INVACARE 9800A BOOM RETRO-FIT
KIT NO.1060539
Installation, Assembly and Operating Instructions
SAFETY SUMMARY
WARNING/CAUTION notices apply to hazards and
unsafe practices which could result in personal
injury and/or property damage.
WARNING
Check all parts for shipping damages before using.
In case of damage, DO NOT use the equipment.
Contact the Dealer/Carrier for further instructions.
DO NOT operate this equipment without first
reading and understanding this instruction sheet. If
you are unable to understand the warnings and
instructions, contact a healthcare professional
(Doctor-Therapist) before attempting to use this
equipment - otherwise injury or damage may result.
After making adjustments, always make sure that
all parts are properly tightened BEFORE using this
patient lift.
DISASSEMBLY OF THE EXISTING SWIVEL
BAR, PUMP AND BOOM ASSEMBLIES
Swivel Bar Removal (FIGURE 1)
Pump Removal (FIGURE 2)
1. Remove the nut and bolt that secure the pump
to the boom assembly.
Nut
Boom
NOTE: Not necessary
to disconnect.
Bolt
FIGURE 2 - PUMP- BOOM DISCONNECT
1. Remove the swivel bar from the eye at the end
of the boom.
FIGURE 1 - SWIVEL BAR REMOVAL
Boom Removal (FIGURE 3)
1. Remove the bolt and locknut that secures the
existing boom to the TOP of the mast assembly.
FIGURE 3 - BOOM REMOVAL
1
ASSEMBLY OF THE EXISTING PUMP, NEW
BOOM AND SWIVEL BAR ASSEMBLIES
Swivel Bar Attachment with Mounting
Bracket (FIGURE 5)
Pump Assembly
1. Using the nut and bolt provided, assemble the
pump to the new boom. Refer to FIGURE 2.
Boom Assembly (FIGURE 4)
1. Align the holes of the boom assembly and mast assembly mounting bracket. Insert the bolt
completely through the holes of the mast
assembly mounting bracket and the boom
assembly.
NOTE: Be sure that the bolt is completely through
the holes of the mast assembly mounting bracket
and boom assembly; and they are not riding on the
shoulder of the bolt.
CAUTION
The boom and mast when assembled MUST
pivot easily.
2. Tighten the bolt and nut using an Allen wrench and
an adjustable wrench.
NOTE: The nut and bolt can be tightened
completely and the boom assembly will pivot easily
if the mounting hardware is aligned properly when
the boom assembly is secured to the mast. To check
for correct alignment of the hardware, use an Allen
wrench and turn the bolt by hand (it should rotate
without a lot of force being applied). If excessive force
is necessary, the bolt is NOT aligned properly.
Disassemble and repeat assembly procedures until
proper alignment is obtained.
1. Pull the pin up through the swivel bar pad.
2. Line-up hole in pin with mounting bracket holes
and secure with shoulder screw and nut
provided.
NOTE: The swivel bar has hooks on both ends
to accept the hardware used to attach the sling
that supports the patient during lift. The swivel
bar pad should remain in place during normal
use.
Nut
Pin
Shoulder
Screw
Swivel Bar
Pad
NOTE: Make sure Pin and Washers are completely
pushed UP through the Swivel Bar Pad.
FIGURE 5 - SWIVEL BAR ATTACHMENT
WITH MOUNTING BRACKET
NOTE: Refer to Owner's Manual for proper
operation of your patient lift.
FIGURE 4 - NEW BOOM INSTALLATION
899 Cleveland Street
Form No. 95-166 Part No. 1060540 Printed in U.S.A.
P.O. Box 4028
Elyria, Ohio 44036-2125
Phone 1-(800)-333-6900