IntriCon RX92369 User Manual

Generic / Preliminary
Sirona
USER MANUAL
ORDER OF A PHYSICIAN.
Carefully read all instructions prior to use. Observe all warnings and precautions noted in these directions. Failure to do so may result in patient complications.
IntriCon Datrix
340 State Place
Escondido, CA 92029
Sirona User Manual Rev D02
-1-
Table of Contents
Notices ................................................................................................................................ 3
User Safety Information...................................................................................................... 4
Warnings............................................................................................................... 4
Cautions ................................................................................................................ 5
Notes:.............................................................................................................................. 5
Section 1: Introduction....................................................................................................... 6
Section 2: Getting Started.................................................................................................. 6
Front Panel.......................................................................................................................... 7
Section 3: Initial Device Setup ........................................................................................... 8
Section 4: Attaching Recorder to Patient............................................................................ 8
Patient Cable...................................................................................................................9
Patient Preparation.......................................................................................................... 9
Patient Hookup................................................................................................................9
Section 5: Instructions for Patient.................................................................................... 10
RECORD Button .......................................................................................................... 11
SEND Button................................................................................................................ 11
SIRONA Patient Interface Light Signals...................................................................... 13
Section 6: Device Maintenance ....................................................................................... 14
Inspection and Cleaning................................................................................................ 14
Testing...........................................................................................................................14
SIRONA DATASHEET................................................................................................... 17
Service/Technical Support:............................................................................................... 18
Sirona User Manual Rev D02
-2-
Notices
Conventions Used in this Manual
WARNING Warning statements describe conditions or actions that can result in
CAUTION Caution statements describe conditions or actions that can result in
NOTE Notes contain additional information on usage.
Manufacturer’s Responsibility
IntriCon Datrix considers itself responsible for effects on safety and performance only if:
1. Readjustments, modifications or repairs to the IntriCon Datrix Holter/Event recorders are carried out only by IntriCon Datrix authorized personnel.
2. The IntriCon Datrix SIRONA is used as presented in this manual.
The warranty is only valid if you use IntriCon Datrix approved replacement parts and accessories.
User Responsibility
The user of this product is responsible for ensuring the implementation of a satisfactory maintenance schedule. Failure to do so may cause undue failure and possible health hazards.
Equipment Identifica tion
IntriCon Datrix equipment is identified by a serial number on the back of the device. Take care not to deface these numbers.
Copyright and Trademark Notices
This document contains information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced or translated to another language without prior written consent of IntriCon Datrix.
Other Important Information
IntriCon Datrix reserves the right to change or amend this manual at anytime without notice.
IntriCon Datrix makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. IntriCon Datrix shall not be liable for errors or omissions that may appear in this document. IntriCon Datrix makes no commitment to update or to keep current the information contained in this document.
Before using the IntriCon Datrix SIRONA Holter/Event recorder read this manual in its entirety and become thoroughly familiar with the contents.
personal injury or loss of life.
damage to the equipment or loss of data.
AND
Sirona User Manual Rev D02
-3-
User Safety Information
Intended Use
The SIRONA APETS recorder is a small, portable, digital Holter/Event recorder intended for use by medical professionals to acquire ECG data from a single patient in a clinical, point of care or outpatient setting. ECG data is first recorded to a Secure Digital (SD) card and then transferred to a Holter/Event analysis system for review by a physician or other qualified professional.
Explanation of Symbols
IP54
READ MANUAL FIRST
KEEP AWAY FROM MOISTURE
TYPE BF DEVICE
DC CURRENT
NON-IONIZING RADIATION
ELECTRONIC EQUIPMENT DISPOSE OF PROPERLY
MANUFACTURER / MANUFACTURE YEAR
Warnings
1. This device captures and presents data reflecting a patient’s physiological condition that when reviewed by a trained medical professional can be useful in determining a diagnosis. However, the data should not be used as sole means for determining a patient’s diagnosis.
Sirona User Manual Rev D02
-4-
2. Use of accessories other than those recommended by IntriCon Datrix may compromise product performance.
3. To maintain designed operator and patient safety, any peripheral equipment and accessories that can come in direct patient contact must be in compliance with IEC 60601-1.
4. Hardware is designed to meet or exceed IEC 60601-1-2, however some environmental electrical interference may cause an artifact in the ECG. The quality of ECG signals may be adversely affected by electromagnetic interference from environmental sources resulting in non-physiological waveforms with the potential for misinterpretation.
5. This device is not intended for use during an MRI.
6. Before performing defibrillation or applying any high frequency surgical equipment to a patient, remove SIRONA leads and electrodes from the chest area. Cable leads or electrodes trapped under defibrillator pads or paddles during defibrillation or electrodes in contact with high frequency electrosurgical equipment can cause patient burns.
7. Once one or more SIRONA patient leads are connected to a patient, do not allow patient leads to meet with any grounded or live parts. Contact could cause unacceptable levels of electrical current to flow to the patient.
8. The equipment is not intended for infants weighing less than 10 kgs.
Cautions
1. Although the plastic enclosure is designed for a clinical environment and can resist moisture, neither the device nor patient cables should be subjected to autoclaving or steam cleaning.
2. Recommended cleaning procedure is to wipe the exterior surfaces with a cloth dampened with warm water and mild detergent solution and then dry with a clean soft cloth.
3. No serviceable parts are inside. The case cannot be opened without destroying it.
4. Do not pull or stretch patient cables, as this could result in mechanical and/or electrical failures. Store patient cables after use by forming them into a loose loop.
5. Align patient cable connector key and SIRONA key before plugging in patient cable. Forcing misaligned connectors can damage connector pins.
6. Avoid shock or sudden impact.
Notes:
1. Excessive patient movement could interfere with the operation of the device.
2. Proper patient preparation is important to successful application of ECG electrodes and operation of the device.
3. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications not expressly approved by IntriCon for compliance could void the user’s authority to operate the equipment.
4. This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the wireless transmission off and on,then the user is encouraged to consult the dealer.
Formatted: Bullets and Numbering
Sirona User Manual Rev D02
-5-
Section 1: Introduction
This manual is written for clinical professionals. It is assumed that the reader has a working knowledge of medical terminology and procedures as required for monitoring cardiac patients.
Purpose of the User Manual
The User Manual describes how to safely operate the SIRONA Holter/Event recorder, Smart
Dock and linkStream. In the manual, the following are described:
Preparing the device for use Understanding and using the keyboard Acquiring and storing ECG data Transmitting stored ECG data Maintenance
System Description
The SIRONA Holter/Event recorder is a portable, battery-operated ambulatory ECG recorder used by trained technicians to collect ECG data from patients in a clinical, point-of-care or outpatient setting. SIRONA provides ECG waveform analysis and automated transmission of ECG data for review by a physician or other qualified professional. ECG data is stored on an internal SD card and can be accessed through the Smart Dock interface or by transmitting the data over a telephone line. Smart Dock is an interface between a PC based application called
Sirona Viewer and the SD card in Sirona. Smart Dock has two functions –
linkStream is an accessory which provides the convenience of checking lead hook-up wirelessly.
1) Charge the battery in Sirona
2) Access the data stored on the SD card in Sirona
Deleted:
Deleted:
Section 2: Getting Started
Batteries
The Sirona uses an internal lithium polymer battery that is not accessible in the field. The battery is designed to last for the life of the product.
Caution: THE BATTERIES USED IN THIS DEVICE MAY PRESENT A FIRE OR
CHEMICAL BURN HAZARD IF MISTREATED. DO NOT DISASSEMBLE, HEAT
ABOVE 100 C (212 F) OR INCINERATE.
Caution: DISPOSE OF ALL BATTERIES IN ACCORDANCE WITH ALL
APPLICABLE LOCAL REGULATIONS
KEEP AWAY FROM CHILDREN.
Sirona User Manual Rev D02
-6-
Deleted: IRONA
Front Panel
LED
The Sirona front panel contains two buttons and five LED lights (Light Emitting Diodes) as follows:
Record Button Send Button Record Green or Red LED Lead-off Red Battery Green or Red LED
LED
Button
Sirona User Manual Rev D02
-7-
Loading...
+ 15 hidden pages