ISL88731EVAL2Z and ISL88731CEVAL2Z Evaluation
Boards Setup Procedure
Introduction
The ISL88731 and ISL88731C are highly integrated battery
charger controllers for Li-ion/Li-ion polymer batteries with
SMBus interface controlling charge voltage, charge current and
input current limiting. The evaluation board includes a USB to
SMBus interface to allow users to control the ISL88731 or
ISL88731C from a Personal Computer.
What’s Inside
This Evaluation Board Kit contains the following materials:
• Qty (1) ISL88731EVAL2Z or ISL88731CEVAL2Z Evaluation
Board
• Qty (1) USB cable
ISL88731(C)EVAL2Z
ISL88731 EVAL2Z REVB
AC ADAPTER
AC ADAPTER
INPUT
AC ADAPTER
AC ADAPTER
OR
ADJUSTABLE DC POWER
DC POWER SUPPLY
OR
SUPPLY
5V –20V, 5A
+
+
-
-
INPUT
+
-
INTERSIL
INTERSIL
What is Needed
The following materials are recommended to perform testing:
• One AC adaptor or one adjustable 25V/6A DC power supply
• Two adjustable electronic loads with constant current mode
and constant voltage mode
•One Volt Meter
• One 500MHz four-channel oscilloscope with voltage and
current probes
• A Personal Computer (PC) with ISL88731(C) control
software and drivers (all available on the self extracting
archive file ISL88731_v3.0_Installer.exe available for
download from www.Intersil.com)
Note: The National Instruments Runtime Engine must be
installed before connecting the evaluation board to your PC.
The Runtime Engine contains driver files required by the New
Hardware Wizard.
It will create a folder (“ISL88731 Installer”) in the folder that
holds the self extracting file. A dialog box will give you the
opportunity to change the location of the extracted files.
Make a note of the location of the folder. You will need to direct
the New Hardware Wizard to this folder for the driver.
Software Installation
Both the ISL88731(C) control program and the National
Instruments run time engine are installed from the “ISL88731
Installer” folder. Navigate to the folder and double click the
“AutoRun.exe” file. You should see the window as shown in
Figure 2.
Click on the “Install or Uninstall ISL88731 Control Software” to
run the Installation Wizard.
You should now see a window as shown in Figure 3.
FIGURE 2. ISL88731(C)EVAL2Z CONTROL SOFTWARE INSTALLATION MENU
FIGURE 3. ISL88731(C)EVAL2Z CONTROL SOFTWARE INSTALLATION WIZARD
2
AN1404.2
June 27, 2011
Page 3
Application Note 1404
In most PCs the National Instruments Runtime Engine will be
installed automatically after the ISL88731(C) control software.
You should see the window as shown in Figure 4.
FIGURE 4. NATIONAL INSTRUMENTS RUNTIME ENGINE INSTALLATION WIZARD
Follow the instructions to install the Runtime Engine. If it does
not run automatically, click “Install National Instruments
Runtime Engine” on the ISL88731(C)EVAL2Z Control Software
Installation Menu (see Figure 2). After installation, click on the
Close Menu button.
Driver Installation
Connect the ISL88731(C)EVAL2Z to the USB port of your PC.
Note: The SMBus part of the ISL88731(C) and the USB-SMBus
interface are powered by the 5V from the USB port on the PC.
It is not necessary to have other power supplies connected to
the evaluation board to read or write to the ISL88731(C).
When WINDOWS detects new hardware, the new hardware
wizard should appear. Direct it to look for the driver
information file in the “ISL88731 Installer” folder (extracted
from the downloaded file). In some systems, it may be
necessary to specify the driver information file
(USB-I2C_WDM.inf).
Run the ISL88731 Application
ISL88731(C) control software should now appear in your
Start/All Programs list. Click on it to run the application.
3
AN1404.2
June 27, 2011
Page 4
Application Note 1404
Figure 5 shows a screen shot of the ISL88731(C) control software user interface.
FIGURE 5. ISL88731(C)EVAL2Z CONTROL SOFTWARE SCREEN SHOT
4
AN1404.2
June 27, 2011
Page 5
Application Note 1404
Enable the Output
When power is applied to the ISL88731(C), the Charge Current
and Charge Voltage Registers are 0x0000 and the output is
disabled. The output is enabled by writing valid, non-zero
values to these 2 registers. See the ISL88731(C) datasheet for
valid values.
The ISL88731(C) is also disabled if the ACIN pin is below 3.2V.
An Input voltage >13.1V will pull ACIN above 3.2V. The
minimum input voltage can be adjusted on the evaluation
board by changing R
or R7 (see Figure 10).
6
Description of the User Interface Controls
and Display Information
Clicking the “Write All” button will write the indicated SMBus
sequences to Registers 14, 15 and 3F in the ISL88731(C).
Clicking on the “Read All” button will read all 5 registers in the
ISL88731(C) and display the results in the grey indicators
below the “Read All” button.
If the device and the interface are operating normally, the
VendorID (reg FE) and the DeviceID (reg FF) will display as
follows:
If the VendorID (reg FE) and DeviceID (reg FF) remain 0000 it
indicates a communication problem.
If the Visa resource name is blank, it indicates a problem with
the USB communication.
Clicking on the “RESET INTERFACE” button may correct the
problem. If not, try closing the software, remove power and
USB connections from the evaluation board, then reconnect
and restart.
Note the register contents in the user interface are updated
ONLY when the “Read All” button is clicked.
The “Write Charge Current” button sends only the charge current
command to the ISL88731(C). Other registers are unchanged.
The small white window next to “Charge Current” accepts
inputs in Amps, and converts them the hexadecimal (hex)
value to be written to the ISL88731(C) in the grey window.
the enter key, the display will change to 4.096. The value can
be incremented and decremented with the up/down arrows at
the left of the window.
The Sense R inputs are used to calculate the hexadecimal
(hex) values and the SMBus sequences for the Charge Current
limit and Input Current limit entered in the white boxes. The
evaluation board is built with 10mΩ sense resistors. If you
change them to another value, enter the new value to get
correct calculations.
After entering a number in the white box, hitting the enter key
or clicking on another area will enter the value and the
resulting hex value and the SMBus sequence will be calculated
and displayed.
.
S = the Start condition, A = the Acknowledge, P = the Stop
condition.
Note: The LO byte is sent first. The HI byte is sent second.
When writing 0x41A0, 0xA0 is written first and 0x41 is sent
second.
Important Note: The SMBus commands are NOT sent until one
of the WRITE buttons is clicked (i.e. Write All or Write Charge
Current).
The “Write Charge Voltage” button writes only the Charge
Voltage register. Other registers are unchanged.
The small white window accepts Charge Voltage (in Volts) and
calculates the hexadecimal value and the SMBus sequence to
write to the ISL88731(C). The SMBus command is sent when
the “Write Charge Voltage” or “Write All” button is clicked. See
Figure 8.
The “Write Input Current” button writes only the Input Current
register. Other registers are unchanged. The small white
window accepts Input Current Limit in Amps and calculates
the hex value and the SMBus sequence to be written to the
ISL88731(C). The Sense R input Is used to calculate the hex
value and the SMBus sequence. The SMBus command is sent
when the “Write Input Current” or “Write All” button is clicked.
See Figure 9.
The ISL88731(C) will shut down unless it receives Charge Voltage
or Charge Current commands every 175s. This is a safety feature
that prevents overcharging batteries if bus master has stopped
functioning. The control software includes the function to write
the Charge Current register every 60 seconds, which can be
enabled or disabled as shown in Figure 6 and Figure 7.
The hex value depends on the value entered in Sense R.
The value in the white window will be forced to the nearest
valid number that can be produced by the DAC in the
ISL88731(C). That is, after entering 4 and hitting the enter key,
the display will change to 3.968. After entering 4.1 and hitting
5
FIGURE 6. ENABLE WRITING CHARGE CURRENT REGISTER EVERY
60 SECONDS
FIGURE 7. DISABLE WRITING CHARGE CURRENT
REGISTER EVERY 60 SECONDS
AN1404.2
June 27, 2011
Page 6
Application Note 1404
FIGURE 8. WRITE CHARGE VOLTAGE BLOCK
FIGURE 9. WRITE INPUT CURRENT BLOCK
Interface Troubleshooting
When the evaluation board is connected to the PC by a USB
cable and the software is running, the software assigns a VISA
resource name.
If this window is blank, it indicates a problem with
communication between the PC and the micro controller on
the evaluation board. Check and correct the connection of the
USB cable to your PC and the evaluation board. Click on the
“RESET INTERFACE” button.
.
If you still get no VISA resource name, Close and restart the
ISL88731(C) control software.
If you still get no VISA resource name, Open the Device
Manager (Control Panel/System/Device Manager) with the
ISL88731(C)EVAL2Z evaluation board connected to your USB
port. You should see a device named “NI-VISA USB Device”. If it
is not in the list of devices, the driver may not be correctly
installed. Run the New Hardware wizard.
If you still get no VISA resource name, contact your Intersil
Field Application Engineer for assistance.
Intersil Corporation reserves the right to make changes in circuit design, software and/or specifications at any time without notice. Accordingly, the reader is
cautioned to verify that the Application Note or Technical Brief is current before proceeding.
For information regarding Intersil Corporation and its products, see www.intersil.com
15
FIGURE 17. BOTTOM SILK SCREEN (BOTTOM VIEW)
AN1404.2
June 27, 2011
Page 16
Mouser Electronics
Authorized Distributor
Click to View Pricing, Inventory, Delivery & Lifecycle Information:
Intersil: ISL88731CEVAL2Z
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.