikelytocauseanyinterferenceinnearby
electronicequipment.
Thisproductissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
10.3ElectromagneticImmunity
Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthis
productshouldassurethatitisusedinsuchanenvironment.
Immunitytest
Electrostatic
discharge(ESD)
IEC61000-4-2
Electrostatic
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Test/Compliancelevel
±8kVcontact
±2kV,±4kV,±8kV,±
15kVair
±2kVforpowersupply
lines;100kHzrepetition
frequency
±1kVforinput/output
lines;100kHzrepetition
frequency
±1kVlinetoline
±2kVlinetoearth
Electromagneticenvironment–guidance
Floorsshouldbewood,concreteorceramictile.Ifoorsarecoveredwith
syntheticmaterial,therelativehumidityshouldbeatleast30%.
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
1563578-H31

Invacare®Alegio™NG
Immunitytest
Voltagedips,short
interruptionsand
voltagevariations
onpowersupply
inputlines
IEC61000-4-11
Powerfrequency
(50/60Hz)
magneticeld
IEC61000-4-8
ConductedRF
IEC61000-4-6
RadiatedRF
IEC61000-4-3
Test/Compliancelevel
<0%UTfor0,5cycleat
45°steps
0%UTfor1cycles
70%UTfor25/30cycles
<5%UTfor250/300
cycles
30A/m
3V
150kHzto80Mhz
6V
inISM&amateurradio
bands
10V/m
80Mhzto2,7GHz
385MHz-5785MHz
testspecications
forimmunitytoRF
wirelesscommunication
equipmentrefertotable
9ofIEC60601-1-2:2014
Electromagneticenvironment–guidance
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.Iftheuserofthisproductrequirescontinuedoperationduring
powermainsinterruptions,itisrecommendedthattheproductispowered
fromanun-interruptiblepowersupplyorabattery.
UTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.
Powerfrequencymagneticeldsshouldbeatlevelscharacteristicofa
typicallocationinatypicalcommercialorhospitalenvironment.
Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio
(cellular/cordless)telephonesandlandmobileradios,amateurradio,AM
andFMradiobroadcastandTVbroadcastcannotbepredictedtheoretically
withaccuracy.ToassesstheelectromagneticenvironmentduetoxedRF
transmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthe
measuredeldstrengthinthelocationinwhichthisproductisusedexceeds
theapplicableRFcompliancelevelabove,thisproductshouldbeobserved
toverifynormaloperation.Ifabnormalperformanceisobserved,additional
measuresmaybenecessary,suchasreorientingorrelocatingthisproduct.
Interferencemayoccurinthevicinityofequipmentmarkedwiththe
followingsymbol:
PortableandmobileRFcommunicationsequipmentshouldbeusedno
closerthan30cmtoanypartofthisproductincludingcables.
Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreection
fromstructures,objectsandpeople.
10.3.1TestspecicationsforimmunitytoRFwirelesscommunicationsequipment
IEC60601-1-2:2014—Table9
TestFrequency
a)
Band
(MHz)
(MHz)
385380-390TETRA400
450430-470
710
745
704-787
788
810
870
800-960
930
Service
GMRS460,
FRS460
LTEBand13,17
GSM800/900,
TETRA800,
iDEN820,
CDMA850,LTE
Band5
a)
Modulation
Pulse
modulation
18Hz
FM
±5kHz
deviation
1kHzsine
Pulse
modulation
217Hz
Pulse
modulation
18Hz
b)
Maximum
Distance(m)
power(W)
b)
c)
1.80.327
20.328
b)
b)
0.20.39
20.328
Immunitytest
level(V/m)
321563578-H

Electromagneticcompatibility(EMC)
TestFrequency
a)
Band
(MHz)
Service
a)
Modulation
(MHz)
b)
Maximum
Distance(m)
power(W)
GSM1800;
1720
1845
1700-1990
1970
CDMA1900;
GSM1900;
DECT;L TEBand
1,3,4,25;
modulation
217Hz
Pulse
b)
20.328
UMTS
Bluetooth,
24502400-2570
WLAN,802.11
b/g/n,RFID
2450,LTE
modulation
217Hz
Pulse
b)
20.328
Band7
5240
5500
5100-5800
5785
WLAN802.11
a/n
modulation
217Hz
Pulse
b)
0.20.39
Ifnecessarytoachievetheimmunitytestlevel,thedistancebetweenthetransmittingantennaandtheMedical
ElectricalEquipmentorSystemmaybereducedto1m.The1mtestdistanceispermittedbyIEC61000-4-3.
a)
Forsomeservices,onlytheuplinkfrequenciesareincluded.
b)
Thecarriershallbemodulatedusinga50%dutycyclesquarewavesignal.
c)
AsanalternativetoFMmodulation,50%pulsemodulationat18Hzmaybeusedbecausewhileitdoesnotrepresentactual
modulation,itwouldbeworstcase.
Immunitytest
level(V/m)
1563578-H33

Notes

Notes

UnitedKingdom&Ireland:
InvacareLimited
PencoedTechnologyPark,
Pencoed
BridgendCF355AQ
Tel:(44)(0)1656776200
uk@invacare.com
www.invacare.co.uk
InvacarePortugal,Lda
RuaEstradaVelha949
4465-784LeçadoBalio
Portugal
InvacareUKOperations
Limited
Unit4,PencoedTechnology
Park,Pencoed
BridgendCF355AQ
UK
1563578-H2023-02-23
MakingLife’sExperiencesPossible®
*1563578H*